Small field dosimetry, a clinical perspective, Indra J Das, AAPM-2012 Small Field Dosimetry: A Clinical Perspective Indra J. Das, PhD, FAAPM, FACR, FASTRO Department of Radiation Oncology Indiana University School of Medicine Indianapolis, IN, USA Wrong detector used for IJD/Brazil-2012 BrainLab cone calibration IJD/Brazil-2012 TG-155 Approved Task 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. IJD/Brazil-2012 Treatment Fields What is a Small Field? Magna-Fields Lack of charged particle Traditional Fields 200x200 cm2 40x40 cm2 Small Field 4x4 cm2 0.3x0.3 cm2 Collaborate with the new task group (Non-compliant/IAEA) on absolute dosimetry to ensure that there is no overlap between the two task groups, but rather are complementary to each other. Review and summarize literature on dosimetry of small fields irrespective of the origin and treatment modality. Provide overview of the issue of CPE for the small field dosimetry in homogeneous and inhomogeneous media. Provide meaningful information on the spectrum and shift in beam energy from Monte Carlo. Provide radiation parameters (men/r, S/r, etc) for small field dosimetry from published literature from Monte Carlo. Provide suitability of specific detectors with respect to perturbations and signal to noise ratio. Provide available information on the correction and perturbation factors in detectors. Provide guidelines in measurement methods for modeling the treatment planning systems for small fields. Provide suitability of algorithms based on measurement for beam modeling in small fields especially in inhomogeneous medium. Provide error analysis and limit of uncertainty in the measurements. Provide guidelines and recommendations for accurate determination of dosimetric data for small fields. IJD/Brazil-2012 ■ 4x4 cm2 Advance Therapy Fields SRS/SRT Gamma Knife Cyber-Knife Tomotherapy IMRT IJD/Brazil-2012 ■ Dependent on the range of secondary electrons Photon energy Collimator setting that obstructs the source size Detector is comparable to the field size IJD/Brazil-2012 Small field dosimetry, a clinical perspective, Indra J Das, AAPM-2012 Source Size Definition of Small Fields 90%, 70%, 50%, 30%, 10% iso-intensity line IJD/Brazil-2012 Jaffray et al, Med Phys 20, 1417-1427 (1993) Das et al, Med. Phys. 35, 206-215, 2008 IJD/Brazil-2012 Small Field Dosimetry Problem Dosimetry Total scatter factor from different institutions 1.02 Absolute 1.00 0.98 Dose ■ Institutional variability in 6 MV Radionics SRS dosimetry Depth Dose ■ – [D(r,d)/D(r,dm)] TMR Profiles Output, Scp (total scatter factor) ■ ■ ■ – [D(r)/D(ref)] Cone Factor (St) 0.96 Relative 0.94 WEH U-Arizona Temple U U Penn Erlanger Serago et al. Fan et al. Zhu et al. 0.92 0.90 0.88 0.86 12% diff 0.84 0.82 0.80 10 IJD/Brazil-2012 Dosimetric Variation with Detectors 15 20 25 30 35 40 Cone Diameter (mm) Das et al, J Radiosurgery, 3, 177-186, 2000 IJD/Brazil-2012 Radiation Measurements Total scatter factor with various detectors Charged particle equilibrium or electronic equilibrium 1.00 0.98 ■ ■ Cone Factor (St) 0.96 0.94 0.92 Photon energy and spectrum Diamond CEA (Film) Kodak(Film) TLD Pinpoint(par) Pinpoint(per) 0.125ion(par) 0.90 0.88 0.86 ■ ■ ■ ■ 0.80 15 20 25 30 35 Field size Off axis points like beamlets in IMRT Changes men/r and S/r Detector size 0.82 10 Change in spectrum ● ● 0.125ion(per) MC(1mm) MC(5mm) 14% 0.84 Range of secondary electrons Medium (tissue, lung, bone) 40 Volume Signal to noise ratio Cone Diameter (mm) Das et al, J Radiosurgery, 3, 177-186, 2000 IJD/Brazil-2012 IJD/Brazil-2012 Small field dosimetry, a clinical perspective, Indra J Das, AAPM-2012 IAEA/AAPM proposed pathway CPE & Electron Range CPE, Charged Particle Equilibrium Electron range= dmax in forward direction Electron range in lateral direction ■ Nearly energy independent ■ Nearly equal to penumbra (8-10 mm) Field size needed for CPE Lateral range 16-20 mm ■ ■ dmax Alfonso, et