RSA-PE-18 INTRODUCTION......................................................................................................................1 INSPECTION PROCEDURE ..................................................................................................2 INSTALLATION.......................................................................................................................3 Removing the Lambda 18 Compartments .......................................................................3 Installing the Accessory...................................................................................................4 Optical Alignment............................................................................................................4 Energy Scan .....................................................................................................................6 OPERATING PROCEDURES.................................................................................................7 Background Correction - An Overview...........................................................................7 Measurement Procedure - 8°/Hemispherical Reflectance Factor....................................10 Measurement Procedure - Specular Subtraction..............................................................11 Measurement Procedure - Transmittance ........................................................................11 Measuring Large Samples ...............................................................................................12 OPERATING SUGGESTIONS................................................................................................13 Background Correction - Scan Parameters......................................................................13 Zeroline Correction..........................................................................................................13 Dry Nitrogen Purge..........................................................................................................14 Creating a Data File for Reflectance Standards...............................................................15 Optional Sampling Accessories .......................................................................................16 MAINTENANCE.......................................................................................................................21 General Information.........................................................................................................21 Mirror Cleaning Procedure ..............................................................................................21 TECHNICAL NOTES...............................................................................................................22 Measurement of Transmittance .......................................................................................22 Specular Subtraction ........................................................................................................22 Use of Proper Reflectance Standards...............................................................................23 RECOMMENDED READING.................................................................................................24 AQ-00148-000, Rev. 2 APPENDIX.................................................................................................................................25 Typical 8°/Hemispherical Reflectance Factor for an SRS-99 .........................................25 AQ-00148-000, Rev. 2 INTRODUCTION The RSA-PE-18 is a six-inch (150 mm) diameter integrating sphere accessory to the Perkin-Elmer Lambda 18 spectrophotometer. The integrating sphere is constructed from Labsphere's proprietary Spectralon® material. The accessory allows diffuse reflectance and transmittance measurements of materials. The RSA-PE-18 is essentially an optical bench which includes double beam transfer optics along with the integrating sphere. The integrating sphere provides sample mounting ports for reflectance and transmittance measurements, a reference port for reflectance standards, and a photomultiplier tube (PMT) detector with preamplifier. Optional center-mounted sample holders permit reflectance measurements to be performed for beam incident angles variable from 0° to 70°. This manual is intended to provide installation procedures as well as measurement procedures for the RSA-PE-18 accessory. The "Operator's Manual" of the Lambda 18 should be referenced in conjunction with this manual. Cuvette Holder Position Attenuator Positions Lens Assembly Position Figure 1. Top View of RSA-PE-18 Accessory Showing Placement of Optional Components AQ-00148-000, Rev. 2 1 INSPECTION PROCEDURE The RSA-PE-18 was thoroughly inspected and calibrated before shipping and should be ready to operate after completing the set-up instructions. It is packaged and shipped in a reinforced cardboard carton. Carefully check the assembly after unpacking it for any damage that may have occurred during shipping. If there is any such damage, file a claim immediately with the involved carriers and contact the Labsphere Customer Service Department at (603) 927-4266. AQ-00148-000, Rev. 2 2 INSTALLATION Removing the Lambda 18 Compartments Before the RSA-PE-18 can be installed the normal sample and detector compartments of the Lambda 18 must be removed. The following tools are required: (1) flat-head screwdriver (1) 3 mm hex driver 1. Ensure that the Lambda 18 is powered OFF and disconnected from the line voltage input. 