Certificate of Compliance Certificate: 2593926 Master Contract: 162767 Project: 70022728 Date Issued: April 2, 2015 Issued to: KROHNE, Inc. 7 Dearborn Rd Peabody, MA 01960 USA Attention: Rick Loiselle The products listed below are eligible to bear the CSA Mark shown with adjacent indicators 'C' and 'US' for Canada and US or with adjacent indicator 'US' for US only or without either indicator for Canada only. Behzad Nejad Issued by: Behzad Nejad, P.Eng. PRODUCTS CLASS 2258 83 CLASS 2258 03 - PROCESS CONTROL EQUIPMENT-Intrinsically Safe and NonIncendive - Systems-For Hazardous Locations-Certified to U.S. Standards - PROCESS CONTROL EQUIPMENT - Intrinsically Safe and Non Incendive Systems - For Hazardous Locations Class I, Groups C, D; Class II, Groups E, F, G; Class III: - Ultrasonic Compact Flowmeter, models OPTISONIC 3400 C-DIV1; rated input:100-230Vac, 22VA, 50/60Hz or 24Vac/dc, 22VA/12W or 12-24Vdc, 12W; Explosion-proof with non-intrinsically safe input/output. Temperature Code T6…T3; Tamb: - 40˚C to +60˚C; Enclosure Type 4X and 6; DUAL SEAL; MWP 25 MPa (250 Bars). Notes: 1.The OPTISONIC 3400 C-DIV1 consists of the OPTISONIC 3000-DIV1 series Primary Head and the UFC 400-DIV1 VN35abcdefghijklkmop Signal Converter. - Ultrasonic Compact Flowmeter, models OPTISONIC 3400 C/i-DIV1; rated input:100-230Vac, 22VA, 50/60Hz or 24Vac/dc, 22VA/12W or 12-24Vdc, 12W; Explosion-proof with intrinsically safe input/output; Temperature Code T6…T3 ; Tamb: - 40˚C to +60˚C; Enclosure Type 4X and 6; DUAL SEAL; MWP 25 MPa (250 Bars). DQD 507 Rev. 2012-05-22 Page: 1 Certificate: 2593926 Master Contract: 162767 Project: 70022728 Date Issued: April 2, 2015 Notes: 1.The OPTISONIC 3400 C/i-DIV1consists of the OPTISONIC 3000-DIV1 series Primary Head and the UFC 400i-DIV1 VN35abcdefghijklkmop Signal Converter. 2. Intrinsically safe when installed per manufacturers Control Drawings 8.30736.30 and drawing 8.30736.31 and 8.30736.32 3. Intrinsically safe input and output field wiring are to be separated shielded wire pairs. 4. Non-intrinsically safe I/O options are restricted from being used in the UFC 400 electronic unit when intrinsically safe I/O options are installed. - Ultrasonic Signal Converter, models UFC 400 F-DIV1 or UFC 400 F/i-DIV1 VN35abcdefghijklkmop; rated input: 100-230Vac, 22VA, 50/60Hz or 24Vac/dc, 22VA/12W or 12-24Vdc,12W; Explosion-proof with nonintrinsically safe or intrinsically safe input/output; Temperature Code T6; Tamb: - 40˚C to +60˚C; Enclosure Type 4X and 6. Notes: 1: The UFC 400 F/...-DIV1 signal converter must be used with the OPTISONIC 3000 F/...-DIV1 series flow sensor or with the UFS 3000 F/...-DIV1 series which is approved under FM project 3016332/CSA 1515313. 2. Intrinsically Safe when installed per manufacturers Control Drawings 8.30736.30, 8.30736.31 and 8.30736.32. 3. Intrinsically safe input and output field wiring are to be separated shielded wire pairs. 