LNG SAMPLE VAPORIZER ISOSAMPLE 8100 for ON LINE GC SAMPLING Recording Q/N : 0808647 Rev 0 Contractor : Project: To be specified Q/N: 0808647 OPTA-PERIPH 1 LNG SAMPLING SYSTEM DATE 14/08/2008 REV. 0 1.0 INTRODUCTION This Quotation defines the TECHNICAL CHARACTERISTICS of the ISO 8100 LNG Sampling System for the xxxx loading Plant. The online system for LNG is composed of an efficient online sample probe and a vaporizer. This device can be provided together with any existing online Gas Chromatograph for LNG. Additionally a sampling system (preferably automatic) is generally installed to collect samples for the purpose of laboratory measurements by contractors and third party. The sample probe and vaporizer are the key elements to ensure integrity of the sample and to obtain the performance of accuracy with 9.0 kJ/kg Gross Calorific Value and 3.0*10 -4 kg/m3 Density maximal random error. When a sampling is required, the holder cylinder described in this quotation is a prerequisite for compliance with existing standards. The first part of system is the sample take-off probe: 2.0 CODES & STANDARDS: The probe vacuum isolated and the vaporizer are complying with following standards: - Sample probe resonant frequency calculation: -BS IEC 61831:1999 for on line velocity <7 m/s -API RP 551 for on line velocity > 7 m/s -Main welding test: APAVE certified under CODAP 2000- Rev 04-04-IA4--ASME V- APIRP 551 - Probe subcooling degree calculation: ISO 8943-2007 -Vaporizer :- capacity calculation: ISO 8943-2007 - Ex-proof ATEX II 2G EExde II C T4 THI - Probe & vaporizer pressure test: PED 97 23 EC - calculations approved Bureau Veritas - Suitability testing of discontinuous sampling with lab and process GC by supplier: EN 12838- Class A of accuracy with 54 kJ/kg Gross Calorific Value and 18*10 -4 kg/m3 Density maximal random error. - Suitability testing of continuous sampling with ISOSAMPLE 8100 piston cylinder holder and process GC by supplier: EN 12838- Class A of accuracy with 9.0 kJ/kg Gross Calorific Value and 3.0*10 -4 kg/m3 Density maximal random error. OPTA-PERIPH 2 LNG SAMPLING SYSTEM DATE 14/08/2008 REV. 0 3. 0 SAMPLE TAKE - OFF PROBE PRINCIPLE: The sample take-off is composed with a probe inserted directly at the centre of the LNG product pipe, double isolating valves, and a capillary tube carrying the LNG sample from the probe towards the vaporizer (see Figure 0-1). Figure 0-1: Picture of the ISO 8100 sample probe This probe has been designed to provide a Thermal insulation of the sample takeoff probe by high vacuum maintained below 1.10-4 torr. THI The conjunction of this ultra-low residual pressure minimizing heat leak by conduction and convection with layer of radiation shields cryogenic constitutes a super insulation system providing apparent thermal conductivity (k) values below 0.4 milliwatt per meter-kelvin (mW/m-k), therefore the enthalpy rise can be controlled to be inferior of the subcooling degree and fractionation will not have occurred in the sampling line. The annexure 1 of this document determines the degree of subcooling reproduced from the ISO 8943 standard compared to the performances of the ISO 8100 Sample probe. 4.0 VAPORIZER PRINCIPLE: The LNG sample has to be totally vaporized before being sent to the online Gas Chromatograph. In particular heavy components of the LNG shall not remain in the vaporizer. In the ISO 8100 device, vaporization is made in supercritical state, eliminating the risk of fractioned vaporization. There are two important points related to the phase envelope, namely Cricondenbar and Cricondentherm. These points represent respectively the maximum pressure point and temperature point at which the LNG exists in two phases. Figure 0-2 here under displays the pressure/temperature diagram for a LNG containing approximately 90% methane, where: OPTA-PERIPH 3 LNG SAMPLING SYSTEM DATE 14/08/2008 REV. 0 B T C Cricondenbar, in bar Cricondentherm, in °C Critical point, in bar Figure 0-2: Pressure/Temperature Diagram The objective is then to transform LNG from liquid to gas state in supercritical conditions at very high pressure (here above 75 bars), such that the LNG sample goes directly into the desired gas state. Such transformation is represented on Figure 0-2 by the dotted line. THI Within the ISO 8100 sample