Recent Advances in Sample Preparation for Accelerated Solvent Extraction SM Rahmat Ullah, Kannan Srinivasan, and Chris Pohl Thermo Fisher Scientific, Sunnyvale, CA, USA Overview Methods Purpose: To demonstrate the utility of a new moisture absorbing polymer as a drying agent for extraction applications. Polymer used Methods: Moisture absorption capacity of a new polymer was studied. The in-cell moisture removal was performed in an extraction setup in an accelerated solvent extraction instrument at high temperature and pressure. Analytes in the extract were measured by high-pressure liquid chromatography (HPLC) and gas chromatography (GC). Results: The moisture absorbing polymer and the diatomaceous earth can remove moisture when mixed with wet sample for in-cell, in-line extraction method. An addition of a small amount of polymer in the collection bottle can also result in complete water removal from the collected extract. Introduction Analyses of organic pollutants are becoming increasingly important, and often with the need to isolate and analyze trace organic compounds from a variety of matrices such as soil, sediment, fruits, and vegetables. In this regard, sample pretreatment constitutes an important step prior to analysis. The purpose of the sample pretreatment step is to selectively isolate or concentrate the analytes of interest from matrix components and present a sample suited for routine analysis by established analytical techniques such as gas chromatography or high-pressure liquid chromatography. Typical sample pretreatment steps include techniques such as solid phase extraction, liquid-liquid extraction, solid-liquid extraction, dilution, evaporation, distillation and the like. Accelerated solvent extraction is a technique used for extracting the analytes of interest from a solid, semisolid or liquid sample by performing extraction using an organic solvent at elevated temperature and pressure. The elevated pressure also elevates the boiling temperature of the solvent, thereby allowing faster extraction to be conducted at relatively higher temperatures. The benefit of a relatively high temperature extraction is primarily speed; thus the extraction process is significantly faster than traditional methods such as Soxhlet extraction. In some samples containing moisture or water such as soil samples or food samples (fruits, vegetables, etc.) an additional step may be needed either before the extraction to remove the moisture from the samples or post extraction to remove the moisture from the extracted solvent (containing the extracted analytes). Sample drying can be accomplished in several ways such as air drying and oven drying prior to extraction. However, these approaches are not suited when analyzing volatile or semivolatile components as they would be removed from the sample prior to extraction or analysis. Another common method for moisture removal is by using salts such as sodium sulfate, calcium chloride, magnesium sulfate, calcium sulfate and the like. These salts tend to associate to water molecules to form hydrated salts. Sodium sulfate for example tends to clump together when water is present. Sodium sulfate is not suitable for in-cell moisture removal and accelerated solvent extraction. Sodium sulfate can dissolve in hot solvent to a certain extent and can precipitate downstream in some instances clogging the outlet frit, tubes and valves. Moreover, sodium sulfate becomes an aggregate hard lump upon water absorption and is not easy to process during sample preparation for in-cell moisture removal and extraction (1). Polymers have been used for moisture removal such polymers have been designated super absorbent polymers (2). The most common polymer is the sodium salt of polyacrylic acid. Although this polymer removes water by absorbing it into the polymer matrix, the water absorbing capacity decreases as the ionic strength increases. Another limitation of the polymer is poor water absorbing property under high temperature conditions. Yet another limitation of this polymer is that it becomes a hard plug inside the extraction cell (1). The present research solves these issues. Here, the authors synthesized a water absorbent polymer comprising a copolymer of a basic monomer and an acidic monomer. This combination is suitable for moisture removal under high ionic strength conditions. Results shown here also demonstrate that when the polymer is mixed with diatomaceous earth (DE) the water removal efficiency increases significantly. Different formats of using the polymer such as in-cell, in-vial and a combination of the two are discussed. 