ZXR LED Down Light Med-Vizion™ MRI Interior LED Lighting System Installation Manual for MRI Applications ETS-Lindgren Inc. reserves the right to make changes to any products herein to improve functioning or design. Although the information in this document has been carefully reviewed and is believed to be reliable, ETS-Lindgren does not assume any liability arising out of the application or use of any product or circuit described herein; nor does it convey any license under its patent rights nor the rights of others. All trademarks are the property of their respective owners. © Copyright 2014 by ETS-Lindgren Inc. All Rights Reserved. No part of this document may be copied by any means without written permission from ETS-Lindgren Inc. Trademarks used in this document: The ETS-Lindgren logo is a registered trademark and Med-Vizion is a trademark of ETS-Lindgren Inc.; Scotch-Brite is a trademark of 3M. Revision Record MANUAL, MANUAL,MEDVIZION ZXR | Part #206472, Rev. B Revision Description Date A Initial Release September, 2014 B Added Appendix C: Warranty October, 2014 ii www.ets-lindgren.com Table of Contents Notes, Cautions, and Warnings ........................................................................ iv 1.0 Introduction ................................................................................................... 5 Before You Begin Installation ............................................................................................. 5 Other Important Safety Requirements and Precautions .................................................... 5 Standard Configuration ....................................................................................................... 6 Optional Items..................................................................................................................... 7 Required Items (To Be Supplied by Customer).................................................................. 7 Approvals ............................................................................................................................ 8 UL/cUL ......................................................................................................................... 8 CE ................................................................................................................................ 8 Local Authority ............................................................................................................. 8 2.0 Pre-Installation Tasks ................................................................................... 9 Product Delivery and Inspection ......................................................................................... 9 Site Preparation .................................................................................................................. 9 Verification Before Installation ............................................................................................ 9 Electrical Requirements .................................................................................................... 10 3.0 Installation ................................................................................................... 11 Required Tools and Materials (To Be Supplied by Customer) ......................................... 11 Med-Vizion Fixture Installation ......................................................................................... 12 Trim Ring Installation ........................................................................................................ 15 EMI Facility Filter Installation (To Be Supplied by Customer) .......................................... 15 To Mount RF Filter to Shield ...................................................................................... 17 Digital Dimmer Module Installation (Optional) .................................................................. 17 Appendix A: Glossary of Terms ...................................................................... 19 Appendix B: Typical Wiring Schematics–ZXR System .................................. 21 Non-Dimming Installation ................................................................................................. 21 Optional Dimmer Installation ............................................................................................ 21 Dimmer Module/Power Supply Filter: Included with Dimmer ........................................... 22 Dimmer Wiring–External to MRI Room ............................................................................ 22 Dimmer Wiring–Internal to MRI Room ............................................................................. 23 ZXR, LED, Lamp Operation .............................................................................................. 25 Appendix C: Warranty ...................................................................................... 27 Scope and Duration of Warranties ................................................................................... 