Instruction Manual INSUFFLATOR Highflow 45 2235 GA-A 274

Instruction Manual
INSUFFLATOR Highflow 45
2235
GA-A 274 / en /
2012-12 V1.0 / PDG 00-0000
Important general notes and instructions
Make sure that this product is used only as intended and described in this instruction manual, by adequately trained
and qualified medical personnel, and that maintenance and repair are only carried out by authorized experts.
Use the product only in the combinations and with the accessories and spare parts specified in this instruction manual.
Use other combinations, accessories and replacement parts only if they are expressly intended for the planned applic­
ation and if the performance characteristics and safety requirements are not impaired. Do not alter the product in any
way.
Reprocess the products before every application and before returning them for repairs as required by the instruction
manual in order to protect the patient, user and others.
This manual is an integral part of the product and must be stored in such a way that it is accessible at any time during
its entire life cycle. This manual must be passed on to any subsequent owner.
Immediately upon receipt, check the product and its accessories for completeness and possible damage. Should the
shipment give right to complaints, please inform the manufacturer or supplier immediately.
Subject to technical changes!
Due to ongoing developments, the illustrations and technical data may deviate slightly.
Safety instructions and levels of danger
Symbols
Level of danger
WARNING!
Failure to observe can result in death or serious injury.
CAUTION!
Failure to observe can result in slight injury or damage to the product.
.
IMPORTANT!
.
NOTE!
Failure to observe can result in damage to the product or surroundings.
Tips for optimum use and other useful information.
GERMANY
RICHARD WOLF GmbH
75438 Knittlingen
Pforzheimerstr. 32
Telephone: +49 70 43 35-0
Telefax:
+49 70 43 35-300
MANUFACTURER
[email protected]
www.richard-wolf.com
BELGIUM / NETHERLANDS
N.V. Endoscopie
RICHARD WOLF Belgium S.A.
Industriezone Drongen
Landegemstraat 6
9031 Gent Drongen
Telephone: +32 92 80 81 00
Telefax:
+32 92 82 92 16
[email protected]
www.richard-wolf.be
Marketing Office
U.A.E
RICHARD WOLF Middle East
P.O. Box 500283
AL Thuraya Tower 1
9th Floor,
Room 904, Dubai
Telephone: + 9 71 43 68 19 20
Telefax:
+ 9 71 43 68 61 12
[email protected]
www.richard-wolf.com
GA-A 274
USA
RICHARD WOLF
Medical Instruments Corporation
353 Corporate Woods Parkway
Vernon Hills, Illinois 60061
Toll Free: 001 (800) 323 - 9653
Phone: 001 (847) 913 - 1113
Fax:
001 (847) 913 - 1488
[email protected]
www.richardwolfusa.com
FRANCE
RICHARD WOLF France S.A.R.L.
Rue Daniel Berger
Z.A.C. La Neuvillette
F-51100 Reims
Telephone: +33 3 26 87 02 89
Telefax:
+33 3 26 87 60 33
UK
RICHARD WOLF UK Ltd.
Waterside Way
Wimbledon
SW17 0HB
Telephone: + 44 20 89 44 74 47
Telefax:
+ 44 20 89 44 13 11
[email protected]
www.richardwolf.uk.com
[email protected]
[email protected]
www.richard-wolf.com
AUSTRIA
RICHARD WOLF Austria
Ges.m.b.H.
Wilhelminenstraße 93 a
A-1160 Vienna
Telephone: +43 14 05 51 51
Telefax:
+43 14 05 51 51 45
INDIA
RICHARD WOLF India Private Ltd.
JMD Pacific Square
No. 211 A, Second Floor
Behind 32nd Milestone
Gurgaon - 122 001
National Capitol Region
Telephone: + 91 12 44 31 57 00
Telefax:
+ 91 12 44 31 57 05
[email protected]
www.richard-wolf.com
0
Contents
1
1.1
1.2
1.3
1.4
1.4.1
1.4.2
1.5
1.5.1
1.5.2
1.5.3
1.6
1.7
General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Indications and field of use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Contraindications and side effects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Contraindications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Side effects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Combinations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Instruments for INSUFFLATOR Highflow 45 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Potential equalization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Requirements for products / components of a combination . . . . . . . . . . . . . . . . . . . . . . . 7
Electromagnetic compatibility (EMC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Connection diagram - in dialog mode (option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2
2.1
2.2
Illustration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Front view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Rear view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3
3.1
3.2
3.2.1
3.2.2
3.3
3.3.1
3.3.2
3.4
3.4.1
3.4.2
3.4.3
Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Language settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CO2 connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Preparation for CO2 operation with gas bottle/cylinder . . . . . . . . . . . . . . . . . . . . . . . . . . .
Preparation for CO2 operation using centralized gas supply . . . . . . . . . . . . . . . . . . . . . .
Connecting the insufflation tube set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Reprocessable insufflation tubes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Disposable insufflations tube . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Highflow insufflation tubes with heating wire . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Reprocessable highflow insufflation tube with heating wire . . . . . . . . . . . . . . . . . . . . . . .
Disposable highflow insufflation tube with heating wire . . . . . . . . . . . . . . . . . . . . . . . . . . .
Disposable tube set for smoke gas evacuation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
13
14
14
14
15
15
16
16
17
17
18
18
4
4.1
4.2
4.2.1
4.2.2
4.2.3
4.2.4
Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Visual check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Function check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Device function after switching on the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Checking the gas supply (status message) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Manual checks of the control function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Checking the pressure and flow control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
19
19
19
19
20
20
20
5
5.1
5.2
5.2.1
5.2.2
5.2.3
5.2.4
5.2.5
5.2.6
5.3
Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Operating principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Controls and modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Operation and display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Dispaly of remaining CO2 gas supply in the status display . . . . . . . . . . . . . . . . . . . . . . .
Gas consumption in status display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Indicator for gas heating (only 2235021, 2235031 / 2235621, 2235631) . . . . . . . . . . . .
Footswitch for smoke gas evacuation (option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Video on-screen display with dialog function (option) . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Control sequence and menu structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
22
22
22
22
22
22
23
23
23
24
GA-A 274
I
5.3.1
5.3.2
5.4
5.5
5.6
5.7
5.8
5.8.1
5.9
5.10
5.10.1
5.11
5.11.1
5.11.2
5.12
5.12.1
5.12.2
5.12.3
5.12.4
5.12.5
5.12.6
5.12.7
5.12.8
5.13
5.13.1
5.13.2
5.13.3
5.13.4
5.13.5
5.13.6
5.13.7
Overview - control sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Overview - menu structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Menu levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Menu control and touchscreen functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Selection of main level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Selection of device modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Selection of Richard Wolf profiles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Changing the standard profiles (default profiles) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Profiles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Device parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Adjusting the device parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Display of actual and nominal values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
General service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Operation of INSUFFLATOR Highflow 45 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Important note for using the reprocessable highflow insufflation tube with heating wire
Notes and instruction for using smoke gas evacuation . . . . . . . . . . . . . . . . . . . . . . . . . . .
Creating a pneumoperitoneum . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Highflow device mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Measures after carrying out insufflation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Possible alarms during the application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Power failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Taking out of service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Overview of operator messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Structure of operator messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Message types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Functionality of control elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Operator's messages (green) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
LOW error (turqoise) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
MEDIUM error (yellow) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
HIGH error (red) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
24
25
26
26
27
28
28
28
29
31
31
32
33
33
34
34
35
35
35
36
36
36
36
37
37
37
37
38
38
40
40
6
6.1
6.2
6.3
6.4
Operation within the core system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Functional principle within the core system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Combining and controlling the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Setting the system ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Connection to the core system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
41
41
41
41
42
7
7.1
7.2
7.2.1
7.2.2
7.3
7.4
7.4.1
7.5
7.5.1
7.5.2
Reprocessing and maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Reprocessing of device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Reprocessing of reusable insufflation tubes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Sterilization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Reprocessing of reusable highflow insufflation tube with heating wire . . . . . . . . . . . . . .
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Maintenance intervals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Quarterly check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Measuring and test equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Checking the pressure and flow control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
43
43
43
44
44
44
44
44
44
44
45
II
GA-A 274
Contents
7.5.3
Leakage test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
8
8.1
8.1.1
8.2
8.2.1
8.2.2
8.2.3
8.2.4
8.3
8.4
8.5
8.5.1
8.5.2
Technical description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Fault in the core system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Application parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Pressure and flow values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Optional device variants . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Operating, storage, transport and shipping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Spare parts and accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Replacing parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Device fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Disposal of product, packaging material and accessories . . . . . . . . . . . . . . . . . . . . . . . .
GA-A 274
46
46
46
47
47
48
48
49
49
50
51
51
51
III
1
General information
1.1
Symbols
Symbols
Designation
Attention, Caution
Follow the instruction manual
OFF (no power supply, separation from mains)
ON (connection to the mains / line voltage)
Potential equalization
Power / Mains fuse
μ
Alternating current (AC)
TYPE CF APPLIED PART
Hygiene filter
Gas heating
Alarm
Signal sound
Signal sound switched off
Footswitch
Input connector for smoke gas evacuation / data input
Output connector for insufflation / data output
Recycle the product separately. Do not discard together with other waste.
REF
Order number
Lot designation
Manufacturer
Manufacturing date
Quantity
Use-by date
1
GA-A 274
Symbols
Designation
Do not reuse.
Latex-free
Contains phthalates or phthalates are present: Diethylhexyl phthalate (DEHP)
-25_C
Maximum temperature range for shipment, transport and storage
+ 50_C
Sterilized with ethylen oxide
Steam sterilization - fractional method
XX°
XX°
Permissible temperature range
XXhPa
Permissible atmospheric pressure range
XXhPa
XX%
Permissible humidity range
XX%
Do not use mobile phone / radio intercom
A Registered Trademark of a Recognized Testing Laboratory, confirm the compliance to the standard
of Medical Electrical Equipment CAN/CSA C22.2 No.60601-1 (c) and ANSI/AAMI ES60601-1 (us)
Identification in conformity with medical devices directive 93/42/EEC only valid if the product and/or
packaging is marked with this symbol. Products of category IIa and above, as well as sterile products
or products with measuring function of category I, are additionally identified with the code no. of the
notified body (0124).
