MTc-Hl-Fs

technical data
High precision Weigh Cell | MTC-HL-FS
technical data
High precision Weigh Cell | MTC-HL-FS
descriptions
FEATURES
OPTIONS
The MTC-HL-FS Weigh Cells cover the weighing range up
SS Weighing range up to 250 g
SS Option 10: RS 422 instead of CAN interface
to 250 g. They have been specifically designed as compact
SS Additive dead load range up to 100 g
SS Option 11: Bus operation
sensor components and are especially suited for applica-
SS Active Vibration Compensation (AVC)
SS Option 12: Higher display resolution
tions in the pharmaceutical industry.
SS High or flat type structured shape
SS Option 13: Binary I/O channels (4-fold)
SS Protection class up to IP 44
SS Option 14: Filling algorithm
The Weigh Cell comes standard equipped to supply final
SS Binary I/O channels with customer-specific functions
SS Option 25: Dead load compensation
weighing values via a CAN interface as ready to connect
SS Interface 1: CAN
SS Option 28: Measuring range changeover
modular components. The integrated software filters can
SS Interface 2: RS 232 for service and configuration
SS Option 32: Built in test weight for high structured
be configured multifunctional, enabling optimal adap-
SS Sampling rate 1 ms
shape (flat structured shape on request)
tation of the weighing system to the respective ambient
conditions. The Weigh Cells extensive command set facilitates simplified control engineering integration.
Complemented by a variety of options as well as different
weighing and dead load ranges, the Weigh Cell can also
fulfill very specific customer requirements in addition to
With the optional CFI (CAN Fieldbus Interface) it is possible
TO BE USED IN
to connect to a variety of industrial fieldbus systems.
Please see specifications on our website.
SS Filling machines for vials, syringes, etc.
SS Packaging machines
SS Capsule and tablet weighing machines
the standard specifications.
MTC-HL-FS systems, in conjunction with weighing range
and structured shape, are always customized and individual.
weight
Concerning track distance and number of tracks, all
Conventional load cell
The structured shapes and variants shown on the following
pages are only exemplary.
Wipotec Weigh Cell
High level load cell
settling time
technical data
all data in summary | MTC-HL-FS
Weigh Cell type
MTC-HL-250/100-FS
Weighing range
* 250 g
4)
Electrically adjustable dead load range
Display value (d)
3)
Display value (d) with higher resolution (Option)
Linearity ≤
1)
Repeatability (S) ≤
50 - 100 g
0.002 g
0.001 g
± 0.002 g
0.001 g
Maximum diameter of weighing platform
40 mm
Settling time (on 1 ‰ of final value)
< 200 ms
1) 2)
Protection class of the Weigh Cell
IP 44
Temperature range
+5 °C to +30 °C
Power supply
24 V DC, ± 5 %
Interface 1
CAN, RS 422
Interface 2
RS 232
Housing material
Aluminum hard coated
1) Depends on the preference settings and on setup and ambient conditions of the system | The absolute repeatability is +/- 3 s
2) Weighing time = settling time plus (adjustable) measuring time
3) Temperature range: +10 °C to +30 °C
4) From weighing platform mounting bolt
* Measuring ranges 100 g / d=0.5 mg and 50 g / d=0.2 mg possible with measuring range changeover in conjunction with higher display resolution
Customer specific variations on request
MTC-FS
Variantenübersicht
technical data
dimensions | MTC-HL-FS
Headquarters
Wipotec GmbH
Adam-Hoffmann-Str. 26
67657 Kaiserslautern, Germany
T +49.631.34146-0
F +49.631.34146-8640
[email protected]
www.wipotec.com
Italy
Wipotec Italia s.r.l.
Via Antonio Gramsci, 18
20016 Pero (MI), Italy
T +39.02.73952424
F +39.02.45508075
[email protected]
www.wipotec.it
USA
Wipotec North America
700 Old Roswell Lakes Parkway,
Suite 200
Roswell, Georgia 30076, USA
T +1.770.971.5414
F +1.770.509.5524
[email protected]
www.wipotec.com
All measurements in mm
Subject to technical modifications
2014-11-EN-MTC-HL-FS