al. Med Phys 35, 5179-5186 (2008) IJD/Brazil-2012 Relative Dosimetry Why So Much of Fuss? msr , f ref Dwf ,msrQmsr = M Qf msr N DW ,Qo kQ,Qo kQf msr ,Q msr ( ( )( )( Reference (ref) conditions cannot be achieved for most SRS devices (cyberknife, gammaknife, tomotherapy etc) Machine Specific reference (msr) needs to be linked to ref Ratio of reading (PDD, TMR, Output etc) is not the same as ratio of dose ) ) fclin Dwf clin M Qf clin M Qf clin ,Qclin / M Qclin clin clin clin , f msr = k Qfclin ,Qmsr f msr f msr f msr M Qmsr Dw , Qmsr / M Qmsr M Qf msr msr f msr Ω Qfclin = clin Qmsr (D (D f msr w ,Qclin f clin w , Qmsr , f msr kQf clin = clin ,Qmsr , f msr kQfclin = clin ,Qmsr )/ (M )/ (M f clin w ,Qclin f clin w ,Qmsr ) (Output ) ) = (Re ading ) IJD/Brazil-2012 D1 M 1 ≠ D2 M 2 rel rel [ D1 M 1 clin , f msr = • kQfclin ,Qmsr D2 M 2 ( S w,air ) fclin ⋅ Pfclin ( S w,air ) fmsr ⋅ Pmsr ] IJD/Brazil-2012 IJD/Brazil-2012 1.E+0 6 MV; Central Axis AIR, at exit 1.1 0.8 0.7 Scanditronix-SFD Scanditronix-PFD Exradin-A16 PTW-Pinpoint PTW-0.125cc PTW-0.3cc PTW-0.6cc PTW-Markus 1.1 Wellhofer-IC4 0.5 0.4 0.3 0.2 0.1 1.E-2 0.5 cm 1.95 1.90 1.E-4 15 MV; Central Axis 0.0 1.0 2.0 3.0 4.0 E (MeV) 5.0 6.0 7.0 1.0 1.E+1 0.0 1 2 3 4 5 6 7 8 0.9 9 Field Size (cm) WATER, at dmax 11 0.8 Relative dose at d max Das et al, TG-106, Med Phys, 35, 4186, 2008 10 (a) 1.E+0 5 cm, primary 0.7 Photon Fluence 0 6 MV 2.00 1.E-3 Scanditronix-SFD Scanditronix-PFD Exradin-A16 PTW-Pinpoint PTW-0.125cc PTW-0.3cc PTW-0.6cc PTW-Markus Wellhofer-IC4 0.6 0.5 0.4 0.3 0.2 0.1 1.E-1 2 3 4 5 6 7 8 9 Field Size (cm) IJD/Brazil-2012 1.80 1.75 1.70 1.65 1.60 0.5 cm, primary 1.55 1.E-2 1.50 0 1.E-3 1 2 3 4 5 6 Diameter of Cone (cm) 1.E-4 0.5 cm, scatter Verhaegen, Das, Palmans, PMB, 43, 2755-2768, 1998 1.E-5 1 1.85 5 cm, scatter 0.0 0 Average Energy (MeV) 0.6 Spectra & Effective Energy from SRS Cones (0.5-5 cm) 1.E-1 Photon Fluence 0.9 Relative dose at d max (c) 5 cm Field Size Limit for Accurate Dose Measurements with Available Detectors 1.0 10 11 0.0 1.0 2.0 3.0 E (MeV) 4.0 5.0 6.0 IJD/Brazil-2012 Small field dosimetry, a clinical perspective, Indra J Das, AAPM-2012 Radiological Parameters IJD/Brazil-2012 Cyber Knife Dosimetry Sanchez-Deblado et al, Phy. Med. Biol., 48, 2081, 2003 IJD/Brazil-2012 Francescon et al, Med. Phys. 25(4), 503, 1998 0.5% MC MC MC MC MC MC IJD/Brazil-2012 F. Araki, Med. Phys. 33, 2955-2963 (2006). IJD/Brazil-2012 Errors in Measured Reading Correction Factors Correction Factor depends on: Field size Source size (FWHM) Detector type Francescon, et al Med Phys 35, 504, 2008 IJD/Brazil-2012 Kawachi et al, Med Phys, 35, 4591-4598, 2008 IJD/Brazil-2012 Small field dosimetry, a clinical perspective, Indra J Das, AAPM-2012 f clin , f msr Published data on kQclin ,Qmsr Detector Density Effect 1.16 Siemens; PTW diode 60012 clin , f msr kQf clin , Qmsr 1.12 Correction factor 1.14 1.06 Elekta; PTW diode 60012 Siemens; Exradin A16 Elekta; Exradin A16 Siemens; Sun Nuclear Dedge 1.10 "Elekta; Sun Nuclear Dedge" Siemens; PTW Pinpoint 31014 1.08 1.04 Elekta; PTW Pinpoint 31014 Siemens; PTW microLion 1.02 Elekta; PTW microLion 1.00 0.98 0.96 0.94 0.92 0.90 0 5 10 15 Francescon et al Med Phys,38, 6513, 2011 , f msr kQf clin clin ,Qmsr 20 25 30 Field size (mm) 35 IJD/Brazil-2012 Scott et al, Phys Med Biol 57 4461–4476, 2012 IJD/Brazil-2012 , f msr kQf clin of Linear Accelerators clin ,Qmsr Correction Factor vs Ion Chambers Med. Phys. 38 (12), 6513-6527, 2011 Chung et al , Med Phys, 37, 2404-2413, 2010 , f msr kQf clin clin ,Qmsr IJD/Brazil-2012 Cyber Knife Radiotherapy and Oncology 100 (2011) 429–435 IJD/Brazil-2012 Tomotherapy kmsr Reference Dosimetry Reference: 5x10 cm2, 85 cm SSD, 10 cm Pantelis et al, Med Phy. 37, 2369-2379, 2010 IJD/Brazil-2012 Sterpin et al, Med Phys, 39, 4066, 2012 IJD/Brazil-2012 Small field dosimetry, a clinical perspective, Indra J Das, AAPM-2012 Depth Dose & Source Size kQ is