1.2
Intended use
The INSUFFLATOR Highflow 45 (2235) serves to create and maintain a
pneumoperitoneum using CO2 gas. It can be used for both diagnostic and
surgical laparoscopy.
In conjunction with integrated smoke gas evacuation (2235011/2235031)
and the "Continuous Flow" (2235104) software module, the system also
serves to distend the colon (rectal cavity and lower portion of the sigmoid)
in a pressure-controlled manner for the purpose of Transanal Endoscopic
Microsurgery (TEM).
This product is designed exclusively for use by specialized medical per­
sonnel and may only be applied by medically qualified and adequately
trained doctors.
GA-A 274
2
1.3
Indications and field of use
The INSUFFLATOR Highflow 45 (2235) is primarily used in diagnostic
and surgical laparoscopy but can also be used for TEM using the integ­
rated smoke gas evacuation (2235011/2235031) and the "Continuous
Flow" (2235104) software module. For TEM applications please follow
the latest instruction manual for the TEM system.
On the basis of the patient’s general condition, the user in charge must
decide whether the planned use is possible or not. For further notes and
instructions please refer to the latest medical literature.
CAUTION!
In case of therapeutic applications an adequate backup device with the
same capabilities must be available should the device fail.
CAUTION!
It is not allowed to use mobile phones or similar devices while the INSUF­
FLATOR Highflow 45 is used.
.
NOTE!
We recommend reading relevant literature for the planned application.
1.4
Contraindications and side effects
1.4.1
Contraindications
The use of this device is contraindicated whenever laparoscopy is con­
traindicated.
WARNING!
The device has a highflow rate.
The device is not suitable for hysteroscopy.
This device must not be used for distending the cavum uteri.
Contraindications resulting from the patient's general condition and de­
scribed in the relevant literature must be observed.
This device is contraindicated in the case of hysteroscopic insufflation do not use this device for intrauterine distension.
Absolute contraindications:
D Hypovolemic shock represents an absolute contraindication with the
exception of laparoscopic surgeons who have adequate experience in
procedures for blood extraction from the abdomen and for finding and
stopping a hemorrhage site and can execute them safely and so can
achieve hemostasis quicker than by laparotomy.
D In the case of severe inflation of the intestine during intestinal obstruc­
tion, increased possibility of adhesions in the pelvis and in the abdomen
as well as in the case of an elevated diaphragm, the required insuf­
flation is contraindicated and the space available for a visualization is
severely limited and the probability of injuring the intestine is severely
increased.
D A further contraindication is the existence of a large mass in the pelvis
or in the area of the pelvis and abdomen (which in general extends to
the umbilicus or above). There is an increased risk of injuring this mass,
and there is a high probability that endoscopic visualization is poor and
that the extension and the nature of the mass cannot be determined. A
mass of this size would probably have to be removed by a laparotomy.
3
GA-A 274
D A type 4 cardiac decompensation also represents a contraindication as
the insufflation and its positioning could cause irreversible arrhythmia to
the heart which is filled with congested blood and cause cardiac failure.
Relative contraindications:
D Septic peritonitis
D Pelvic abscess
D Multiple interventions in the abdomen
D Diaphragmatic hernia
D Obesity
D Pregnancy
TEM is contraindicated where a general contraindication against surgical
interventions, a severe infection or severe metabolic disorders exist.
On the basis of the patient's general condition the doctor in charge must
decide whether the planned use is possible or not. For further notes and
instructions please refer to the latest medical literature.
GA-A 274
4
1.4.2
Side effects
Warnings:
Metabolic acidosis and the resulting cardiac arrhythmia. Intraabdominal
pressures of more than 20 mmHg (mercury column) for an extended
period of time must be avoided.
This may lead to the following complications:
D Reduced breathing with impaired diaphragm excursion
D Reduced venous backflow
D Reduced cardiac output
D Acidosis
Excessive absorption of CO2 is either due to excessive flow or excessive
pressure. The abdomen can be sufficiently distended with pressures
between 15 - 20 mmHg (mercury column). Hardly ever abdominal pres­
sures of more than 20 mmHg are required. At these values only limited
absorption should occur. Pressures of more than 20 mmHg (mercury
column) are actually never necessary and will lead to increased and
faster absorption. Adequate artificial respiration will help to avoid prob­
lems with regard to CO2.
Z Idiosyncratic reactions. In patients suffering from sickle-cell disease or
pulmonary insufficience the use of these devices can lead to an in­
creased risk of a metabolic imbalance due to excessive CO2 absorp­
tion.
Z Hypothermia. Highflow rates bear the potential risk of hypothermia.
Precautions:
It is strongly recommended using a hydrophobic bacteria filter to prevent
the transmission of microorganisms between patients.
5
GA-A 274
1.5
Combinations
.
IMPORTANT!
In addition to this instruction manual, follow the manuals for the products
used in combination with this product.
WARNING!
Danger of life-threatening gas or air embolism.
Purge the connection tubes/hoses with a sufficient amount of CO2 before
use.
The use of devices using additional gaseous media in combination with
the INSUFFLATOR Highflow 45 is exclusively within the responsibility of
the user.
WARNING!
Danger of intrabdominal excess pressure when using a second gas
source.
When using argon plasma coagulators an additional visual or manual
pressure monitoring device of the user is mandatory because the
pressure monitoring function / pressure relief of the INSUFFLATOR High­
flow 45 can be inactive as a result of a kinked insufflation tube or a closed
instrument stopcock.
The preselected gas flow rates of the argon plasma coagulator should not
exceed 2 l/min. Activate the argon plasma coagulator only briefly.
1.5.1
Instruments for INSUFFLATOR Highflow 45
The INSUFFLATOR Highflow 45 can be combined with the instruments
used during a diagnostic and/or surgical laparoscopy and/or TEM.
Follow the instruction manuals of the products used in conjunction with
this product.
On principle, the INSUFFLATOR Highflow 45 can be combined with all
trocar sleeves available in the market.
GA-A 274
6
1.5.2
Potential equalization
The potential equalization cable establishes a direct connection between a medical electrical de­
vice and an equipotential bonding rail.
It serves to equalize differences in potential between enclosures of electrical equipment and firmly
installed conductive parts in the patient environment.
1.5.3
Requirements for products / components of a combination
1.5m
1.5m
The general requirements depend on whether the products /
components are inside or outside the patient environment.
1.5m
IEC 60601-1, 3. Ed.
Medically used room
inside the
patient environment
MP
MP
~
~
MP
MP
~
outside the
patient environment
-
NMP
~
~
Verification of the total patient leakage current
**
NMP
MP
MP
-
Requirements / measures
Leakage currents to section 16.6
IEC 60601-1:2005 / EN 60601-1:2006 *
**
MP
~
Non-medically
used room
-
Verification of leakage currents
**
a) additional protective earth connection
(consult the corresponding manufacturer),
-
or
NMP
b) additional isolating transformer for medical applications **
~
MP
NMP
~
MP / NMP
MP
~
**
MP
NMP
~
~
MP
MP / NMP
~
~
additional "isolating transformer" to IEC/ EN60601-1
**
Functional connection
MP
NMP
Verification of leakage currents
a) no plugs with metal housing, or
b) additional isolation device (to avoid voltage differentials)
-
~
Verification of leakage currents
a) common protective earth connection, or
b) additional protective earth connection at MP (clarify with the corre­
ponding manufacturer), or
c) additional isolation device (to avoid voltage differentials), or
d) no plugs with metal housing in the patient environment
additional isolating device to IEC/
EN 60601-1
Multiple socket strip
Power supply grid
= medical electrical device to IEC/ EN 60601-1, ANSI/AAMI es60601-1, CSA C22.2 No. 60601-1 08
= non-medical electrical device in accordance with product-specific IEC/EN/UL standards
*
**
When connected via a multiple socket strip under standard conditions the earth leakage current of the power bar must not exceed 5 mA.
e.g. Richard Wolf video cart with "isolating transformer"
7
GA-A 274
.
1.6
IMPORTANT!
The persons combining products to form a system are responsible for not impairing the system's compliance with
performance and safety requirements, and that the technical data and the intended use are adequately fulfilled.
Possible electromagnetic or other interference that may occur between the product and other products can cause
faults or malfunctions.
When selecting the system components, make sure that they meet the necessary requirements of the medical envi­
ronment they are used in, in particular IEC/ EN 60601-1 (3. Edition IEC/EN 60601-1, section 16). In case of doubt
contact the manufacturer(s) of the system components.
Do not touch connectors for electrical connections between various components (such as signal input connectors
and signal output connectors for video signals, data exchange, controls etc.) and the patient at the same time.
Electromagnetic compatibility (EMC)
Please observe the following:
The device/system in the following referred to as product always related to INSUFFLATOR Highflow 45.
The product does not have any performance features classified as essential performance features in accordance with IEC/EN 60601-1.
Guidelines and manufacturer's declaration - Electromagnetic emissions
The product is intended for use in the environment specified below. The user must assure that the product is used in such an environment.
Emissions measurement / test
Compliance
Electromagnetic environment - Guidelines
HF emissions to CISPR 11
Group 1
The product uses HF energy for its internal function.
The HF emission level is extremely low and it is not likely to cause any
interference in nearby electronic equipment.
HF emissions to CISPR 11
Class B
Harmonic emissions
to IEC 61000-3-2
Class A
In conformity with IEC 61000-3-3 “Emission of voltage fluctuations
/ flicker"
The product is suitable for use in all establishments, including domestic
establishments. This also includes establishments directly connected to
the public low voltage power supply network that supplies buildings used
for domestic purposes.
Guidelines and manufacturer's declaration - Electromagnetic immunity
The product is intended for use in the environment specified below. The user must assure that the product is used in such an environment.
Compliance
Immunity tests
IEC 60601 test level
Electromagnetic environment - Guidelines
Electrostatic discharge (ESD)
to IEC 61000-4-2
 6 KV contact discharge
 8 KV air discharge
Yes
Floors should be wood, concrete or ceramic tile.
With floors made of synthetic material, the relative
humidity of the ambient air must be at least 30%.
Electrical fast transience, bursts
to IEC 61000-4-4
 2 KV for power supply lines
 1 KV for input and output
lines
Yes
Mains/line power quality should be that of a typical
commercial or hospital environment.