27 www.ets-lindgren.com iii Warranty Exclusions ......................................................................................................... 28 Buyer’s Remedies ............................................................................................................ 28 List of Figures Figure 1: Med-Vizion LED Lighting System ................................................................................ 6 Figure 2: ZXR Light Engine ....................................................................................................... 12 Figure 3: Bar Hangers inserted through Plaster Frame ............................................................ 12 Figure 4: Adjustable Bar Hanger End ....................................................................................... 13 Figure 5: Junction Box Cover .................................................................................................... 13 Figure 6: Non-Ferrous Fitting on Junction Box (not user serviceable)...................................... 14 Figure 7: Junction Box Wiring Connections .............................................................................. 14 Figure 8: Trim Ring Modification ............................................................................................... 15 Figure 9: EMI Facility RF Filter (ELUL 2030) ............................................................................ 16 Figure 10: EMI Facility Filter Installation Procedure ................................................................. 16 Figure 11: Dimmer Input/Output RJ45 Plugs ............................................................................ 17 Figure 12: For Non-Dimming System Installation ..................................................................... 21 Figure 13: Dimming System Installation–External to MRI Room .............................................. 23 Figure 14: Dimming System Installation–Internal to MRI Room ............................................... 24 Notes, Cautions, and Warnings Note: Denotes helpful information intended to provide tips for better use of the product. CAUTION: Denotes a hazard. Failure to follow instructions could result in minor personal injury and/or property damage. Included text gives proper procedures. WARNING: Denotes a hazard. Failure to follow instructions could result in SEVERE personal injury and/or property damage. Included text gives proper procedures. iv www.ets-lindgren.com 1.0 Introduction This manual provides instructions to install the ETS-Lindgren Med-Vizion™ MRI Interior LED Lighting System. All Med-Vizion system components are designed for indoor use and installation only. Note: For assistance during the installation process or operations thereafter, please contact ETS-Lindgren. Before You Begin Installation For the safe handling, installation, and operation of the Med-Vizion system, a thorough review and understanding of the information in this manual must be completed before starting the installation process. WARNING: There are no serviceable components in the Med-Vizion system. Failure to properly install the Med-Vizion system per the instructions in this manual will void your warranty. Attempting to repair or alter the Med-Vizion system in any way will void your warranty. WARRANTY WARNING: Additional supports and/or hangers for the drop ceiling grids on Med-Vizion fixture(s) are recommended and necessary in earthquake zones or when required by local/state safety codes. Always install Med-Vizion according to all local, state and national codes. Other Important Safety Requirements and Precautions WARNING: Power to the Med-Vizion system must be de-activated before attempting to wire or service this product at any time. AC power is removed by activation of the wall power switch. AC power cannot be removed by use of the optional dimmer panel. Make sure that all required safety equipment is present and all workers are familiar with the local safety codes. Observe proper precautions when working in an MRI suite. Always assume the magnet is active. www.ets-lindgren.com Introduction 5 Installation requires one separate 120 VAC branch circuit (rated at 20 Amps) to power up to 28 ZXR fixture(s). Class 2 wiring (e.g. for dimmer cables between the ZXR units). The Med-Vizion system is not intended for use in air handling spaces. Standard Configuration The following components are included with the Med-Vizion system: Med-Vizion Light Engine Assembly and Plaster Frame/Driver Assembly Rails, Trim Ring, and Lens Installation Instructions Light Engine Assembly RF Shielded Driver Assembly Plaster Frame/Driver Assembly Figure 1: 6 Introduction Med-Vizion LED Lighting System www.ets-lindgren.com Optional Items The following components are included with the Med-Vizion system dimmer option: Med-Vizion Dimmer Module Med-Vizion Dimmer Power Supply Low-voltage Communication/Power Filter Required Items (To Be Supplied by Customer) WARNING: All components supplied by the installer for use inside an MRI room must be non-ferrous. The following components are not supplied by ETS-Lindgren and must be provided