2.007 CLEAN CBEF IU MICRO-G AND SAMPLE TRAY

2.007 CLEAN CBEF IU MICRO-G AND SAMPLE TRAY
(CBEF/18S-ALL/FIN/Payload/HC)
Page 1 of 6 pages
OBJECTIVE:
Clean up inside of Cell Biology Experiment Facility (CBEF) Incubator Unit (IU) Micro-G.
LOCATION:
Saibo Rack (JPM1A2)
DURATION:
25 minutes
10 minutes for preparation
15 minutes for Step 1 through 5
CREW:
One
PARTS:
BZK Wipes P/N BCCK003
MATERIALS:
Medium Glove or Large Glove P/N LSE-GV001 or LSE-GV002
1.
PULLING OUT CBEF IU SAMPLE TRAY
NOTE
To maintain cleanliness, gloves must be worn when working
inside CBEF Incubator Unit.
1.1
Don gloves.
Figure 1. CBEF IU Handle Stopper
11 MAR 09
PGCF1044.xml
2.007 CLEAN CBEF IU MICRO-G AND SAMPLE TRAY
(CBEF/18S-ALL/FIN/Payload/HC)
CBEF IU
Micro-G
1.2
Page 2 of 6 pages
Unwrap Velcro strap from CBEF IU Micro-G Door Handle.
Refer to Figure 1.
1.3
Slowly open CBEF IU Micro-G door.
Engage CBEF IU Micro-G Door Handle to CBEF IU Door
Handle Stay.
Figure 2. GRIP POINT of CBEF IU Micro-G Sample Tray
1.4
Grab GRIP POINT.
Refer to Figure 2.
NOTE
It is required to pull out CBEF IU Micro-G to hard stop. If CBEF
IU Micro-G is latched halfway before hard stop, it will make it
difficult to release side lock later.
1.5
2.
11 MAR 09
Pull out until hard stop of CBEF IU Micro-G Sample Tray.
CLEANING UP CBEF IU MICRO-G SAMPLE TRAY
PGCF1044.xml
2.007 CLEAN CBEF IU MICRO-G AND SAMPLE TRAY
(CBEF/18S-ALL/FIN/Payload/HC)
Page 3 of 6 pages
Figure 3. Switch for CAN-1 to CAN-6, left and right wall, ceiling and floor of CBEF IU Micro-G
2.1
Safing
✓sw CAN-1 – OFF
✓sw CAN-2 – OFF
✓sw CAN-3 – OFF
✓sw CAN-4 – OFF
✓sw CAN-5 – OFF
11 MAR 09
PGCF1044.xml
2.007 CLEAN CBEF IU MICRO-G AND SAMPLE TRAY
(CBEF/18S-ALL/FIN/Payload/HC)
Page 4 of 6 pages
✓sw CAN-6 – OFF
Refer to Figure 3.
2.2
Wipe CBEF IU Micro-G Sample Tray except connectors (6) and
CBEF IU Temperature/Humidity Sensor with BZK Wipes.
Refer to Figure 3.
2.3
Wipe GRIP POINT with BZK Wipes.
Refer to Figure 3.
NOTE
CBEF IU Micro-G String Fan connector and cable are
damageable.
2.4
Wipe right wall except CBEF IU Micro-G String Fan cables, and
Temperature/Humidity Sensor with BZK Wipes.
Refer to Figure 3.
2.5
Wipe bottom floor with BZK Wipes.
Refer to Figure 3.
2.6
Wipe left wall with BZK Wipes except cable.
Refer to Figure 3.
2.7
Wipe ceiling with BZK Wipes.
Refer to Figure 3.
3.
11 MAR 09
STOWING CBEF IU SAMPLE TRAY
PGCF1044.xml
2.007 CLEAN CBEF IU MICRO-G AND SAMPLE TRAY
(CBEF/18S-ALL/FIN/Payload/HC)
Page 5 of 6 pages
Figure 4. Slide locks of CBEF IU Micro-G Sample Tray
3.1
Wipe slide locks with BZK Wipes.
Refer to Figure 4.
NOTE
Slide locks are located in the middle of slide rail of CBEF IU
Micro-G Sample Tray (Locations are above CAN-4 marking
and below CAN-6 marking). It is required to release slide lock
simultaneously by pushing in. Refer to Figure 4.
3.2
11 MAR 09
Release CBEF IU Micro-G Sample Tray slide locks (one each
top and bottom) by simultaneously pressing slide locks and
slide in a little until unlatched.
Refer to Figure 4.
***********************************************************************
PGCF1044.xml
2.007 CLEAN CBEF IU MICRO-G AND SAMPLE TRAY
(CBEF/18S-ALL/FIN/Payload/HC)
Page 6 of 6 pages
*
If sample tray is latched in the middle of the way,
*
*
*
Fully pull out until hard stop, then try again.
*
*
***********************************************************************
4.
3.3
Grab GRIP POINT.
3.4
Slide in CBEF IU Micro-G Sample Tray.
WIPING AND CLOSING CBEF IU MICRO-G DOOR
Figure 5. Rear surface of CBEF IU Micro-G door
4.1
Wipe rear surface of CBEF IU Micro-G door except for cable
and ventilation valve with BZK Wipes.
Refer to Figure 5.
4.2
Close CBEF IU Micro-G door.
Securely wrap Velcro strap around the CBEF IU Micro-G Door
Handle.
4.3
5.
Doff gloves.
CLOSE OUT
5.1
Update IMS or notify SSIPC to update IMS with stowage
location.
5.2
Trash following items.
Trash:
BZK Wipes
Medium Glove or Large Glove
11 MAR 09