JCT JCE-2 Ex Midi Sample Gas Cooler Innovation on Gas Analysis

Innovation
on Gas Analysis
JCE-2 Ex Midi Sample Gas Cooler
JCT
Analysentechnik
GmbH
Gas Sampling
Probes
Heated Sample
Lines
Sample Gas Coolers
Gas
Converters
Accessories
APPLICATION
FEATURES
• For continuous gas conditioning in
processes with dust and humidity
• Removing water from gas sample
• Reducing temperature of gas
• Extractive gas analysis
• Emission monitoring and process
monitoring
• For ATEX Zone II category 3
• Up to 2 separate gas streams
• Continuously running compressor with
hot gas by pass cooling system
• Built-in condensate pump(s)
• Proportional temperature control
• Digital temperature display
• Full protection coating of all copper
parts
• Stable dew point
BENEFITS
• Very low dissolution rates of water
soluble gases
• Safety for analysers
• Small dead space in heat exchanger
• Long lifetime also in harsh conditions
• Rapid availability
• Low and comfortable maintenance
www.jct.at
TECHNICAL DATA
Cooling principle
Compressor with hot gas bypass technology
Number of gas paths
1-2
Operation
PVDF
Stainless steel
Sample gas flow rate Mono*
max. 250 l/h
max. 500 l/h
Sample gas flow rate Dual*
max. 2 x 125 l/h
max. 2 x 250 l/h
max. +140°C
max. +180°C
max. +70°C
max. +80°C
Heat exchanger material
Gas temperature sample gas inlet*
Dew point sample gas inlet*
Dew point sample gas outlet*
+3°C
Dew point stability
(for constant inlet conditions)
±0,3 K
Ambient temperature*
+5....+40°C
Cooler capacity
Operating pressure without condensate pump
90 W
160 W
1,5 bar abs.
100 bar abs.
Operating pressure with condensate pump
0,5 ... 1,5 bar abs.
Ready for operation
< 20 min
Pressure drop at max. flow rate Mono
20 mbar
47 mbar
Pressure drop at max. flow rate Dual (each gas path)
4,5 mbar
12 mbar
Construction
Dimensions over all (housing type S)
278 x 312 x 360 mm ( W x H x D )
Dimensions over all (housing type Q)
403 x 312 x 235 mm ( W x H x D )
Mounting position
Stand alone or wall mount
Alignment
horizontal
Weight
approx. 19 kg
Housing, colour
Sheet steel 1,5 mm, powder coated, RAL 7035
Dead volume Mono
67 ml
67 ml
Dead volume Dual
55 ml
55 ml
Connection sample gas/ condensate outlet
PVDF ID/OD 4/6 mm hose fitting
Connection condensate outlet on bottom
ID/OD 10/12 mm
Type of protection
Approval / Sign
ID/OD 10/12 mm
Ex II 3G Ex ma IIA T3
Ex II 3D Ex ma IIIB T180°C
(IEC / EN60079)
CE
Electrical
Supply voltage
Option
Power consumption (depending on load, ambient temperature and configuration)
230 VAC 50/60 Hz
115 VAC 50/60 Hz
Supply connection / connection status contact
approx. 2,3 m open wire ends
Fusing
approx. 190 VA
External on installation site, fuse characteristic C:
230 VAC 6A; 115 VAC 10A
Protection class
IP 20 (EN 60529)
On - time
100 %
Alarm set points
<0 / >+10°C
Status / alarm relay
Volt free change over contact
Switching capacity relays
max. 230 VAC/2A; min. 5 VADC 5mA
*...may be affected by other values
Page 2
GAS FLOW DIAGRAM
A
Gas heat exchanger
B
Condensate pump
ORDERCODE
Housing
Heat exchanger
Condensate pump
Power supply
Options
for rear panel mounting
S
for side panel mounting
Q
1 Mono PVDF heat exchanger
1
1 Mono SS heat exchanger
2
1 Mono glass heat exchanger
3
1 DUAL PVDF heat exchanger
4
1 DUAL SS heat exchanger
5
Without condensate pump JSR25
0
1 condensate pump JSR25
1
2 condensate pumps JSR25
2
230 VAC 50 / 60 Hz
1
115 VAC 50 / 60 Hz
3
Standard version
X
Special version (on request)
1
Order Code:
JCE-2.
Scope of delivery: device, operating manual
Page 3
DIMENSIONS
All dimensions in mm
Specifications are subject to change without notice.
PDS_E_JCE-2_v1.1