by the customer to complete the installation process: Class 1 Conduit/ box for incoming mains power wiring Class 1 Conduit and fittings for the wiring between the Med-Vizion fixture(s) and the EMI facility filter EMI facility filter, minimum ratings: 277 supplies up to 28 lights) ZXR system power switch, a standard SPST light switch rated for 20A, or equivalent Ceiling grid components Drop ceiling grid supports, hangers, or other hardware as required by National and Local Building Codes All cables, including all RJ45 and RJ-45 cables shown in Figure 12 on page 21 and Figure 13 on page 23 www.ets-lindgren.com VAC VAC /120 , =/>20A (one circuit Introduction 7 Approvals UL/CUL The Med-Vizion system is constructed as an Indoor Recessed Down light per UL 1598, LUMINAIRE STANDARD (both US and Canadian Requirements). CE The Med-Vizion system is compliant with all applicable European directives. Technical Construction Files are available upon request. LOCAL AUTHORITY The subcontractor/installer should secure permits with the appropriate authorities. 8 Introduction www.ets-lindgren.com 2.0 Pre-Installation Tasks CAUTION: Before connecting any components, follow the information provided in Introduction on page 5. Product Delivery and Inspection Note: Customer is responsible for any damage not reported within fifteen (15) days of receipt of shipment. Upon delivery, immediately unpack the Med-Vizion™ MRI Interior LED Lighting System. Inspect the product to ensure that nothing is damaged and that all components have been received. Immediately notify the freight company of any damaged components. Damaged product must not leave the loading dock until the shipper can verify claim. Site Preparation Before beginning site work, notify the business or construction manager of the following: Scope of work; include length of installation, any disruptions to electrical service, and what specific hours of the day the installation is to be done. Any safety requirements or conditions specific to the installation site. Verification Before Installation WARNING: Verify load capability as instructed in step 2 following this warning. Any ceiling grid, ZXR fixture(s), or tile(s) falling onto a person(s) or equipment in a room where the ZXR assembles have been installed may cause serious injury or damage. 1. Clearance: A minimum clearance of 9 inches above the surface of the drop ceiling is required for installation for every Med-Vizion light box. The ZXR assembly rises 6 inches above the ceiling tile when installed. 2. Load capability: The ceiling grid must be capable of supporting the combined weight of the ZXR fixtures. The installer is responsible for verifying the load capability of the support grid. www.ets-lindgren.com Pre-Installation Tasks 9 3. Condition of the magnet: Confirm if the magnet has been energized. Active MRI magnets produce extremely powerful magnetic fields that can produce serious injury or death if proper safety precautions are not followed. Electrical Requirements Circuits must be wired in accordance to all state and local electrical codes. If any ZXR lamp senses an over-temperature condition, it will shut OFF. In order to restore normal operation it will be necessary to reset the system by turning the ZXR system power switch OFF, then ON again. If an optional dimmer control is installed, turning the lamps off with the dimmer control WILL NOT restore the lamp system properly. If a dimmer module is made part of the Med-Vizion system, a standard 120VAC outlet is required for dimmer power. The dimmer module power supply must be installed outside of the MRI room. 10 For dimmer modules installed outside of the MRI room (preferred location): A separate facility power filter is not required for the dimmer module power supply. For dimmer modules installed inside of the MRI room: Route the power supply output through a PN254757 1amp/250V facility filter by means of RJ45, RJ45 cable to the dimmer module location. Commercial, incandescent, lighting dimmers, or fan speed controls (electronic, phase angle) are not compatible with the ZXR lighting system. Electronic, phase angle dimmer controls are not compatible with facility RF electric filters nor will they operate the individual ZXR LED lamps. Pre-Installation Tasks www.ets-lindgren.com 3.0 Installation WARNING: All tools must be approved for use in an MRI suite; always assume the magnet is active. CAUTION: Before installing any components, follow the safety information provided in Introduction on page 5. Note: Before installing any components, complete the steps in Pre-Installation Tasks on page 9. Required Tools and Materials (To Be Supplied by Customer) The following items are recommended for the installation of the Med-Vizion™ MRI Interior LED Lighting System: Tape measure and ladder(s) Wire strippers Screwdrivers appropriate for hardware Drill with hole-forming bit or saw appropriate for through-wall EMI dimmer filter installation (optional) Channel locks or adjustable wrench for EMI filter nut (optional) Additional grid ceiling support wires as needed (must be non-ferrous) www.ets-lindgren.com Installation 11 Med-Vizion Fixture Installation Figure 2: ZXR Light Engine 1. With all wiring and conduit installed per the site architect’s direction, remove the ceiling tile in which the Med-Vizion system is to be installed. 