RNA CVC 123 Calibration Verification Controls

RNA CVC 123
Calibration Verification Controls
How to Locate a Calibration Control Lot
1.
Select Bookmark tab on left:
2.
Locate RNA Kit #:
x Select appropriate kit
x Please note specific sheets for mmHg vs kPa, or the H4 or GL cartridge
3.
To Find a Specific Level and Lot:
x Open file by selecting the “+” icon
x The expanded list of control levels will display:
IRMA TruPoint®
Calibration Verification Control
For use with the RNA Medical, Inc.
CVC 123 Calibration Verification Controls
RNA Medical CVC 123
Calibration Verification
Control
Lot 212319
Level 1 Lot 14116
Level 2 Lot 11522
Level 3 Lot 11621
Level 4 Lot 11722
Level 5 Lot 14217
EXP. 11/14
For use with CC, BG, and H3 Cartridges
Expected Values Chart *
LEVEL
1
2
3
4
5
LEVEL
1
2
3
4
5
LOT
NUMBER
14116
11522
11621
11722
14217
LOT
NUMBER
14116
11522
11621
11722
14217
pH
pCO2 mmHg
pO2 mmHg
6.842 – 6.902
7.126 – 7.186
7.408 – 7.468
7.606 – 7.666
7.849 – 7.909
80.1 – 104.1
64.1 – 80.1
36.2 – 46.2
16.4 – 26.4
6.2– 16.2
36.1 – 66.1
69.9 – 89.9
104.4 – 124.4
147.1 – 167.1
425.2 – 545.2
Na+ mM
K+ mM
iCa mM
84.8 – 94.8
107.6 – 117.6
128.0 – 138.0
155.2 – 165.2
164.1 – 174.1
9.16 – 12.16
1.44 – 2.44
3.83 – 4.83
5.76 – 6.76
1.19 – 2.19
2.15 – 3.15
1.27 – 1.47
0.97 – 1.17
0.43 – 0.63
ORL**
The values on the Expected Values Chart are based on the results of multiple
sample analysis. Samples were drawn with a 3cc syringe, 15 gauge plastic
needle tip, after equilibration at 22°C (72°F) for a minimum of 8 hours.
Use the expected values for each parameter as a guide in evaluating
performance. Since performance is subject to sample temperature and
environmental barometric pressure, ITC recommends that each institution
establish its own expected values and acceptable limits. The mean values
established at your institution should fall within the expected ranges.
*NOTE: These values are for use with the IRMA TruPoint only.
**ORL = Out of Range Low
Calibration Verification Control 123 212319_CC 1211
IRMA TruPoint®
Calibration Verification Control
For use with the RNA Medical, Inc.
CVC 123 Calibration Verification Controls
RNA Medical CVC 123
Calibration Verification
Control
Lot 212319
Level 1 Lot 14116
Level 2 Lot 11522
Level 3 Lot 11621
Level 4 Lot 11722
Level 5 Lot 14217
EXP. 11/14
For use with Glucose (GL) Cartridges
Expected Values Chart *
LEVEL
LOT NUMBER
Na+mM
K+mM
Cl-mM
1
2
3
4
5
14116
11522
11621
11722
14217
85.7 - 95.7
107.3 - 117.3
128.4 - 138.4
155.8 – 165.8
164.2 – 174.2
8.95– 11.95
1.54– 2.54
3.79 – 4.79
5.63 – 6.63
1.25 – 2.25
52.7 – 62.7
64.3 – 74.3
86.9 – 96.9
120.2 – 130.2
134.0 – 144.0
Glucose
mg/dL
375- 475
50 – 80
163 – 203
256 – 316
ORL**
The values on the Expected Values Chart are based on the results of multiple
sample analysis. Samples were drawn with a 1mL syringe, 15 gauge plastic
needle tip, after equilibration at 22°C (72°F) for a minimum of 8 hours.
Use the expected values for each parameter as a guide in evaluating
performance. Since performance is subject to sample temperature and
environmental barometric pressure, ITC recommends that each institution
establish their own expected values and acceptable limits. The mean values
established at your institution should fall within the expected ranges.