2 Recent Advances in Sample Preparation for Accelerated Solvent Extraction A proprietary moisture absorbing po Sample preparation Sample containing moisture was us polymer or the polymer was combin Dionex™ ASE™ extraction cell for t Accelerated Solvent Extraction syste Extraction using the Dionex ASE Sample extraction at high temperatu 350 Accelerated Solvent Extractor s Dionex ASE 350 extraction condition Pressure: 1500 psi Temperature: 100 ⁰C- 150 ⁰C Static time: 5-10 min Cycles: 1-3 Flush: 10-75% Purge: 120 sec Solvent: Hexane or 1/1 Acetone/Dic Measurement of moisture remova Known amounts of water were spike measure the water removal capacity solvent. Similarly, water removal cap temperature for comparison purpose Water removal capacity of polyme The water removal capacity of the p water present in the cell without any Liquid chromatography A P680 pump, PDA-100 photodiode used for chromatographic separatio Chromatography Data System softw column was a Thermo Scientific ™ flow rate was 1.0 mL/min. Separatio mM acetic acid/ammonium acetate B changed from 25% to 70% over 1 Gas Chromatography A GC with Flame Ionization Detecto hydrocarbon (PAH). The Chromeleo .The separation column was a Therm The helium carrier gas flow rate wa min), 25 °C/min to 140 °C , then 10 a new moisture absorbing polymer as a drying ty of a new polymer was studied. The in-cell extraction setup in an accelerated solvent ure and pressure. Analytes in the extract were matography (HPLC) and gas chromatography mer and the diatomaceous earth can remove for in-cell, in-line extraction method. An addition of ction bottle can also result in complete water oming increasingly important, and often with the nic compounds from a variety of matrices such as In this regard, sample pretreatment constitutes an urpose of the sample pretreatment step is to nalytes of interest from matrix components and alysis by established analytical techniques such as liquid chromatography. Typical sample such as solid phase extraction, liquid-liquid on, evaporation, distillation and the like. hnique used for extracting the analytes of interest e by performing extraction using an organic solvent The elevated pressure also elevates the boiling owing faster extraction to be conducted at relatively elatively high temperature extraction is primarily ignificantly faster than traditional methods such as or water such as soil samples or food samples tep may be needed either before the extraction to or post extraction to remove the moisture from the cted analytes). Sample drying can be accomplished oven drying prior to extraction. However, these zing volatile or semivolatile components as they or to extraction or analysis. removal is by using salts such as sodium sulfate, calcium sulfate and the like. These salts tend to ydrated salts. Sodium sulfate for example tends to l moisture removal and accelerated solvent e in hot solvent to a certain extent and can ces clogging the outlet frit, tubes and valves. aggregate hard lump upon water absorption and is paration for in-cell moisture removal and extraction removal such polymers have been designated st common polymer is the sodium salt of removes water by absorbing it into the polymer ecreases as the ionic strength increases. Another absorbing property under high temperature s polymer is that it becomes a hard plug inside the ch solves these issues. absorbent polymer comprising a copolymer of a er. This combination is suitable for moisture removal esults shown here also demonstrate that when the arth (DE) the water removal efficiency increases