2. Lift the cover to the detector compartment. 3. Detach the gray detector connector, as well as the chrome stepping motor connector, by pulling straight back. The detector connector is often latched into position. To unlatch, use the flat-head screwdriver to push the latching mechanism from right to left. 4. Remove the detector compartment's sheet metal cover by removing the 3 mm hex socket cap screws which hold it in place. Two of these screws are located at the rear of the cover. The third is centrally located at an L-bracket on the lower front of the cover. 5. Fully loosen, but do not remove, the four 3 mm hex socket cap screws located in recesses at the corners of the detector compartment. 6. Grasp the detector compartment firmly and remove by carefully lifting the back end first and then the front. 7. Lift the cover to the sample compartment. 8. Remove the beam path windows by grasping the edges of the magnetic window mounts and pulling straight back. 9. Remove the 10 mm cuvette holders by loosening the eight Phillips screws located at the corners of both holders and then lifting straight up. 10. Fully loosen, but do not remove, the four 3 mm hex socket cap screws located in recesses at the corners of the sample compartment. 11. Remove the sample compartment by grasping firmly and carefully lifting straight up. AQ-00148-000, Rev. 2 3 Installing the Accessory 1. 2. 3. 4. 5. Ensure that the Lambda 18 is powered OFF and disconnected from the line voltage input. Slide the RSA-PE-18 assembly into the vacated compartment area until it sets firmly onto the mounting studs of the Lambda 18 baseplate. The connector access holes will align when the assembly is seated properly. Secure the assembly by tightening the four (3 x 40 mm) captive socket cap screws provided using the 3 mm hex driver. The screws fit through the access holes located on the baseplate of the transfer optics enclosure. Attach the detector connector to the Lambda 18. The PMT connector (the 11-pin connector with a central Teflon sleeve) can be damaged quite easily. Insert this connector gently, ensuring the pins are aligned with their receptor holes. If the center sleeve becomes depressed or bent, the RSA-PE-18 will not initialize the Lambda 18 properly and the PMT function will be impaired. Secure the connector by sliding the latching mechanism from left to right. CAUTION: DO NOT FORCE THE CONNECTOR! Optical Alignment The transfer optics of the RSA-PE-18 were factory pre-aligned prior to shipment. However, it is generally good practice to ensure that misalignment did not occur in transit. 1. 2. 3. 4. AQ-00148-000, Rev. 2 Place one Spectralon standard (or other white reflectance standard) in the sample reflectance port and one in the reference reflectance port. Power ON the Lambda 18. Initialize the operating software. Upon completion of the Lambda 18's self-test routine, the software should display a wavelength of approximately 652 nm and report its status as READY. This indicates that there was sufficient sample beam energy to perform the wavelength calibration during initialization. Go to step 10). If the instrument does not initialize, the message "wavelength calibration failed" appears or the instrument status remains BUSY for an extended period, a misalignment may have occurred. 4 Note: It is also possible that the PMT connector was not properly installed. Power OFF the Lambda 18. Inspect the PMT plug to ensure that none of the pins are depressed into the connector. Re-connect the PMT and power ON the Lambda 18 once again to attempt initialization. 5. 6. 7. 8. Power OFF the Lambda 18 and remove the RSA-PE-18 accessory. Replace the standard sample and detector compartments. Power ON the instrument and initialize the software. Note ignition of lamp D2. Expect to see 652 nm displayed following a proper initialization. Choose single beam operation (reference beam). Open the SLIT width to 5.0 nm. Enter a PMT GAIN value of 0.1 (the minimum value). In Perkin-Elmer's UVCSS software, one types the command SERV to select single beam operation. The PMT GAIN value is set from the SCAN methods page. In UVDM software, simply open the INSTRUMENT and SCAN windows to select single beam operation and to set the PMT GAIN value. CAUTION: DO NOT PROCEED IF UNABLE TO ACCESS SINGLE BEAM MODE AND SET THE PMT GAIN TO THE MINIMUM VALUE. 