QuickTrace™ M-8000 Installing the Sample Probe Transfer Line Without an ENC-500 enclosure

QuickTrace™ M-8000
Installing the Sample Probe Transfer Line
Without an ENC-500 enclosure
When installing the sample probe / probe transfer line without a CETAC ENC-500, tube keeper 1 is deployed as
normal and tube keeper 2 is deployed with a minimal gap.
1
Create the service loop.
2
Attach tube keeper 1 at the service loop / drive cable sheath interface as shown in Figure 1.
3
Attach tube keeper 2 about ¼ to ½ inch (5 – 10 mm) from tube keeper 1.
4
Connect the free end of the transfer line to channel 2 of the on-board peristaltic pump as shown in Figures
3 and 4.
Notes:
• The length of the transfer line should allow the tubing to float over the top of the sample without getting
caught in the sample tubes (see figure 5).
• Ensure the service loop does not kink when the probe is in the full up position (see Figure 1).
• Ensure the service loop is long enough that there is no tension on the transfer tubing when the probe is in
the full down position (see Figure 2).
• The tube keeper placement for the service loop can be tested by manually rotating the Z-drive rotor; see
the Z-Drive Assembly Quick Installation Guide.
Tube keeper 2
5-10 mm gap
Tube keeper 1
No tension on
sample line
Service loop
Figure 1: Transfer line service loop, probe up
Figure 2: Transfer line with probe in full down position
Figure 3: Peristaltic pump connection (side view)
Figure 4: Peristaltic pump connection (front view)
No tension on
sample line
Figure 5: Transfer line view with probe position in farthest standard position
2
With an ENC-500 enclosure
When installing the sample probe / probe transfer line with a CETAC ENC-500, tube keeper 1 is deployed as
normal and tube keeper two is also deployed in the normal position.
1
Create the service loop.
2
Attach tube keeper 1 at the service loop / drive cable sheath interface as shown in Figure 6.
3
Attach tube keeper 2 about 7-9 inches (17-23 cm) from tube keeper 1, as shown in Figure 6.
4
Connect the free end of the transfer line to the bulkhead feed through with the supplied PEEK nut and
Teflon® ferrule.
Notes:
• Ensure the service loop does not kink when the probe is in the full up position (see Figure 6).
• Ensure the service loop is long enough that there is no tension on the transfer tubing when the probe is in
the full down position (see Figure 2).
• The tube keeper placement for the service loop can be tested by manually rotating the Z-drive rotor; see
the Z-Drive Assembly Quick Installation Guide.
Tube keeper 2
17-23 cm gap
To bulkhead
Tube
keeper 1
Service
loop
Figure 6: Transfer line retention with an ENC-500 enclosure, shown with
probe up
DuPont™ Teflon® is a trademark or registered trademark of E.I. du Pont de Nemours and Company.
© 2011 CETAC Technologies / Printed in USA / Document part number 610126 rev1
3