probe and vaporizer, the following operations are taking place (see Figure 0-3): • The temperature rises from -160°C to -150°C in the capillary tube, at LNG process pressure (minus some pressure drops in the sample line) • A supercritical cell featuring pressure reductor/check valve integrated at the inlet of vaporizer coil has been designed for flashing at that point the liquid sample without back pressure effect to process line • The vaporization in this cell (indicated as "critical point in the figure 2-4) of approximately 0.5 cc increases the temperature and the pressure, and leads to conditions of approximately 80 bars and -100°C • The temperature is controlled at +65°C in the vaporizer, in order to obtain approximately +55°C at the outlet of the vaporizer OPTA-PERIPH 4 LNG SAMPLING SYSTEM DATE 14/08/2008 REV. 0 Figure 0-3: Sample probe/Vaporizer profile THI Similarly to the sampling probe, no fractioned vaporization is taking place in the vaporizer, ensuring that the required sample quality and stability is maintained carefully before being sent to the Gas Chromatograph. OPTA-PERIPH 5 LNG SAMPLING SYSTEM DATE 14/08/2008 REV. 0 5.0 SYSTEM DESCRIPTIVE: First segment probe: The sampling probe must be located on the top of horizontal part of pipe to withdraw the LNG from a penetration in pipe. The probe is designed with variable diameter for non resonant compliance . This tube constitutes the external part of the vacuum jacket downside the flange, the Internal sampling tube is sized 1/8" OD . The flange item 2 is sized 2 to 4" 150 to 600 LBS according to the user requirements and the upside vacuum jacket item 15 is welded on flange and assembled with high vacuum flanged connections. 3 15 2 Diagram 3 THI A first sample isolation core valve item 3 is supplied and encapsulated in the vacuum jacket, this valve is only accessible by opening the flange cap to isolate the system from pipeline if any maintenance work is required, thus, the downstream part of system can be removed and the vacuum re-established in this pre-insulation section without process service disturbance such as outage of the loading line. A valve drive special tool is supplied with the probe. Second segment probe: A capillary assembly item 4 is provided for breaking equilibrium between the sample and pipeline pressures and the main isolating actuated valve item 5 is supplied with extended bonnet and pneumatic actuator item 9. This actuator is operated via the Control unit installed near of the vaporizer cabinet to switch ON/OFF the valve (normally closed) automatically for alarm or manually. The valve is automatically turned OFF if the temperature TT5 at outlet of vaporizer is dropping below - 20°C because of vaporizer failure. OPTA-PERIPH 6 LNG SAMPLING SYSTEM DATE 14/08/2008 REV. 0 5 4 Diagram 4 9 The check valve item 7 located on critical point ( refer to Fig 2-4) is operating as pressure barrier between the liquid/vapor phases e.g: critical point (at which the liquid state of LNG ceases to exist). Temperature measurement transmitter thermocouple K item TT2 (with 1 meter coil in the vacuum jacket to avoid heat leak by conduction) is provided downstream of the check valve at the inlet of vaporizer coil where the sample is transformed from liquid to gas phase between -80 and -130°C approx. 3. VAPORIZER: The coil exchanger item 16 is a 2m x 3 mm ID tubing and the assembly is so constructed for providing sample flushing to avoid fractionated distillation effect and the heavier components of the LNG will not remain in the coil. Three temperature points are monitored: TT2: thermocouple K Transmitter: measuring the inlet temperature of LNG, a 1 m extension capillary is provided in the vacuum jacket for thermal insulation. TT3:PT100 (inserted in well item 16) measuring the heat exchanger temperature controlled by Application PC Controller. TT5: pneumatic thermoswitch measuring the natural gas at outlet. TT5 THI . TT2 OPTA-PERIPH 7 LNG SAMPLING SYSTEM DATE 14/08/2008 REV. 0 All these signal temperature are entering in the Control unit for processing the application. An high temperature trip switch TAH4 calibrated at 135°C for class T4 is supplied for safety redundancy and a pressure alarm pneumatic switch PT4 is supplied as well on natural gas at outlet of vaporizer. The Control