9. 10. 11. 12. 13. 14. 15. 16. AQ-00148-000, Rev. 2 With the Lambda 18 power remaining ON, remove the standard sample and detector compartments and install the RSA-PE-18 accessory . Open the accessory cover. GOTO a wavelength value of 0 nm. This should produce white light in the instrument beams, visible when placing a piece of white paper in the beam paths. (The sample beam is closest to the front of the Lambda 18 and has two mirrors as transfer optics in the RSA-PE-18.) Remove the plug at the top of the integrating sphere. View the rectangular images at the sample and reference reflectance ports. The images should be approximately centered on the white standards placed at each reflectance port. If not centered, adjust the mirror alignment. Align one mirror at a time. In the sample beam, place a small piece of white paper over the face of the second mirror in the beam path. Adjust the first mirror to align the beam cross-section on the second mirror. Turn only one thumbscrew at a time. Adjust the second mirror of the sample beam to center the image on the sample reflectance port of the integrating sphere. Repeat this procedure for the reference beam. Adjust one mirror at a time. Ensure that the beam is centered on each subsequent mirror until a centered image appears on the sphere port. 5 17. Replace the plug onto the top of the integrating sphere. 18. When alignment is complete, power OFF. Connect the detector if necessary. Close the accessory cover. Repeat the initialization procedure. Energy Scan A single beam energy scan prior to sample measurements is recommended. The energy scan provides an indication that the RSA-PE-18 is functioning properly with the Lambda 18 on which it is installed. The actual energy scan performed on a particular accessory is included in shipment. The results of the following procedure should be similar. Actual results are particular to each RSA-PE-18, the Lambda 18 on which the RSA-PE-18 is operated, and the white standard used in the sphere sample port. 1. 2. 3. 4. Choose single beam operation (sample beam). In Perkin-Elmer's UVCSS software, type SERV to select single beam operation. Instrument parameters are set from the SCAN methods page. In UVDM software, simply open the INSTRUMENT and SCAN windows to select single beam operation and to set the instrument parameters. Place one Spectralon standard (or other white reflectance standard) in the sample reflectance port and one in the reference reflectance port of the integrating sphere. Set the wavelength range to start at 499 nm and end at 498 nm. Set the GAIN value to 8.0 nm. The LIVE DISPLAY should reveal an intensity value of 80.0 I at 499 nm. If the value is too large, return to the SCAN parameters and decrease the GAIN value and return to the LIVE DISPLAY. Conversely if the value is too small, increase the GAIN value. Repeat this procedure until the optimum GAIN value is found to produce the intensity value of 80 (+/- 2). Note: If a SLIT value greater than 2.0 nm or a GAIN value greater than 11.0 is required to obtain 80 I units, check the mirror alignment. If the RSA-PE-18 mirror alignment is satisfactory, it may be that the optics of the Lambda 18 are in need of alignment or at least servicing. See your local Perkin Elmer technical service representative. 5. AQ-00148-000, Rev. 2 With satisfactory SLIT and GAIN settings obtained, run a scan over the wavelength range of 850 nm to 200 nm. The result should look similar to the one in the document that was included with shipment. 6 OPERATING PROCEDURES Background Correction - An Overview Prior to performing either reflectance or transmittance measurements, it is necessary to perform an instrument background correction. This determines a 100% line for measurement. White standard materials need to be placed in both the sample and reference ports of the integrating sphere during background corrections. Calibrated Spectralon® standards provide durable, reproducible working reflectance standards. The calibrated standards are recommended for performing most background corrections. If the standards become soiled, they may be rinsed with distilled water and thoroughly dried. If the standards become badly contaminated, they may be lightly sanded using 220-240 grit emery cloth under a stream of running water. This treatment in no way affects the performance of the material. Typical reflectance data for an SRS-99 calibrated standard appears in Table A of the Appendix. The SRS-99 is not included with the accessory unless it is ordered. For transmittance measurements, both reflectance ports need only be filled with the (2) USRS-99-010 uncalibrated Spectralon standards supplied with the accessory. The background correction produces an absolute 100% T reference. Any reflectance measurement using an integrating sphere is relative to the reference standard used. This means that the 100% R line displayed on the Lambda 18 is