not Constant in Small Field Kawachi et al, Med Phys, 35, 4591-4598, 2008 IJD/Brazil-2012 Sham et al, Med Phys, 35, 3317-3330, 2008 Dose and Penumbra with Spot Size Scott et al, Med Phys, 36, 3132, 2009 IJD/Brazil-2012 Profile & Source Size Sham et al, Med Phys, 35, 3317-3330, 2008 0.5 cm, 0.25 cm3 lung 0.9 Perturbations of the detector Relative Dose 0.8 Simple scaling based on density M. K. Woo, and J. R. Cunningham, "The valididty of density scaling method in primary electron transport for photon and electron beams," Med. Phys. 17, 187-194 (1990). ■ Ho m o ge ne o us M o nte C a rlo EP L TM R R a tio P o we r La w TMR C o nvo lutio n C C C Wa te r 1.0 Range of secondary electrons ■ IJD/Brazil-2012 1.1 Effect of Inhomogeneity ■ IJD/Brazil-2012 0.7 0.6 0.5 T. Mauceri, and K. R. Kase, "Effects of ionization chamber construction on dose measurements in heterogeneity," Medical Physics 14, 653-656 (1987). R. K. Rice, J. L. Hansen, L. M. Chin, B. J. Mijnheer, and B. E. Bjarngard, "The influence of ionization chamber and phantom design on the measurement of lung dose in photon beam," Medical Physics 15, 884-890 (1988). IJD/Brazil-2012 0.4 0.3 0.2 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 Depth (cm) Jones & Das, Med. Phys. 30, 296, 2003 Jones & Das, Med. Phys. 32, 766, 2005 IJD/Brazil-2012 Small field dosimetry, a clinical perspective, Indra J Das, AAPM-2012 TG-155 Recommendation Summary Dosimetric measurements should be carried out with more than one detector system. Small volume detector should be used that has minimum energy, dose and dose rate dependence as discussed in TG-120 and Report No103 should be used. Stereotactic diodes or electron diodes are recommended for field sizes < 1x1cm2 Micro chambers are best suited for dosimetric measurements for field sizes > 1x1 cm2 however, signal to noise as well as polarity effect should be evaluated. The quality of electrometer and triaxial cable as well as any connector and cables need to be of high quality. Stereotactic diode with micron size sensitive volume should be the detector of choice for measurements in beams in radiosurgery. The energy spectrum does vary in small fields such as SRS, and IMRT but these changes result in insignificant variations in stopping power ratios when compared to those of the reference field used in dosimetry codes of practice. The treatment planning system performance should be carefully validated when used for the treatment planning incorporating small fields. Although pencil beam and convolution/superposition dose engines are expected to perform well in small field treatment geometries and in almost homogeneous media, dose engines based on the Monte Carlo method are the most accurate method for modelling dose from small fields in heterogeneous media. The calculation grid size should be significantly smaller (~1/10) compared to the field size. Small field dosimetry should have an independent audit by a different physicist either internal or external like Radiological Physics Center verification. IJD/Brazil-2012 Small volume detector should be used that has minimum energy, dose and dose rate dependence. Micro-ion chambers are best suited for small field dosimetry; however, signal to noise should be evaluated. Stereotactic diode are ideally suited for radiosurgery beams. If field size is small compared to detector measurements should be performed at a greater source to surface distance with proper correction. IJD/Brazil-2012 -Summary Energy spectrum does vary in small fields such as SRS, and IMRT, however, its impact is not significant. Stopping power ratio in small fields for most ion chambers is relatively same as the reference field. Spot check and verification of smaller fields should be carried out with at least another independent method (TLD, film, MC, etc). Stay tuned to newer data and IAEA and AAPM TG guidelines. IJD/Brazil-2012 Thanks IJD/Brazil-2012
© Copyright 2024