System Components for Sample Gas Sampling Data Sheet 10/235.10 EN March 2004 A gas analysis system usually consists of two groups of devices: • The gas analyzer measures the concentrations of the sample gas components and converts them into an electrical signal. • Supplementary devices sample the sample gas from the process, condition it and feed it in the gas analysis system. Contents In most cases, the sample gas sampled from the process cannot be processed by the gas analyzer without further conditioning. Too high a dust content, temperature and dew point as well as too high or too low pressure, and not least interference components in the sample gas can impair the operational capability of the gas analyzer and falsify the measuring result. Supplementary devices such as the sampling probe, the sample gas line, the sample gas cooler and pumps and filters ensure that the sample gas inlet conditions of the connected gas analyzers are satisfied and faultless measuring results are obtained independently of the process sequence and the local conditions. Modular Sample Gas Probe System FE2 Filter Unit Annular Heating Element Heating Sleeve PFE2 Filter Unit PFE3 Filter Unit Selection Table for Probe Tube Types and Probes Type K Probe Tube Type 40 Probe Tube Type 42 Probe Tube (heated) Type 60 Probe Tube (Cement Flue Gas Probe) Probe 2 Probe 1, Probe 11 Ceramic Filter Process Pressure Reducers Heated Sample Gas Lines CGWB 13 Heated Sample Gas Line TBL 01, TBL 01-E Heated Sample Gas Lines Electronic Temperature Controller and Switching Relay Page 2 3 4 5 6 13 15 16 17 18 19 20 21 22 23 25 26 28 30 Modular Sample Gas Probe System Easily adaptable to different sampling conditions thanks to modular construction Components: • FE2, PFE2, PFE3 filter units • Probe tube types K, 40, 42, 60 • Annular heating element • Heating sleeve Probe tubes and filter unit heatable Connections for test gas and purging Filter with high retention rate and long service life Easy filter replacement Usable in emission measuring systems conforming to 13th BImSchV and 17th BImSchV (Federal Immission Protection Law) 2 System Components for Sample Gas Sampling 10/23-5.10 EN March 2004 FE2 Filter Unit Use The FE2 filter unit is a component of the modular gas sampling probe system. It is used for filtering gases containing dust. Process Gas Sampling Conditions Pressure pabs = 50 to 600 kPa Temperature max. 200 °C (coated max. 180 °C) Flow Rate 30 to 500 l/h, relative to 100 kPa and 0 °C Pressure Drop approx. 0.6 hPa at 100 l/h Dust Content max. 3 g/m3 (application recommendation); max. 40 g/m3 with probe tube type 40 with blow-back device; max. 2,000 g/m3 with probe tube type 60 Filter Characteristics Connection Ports Filter Element Surface filter Sample Gas / Test Gas G 1/4 internal thread to DIN ISO 228/1 Filter Fineness 0.5 µm Compressed Air (2 Ports) G 1/2 internal thread to DIN ISO 228/1 Retention Rate 99.98 % Probe Tube G 3/4 internal thread to DIN ISO 228/1 Seal Integrity 10–4 hPa l/s Compressed Air Supply for Blowback Dead Volume 280 ml (See PFE2 Filter Unit, page 6) Materials of the Gas-carrying Parts Casing and Connections Special steel Mat. No. 1.4571 Flange Special steel Mat. No. 1.0718 (optionally special steel Mat. No. 1.4571) Filter Aluminum oxide surface filter Coating Thermoset plastic, resistant up to 180 °C Installation Flange DN 65, PN 6, Form B to DIN 2573, flange gasket A; positioning of holes: 0°, 45°, 90°, 135°, 180°, 225°, 270°, 315°, 360° Mounting Flange connect or bolt to the probe tube Instrument Air Free of dust, water and oil Pressure max. 600 kPa Ordering Information Housing SS, flange 1.0718 Housing SS, with coating Housing SS, 1 Pt 100 Housing SS, 2 Pt 100 Housing SS, with coating, 1 Pt 100 Housing SS, with coating, 2 Pt 100 Housing and flange SS Housing and flange SS, coated Housing and flange SS, 1 Pt 100 Housing and flange SS, 2 Pt 100 Housing and flange SS, coated, 1 Pt 100 Housing and flange SS, coated, 2 Pt 100 Catalog No. 23042-5-8018413 23042-5-8018416 23042-5-8018402 23042-5-8018404 23042-5-8018406 23042-5-8018408 23042-5-8018414 23042-5-8018417 23042-5-8018403 23042-5-8018405 23042-5-8018407 23042-5-8018409 Weight approx. 8.0 kg 10/23-5.10 EN March 2004 System Components for Sample Gas Sampling 3 Annular Heating Element Use For heating the FE2 filter unit, which is a component of the modular gas sampling probe system. Heating is necessary if condensation is likely from moist gases and there is consequently risk of the filter becoming clogged and corroded. Data Power Supply 115 V AC or 230 V AC, 50 to 60 Hz Heater Output 200 W, 100 W or 50 W – depending on setting The heating resistors are factory-set for heat output of 100 W Temperature of Heater (unregulated) approx. 120 °C to 180 ° C (measured at ambient temperature of 20 °C, still air) Ambient Temperature 0 to 55 °C (still air) Degree of Protection of Housing IP 50 Materials Housing: aluminum alloy, heating surface: brass Dimensional Drawing (dimensions in mm) Weight approx. 