incorrect. Display data may even exceed 100% if the sample has a higher absolute reflectance than the reference material. It is recommended that an SRS-99-010 calibrated Spectralon standard or other calibrated reflectance standard be used in the sample port during the background correction. Subsequent sample reflectance data are corrected by the actual values reported in the calibration certificate for the SRS-99-010 standard. The following derivation explains the correction. In a double beam, ratio recording spectrophotometer, the background correction is intended to balance the amount of energy in each beam. During the background correction (Bc), the ratio of the sample beam energy (Es) to the reference beam energy (Er) is recorded into instrument memory: B c= Es Er AQ-00148-000, Rev. 2 7 In a transmittance measurement, the transmittance of the sample (Ts) interrupts the amount of energy in the sample beam, producing the following ratio: EsT s Er The Lambda 18 displays a result which is this ratio divided by the background correction: D ISPLA Y V A LU E= EsT s 1 EsT s Er * = * = Ts Er B c Er Es In a reflectance measurement using what is often referred to as a comparison type integrating sphere, the reflectance of the sample (Rs) placed at the sample port interrupts the sample beam energy and the reflectance of the reference material (Rr) placed in its port interrupts the energy in the reference beam to produce the following ratio: EsR s ErR r The Lambda 18 displays: D ISPLA Y V A LU E= EsR s 1 * = ErR r B c EsR s Er R s * = ErR r Es R r Therefore, the DISPLAY VALUE on the Lambda 18 must be multiplied by the known reflectance of the reference material in order to obtain the actual reflectance of the sample. Please note that the method just described has used the reference port of the integrating sphere for the reference material. The validity of the ratio used to describe the background correction has a strict dependence. Two samples of matched reflectance value must be used in each reflectance port during the background correction. If matched samples are not available (which is often the case over a wide spectral range), an alternative method places the reference material in the sample port during the background correction. A matched white material is not necessary for the reference port. AQ-00148-000, Rev. 2 8 The background ratio becomes: B c= EsR r Er The sample replaces the reference material and the following ratio results: EsR r Er The Lambda 18 displays: D ISPLA Y V A LU E = EsR s Er * 1 EsR s Er R s = * = Bc Er EsR r R r Once again the DISPLAY VALUE is the ratio of the material reflectances. The advantage of this method is that matched samples are not required for the background correction. The method does require that the material placed at the reference port remain in place for the sample measurement. AQ-00148-000, Rev. 2 9 Measurement Procedure - 8°/Hemispherical Reflectance Factor 1. Set all Instrument Parameters as desired. Note: It is recommended that scans be recorded as %T (or %R if available) on the ordinate scale. The SLIT should be set as wide as possible for the best signal-to-noise ratio performance. 2. 3. 4. Perform an instrument Background Correction (review III-A). Place the sample to be measured in the sample port. Obtain the measurement data. Multiply the instrument data by the reflectance of the standard reference material used. For example: DISPLAY VALUE Rr Rs .650 x .985 = .640 (64% R) Note: If the system you ordered included an SRS-99-010, the reference data for this calibrated standard can be converted to a Lambda 18 data file in either UVCSS or UVDM software using DWRITE function (see section IV-F). The Arithmetic functions of the software can then be utilized to multiply spectra. AQ-00148-000, Rev. 2 10 Measurement Procedure - Specular Subtraction The integrating sphere of the RSA-PE-18 is fitted with a removable plug located at a port which corresponds to an 8° angle of reflection (i.e.- the specular direction). The port is adjacent to the sample transmittance port. The port may be fitted with a light trap to exclude the specular (mirror-like) component reflected from a sample. 1. 2. 3. 4. 5. Set all Instrument Parameters as desired. Perform an instrument Background Correction (review III-A). Replace the specular port plug with the light trap. The port features a detaining lever which is keyed to properly secure both the plug and light trap when mounted. Place the sample to be measured in the sample port. Obtain the measurement data. Multiply the instrument data by the reflectance of the standard reference material used. Note: Measurements performed using specular subtraction light traps should always be interpreted with caution. A light trap of finite size is not always effective for the complete extinction of a broadened specular peak. Also, diffuse reflections distributed within a few degrees of specular may inadvertently be excluded from measurements as well. Specular exclusion from diffuse reflectance measurements is required in the colorimetry of glossy sample surfaces as well as in Kubelka-Munk analysis of materials. Measurement Procedure - Transmittance 1. 