AA3 - 12 Month Maintenance Checklist Changing XY-2/3 Sample Probe and Sample Line NOTE: The XY2 Sampler Probe is p/n: 168+B532-03. The XY3 Sample Probe is p/n: 168+B582-03. They are NOT interchangable!! Remove Probe and Sample Line - Disconnect sample line from probe and connector/splitter - Unscrew large white nylon nut at the top of the probe - Remove probe and sample line Prepare New Probe - Sleeve probe to sample line (p/n: 562-2015-01) with a short piece of tygon (p/n: 116-0536-09) Install New Probe - Remove large white nylon nut from new probe - Insert Probe into holder - Slip nylon nut over probe before fully putting probe in place - Direct sample line to go to the left of the sample arm - Carefully center probe in lower guide - Tighten the nylon nut Route New Sample Line - Starting at the probe, route the new sample line to the left and then down through the zip tie - Leaving a small amount of slack, insert the sample line through the 2 clips on the back of the arm - Route the sample line across the top of the sampler arm or around back of the sampler to a clip placed on the side of the sampler - Keep sample line as short as possible by keeping the sampler as close to the pump as possible - Turn on the sampler and allow it to go through initialization - Adjust length for slack as needed and cut sample line - Attach to sample pump tube using connector/splitter CSM-AA3YR rev 2 5/11 SEAL Analytical - Mequon Technology Center 10520-C Baehr Road Mequon, WI 53092 262-241-7900 www.Seal-us.com Copyright © 2011 SEAL Analytical Reproduction or distribution of this document is expressly forbidden without written permission from SEAL Analytical.