1.7 kg Electrical Connections Terminals via cable gland PG 13.5 The electrical supply conductor must be heat resistant (≥ 200 °C). Installation Two half shells are placed round the FE2 filter unit to be heated and closed with a wing screw. Option Regulated Heating with 1 x Pt 100 Heating Circuit Ordering Information Ring heater 230 V, 50/60 Hz Ring heater 115 V, 50/60 Hz Ring heater with Pt 100, 230 V, 50/60 Hz Ring heater with Pt 100, 115 V, 50/60 Hz Catalog No. 23029-5-7865312 23029-5-8008813 23029-5-7865316 23029-5-7865317 230 / 115 V AC 200 W 4 System Components for Sample Gas Sampling 230 / 115 V AC 100 W 230 / 115 V AC 50 W 10/23-5.10 EN March 2004 Heating Sleeve Regulated Heating Sleeve for FE2 Filter Unit Heater Micanite Holding Temperature max. 350 °C Temperature Control max. 200 °C (Pt 100) Power Supply 230 V AC, optionally 115 V AC Power Consumption 250 VA Connection Terminal box with PG 9 bolted connection, protection level IP 54 Supply Cable 3 x 1 mm2 with silicone sheath, length 1 m Ambient Temperature –20 to +45 °C Relative Humidity ≤ 75 % yearly mean; infrequent slight condensation permissible Weight approx. 1 kg Essential Accessories Insulating Mat Temperature Controller (see page 30) Ordering Information Heated sleeve, 230 V, 50/60 Hz, 300 VA Heated sleeve, 115 V, 50/60 Hz, 300 VA 10/23-5.10 EN March 2004 Catalog No. 23043-5-8018410 23043-5-8018411 System Components for Sample Gas Sampling 5 PFE2 Filter Unit Use The PFE2 filter unit consists of the FE2 filter unit with a regulated heating sleeve, installed in a special-steel protective case. It is used to filter dust-containing gases and is a component of the modular gas sampling probe system. When used for gases with high dust content, the PFE2 can optionally be fitted with valves for automatic cleaning of the filter element in the FE2 filter unit. Protective Case Dimensions 450 x 450 x 400 mm (H x W x D) Material Special steel, Mat. No. 1.4301 Degree of Protection of Housing IP 54 Ambient Temperature –20 to +45 °C Weight approx. 18 kg Control Valve Options Solenoid Valve 3/2-way solenoid valve – Heating sleeve, regulated (Pt 100), with insulating mat – Annular heating element, unregulated – Filter fineness 0.3 µm Nominal Diameter DN 2.0 Pressure Range 600 kPa (6 bar) Pneumatic Valve Connections G 1/8 internal thread DIN ISO 228/1 Diaphragm Valve 2/2-way straight-way valve Material Brass Operating Pressure / Control Pressure 600 kPa (6 bar) Electrical Connections Standard connector Nominal Diameter DN 12 Connections G 1/2 internal thread DIN ISO 228/1 Control air: G 1/8 internal thread DIN ISO 228/1 Material Brass Mode of Operation Closed in the normal position Power Supply 115 V AC or 230 V AC, 50 to 60 Hz Power Consumption 8 VA Mode of Operation Closed in normal position Ambient Temperature 0 to +100 °C Ambient Temperature 0 to +100 °C 6 System Components for Sample Gas Sampling 10/23-5.10 EN March 2004 PFE2 Filter Unit Ordering Information Catalog No. 23028-0↓ Design without protection box with protection box IP 54 (450 x 450 x 400 mm) with protection box, IP 54, anti-freeze heating with protection box, IP 54, ambient temp. max. 65 °C with protection box IP 54 (330 x 203 x 270 mm) ↓ ↓ ↓ ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 1 2 3 4 5 Filter housing (equal FE 2) Stainless steel Stainless steel, with coating, max. 180 °C Housing and flange SS Housing and flange SS, coated Stainless steel with 0,3 µm filter element Stainless steel, with coating, max. 180 °C, with filter element <0,3 µm (filt. fineness) Housing and flange SS, with filter element <0,3 µm (filtration fineness) Housing and flange SS, coated, with filter element <0,3 µm (filtration fineness) 1 2 3 4 5 6 7 8 Heating Without with heating unit and insulation (max. 200 °C) with ring heater, unregulated 1 2 3 Control / Monitoring Without with 1x Pt 100 with 2x Pt 100 1 2 3 Back purging unit (without compressed air unit) without reverse purging with pneumatic valve with 3x pneumatic-, 1x air-bleed-valve(s) with check valve for AC-FNT 0 1 2 3 Power supply 230 V, 50 ... 60 Hz 115 V, 50 ... 60 Hz 1 2 ↓ Catalog No. 23028-0- 10/23-5.10 EN March 2004 ↓ 0 ↓ System Components for Sample Gas Sampling ↓ ↓ ↓ ↓ ↓ 7 8 2x Pg21 3x Pg16 1x Pg13,5 1x Pg11 Take holes for cable glands on the left into account Terminal box with cable glands Use as required Cover swivels upwards Electrical connections System Components for Sample Gas Sampling Flange plate with pipe clamp 2 1/2" RST 37-2 hot galvanized DIN 3568 for mounting at probe mounting adapter Elongated holes for M8 safety collar bolts Compressed air Purging Bulkhead fitting 10/12x1 mm Test gas connection Bulkhead fitting 4/6x1 mm Control air Bulkhead fitting 4/6x1 mm Sample gas connection PFE2 Filter Unit Probe Protective Case Dimensional Diagram (dimensions in mm) 10/23-5.10 EN March 2004 PFE2 