Surge voltage (surges)
to IEC 61000-4-5
 1 KV line to line
voltage
 2 KV line to ground
voltage
Yes
Mains/line power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short interruptions and sup­
ply voltage variations
to IEC 61000-4-11
Voltage dip for 0.5 cycle
> 95% UT *
Voltage dip for 5 cycles
60% UT *
Voltage dip for 25 cycles
30% UT *
Voltage dip for 5 sec
> 95% UT *
Yes
Mains/line power quality should be that of a typical
commercial or hospital environment. If the user of
the product requires continued operation during
power mains/line interruptions it is recommended
that the product be powered from an uninterrupti­
ble power supply or battery.
Power frequency (50/60 Hz) magnetic
field,
to IEC 61000-4-8
3 A/m
Yes
Power frequency magnetic fields should be at le­
vels characteristic of a typical location in a com­
mercial or hospital environment.
* NOTE: UT is the line / mains voltage prior to application of the test level.
GA-A 274
8
Guidelines and manufacturer's declaration - Electromagnetic immunity for products that are not life-supporting
The product is intended for use in the environment specified below. The user must assure that the product is used in such an environment.
Immunity tests
IEC 60601 test level
Compliance
Electromagnetic environment - Guidelines
Portable and mobile RF communications equipment should
be used no closer to any part of the product, including ca­
bles, than the recommended separation distance calculated
from equation applicable to the frequency of the transmitter.
Recommended separation distance:
d = 1.2 p P
Conducted HF interference
to IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
Radiated HF interference
to IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
d = 1.2 p P for 80 MHz to 800 MHz
Yes
d = 2.3 p P for 800 MHz to 2.5 GHz
P = Nominal power output rating of the transmitter in watts
(W)
(according to the transmitter manufacturer)
d = recommended separation distance in meters (m)
Field strengths from fixed RF transmitters, as determined by
an electromagnetic site survey1, should be less than the
compliance level in each frequency range2.
Interference may occur in the vicinity of devices with the
following symbol:
REMARKS:
At 80 MHz and 800 MHz the higher frequency range applies. These guidelines may not apply in all situations. The propagation of elec­
tromagnetic waves is affected by absorption and reflexion from buildings, objects and people.
1 = The field strength of stationary transmitters (e.g. base station for radio phone, earth to earth radio stations,, amateur radio stations, radio
and television transmitters) cannot be exactly predicted in theory. To assess the EMC environment due to fixed transmitters an electromagnetic
site survey should be conducted. If the measured field strength in the location in which the product is used exceeds the applicable compliance
level above, the product should be observed to verify normal operation.
If abnormal performance is observed, additional measures may be required, such as reorienting or relocating the product.
2 = Over the frequency range between 150 kHz and 80 MHz the field strength should be below 3 V/m.
Recommended separation distances between portable and mobile HF telecommunication devices and devices which
are not life-supporting
The product is intended for use in an electromagnetic environment where HF disturbances are controlled.
The user can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile HF telecommunica­
tions equipment and the product.
Rated nominal output power of the­
transmitter (W)
Separation distance as a function of transmitter frequency (m)
150 kHz to 80 MHz
d = 1.2 p P
80 MHz to 800 MHz
d = 1.2 p P
800 MHz to 2.5 GHz
d = 2.3 p P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a nominal output power not listed in the table above, the recommended separation distance (d) in meters (m) can be
determined using the applicable equation (observe frequency). P = nominal power of the transmitter in Watts (W).
REMARKS:
At 80 MHz and 800 MHz the higher frequency range applies. These guidelines may not apply in all situations. The propagation of elec­
tromagnetic waves is affected by absorption and reflexion from buildings, objects and people.
9
GA-A 274
1.7
Connection diagram - in dialog mode (option)
1.1
1.1
1
2
Legend
Direct connection
1
1.1
2
.
CAN BUS connection cable (option)
CAN BUS termination (option)
Video cable to meet the signal requirements of the camera controller used
IMPORTANT!
The optional dialog function is only possible is only possible with enabled software module 3 of INSUF­
FLATOR Highflow 45.
Connection of the CAN BUS connection cable (1) only with interactive camera controller.
If the CAN BUS connection cable is connected, the free CAN BUS connectors of INSUFFLATOR Highflow 45 and
the camera controller must be terminated with CAN BUS terminating resistors (1.1).
GA-A 274
10
2
Illustration
2.1
Front view
6
5
4
3
2
1.1
1
Legend
1
Power ON/OFF switch
1.1
Power ON/OFF LED
2
Touchscreen
3
Connector for disposable highflow insufflation tube with heating wire
' only 2235021, 2235031 / 2235621, 2235631
4
Connector for reusable highflow insufflation tube with heating wire
' only 2235021, 2235031 / 2235621, 2235631
5
Connector for insufflation tube
(type CF applied part)
6
Connector for smoke gas evacuation
(type CF applied part)
' only 2235011, 2235031 / 2235611, 2235631
11
GA-A 274
2.2
Rear view
13
14
12
15
11
10
8
7
9
16
FEDERAL LAW RESTRICTS THIS UNIT TO BE
USED OR SOLD, EXCEPT UNDER THE SUPERVISION OF A MEDICAL DOCTOR.
Legend
7
Potential equalization connector
8
Connector for footswitch for smoke gas evacuation
' only 2235011, 2235031 / 2235611, 2235631
9
Service interface
10
Silencer
' only 2235011, 2235031 / 2235611, 2235631
11
Power input connector with fuse holder
12
Fuse plate
13
Device connector for gas supply
14
CAN BUS interface
15
Mounting points for bottle holding bracket
16
Identification plate
GA-A 274
12
3
Setup
WARNING!
The device is not protected against explosions.
Explosion hazard.
Do not operate this device in areas where there is a danger of explosion.
WARNING!
Danger if a power supply without protective earth is used.
Danger of electric shock!
Connect the device only to a power supply with protective earth connec­
tor.
.
NOTE!
The line / mains voltage must correspond with the voltage indicated on
the identification plate. Connect the device only via the supplied power
cable or a power cable with the same specifications.
Do not block any ventilation slots!
CAUTION!
Danger of faults and malfunctions.
To guarantee the safety of the user, patient and others, use only acces­
sories and spare parts as specified by the manufacturer of this product.
Other accessories or spare parts can cause the emission of increased
electromagnetic radiation or reduced immunity against interference.
.
IMPORTANT!
Medical electrical equipment is subject to special precautions with regard
to electromagnetic compatibility (EMC).
Make sure you follow the notes on EMC for installation and operation.
Medical electrical devices can be influenced by mobile HF communication
devices.
If it is necessary to stack the devices or place them next to each other
and HF interference is observed, make sure that the devices are ope­
rated as specified in the intended use.
WARNING!
Use the device only with the insufflation tubes and filters specified by the
manufacturer. Sterilize the insufflation tube before each use.
Danger of infection due to unsterile accessories and possible malfunction
(loss of gas pressure or gas flow) when using spare parts supplied by
other manufacturers or in the case of wrong assembly.
CAUTION!
Disposable items have been designed for one single use only and are
sterilely packed. Do not use disposable items if:
' the sterile packaging is damaged
' the use-by-date has been exceeded
13
.
NOTE!
For use, place the device on a level surface.
.
NOTE!
Remove the protective foil from the touchscreen before the first use.
GA-A 274
3.1
Language settings
Fig. 1
The language is selected in the Device parameters menu (see section 5).
When switching on the device for the first time (setup) as well as during a
new start after having reset the device to "factory settings", the required
language must be selected.
' Language selection for the “Standard profile”.
' For "User-specific operator profile" it is possible to define a separate
language.
If a wrong language was selected and it is impossible to select the
language selection menu in this language, proceed as follows:
Z Switch off the device and wait for at least 60 seconds.
Z Switch on the device and press the device name on the touchscreen
until the language selection menu is displayed.
Z Let go of the button and select the required language.
' Language selection for the “Standard profile”.
3.2
CO2 connection
3.2.1
Preparation for CO2 operation with gas bottle/cylinder
WARNING!
Make sure that no liquid CO2 enters the INSUFFLATOR Highflow 45 as a
result of incorrect positioning of the gas bottle/cylinder!
This will endanger the patient as a result of possible malfunctions or fail­
ure of the INSUFFLATOR Highflow 45 during the surgical intervention.
Operate the gas bottle/cylinder only in vertical position with the valve on
top.
.
18
Z Use mounting screws (15) to attach the bottle holding bracket (18) to
the device if required
' Optionally, large gas bottles/cylinders can be fastened to a device
cart.
15
.
GA-A 274
IMPORTANT!
Observe the manual for the pressure reducing valve.
IMPORTANT!
When using gas bottles with more than 2 l of volume, use a separate
holding bracket (do not attach to the INSUFFLATOR Highflow 45) and
secure the gas bottle/cylinder against falling over.
14
Z Place the gas bottle (max. 2 l of volume) into the bottle holding bracket
and secure with the belt (18.1).
18.1
20.1
20
19.1
Z Connect the pressure reducing valve (20) to the gas bottle (19) and
tighten the union nut (20.1) with your hand.
19
Z Connect the CO2 pressure tube (21) to the pressure reducing valve
(20) and tighten the union nut (21.1) with your hand.
21.1
Z Open the bottle valve (19.1) by turning counterclockwise.
21
3.2.2
Preparation for CO2 operation using centralized gas supply
Connection to a centralized gas supply
17.1
Z Connect the CO2 pressure tube (17) to the device connector for gas
supply and hand-tighten the union nut (17.1).
' Then connect the CO2 pressure tube (17) to the wall connector for
centralized gas supply.
17
3.3
Connecting the insufflation tube set
CAUTION!
Contamination of the device by a backflow of gas or body fluid.
There must always be a hygiene filter between the device connector and
the insufflation tube.
.
15
IMPORTANT!
To connect and disconnect the insufflation tube always hold it by the con­
nectors.
GA-A 274
3.3.1
Reprocessable insufflation tubes
D 8170.101: for Veress cannula/RIWO-ART trocar sleeve, dia. 10 mm
(B1) + further models with luer lock fitting
D 8170.232: for high-flow trocar sleeves, dia. 10 mm (B1, B2)
Cut off the luer lock fitting on the insufflation tube and
connect the end of the tube onto the connecting piece (x).