2. 3. 4. Place a white standard material (USRS-99) in each reflectance port. Set all Instrument Parameters as desired. Perform an instrument Background Correction. Place the sample to be measured in the sample port. Obtain the measurement data. Note: When measuring the transmittance of large sized samples, ensure that the reference beam is not obscured by any part of the sample. Samples up to 150 mm wide and 115 mm high will fit inside the accessory enclosure. AQ-00148-000, Rev. 2 11 Measuring Large Samples The sample reflectance port of the integrating sphere is located on the extreme right-hand edge of the accessory to accommodate the measurement of large objects. The hinged sheet metal enclosure surrounding the port, as well as the port's sample holder, can be removed to obtain access for large samples. 1. 2. 3. Using a 3 mm hex driver, remove the three hex socket cap screws located beneath the enclosure. Using a 0.9 mm hex driver, loosen but do not remove the set screws in the flat plate of the sample holder. Remove the sample holder to expose the port frame. Note: The sample surface should sit flush against the port frame to make a good light seal with the port. If the sample does not make an adequate seal, cover the sample and port with a dark cloth or darken the room. Ambient light from the room or even the Lambda 18 light source compartment can add noise to the spectra or even saturate the PMT within the integrating sphere. To check the quality of the light seal, choose servo slit (PMT region) instrument operation. Turn off the light sources of the Lambda 18 (via software). Increase the PMT gain to a value near 20. GOTO a wavelength less than 860 nm (PMT region). Shine a flashlight around the sample port area with the sample in place. If there is a rapid clicking sound (the slit servo mechanism), there is a light leak. Note the location and improve the light seal. AQ-00148-000, Rev. 2 12 OPERATING SUGGESTIONS Background Correction - Scan Parameters It is suggested that the background correction scan start and end at wavelengths at least 5 nm beyond those desired for sample scans. This reduces scan noise at the extreme ends of spectra obtained with the RSA-PE-18. Therefore, if the measurement range of interest is 300 nm - 850 nm, background correct from 295 nm - 855 nm. Following a background correction, the Lambda 18 allows the scan range to be shortened without requiring a new correction. However, if any other parameters are changed such as scan speed, slit or gain values then a new background correction is required. Zeroline Correction An integrating sphere is sensitive to small-angle scatter from the sample beam coupling optics. The scattered "halo" surrounds the sample port and has the effect of mixing a small proportion of the sphere wall reflectance with that of the sample. This means that the 0% R line displayed on the Lambda 18 can be incorrect. This is especially applicable when using masking techniques to measure very small samples. The sample beam illumination will overfill the mask aperture. The "halo effect" can be measured after a background correction by placing a light trap in the sample reflectance port. The light trap should approximate a sample of 0% reflectance. The actual sample reflectance factor, Rs, relative to the perfect diffuser is computed as follows: R s =R r* SA M PLE -ZERO LIN E 100% -ZERO LIN E Where Rr is the reflectance value of the white standard SAMPLE is the sample data displayed on the Lambda 18 ZEROLINE is the data from the light trap measurement. Note: The zeroline data need only be measured periodically. It should be re-measured after changing Instrument Parameters or reflectance standards. AQ-00148-000, Rev. 2 13 Dry Nitrogen Purge The RSA-PE-18 may be equipped with a brass purge inlet connector located at the rear quarter of the accessory. A dry nitrogen purge of the RSA-PE-18 is recommended if spectra free from absorbance bands due to oxygen and water vapor are desired. Both the Lambda 18 and the RSA-PE-18 should be purged concurrently. The nitrogen source outlet should be fitted with a Y-connector. One branch of the Y-connector should lead to the purge inlet of the Lambda 18 and the other to the inlet on the RSA-PE-18. When purging is to begin, the RSA-PE-18 should be prepared by making sure all the integrating sphere ports are empty, including the center mount access port at the top. The instrument should then be initially purged for 10 minutes at a flow rate of about 10 L/min. After the initial purge, the flow rate may be reduced to 6-8 L/min. When samples are introduced into the RSA-PE-18 compartments, the nitrogen flow should remain on. After the samples are mounted in