Filter Unit PFE2 Filter Unit: Pneumatics Connections for 2-Stage Back-Purging 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 10/23-5.10 EN March 2004 Terminal box Probe protective case B Bulkhead fitting compressed air 4 to 6 bar E Plug (test gas connection) Pilot valve, filter cleaning -Y2.1 Diaphragm valve, filter cleaning -Y2.2 C Sample gas connection Pilot valve, pulsed compressed air -Y1.1 Diaphragm valve, pulsed compressed air -Y1.2 Bulkhead fitting 6 mm control air 2…6 bar Bulkhead fitting 12 mm compressed air 4…6 bar Heated sample gas line Power supply for heated sample gas line FE2 filter unit Solenoid valve, venting -Y4 Diaphragm valve, tube and filter surface cleaning -Y3.2 Pilot valve, tube and filter surface cleaning -Y3.1 System Components for Sample Gas Sampling 9 PFE2 Filter Unit Probe Protective Case, PFE2 Filter Unit Terminal Diagram, Terminal Block -X1 (Standard) Probe protective case Terminal box, strip -X1 1 1st Pt 100 FE2 filter unit 2 Pt 100 three wire connection IEC 751 Cl. B 3 4 1st Pt 100 Type 42 probe tube 5 Pt 100 three wire connection IEC 751 Cl. B 6 7 8 2nd Pt 100 FE2 filter unit (fail safe circuit) 9 Pt 100 three wire connection IEC 751 Cl. B 10 11 2nd Pt 100 Type 42 probe tube (fail safe circuit) 12 Pt 100 three wire connection IEC 751 Cl. B 13 14 Heater FE2 filter unit Heater Type 42 probe tube L 15 N 16 PE 17 L 19 L 19 N 21 N 21 PE 18 PE 18 20 Heater Type 42 probe tube 10 230 V 50-60 Hz or 115 V 50-60 Hz 230 V 50-60 Hz 20 N 22 N 22 PE 23 PE 23 System Components for Sample Gas Sampling 115 V 50-60 Hz 10/23-5.10 EN March 2004 PFE2 Filter Unit Probe Protective Case, PFE2 Filter Unit Terminal Diagram, Terminal Block -X2 (Standard) Probe protective case Terminal box, strip -X2 1 1st Pt 100 sample gas line 2 3 Pt 100 three wire connection IEC 751 Cl. B 4 2nd Pt 100 sample gas line (fail-safe circuit) 5 Pt 100 Three wire connection IEC 751 Cl. B 6 7 Heater sample gas line wattage acc. to length max. 3600 W L 8 N 9 PE 10 Anti-frost heater for system shut down approx. 5 °C at ambient temperature -20 °C 30 W Internal temperature approx. 100 °C Use heat-resistant cables! L 11 N 12 PE 13 10/23-5.10 EN March 2004 System Components for Sample Gas Sampling 230 V 50-60 Hz or 115 V 50-60 Hz 230 V 50-60 Hz or 115 V 50-60 Hz The heater operates unregulated. Switch on from outside only when system shut down 11 PFE2 Filter Unit Probe Protective Case, PFE2 Filter Unit Terminal Diagram, Terminal Block -X3 (for Cleaning) Terminal box, strip -X3 Probe protective case L PE N R Pilot valve -Y2.1 filter cleaning A 1 2 P Extended cleaning Pilot valve -Y3.1 cleaning of filter surface and tube R A 230 V 50-60 Hz or 115 V 50-60 Hz 10 L PE N 11 L PE N 12 L PE N 13 PE 9 3 L PE N 10 PE 9 1 2 4 P R Pilot valve -Y1.1 Pulsed compr. air A 5 6 P P Pilot valve -Y4 Venting R 7 8 B Probe / flange Internal temp. approx. 100 °C Use heat-resistant cable Remove with equipotential bonding Dimensional Drawing: Mounting of Probe Protective Case with PFE2 Filter Unit (dimensions in mm) X α Minimum distance xmin depending on mounting angle α: α xmin / mm 12 10° 229 15° 248 20° 268 25° 287 30° 307 35° 324 System Components for Sample Gas Sampling 10/23-5.10 EN March 2004 PFE3 Filter Unit Design The PFE3 filter unit consists of a modified FE2 filter unit with a self-regulating heater. It is installed in a stainless steel protective case. Process Gas Sampling Conditions Pressure pabs = 50 to 600 kPa Temperature max. 200 °C Flow Rate 30 to 500 l/h, relative to 100 kPa and 0 °C Pressure Drop approx. 0.6 hPa at 100 l/h Dust Content max. 3 g/m3 (application recommendation) Filter Characteristics Filter Element Surface filter Installation Filter Fineness 0.5 µm (optional 0.3 µm) Flange DN 65, PN 6, Form B to DIN 2527; positioning of holes: 45°, 135°, 225°, 315° Seal Integrity 10–4 hPa l/s Mounting Flange connect or bolt to the probe tube Dead Volume 280 ml Dimensions 50 / 20 x 135 mm Connection Ports Heater Test Gas Bulkhead fitting 6 mm (option) Sample Gas Pipe fitting 6 mm Type Self-regulating heater (400 W) Probe Tube G 3/4 internal thread to DIN ISO 228/1 Temperature 180 °C (ambient temperature 20 °C) Temperature Alarm approx. 150 °C Protective Case Dimensions 270 x 203 x 330 mm (H x W x D) Power Supply 110 to 240 V AC Material Special steel Mat. No. 1.4301 Option Annular heating element (see page 4) Degree of Protection of Case IP 54 Materials of the Gas-carrying Parts Casing, Flange and Connections Special steel Mat. No. 1.4571 Seals Viton 10/23-5.10 EN March 2004 Weight approx. 