4. Non-intrinsically safe I/O options are restricted from being used in the UFC 400 electronic unit when intrinsically safe I/O options are installed. The codes are as follows where: a = Standard: 0 or 4 b = Type Display: 4 or H c = Power Supply: 1, 4 or A d = Approval: B e = Cable connection: 4 or R f = Instruction / program language: Not safety relevant g = Custody transfer: Not safety relevant DQD 507 Rev. 2012-05-22 Page: 2 Certificate: 2593926 Master Contract: 162767 Project: 70022728 Date Issued: April 2, 2015 h = Process diagnostics: Not safety relevant i = Converter Housing: 1, 2, 3 or 4 j = Private Label: 0 or 1 k = Communication: 1, 2, 3, 4, 6, 7, 8, B, C or E l = 1st IO Module: 0, 1, 2, 8, A, B, C, E, F, G, H, K, P or R m = 2nd IO Module: 0, 8, A, B, C, E, F, G, H, K, P or R n = Connection Box: 0 o = Tagplate: Not safety relevant p = Instruction language: 3 (or translated manual required) Class I, Groups A,B,C, D; Class II, Groups E, F, G; Class III: - Ultrasonic flow sensor, model OPTISONIC 3000 F/…-DIV1 VN61abcdefghijklkmopqrst; OPTISONIC 3000 F/…-DIV1 VN62abcdefghijklkmopqrst; OPTISONIC 3000 F/…-DIV1 VN63abcdefghijklkmopqrst ; Temperature Code: T6..T2; Tamb: - 40 to +70˚C; Enclosure Type 4X and 6; DUAL SEAL; MWP 25 MPa (250 Bars). Notes: 1. The OPTISONIC 3000 F/...-DIV1 series flow sensor must be used with UFC 400 F/...-DIV1signal converter. 2. Intrinsically Safe when installed per manufacturers Control Drawing 8.30736.30. 3. The Ultrasonic flow sensor models include the OPTISONIC 3000 F-DIV1 (default version); OPTISONIC 3000 F/LT-DIV1 (low temperature version); OPTISONIC 3000 F/HJ-DIV1(default version with heating jacket); OPTISONIC 3000 F/XXT-DIV1(extra extended temperature version); OPTISONIC 3000 F/XXT/HJ-DIV1 (extra extended temperature version with heating jacket); and OPTISONIC 3000 F/HV-DIV1 (high viscosity). The codes are as follows where: a = Standard: 0 or 4 b = Nominal diameter: 4, 5, 6, 7, 8, A, B, C, D, E, F, G, H, K, L, M, N, P, R, S, T, U, V, W, X or Y c = Nominal pressure: A, B, D, E, F, X or Y d = Approval: B DQD 507 Rev. 2012-05-22 Page: 3 Certificate: 2593926 Master Contract: 162767 Project: 70022728 Date Issued: April 2, 2015 e = System design/ cable connector: 1, 2, 4, A, R or U f = Converter model: 0, C or D g = Tube / Flange material: A, B, C, E, F, G, H, K, L, M or N h = Cable length: 0, 1, 2, 3, 4 or 5 i = Calibration: Not safety relevant j = Housing: 0 or 2 k = Private Label: 0 or 1 l = Finish: Not safety relevant m = Protection: 0 or 2 n = Construction requirements: Not safety relevant o = QA / QC requirements: Not safety relevant p = Custody transfer: Not safety relevant q = Tag plate: Not safety relevant r = Instruction manual language: 3 (or translated manual required) s = Process temperature: 0, 1 or 2 t = Extended application: 0, 1 or 2 CLASS 2258 83 -PROCESS CONTROL EQUIPMENT - Intrinsically Safe and Non-Incendive Systems - For Hazardous Locations – US Standards Class I, Groups A, B, C, D; Class II, Groups E, F, G; Class III: - Ultrasonic Compact Flowmeter, models OPTISONIC 3400 C-DIV1; rated input: 100-230Vac, 22VA, 50/60Hz or 24Vac/dc, 22VA/12W or 12-24Vdc, 12W; Explosion-proof with non-intrinsically safe input/output. Temperature Code T6…T3; Tamb = - 40˚C to +60˚C; Enclosure Type 4X and 6; DUAL SEAL; MWP 25 MPa (250 Bars). Notes: DQD 507 Rev. 2012-05-22 Page: 4 Certificate: 2593926 Master Contract: 162767 Project: 70022728 Date Issued: April 2, 2015 1. The OPTISONIC 3400 C-DIV1 consists of the OPTISONIC 3000-DIV1 series Primary Head and the UFC 400-DIV1 VN35abcdefghijklkmop Signal Converter. - Ultrasonic Compact Flowmeter, models OPTISONIC 3400 C/i-DIV1; rated input: 100-230Vac, 22VA, 50/60Hz or 24Vac/dc, 22VA/12W or 12-24Vdc, 12W; Explosion-proof with Intrinsically Safe input/output; Temperature Code T6…T3; Tamb: -40˚C to +60˚C; Enclosure Type 4X and 6; optional DUAL SEAL; MWP 25 MPa (250 Bars). Notes: 1. The OPTISONIC 3400 C/i-DIV1consists of the OPTISONIC 3000-DIV1 series Primary Head and the UFC 400i-DIV1 VN35abcdefghijklkmop Signal Converter. 