Removal of Multiple High-Abundant Proteins from Monkey Plasma for Proteomics Sample Preparation Application

Removal of Multiple High-Abundant
Proteins from Monkey Plasma for
Proteomics Sample Preparation
Application
Proteomics
Author
Nelson Guerreiro and Stephane Charmont
Novartis Pharma AG
BioMarker Development
CH 4002 Basel, Switzerland
Abstract
The plasma proteome constitutes a highly complex array
of circulating proteins, and is a rich source of potential
diagnostic and prognostic biomarkers. The application of
proteomic technologies for biomarker discovery has been
limited by the presence of a few highly-abundant proteins,
which constitute over 80% of the total plasma proteins.
The ability to effectively remove high-abundant proteins in
plasma is essential in order to increase the detection sensitivity of proteins present in low abundance. In this study
we show the applicability of the Agilent Multiple Affinity
Removal System for the immunodepletion of high-abundant
target proteins from monkey plasma prior to separation by
two-dimensional gel electrophoresis (2DGE). The removal
of six highly-abundant plasma proteins enhanced the
detection and resolution of less-abundant proteins by
2DGE. In conclusion, efficient and selective immunodepletion of plasma proteins prior to proteomic separation can
increase the likelihood of detecting novel biomarkers
Introduction
Blood plasma is a rich source of circulating molecules derived from many tissues and processes that
can act as indicators of a physiological condition.
Currently in the drug development and risk
assessment process there is a growing interest in
the discovery and use of biomarkers in plasma.
The noninvasive collection of plasma is particularly amenable for monitoring treatment-related
effects in a clinical setting. Many of the rapidly
evolving technologies of proteomics are being used
to profile the dynamic changes in the plasma proteome associated with treatment or disease.
Plasma, however, is exceptionally difficult to analyze due to a dynamic range of more than 10 orders
of magnitude in concentration that separate the
highest- and lowest-abundant proteins. This is further compounded by the presence of a few highabundant proteins such as albumin and
immunoglobulins, which together may account for
as much as 80% of the total plasma protein. A
larger number of proteins, including potential biomarkers, are present at far lower concentrations.
Removal of abundant serum proteins will help
increase the relative amounts of less-abundant
proteins to the detectable dynamic range of the
various proteomic platforms.
The Multiple Affinity Removal System developed
by Agilent has been shown to simultaneously
remove multiple high-abundant proteins from
human plasma/serum in a single step with high
specificity. The Multiple Affinity Removal System
is comprised of an affinity column packed with
immobilized affinity-purified polyclonal antibodies
designed for the removal of human albumin, transferrin, IgG, IgA, haptoglobin, and antitrypsin. In
drug development and risk assessment, various
1000
800
600
400
200
Flow-through
Low-abundant
proteins
0.0
mAU
5.0
2
10.0
Flow-through
Low-abundant
proteins
50
30
20
10
15.0
20.0
Bound
High-abundant
proteins
B
60
Injection - 200 µL
Flow-through fractions from four injections were
pooled into a spin concentrator with 5 kDa molecular weight cutoff. The samples were spun at
7500 × g at 10 °C. Buffer exchange was performed
by three rounds of spinning of 4 mL of buffer containing 7 M urea, 2 M thiourea, 4% CHAPS, 1% DTT
and PMSF. Protein content was determined using
Bradford assay (BioRad). Concentrated protein
fractions with 250 µg of proteins were diluted to
350 µL in IEF buffer (7 M urea, 2 M thiourea,
4% CHAPS, 1% DTT, 2% IPG 3–10 pharmalytes) and
applied to an IPG strip for overnight in-gel sample
rehydration. IPG gel strips were focused for
67500 Vh at 20 °C using the Multiphor II unit
(Amersham, Uppsala, Sweden). The second dimension (SDS-PAGE) was carried out on vertical
Monkey
0
40
Preparation of Protein Fractions for 2DGE
Bound
High-abundant
proteins
A
Re-equilibration 0% B - Buffer A
mAU
1200
25.0
30.0
Human
Re-equilibration 0% B - Buffer A
High-abundant protein removal from crude
monkey and human plasma was performed according to a standard liquid chromatography (LC) protocol provided with the Agilent Multiple Affinity
Removal System. Briefly, crude plasma samples
were diluted five times with Buffer A containing
protease inhibitor PMSF and filtered through
0.22-µm spin filters by spinning at 16000 × g. A
200-µL sample of diluted plasma was used for
injection in Buffer A (0% B) at a flow rate of
0.5 mL/min for 10 min. After the flow-through
fraction was collected, the bound fraction was
eluted with Buffer B (100% B) at a flow rate of
1 mL/min for 7 min. Collected flow-through
fractions were stored at –20 °C until use.
