300 Held Drive Northampton Pa 18067 (USA) 610-262-6090 TSB#: 4150-1 Technical Service Bulletin Title: Alarm Indicator LED Color Backwards Product Model#: PM4150 Series Date: 05/05/2015 Product Name: Easy Pulse POC Bulletin Classification* (Check Only One) Required / Mandatory Recommended Problem / Issue: Affected Serial Numbers: 10020339 thru 10021634 Precision Medical has identified a problem with the PM4150 Series EasyPulse Portable Oxygen Concentrators, serial numbers 10020339 thru 10021634. The Amber and Red Alarm Indicator LEDs in the device’s Front Panel are switched, meaning the Red LED comes on when the Amber LED should and the Amber LED comes on when the Red LED should. Correct Indicators Green Ready for use Amber Caution - Prompt response by the OPERATOR is required Red QP737F3 Warning - Immediate response by the OPERATOR is required Page 1 of 2 Rev 0 300 Held Drive Northampton Pa 18067 (USA) 610-262-6090 Technical Service Bulletin Incorrect Indicators Green Ready for use Amber Warning - Immediate response by the OPERATOR is required Red Caution - Prompt response by the OPERATOR is required This error DOES NOT create a hazard or any unsafe condition to the end user Correction / Resolution: This problem can only be corrected with an update to the device internal firmware. For the firmware to be updated, the device must be returned to Precision Medical. This correction is completely voluntary and Precision Medical will update the firmware at no cost. * Recommended – Correction is voluntary, change does not affect safety or performance. QP737F3 Page 2 of 2 Rev 0