2. Intrinsically Safe when installed per manufacturers Control Drawings 8.30736.30 and drawing 8.30736.31 and 8.30736.32 3. Intrinsically safe input and output field wiring are to be separated shielded wire pairs. 4. Non-intrinsically safe I/O options are restricted from being used in the UFC 400 electronic unit when intrinsically safe I/O options are installed. - Ultrasonic Signal Converter, models UFC 400 F-DIV1 or UFC 400 F/i-DIV1 VN35abcdefghijklkmop; rated input 100-230Vac, 22VA, 50/60Hz or 24Vac/dc, 22VA/12W or 12-24Vdc, 12W; Explosion-proof with with nonintrinsically safe or intrinsically safe input/output having Entity Parameters; Temperature Code T6; Tamb = -40˚C to +60˚C; Enclosure Type 4X and 6. Notes: 1: The UFC 400 signal converter must be used with the OPTISONIC 3000-DIV1 series flow sensor or with the UFS 3000 F/...-DIV1 series which is approved under FM project 3016332/CSA 1515313. 2. Intrinsically Safe when installed per manufacturers Control Drawings 8.30736.30 and drawing 8.30736.31 and drawing 8.307636.32. 3. Intrinsically safe input and output field wiring are to be separated shielded wire pairs. 4. Non-intrinsically safe I/O options are restricted from being used in the UFC 400 electronic unit when intrinsically safe I/O options are installed. The codes are as follows where: DQD 507 Rev. 2012-05-22 Page: 5 Certificate: 2593926 Master Contract: 162767 Project: 70022728 Date Issued: April 2, 2015 a = Standard: 0 or 4 b = Type Display: 4 or H c = Power Supply: 1, 4 or A d = Approval: 4 e = Cable connection: 4 or R f = Instruction / program language: Not safety relevant g = Custody transfer: Not safety relevant h = Process diagnostics: Not safety relevant i = Converter Housing: 1, 2, 3 or 4 j = Private Label: 0 or 1 k = Communication: 1, 2, 3, 4, 6, 7, 8, B, C or E l = 1st IO Module: 0, 1, 2, 8, A, B, C, E, F, G, H, K, P or R m = 2nd IO Module: 0, 8, A, B, C, E, F, G, H, K, P or R n = Connection Box: 0 o = Tagplate: Not safety relevant p = Instruction language: 3 (or translated manual required) - Ultrasonic flow sensor, model OPTISONIC 3000 F/…-DIV1 VN61abcdefghijklkmopqrst; OPTISONIC 3000 F/…-DIV1 VN62abcdefghijklkmopqrst; OPTISONIC 3000 F/…-DIV1 VN63abcdefghijklkmopqrst ; Temperature code: T6 …T2; Tamb: - 40 to +70˚C; Enclosure Type 4X and 6; optional DUAL SEAL; MWP 25 MPa (250 Bars). Notes: 1. The OPTISONIC 3000 F/…-DIV1 series flow sensor must be used with UFC 400 F/...-DIV1signal converter. 2. Intrinsically Safe when installed per manufacturers Control Drawing 8.30736.30. 