A representative chromatogram of protein depletion from monkey plasma on a 4.6 mm × 100-mm
Agilent Multiple Affinity Removal column is shown
in Figure 1A. The resulting chromatogram is
almost identical to that obtained for human plasma
(Figure 1B) under the same running conditions.
Elution 100% B
Immunoaffinity Chromatography
Results and Discussion
Elution 100% B
Experimental
gradient slab gels (9%–18% T) using the ISO-DALT
system. The gels were stained with SyproRuby®
(Molecular Probes, Eugene OR) following standard
procedures.
Injection - 200 µL
animal models are used in preclinical safety to
monitor the safety of compounds before going into
man. The identification of circulating biomarkers
from these animal models, which are indicative of
an adverse effect, can be of high value to treatment
monitoring of patients in clinical trials. In this
application, we have investigated the applicability
of the Multiple Affinity Removal System for the
removal of high-abundant proteins in plasma samples collected from cynomolgus monkey
(macaca fascicularis).
0
0.0
Figure 1.
5.0
10.0
15.0
20.0
25.0
30.0
Chromatogram of the affinity removal of highabundant proteins from A) monkey plasma and
B) Human plasma. A 200-µL sample of 5 × diluted
plasma was used for injection in Buffer A (0% B) at
a flow rate of 0.5 mL/min for 10 min on a
4.6 mm × 100-mm Multiple Affinity Removal Column.
Depleted protein fractions (flow-through) were collected from 2–4 min. The bound fractions were
eluted with Buffer B (100% B) at a flow rate of
1 mL/min for 7 min.
The flow-through fraction and bound fraction of
monkey plasma were collected and concentrated.
A volume of sample containing 250 µg of protein
was loaded for 2DGE. Figure 2 shows the 2D protein profile of the flow-through fraction and the
bound fraction compared to crude serum. Circled
areas indicate where the targeted high-abundant
proteins are removed. Proteins that bind to the
column are clearly absent from the flow-through
fraction indicating a high degree of efficiency in
the performance of the affinity column. However,
spots possibly representing haptoglobin could still
be seen in the flow-through fraction of monkey
plasma albeit at much lower intensities when compared to the bound fraction. The identity of these
spots has not been confirmed by mass spectrometry (MS). Furthermore, numerous additional spots
were detected in the bound fraction and are likely
to represent albumin fragments.
the staining intensity of several protein spots. In
addition, many less-abundant proteins that were
obscured by comigration with the high-abundant
target proteins are now visible. Figure 3 shows a
2D region in a gel displaying improved resolution
and detection of proteins after immunodepletion
of monkey and human plasma samples.
Results from this study demonstrate that the
Agilent Multiple Affinity Removal System for
human proteins can be applied for effective
removal of high-abundant proteins from monkey
plasma. The removal of these highly-abundant
plasma proteins enhances the detection of less
abundant proteins by 2DGE, and aids the display
of proteins that comigrate and are otherwise
obscured by the widely spread pattern of highlyabundant proteins. In addition, computer-aided
image analysis of the 2D profiles is improved, particularly with gel-to-gel matching and quantitation.
Concomitant to the removal of the six abundant
proteins there was a significant enhancement in
A) Flow-through fraction
B) Crude plasma
Antitrypsin
C) Bound fraction
Albumin
Transferrin
IgG heavy
Haptoglobin
Ig light
Figure 2.
2DGE profile of monkey plasma before and after immunodepletion with the Multiple Affinity Removal System. Gel
images represent A) flow-through fraction B) crude plasma and C) bound fraction. Regions corresponding to albumin,
transferrin, IgG, IgA, haptoglobin, and antitrypsin are circled. Enlarged areas show increased resolution. In the first
dimension, proteins were focused on IPG strips (pH 4–7 linear, 18 cm) for 67500 Vh. In the second dimension, proteins
were separated on vertical gradient slab gels (9%–18%). The gels were stained with SyproRuby .
3
www.agilent.com/chem
A
Human
Flow-through fraction
Crude plasma
Bound fraction
B
Monkey
Figure 3.
Improved resolution and detection of proteins after immunodepletion. Corresponding regions of the 2D gel from
A) human and B) monkey plasma samples after immunodepletion.
Additional data to the above mentioned will be
released in a future journal publication
Nelson Guerreiro and Stephane Charmont are
Principal Scientists at Novartis Pharma AG,
BioMarker Development in Basel
Switzerland.
For More Information
For more information on our products and services,
visit our Web site at www.agilent.com/chem/affinity
Agilent shall not be liable for errors contained herein or for incidental or consequential
damages in connection with the furnishing, performance, or use of this material.