B2
(X)
B1
B1
B2 B1
3.3.2
B1
B1
B2
B2
Disposable insufflations tube
CAUTION!
The disposable insufflation tube is designed for one single use and is sterilely
packed.
Do not use single-use insufflation tubes if:
- the sterile packaging is damaged
- the use-by date has been exceeded
D 4170.501: for Veress cannula/RIWO-ART trocar sleeve, dia. 10 mm
and high-flow trocar sleeve, dia. 10 mm (B) + others with luer lock
connector
Cut off the luer lock fitting on the insufflation tube and connect
the end of the tube onto the connecting piece (x).
(X)
B
GA-A 274
B
B
16
3.4
Highflow insufflation tubes with heating wire
Optionally, a reprocessable high-flow insufflation tube with heating wire
as well as a disposable insufflaction tube with heating wire and integrated
hygiene filter can be connected to the device.
The plugs are encoded accordingly and can only be connected to the re­
spective socket provided on the device.
For gas heating mode see section 5.
3.4.1
Reprocessable highflow insufflation tube with heating wire
.
IMPORTANT!
After autoclaving, the reprocessable highflow insufflation tube with heat­
ing wire must be allowed to cool down for at least 30 minutes. Otherwise,
an overtemperature alarm is triggered after connecting the high-flow in­
sufflation tube to the device.
The reprocessable highflow insufflation tube with heating wire is encoded
and only allows a limited number of applications.
The Device parameters - Device info menu allows the display of the num­
ber of applications already carried out as well as the maximum permiss­
ible number of applications (counts up, e.g. 100 of 100 = end of service
life).
In order to check the number of applications already carried out, the in­
sufflation tube can plugged on briefly without causing the service life to
end.
For manual control, the reusable/reprocessable insufflation tube is fitted
with tear-off tabs. If there is no tear-off tab left on the tube, the reprocess­
able insufflation tube must not be sterilized again.
17
GA-A 274
3.4.2
Disposable highflow insufflation tube with heating wire
The disposable highflow insufflation tube with heating wire is encoded
and only allows one single use.
Go to the Device parameters - Device info menu to display whether the
insufflation tube can still be used (counts up, 1 of 1 = end of service life).
3.4.3
Disposable tube set for smoke gas evacuation
.
GA-A 274
IMPORTANT!
The disposable tube set for smoke gas evacuation must only be used for
evacuating smoke gas and not for suctioning out liquids.
18
4
Checks
.
4.1
IMPORTANT!
Run through the checks before and after each use.
Do not use the products if they are damaged or incomplete or have loose
parts.
Return damaged products together with any loose parts for repair.
Do not attempt to do any repairs yourself.
Visual check
Z Check the device and accessories for damage, loose or missing parts,
hygiene and completeness.
Z Check all connection cables, tubes and hoses for damage.
Z Any inscriptions, lettering or labeling necessary for the safe intended
use must be legible.
' Missing inscriptions, lettering or labeling leading to wrong handling or
reprocessing must be reinstated.
4.2
Function check
.
IMPORTANT!
Before you perform the function checks, make sure that the devices are
in perfect technical condition and are set up correctly. This must also be
guaranteed within the scope of the visual check.
.
IMPORTANT!
Before each use, check the INSUFFLATOR Highflow 45 manually for
proper control functions.
This device is equipped with automatic function check routines which
automatically detect and report various types of faults.
The device detects faults and errors that occur during the self test or
during operation.
4.2.1
Device function after switching on the device
Z Switch on the device with the power switch.
' The LED on the power switch lights up.
' When the self test starts also listen to a signal being sounded for
checking the speaker function.
' If during the selftest the gas supply is insufficient, a device message
is displayed and the device is not ready for operation.
' After a successful self test the device is ready for operation.
' The mode selection is displayed.
' Select a suitable profile that suits the mode (see section 5.6).
19
GA-A 274
4.2.2
Checking the gas supply (status message)
Z The CO2 gas supply is displayed in the status display.
' Check that the gas bottle is open and there is sufficient CO2 supply.
' For displaying the fill levels on the status display see section 5.2.2.
4.2.3
Manual checks of the control function
Veress cannula (if used):
Z Check the Veress cannula for patency (free passage) and function
(snap-in mechanism).
Pressure preselection:
Z Change the nominal pressure using the pressure (+) and (-) buttons.
' When insufflation is on, the preselected pressure (nominal value) is
displayed for approx. 2 s.
Flow control (checking the flow):
Z Use the flow (+) and (-) buttons to change the nominal flow.
' The preselected flow rate (nominal value) is displayed for approx. 2 s
when insufflation is switched on.
' In baby mode (option) high flow cannot be selected.
4.2.4
Checking the pressure and flow control
Primary insufflation:
Z Preselect the following nominal pressure: 12 mmHg (mercury column)
Z Connect the hygiene filter and insufflation tube to the insufflation con­
nector.
Z Start insufflation.
' The current flow rate must agree with the preselected flow rate of 1
l/min.
' Permissible tolerance: ¦ 0.5 l/min
Z Close the insufflation tube with your thumb.
' As soon as the current pressure exceeds the preselected pressure,
the device must release pressure.
Z Switch off insufflation and open the insufflation tube.
GA-A 274
20
Highflow:
Z Nominal pressure to be preselected: 12 mmHg
Flow rate to be preselected: 10 l/min
Z Select highflow.
' First switch on insufflation, then select highflow.
' The current flow rate must agree with the preselected flow rate of 10
l/min.
' Permissible tolerance: ¦ 2.0 l/min
Z Close the insufflation tube with your thumb.
' As soon as the current pressure exceeds the preselected pressure,
the device must release pressure.
Z Open the insufflation tube.
Z Switch back from highflow to primary insufflation.
' After 1.5 s at the latest, a flow rate of 1 l/min ¦0.5 l/min must be dis­
played.
Z Switch off insufflation.
21
GA-A 274
5
Use
5.1
Operating principle
A pneumoperitoneum is established by insufflating CO2 gas into the ab­
dominal cavity. The gas is supplied to the INSUFFLATOR Highflow 45
either from CO2 bottles/cylinders or from a centralized gas supply. The
gas pressure and gas flow rate are controlled by INSUFFLATOR High­
flow 45 in accordance with the specifications of the user.
Microprocessor control and display of the pressure and flow values as
well as several independent safety functions guarantee safe insufflation.
To heat the gas to body temperature, we recommend one of the following
heating tubes (models 2235021, 2235031 / 2235621, 2235631).
5.2
Controls and modes
5.2.1
Operation and display
5.2.2
Dispaly of remaining CO2 gas supply in the status display
Z The remaining CO2 gas supply is displayed in the status display.
' If during the selftest the gas supply is insufficient, a device message
is displayed and the device is not ready for operation.
The device is controlled via the touchscreen and the displays and mes­
sages are also displayed on the touchscreen.
The following fill levels of the gas bottle/cylinder are displayed:
5.2.3
Display
Description
Lights up green
Sufficient gas supply
Lights up yellow
Have spare gas bottle/cylinder ready or
display when connected to centralized
gas supply
Blinks red
Change the gas bottle or in the case of
centralized gas supply: check connection
Gas consumption in status display
The gas consumption is displayed in liters on the touchscreen under
status display and deleted automatically during switch-off (> 1 min).
Reset serves to reset the consumption indicator.
GA-A 274
22
5.2.4
Indicator for gas heating (only 2235021, 2235031 / 2235621, 2235631)
Z When connecting a highflow insufflation tube with heating wire, the
CO2 gas is heated to body temperature.
The following operation modes are displayed in the status display:
5.2.5
Display
Description
Blinks green
Heating
Lights up green
Operating temperature
Blinks red
Overtemperature
Errors
Heating module deactivated
Symbol empty
No heating tube connected
Footswitch for smoke gas evacuation (option)
The footswitch serves to switch smoke gas evacuation ON / OFF.
The smoke gas evacuation pump is only active at pressures > 3 mmHg.
5.2.6
Video on-screen display with dialog function (option)
In conjunction with an interactive (dialog-capable) camera controller, the
dialog function allows the display of device parameters on the monitor.
The above requires a connection between the devices using a CAN BUS
connection cable (see connection diagram in section 1).
For the dialog function optional software module3 must be activated.
The on-screen display of the device parameters on the monitor is the
standard setting (on-screen display can be deactivated in the service
menu).
23
GA-A 274
5.3
Control sequence and menu structure
5.3.1
Overview - control sequence
Mode selection and user-specific profile generated
Switch on the device
Selftest
Mode selection
Profile selection
Main level
Menu level
Mode selection and no user-specific profile generated
Switch on the device
Selftest
GA-A 274
Mode selection
Main level
(standard settings)
Menu level
24
5.3.2
Overview - menu structure
Line/
mains power
ON
User
Selftest
OK
Not OK - error message
Mode selection
Standard profile
Profile selection (option)
User-specific profile
Main level
Main parameters
Menu level
Select
Standard profile
Selection
Generate new profile
Dialog profile name
Change profile name -
Dialog profile name
Store profile parameters
Delete profile
Device-specific menu items
Delete dialog
e.g. parameter control
Device parameters
Display
Settings
Audio
Settings
Device info
Display of device data
Language
Settings
Factory setting
Dialog resetting
Service
Authorized persons
25
General service
Parameters, resets etc.
Device service (protected)
Dialog password input
GA-A 274
5.4
Menu levels
The device menu is subdivided into 4 areas:
D
D
D
D
5.5
Main level for user
Menu level for user
General service area for user
Device service for authorized persons - with password only
Menu control and touchscreen functions
.
IMPORTANT!
When controlled via the touchscreen, it is sufficient to touch the monitor
surface slightly.
Do not touch the surface of the touchscreen monitor with sharp,
pointed or contaminated objects as this will lead to reduced image
quality or a damaged surface.
.
NOTE!
The illustrations of the user interfaces are only exemplary and vary
depending on the use.
XXXX
XXXX
Touchscreen:
The touchscreen monitor is used for information output and the input or
selection of parameters.
XXXXXXXX
XXXXXXXXX - XXXX X
XXXXX
Touchscreen areas:
Area for displays and settings:
When the displayed function is active, the displayed numerical value is
shown in red. The corresponding numerical value is adjusted using the
(+) and (-) buttons or the slide control.