Unit provides a temperature control of exchanger TT3 at 65°C plus a temperature interlock system for auto-shut-off sample take-off valve if outlet temperature T5 drops to -20°C ( normal is +55 °C) or if pressure at vaporizer outlet exceeds 8 bar-g. As per our recommendation a fast loop Mass Flow Controller FIC1 can be supplied by the GC sampling integrator upstream of the GC; this MFC can be processed from our Control Unit to control the TT2 temperature at coil inlet below -75°C, range of flow rate is between 1 000 and 1500 SL/H. The capacity of the LNG vaporizer has been designed to gasify the whole volume of LNG which is being withdrawn for sampling For this application the subcooling degree compliance has been calculated for a total gas vaporized volume of 1 000 SL/H. Vaporizer integration: The vaporizer is contained in a stainless steel cabinet, protecting code is ATEX II 2G EEx de IIBT4 and heating capacity is 0.5 kW. The actuated valve local manifold is contained in the box item 02. The other items are : -item 01: inter-connection box -item 03: temperature transmitter TT2 -item 04: temperature transmitter TT5 -item 05: absolute pressure transmitter PAL2. THI The vaporizer cabinet is mounted on the probe frame tied to the pipeline. OPTA-PERIPH 8 LNG SAMPLING SYSTEM DATE 14/08/2008 REV. 0 4-. CONTROL UNIT & PNEUMATIC LOGIC: The Control Unit provides a temperature control of exchanger TT3 at 65°C plus a temperature interlock system for auto-shutoff sample take-off valve if outlet temperature TT5 drops to -20°C ( normal is +55 °C) or if pressure at vaporizer outlet exceeds 8 bar-g. As per our recommendation a fast loop Mass Flow Controller FIC1 can be supplied by the GC sampling integrator upstream of the GC; this MFC can be processed from our Control Unit to control the TT2 temperature at coil inlet below -75°C, range of flow rate is between 1 000 and 1500 SL/H. - A Pneumatic logic controller of sample probe valve is provided integrating: . manual valve on/off . remote pneumatic on/off valve control . automatic valve shut-off if TS5 < -20°C . automatic valve off-off if pressure alarm > process pressure + 10 bar-g 5-. ASSEMBLY of PROBE/VAPORIZER: Sample probe must be installed at point in the pipeline where the LNG is in a subcooled condition. The two segments of Probe and the vaporizer must be flanged vertically on top side of pipe line with the vaporizer enclosure supported on same axis. As option a probe/vaporizer mounting frame is supplied and the bellows supplied on the probe second segment allows a line taping moving of +/- 45 mm from axis. The control unit and pneumatic logic plate assemblies are mounted and fully distributed on frame. THI Vacuum generation: On original start-up , the vacuum generation inside of the jacket is operated with an eductor supplied as option. OPTA-PERIPH 9 LNG SAMPLING SYSTEM DATE 14/08/2008 REV. 0 Mounting frame on pipeline Control unit with isolator Plate assembly Pneumatic logic Plate assembly 6-. MAIN SPECIFICATIONS: Pipe-line pressure: 2 to 10 bar-g Design pressure: 18 bar-g Pipe-line temperature: - 160 °C Volume of gas sample vaporized: approx 1000 Sl/h Lag time through sampling probe and vaporizer: 30 sec approx. Thermal conductivity of probe: 0.4 milliwatt per meter-kelvin (mW/m-k) THI Heating capacity of vaporizer: 0.5 Kw Sample outlet connection: 1/4" NPTF Protecting electrical code: ATEX II 2G EEx de IICT4 Enclosure rating: IP65 or NEMA 4x 50 OPTA-PERIPH 10 LNG SAMPLING SYSTEM DATE 14/08/2008 REV. 0 7-. BILL OF MATERIAL: THI 1-SAMPLE TAKE-OFF PROBE -High vacuum -Integrating the core encapsulated valve item 3 and the actuated valve item 5 normally closed -Pneumatic actuator of shut-off valve -Bellows on the probe second segment for probe axis taping moving of +/- 45 mm. -Vacuum jacket isolation by auto-opturation valve at inlet and vacuum valve at outlet . -Flange Type: RF Flange or other as specified -Rating:300lb or other as specified -Size :2" ANSI or other as specified Maximum process pressure: 16 barg as standard or 150 barg for the high pressure model -Sampling tube diameter and length determined for non resonant compliance -Orientation: on top -Material: all parts in S S 316L. . Options: 1- Dual pitot inlet orifice for direct and reverse flow sampling 2-Vacuum gauge ranged -1 to 3 bar-g 