*NOTE: These values are for use with the IRMA TruPoint only.
**ORL = Out Of Range Low
Calibration Verification Control 123 212319_GL 0112.doc1014
IRMA TruPoint®
Calibration Verification Control
For use with the RNA Medical, Inc.
CVC 123 Calibration Verification Controls
For use with H4 Cartridges
RNA Medical CVC 123
Calibration Verification
Controls
Lot 212319
Level 1 Lot 14116
Level 2 Lot 11522
Level 3 Lot 11621
Level 4 Lot 11722
Level 5 Lot 14217
EXP. 11/14
Expected Values Chart *
LEVEL
LOT NUMBER
Na+mM
K+mM
CI-mM
1
2
3
4
5
14116
11522
11621
11722
14217
84.7 – 94.7
110.2 – 120.2
127.3 - 137.3
155.7 – 165.7
164.9 – 174.9
9.40 – 12.40
1.58 - 2.58
3.63 - 4.63
5.37 - 6.37
1.24 - 2.24
59.9 -69.9
68.6 - 78.6
85.5 – 95.5
108.8 – 118.8
119.2 – 129.2
BUN
mg/dL
ORL**
7 - 13
27 - 37
46 - 62
ORL**
The values on the Expected Values Chart are based on the results of multiple
sample analysis. Samples were drawn with a 3cc syringe, 15 gauge plastic
needle tip, after equilibration at 22°C (72°F) for a minimum of 8 hours.
Use the expected values for each parameter as a guide in evaluating
performance. Since performance is subject to sample temperature and
environmental barometric pressure, ITC recommends that each institution
establish their own expected values and acceptable limits. The mean values
established at your institution should fall within the expected ranges.
*NOTE: These values are for use with the IRMA TruPoint only.
**ORL = Out Of Range Low
CVC 123_H4_212319_1114
IRMA TruPoint®
Calibration Verification Control
For use with the RNA Medical, Inc.
CVC 123 Calibration Verification Controls
For use with Lactate (LA) Cartridges
RNA Medical CVC 123
Calibration Verification
Control
Lot 212319
Level 1 Lot 14116
Level 2 Lot 11522
Level 3 Lot 11621
Level 4 Lot 11722
Level 5 Lot 14217
EXP. 11/14
Expected Values Chart *
LEVEL
LOT NUMBER
1
2
3
4
5
14116
11522
11621
11722
14217
Lactate
mmol/L
ORH**
0.5 – 1.1
2.2 – 3.2
5.6 – 8.6
ORH**
The values on the Expected Values Chart are based on the results of multiple
sample analysis. Samples were drawn with a 1mL syringe, 15 gauge plastic
needle tip, after equilibration at 22°C (72°F) for a minimum of 8 hours.
Use the expected values for each parameter as a guide in evaluating
performance. Since performance is subject to sample temperature and
environmental barometric pressure, ITC recommends that each institution
establish their own expected values and acceptable limits. The mean values
established at your institution should fall within the expected ranges.
*NOTE: These values are for use with the IRMA TruPoint only.
**ORH= Out Of Range High
Calibration Verification Control 123_LA_212319_1114.doc
IRMA TruPoint®
Calibration Verification Control
For use with the RNA Medical, Inc.