Status area:
For approx. 10 s the corresponding status message appears "!" and an
acoustic signal (option) is sounded (e.g. when the lowest or highest
permissible value of the adjustment is reached).
Device parameters:
Indicator for additional parameters, e.g. gas consumption in case of an
insufflator.
Line for main functions:
Calling up the menus, switching on and off device functions, save, delete
etc.
XXX
XXXXXXX
0
100
Status area
Line for main functions
A
B
Navigation:
Navigation buttons A = up/down between individual menu items (single
steps).
Navigation buttons B = up/down by page or paragraph.
Please note:
The navigation buttons are only active if further menu items can be
called up which are not visible on the first menu page.
GA-A 274
26
Control elements:
Parameter
designation
0
Menu
Selected: Button is highlighted, letters are black (button depressed).
Menu
Not selected: Button is dark, letters are white (button not depressed).
Menu
Selectable: Letters are white.
Menu
Not selectable: Letters are gray.
20
Unit
100
Back
Display and parameter setting:
The parameters are displayed as bars and as numbers.
With the (+) and (-) buttons the value can be increased or decreased
and, optionally changed using the slide control (nominal value).
When you keep the (+) / (-) button depressed, the value changes incre­
mentally.
XXXX
Menu overview - calling up the menus:
The Menu button serves to switch from the main level to menu level.
In the menus, any changes in the settings are implemented immediately
after the input.
You may exit the menu items with the Back button.
XXXX
XXXXXXXX
X mode - Pro­
file
Back
5.6
Selection of main level
The selection on the main level depends on the device version of the
optional software modules and on the device mode. See table in section
8.2.3.
.
27
GA-A 274
5.7
Selection of device modes
Device mode
Description
Device operating mode
In a device with several operating modes, the required mode must be selected each
time after the device was started.
The listing only contains the profiles that are assigned to the mode operation.
Optionally, the device modes can be combined (see section 8.2.3).
Default
In the case of primary insufflation via a Veress cannula (1l/min).
Highflow
In the case of minilaparotomy – primary insufflation via Veress cannula is not applica­
ble.
Baby mode (option)
Special mode for newborns and babies. Precise pressure and generation of flow in
steps of 0.1 l.
TEM (option)
Continuous flow mode with a maximum flow of 8 l/min (only in conjunction with smoke
gas evacuation).
Continuous flow (option)
Continuous flow mode for laparoscopy/thoracic surgery at a maximum flow of 8 l/min.
5.8
Selection of Richard Wolf profiles
Richard Wolf profile
Description
Richard Wolf profiles:
The Richard Wolf profiles represent recommended settings for the corresponding field
of application. The listing only contains the profiles that are assigned to the mode of
operation.
Except for the standard profile (default profile), the profiles can be deleted. New selfcreated profiles can be incorporated at any time.
Default
Recommended settings for standard laparoscopy.
Gynecology
Recommended settings for laparoscopic interventions in gynecology.
Urology
Recommended settings for laparoscopic interventions in urology.
5.8.1
Changing the standard profiles (default profiles)
Profiles
Description
Standard profiles
The profiles can be selected in the Operation - Profiles menu.
After switch-off, the settings are reset to the initial values.
Richard Wolf profiles
The preconfigured Richard Wolf profiles (e.g. urology, gynecology) can be changed or
deleted as required. After a reset to the "factory setting", the stored profiles are
reinstated.
Resetting the user-specific
profiles
During a reset to the factory settings, the custom profiles are deleted and the Richard
Wolf profiles are reinstated.
GA-A 274
28
5.9
Profiles
Standard profile
1.
Dr. Bones
2.
TEP hernia
3.
New profile
Delete profile
Save profile
Back
Custom profiles:
In the Profile menu, additional custom profiles can be generated and
edited.
Please note:
In order to store changed parameters, open the user profile again (dis­
play in red letters) and confirm the settings with Save profile (see the
following overviews).
Overview - Generating/editing and deleting profiles
Select
Single step
Active
Selection possible
Up/down by page or paragraph
Not active
Selection not possible
Delete profile
Change profile name
Generate profile
Standard setting
profile
Standard
Standardsetting
profile
Standard
Standardsetting
profile
Standard
1.
New profile
1.
Dr. Bones
1.
Dr. Bones
2.
New profile
2.
TEP hernia
2.
TEP hernia
3.
New profile
3.
Save profile
Please enter profile
name:
2
1
3
F
Q
A
Y
5
C
D
X
Dr. Bones
4
E
S
6
T
F
C
Save profile
Back
Delete profile
Z
G
V
C
U
H
B
OK
Accept new profile
name
K
M
P
L
A
Ä
Y
-
E
S
C
D
X
5
T
F
C
6
Z
G
V
C
U
H
B
N
K
M
P
L
Back
Delete profile
Do you really want to delete the profile?
´
0
O
I
Y
OK
Cancel
9
8
7
New profile
Save profile
TEP hernia
4
3
F
Q
Ü
Ö
,< .
2
Back
Delete profile
Please enter profile
name:
1
´
0
O
I
Y
N
9
8
7
3.
New profile
Ü
Ö
,< .
Cancel
Accept changed profile
name
Ä
OK
Delete profile
Cancel
Cancel
delete
Select different profile
Standardsetting
profile
Standard
1.
2.
3.
Save profile
Dr. Bones
TEP hernia
New profile
Delete profile
Back
The selected profile is
accepted.
29
GA-A 274
Select
Overview - Changing the settings in a profile
Active
Selection possible
Example: Change display brightness
Not active
Selection not possible
Select
profile
Standard profile
Select
Device-specific functions
1.
Dr. Bones
Device parameters
2.
TEP hernia
3.
Dr. Muster
Save profile
Back
Delete profile
Back
Display
Select
Device-specific functions
Audio
Device info
Device parameters
Language
Dr. Bones
Dr. Bones
Back
Back
Adjust
brightness
Display
Adjust brightness
min.
80
max.
Automatic brightness control
Back
Please note:
After changing a parameter, the setting must be saved
in the "Profile" menu.
Display
Audio
A red profile name and the red "save profile" button
indicate that the changed setting has not yet been
saved.
Any settings that haven't been saved are lost when
changing between the profiles or in case of a restart.
Device info
Language
Dr. Bones
Back
Select
Standard profile
1.
Dr. Bones
Device-specific functions
2.
TEP hernia
Device parameters
3.
Dr. Muster
Dr. Bones
Save profile
Delete profile
Back
Back
Accept setting
GA-A 274
30
5.10
Device parameters
Display
Overview of device parameters:
Via the Device parameters menu, call up the list of submenus for chan­
ging the parameters.
Further parameters, if available, can be displayed using the navigation
buttons.
Audio
Device info
Language
Back
5.10.1
Adjusting the device parameters
Display
Adjust brightness
min.
80
max.
Automatic brightness control
Back
Volume
XXX
min.
50
max.
Acoustic operator's notes
Button sound
Back
Adjusting the touchscreen brightness:
In the Display submenu, the brightness or automatic brightness control
of the touchscreen can be selected.
The brightness of the touchscreen is adjusted using the (+) and (-) but­
tons or the slide control, respectively.
In the "automatic brightness control" setting manual adjustment is not
possible. In this mode, the brightness of the touchscreen is automatically
adapted to the brightness of the environment. This avoids poor legibility
(too dark) or glare (too bright) of the touchscreen.
Volume adjustment:
In the Audio submenu, the volume of the key tones, acoustic operator's
notes and alarms can be adjusted.
To adjust the volume of the audio signals, use the (+) and (-) buttons or
the slide control.
With the Key tone button or the acoustic operator's notes button acoustic
signals for the operation of the keys or operator's notes can be switched
off and on. The alarms cannot be switched off.
RICHARD WOLF GMBH
Device names:
Highflow 45
Serial number:
XXXXXXXXXX
Device number:
XXXX
MAC address:
XX-XX-XX-XX-XX-XX
Device info:
The device parameters are displayed in the Device info submenu.
Back
31
GA-A 274
British English
US English
Language selection:
The "Language" submenu displays the selected language. Here you may
select the required language.
The navigation buttons can be used to select other languages.
German
Français
Back
Selected menu language:
Select language and confirm with OK or return to language selection
using Cancel. If there is no input for 5 seconds the window will close
automatically.
XXXXX
OK
Cancel
Do you want to reset the device to factory
setting?
OK
5.11
Cancel
Factory setting:
The Factory settings submenu resets all settings to the delivery state.
Confirming the factory settings with OK resets all settings and deletes all
Profiles (custom settings) and automatically the device is restarted.
After the restart the language selection is displayed. This language
setting defines the language of the "Standard profile".
Display of actual and nominal values
Display of numerical value in red = actual value
Display of numerical value in gray = nominal value
Bar indicator = actual value
The nominal value cannot be adjusted via this slider.
When touched, the nominal value is displayed in gray as a numerical
value.
The nominal can be adjusted via this slider.
When touched, the nominal value is displayed in gray as a numerical
value.
GA-A 274
32
5.11.1
Service
.
General service
Device service (protected)
! Message text !
Back
5.11.2
NOTE!
All menu screens in the service area are exclusively in English.
General service:
Service area for the user.
Device-related display, resets and settings.
Please note:
To leave the service area the device must be switched off and then on
again.
Device service (protected):
Only for authorized persons with a password.
General service
Device info and "Device Info" resets:
Display of device-specific parameters (e.g. type/SN).
Information - Mainboard "Info WXB":
Display of software/hardware information for the mainboard.
Information - Function board "Info FB":
Display of software/hardware information for the function board.
"System Settings":
Setting the system ID for the core system (standard = 1).
Start and implementation of the touchscreen calibration.
"Logfile" protocol:
Certain events are recorded (e.g. resets, updates, faults etc.).
Activation of software modules:
Optional software modules can be activated at a later point in time.
The license code for activation can be obtained from the manufacturer.
For this purpose it is necessary to specify MAC address (Device Info) of
the device. The license code generated by the manufacturer must be
entered via the keyboard.
Calibrating the touchscreen:
Crosshairs appear consecutively in the corners of the display. Try to
press on the crosshairs as precisely as possible.
When all 4 crosshairs have been actuated, the calibration can be saved
with the Confirm Touch Calibration button.
