PRE- BID T.E. NO.: GMSCL/EQUIPMENT PURCHASE/RADIOLOGY

PRE- BID
T.E. NO.:
GMSCL/EQUIPMENT PURCHASE/RADIOLOGY-2/2015-16
Name of
Item :
ECG Machine 3 channel with Print and Digital view and Backup
[A]
1
2
3
i)
ii)
iii)
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
a)
b)
32
33
a)
b)
Date:28/05/2015
Time:11:00 hrs
Required Specifications
Technical Specification of 3-Channel ECG Machine with print & digital view
The ECG machine should operate on mains (220 V - 50 Hz) and rechargeable battery.
The recorder should run minimum of 4 hours on fully charged battery.
It should provide facility to record following reads.
Standard Lead (the limb reads or bipolar limb reads) : I, II & III.
Augmented Limb Leads : AVL, AVR & AVF.
Chest Leads (the unipolar or v-leads) : From v1, to v6.
Should be provided with terminal for a good earth connection to precrude Electrical
disturbances while recording. Electrodes of paediatric sizes for use in Newborn, infants, and
paediatric patients must be provided
It should record on standard universal thermal printer paper available in India.
It should record the paper at a speed of 25mm per second.
Post sale services and maintenance should be available.
Should have an LCD display of atleast 3”. with good resolution
should be able to be connected HIMS/LAN (optional)
Should be able to display 3+R channel
Should have simultaneous acquisition of 12 lead ECG.
Should not be more than 5 kg and portable
Should have Rhythm lead recording facility.
Should accept Z fold thermal paper with a minimum width of 60_100 mm
Recording speed selection of 5, 25 & 50 mm/sec should be provided
Should have Manual, automatic recording modes.
Should have different filters like Baseline filter, EMG filter & AC filter
Should have a sampling rate of 1000 samples/second/channel
Automatic adjustment of baseline for optimal recording
Should have QRS and pacer detection
Should have indication for Filter, Fault, print Error identification, artifacts, Low Battery
indication
Should have visual indication of Lead Disconnection.
Data storage for last patient.
Should have self-test facility
Should have sensitivity of 5, 10, 20 mm/mV.
DefibriIIation protection should be provided.
Mains and built-in rechargeable battery.
Should able to Print 35 Nos ECG on Battery
Manufacturer should be ISO 9000
Approved or CE or USFDA
Supplier should Supply : 10 ECG cable -01 Nos,ECG clamp Electrode 4Nos,
ECG chest Electrode-6 Nos, ECG paper Role-50 Nos,
ECG Jelly Bottle-25 Nos, operating Manual (English) free of charge at the time of delivery
INCLUDING Paediatric accessories
Supplier should give Training for operation & proper use of equipment to teaching -non
teaching operating staff at the time of Installation of the equipment
INSTALLATION: In case it is found that equipment/accessories, spares accompanying the
equipment has been damaged during transit or found to be wrongly supplied, then it is the
responsibility of the Tenderer to replace the equipment totally at his own risk and cost.
Warranty -Should be at least 24 (Twenty Four) Months Guarantee/Warranty
Service center: Tenderer should have company Service Engineer & Authorized Service
1
PRE- BID
T.E. NO.:
GMSCL/EQUIPMENT PURCHASE/RADIOLOGY-2/2015-16
Name of
Item :
ECG Machine 3 channel with Print and Digital view and Backup
[A]
Date:28/05/2015
Time:11:00 hrs
Required Specifications
Franchise Network in entire Gujarat. Service Call Should be attended within 48 Hours.
Warranty for 2 years & CMC price next 5 years, which should include cost of labour spares,
entire equipment. The CMC cost will be taken into consideration for price comparision.
2
PRE- BID
T.E. NO.:
GMSCL/EQUIPMENT PURCHASE/RADIOLOGY-3/2015-16
Name of Item :
Mobile X-Ray Machine
Date: 28/05/2015
Time:11:00 hrs
[A]
Required Specifications
1
2
Should be a floor stand unit with scissor arm, fiber wheels and locking system.
Should have a counter balanced arm mechanism./ tension spring mechanism
Should have a X-ray tube current of minimum 4 to 8 mA in steps of 1mA and 50 to 70 KV in steps
of 1KV
Should have a constant potential minimum 100Khz high frequency X-ray generator.
Machine should be portable and easy to move (floor mount)
Should have an exposure timer of minimum .04 to 4 sec in steps of 10 m sec
The x ray unit should have option for selecting exposure parameters for Adult & Child
X-ray tube head should have swing angulations of at least 290 degree in the vertical plane and
360 degree continuous rotations in the horizontal plane.
X-ray tube head should have angle indication.
Should be supplied with cones.
Should have latest and valid AERB Certificate of approval.
Warranty for 2 years & CMC price next 5 years, which should include cost of labour spares,
entire equipment. The CMC cost will be taken into consideration for price comparision.