2. Find the center of the ceiling tile and cut a 6-3/4” diameter hole. Bar Hangers Figure 3: 12 Installation Bar Hangers inserted through Plaster Frame www.ets-lindgren.com 3. Insert two retractable bar hangers through the slots in the end of the plaster frame. 4. Remove ceiling tile adjacent to the space where the Med-Vizion lamp will be located. Extend the end of the bar hangers to the sides of the ceiling tile and replace the tile with the Med-Vizion lamp back into the ceiling grid by setting the hanger ends onto the ceiling grid. The cut out of the hanger end interlocks with the grid framework. Sheet Metal Screw Hole Hanger Notch Figure 4: Hanger End Tabs Adjustable Bar Hanger End 5. Use sheet metal screws to further anchor the hanger ends to the ceiling grid. As an alternative, the triangular tabs on the hanger ends can be bent over and interconnected with the ceiling grid to secure the hanger ends. Figure 5: Junction Box Cover 6. Remove junction box cover plate, along with desired knockout for the filtered AC mains. www.ets-lindgren.com Installation 13 Figure 6: Non-Ferrous Fitting on Junction Box (not user serviceable) 7. Use EMT conduit/THHN wire and non-ferrous fitting for AC power to the junction box. Do not use any flex metal conduit light drops (whips) to the individual lamps. All AC power from the facility RF filter to each lamp must be in EMT conduit. Figure 7: Junction Box Wiring Connections 8. Wire incoming Hot, Neutral and Ground (3) wires into the junction box and replace cover plate on the junction box. 14 Installation www.ets-lindgren.com Trim Ring Installation Slide the trim ring into the fixture housing and position the spring wire clips within the brackets located on the inside wall of the housing. Note: If ceiling tile thickness is less than 0.75”, modify the spring wire clip by bending it as shown in Figure 8. This will allow for additional space when working with shallow ceiling tiles. Figure 8: Trim Ring Modification EMI Facility Filter Installation (To Be Supplied by Customer) An EMI facility filter is designed to prevent EMI (Electromagnetic Interference) from getting inside the MRI room. The EMI facility filter is NOT supplied as part of the Med-Vizion system components and is not necessary for non-MRI applications. The EMI filter and mounting hardware is supplied by the customer or specified RF shield subcontractor. The EMI filter functionally eliminates electromagnetic interference from entering the room. Mount the EMI facility filter according to approved system layout documentation. The interconnection Class 1 wiring (Lamp-To-Lamp) is customer supplied and must meet local electrical code specifications. Refer to installation wiring diagram for ampacity requirements. www.ets-lindgren.com Installation 15 Load end – RF shielded compartment Line end – punch can to accept conduit Figure 9: EMI Facility RF Filter (ELUL 2030) Figure 10: EMI Facility Filter Installation Procedure 16 Installation www.ets-lindgren.com TO MOUNT RF FILTER TO SHIELD 1. Locate space in penetration panel area to place filter. 2. Using a hole saw that has the same O/D as the chase nipple O/D, gently cut through the copper shielding. Tip: Hand reverse rotate the hole-saw to cut through the copper. Once you have cut through the copper, proceed to cut through the shield panel. 3. Using the hole saw cut through the plywood. 4. Clean off the copper surface where the chase nipple will pass through the shield using Scotch-Brite™ pad or fine steel wool. Do not use sandpaper. 5. Mount filter; install a brass flange nut to one end of the chase nipple. Insert chase nipple through shielded (top compartment). On the inside of the shield secure the chase nipple with second brass flange nut, snug up, do not over tighten. 6. Secure bottom of the filter to the shield using 3/4" lag screws. 7. Make up electrical connections, connect ground wire to shield buss bar only. Digital Dimmer Module Installation (Optional) Note: Each dimmer module can support up to 16 ZXR lamps. Figure 11: Dimmer Input/Output RJ45 Plugs www.ets-lindgren.com Installation 17 Use the following directions if the lighting system will use a digital dimmer module. 1. If dimmer is located outside of the shield: Run a RJ45 cable from the dimmer power supply to the dimmer module. Run a RJ45 cable from the dimmer module to the dimmer module RF filter. Run a RJ45 cable from the shielded compartment of the filter, through the chase nipple, to the first lamp in a lighting circuit. Series connect RJ45 cables from the initial lamp to each successive lamp on the circuit. 2. If the dimmer module is located inside of the shield: Locate the dimmer module power supply outside the shield. Run a RJ45 cable from the power supply to the dimmer module RF filter. Run a RJ45 cable from the RF filter to the dimmer module. Run a RJ45 cable from the dimmer module to the first lamp in a lighting circuit. Series connect RJ45 cables from the initial lamp to each successive lamp on the circuit. 