Industrial X-Ray Chiller Manual XRCA-3001-WA For OEM COMET AG Herrengasse 10, CH-3175 Flamatt Tel. +41 31 744 9000, www.comet-xray.com XRCA-3001-WA, Doc.-No. 50005554 Page 1 of 16 Document history Version 1.0 2.0 3.0 4.0 Date 03/31/2010 07/23/2010 10/30/2012 07/01/2013 07/30/2014 COMET AG Herrengasse 10, CH-3175 Flamatt Tel. +41 31 744 9000, www.comet-xray.com Author M. Schmid St. Haferl R. Moser R. Moser R. Moser Amendment(s) Layout/structure Status Released Layout/structure Update Chapter 2+3 Outline drawing added XRCA-3001-WA, Doc.-No. 50005554 Page 2 of 16 Caution: Read the manual before setting to work! Always keep cooling unit in upright position and do not shock Power supply 230 V 50/60 Hz / 208 V 50/60 Hz! Run cooling unit always with correct coolant level and clean filter regular otherwise damage of pump may occur! Only use cooling hoses with sufficient pressure and coolant resistance! Never operate damaged or leaking equipment! Before starting any service work disconnect the cooling unit from power supply! Never carry out any work at the refrigeration cycle, this work must be carried out at manufacturer site! COMET AG Herrengasse 10, CH-3175 Flamatt Tel. +41 31 744 9000, www.comet-xray.com XRCA-3001-WA, Doc.-No. 50005554 Page 3 of 16 Contents 1. Description ................................................................................................................ 5 2. Technical data ........................................................................................................... 6 3. Performance chart .................................................................................................... 7 4. Setting to work .......................................................................................................... 8 5. Settings ..................................................................................................................... 9 6. Maintenance ........................................................................................................... 12 7. Outline Drawing ...................................................................................................... 13 8. Circuit Diagrams ..................................................................................................... 14 9. Declaration of conformity ........................................................................................ 16 Appendix: Flow diagram and Circuit diagram COMET AG Herrengasse 10, CH-3175 Flamatt Tel. +41 31 744 9000, www.comet-xray.com XRCA-3001-WA, Doc.-No. 50005554 Page 4 of 16 1. Description The cooling unit WLK 32 is designed to provide a continues flow of cooling fluid at a constant temperature. Water circulates between the cooling unit and the heat source. The water is chilled by an air-cooled refrigeration system (refrigerant R 134A).The cooling unit has a main switch to start the refrigeration system. The water level in the reservoir is supervised by a level switch. If the level drops below minimum, the refrigeration cycle stops and the pilot lamp "Water level low" is activated and the safety circuit is activated. The maximum cooling capacity of the cooling unit depends on the ambient and the fluid temperature. The cooling capacity is 3000W related to 40 °C ambient temperature and +25 °C ± 2°C fluid temperature. Cooling hoses are connected to the cooling unit via quick connectors. Water inlet and outlet are marked with symbols: Inlet: O Outlet: O . The water outlet temperature is monitored by an electronic temperature controller. This controller is provided with additional contacts to prevent freezing and for temperature maximum signal. The signals "temperature maximum" and "water level low" are combined in a potential free safety circuit. The water pressure of the pump is limited by an adjustable bypass valve. Note: If the ambient temperature is below the adjusted anti freezing temperature the refrigeration system does not operate! Anti freezing, temperature maximum and bypass valve is factory adjusted according to specification. COMET AG Herrengasse 10, CH-3175 Flamatt Tel. +41 31 744 9000, www.comet-xray.com XRCA-3001-WA, Doc.-No. 50005554 Page 5 of 16 2. Technical data Physical dimensions Length: Width: Height: Weight: Coolant capacity: 590 mm 612 mm 925 mm 115 kg without coolant 14 l Performance data Cooling capacity: Water flow: Voltage rating: Current consumption: Noise level: Airflow at 50Hz (60Hz): Safety class: 3000 W at 40°C ambient temperature 6.0 l/min at 4 bar 230 V / 50/60 Hz (switchable to 208 V / 50/60 Hz) < 9.3 A 65 dB(A), distance 1 m in any direction 1700 m3/h (2200 m3/h) IP 33 Environmental specifications Ambient temperature: Storage temperature: Air humidity: + 5 °C ... + 40°C - 25°C... + 70°C (storage without water) 20 % ... 90 % non condensing Refrigeration cycle Refrigerant: Capacity: R 134A 1.3 kg Settings Maximum pressure Water outlet temperature (T1): Anti freezing (T2): Temperature MAX (T3): Flow switch open: Flow switch close: COMET AG Herrengasse 10, CH-3175 Flamatt Tel. +41 31 744 9000, www.comet-xray.com 6.7 bar +25°C + 5°C +35°C < 4 l/min > 4.2 l/min XRCA-3001-WA, Doc.-No. 50005554 Page 6 of 16 3. Performance chart COMET AG Herrengasse 10, CH-3175 Flamatt Tel. +41 31 744 9000, www.comet-xray.com XRCA-3001-WA, Doc.