17 kg Ambient Temperature –20 to +45 °C System Components for Sample Gas Sampling 13 PFE3 Filter Unit Dimensional Drawing, Gas Connections (dim. in mm) Electrical Connections 204 3 1 2 1 PE 3 4 270 2 PE 5 6 3 7 2 8 2x M16 9 4 50 1 10 11 12 5 13 3x M20 2x M16 1 Test gas connection (optional) 6 mm bulkhead fitting 2 Sample gas connection 6 mm pipe fitting 3 Check valve for ACF-NT (optional) 1 Heater for filter 2 Heater for sample gas line 3 Pt 100 (optional) 4 Pt 100 sample gas line 5 Temperature monitor Ordering Information Catalog No. 23046-0↓ Design with protection box IP 54 (330x 203x 270 mm) with protection box IP 54 (330x 203x 270 mm) and check valve for ACF-NT ↓ ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 1 2 Filter housing (without connections for back purging) Housing stainless steel (filtration fineness 0,5 µm) Housing and flange stainless steel (filtration fineness 0,5 µm) Housing stainless steel with filter element <0,3 µm (filtration fineness) Housing and flange stainless steel, with filter element <0,3 µm (filtration fineness) 1 2 3 4 Heating With self-regulating heating and temperature alarm (110/220 V, AC) with ring heater unregulated 1 2 Power supply 230 V AC 115 V AC 1 2 ↓ Catalog No. 23046-0- 14 ↓ System Components for Sample Gas Sampling ↓ ↓ ↓ 10/23-5.10 EN March 2004 2) Probe H water cooled Probe 2 heated 1) 2) < 180 °C Probe 60 S water cooled Probe 2 unheated 1) 1) < 130 °C Probe 2 partially heated Probe tube type 60 water cooled Probe tube type 42 SS heated Probe tube type 40 silicone carbide Probe tube type 40 SS 1.4893 Probe tube type 40 SS 1.4571 coated 1) 2000 g/m3 Probe tube type K PVDF 40–100 g/m3 Max. temperature Probe tube type 40 SS 1.4571 Selection Table for Probe Tube Types and Probes 1) 1) 1) 1) 1) 3) 3) 180–450 °C 180–900 °C 1300 °C 1400 °C < 50 mg/m3 Dust content < 1000 mg/m3 < 3 g/m3 > 3 g/m3 Water vapor dew point 1) 1) 1) 1) < 65 °C > 65 °C Acidic aerosols (acid dew pt. 130 °C) Flow velocity < 10 m/s > 10 m/s Technical data see ... p. 16 p. 17 p. 17 p. 17 p. 17 p. 18 p. 19 p. 20 p. 20 p. 20 1) Back purging required 2) with regulated heating sleeve 3) “ACK” data sheet (Document No. 10/23-8.40 EN) Lead-Through Tube Installation Brickwork with metal sheeting 1 2 5 3 Isolated sheet-metal channel 1 2 Brickwork 5 3 1 4 2 5 3 α Flow direction of process gas 1 Lead-through tube 2 Mounting flange DN 65, PN 6, Form B to DIN 2573 3 Gasket 4 Welded-on rectangular block 5 Sampling probe flange Minimum distance from wall xmin depending on mounting angle α: α xmin / mm 10° 133 15° 138 20° 143 25° 147 30° 151 35° 153 Dimensions in mm 10/23-5.10 EN March 2004 System Components for Sample Gas Sampling 15 Type K Probe Tube Use Dimensional Drawing (dimensions in mm) Sampling of low-dust but strongly corrosive process gases up to 130 °C in conjunction with the FE2 / PFE2 filter unit. Area of application includes gas sampling in chemical industry and in power stations after the flue gas scrubber. 1 Process Gas Sampling Conditions Pressure pabs = max. 500 kPa (5 bar) 1 Temperature max. 130 °C 4 2 Flow Rate 30 to 300 l/h, relative to 1 bar, 0 °C, dry gas 1 Sample gas outlet 2 Outer ∅ 20 mm, inner ∅ 16 mm for L1 ≤ 1000 mm Outer ∅ 32 mm, inner ∅ 16 mm for L1 ≥ 1500 mm Pressure Drop ≤ 10 hPa at 250 l/h Flow max. 10 m/s Ordering Information Ambient Temperature –20 to +80 °C 90% Time (T90) < 2 s at 250 l/h Dust Content max. 50 mg/m3 in conjunction with the FE2 / PFE2 filter unit Length 500 mm Length 1000 mm Length 1500 mm Length 2000 mm Length 2500 mm Catalog No. 23025-4-0800857 23025-4-0800858 23025-4-0800859 23025-4-0800860 23025-4-0800861 Mounting Flange DN 65, PN 6, Form B, to DIN 2573 positioning of holes: 45°, 135°, 225°, 315° Mounting Angle any; 8° to 90° downwards from the horizontal is recommended (see dimensional drawing) Sample Gas Outlet G 1/4 internal thread DIN ISO 228/1 Materials of the Gas-carrying Parts The gas sampling probe K consists of PVDF, mounting flange included. It is available in several lengths. From 1,500 mm fitting length upwards the probe includes a metal support tube hermetically enclosed in plastic tubes. 16 System Components for Sample Gas Sampling 10/23-5.10 EN March 2004 Type 40 Probe Tube Use Dimensional Drawing (dimensions in mm) For sampling hot dust-containing gases in conjunction with the FE2 / PFE2 filter unit Temperature SS Mat. No. 1.4571 SS Mat. No. 1.4571 (coated) SS Mat. No. 1.4893 Silicone carbide (max. length 1.000 mm) max. 300 l/h Pressure Drop approx. 10 hPa at 30 to 90 l/h 17,2 ø L1 max. 450 °C max. 180 °C max. 900 °C max. 1.300 °C Flow Rate G3/4 16 Special Steel Material No. 1.4893 (max. 900 °C) 21,4 ø pabs = 50 to 500 kPa (0,5 to 5 bar) 16,8 ø Pressure 12,6 ø Process Gas Sampling Conditions Special Steel Material No. 1.4571 (max. 450 °C) G3/4 Dust Content max. 3 g/m3 (application recommendation); max. 40 g/m3 with blow-back device in conjunction with FE2 / PFE2 filter unit L1 16 Silicon-Carbide (max. 1300 °C) 34 ø Mounting Angle any; 8° to 90° downwards from the horizontal is recommended. It is imperative to install the silicon carbide probe tube vertically. 