18 Installation www.ets-lindgren.com Appendix A: Glossary of Terms Med-Vizion Light Fixture The mechanical/electrical subassembly comprised of the ceiling plaster frame, LED driver box, and AX mains wiring J-box. See Figure 1 on page 6. Facility AC Power RF Filter An RF filter assembly designed to prevent electromagnetic interference (EMI) from getting inside the MRI room. VAC Provides 120 power directly to each individual ZXR lamp fixture. This filter is made part of the primary RF shield installation and is not supplied with the ZXR lamp system. Med-Vizion System Power Switch The electrical disconnect switch wired into the mains feed VAC to the facility filter that supplies 120 power to the ZXR lighting circuit. Med-Vizion Dimmer Module Electronic controller that sends control information to each ZXR lamp which is proportional to the selected setting of the dimmer module. Med-Vizion Dimmer EMI Communication Filter An RF filter assembly designed to prevent electromagnetic interference (EMI) from entering the MRI room. This filter is supplied as part of the optional Med-Vizion dimmer application. Used for either the dimmer module or the dimmer power supply depending on the location of the dimmer module. Med-Vizion Dimmer Interconnect Cables RJ45/RJ45 interconnecting cables between the dimmer module, dimmer filter, and ZXR lamp fixtures. Dimmer Power Supply Provides 12 www.ets-lindgren.com VDC power to the dimmer module. Glossary of Terms 19 This page intentionally left blank. 20 Glossary of Terms www.ets-lindgren.com Appendix B: Typical Wiring Schematics–ZXR System Non-Dimming Installation The ZXR system will install as a typical interior MRI lighting system. It will utilize a standard EMI facility power filter for each independent lighting circuit. The following schematic depicts a typical single circuit installation. Additional independent lighting circuits will duplicate this schematic. Figure 12: For Non-Dimming System Installation Optional Dimmer Installation The optional dimmer components that are interconnected to the Med-Vizion system are made up of a Remote Dimmer Panel (i.e. Dimmer Module), powered by a 12VDC wall adapter (included), and Dimmer Facility Filter. They are directly wired into the Med-Vizion system as shown in the pages that follow. www.ets-lindgren.com Typical Wiring Schematics–ZXR System 21 Dimmer Module/Power Supply Filter: Included with Dimmer The dimmer EMI communication/power filter function is the same as the facility power filters, in that it is designed to prevent EMI (Electromagnetic Interference) from getting inside the MRI room. Note that this filter is required when: Dimmer module external to the MRI room: Filter needed to provide control information from the dimmer module to each ZXR lamp circuit. Dimmer module internal to the MRI room: Filter needed to provide 12VDC power from the externally installed 12VDC power supply to the dimmer module installed within the MRI room. The same 250747 1amp/240V communication/power filter is used for either of the two dimmer locations detailed above. Dimmer Wiring–External to MRI Room The dimmer wiring arrangement differs depending on whether the control is to be located in the control room or in the MRI suite. 1. If the dimmer control is to be located in the control room, position it at any convenient location using a plastic single-gang switch box, minimum 3’’ deep. 2. Locate the power supply adapter at any convenient nearby 120VAC power outlet. Route RJ45 cable from the power supply to the dimmer control box and route RJ45 cable to the dimmer control box and connect to either connection point on dimmer. 3. Install dimmer filter to the outside of the shield wall and attach install nipple to the opposite side of panel with hex nut. 4. Important note: All RJ45 cables are to use the 568-B configuration. Pull a standard RJ45 wire between the dimmer location and the signal filter located in the equipment room. Since this is a Class 2 circuit, it is not necessary to run conduit for this cable. 5. Connect Female RJ45 pigtail (Red) on the outside of room to the power filter, choose the two wires labeled PWM Dimming Signal. The wire colors should be Green (-) and Green/White (+). Terminate both ring terminals on the corresponding studs on the filter. (Green/white) Line and (Green) Neutral studs. The two remaining wires labeled 12VDC Power will not be used. Repeat the process inside the room using the (blue) RJ45 pigtail connector (provided) to the filter inside the MRI room. Pull a standard Cat 5 patch cord from the RJ45 pigtail connector to the nearest ZXR lamp. Connect all remaining lamps in daisy chain fashion. See Figure 13 on page 23. 