QP737F3 - Technical Service Bulletin

300 Held Drive  Northampton  Pa  18067  (USA)  610-262-6090
TSB#: 4150-1
Technical Service
Bulletin
Title: Alarm Indicator LED Color Backwards
Product Model#: PM4150 Series
Date: 05/05/2015
Product Name: Easy Pulse POC
Bulletin Classification* (Check Only One)
Required / Mandatory
Recommended
Problem / Issue:
Affected Serial Numbers: 10020339 thru 10021634
Precision Medical has identified a problem with the PM4150 Series EasyPulse Portable Oxygen Concentrators, serial
numbers 10020339 thru 10021634.
The Amber and Red Alarm Indicator LEDs in the device’s Front Panel are switched, meaning the Red LED comes on
when the Amber LED should and the Amber LED comes on when the Red LED should.
Correct Indicators
Green Ready for use
Amber Caution - Prompt response by the OPERATOR is required
Red
QP737F3
Warning - Immediate response by the OPERATOR is required
Page 1 of 2
Rev 0
300 Held Drive  Northampton  Pa  18067  (USA)  610-262-6090
Technical Service
Bulletin
Incorrect Indicators
Green Ready for use
Amber Warning - Immediate response by the OPERATOR is required
Red
Caution - Prompt response by the OPERATOR is required
This error DOES NOT create a hazard or any unsafe condition to the end user
Correction / Resolution:
This problem can only be corrected with an update to the device internal firmware.
For the firmware to be updated, the device must be returned to Precision Medical.
This correction is completely voluntary and Precision Medical will update the firmware at no cost.
*
Recommended – Correction is voluntary, change does not affect safety or performance.
QP737F3
Page 2 of 2
Rev 0