place, the top port plug replaced, and the accessory cover closed, the instrument should be allowed to purge for 5-10 minutes before the measurement data is acquired. This should be repeated for each new sample. Note: It is recommended that only high purity (>99.999%) dry nitrogen be used to purge the RSA-PE-18. AQ-00148-000, Rev. 2 14 Creating a Data File for Reflectance Standards Reflectance data collected by the Lambda 18 using the RSA-PE-18 must be multiplied by the reported reflectance values of the standard reference material used. Both UVCSS and UVDM software packages feature arithmetic functions to multiply entire spectra. A data file containing the reflectance standard's data which would be recognized by the instrument software has to be created in ASCII format using MS-DOS™ line editor commands or compatible software such as Lotus 1-2-3™, Excel™ or Wordstar™. The most important aspect of the ASCII data file to UVCSS or UVDM is the file format, especially for the header. The best approach is to take an existing data file and convert it to ASCII format using the JCAMP command (the file will have a .DX or .DA filename extension). This file can then be examined and edited by one of the software systems previously identified. It may be desirable to first create a "ready-to-edit" data file in UVCSS or UVDM which would already contain relevant header information and data intervals. For example, run a 100% scan immediately following a background correction with the scan range and data interval initially selected to correspond with the available reference data. Reference data for wavelength intervals smaller than those reported in the accompanying calibration certificate can be determined by interpolation. After editing the file's reflectance data at each wavelength, substituting the values for the reflectance standard, save the file in ASCII for mat (be sure to use the correct .DX or .DA filename extension). The file is now ready to be converted back into the proper data file format for either UVCSS or UVDM. Please consult your software manual or your local Perkin Elmer technical applications specialist for more assistance in editing instrument data files. AQ-00148-000, Rev. 2 15 Optional Sampling Accessories Center Mounted Samples - Variable Angle Reflectance A center mounted sample holder is an optional accessory which places samples at the center of the sphere enabling measurements of hemispherical reflectance as a function of incident angle. The standard holder can accommodate samples up to 50 mm (W) x 31.75 mm (H) x 12 mm (D). See Figure 2. An optional clip style holder is also available for thin flexible samples. 1. 2. 3. 4. 5. 6. 7. AQ-00148-000, Rev. 2 Load the reflectance standard into the jaws with the reflecting surface even with the front edge of the jaws. This will place the front surface at the rotational axis of the sample holder. Tighten the jaws using the thumbscrew at the top. Open the lid of the RSA-PE-18 and remove the top plug. Carefully insert the sample holder into the vacated port. Align the mounting pins in the holder with the holes in the port adapter. Set the angle of incidence by the indicator inscribed on the rotary stage of the sample holder. The 0° location indicates normal incidence (i.e.- the sample face is perpendicular to the beam path). Set all Instrument Parameters as desired. Perform an instrument Background Correction. Replace the reflectance standard with the sample to be measured in the sample holder. Obtain the measurement data. 16 Figure 2. Center mounted Sample Holder CAUTION: Care should be taken when handling the center sample holder. Do not handle the coated surfaces with bare hands. In addition, take care when inserting or removing the center mount from the integrating sphere. The center mount must be removed slowly, lifting straight up to avoid damaging the interior of the integrating sphere. When using the center mount sample holder for the reflectance of samples which also transmit, the transmitted component must be extinguished. This can be accomplished by backing the sample with a light trap or black felt cloth. The specular subtraction port can also be used for center mounted samples. The specular component corresponding to a sample beam incident angle of 16° is excluded by the light trap placed at the specular subtraction port. (Please note that the specular subtraction port excludes the 8° component for sample port measurements, section III-C). At normal incidence (0°) the specular component retroreflects and is excluded inadvertently by the sample beam entrance port of the integrating sphere. Measuring The Reflectance of Very Small Samples - Small Spot Accessory The cross section of the sample beam at the sample reflectance port is approximately 9 mm wide by 19 mm high. Samples surfaces should usually be large enough to fill the entire sample port aperture or at least large enough to fill the beam cross