The Restart button serves to restart the calibration.
The Cancel button stops calibration.
33
GA-A 274
5.12
Operation of INSUFFLATOR Highflow 45
CAUTION!
Danger of infection due to unsterile accessories.
Sterilize the reprocessable containers and accessories as described in
section 6.
Discard used disposable items after each use as required by the laws
and regulations valid in your country.
Follow the sterilization rules for the different products!
CAUTION!
The products have only limited strength!
Applying excessive force will lead to damage, impair the function and
endanger the patient.
Immediately before and after each use, check the products for damage,
loose parts and completeness.
Make sure that no missing parts remain in the patient.
Do not use the products if they are damaged or incomplete or have loose
parts.
5.12.1
.
IMPORTANT!
Before you start laparoscopy observe and carry out the measures de­
scribed in section 3 "Setup" and section 4 "Checks".
.
IMPORTANT!
Before each use, check the INSUFFLATOR Highflow 45 manually for
proper control functions.
Important note for using the reprocessable highflow insufflation tube with heating wire
The reprocessable highflow insufflation tube with heating wire heats the
insufflation gas to body temperature.
In the case of low gas flow rates (0.1 l/min - 1 l/min) the tube set temper­
ature on the outside may exceed 50 °C!
The actual gas flow rate depends on the overpressure generated in the
abdomen. The preset maximum nominal gas flow is only reached as long
as no or only a very slight overpressure is measured in the abdomen.
Otherwise, the gas flow is of secondary importance relative to the pres­
sure, i.e. it may be reduced to less than 1 l/min during the operation even
though higher settings were chosen.
To avoid possible injuries of the patient, user or others, make sure that
the following precautions are observed:
D Make sure that the tube set is not connected to the patient for more
than 1 minute.
D Avoid direct contact between the tube set and the patient.
D Before the operation, place the tube set in such a way that it either
does not touch the patient at all or that the contact surface is well insu­
lated (several layers of sterile cloth or gauze between the tube set and
the patient).
GA-A 274
34
5.12.2
Notes and instruction for using smoke gas evacuation
.
IMPORTANT!
The disposable tube set for smoke gas evacuation must only be used for
evacuating smoke gas and not for suctioning off liquids.
D Smoke gas evacuation can only be activated in the "Highflow" device
mode.
D Smoke gas evacuation is not available in Baby mode (option).
D The smoke gas evacuation pump is only switched on at pressures > 3
mmHg and is switched off automatically at pressures < 3 mmHg.
D Smoke gas evacuation can be activated and deactivated via the touch­
screen using the Evacuation button or via the footswitch (option).
5.12.3
Creating a pneumoperitoneum
Primarily the hospital's own specifications and routines must be followed
(primary insufflation via Veress cannula or minilaparotomy (open ac­
cess)).
Firstly select the device mode:
"Standard" device mode (access with Veress cannula)
D Start primary insufflation by pressing the Insufflation button.
D Press the Highflow button to switch to highflow insufflation.
"Highflow" device mode (access via minilaparatomy – open access)
D Start highflow insufflation using the Insufflation button.
5.12.4
Highflow device mode
WARNING!
Danger of gas embolism.
The "Highflow" device mode should only be used under endoscopic view.
.
35
NOTE!
If insufflation at a highflow rate is used over a prolonged period of time,
the CO2 gas supply may cool down and cause a transitional drop in the
gas pressure of the supply container. At the same time an operator's note
is displayed.
When insufflation is stopped, the original gas pressure will build up again
in the supply tank/cylinder.
GA-A 274
5.12.5
Measures after carrying out insufflation
.
IMPORTANT!
Close the gas bottle/cylinder valve after the application. Any leaks at
bottle/cylinder valve or on the connection can cause the gas bottle/cy­
linder to be completely emptied after some time.
Z Close the gas bottle/cylinder valve.
Z Remove and discard the hygiene filter.
Z Discard or reprocess, respectively, the insufflation tube.
5.12.6
Possible alarms during the application
Overpressure alarm:
Possible causes:
D The patient strains muscles as the anesthetic effect gets weaker.
D The user exerts pressure onto the abdomen.
D Additional gas or liquid flow into the abdomen.
Gas supply alarm:
D See section 5.2.2
5.12.7
Power failure
5.12.8
Taking out of service
In the case of a power failure (< 30 s), the selected modes of operation
are saved (and remain active). After reestablishing the power supply, you
may continue with the setting previously selected.
Z To take the device out of service, switch off the power switch and dis­
connect the device from the power supply / mains.
GA-A 274
36
5.13
Overview of operator messages
5.13.1
Structure of operator messages
There are 4 different types of messages.
The device messages are displayed in the corresponding color together with a
symbol and the associated message number, as required by the message type.
For certain messages the core system will only display the "Follow messages on
device" note.
Please note:
In the tables, only the last 3 digits of the message number are displayed.
Message text
XXX-XXX-XXX
optional fault correction
OK
5.13.2
Message types
Symbol
Color
Green
Turqoise
Yellow
Red
5.13.3
Message type
Operator's note
LOW error
MEDIUM error
HIGH error
Description
Informs about certain events.
(Audio signal can be activated/deactivated in the
"Audio" menu).
Informs the user about imminent dangers.
Error of low priority.
Informs the user about imminent dangers.
Error of medium priority.
Indicates severe device errors which exclu­
des any further control of the device.
High-priority faults.
-------
1 alarm pulse
Symbol does not
blink.
3 alarm pulses
Symbol blinks
10 alarm pulses
Symbol blinks
Functionality of control elements
Control element
No button
OK
Description
The message window is displayed until the conditions are fulfilled.
The message window can be exited.
The operation of the device may continue but with limited functions.
Audio
The audio alarm signals are activated.
The audio alarm signals can be switched off by pressing the button.
Audio
The audio alarm signals are deactivated.
Pressing the button switches on the acoustic alarm signals.
Service
37
Alarm sequence
Once this button is displayed, you can only switch to the "Service" submenu.
The device can no longer be operated.
GA-A 274
5.13.4
Operator's messages (green)
Message
number
Message text
xxx-xxx-005
Connection to core system established
xxx-xxx-006
Possible cause
Corrective action
During the runtime a connection
to the core system has been
established
' Status message
Connection to core system disconnected
During the runtime the connection
to the core system was inter­
rupted
' Status message
xxx-xxx-007
Language switchover by core system successful
Language switchover by the core
system
' Status message
xxx-xxx-008
Upper range limit reached
-------
' Status message
xxx-xxx-009
Lower range limit reached
-------
' Status message
xxx-xxx-010
Please wait, the device parameters are being
reset!
-------
' Status message
xxx-xxx-011
Memory access, please wait
-------
' Status message
xxx-xxx-012
Please wait, profile is being deleted
-------
' Status message
xxx-xxx-013
Please wait, a new profile is generated
-------
' Status message
Maximum number of profile entries is reached
xxx-xxx-014 To generate a new profile, an existing profile must Maximum profile entries reached
be deleted.
xxx-xxx-015
Invalid profile name
Please enter profile name again. The name must
contain at least 1 alphanumeric character.
xxx-xxx-016
The profile name entered already exists
Please enter a different name
xxx-xxx-018
Incorrect profile name input
' Carry out message
Double profile name
' Carry out message
To exit the service menu, a device restart is requi­ Message when logging into the
red.
service area
xxx-xxx-019
Wrong service password
' Carry out message
' Status message
Password spelling error
' Status message
xxx-xxx-020
Please wait
-------
' Status message
xxx-xxx-021
Function cannot be activated
-------
' Status message
xxx-xxx-027
Please restart device
A device restart is necessary
' Carry out message
Gas supply interrupted
CO2 pressure sensor defective
' Open gas bottle/cylinder or
connect full gas bottle/cylinder
' If this message occurs
repeatedly although the gas
bottle/cylinder is full and the
valve is open, contact the ser­
vice department
Possible cause
Corrective action
EEPROM memory error
' Confirm with OK button automatic device restart
' If fault occurs again, contact
service department
Device temperature too high
' Allow the device to cool
down
' Carry out a device restart
' If fault occurs again, contact
service department
Device temperature too low
(outside the permissible operating
conditions)
' Wait until device temperature
meets the required operating
conditions
Temperature sensor defective
' Contact service department.
xxx-xxx-436
xxx-xxx-437
5.13.5
Insufficient gas supply
Please check the gas supply
LOW error (turqoise)
Message
number
Message text
xxx-xxx-017
Profile data not correctly initialized
Press "OK" to delete the profile data and set to
default values.
xxx-xxx-028
xxx-xxx-029
Device temperature too high
Please allow the device to cool down
xxx-xxx-030
Device temperature too low
Please allow the device to warm up
xxx-xxx-031
Device error - Temperature sensor defective
GA-A 274
38
Message
number
Message text
xxx-xxx-033
Environment light sensor defective
Automatic brightness control of display is swit­
ched off.
xxx-xxx-034
to
xxx-xxx-036
Device error - No connection to core system pos­
sible
xxx-xxx-037
Possible cause
Corrective action
Environment light sensor defec­
tive
' Contact service department.
Electronics error
' Carry out a device restart
' If fault occurs again, contact
service department
Faulty touchscreen calibration
The default calibration is used, for recalibration
switch to service menu.
Saved EEPROM values of the
touch calibration are not within
the calibration range
' Carry out recalibration on the
service level (general service).