23 ø G 3/4 external thread DIN ISO 228/1 15 ø Installation G3/4 105 L1 16 Ordering Information Catalog No. 23036-0↓ Version Stainless steel (1.4571), up to 500 °C Stainless steel (1.4571), coated, up to 180 °C Silicon carbide up to 1300 °C, max. 1000 mm Stainless steel (1.4893), up to 900°C Stainless steel (1.4571), up to 300 °C, heated (for cement) 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 1 2 4 5 6 Nominal length 500 mm 1000 mm 1500 mm 2000 mm 2500 mm 3500 mm (only for heated version) 4000 mm (only for heated version) 4500 mm (only for heated version) 1 2 3 4 5 6 7 8 ↓ Catalog No. 23046-0- 10/23-5.10 EN March 2004 ↓ System Components for Sample Gas Sampling ↓ 17 Type 42 Probe Tube (heated) Use Probe Tube Heater • Low gas temperature • High moisture content in gas • Corrosive components in the process gas Heat Output nominal length 1.000 mm: 2 x 200 W nominal length 1.500 mm: 2 x 300 W nominal length 2.000 mm: 2 x 400 W Typical places of use are: • Sampling points in wet desulphurization plants (crude gas, clean gas, part-cleaned clean gas) • Waste incineration plants • Auxiliary boilers and stack emission measurements Nominal Voltage 230 V / 115 V AC, 50 to 60 Hz Control Control temperature max. 200 °C Process Gas Sampling Conditions Temperature Sensor Pt 100 Pressure pabs = 200 kPa (2 bar) Continuous Operation at 200 °C flue gas temperature 450 °C Temperature < 200 °C Ordering Information Flow Rate max. 250 l/h Length 1000 mm Length 1500 mm Length 2000 mm Flow Velocity max. 12 m/s (> 1,000 mm max. 7 m/s) Catalog No. 23041-4-8018392 23041-4-8018393 23041-4-8018394 Dust Content max. 3 g/m3 (application recommendation); max. 40 g/m3 with blow-back device in conjunction with FE2 / PFE2 filter unit T90 Time approx. 0.4 s at 250 l/h Dimensional Drawing (dimensions in mm) and Terminal Diagram 40 Heater 40 x 2 1000/1500/2000 mm 25 x 1,5 160 Pt100 12345 Temperature Controller Connectionr 18 System Components for Sample Gas Sampling 10/23-5.10 EN March 2004 Type 60 Probe Tube (Cement Flue Gas Probe) Use Cooling Water Supply For sampling hot cement flue gas with high dust content for CO, CO2, NO and SO2 analysis in “dry” cement production process (e.g. recuperator furnaces, furnaces with grate coolers), at rotary kiln inlet and/or after precalcination. Quality Drinking water quality with approx. 30 % anti-freeze if required Outlet Temperature min. 50 °C (or > dew point of the cement flue gas), max. 85 °C Process Gas Sampling Conditions Pressure pabs = max. 400 kPa (4 bar) Pressure pabs = max. 200 kPa (2 bar) Quantity Circulated approx. 3 m3/h (closed system) Temperature max. 1,200 °C for length = 3,500 mm max. 1,250 °C for length = 3,000 mm max. 1,300 °C for length = 2,500 mm max. 1,350 °C for length = 2,000 mm max. 1,400 °C for length = 1,500 mm Type 60 S Probe Tube Model with reinforced protection tube and flange DN 150, PN 16 Flow Rate max. 250 l/h Ordering Information Dust Content max. 2,000 g/m3 with periodical compressed air cleaning in conjunction with FE2 filter unit Length 1500 mm Length 2000 mm Length 2500 mm Length 3000 mm Length 3500 mm Length 1500 mm, reinforced protection tube Length 2000 mm, reinforced protection tube Length 2500 mm, reinforced protection tube Length 3000 mm, reinforced protection tube Length 3500 mm, reinforced protection tube Type 60 S, length 1500 mm Type 60 S, length 2000 mm Type 60 S, length 2500 mm Type 60 S, length 3000 mm Type 60 S, length 3500 mm Dead Time (Tt) for Length = 3,500 mm approx. 40 s at 60 l/h, approx. 9 s at 250 l/h 90% Time (T90) for Length = 3,500 mm approx. 82 s at 60 l/h, approx. 17 s at 250 l/h Pressure Drop approx. 10 hPa Materials of the Gas-carrying Parts Special steel, Material No. 1.4541 Mounting A flanged probe connector (internal diameter 200 mm) should be installed by the customer. The mounting flange is essential for installing the probe tube. Catalog No. 23038-4-0856831 23038-4-0856832 23038-4-0856833 23038-4-0856834 23038-4-0856835 23038-4-8018478 23038-4-8018479 23038-4-8018480 23038-4-8018481 23038-4-8018482 23038-4-0730642 23038-4-0730643 23038-4-0730644 23038-4-0730645 23038-4-0730646 Dimensional Drawing (dimensions in mm) ~60 3 2 A ø60,3 ~70 1 35 ø89 A 90 L2 L3 3 L1 10/23-5.10 EN March 2004 System Components for Sample Gas Sampling 1 Sample gas inlet ports 2 Thickened shank of probe tube 3 Cooling water connections L1 Probe tube length L2 Length of thickened shank: 800 mm for L1 ≤ 2,000 mm 1,300 mm for L1 ≥ 2,500 mm L3 200 mm with reinforced protection tube 19 Probe 2 Use Wherever large volumes of gas are sampled, gas pressure is raised and there is a high moisture or dust content in the gas. Process Gas Sampling Conditions Pressure pabs = max. 400 kPa (4 bar) Temperature max. 450 °C (probe tube unheated) max. 450 °C (probe tube partially heated) max. 130 °C (probe tube heated Flow Rate max. 2,000 l/h Dust Content max. 20 g/m3 (application recommendation), max. 100 g/m3 with purging Purging Pressure max. 600 kPa (6 bar) Heating Temperature of Inlet Filter ≥ 200 °C (at process gas temperature of 20 °C) 90% Time (T90) < 2 s at 250 l/h Filter Fineness 0.5 µm Sample Gas Outlet G 1/4 