Relaxation times

Institute for Biomedical Engineering
EXCITE
Afternoon Hands-On MRI Sessions: fMRI & DTI
Institute for Biomedical Engineering
Contrast in MRI - Relevant Parameters
Relaxation times:

T1
Spin-lattice relaxation time (longitudinal relaxation time)
Return of spin system to equilibrium state

T2
Spin-spin relaxation time (transverse relaxation time)
Loss of phase coherence due to fluctuations of interacting
spins (‘phase memory time’)

T2*
Decay time of free induction decay
Signal loss due to magnetic field inhomogeneity (difference
in magnetic susceptibility)

ADC Apparent diffusion coefficient
Signal loss due to diffusion of water molecules in an
inhomogeneous magnetic field

k
water exchange rate
Exchange of water between macromolecule bound fraction and
bulk (free) water
Institute for Biomedical Engineering
Relations and Limitations
Sensitivity:
Signal-to-Noise Ratio (SNR)
Spatial
resolution
Temporal
resolution
Signal: magnetization (number of spins, magnetic field strength, …. )
Noise: thermal noise of receiver system, physiological noise, …
Institute for Biomedical Engineering
MRI Contrast
IBT
Institute for Biomedical Engineering
Relaxation times
 MRI delivers good soft tissue contrast
 Tissue specific magnetic parameters for contrast
generation
 T2 / T2*: how fast is signal lost after excitation
 T1: how fast is magnetization gained back after excitation for next
experiments
 Sequence parameters and sequence type determine
contrast
Institute for Biomedical Engineering
The NMR signal
+1.0
Mz
My
Mx
1-exp(-t/T1)
Relaxation
Mi(t)/Meq
Relaxation
0.0
exp(-t/T2*)
-1.0
0
0.5
time (s)
1.0
0
0.5
time (s)
1.0
0
0.5
time (s)
1.0
Institute for Biomedical Engineering
T1 weighting
 Relevant parameters:
 Repetition time (TR) = time between two excitations
 Flip angle -> how much magnetization is left for next excitation
 Strong T1 weighting for large flip angle and short TR
Mz
MzA
θ
T1 Relaxation
during TR
MzB
Mxy
Institute for Biomedical Engineering
T1 weighting: Example




Two metabolites with T1=500ms (blue) and T1=250ms (red)
Flip angle: 60°
Signal proportional to DMz
TR=3000ms
Mz
time
01.04.2010
IBT
8
Institute for Biomedical Engineering
T1 weighting: Example




Two metabolites with T1=500ms (blue) and T1=250ms (red)
Flip angle: 60°
Signal proportional to DMz
TR=300ms
Mz
time
01.04.2010
IBT
9
Institute for Biomedical Engineering
T1 weighting: Example