section. For sample surfaces smaller than the beam cross-section, a mask can be used to reduce the port aperture to a size small enough to accommodate any sample. [A. Brunsting, R.S. Hernicz and A.J. Dosmann, "Small Area Measurements of Diffuse Reflectance from 410 to 700 nm," Applied Optics 23(23), 4218-4225, 1984.] AQ-00148-000, Rev. 2 17 Labsphere offers an additional optical assembly for reducing the beam cross-section at the three sample locations on the integrating sphere. The assembly consists of a four position wheel, three of which are fitted with various focal length lenses. The lenses are fused silica to operate over the entire spectral range of the RSA-PE-18. The focal lengths vary according to the sample position on which to reduce the beam size. The RSA-PE-18 also needs to be fitted with an iris diaphragm in each beam to further reduce the beam cross-section. Locate the holes in the baseplate for the two attenuators. See Figures 1 and 3. Align the feet of one attenuator with one set of hole such that when the attenuator is fully open, the adjustment handle points towards the back of the accessory. Carefully tip the accessory to access the bottom panel of the accessory. Place the hex screw in the hole between the two feet of the attenuator and tighten to secure the attenuator. Repeat with the second attenuator. Locate the holes in front of the entrance port of the sphere for the lens holder assembly. See Figures 1 and 3. Align the feet of the lens holder assembly with the holes and position the lens holder assembly such that the base of the assembly faces away from the sphere. Tighten the hex screw from the top of the base plate to secure the lens holder assembly to the accessory. Attenuators Lens Assembly Figure 3. Top View of Accessory With Small Spot Kit AQ-00148-000, Rev. 2 18 T Figure 4. Lens Holder Assembly If your RSA-PE-18 is equipped with the beam reducing optics, the following procedure applies: 1. 2. 3. GOTO a wavelength value of 0 nm. This should produce white light in the instrument beams, visible when placing a piece of white paper in the beam paths. Opening the slit width increases the brightness. Place the sample to be measured in the appropriate sample holder of the integrating sphere. Turn the lens wheel so that the letter indicator at the top of the wheel corresponds to the sample position. Indicator O T C R 4. 5. AQ-00148-000, Rev. 2 Sample Position All, lens, normal beam Transmittance, front of sphere Center Mount Reflectance, rear of sphere Close the sample beam iris diaphragm until the beam cross section is contained within the boundaries of the sample surface (i.e.-no overlap). Adjust the reference beam iris to approximately the same diameter as the sample beam iris. This balances the energy in both beams. 19 Note: The live display of the Lambda 18 can also be used, adjusting the reference beam iris to obtain a reading of 100%T or 0A after first setting the sample beam iris for the sample to be measured. Cuvette Holders for the Sample and Reference Beam Locate the holes in the baseplate in the reference beam path in front of the last mirror. Align the dowels on the reference beam cuvette holder with the holes in the baseplate and secure the assembly. See Figures 1 and 5. Place the sample cuvette in the cuvette holder. The sample beam cuvette holder slides into the transmittance port holder. Cuvette Holder Dowels Figure 5. Cuvette Holder Specular Light Trap Raise the detaining lever and carefully remove the specular port plug. Gently insert the specular light trap and secure with the detaining lever. AQ-00148-000, Rev. 2 20 MAINTENANCE General Information When not in use, the RSA-PE-18 should be stored carefully in a controlled environment. Dust and moisture may adversely affect its performance. PRECAUTIONS • Never disconnect or connect the detector cable with the instrument powered on. • Do not remove the cover to the detector preamplifier electronics located underneath the integrating sphere. • Do not handle the photodetector or mirrors. • Do not allow foreign objects to enter the integrating sphere. • Clean dust and particulate debris out of the sphere using a very gentle air blower. Mirror Cleaning Procedure The transfer optics mirrors should never be touched or handled. However, the mirrors do have a protective magnesium fluoride overcoat to allow cleaning if necessary. The following procedure is recommended: 1. Remove mirror fixture. 2. Spray the mirror with Fantastic Spray Cleaner. 3. Rinse the mirror with distilled water. 4. Repeat until the water sheets off the mirror. 5. Blow dry with a stream of clean nitrogen. 