the polymer such as in-cell, in-vial and a Methods Results Polymer used Moisture absorbing capability A proprietary moisture absorbing polymer was used here. The amount of moisture absorbing po absorbing one gram of water at room Sample preparation Sample containing moisture was used or a spiked sample with water was added to the polymer or the polymer was combined with DE and loaded into the Thermo Scientific™ Dionex™ ASE™ extraction cell for the extraction using the Thermo Scientific Dionex ASE Accelerated Solvent Extraction system. Extraction using the Dionex ASE 350 System Sample extraction at high temperature and pressure was performed using Dionex ASE 350 Accelerated Solvent Extractor system. Dionex ASE 350 extraction conditions: The moisture removal formats a) In-cell (in-line) moisture removal: The moisture absorbing polymer can in-cell moisture removal and extractio moisture removal the polymer is used Dispersant for ASE. The ASE Prep DE accelerated solvent extraction, therefo customer’s current practices. b) In-vial (off-line) moisture removal: The amount of moisture absorbing po of water at room temperature. A simp (proportional to the amount of water p complete water removal from the extr needed is 4 g to absorb one gram of polymer in an aliquot basis. Pressure: 1500 psi Temperature: 100 ⁰C- 150 ⁰C Static time: 5-10 min Cycles: 1-3 Flush: 10-75% Purge: 120 sec Solvent: Hexane or 1/1 Acetone/Dichloromethane Measurement of moisture removal capacity at room temperature Known amounts of water were spiked as the sample and was mixed with the polymer to measure the water removal capacity at room temperature in the presence of organic solvent. Similarly, water removal capacity of sodium sulfate was measured at room temperature for comparison purposes. c) Combination Mode: In this mode, the in-cell moisture rem removal (off line). If some breakthroug addition of a small amount of polymer moisture removal. In fact a small amo there would be no moisture present in for samples with unknown moisture c Water removal capacity of polymer and polymer-DE The water removal capacity of the polymer was measured as the maximum amount of water present in the cell without any breakthrough of the water into the collection bottle. Liquid chromatography A P680 pump, PDA-100 photodiode array detector and a chromatographic oven were used for chromatographic separation. Thermo Scientific™ Dionex™ Chromeleon™ 6.8 Chromatography Data System software was used for data acquisition. The separation column was a Thermo Scientific ™ Acclaim ™ Polar Advantage C16 5µm 4.6 x 150 mm, flow rate was 1.0 mL/min. Separation was based on gradient elution of A comprising 25 mM acetic acid/ammonium acetate and B comprising of acetonitrile where composition of B changed from 25% to 70% over 17.5 min. Detection wavelength was 280 nm. Gas Chromatography A GC with Flame Ionization Detector (FID) was used for the analysis of polyaromatic hydrocarbon (PAH). The Chromeleon 6.8 CDS software was used for data acquisition .The separation column was a Thermo Scientific ™ TR-5MS 30 m x 0.25 mm x 0.25 5µm. The helium carrier gas flow rate was 1.5 mL/min. The temperature profile was 65 °C (1 min), 25 °C/min to 140 °C , then 10 °C/min to 290 °C. The run time was 40 minutes. In-cell moisture removal A combination of polymer and DE (1:1 Dionex ASE 350 system at various ex used as the extraction solvent. Result efficiency dropped as the temperature was still feasible even at 150 °C. The agent required for a given sample wit sediments, fruits and vegetable with r temperature. Table 1. In-cell water removal capa temperatures and cell sizes using t Accelerated solvent extraction temperature, ⁰C Total water present in the cell, g 100 5.05 125 2.54 150 2.15 100 10.05 125 5.05 150 2.09 100 15.15 125 8.14 150 2.14 Thermo Scientific Poster Note • PN70546_E 03/13S 3 Results Analyte recovery for in-cell moist Analyte recovery from the extraction phenols, two anilines and a neutral a presence of 8 g of the polymer and a mL standard solution in acetonitrile a was then extracted using the Dionex dichloromethane: acetone solvent a was evaporated to 10 mL under nitro µg/mL for a 10 mL extract. Moisture absorbing capability mer was used here. or a spiked sample with water was added to the with DE and loaded into the Thermo Scientific™ extraction using the Thermo Scientific Dionex ASE . 