CVC 123 Calibration Verification Controls
RNA Medical CVC 123
Calibration Verification
Control
Lot 212320
Level 1 Lot 24117
Level 2 Lot 21523
Level 3 Lot 21622
Level 4 Lot 21723
Level 5 Lot 24219
EXP. 05/15
For use with CC, BG, and H3 Cartridges
Expected Values Chart *
LEVEL
1
2
3
4
5
LEVEL
1
2
3
4
5
LOT
NUMBER
24117
21523
21622
21723
24219
LOT
NUMBER
24117
21523
21622
21723
24219
pH
pCO2 mmHg
pO2 mmHg
6.848 – 6.908
7.136 – 7.196
7.409 – 7.469
7.616 – 7.676
7.850 – 7.910
79.5 – 103.5
62.8 – 78.8
37.9 – 47.9
16.2 – 26.2
6.2– 16.2
28.4 – 58.4
66.1 – 86.1
102.5 – 122.5
147.9 – 167.9
425.5 – 545.5
Na+ mM
K+ mM
iCa mM
85.2 – 95.2
107.6 – 117.6
129.9 – 139.9
158.1 – 168.1
164.8 – 174.8
9.19 – 12.19
1.48 – 2.48
3.86 – 4.86
5.88 – 6.88
1.23 – 2.23
2.17 – 3.17
1.28 – 1.48
0.99 – 1.19
0.43 – 0.63
ORL**
The values on the Expected Values Chart are based on the results of multiple
sample analysis. Samples were drawn with a 3cc syringe, 15 gauge plastic
needle tip, after equilibration at 22°C (72°F) for a minimum of 8 hours.
Use the expected values for each parameter as a guide in evaluating
performance. Since performance is subject to sample temperature and
environmental barometric pressure, ITC recommends that each institution
establish its own expected values and acceptable limits. The mean values
established at your institution should fall within the expected ranges.
*NOTE: These values are for use with the IRMA TruPoint only.
**ORL = Out of Range Low
IRMA TruPoint®
Calibration Verification Control
For use with the RNA Medical, Inc.
CVC 123 Calibration Verification Controls
For use with Glucose (GL) Cartridges
RNA Medical CVC 123
Calibration Verification
Controls
Lot 212320
Level 1 Lot 24117
Level 2 Lot 21523
Level 3 Lot 21622
Level 4 Lot 21723
Level 5 Lot 24219
EXP. 05/15
Expected Values Chart *
LEVEL
LOT NUMBER
Na+mM
K+mM
Cl-mM
1
2
3
4
5
24117
21523
21622
21723
24219
86.2 – 96.2
107.1 – 117.1
129.3 – 139.3
156.3 – 166.3
162.5 – 172.5
8.66 – 11.66
1.51 - 2.51
3.72 - 4.72
5.54 - 6.54
1.26 - 2.26
51.9 -61.9
63.3 - 73.3
86.3 – 96.3
117.6 – 127.6
128.9 – 138.9
Glucose
mg/dL
356- 456
56 – 86
166 – 206
264 – 324
ORL**
The values on the Expected Values Chart are based on the results of multiple
sample analysis. Samples were drawn with a 1mL syringe, 15 gauge plastic
needle tip, after equilibration at 22°C (72°F) for a minimum of 8 hours.
Use the expected values for each parameter as a guide in evaluating
performance. Since performance is subject to sample temperature and
environmental barometric pressure, ITC recommends that each institution
establish their own expected values and acceptable limits. The mean values
established at your institution should fall within the expected ranges.
*NOTE: These values are for use with the IRMA TruPoint only.
**ORL = Out Of Range Low
Calibration Verification Control 123 Glucose GL_212320_0515
IRMA TruPoint®
Calibration Verification Control
For use with the RNA Medical, Inc.
CVC 123 Calibration Verification Controls
For use with H4 Cartridges
RNA Medical CVC 123
Calibration Verification
Controls
Lot 212320
Level 1 Lot 24117
Level 2 Lot 21523
Level 3 Lot 21622
Level 4 Lot 21723
Level 5 Lot 24219
EXP. 05/15
Expected Values Chart *
LEVEL
LOT NUMBER
Na+mM
K+mM
CI-mM
1
2
3
4
5
24117
21523
21622
21723
24219
84.8 – 94.8
106.9 – 116.9
128.6 – 138.6
154.0 – 164.0
159.3 – 169.3
9.63 – 12.63
1.61 - 2.61
3.91 - 4.91
5.71 - 6.71
1.22 - 2.22
58.1 -68.1
69.7 – 79.7
85.6 – 95.6
112.7 – 122.7
121.5 – 131.5
BUN
mg/dL
ORL**
6 - 12
28 - 38
47 - 63
ORL**
The values on the Expected Values Chart are based on the results of multiple
sample analysis. Samples were drawn with a 3cc syringe, 15 gauge plastic
needle tip, after equilibration at 22°C (72°F) for a minimum of 8 hours.