22 Typical Wiring Schematics–ZXR System www.ets-lindgren.com Note: Power supply lead and fixture lead can be plugged into either connection point on switch. Figure 13: Dimming System Installation–External to MRI Room Dimmer Wiring–Internal to MRI Room 1. If the dimmer control is to be located within the shield room, position it at any convenient location using a plastic single-gang switch box, minimum 3’’ deep. 2. Locate the power supply adapter external to the MRI room at any convenient nearby 120VAC power outlet. Power supply should be placed as close as possible to the location of the facility filters at the MRI penetration panel. Locating the specific power supply outlet near the magnet’s penetration panel is highly recommended. Route RJ45 patch cable to the power filter plug into the (Red) RJ45 pigtail (provided). Note: The dimmer power supply cannot be installed inside the MRI room. www.ets-lindgren.com Typical Wiring Schematics–ZXR System 23 3. Attach the RJ45 pigtail to the filter; choose the 2 wires Labeled 12VDC Power. The wire colors should be Blue and Blue/White. Terminate both ring terminals on the corresponding studs on the filter. (Blue/white) Line and (Blue) Neutral studs. The two remaining wires labeled PWM Dimming Signal will not be used. Make the same connections on the inside of the filter using the (Blue) RJ45 pigtail; route a patch cord to the dimmer, plug into either RJ-45 connector on the dimmer control. 4. Important note: All RJ45 cables are to use the 568-B configuration. Route a RJ45 cable from dimmer to the nearest ZXR lamp into the dimmer control box. Plug it into RJ-45 connector on back of the dimmer control panel. Connect all remaining lamps daisy chain fashion. See Figure 14 below. Figure 14: Dimming System Installation–Internal to MRI Room 24 Typical Wiring Schematics–ZXR System www.ets-lindgren.com ZXR, LED, Lamp Operation 1. Verify proper lamp operation by pressing the ON button. All lamps should light to full ON. 2. Press the down arrow and the lamps should gradually begin to dim. They will hold the present level when you release the button. 3. To save this setting, press the Save button. The blue windows will flash three times to acknowledge the level is saved. 4. Pressing OFF should extinguish the lamps; Pre-set should restore the saved level; ON should return to full brightness. www.ets-lindgren.com Typical Wiring Schematics–ZXR System 25 This page intentionally left blank. 26 Typical Wiring Schematics–ZXR System www.ets-lindgren.com Appendix C: Warranty Scope and Duration of Warranties Seller warrants to Buyer that the Products to be delivered hereunder will be (1) free from defects in material, manufacturing workmanship, and title, and (2) conform to the Seller’s applicable product descriptions and specifications, if any, contained in or attached to Seller’s quotation. If no product descriptions or specifications are contained in or attached to the quotation, Seller’s applicable product descriptions and specifications in effect on the date of shipment shall apply. The criteria for all testing shall be Seller’s applicable product specifications utilizing factory-specified calibration and test procedures and instruments. All product warranties, except the warranty of title, and all remedies for warranty failures are limited to three years. Product Warranted Duration of Warranty Period Med-Vizion™ MRI Interior LED Lighting System Three Years Any product or part furnished to Buyer during the warranty period to correct a warranty failure shall be warranted to the extent of the unexpired term of the warranty applicable to the repaired or replaced product. The warranty period shall commence on the date the product is delivered to Buyer; however, if Seller assembles the product, or provides technical direction of such assembly, the warranty period for such product shall commence on the date the assembly of the product is complete. Notwithstanding the foregoing, in the event that the assembly is delayed for a total of thirty (30) days or more from the date of delivery for any reason or reasons for which Seller is not responsible, the warranty period for such product may, at Seller’s options, commence on the thirtieth (30th) day from the date such product is delivered to Buyer. Buyer shall promptly inspect all products upon delivery. No claims for shortages will be allowed unless shortages are reported to Seller in writing within ten (10) days after delivery. No other claims against Seller will be allowed unless asserted in writing within thirty (30) days after delivery (or assembly if the products are to be assembled by Seller) or, in the case of alleged breach of warranty, within the applicable warranty period. www.ets-lindgren.com Warranty 27 Warranty Exclusions Except as set forth in any applicable patent indemnity, the foregoing warranties are exclusive and in lieu of all other warranties, whether written, oral, express, implied, or statutory. EXCEPT AS EXPRESSLY STATED ABOVE, SELLER MAKES NO WARRANTY, EXPRESS OR IMPLIED, BY STATUTE OR OTHERWISE, WHETHER OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE OR USE OR OTHERWISE ON THE PRODUCTS, OR ON ANY PARTS OR LABOR FURNISHED DURING THE SALE, DELIVERY OR SERVICING OF THE PRODUCTS. THERE ARE NO WARRANTIES WHICH EXTEND BEYOND THE DESCRIPTION ON THE FACE HEREOF. Warranty