3- Air vacuum jet ejector fitted on vacuum valve (only one air ejector required for all the LNG PROBES of the plant) 4- Mounting frame on the process flange: material in SS 304 ; this option is to line-up the probe and vaporizer on process flange 5- Mouting of Control Unit and Pneumatic Logic on frame with complete pneumatic and electric distribution Note: the weight of probe / vaporizer to be added to stand weight is 95 Kg. II:LNG VAPORIZER / CONTROLLER: - Vaporizer to be flanged at the extremity of sample probe spool. - Power supply: 240 VAC - 50Hz - Heating Capcity:500 Watts - Classification:ATEX II 2G EEx de IICT4 -Maximum process pressure: 150 barg - Cabinet Material:304 Stainless Steel - Monitoring of three temperature points: TS 1: thermocouple K: measuring the inlet temperature of LNG, extension capillary is provided in the vacuum jacket for thermal insulation. .TS2:PT100 measuring the vaporizer heat exchanger temperature .TS 5: pneumatic temperature switch : monitoring the natural gas at outlet. -As safety redundancy an high temperature trip switch ,calibrated at 135°C for class T4 is integrated in the vaporizer heater block. - Pressure alarm pneumatic switch fitted on the natural gas at outlet - Pneumatic logic controller of sample probe valve integrating: . manual valve on/off OPTA-PERIPH 11 LNG SAMPLING SYSTEM DATE 14/08/2008 REV. 0 . remote pneumatic on/off valve control . automatic valve shut-off if TS5 < -20°C . automatic valve off-off if pressure alarm > process pressure + 10 bar-g III-TEMPERATURE CONTROL UNIT integrating: . Vaporizer PID temperature indicator controller:0 to 100 °C (usual set point: 65°C) . TS1 température indicator . Protection: ATEX II 2G EEx di IICT6 . Enclosure in aluminum painted . Power supply: 240 VAC - 50Hz & 24 VDC Option for Exd 24 VDC generation if not available on site Delivery: 24 weeks after order receipt THI IV- SAMPLING RECOMMENDATIONS: At outlet of vaporizer in by-pass to BOG line, the flow-rate is to be controlled at 1000 SL/H, this is not included in our quotation. Sampling system supplied by other shall involve a Mass Flow Controller type EMERSON BROOKS Type 5816N/36 EX (item 1 in the typical spot sampling diagram attached) the optimized process control is to control the flow from 1000 to 1500 SL/H in order to keep the temperature TS1 at the lowest value achievable (usually between -130 to -80°C). The vaporizer has been designed for a maximum flow-rate of 2000 SL/H. OPTA-PERIPH 12 LNG SAMPLING SYSTEM DATE 14/08/2008 REV. 0 LNG AUTOSAMPLER ISOSAMPLE THI for AUTOMATIC FILLING of LABORATORY CYLINDERS OPTA-PERIPH 13 LNG SAMPLING SYSTEM DATE 14/08/2008 REV. 0 1.0 INTRODUCTION This Quotation defines the TECHNICAL CHARACTERISTICS of the ISO 8100 LNG Sampling System. This supply is related to the LNG Charge Continuous AutoSampling System for waterless-type gas sample holder as recommended in the latest version of ISO 8943-2007 : The system features the waterless-type sample holder recommended and this unit (typically described with rubber membrane in the ISO standard ) is built within the Constant Pressure/Floating Piston (CP/FP) principle far more robust with a guarantee of 5 years operation without maintenance. As well the CP/FP cylinder is treated with sulfur passivation for samples containing sulfur substances to be analysed . Notes: Charge autosampling is related to grabbing operated during full unloading operation, thus, the sample collected is representative of complete transfer. It has been assumed that the requisition is not for Intermittent sampling for CP/FP sample container not dedicated to charge sampling application but spot autosampling. Intermittent sampling requires a cleaning procedure of containers ( para B.4.1 of ISO 8943/2007) involving trace gas analysis too heavy for useful operation. 2.0 ARCHITECTURE OF SYSTEM THI The system involves the Probe vaporizer quoted in the first section, the Autosampler Model 8100 installed in a shelter near from the GC and the software implemented in the Application PC controller. Two control architectures can be supplied: Sampler Application Controller is a PC in control room (safe area) with local multiplexer / screen for zone 1 area (represented on diagram 1). Or Sampler Controller is a PLC supplied for zone 1 area. 50 OPTA-PERIPH 14 LNG SAMPLING SYSTEM DATE 14/08/2008 REV. 0 In both cases Sampler Controller features a MODBUS or OPC data-link to customer DCS. Additionally, data's acquisition and fast loop flow-control of existing Probe vaporizer (Cegelec) is provided as well as GC (Emerson) data's acquisition for statistics and validation of complete system at the end of unloading. 