0 System and pressure was performed using Dionex ASE em. The amount of moisture absorbing polymer needed is approximately 0.20 g for absorbing one gram of water at room temperature. The moisture removal formats a) In-cell (in-line) moisture removal: The moisture absorbing polymer can remove moisture when mixed with wet sample for in-cell moisture removal and extraction. For improved flow characteristics and improved moisture removal the polymer is used in conjunction with Diatomaceous Earth (DE) Dispersant for ASE. The ASE Prep DE is normally used with the sample when pursuing accelerated solvent extraction, therefore adding the polymer to this setup maintains the customer’s current practices. b) In-vial (off-line) moisture removal: The amount of moisture absorbing polymer needed is 0.20 g for absorbing one gram of water at room temperature. A simple addition of a small amount of polymer (proportional to the amount of water present) in the collection bottle can result in complete water removal from the extract. In comparison the amount of sodium sulfate needed is 4 g to absorb one gram of water. Further there is no need to add the polymer in an aliquot basis. romethane apacity at room temperature as the sample and was mixed with the polymer to t room temperature in the presence of organic city of sodium sulfate was measured at room ray detector and a chromatographic oven were Thermo Scientific™ Dionex™ Chromeleon™ 6.8 e was used for data acquisition. The separation claim ™ Polar Advantage C16 5µm 4.6 x 150 mm, was based on gradient elution of A comprising 25 d B comprising of acetonitrile where composition of 5 min. Detection wavelength was 280 nm. FID) was used for the analysis of polyaromatic 6.8 CDS software was used for data acquisition Scientific ™ TR-5MS 30 m x 0.25 mm x 0.25 5µm. 1.5 mL/min. The temperature profile was 65 °C (1 /min to 290 °C. The run time was 40 minutes. Analyte recovery for in-vial moisture using the same concentration level o water (5 mL) was spiked into the 10 polymer or by sodium sulfate. The re recoveries were obtained for moistu sulfate. Table 2. Recovery of analytes. Analytes c) Combination Mode: In this mode, the in-cell moisture removal (in-line) is followed by in-vial moisture removal (off line). If some breakthrough of water is observed in the extract then addition of a small amount of polymer in the collection bottle can result in complete moisture removal. In fact a small amount in the collection vessel always ensures that there would be no moisture present in the samples. This mode is particularly useful for samples with unknown moisture content. and polymer-DE mer was measured as the maximum amount of eakthrough of the water into the collection bottle. The 10 mL extract was analyzed usi above experiment are shown in Tab acceptable performance of the polym acceptance criteria of ± 30% as per there is no detrimental effect of usin analytes. Total level 10 m 1 2, 4-Dinitrophenol 3 2 Phenol 3 A combination of polymer and DE (1:1) was tested for in-cell water removal using a Dionex ASE 350 system at various extraction temperatures and cell sizes. Hexane was used as the extraction solvent. Results are shown below in Table 1. The water removal efficiency dropped as the temperature increased, nevertheless practical water removal was still feasible even at 150 °C. The data provides guideline on the amount of drying agent required for a given sample with a known range of water content such as sediments, fruits and vegetable with respect to a chosen cell size and extraction temperature. 3 P-Toluidine 3 4 4-Nitrophenol 3 5 2-Chlorophenol 3 6 4-Ethylaniline 3 7 4-Chloroaniline 3 8 2-Nitrophenol 3 Table 1. In-cell water removal capacity of the polymer-DE at different temperatures and cell sizes using the Dionex ASE 350 system. 