internal thread DIN ISO 228/1 Test Gas / Purge Air Inlet G 3/8 internal thread DIN ISO 228/1 Electrical Connections Harting connector, IP 55 Probe Tube Heating Heat Output (unregulated) Nominal length 1,200 mm: 650 W Nominal length 1,700 mm: 950 W partially heated (for heating of the wall bushing): 255 W Nominal Voltage 230 V / 115 V AC, 50 to 60 Hz Ordering Information Catalog No. 23022-4-0768301 23022-4-0768302 23022-4-0768303 23022-4-0768304 23022-4-0768305 23022-4-0768306 23022-4-0768307 23022-4-0768308 1200 mm, 220 V, 650 W 1700 mm, 220 V, 950 W 1200 mm, 115 V, 650 W 1700 mm, 115 V, 950 W 1200 mm, 220 V, 255 W, partially heated 1700 mm, 220 V, 255 W, partially heated 1200 mm, 115 V, 255 W, partially heated 1700 mm, 115 V, 255 W, partially heated Material of the Gas-carrying Parts Support Tube, Heating Rod, Connecting Pipe Special steel, Material No. 1.4571 Dimensional Drawing and Connection Diagram Z Inlet Filter Aluminum oxide Seals Ceramic fiber 650 1 Flange Special steel, Material No. 1.4581 α Z Mounting 1=N 2 = L1 Flange DN 65, PN 6, Form B, to DIN 2573 position of holes: 45°, 135°, 225°, 315° 4 260 Mounting Angle any; 10° to 35° downwards from the horizontal recommended (take direction of flow into account), see dimensional diagram 20 1 80 1 2 3 4 Filter mounting Gas sampling probe tube Sample gas outlet or test gas / purge air inlet Harting connector System Components for Sample Gas Sampling 10/23-5.10 EN March 2004 Probe 1, Probe 11 Probe 1 Probe 11 Use in interior air, garage and tunnel monitoring for the gases carbon monoxide (CO) and oxygen (SO2) Use in interior air, garage and tunnel monitoring, including corrosive atmospheres. The materials used ensure that with the gases normally occurring no adsorption occurs. Technical Data Technical Data Sampling Pressure Atmospheric pressure Sampling Pressure Atmospheric pressure Gas Temperature –25 to +50 °C Gas Temperature –25 to +50 °C Sample Gas Flow Rate max. 120 l/h Sample Gas Flow Rate max. 70 l/h Pressure Drop with New Filter ≤ 10 hPa at 120 l/h Pressure Drop with New Filter ≤ 10 hPa (mbar) at 70 l/h Dust Content in the Air approx. 100 mg/m3 Service Life of Filter (estimated) filter constant (2500) d (days) = flow rate (l/h) x dust concentration (mg/m3) Sample Gas Connection Port G 1/8 internal thread DIN ISO 228/1 or nozzle for flexible tube of internal diameter 4 mm Materials Silicone carbide (filter), aluminum (end seals), PVC (nozzle) Transport and Storage Temperature –25 to +70 °C Mounting On wall or suspended from ceiling Ordering Information Ordering Information Catalog No. 23011-5-5862922 Room air probe 1 Dimensional Drawing (dimensions in mm) Room air probe 11 Catalog No. 23021-4-0859585 Dimensional Drawing (dimensions in mm) ~ ~ 104 ~ ~ 83 Materials Polycarbonate (probe head), aluminum oxide (filter) Transport and Storage Temperature –25 to +70 °C Mounting On wall or suspended from ceiling ~ ~140 Sample Gas Connection Port Bolted connection for flexible pipe with internal diameter 4 mm and external diameter 7 mm 8,4 ~ ~ 57 85 mm wide 10/23-5.10 EN March 2004 System Components for Sample Gas Sampling 21 Ceramic Filter Use For filtering dust from gases in the gas conditioning systems of analyzer systems when the gas sampling probe does not have a filter. The ceramic filter is positioned upstream the other devices in the sample gas conditioning chain. The ceramic filter can be heated with the annular heating element if condensation from the sample gas is likely. Technical Data Flow Rate (Air) max. 300 l/h Gas Pressure pabs = 20 kPa to 1.7 MPa (0.2 to 17 bar) at 20 °C filter temperature, max. 200 kPa (2 bar) at 180 °C Pressure Drop 0.1 to 0.85 hPa at 30 to 250 l/h (air) Dynamic Characteristics at Flow rate (air) 30 l/h 3.7 s Dead time Tt 90% time T90 11.5 s 60 l/h 2.3 s 5.5 s 90 l/h 1.7 s 3.9 s Filter Fineness 0.5 µm Dimensional Drawing (dimensions in mm) Construction, Mounting Materials of the Gas-carrying Parts Housing: steel, St 37 K or special steel, Material No. 1.4571; Filter: aluminum oxide; Seals: FPM, Cu, Al2O3 fiber Gas Connection Ports G 1/4 internal thread DIN ISO 228/1 or pipe connectors for pipe of internal diameter 4 mm Filter Surface approx. 212 cm2 Weight approx. 2.1 kg Filter Heating Annular heating element, see page 4 Mounting On wall with angle bracket Ordering Information Ceramic filter, steel version Ceramic filter, stainless steel version 22 Catalog No. 23031-5-8018177 23031-5-8018178 System Components for Sample Gas Sampling 10/23-5.10 EN March 2004 Process Pressure Reducers Use If the pressure of the process gas is too high for the analyzer, a pressure reducer is required to lower the pressure. The type of pressure reducer used is determined by the properties of the process gas and the required sample gas pressure. Technical Data Casing: Brass Casing: Brass Casing: Special steel Casing: Special steel Single-stage with one manometer (outlet) max. 1.5 MPa (15 bar) 0.8 to 10 kPa (8 to 100 mbar) Single-stage without manometer max. 3.5 MPa (35 bar) 10 to 70 kPa (0.1 to 0.7 bar) 80 kPa (0.8 bar) – 25 to 250 l/h Special steel EPDM Special steel Single-stage with two manometers (inlet / outlet) max. 3.5 MPa (35 bar) 10 to 70 kPa (0.1 to 0.7 bar) 80 kPa (0.8 bar) 0 to 4 MPa (0 to 40 bar) 0 to 150 kPa (0 to 1.5 bar) 25 to 250 l/h Special steel EPDM Special steel approx. 