coverage does not include any defect or performance deficiency (including failure to conform to product descriptions or specifications) which results, in whole or in part, from (1) negligent storage or handling of the product by Buyer, its employees, agents, or contractors, (2) failure of Buyer to prepare the site or provide an operating environmental condition in compliance with any applicable instructions or recommendations of Seller, (3) absence of any product, component, or accessory recommended by Seller but omitted at Buyer’s direction, (4) any design, specification, or instruction furnished by Buyer, its employees, agents or contractors, (5) any alteration of the product by persons other than Seller, (6) combining Seller’s product with any product furnished by others, (7) combining incompatible products of Seller, (8) interference with the radio frequency fields due to conditions or causes outside the product as furnished by Seller, (9) improper or extraordinary use of the product, or failure to comply with any applicable instructions or recommendations of Seller including maintenance, calibration and cleaning procedures and intervals, or (10) acts of God, acts of civil or military authority, fires, floods, strikes or other labor disturbances, war, riot, or any other causes beyond the reasonable control of Seller. This warranty does not include (1) batteries, (2) cables, (3) gasket, (4) fingerstock, or any item that is designed to be consumable. Seller does not warranty products of others which are not included in Seller’s published price lists. Buyer’s Remedies If Seller determines that any product fails to meet any warranty during the applicable warranty period, Seller shall correct any such failure by either, at its option, repairing, adjusting, or replacing without charge to Buyer any defective or nonconforming product, or part or parts of the product. Seller shall have the option to furnish either new or exchange replacement parts or assemblies. Warranty service shall be performed at the Seller’s factory, or the Buyer’s site at the sole discretion of the Seller. Within the warranty period, the Buyer shall be responsible for all transportation to the Seller’s factory, and the Seller shall be responsible for transportation of goods to the Buyer’s site. 28 Warranty www.ets-lindgren.com Within the contiguous 48 United States, warranty service performed during the applicable warranty period will be performed without charge to Buyer during Seller’s normal business hours. After the warranty period, service will be performed at Seller’s prevailing service rates. Subject to the availability of personnel, after-hours service is available upon request at an additional charge. Outside the contiguous 48 United States, travel and per diem expenses, when required, shall be the responsibility of the Buyer, or End User, whichever is applicable regardless of the warranty period. The remedies set forth herein are conditioned upon Buyer promptly notifying Seller within the applicable warranty period of any defect or non-conformance and making the product available for correction. The preceding paragraphs set forth Buyer’s exclusive remedies and Seller’s sole liability for claims based on failure of the products to meet any warranty, whether the claim is in contract, warranty, tort (including negligence and strict liability) or otherwise, and however instituted, and, upon the expiration of the applicable warranty period, all such liability shall terminate. IN NO EVENT SHALL SELLER BE LIABLE TO BUYER FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OF ANY KIND ARISING OUT OF, OR AS A RESULT OF, THE SALE, DELIVERY, NON-DELIVERY, SERVICING, ASSEMBLING, USE OR LOSS OF USE OF THE PRODUCTS OR ANY PART THEREOF, OR FOR ANY CHARGES OR EXPENSES OF ANY NATURE INCURRED WITHOUT SELLER’S WRITTEN CONSENT DESPITE ANY NEGLIGENCE ON BEHALF OF THE SELLER. IN NO EVENT SHALL SELLER’S LIABILITIES UNDER ANY CLAIM MADE BY BUYER EXCEED THE PURCHASE PRICE OF THE PRODUCT IN RESPECT OF WHICH DAMAGES ARE CLAIMED. This agreement shall be construed in accordance with laws of the State of Texas. In the event that any provision hereof shall violate any applicable statute, ordinance, or rule of law, such provision shall be ineffective to the extent of such violation without invalidating any other provision hereof. Any controversy or claim arising out of or relating to the sale, delivery, non-delivery, servicing, assembling, use or loss of use of the products or any part thereof or for any charges or expenses in connection therewith shall be settled in Austin, Texas by arbitration in accordance with the Rules of the American Arbitration Association, and judgment upon the award rendered by the Arbitrator may be entered in either the Federal District Court for the Western District of Texas or the State District Court in Austin, Texas, all of the parties hereto consenting to personal jurisdiction of the venue of such court and hereby waive the right to demand a jury trial under any of these actions. www.ets-lindgren.com Warranty 29
© Copyright 2024