9 2, 4Dichlorophenol 3 10 2, 4, 6Trichlorophenol 3 In-cell moisture removal Accelerated solvent extraction temperature, ⁰C Total water present in the cell, g Drying agent (Polymer and DE) Cell size, mL Water observed in the collection bottle, g 100 5.05 2 g each 34 No 125 2.54 2 g each 34 No 150 2.15 2 g each 34 No 100 10.05 4 g each 66 No 125 5.05 4g each 66 No 150 2.09 4 g each 66 No 100 15.15 6 g each 100 No 125 8.14 6 g each 100 No 150 2.14 6 g each 100 No 4 Recent Advances in Sample Preparation for Accelerated Solvent Extraction Polyaromatic hydrocarbon (PAH) re cell (in-line) moisture removal and e clean soil sample was spiked with P 30% moisture) by adding water. The polymer and DE and loaded into a 3 using the Dionex ASE 350 instrume acetone) at an extraction temperatu under nitrogen stream at 40⁰C. The final extract volume of 1 mL. The extract was analyzed using a G experiment are shown in Table 3. A it was still deemed acceptable base Method 8270. The extraction condit these two PAHs, naphthalene and a Analyte recovery for in-cell moisture removal and in-vial moisture removal Analyte recovery from the extraction using the polymer was studied using a mixture of phenols, two anilines and a neutral analyte. An in-cell extraction was pursued in the presence of 8 g of the polymer and a spike with a standard solution that contained 1.5 mL standard solution in acetonitrile and 8.5 mL of water. The spiked polymer sample was then extracted using the Dionex ASE 350 instrument and a 1:1 ratio of dichloromethane: acetone solvent at an extraction temperature of 100⁰ C. The extract was evaporated to 10 mL under nitrogen stream at 40⁰C. The spike level was 30 µg/mL for a 10 mL extract. mer needed is approximately 0.20 g for mperature. move moisture when mixed with wet sample for For improved flow characteristics and improved conjunction with Diatomaceous Earth (DE) s normally used with the sample when pursuing e adding the polymer to this setup maintains the mer needed is 0.20 g for absorbing one gram addition of a small amount of polymer sent) in the collection bottle can result in t. In comparison the amount of sodium sulfate ter. Further there is no need to add the The 10 mL extract was analyzed using an HPLC instrument The results from the above experiment are shown in Table 2. The per cent recovery data showed acceptable performance of the polymer for these test analytes based on an acceptance criteria of ± 30% as per EPA Method 8270. Moreover, it also indicates that there is no detrimental effect of using this polymer for in-cell extraction for these test analytes. Analyte recovery for in-vial moisture removal at room temperature was also studied using the same concentration level of analytes in a 10 mL solvent. A known amount of water (5 mL) was spiked into the 10 mL solvent. The water was removed either by the polymer or by sodium sulfate. The results are shown in Table 2. Comparable recoveries were obtained for moisture removal by the polymer compared to sodium sulfate. Table 2. Recovery of analytes. Analytes al (in-line) is followed by in-vial moisture of water is observed in the extract then the collection bottle can result in complete nt in the collection vessel always ensures that he samples. This mode is particularly useful tent. was tested for in-cell water removal using a action temperatures and cell sizes. Hexane was are shown below in Table 1. The water removal ncreased, nevertheless practical water removal ata provides guideline on the amount of drying a known range of water content such as pect to a chosen cell size and extraction y of the polymer-DE at different e Dionex ASE 350 system. ying agent olymer and E) Cell size, mL Water observed in the collection bottle, g 2 g each 34 No 2 g each 34 No 2 g each 34 No 4 g each 66 No 4g each 66 No 4 g each 66 No 6 g each 100 No 6 g each 100 No 6 g each 100 No Total spike level in 10 mL In-cell moisture removal by the polymer In-vial moisture removal by the polymer In-vial moisture removal by the sodium sulfate µg % recovery % recovery % recovery 1 2, 4-Dinitrophenol 300 81.01 82.96 