-No. 50005554 Page 7 of 16 4. Setting to work the cooling unit must be positioned in an upright and free-standing position for sufficient air circulation use locking brake at caster remove top cover connect hoses with cooling unit remove cap of reservoir and fill the reservoir with water about 3 cm below thread For ambient temperatures lower than 10°C use antifreeze. We recommend Ethylenglycol with tap water in a mix of 2:3 check coolant level and refill if necessary Establish electrical connection Note: Operate equipment only with correct coolant level, otherwise danger of pump destruction! switch on the equipment for some minutes for the ventilation of the circuit replace plastic cap of reservoir by applied brass cap mount cover Operation on different mains voltages The installed transformer makes operation on either 208V 50/60Hz or 230V 50/60Hz possible. COMET AG Herrengasse 10, CH-3175 Flamatt Tel. +41 31 744 9000, www.comet-xray.com XRCA-3001-WA, Doc.-No. 50005554 Page 8 of 16 5. Settings Note: The temperature controller and the bypass valve are adjusted according specification. Settings of water outlet temperature, temperature maximum (parameter P1), anti freezing, and bypass valve should not be changed. If it is necessary to change these settings refer to the following: Set maximum pressure: Basic adjustment < 6,7 bar Solve sealing Increasing maxium pressure: Turn set screw clockwise reducing maximum pressure: turn set screw anti-clockwise Set flow switch Unscrew fixing screw of contactor head Scale up +: move contactor head in the direction of the oil flow Scale down -: move contactor head against the direction of the oil flow Screw on fixing screw of contactor head COMET AG Herrengasse 10, CH-3175 Flamatt Tel. +41 31 744 9000, www.comet-xray.com XRCA-3001-WA, Doc.-No. 50005554 Page 9 of 16 Temperature controller Description of indicating LED's: LED 1: LED on: cooling on LED off: "heating" of refrigeration circuit LED 2: LED on: Temperature maximum exceeded LED off: LED 3: LED on: cooling unit ready LED off: Temperature below anti freezing, running of refrigeration system Impossible Water outlet temperature (rated value) T1 Press SET-key and keep it pressed (rated value is shown on display) Release SET-key (actual value is shown on display) Note: Hysteresis of T3 is adjusted at ± 4°C symmetrical to rated value Anti freezing T2 Press keys (UP) and (DOWN) simultaneously for 4 seconds, P1 is shown on display Select to parameter P30 with key (UP) Press SET-key and keep it pressed (value of T2 is shown on display) Note: Hysteresis of anti freezing T2 is adjusted at 4°C. If temperature drops below anti freezing temperature (+5 °C) refrigerating unit runs only if temperature exceeds +9 °C. COMET AG Herrengasse 10, CH-3175 Flamatt Tel. +41 31 744 9000, www.comet-xray.com XRCA-3001-WA, Doc.-No. 50005554 Page 10 of 16 Temperature maximum T3 4 seconds, P1 is shown on display Press SET-key and keep it pressed (value of P1 is shown on display) Note: Hysterese of T3 adjusted at ± 2,5°C symmetrical to parameter P1. Example: Value P1: 32.5°C Temperature maximum T3: fault indication ON: 35.0°C fault indication OFF: 20.0°C Release SET-key (P1 is shown on display) to run-mode (if there is no changing of parameters, the temperature controller returns automatically into run-mode after 45 seconds). COMET AG Herrengasse 10, CH-3175 Flamatt Tel. +41 31 744 9000, www.comet-xray.com XRCA-3001-WA, Doc.-No. 50005554 Page 11 of 16 6. Maintenance In order to achieve maximum cooling capacity keep the condenser of the cooling unit clean. Regularly once a week check this heat exchanger and if necessary clean it: Clean fins of heat exchanger: Remove dust by forcing it out in the opposite direction from which it entered. If compressed air is available, direct the air against the inside of the condenser. The water circuit is an open system. Evaporation of water may occur. Check water level regularly and refill if necessary. Check antifreeze. About every 3 months check the filter of the pump for clean condition, if necessary more often. For this proceed as following: Note: If the filter is not in clean condition damage of pump and motor may occur! Disconnect the cooling unit from mains Remove cover Close stop valve Unscrew filter of the pump and check the filter for impurity Notice: Water will run out of pump. Collect the water with suitable vessel If necessary clean the filter Screw the filter into the pump Check water level and refill if necessary Remove spilled out water Open stop valve Attention: Never operate cooling unit when stop valve is closed, otherwise damage of pump may occur! Connect the cooling unit to the mains and run cooling unit about 10 minutes to deaerate cooling circuit. Check filter for leakage Mount cover of the cooling unit COMET AG Herrengasse 10, CH-3175 Flamatt Tel. +41 31 744 9000, www.comet-xray.com XRCA-3001-WA, Doc.-No. 50005554 Page 12 of 16 7. Outline Drawing COMET AG Herrengasse 10, CH-3175 Flamatt Tel. +41 31 744 9000, www.comet-xray.com XRCA-3001-WA, Doc.-No. 50005554 Page 13 of 16 8. Circuit Diagrams Electrical diagram COMET AG Herrengasse 10, CH-3175 Flamatt Tel. +41 31 744 9000, www.comet-xray.com XRCA-3001-WA, Doc.-No. 50005554 Page 14 of 16 Block diagram COMET AG Herrengasse 10, CH-3175 Flamatt Tel. +41 31 744 9000, www.comet-xray.com XRCA-3001-WA, Doc.-No. 50005554 Page 15 of 16 9. Declaration of conformity COMET AG Herrengasse 10, CH-3175 Flamatt Tel. +41 31 744 9000, www.comet-xray.com XRCA-3001-WA, Doc.-No. 50005554 Page 16 of 16
© Copyright 2024