3
4
5
6
7
8
9
10
11
1
PRE- BID
T.E. NO.:
GMSCL/EQUIPMENT PURCHASE/RADIOLOGY-1/2015-16
Name of
Item :
X-Ray Digital Machine 500 mA
[A]
Date: 28/05/2015
Time:11:00 hrs
Required Specifications
(40KW) 500 mA HIGH FREQUENCY X-RAY MACHINE
(A)
1
2
3
4
5
6
(B)
1
2
3
4
5
6
7
8
9
10
11
12
13
(C)
1
2
3
4
(D)
1
(E)
1
(F)
1
High frequency X-Ray machine suitable for general radiography.
X-RAY GENERATOR:
High Frequency X-Ray Generator having frequency of 50KHz or more should be
provided.
Power output of generator should be 40KW or more.
Radiographic KV Range should be 40 to 125KV.
mA Range (Rad.): 450mA or more.
Exposure time (Rad.): 1ms to 3Sec.
mAs Range (Rad.): 0.5 to 200mAs.
CONTROL:
A very compact, Soft Touch Control Panel having following functions & indications
should be provided. The panel can be supplied in Floor or Wall mount with Spill Proof
design.
Following features should be available on the control panel.
Machine ON/OFF Switch.
Digital Display of KV & mAs.
KV & mAs increase and decrease switches.
Tube focal spot selection Switch.
Ready and X-Ray on switch with Indicators
Bucky Selection Switch.
Self diagnostic Programme with Indicators for Earth fault error, KV error, filament error
& Tube’s Thermal Overload.
Anatomical Programming Radiography (i.e. APR) should be provided in which KV and
mAs are automatically selected depending upon the physique of the patient and part of
the body to be X-Rayed.
Anatomical Programming should be of minimum 200 programmes.
A dual action hand switch with retractable cord should be provided for Radiation
Protection of Operator. There should be provision for a infraRED cordless Exposure
switch also.
There should be provision of auto shut off of Control if no key is pressed for 10Min.
X-RAY TUBE:
One No. Dual focus Rotating Anode X-Ray tube thermally protected.
Anode heat storage capacity of tube should be more than 100KHU.
One Pair of 8 meter H.V. Cable.
One No. Collimator with auto shut off facility should be provided.
HV TANK:
A very compact H.V. Tank filled with high dielectric transformer oil should be provided.
The H.V. Tank should contain H.V. transformer, Filament Transformers, H.V. Rectifiers
& H.V. Cable receptacles.
TUBE STAND:
Floor to Ceiling Stand with Counter Balanced Tube Head (Rotatable ± 180 Degree), 360
Degree Rotatable; mounted on Floor Ceiling Rails for convenient movements should be
provided.
TABLE:
Horizontal table with 6-way movement of the tabletop should be provided.
1
PRE- BID
T.E. NO.:
GMSCL/EQUIPMENT PURCHASE/RADIOLOGY-1/2015-16
Name of
Item :
X-Ray Digital Machine 500 mA
[A]
Date: 28/05/2015
Time:11:00 hrs
Required Specifications
(40KW) 500 mA HIGH FREQUENCY X-RAY MACHINE
2
3
4
5
6
(G)
1
2
3
(H)
(I)
1
2
3
4
5
Longitudinal movement of tabletop should be more than 400mm & transverse
movement should be more than 160mm. It should have height adjustment facility.
The Table should consists of motorized reciprocating bucky with Imported Grid of size
17 ¼” x 18 7/8” & of Ratio 8:1 – 85 lines/inch.
The Bucky should cover the entire length of the table and should be locked at any
desired position by an Electromagnetic lock.
The tabletop should be made of low radiation absorption, waterproof material.
Table Accessories like stainless steel cassette tray, Compression band should be
provided.
VERTICAL BUCKY STAND:
Vertical Bucky Stand with oscillating Grid of Ratio 8:1, 85 lines/inch is to be provided.
The Bucky should move up & down & should be equipped with a stainless steel
cassette tray.
The stand should be floor-mounted type & can accommodate cassettes up to 14” X
17”. The Bucky is tilted in 6 steps of 15 degree Angle each for various Radiographs.
POWER REQUIREMENT: The unit should be operable on 3 Phase, 440Volts AC 50Hz
with line resist less than 0.4 Ohms. Line Regulation +10%.
OTHER REQUIREMENTS:
Unit should be approved by B.I.S. (Bureau of Indian Standards) for Mechanical &
Electrical Safety.
The unit should be approved by AERB
Offered Model should be European CE ROHS and US FDA for complete system. Please
attach certificate copy along with tender document.
The X ray generator, X ray tube, Collimator and Table Bucky Should be form same
manufacture.
The company should have a local Service center.
Warranty for 2 years & CMC price next 5 years, which should include cost of labour
spares, entire equipment. The CMC cost will be taken into consideration for price
comparision.
2