3. The Ultrasonic flow sensor models include the OPTISONIC 3000 F-DIV1 (default version); OPTISONIC 3000 F/LT-DIV1 (low temperature version); OPTISONIC 3000 F/HJ-DIV1(default version with heating jacket); OPTISONIC 3000 F/XXT-DIV1(extra extended temperature version); OPTISONIC 3000 F/XXT/HJ-DIV1 (extra extended temperature version with heating jacket); and OPTISONIC 3000 F/HV-DIV1 (high viscosity). DQD 507 Rev. 2012-05-22 Page: 6 Certificate: 2593926 Master Contract: 162767 Project: 70022728 Date Issued: April 2, 2015 The codes are as follows where: a = Standard: 0 or 4 b = Nominal diameter: 4, 5, 6, 7, 8 , A, B, C, D, E, F, G, H, K, L, M, N, P, R, S, T, U, V, W, X or Y c = Nominal pressure: A, B, D, E, F, X or Y d = Approval: 4 e = System design/ cable connector: 1, 2, 4, A, R or U f = Converter model: 0, C or D g = Tube / Flange material: A, B, C, E, F, G, H, K, L, M or N h = Cable length: 0, 1, 2, 3, 4 or 5 i = Calibration: Not safety relevant j = Housing: 0 or 2 k = Private Label: 0 or 1 l = Finish: Not safety relevant m = Protection: 0 or 2 n = Construction requirements: Not safety relevant o = QA / QC requirements: Not safety relevant p = Custody transfer: Not safety relevant q = Tag plate: Not safety relevant r = Instruction manual language: 3 (or translated manual required) s = Process temperature: 0, 1 or 2 t = Extended application: 0, 1 or 2 APPLICABLE REQUIREMENTS C22.2 No. 25-1966 (R2009) DQD 507 Rev. 2012-05-22 - Page: 7 Certificate: 2593926 Master Contract: 162767 Project: 70022728 Date Issued: April 2, 2015 C22.2 No. 30-M1986 (R2012) - CAN/CSA-C22.2 No. 94-M91 (R2011) CAN/CSA-C22.2 No. 157-92 (R2012) - CAN/CSA-C22.2 No. 61010.1-04 - FM 3600 – 2011 - FM 3610 – 2010 - FM 3615 – 2006 - FM 3616 – 2011 Enclosures for Use in Class II, Groups E, F and G Hazardous Locations Explosion-Proof Enclosures for Use in Class I Hazardous Locations Special Purpose Enclosures - Intrinsically Safe and Non-Incendive Equipment for Use in Hazardous Locations Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use, Part 1: General Requirements Electrical Equipment for Use in Hazardous (Classified) Locations, General Requirements Intrinsically Safe Apparatus for Use in Class I, II & III, Division 1, and Class I, Zone 0 & 1 Hazardous (Classified) Locations Explosion-proof Electrical Equipment, General Requirements Dust-Ignitionproof Electrical, General Requirements Electrical Equipment for Measurement, Control, and Laboratory Use Requirements for Process Sealing Between Electrical Systems and Flammable or Combustible Process Fluids Enclosures for Electrical Equipment - FM 3810 - 2005 - ANSI/ISA 12.27.01-2003 - UL 50 – 11th Edition - MARKINGS General: - Submittor’s name, tradename and/or CSA File No. LR ; - Model designation; - Hazardous Location designation; - Serial no.; - CSA Monogram; DQD 507 Rev. 2012-05-22 Page: 8 Certificate: 2593926 Master Contract: 162767 Project: 70022728 Date Issued: April 2, 2015 - Temperature Code with Tamb.; - The words “DUAL SEAL” (except on converter models) for version with burst disk - The maximum working pressure (MWP) rating, in Pascal, PSI or Bar - Reference to Installation Drawing; Markings, “Warning: Substitution of components may impair intrinsic safety” and “Avertissement: La substitution de composants peut compromettre la Sécurité Intrinsèque”; Models: OPTISONIC 3400 compact and UFC 400 - Markings, “Open circuit before removing cover” and “Ouvrir le circuit avant d’énlever le couvercle”; Markings, “Cover joints must be cleaned before replacing cover” and “Nettoyer les surfaces de contact de couvercle avant de le remettre en place”; Models: OPTISONIC 3000 - Markings “Intrinsically Safe” DQD 507 Rev. 2012-05-22 Page: 9
© Copyright 2024