3. 0 SAMPLE TAKE - OFF PROBE SUPERVISION: Sample vacuum thermal isolation probe with cryogenic valve is installed at point in the pipeline where the LNG is in a subcooled condition. The degree of subcooling at a sampling point will be ascertained by observation of the temperature and pressure of the LNG at that point and comparing the temperature with the B.P. of the LNG at the same pressure as calculated from the composition of the LNG. It has been assumed that the pipeline temperature TT1, flowrate FI1 and pressure PT1 sensors/transmitters in pipe-line at sample take-off are supplied by others . 4. 0 CONTINUOUS CHARGE AUTOSAMPLING SYSTEM: A Charge aggregate autosampling system is provided with automatic/manual interchange sequential control system as well as time programmed automatically purging and filling for complete unloading transfer sampling. The sampling start up automatically when the monitored pipe-line flow rate (FI1) reaches the nominal programmed value according to the ISO 8943/2007 para 7.2 "Consecutiveness of sampling "and sampling is operated from flow stabilisation to decresed flow rate before end of transfer. Suspension of sampling for sudden change in flow rate is auto-operated according to para 7.3. Bombs are filled in duplicate; and usually , 2 bombs of 0,5 litres each are supplied per stream sampled. The grab procedure is automated by the PC Controller. 4.1 Mass Flow Controller: FIC THI Usually two FIC are provided in the system: 1-FIC1 (not included in this quotation) controls the vaporizer fast loop flowrate returning to the BOG line. The flow set point is controlled by the Application PC controller : minimum set point of 1000 Sl/h to comply with subcooling degree requirement can be increased up to 1800 Sl/h if the temperature TT2 expected to be at -80 °C at vaporizer coil inlet is rising . Would this function been unable to return to cooling conditions , the vaporizer temperature controller set point is changed from 65 to 20 °C. OPTA-PERIPH 15 LNG SAMPLING SYSTEM DATE 14/08/2008 REV. 0 2- FIC2 is provided for controlling the flow to the charge holder cylinder at 3 to 10 SL/H (refer to annex 2 on base of 6 to 12H unloading) To compensate the Joule-Thomson effect, the autosampler cabinet is electrically heated. 4.2 Back Pressure Regulator: item 18-1 The pressure at outlet of vaporizer is the same than on pipe-line pressure usually between 2.5 and 6 barg. The gas flow-rate on fast loop controlled by the FIC1 is released to Marine vapor recovery line header. This back pressure regulator function is only for safety and set point is adjusted at 8 barg. 4.3 System inerting manifold: to GC from vapo THI He Inerting manifold is automatically operated at the first commissioning sampling to remove any trace of air in the system or further to a maintenance operation with material dismounting. As well this inerting function can be involved in the plant fire prevention procedure. to sampling OPTA-PERIPH 16 LNG SAMPLING SYSTEM DATE 14/08/2008 REV. 0 5. CHARGE HOLDER CYLINDER: One constant pressure floating piston (CF/FP) holder is installed between vaporizer and charge portable sampling containers train For this application the holder cylinder volume has been designed at 25 L The holder piston cylinder features PTFE seals double piston barrier with a piston scrapper to auto-clean the cylinder polished wall between two samplings. Original purging is compliant with the ISO 8943-2007 section 6 6 4 and 7.4.1-d : this purging is made through the piston shaft bore . On the "ISOSAMPLE 8100" this procedure is fully automated: the buffer gas chamber been pre charged by compressed air with piston in down position, the sample chamber is reduced to the nominal zero volume before of sampling. Then the sample purge valve is actuated to open for purging the tiny dead volume between piston/end plate through the piston shaft venting to atmosphere. For a sampling line from vaporizer of 30 m with 3 mm