Two metabolites with T1=500ms (blue) and T1=250ms (red)
Flip angle: 60°
Signal proportional to DMz
TR=100ms
Mz
time
01.04.2010
IBT
10
Institute for Biomedical Engineering
T2 / T2* weighting
 Relevant parameter:
 Echo time (TE) = time between excitations and data acquisition
 Strong T2 weighting for long TE
Mxy
t / ms
TEshort TEmedium
TElong
Institute for Biomedical Engineering
Proton density weighting
 Intensity scales with number of signal generating nuclei per
volume element
 Keep influence of relaxation times small:
 Short TE -> small effect of T2 / T2* on signal
 Long TR -> small effect of T1
Institute for Biomedical Engineering
Functional MRI (fMRI)
IBT
Functional MRI (fMRI)
Institute for Biomedical Engineering
 Uses echo planar imaging (EPI) for fast
acquisition of T2*-weighted images.
 Spatial resolution:
 3 mm
 < 200 μm
(standard 1.5 T scanner)
(high-field systems)
EPI
(T2*)
 Sampling speed:
 1 slice: 50-100 ms
dropout
 Problems:
 distortion and signal dropouts in certain regions
 sensitive to head motion of subjects during scanning
 Requires spatial pre-processing and statistical
analysis.
T1
But what is it that makes T2*
weighted images “functional”?
IBT
14
The BOLD contrast
Institute for Biomedical Engineering
REST
neural activity   blood flow   oxyhemoglobin   T2*   MR signal
ACTIVITY
Source: Jorge Jovicich, fMRIB Brief Introduction to fMRI
Institute for Biomedical Engineering
The temporal properties of the BOLD signal
 sometimes shows
initial undershoot
 peaks after 4-6
secs
 back to baseline
after approx. 30
secs
 can vary between
regions and
subjects
Peak
Brief
Stimulus
Undershoot
Initial
Undershoot
Institute for Biomedical Engineering
MRI and Diffusion
IBT
Institute for Biomedical Engineering
Brownian motion
 Molecules or atoms in fluids and gases move freely
 Collisions with other particles causes trembling movement
 Brownian motion: microscopic random walk of particles in
fluids of gases (R. Brown 1827)
 Brownian motion depends on thermal energy, particle
properties and fluid density
Institute for Biomedical Engineering
Diffusion
 Diffusion: irreversible automatic mixing of fluids (or gases)
due to Brownian motion
 Root mean square displacement depends on diffusion
coefficient D and time t: r  2  D t
(A. Einstein)
 Diffusion coefficient D affected by cell membranes,
organelles, macromolecules (Le Bihan 1995)
Institute for Biomedical Engineering
Anisotropy
 Restrictions on water diffusion usually without spherical
symmetry  anisotropic diffusion in biological tissue
 Diffusion tensor (=3x3-matrix) instead of diffusion
coefficient accounts for anisotropic diffusion in 3D
 Principal diffusion direction: direction with largest diffusion
coefficient
r1
r2
r3
Free Diffusion
Restricted Diffusion
ri  2  Di t , i  1, 2,3
Institute for Biomedical Engineering
 Example: nerve fibre
 Diffusion perpendicular to fibre restricted
 Water diffusion indicates white matter organization
Institute for Biomedical Engineering
Diffusion and MRI
 Diffusion leads to signal loss in MRI
Institute for Biomedical Engineering
Diffusion gradients
 Signal attenuation depends on diffusion coefficient and
gradient waveforms
 GE: sum of diffusion weighting gradients zero
 SE: diffusion weighting gradients have equal area
 Single shot techniques freeze out physical motion
90°
diffusion
gradient
180° diffusion
gradient
TE
EPI readout
Institute for Biomedical Engineering
Diffusion weighted imaging DWI
 b-value (=b-factor) describes diffusion weighting
analogous to TE in T2 weighted sequences
 b-value determined by diffusion weighting gradients (i.e.
gradient form, strength, distance)
signal
S  S0 ebD
0 200 400 600 800 1000
b-factor
[s/mm2]
S0: signal without diffusion weighting;
D: diffusion coefficient in direction of gradient
Institute for Biomedical Engineering
DTI
 Ellipsoid represents diffusion tensor
 Fibre structure via map of diffusion anisotropy: calculate
fractional anisotropy (or relative anisotropy or volume
ratio)
PS
MS
MP
S
ADC
l1
M
P
DWIs + Reference
l2
l3
3D
ellipsoid
FA
Colorcoded FA
Institute for Biomedical Engineering
 Principal diffusion coefficient and vector: longest axis of
diffusion tensor
Institute for Biomedical Engineering
 Brain structures via analysis of principle diffusion vectors
Pons
Medulla
Superior longitudinal
fasciculus
Corticospinal tract
Corpus callosum Tapetum
Middle cerebellar
peduncle
Medulla
Optic radiation
Superior cerebella
peduncle
Institute for Biomedical Engineering
MR Angiography
IBT
Institute for Biomedical Engineering
MR Angiography
Image
Slice
Image
Slice
Mz
imaging
saturation
Blood flow
Gradient echo imaging:
Don’t wait for gradient echo
 Bright signal from
unsaturated spins in slice
Saturation: apply 90°
slice-selective pulse
Stationary spins
Inflowing spins
time
01.04.2010
IBT
29