How to source continuous pulses without any measurement

Agilent B2901/02/11/12A Precision Source/Measure Unit
How to source continuous pulses without any measurement
This material shows how to setup the instrument to source continuous pulses without any measurement using the
B2901/02/11/12A, through example to source 2 V continuous pulses with the timing parameters such as 300 us pulse
delay, 200 us pulse width and 500 us source trigger period.
Figure 1 shows the timing chart for the continuous pulse source.
TPD : Pulse Delay TPW : Pulse Width
V or I
TSP
Pulse
Peak Value
TSP : Source Trigger Period
TSP
TSP
Source Value
Output
Off State
TPD
Output ON
TPW
TPD
TPW
TPD
TPW
Time
TIMER Trigger
Figure 1. Timing chart for the continuous pulse source
It is required to abort the instrument when you’d like to stop sourcing continuous pulses.
This material also shows how to abort the instrument to stop continuous pulses.
Setting up the pulse condition
Step 1. Press
source.
to edit the source function, and then select
(1) Press Mode
to set the source function to the voltage
(2) Press VOLTS (V)
Agilent B2901/02/11/12A Precision Source/Measure Unit
Step 2.
Press
to edit the source value, and then enter 0 V to set the source value to 0 V.
(1) Press Source
Step 3.
Press
(2) Enter Source Value
to edit the limit value, and then enter 10 mA to set the limit value to 10 mA.
(1) Press Limit
Step 4.
Press
(2) Enter Limit Value
to change the keys shown in Assist keys, and then press
(1) Press Show Pulse
Step 5.
Press
, then press
(2) Pulse Sub-Panel appears
to set the pulse source ON. After turning on the pulse source, you can
see Source Shape which shows the pulse source mode.
(1) Press ON
to show Pulse Sub-Panel.
Source Shape
Agilent B2901/02/11/12A Precision Source/Measure Unit
Step 6.
Rotate
to select Channel 1 Pulse Parameters and set them up as below.
(Peak: 2 V, Delay: 300 us, Width: 200 us)
Step 7.
Press
to show Trigger Sub-Panel.
(1) Press Show Trigger
Step 8.
Press
to edit the trigger type, and then select
(1) Press
(2) Trigger Sub-Panel appears
to set the trigger type to MANUAL.
(2) Press MANUAL
Agilent B2901/02/11/12A Precision Source/Measure Unit
Step 9.
Rotate
to select the source trigger period and press
to edit it. Then enter 500 us to set the
source trigger period to 500 us.
(1) Rotate and press
Source Trigger Period
Step 10. Rotate
to edit
(2) Enter Source Trigger Period
to select the source trigger source and press
to edit it. Then press
the source trigger source to TIMER.
(1) Rotate and press
Source Trigger Source
to edit
Step 11. If you aren’t on the top of the Function menu, press
Step 12. Press
(2) Press TIMER
repeatedly to return to the top level.
On the top level of the Function menu
Any tree structure can’t be seen
In the middle level of the Function menu
Some tree structure can be seen
and then press
to open Trigger Configuration dialogue.
(1) Press Trigger
(2) Press Config
to set
Agilent B2901/02/11/12A Precision Source/Measure Unit
Step 13. Press
and select
to specify the layer, the trigger parameters of which are configured.
(1) Press
Step 14. Rotate
(2) Press ARM
to select the arm count and press
to edit it. Then press
count to INFINITY.
(1) Rotate and press
ARM Count
Step 15. Press
to edit
(2) Press INF.
to make the modification effective.
(1) Press OK to make the modification effective
to set the arm
Agilent B2901/02/11/12A Precision Source/Measure Unit
Starting sourcing the continuous pulses
Step 16. If you aren’t on the top of the Function menu, press
repeatedly to return to the top level.
On the top level of the Function menu
Any tree structure can’t be seen
In the middle level of the Function menu
Some tree structure can be seen
Step 17. Press
,
,
, and then press
continuous pulses.
(1) Press Trigger
(2) Press Initiate
(3) Press Trans.
(4) Press Ch 1
to start sourcing the
Agilent B2901/02/11/12A Precision Source/Measure Unit
Stopping sourcing the continuous pulses
Step 18. If you aren’t on the top of the Function menu, press
repeatedly to return to the top level.
On the top level of the Function menu
Any tree structure can’t be seen
In the middle level of the Function menu
Some tree structure can be seen
Step 19. Press
,
,
, and then press
continuous pulses.
(1) Press Trigger
(2) Press Abort
(3) Press Trans.
(4) Press Ch 1
to stop sourcing the