2.35 kg max. 50 °C approx. 2.35 kg max. 50 °C 23422-4-0730699 23422-4-0730700 Weight Gas temperature and ambient temperature approx. 1.25 kg max. 50 °C Single-stage with two manometers (inlet / outlet) max. 3.5 MPa (35 bar) 10 to 40 kPa (0.1 to 0.4 bar) 50 kPa (0.5 bar) 0 to 4 MPa (0 to 40 bar) 0 to 60 kPa (0 to 0.6 bar) 25 to 250 l/h Buna rubber Buna rubber Special steel; gascarrying parts: brass approx. 2.65 kg max. 50 °C Catalog No. 23035-5-5864958 23035-5-5864951 Construction Inlet pressure Outlet pressure (adjustable) Safety valve set to – Manometer display Inlet pressure Manometer display Outlet pressure Flow rate Material Diaphragm Material Gasket Material Housing – 0 to 250 hPa (0 to 250 mbar) 25 to 250 l/h Buna rubber Buna rubber Brass – Supplied: Pressure reducer with angle bracket 10/23-5.10 EN March 2004 System Components for Sample Gas Sampling 23 Process Pressure Reducers Dimensional Drawings (dimensions in mm) Pressure reducer single-stage with one manometer (outlet) (Catalog No. 23035-5-5864958) Pressure reducer single-stage without manometer (Catalog No. 23422-4-0730699) x. 30 appro Ø4 79 able) ° (vari 105 GasTech approx. 94 (depending on handwheel position) Pressure reducer single-stage with two manometers (inlet / outlet) (Catalog No. 23035-5-5864951) Pressure reducer single-stage with two manometers (inlet / outlet) (Catalog No. 23422-4-0730700) 118 appro ble) (varia 148 Ø4 ° x. 30 GasTech approx. 94 (depending on handwheel position) 24 System Components for Sample Gas Sampling 10/23-5.10 EN March 2004 Heated Sample Gas Lines Heated Sample Gas Line CGWB 13 Parallel strip heater (holding temperature max. 200 °C) Pt 100 resistance thermometer for control system Flexible sample gas pipe, replaceable (option) Self-regulating anti-frost heater (option) Self-regulating heater (option, holding temperature max. 100 °C) Heated sample gas line TBL 01 Fixed resistor heating conductor (holding temperature max. 200 °C) Pt 100 resistance thermometer for control system Self-regulating heater (option, holding temperature max. 100 °C) Heated sample gas line TBL 01-E Competitive version Pipe ends with polyamide hard caps Selection Table for Heated Sample Gas Lines Heater Self-regulating strip heater (30 °C) Self-regulating strip heater (100 °C) Fixed resistor heating conductor (200 °C) Parallel strip heater Outer covering Options Line can be shortened Gas-carrying line can be replaced Small bending radius Gas connections Electrical connections (heater and Pt 100) Maximum line length Applications CO, O2, CO2 CO, O2, CO2 SO2, NOx, CnHm 10/23-5.10 EN March 2004 CGWB 13 TBL 01 TBL 01-E x x – x Polyvinyl chloride – x x – Polyamide – – x – Polyamide x x – Various options Power fixed, Pt 100 plug 130 m (60 m) – – x Various options Power and Pt 100 plug 100 m (40 m) – – x Stainless steel Power and Pt 100 fixed 40 m Anti-frost heater Holding temp. 100 °C Holding temp. 200 °C – Holding temp. 100 °C Holding temp. 200 °C – – Holding temp. 200 °C System Components for Sample Gas Sampling 25 CGWB 13 Heated Sample Gas Line Technical Data Self-regulating strip heater Regulated parallel strip heater Outer covering PVC Medium-carrying pipe PTFE Integral pipe-end connectors 1) POM cap Frost protected line PVC PTFE POM cap Heater Power supply Heat output Current consumption self-regulating strip heater 230 V or 115 V, 50 to 60 Hz 38 W/m 0.15 A/m Holding temperature Ambient temperature Tamb External diameter self-regulating strip heater 230 V or 115 V, 50 to 60 Hz 16 W/m Making / holding current = 65/25 mA see diagram –30 to +60 °C 30 mm PVC PTFE Metal cap and metal cap with flange CPD heating element, regulated 230 V or 115 V, 50 to 60 Hz 60 W/m 0.3 A/m Flexible sample gas pipe Maximum line length 6 x 4 x 1 mm or 8 x 6 x 1 mm 130 m 6 x 4 x 1 mm or 8 x 6 x 1 mm 90 m Line duct Minimum bending radius Weight Connection lead for heater (fixed) Connection lead for Pt 100 (with plug) PG 42 300 mm approx. 1 kg/m 3 x 2.5 mm2; length 2.5 m PG 42 300 mm approx. 1 kg/m 3 x 2.5 mm2; length 2.5 m 200 °C –30 to +60 °C 40 mm; with replaceable flexible pipe: 45 mm 6 x 4 x 1 mm or 8 x 6 x 1 mm 60 m with replaceable flexible pipe: 25 m, 60 m if laid straight PG 42 300 mm approx. 