6. Replace mirror fixture. AQ-00148-000, Rev. 2 21 TECHNICAL NOTES Measurement of Transmittance There are three types of transmittance measurements that can be performed using a reflectance spectroscopy accessory; total, diffuse and regular transmittance. All measurements follow a background correction with white reference materials in the reflectance ports. Total transmittance is performed with the sample in the transmittance port and the white standards remaining in the reflectance ports. This includes the light transmitted through the sample in all directions. Diffuse transmittance is performed by replacing the white standard in the port diametrically opposite the sample with a black cavity light trap. This excludes light transmitted in the regular sample beam direction, measuring only light scattered through the sample. Regular transmittance is that portion transmitted through the sample without scattering. Regular transmittance is calculated by subtracting the diffuse transmittance from the total. Specular Subtraction The specular component or gloss refers to the mirror-like reflection from an otherwise diffuse surface. This property of a surface is responsible for the degree to which reflected high lights or the images of objects may be seen as superimposed on the surface. In color measurement, the gloss only reflects back the color of the illuminating light source without any color information about the bulk material. In integrating sphere measurements of reflectance, two techniques for subtracting the specular or gloss component are used. The sample beam can be incident at an angle slightly off the normal to the sample port. A light trap can then be positioned at another port across the sphere in the specular direction. In the other technique, the sample beam can be incident normal to the sample surface. The specular component then retroreflects back out of the sphere through the sample beam entrance port. This can make for a simplified and more efficient integrating sphere design; fewer ports are required in the sphere wall. AQ-00148-000, Rev. 2 22 Use of Proper Reflectance Standards The use of proper standards is extremely important in obtaining consistent results in any reflectometer. The term reflectance factor is often used when reporting reflectometer values. The difference between reflectance factor and reflectance lies in the selection of the reference. For the reflectance factor, the perfect reflecting diffuser is used as the reference. The reference for the reflectance is the incident flux. In measurements with any biconical geometry, the reflectance factor can exceed unity (100% R), whereas the reflectance cannot. In any reflectometer, determination of the reflectance can be approximated by measuring the incident flux with a standard of known reflectance. It is also important to choose a standard with similar geometric reflectance characteristics to the samples being measured. This preserves the reflective geometry of the reflectometer between reference and sample readings. Specular standards such as the NIST SRM 2003X or NIST SRM 2011 first surface mirror should be used for measuring the reflectance of specular surfaces in integrating spheres. Samples mounted on the surface behave as part of the sphere. Surfaces that require specular standards would include mirrors as well as polished glass or metal surfaces. Most other samples measured for reflectance will be predominantly diffuse in character. These samples would include most paints, coatings, textiles, ceramics, and natural materials. Recommended standards for measurement of diffuse reflectance include Labsphere's Spectralon™ and Infragold™ Reflectance Standards. Samples which are neither completely specular nor diffuse are more difficult to adequately measure. Sometimes it may be desirable to compare these surfaces to both a specular and a diffuse standard to obtain an average result. AQ-00148-000, Rev. 2 23 RECOMMENDED READING Clarke, F.J.J. and J.A. Compton, "Correction Methods for Integrating Sphere Measurements of Hemispherical Reflectance", Color Research and Application, 11 (4), 253-262, 1986. Erb, W. and W. Budde, "Properties of Standard Materials for Reflection", Color Research and Application, 4 (3), 113-118, 1979. Goebel, D.G., "Generalized Integrating Sphere Theory", Applied Optics, 6 (1), 125-128, 1967. Roos, A., C.G. Ribbing, and M. Bergkvist, "Anomalies in Integrating Sphere Measurements on Structured Samples", Applied Optics, 27 (18), 3828-3832, 1988. Springsteen, A.W., "Physical and Reflectance Properties of Spectralon Diffuse Reflectance Materials," Labsphere Technical Note #4, 1989. Weidner V.R. and J.J. Hsia. "Reflectance Properties of Pressed Polytetrafluoroethylene Powders," J.O.S.A., 71(7) 856-861,1981. Weinder, V. , Technical Document 250-8, "Spectral Reflectance Measurements"; U.S. Dept. of Commerce Riblicalver, 1986. Zwinkels, J.C., "Errors in Colorimetry Caused by the Measuring Instrument", Textile Chemist and Colorist, 21 (2), 23-29, 1989. AQ-00148-000, Rev. 2 24 APPENDIX Typical 8°/Hemispherical Reflectance Factor for an SRS-99 Wavelength (nm) 250 300 400 500 600 700 800 900 AQ-00148-000, Rev. 2 Reflectance .973 .984 .991 .991 .992 .992 .991 .991 25
© Copyright 2024