' If fault occurs again, contact
service department
xxx-xxx-038
to
xxx-xxx-042
Device error - Faulty initialization, device is not
ready
Electronics error
' Carry out a device restart
' If fault occurs again, contact
service department
xxx-xxx-043
Device error - Data transmission
Software error - Internal data
transmision
' Contact service department.
xxx-xxx-061
Device error - Device not ready
Internal operating voltages not
okay
' Carry out a device restart
' If fault occurs again, contact
service department
xxx-xxx-062
Device error - Faulty initialization, device is not
ready
Hardware and software versions
are not compatible
' Carry out a device restart
' If fault occurs again, contact
service department
xxx-xxx-063
Device error - No connection to core system pos­
sible
core processor update faulty
' Carry out a device restart
' If fault occurs again, contact
service department
xxx-xxx-064
Device error - type number
The type number/model number
saved in the device is faulty
' Carry out a device restart
' If fault occurs again, contact
service department
xxx-xxx-066
Device error - Initialization failed
The device is not ready
Electronics error
' Carry out a device restart
' If fault occurs again, contact
service department
xxx-xxx-067
Device error - Data transmission
Software error - Internal data
transmision
' Contact service department.
xxx-xxx-400
Device error - Hardware safety activated
Device error
' Contact service department.
xxx-xxx-403
to
xxx-xxx-404
Device error - Pneumatic component
Electronics error
' Contact service department.
xxx-xxx-405
Device error - Pneumatic test
Pneumatic component defective
' Contact service department.
xxx-xxx-406
Device error - Hardware incompatible
Electronics error
' Carry out a device restart
' If fault occurs again, contact
service department
xxx-xxx-407
Device error - Smoke gas evacuation module
Electronics error
' Contact service department.
xxx-xxx-408
Device error - Electronics
Electronics error
' Contact service department.
xxx-xxx-409
xxx-xxx-410
Device error - Pneumatic component
Pneumatic component defective
' Contact service department.
xxx-xxx-411
xxx-xxx-412
Device error - Electronics
Electronics error
' Contact service department.
xxx-xxx-413
Device error - Gas heating module
Electronics error
' Contact service department.
xxx-xxx-415
Device error - Electronics
Electronics error
' Contact service department.
xxx-xxx-428
Device error - Gas heating module
Electronics error
' Contact service department.
xxx-xxx-433
Device error - Data transmission
Gas heating module is deactivated
Software error - Internal data
transmision
' When actuating the OK but­
ton, the gas heating module is
deactivated
' If fault occurs again, contact
service department
39
GA-A 274
5.13.6
MEDIUM error (yellow)
Message
number
Message text
xxx-xxx-032
Possible cause
Corrective action
Device error - Temperature sensor defective
Temperature sensor defective
' Contact service department.
xxx-xxx-044
Device error - Data transmission
Software error - Internal data
transmision
' Contact service department.
xxx-xxx-068
Device error - Data transmission
Software error - Internal data
transmision
' Contact service department.
xxx-xxx-414
Device error - Gas heating module
Gas heating module is deactivated
Electronics error
' When actuating the OK but­
ton, the gas heating module is
deactivated
' If fault occurs again, contact
service department
xxx-xxx-417
to
xxx-xxx-419
Device error - Pneumatic component
Electronics error
' Contact service department.
xxx-xxx-421
Device error - Smoke gas evacuation module
Electronics error
' Contact service department.
xxx-xxx-422
Device error - Pneumatic component
Pneumatic component defective
' Contact service department.
xxx-xxx-424
Device error - No input pressure
Gas supply interrupted
CO2 pressure sensor defective
' Open gas bottle/cylinder or
connect full gas bottle/cylinder
' If this message occurs
repeatedly although the gas
bottle/cylinder is full and the
valve is open, contact the ser­
vice department
xxx-xxx-425
Device error - Pneumatic component
Pneumatic component defective
' Contact service department.
xxx-xxx-427
Device error - Electronics
Electronics error
' Contact service department.
xxx-xxx-429
xxx-xxx-430
Device error - Gas heating module
Electronics error
' Contact service department.
Possible cause
Corrective action
Device error
' Contact service department.
Overpressure alarm pressure > 30 mmHg (mercury
column) for more than 5 s
' Decrease intraabdominal
pressure
5.13.7
HIGH error (red)
Message
number
Message text
xxx-xxx-401
Device error - Hardware safety activated
xxx-xxx-402
Critical overpressure > 30 mmHg (mercury co­
lumn)
xxx-xxx-416
Device error - Electronics
Electronics error
' Contact service department.
xxx-xxx-420
Device error - Pneumatic component
Electronics error
' Contact service department.
xxx-xxx-423
Device error - Electronics
Electronics error
' Contact service department.
xxx-xxx-426
Device error - Software
Software error
' Contact service department.
GA-A 274
40
6
Operation within the core system
6.1
Functional principle within the core system
In normal operation, the device can be controlled via the core system
using the touchscreen monitor, optional voice control as well as the con­
trols of the device itself.
.
6.2
IMPORTANT!
Should the core system fail, the device can still be controlled manually via
the controls on the device itself. For a description please refer to the in­
struction manual of the respective device.
Combining and controlling the device
Via the integrated CAN BUS interface the device can be integrated in the
Richard Wolf core system .
At the CAN BUS interface only the components approved for use with the
core system can be used.
The components must fulfill the requirements in accordance with the cur­
rent instruction manual for the core system, section on "Possible combi­
nations".
The control computer complies with IEC / EN 60601-1 and may therefore
be used in the patient environment.
Further descriptions of use:
D Follow the latest instruction manual for the core system.
D For the operation, menu structure, software updates and configurations
follow the latest version of the core.software manual.
D For a detailed understanding of the device and menu functions, follow
the instruction manuals of the connected devices.
6.3
Setting the system ID
Z Make sure that the same system IDs are set on the insufflator and on
the camera controller as otherwise these devices cannot communicate
with each other.
' The system ID for this device can be found/set in the Device para­
meters - Service - General service - System Settings menu (see sec­
tion 5).
41
GA-A 274
6.4
Connection to the core system
Touchscreen
monitor
RIWO
CONTROL
core.control
USB
headset connector
CAN BUS
e.g.
INSUFFLATOR
Highflow 45
CAN BUS
e.g.
HIGHLIGHT X300
light projector
CAN BUS
e.g.
Camera
CAN BUS termination
.
IMPORTANT!
The device system must be operated via a "galvanically isolated transformer".
.
IMPORTANT!
The last device in the CAN BUS chain must be terminated using the supplied terminating resistor.
GA-A 274
42
7
Reprocessing and maintenance
7.1
Reprocessing of device
WARNING!
Prevent humidity from entering the device.
Danger of electric shock!
Before reprocessing switch off and disconnect the device from the power
supply / mains.
Clean the device with a soft cloth moistened with surface disinfectant,
alcohol or spirit.
Follow the disinfectant manufacturer's instructions!
7.2
.
NOTE!
For a description of the surface cleaning and disinfecting agents for
devices and universal device carts / trolleys and utility carts approved by
Richard Wolf, please refer to manual GA-J020 "Reprocessing of
RICHARD WOLF Heat-stable Instruments".
.
IMPORTANT!
Make sure that no humidity enters the device. Do not use any cleaning
agents, scouring agents or solvents on this device.
Reprocessing of reusable insufflation tubes
.
IMPORTANT!
Follow the latest notes and instructions in manual GA-J 020, “Reproces­
sing of RICHARD WOLF heat-stable instruments“.
Do not use disinfectants containing peracetic acid without corrosion
protection, phenols or chlorine components for the reprocessing of Ri­
chard Wolf products.
Strictly adhere to the maximum immersion time in the disinfectants used,
as described by the manufacturers.
WARNING!
Do not reprocess disposable items!
The service life of products marked as disposable i.e. for single use only,
has been designed for only one use in or on a single patient.
If disposable items are reprocessed to be used again, a deterioration of
the product quality cannot be excluded, which will endanger the patient,
the user and others.
In this case, the manufacturer can no longer guarantee product safety
and performance.
43
GA-A 274
7.2.1
Cleaning
7.2.2
Sterilization
7.3
Reprocessing of reusable highflow insufflation tube with heating wire
Z Rinse out the tubes with a cleaning gun and dry with compressed air.
Steam sterilization:
Z Steam sterilization using a fractional pre-vacuum procedure at 134 C
(272 F).
Z For the reprocessing follow the manual of the reprocessable highflow
insufflation tube with heating wire used.
7.4
Maintenance
.
IMPORTANT!
In the case of inquiries or in your correspondence please always indicate
the product number and the serial number listed on the identification
plate. Further documentation is available from the manufacturer on re­
quest.
Service hotline:
7.4.1
7.5
7.5.1
Phone:
Fax:
+49-(0)-7043/35-389
+49-(0)-7043/35-1209
Maintenance intervals
.
IMPORTANT!
To avoid any incidents or damage caused by aging and wear it is neces­
sary to service the product and the accessories at adequate intervals.
Depending on the frequency of use, but at least once a year, have an
expert check the functional and operational safety of the equipment.
.
NOTE!
A check may only be carried out by qualified operator staff. If the spe­
cified measured values and tolerances are not met, the device must be
checked by an authorized service technician.
.
IMPORTANT!
The device must not be used if the specified values are not displayed or
functions are not fulfilled.
Quarterly check
Measuring and test equipment
- Insufflation tube (8170.101), original length 2.5 m
- Hygiene filter (4171.111)
- Transparent container (measuring cup)
GA-A 274
44
7.5.2
Checking the pressure and flow control
Primary insufflation:
Z Preselect the following nominal pressure: 12 mmHg (mercury column)
Z Connect the hygiene filter and insufflation tube to the insufflation con­
nector.
Z Start insufflation.
' The current flow rate must agree with the preselected flow rate of 1
l/min.
' Permissible tolerance: ¦ 0.5 l/min
Z Close the insufflation tube with your thumb.
' As soon as the current pressure exceeds the preselected pressure,
the device must release pressure.
Z Switch off insufflation and open the insufflation tube.
Highflow:
D Preselect the following nominal pressure: 12 mmHg (mercury column)
D Preselected the following flow rate: 10 l/min
Z Start insufflation, then switch over to Highflow.
' The current flow rate must agree with the preselected flow rate of 10
l/min.
' Permissible tolerance: ¦ 2.0 l/min
Z Close the insufflation tube with your thumb.
' As soon as the current pressure exceeds the preselected pressure,
the device must release pressure.
Z Open the insufflation tube.
Z Switch back from highflow to primary insufflation.
' After 1.5 s at the latest, a flow rate of 1 l/min ¦0.5 l/min must be dis­
played.
Z Switch off insufflation.
7.5.3
Leakage test
Z Preselect the following nominal pressure: 12 mmHg (mercury column)
Z Fill a transparent container (e.g. measuring cup) with water.
Z Immerse the patient end of the insufflation tube approx. 5 - 10 cm into
the water (5 - 10 cm below the surface).
Z Start insufflation.
' Gas bubbles must be observed at the end of the tube.
Z Switch off insufflation.