86.79 2 Phenol 300 89.99 88.58 88.41 3 P-Toluidine 300 86.72 95.03 93.88 4 4-Nitrophenol 300 96.05 90.96 88.93 5 2-Chlorophenol 300 92.33 87.45 87.16 6 4-Ethylaniline 300 92.82 89.47 88.02 7 4-Chloroaniline 300 94.44 88.39 87.51 8 2-Nitrophenol 300 93.83 92.64 87.39 9 2, 4Dichlorophenol 300 97.50 93.25 89.58 10 2, 4, 6Trichlorophenol 300 97.51 95.20 91.02 Polyaromatic hydrocarbon (PAH) recovery from spiked soil sample was pursued for incell (in-line) moisture removal and extraction in the Dionex ASE 350 system. A 5 g clean soil sample was spiked with PAHs and then the soil was made moist (to about 30% moisture) by adding water. The spiked wet soil sample was mixed with 4 g of 1:1 polymer and DE and loaded into a 34 mL ASE cell. The soil sample was then extracted using the Dionex ASE 350 instrument by solvent (1:1 ratio of dichloromethane: acetone) at an extraction temperature of 100⁰ C. The extract was evaporated to 1 mL under nitrogen stream at 40⁰C. The spike level was calculated as 20 µg/mL for the final extract volume of 1 mL. The extract was analyzed using a GC instrument with FID. The results from the above experiment are shown in Table 3. Although two compounds showed a lower recovery, it was still deemed acceptable based on the acceptance criteria of ± 30% as per EPA Method 8270. The extraction condition yet to optimize to increase the recovery of these two PAHs, naphthalene and acenaphthylene. Table 3. PAH recovery by using GCAnalytes 1 Naphthalene 2 Acenapththyene 3 Acenaphthylene 4 Phenanthrene 5 Anthracene 6 Fluoranthene 7 Pyrene 8 Benzo(a)anthracene 9 Chrysene 10 Benzo(b)fluoranthene 11 Benzo(k)fluoranthene 12 Benzo(a)pyrene 13 Indeno((1,2,3,c,d)pyrene 14 Dibenzo(a,h)anthracene 15 Benzo(g,h,i)perylene Analyte recovery for in-vial (off-line) mo using the same solvent, 1:1 dichlorome PAHs. A known amount of water was a The water was removed either by the p evaporated to 1 mL under nitrogen stre final extract volume of 1 mL. The resul were obtained for in-vial moisture remo Therefore, the usefulness of the polym table. Conclusion The utility of a new polymer for r from a collected extract is shown The new polymer is designed to intended for in-cell (in-line) mois and a combination of both in-cel The unique formulation of the po extraction conditions and is not a polymer is a free-flowing white g ASE Prep DE and used for the m polymer can be easily removed f high-temperature extractions are The polymer overcomes the limi removal and extraction. Investigations are underway to e moisture/water containing sampl References 1. Burford, M. D., Hawthorne, S. B. line supercritical fluid extraction, 2. Determination of Acephate and M Polymer, Analytical Communicat Thermo Fisher Scientific Inc. All rights reserved subsidiaries. This information is not intended to encourage use property rights of others. PO70546_E 03/12S Thermo Scientific Poster Note • PN70546_E 03/13S 5 e removal and in-vial moisture removal sing the polymer was studied using a mixture of alyte. An in-cell extraction was pursued in the pike with a standard solution that contained 1.5 d 8.5 mL of water. The spiked polymer sample SE 350 instrument and a 1:1 ratio of n extraction temperature of 100⁰ C. The extract en stream at 40⁰C. The spike level was 30 an HPLC instrument The results from the 2. The per cent recovery data showed r for these test analytes based on an A Method 8270. Moreover, it also indicates that his polymer for in-cell extraction for these test moval at room temperature was also studied analytes in a 10 mL solvent. A known amount of L solvent. The water was removed either by the lts are shown in Table 2. Comparable removal by the polymer compared to sodium ike In-cell moisture removal by the polymer In-vial moisture removal by the polymer In-vial moisture removal by the sodium sulfate % recovery % recovery % recovery 81.01 82.96 86.79 89.99 88.58 88.41 86.72 95.03 93.88 96.05 90.96 88.93 92.33 87.45 87.16 92.82 89.47 88.02 94.44 88.39 87.51 93.83 92.64 87.39 97.50 93.25 89.58 97.51 95.20 91.02 very from spiked soil sample was pursued for inaction in the