Use the expected values for each parameter as a guide in evaluating
performance. Since performance is subject to sample temperature and
environmental barometric pressure, ITC recommends that each institution
establish their own expected values and acceptable limits. The mean values
established at your institution should fall within the expected ranges.
*NOTE: These values are for use with the IRMA TruPoint only.
**ORL = Out Of Range Low
CVC 123_H4_212320_0515
IRMA TruPoint®
Calibration Verification Control
For use with the RNA Medical, Inc.
CVC 123 Calibration Verification Controls
For use with Lactate (LA) Cartridges
RNA Medical CVC 123
Calibration Verification
Control
Lot 212320
Level 1 Lot 24117
Level 2 Lot 21523
Level 3 Lot 21622
Level 4 Lot 21723
Level 5 Lot 24219
EXP. 05/15
Expected Values Chart *
LEVEL
LOT NUMBER
1
2
3
4
5
24117
21523
21622
21723
24219
Lactate
mmol/L
ORH**
0.6 – 1.2
2.3 – 3.3
5.8 – 8.8
ORH**
The values on the Expected Values Chart are based on the results of multiple
sample analysis. Samples were drawn with a 1mL syringe, 15 gauge plastic
needle tip, after equilibration at 22°C (72°F) for a minimum of 8 hours.
Use the expected values for each parameter as a guide in evaluating
performance. Since performance is subject to sample temperature and
environmental barometric pressure, ITC recommends that each institution
establish their own expected values and acceptable limits. The mean values
established at your institution should fall within the expected ranges.
*NOTE: These values are for use with the IRMA TruPoint only.
**ORH= Out Of Range High
Calibration Verification Control 123_LA_212320
IRMA TruPoint®
Calibration Verification Control
For use with the RNA Medical, Inc.
CVC 123 Calibration Verification Controls
RNA Medical CVC 123
Calibration Verification
Control
Lot 312321
Level 1 Lot 34118
Level 2 Lot 31525
Level 3 Lot 31624
Level 4 Lot 31725
Level 5 Lot 34220
EXP. 08/16
For use with CC, BG, and H3 Cartridges
Expected Values Chart *
LEVEL
1
2
3
4
5
LEVEL
1
2
3
4
5
LOT
NUMBER
34118
31525
31624
31725
34220
LOT
NUMBER
34118
31525
31624
31725
34220
pH
pCO2 mmHg
pO2 mmHg
6.830 – 6.890
7.128 – 7.188
7.408 – 7.468
7.614 – 7.674
7.846 – 7.906
84.4 – 108.4
64.8 – 80.8
37.2 – 47.2
16.4 – 26.4
6.8– 16.8
24.8 – 54.8
63.6 – 83.6
99.4 – 119.4
143.8 – 163.8
416.2 – 536.2
Na+ mM
K+ mM
iCa mM
85.0 – 95.0
107.6 – 117.6
129.1 – 139.1
156.6 – 166.6
163.3 – 173.3
9.12 – 12.12
1.48 – 2.48
3.86 – 4.86
5.80 – 6.80
1.21 – 2.21
2.12 – 3.12
1.25 – 1.45
0.98 – 1.18
0.42 – 0.62
ORL**
The values on the Expected Values Chart are based on the results of multiple
sample analysis. Samples were drawn with a 3cc syringe, 15 gauge plastic
needle tip, after equilibration at 22°C (72°F) for a minimum of 8 hours.
Use the expected values for each parameter as a guide in evaluating
performance. Since performance is subject to sample temperature and
environmental barometric pressure, ITC recommends that each institution
establish its own expected values and acceptable limits. The mean values
established at your institution should fall within the expected ranges.
*NOTE: These values are for use with the IRMA TruPoint only.
**ORL = Out of Range Low
Calibration Verification Control 123 10-25-10_CC Template