' No gas bubbles must emerge from the end of the tube.
Z After the end of the test, the insufflation tube must be removed from
the water container immediately as otherwise vacuum may pull water
into the tube.
45
GA-A 274
8
Technical description
8.1
Troubleshooting
.
IMPORTANT!
If the faults cannot be corrected with the help of this table, please contact
our service department or return the device for repair.
'Do not attempt to do any repairs yourself!
Errors
Possible cause
Corrective action
Device is not functional
Power switch is not on
'Actuate the power switch
Power cable is not connected
'Connect the power cable
Fuse in device is defective
'Replace the fuse
No line voltage/mains
'Check in-house power supply
Insufflation not activated
'Switch on insufflation
Gas bottle/cylinder empty or not open
'Replace or open gas bottle/cylinder
As a result of manipulations of the
user, the abdominal pressure increa­
ses
'Remove cause
Device defective
'Contact the service dept.
No gas flow
Preselected pressure is exceeded
Other devices cause a pressure incre­ 'Remove cause
ase
Pressure won't build up
Smoke gas evacuation won't start
Tube system is leaking
'Check tube and connectors
Flow preselection is to low
'Increase flow preselection
Instrument stopcock is closed
'Open instrument stopcock
Pressure < 3 mmHg
'Generate pressure > 3 mmHg
Evacuation not actuated
'Actuate evacuation
'Open evacuation tube
8.1.1
Fault in the core system
Errors
Possible cause
Corrective action
'Check connections
Device is not logged in the core system
Data transfer error
'Replace CAN BUS cable
'Set to the same system ID
'Contact the service dept.
GA-A 274
46
8.2
Technical data
Voltage
Frequency
Current
rating
A
Fuse
Hz
Power
consumption
VA
Vμ
2235001
2235011
100-240
50 / 60
60
0.6 - 0.2
T 3.15 H
2235021
2235031
100-127 /
220-240
50 / 60
80
0.8 - 0.3
T 3.15 H
2235601 (USA)
2235611 (USA)
100-240
50 / 60
60
0.6 - 0.2
T 3.15 H
2235621 (USA)
2235631 (USA)
100-127 /
220-240
50 / 60
80
0.8 - 0.3
T 3.15 H
Product no.
Electromagnetic compatibility EMC to
IEC / EN 60601-1-2
Medical devices directive 93/42/EEC
Class IIb
Protection against electric shock
Type CF applied part
Safety class to IEC / EN 60601-1 ;
(UL 60601-1 / CSA C22.2 No.601.1 - for USA)
I
Degree of protection against the ingress of liquids
IP 20 (not protected)
Mode / duty factor
Continuous operation
Degree of protection when flammable mixtures are
present
Dimensions
WxHxD
300 mm x 160 mm x 455
Application parameters
Insufflation medium
Input pressures
Intraabdominal pressure
Gas flow rate, primary filling
Gas flow rate highflow
Consumption indicator
Gas temperature with heating tube
47
This device is not protected against explosions
(Do not operate this device in ignitable atmosphere)
8.9 kg (19.5 lbs) to 12.8 kg (28.1 lbs)
(depending on configuration)
Weight
8.2.1
A
med. CO2 (degree of purity > 99.5 %)
4 bar to 7.5 bar
3 mmHg to 25 mmHg (¦ 2 mmHg)
preselectable in increments of 1 mmHg
(mercury column)
1 l/min (¦ 50 %)
Depending on
device mode
2 l/min to 45 l/min (¦ 20 %),
selectable in increments of 1 l/min
0 l to 999 l (¦ 20 %)
37 ºC ¦ 2 ºC
GA-A 274
8.2.2
Pressure and flow values
Device mode
Pressure values min. / max.
Flow values min. / max.
Default
min.: 3 mmHg (mercury column)
max.: 25 mmHg (mercury column)
Primary insufflation (1l/min)
min.: 2 l/min
max.: 45 l/min
Highflow
min.: 3 mmHg (mercury column)
max.: 25 mmHg (mercury column)
min: 2 l/min
max.: 45 l/min
Baby mode (option)
min.: 3 mmHg (mercury column)
max.: 20 mmHg (mercury column)
min.: 0,1 l/min
max.: 2 l/min
TEM (option)
min.: 3 mmHg (mercury column)
max.: 25 mmHg (mercury column)
Primary insufflation (1l/min)
min.: 2 l/min
max.: 8 l/min
Continuous flow (option)
min.: 3 mmHg (mercury column)
max.: 25 mmHg (mercury column)
Primary insufflation (1l/min)
min.: 2 l/min
max.: 8 l/min
Gynecology
min.: 3 mmHg (mercury column)
max.: 25 mmHg (mercury column)
Primary insufflation (1l/min)
min.: 2 l/min
max.: 45 l/min
Urology
min.: 3 mmHg (mercury column)
max.: 25 mmHg (mercury column)
Primary insufflation (1l/min)
min.: 2 l/min
max.: 45 l/min
8.2.3
Interfaces
Gas connection
Connection for hygiene filter
NIST connector to EN 739 for CO2
Conical connector, dia.  15.0 mm, DIN ISO 5356 T1
Power supply connection for highflow insufflation tube
with heating wire, reprocessable
5-pin socket, small
Power supply connection for disposable highflow insuf­
flation tube with heating wire
5-pin socket, large
Connector for smoke gas evacuation
CAN BUS interface
Service interface
GA-A 274
Conical connector, dia. 15.0 mm,
DIN ISO 5356 T1
Sub-D socket, 9-pin
RJ 45
48
8.2.4
Optional device variants
Options
INSUFFLATOR
Highflow 45
BASIC
INSUFFLATOR
Highflow 45
EVAC
INSUFFLATOR
Highflow 45
HEAT
INSUFFLATOR
Highflow 45
EVAC + HEAT
(Basic version)
(Basic version
with
smoke gas evacuation)
(Basic version
with
gas heating)
(Basic version
with
smoke gas evacuation
and gas heating)
2235001
2235011
2235021
2235031
Software module 1
Connection to core
(CAN BUS)




Software module 2
Baby mode








Software module 4
Continuous flow mode
(TEM + LAP max. 8 l/min)




TEM
LAP
----


----


Device mode
Default




Device mode
Highflow




Device mode
Baby mode *
 1)
 1)
 1)
 1)
Device mode
TEM
-----
 2)
-----
 2)
Device mode
Continuous flow
 2)
 2)
 2)
 2)
Software module 3
Video on-screen display with
dialog
(CAN BUS)
 = standard equipment
 = option
 1) = only in conjunction with software module 2
 = possible
----- = not possible
 2) = only in conjunction with software module 4
* = In "Baby mode" evacuation and highflow cannot be activated and are masked off.
8.3
Operating, storage, transport and shipping
Operating conditions
+ 10ºC to + 40ºC , 30% to 75% rel. humidity
atmospheric pressure 700 hPa to 1060 hPa
Storage, transport and shipping conditions
- 20ºC to + 60ºC, 10% to 90% rel. humidity
atmospheric pressure 700 hPa to 1060 hPa
.
NOTE!
To prevent damage during transport or shipment of the products we recommend using the original
packaging material.
49
GA-A 274
8.4
Spare parts and accessories
Product no.
Designation
64268.007
Device fuse T 3.15 H (pack of 10)
2440.03
Power cable (Europe), 3.0 m
8170.101
Insufflation tube, dia.  5.0 mm, length 2.5 m, reprocessable
8170.232
Highflow insufflation tube, dia.  8.0 mm, length 2.5 m, reprocessable
4170.501
Single-use insufflation tube, length 2.5 m, sterile (pack of 10)
8170202
Highflow insufflation tube with heating wire, length 3.0 m, reprocessable, only 2235021, 2235031 /
2235621, 2235631
4170502
Disposable highflow insufflation tube with heating wire and integrated hygiene filter, sterile (pack of
10), only 2235021, 2235031 / 2235621, 2235631
4170503
Disposable tube set for smoke gas evacuation, sterile (pack of 10), only 2235011, 2235031 /
2235611, 2235631
4170504
Smoke gas blow-off filter, sterile (pack of 10)
4171.111
Hygiene filter, single-use item, sterile (pack of 10)
74021.029
CO2 pressure tube, length 1.5 m
74021.030
Connection tube for centralized CO2 gas supply (Dräger), length 5.0 m
74021.035
Connection tube for centralized CO2 gas supply (Air Liquide), length 5.0 m
74007.056
Pressure reducing valve for CO2, ISO 5145
74007.054
Pressure reducing valve for CO2, DIN EN 738-1
74007.055
Pressure reducing valve for CO2, DIN EN ISO 707 (PIN Index)
74002.267
Space seal for bottle/cylinder connection on pressure reducing valve, ISO 5145
2221.951
Bottle holding bracket, houses CO2 bottles, ( 80 to 90 mm)
2221.952
Bottle holding bracket, houses CO2 bottles, ( 100 to 115 mm)
2046.961
0.75 kg CO2 bottle (approx. 350 l CO2) with DIN connector, ( 80 mm)
2030.1031
Single-pedal footswitch for smoke gas evacuation
35100.115
Silencer (pack of 3), only 2235011, 2235031 / 2235611, 2235631
' Further accessories on request
ISO connector (ISO 5145)
DIN connector (DIN EN 738-1)
PIN Index connector (DIN EN ISO 407)
The products can be combined as required provided the relevant technical data and intended uses are observed. For
the total overview please refer to the latest catalog sheets and brochures or contact Richard Wolf or your representative.
GA-A 274
50
8.5
Replacing parts
8.5.1
Device fuses
CAUTION!
The specifications of the device fuses must correspond with the fuse ra­
tings on the identification plate.
Use only the fuses specified in the spare parts list.
L Power input connector with fuse holder
1
2
2
3
Z Switch off the device and disconnect the power cable from the wall
socket and from the power input connector of the device.
Z Unscrew the fuse holder [1] using a screwdriver.
Z Take the fuses [2] out of the fuse holder [1] and replace.
Z Reinsert the fuse holder and push until it clicks in place [3].
8.5.2
51
Disposal of product, packaging material and accessories
For the disposal follow the laws and regulations valid in your country.
' For further information please contact the manufacturer.
GA-A 274