1 kg/m 3 x 2.5 mm2; length 2.5 m – – LIYCI 2 x 0.14 mm2; length 5 m see diagram –30 to +60 °C 43 mm 1) For line lengths > 10 m, only the analyzer side is assembled, an assembly set for the probe side is supplied. Holding Temperature Diagrams Self-regulating strip heater, 6 x 4 x 1 mm PTFE Medium pipe temperature °C Medium pipe temperature °C Frost protected line, 6 x 4 x 1 mm PTFE, 16 W/m, self-regulating 50 40 30 20 - 30 - 20 - 10 0 + 10 + 20 + 30 100 90 80 70 + 40 - 30 - 20 Ambient temperature °C - 10 0 + 10 + 20 + 30 + 40 Ambient temperature °C Dimensional Drawing (dimensions in mm) L= m 300 24 ∅6 11 26 System Components for Sample Gas Sampling 50 60 49 Analyzer Side ∅ 30 ∅6 Probe Side 42 56 10/23-5.10 EN March 2004 CGWB 13 Heated Sample Gas Line Ordering Information Catalog No. 23281-0↓ Version controlled (Pt 100) self-regulated 16 W/m, 30°C (antifreeze) self-regulated 38 W/m, 100°C ↓ ↓ ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 2 3 7 Connection probe side with metal cap for PFE 2 probe: metal cap / analyzer: flange end of line: plastics 2 3 4 Gas conducting line PTFE-tube 6x4x1 mm PTFE-tube 8x6x1 mm 1 2 Power supply 230 V, 50/60 Hz 115 V, 50/60 Hz 1 2 Controlling and monitoring without Pt 100 with 1 x Pt 100 0 1 ↓ Catalog No. 23281-0- 10/23-5.10 EN March 2004 ↓ 0 ↓ System Components for Sample Gas Sampling ↓ ↓ ↓ ↓ 27 TBL 01, TBL 01-E Heated Sample Gas Lines Technical Data TBL 01 Self-regulating strip heater 30 W/m TBL 01 Fixed resistor heating conductor 90 W/m TBL 01-E Fixed resistor heating conductor 90 W/m Outer covering Medium-carrying pipe Integral pipe end connectors Polyamide corrugated tubing PTFE Hard caps, metal caps Polyamide corrugated tubing PTFE Polyamide hard caps Power supply Heat output Current consumption 230 V or 115 V, 50 to 60 Hz 30 W/m I = 0.15 A/m (starting current), I = 65 mA (holding current) see diagram on page 26 –30 to +60 °C 43 mm 6 x 4 x 1 mm or 8 x 6 x 1 mm – 100 m PG 48 200 mm approx. 1 kg/m Screw fitting (fixed for hard cap) Polyamide corrugated tubing PTFE Hard caps, metal caps and metal cap with flange 230 V or 115 V, 50 to 60 Hz 90 W/m I = f(length) Holding temperature Ambient temperature External diameter Flexible sample gas pipe Flexible test gas pipe Maximum line length Line duct Minimum bending radius Weight Electrical connection for heater and Pt 100 200 °C –30 to +60 °C 43 mm 6 x 4 x 1 mm or 8 x 6 x 1 mm 6 x 4 x 1 mm 40 m PG 48 200 mm approx. 1 kg/m Plugs (see dimensional drawing; fixed for hard cap) 230 V or 115 V, 50 to 60 Hz 90 W/m I = f(length) 200 °C –30 to +60 °C 43 mm 6 x 4 x 1 mm or 8 x 6 x 1 mm – 40 m – 200 mm approx. 1 kg/m fixed TBL 01 Dimensional Drawing (dimensions in mm) Analyzer side A Probe side B PG11 50 24 42 ∅6 ∅6 60 49 ∅6 A,B 24 PG7 11 56 11 80 250 300 L= m TBL 01-E Dimensional Drawing (dimensions in mm) End Cap PA SS 53 ∅6 ∅6 SS 1000 1000 120 120 Sensor Location Length max. 40 m 28 System Components for Sample Gas Sampling 10/23-5.10 EN March 2004 TBL 01, TBL 01-E Heated Sample Gas Lines TBL 01 Ordering Information Catalog No. 23283-0↓ Design Resistor heating line 90 W/m self regulating heating line 30 W/m ↓ ↓ ↓ ↓ ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 ↓ 0 1 2 Gas conducting line with PTFE-tube 6x4x1 mm with PTFE-tube 8x6x1 mm 1 3 Material of sheathing with plastic hose of PA (outside and indoors installation) 1 Connections End of line with flange connection End of line: probe: metal cap / analyzer: flange End of line with flanges End of line with plastic tops End of line with 2 metal tops 1 5 6 7 8 Control / Monitoring without Pt 100 with 1 x Pt 100 0 1 Power supply 230 V, 50 ... 60 Hz 115 V, 50 ... 60 Hz 1 2 ↓ ↓ Catalog No. 23283-0- ↓ ↓ ↓ ↓ TBL 01-E Ordering Information Resistor heating line 90 W/m, PTFE 6x4x1, PA caps, 230V Resistor heating line 90 W/m, PTFE 8x6x1, PA caps, 230V Resistor heating line 90 W/m, PTFE 6x4x1, PA caps, 115V Resistor heating line 90 W/m, PTFE 8x6x1, PA caps, 115V 10/23-5.10 EN March 2004 Catalog No. 23284-4-0730695 23284-4-0730696 23284-4-0730697 23284-4-0730698 System Components for Sample Gas Sampling 29 Electronic Temperature Controller and Switching Relay Temperature Controller Control Range Connection Diagram –200 to +850 °C Measurement Input Pt 100 Outputs 2 x Relay 3 A, 250 V AC Power Supply 110 to 240 V AC +10 / –15 %, 48 to 63 Hz Dimensions 48 x 48 mm, C-rail housing Weight 140 g Catalog No. 23345-5-8328859 Switching Relay Load Voltage Urms = 24 to 265 V Max. Load Current Irms = 30 A (Ambient temperature = 25 °C) Weight 200 g Catalog No. 23345-5-8328731 30 System Components for Sample Gas Sampling 10/23-5.10 EN March 2004 The IndustrialIT word mark and all mentioned product names in the form XXXXXIT are registered or pending trademarks of ABB. ABB has Sales & Customer Support expertise in over 100 countries worldwide. The Company’s policy is one of continuous product improvement and the right is reserved to modify the information contained herein without notice. Printed in the Fed. Rep. of Germany (03.04) ABB Automation Products GmbH Analytical Division Stierstaedter Strasse 5 60488 Frankfurt am Main Germany Phone: +49 69 793040 Fax: +49 69 79304566 EMail: [email protected] 10/235.10 EN Rev. 3 © ABB 2004 www.abb.com
© Copyright 2024