Dionex ASE 350 system. A 5 g Hs and then the soil was made moist (to about piked wet soil sample was mixed with 4 g of 1:1 mL ASE cell. The soil sample was then extracted by solvent (1:1 ratio of dichloromethane: of 100⁰ C. The extract was evaporated to 1 mL pike level was calculated as 20 µg/mL for the instrument with FID. The results from the above ough two compounds showed a lower recovery, on the acceptance criteria of ± 30% as per EPA yet to optimize to increase the recovery of naphthylene. Table 3. PAH recovery by using GC-FID. Analytes In-cell moisture removal by the polymer-DE In-vial moisture removal by the polymer In-vial moisture removal by the sodium sulfate % recovery % recovery % recovery 1 Naphthalene 77.8 100.8 96.2 2 Acenapththyene 93.9 99.3 97.5 3 Acenaphthylene 74.5 97.0 95.7 4 Phenanthrene 100.7 103.1 98.6 5 Anthracene 103.1 103.3 99.8 6 Fluoranthene 102.3 110.1 101.0 7 Pyrene 98.8 109.4 100.6 8 Benzo(a)anthracene 96.8 116.4 108.1 9 Chrysene 93.3 117.3 104.0 10 Benzo(b)fluoranthene 96.2 120.3 103.7 106.4 11 Benzo(k)fluoranthene 97.2 119.2 12 Benzo(a)pyrene 85.3 118.0 106.0 13 Indeno((1,2,3,c,d)pyrene 98.1 112.1 105.5 14 Dibenzo(a,h)anthracene 109.7 110.9 105.1 15 Benzo(g,h,i)perylene 105.9 112.5 107.3 Analyte recovery for in-vial (off-line) moisture removal at room temperature was studied using the same solvent, 1:1 dichloromethane: acetone. A 40 mL solvent was spiked with PAHs. A known amount of water was added to the spiked solvent acting as an extract. The water was removed either by the polymer or by the sodium sulfate. The extract was evaporated to 1 mL under nitrogen stream at 40⁰C. The spike level was 20 µg/mL for the final extract volume of 1 mL. The results are shown in Table 3. Comparable recoveries were obtained for in-vial moisture removal by the polymer compared to sodium sulfate. Therefore, the usefulness of the polymer as drying agent was evident from the above table. Conclusion The utility of a new polymer for removing moisture from wet samples as well as from a collected extract is shown here. The new polymer is designed to work with accelerated solvent extraction and is intended for in-cell (in-line) moisture removal, in-vial (off line) moisture removal and a combination of both in-cell and in-vial. The unique formulation of the polymer allows moisture removal under ASE extraction conditions and is not affected by the sample ionic strength. The polymer is a free-flowing white granular material that can be easily mixed with ASE Prep DE and used for the moisture removal applications. Additionally the polymer can be easily removed from the Dionex ASE extraction cell after the high-temperature extractions are complete. The polymer overcomes the limitation of sodium sulfate for in-cell moisture removal and extraction. Investigations are underway to expand the applicability of this polymer to other moisture/water containing samples. References 1. Burford, M. D., Hawthorne, S. B., Miller, D. J. Evaluation of drying agents for offline supercritical fluid extraction, J. Chromatography A, 1993, 657, 413-427. 2. Determination of Acephate and Methamidophos in Foods Using Super-absorbent Polymer, Analytical Communications, 1997, 34, 253-256. Thermo Fisher Scientific Inc. All rights reserved. All trademarks are property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. PO70546_E 03/12S 6 Recent Advances in Sample Preparation for Accelerated Solvent Extraction www.thermoscientific.com/dionex ©2013 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Australia +61 3 9757 4486 Austria +43 1 333 50 34 0 Belgium +32 53 73 42 41 Brazil +55 11 3731 5140 China +852 2428 3282 PN70546_E 03/12S Denmark +45 70 23 62 60 France +33 1 60 92 48 00 Germany +49 6126 991 0 India +91 22 2764 2735 Italy +39 02 51 62 1267 Japan +81 6 6885 1213 Korea +82 2 3420 8600 Netherlands +31 76 579 55 55 Singapore +65 6289 1190 Sweden +46 8 473 3380 Thermo Fisher Scientific, Sunnyvale, CA USA is ISO 9001:2008 Certified. Switzerland +41 62 205 9966 Taiwan +886 2 8751 6655 UK/Ireland +44 1442 233555 USA and Canada +847 295 7500
© Copyright 2024