Guidelines for Using Lyophilized TaqMan® Assays with Master Mix For 7500/7300 System users To run an experiment on your real-time PCR instrument using strip-tubes, one of the following precision holders is required. Table 1. 7300/7500 System Precision Holders Instrument Precision Holders Part No. 7300 and 7500 7500 Fast Precision Plate Holder for 0.2ml tubes and strips, for 7300 and 7500 Systems Precision Plate Holder for 0.1ml tubes and strips, for 7500 Fast System Reference Image 4367033 4403809 Steps: 1. Open the real-time PCR instrument tray and remove the “Precision Plate Holder for plates” from the instrument tray if necessary. 2. Place the “Precision Plate Holder for tubes and strips” (refer to Table 1) in the instrument tray as shown in Figure 1.* Step 1: Open tray and remove holder for plates. Step 2: Insert the holder for strip-tubes. Figure 1. Preparing the real-time PCR system for Lyophilized TaqMan® Gene Expression Assays with Master Mix products. For instructions on how to prepare experiments in strip-tubes, please refer to the “Lyophilized TaqMan® Gene Expression Assays with Master Mix” experiment protocol provided in either the Fast or Standard format product inserts. _______________________________________________________________ * If the adaptor does not fit your PCR instrument, please contact technical support, as some older model instruments are not compatible. http://www.appliedbiosystems.com/support/contact/ For StepOne™ and StepOnePlus™ System users To run an experiment on your real-time PCR instrument using strip-tubes, one of the following adaptor trays is needed. Table 2. StepOne™/StepOnePlus™ System Accessories Instrument Adaptor Trays Part No. StepOne™ StepOnePlus™ StepOne™ MicroAmp® Fast 48-Well Tray Reference Image 4375282 StepOnePlus™ MicroAmp® 96-Well Tray for VeriFlex™ Blocks 4379983 Use only Fast consumables (reaction plates, strip-tubes, and tubes) with the StepOne™ and StepOnePlus™ Systems. Steps: 1. Open the real-time PCR instrument tray. 2. If you have been running plates, place the adaptor tray (refer to Table 2) directly on top of the tray of the real-time PCR instrument in the correct orientation (A1 well position on the top left) as noted in Figure 2. Step 1: Open the tray. Step 2: Place strip-tube adaptor directly on top. Figure 2. Preparing the real-time PCR system for Lyophilized TaqMan® Gene Expression Assays with Master Mix products. For instructions on how to prepare experiments in strip-tubes, please refer to the “Lyophilized TaqMan® Gene Expression Assays with Master Mix” experiment protocol provided in either the Fast or Standard format product inserts. For ViiA™ 7 System users To run an experiment on the ViiA™ 7 Real-Time PCR instrument using strip-tubes, the following trays and adaptors are needed. You can find product details for the Retainer Trays on our website, while ViiA™ 7 adaptors are only bundled with the ViiA™ 7 instruments. Table 3| ViiA™ 7 Trays and Adaptors Instrument ViiA™ 7 Standard 96-well Block ViiA™ 7 Fast 96-well Block Trays and Adaptor Part No. Retainer Tray for Standard 96-well (0.2mL) tubes and strips 4381850 ViiA™ 7 96-Well Tube Adaptor 4459845 Retainer Tray for Fast 96-well (0.1mL) tubes and strips 4379983 ViiA™ 7 Fast 96-Well Tube Adaptor 4462078 Steps: 1. Touch to eject the tray arm. 2. When the tray arm opens, pull the latch, then lift and remove the plate adaptor (Figure 3, Steps 2 and 3). 3. Replace with the appropriate Tube Adaptor (see Table 3). 4. Cap and place the strip-tubes in the corresponding Retainer Tray (Table 3) and place the Retainer Tray with strip-tubes on the Tube Adaptor (See “Loading 0.2mL Strip-Tubes on the Retainer Tray” section on the following page). 5. Make sure “A1” of the Tube Adaptor aligns with “A1” of the Retainer Tray. 6. Touch again to close the tray arm. Steps 2 & 3: Replace the Plate Adaptor Steps 4 & 5: Place the Retainer Tray on Adaptor (Retainer tray in picture is for 0.2mL tube strips) Figure 3. Preparing the ViiA™ 7 Real-Time PCR System for Lyophilized TaqMan® Gene Expression Assays with Master Mix strip-tubes. Loading 0.2mL Strip-Tubes on the Retainer Tray The Retainer Tray for Standard 96-well (0.2mL) tubes and strips (Part No. 4381850) is comprised of two parts; a top unit and a bottom unit. Steps: 1. 2. 3. 4. Snap open the top unit of the Retainer Tray (Figure 4, Step 1). Put the un-capped 0.2mL strip-tubes into the bottom unit of the retainer tray. Snap the top unit of the retainer tray back on top (over the un-capped strip-tubes). Cap the strip-tubes, the optical strip-tube caps should be on top of the top unit of the Retainer Tray. Step 1: Snap open the top unit of the Retainer Tray. Steps 2 to 4: Assemble the tube strip with the Retainer Tray. Figure 4. Preparing and loading Lyophilized TaqMan® Gene Expression Assays with Master Mix strip-tubes into the Retainer Tray. For Research Use Only. Not intended for any animal or human therapeutic or diagnostic use. © 2011 Life Technologies Corporation. All rights reserved. The trademarks mentioned herein are the property of Life Technologies Corporation or their respective owners. TaqMan is a registered trademark of Roche Molecular Systems, Inc. TAwMMG 3.2011 v3
© Copyright 2024