ID tubing electropolished a purging cycle of 1min (programmable) is operated, then the purge valve is closed. Level transmitter is provided for comparison of calculated and actual sample level in the holder or for indication of level when the air motive pressure is to be decreased to a smaller value consecutively to a pressure drop in line below 2.5 barg. THI - During grabbing cycle, the system is controlling a time proportional charging of the holder piston cylinder during a period of 6 to 12H (programmable according to cargo size) . The Mass flow controller FIC2 set point is controlled by the Application PC Controller for adjusting in continuous this proportional charging from 6.75 SL/H (for 6H transfer period ) to 3.375 SL/H (for 12 period) . The loading pressure of floating piston holder is controlled at 1.5 barg (1 barg below the minimum process expected pressure) for charging but in case of process press dropping below the 2.5 barg minimum expected pressure, the loading pressure of floating piston holder is re-adjusted at process pressure less 1 barg. - At the end of each grabbing period , the holder piston cylinder supposed to store 40.5 SL of gas under 1.5 barg (corresponding to a volume of 16.2 L in the holder) is automatically isolated and connected to the charge portable containers circuit for loading a OPTA-PERIPH 17 LNG SAMPLING SYSTEM DATE 14/08/2008 REV. 0 new couple of containers. - The couple of charge containers are loaded during the unload transfer of 10 to 12 H , then the system is stopped waiting for operator new cycle validation once the charge containers have been collected and replaced in the rack . The loading pressure of floating piston holder is controlled at 1.5 barg for charging (if process pressure does not drop below 2.5 barg) and at 4 bar-g for purging + end charging the portable containers. For motive air insulation, the holder piston cylinder features interseals chamber to be filled with a tiny volume of Helium. A small pressure helium ( below 80 g/cm2g) will be controlled.. The LNG Sampling holder capacity is 25L to comply with the sum of the volume required for charging the charge portable cylinders and additional volume required for purging cycle (refer to calculations annex 2). The number of purging cycle will be as per annex D of ISO 10715 and Appendix C of ISO 8943-2007defined according to the final pressure in cylinder : i.e: 8 cycles (programmable) at 4 Kg/cm2g. CP/PF holder loading pressure control: THI The pressure inside of holder is monitored by the PT5 transmitter and the Application PC Controllers operates the charge and discharge solenoid valves 8-1 and 8-2 for adjusting the required pressure. At starting of sampling, the valve 5-6 is opened for the controller to check that a pressure is recovered at the transportation containers outlet thus the container valves have been opened by the operator before of remounting. OPTA-PERIPH 18 LNG SAMPLING SYSTEM DATE 14/08/2008 REV. 0 6. CHARGE TRANSPORTATION CONTAINERS: Charge aggregate sampling containers are provided with automatic/manual interchange sequential control system as well as time programmed automatically purging and filling for complete unloading sampling. . The grab procedure is automated by the Plant PC controller. The accumulated average sample gas of charge unloading is transferred from holder cylinder to portable sampling container and the gas will be controlled in constant pressure of 4 bar-g at final charging.. -Procedure for purging will be as per the fill and empty method refer to Annex D of ISO 10715 and Appendix C of ISO 8943-2007 defined according to the final pressure in cylinder : i.e: 8 cycles (programmable) at 4 Kg/cm2g. The procedure is operated within the following steps: At the end of tansfert sampling - depending of cargo size programmeda couple of new spot cylinder will be fed in parallel by the holder cylinder after a preliminary purge following procedure: THI - Open the vent valve and purge slowly the vent line :2 SL - Open the load valve and purge the line and charge container: 5 SL - Close the discharge valve and load the spot container at 4.Kg/cm2g - Close the charge valve and vent the spot container until it reaches atmospheric pressure - Close the discharge valve and open the load the spot container at 4.Kg/cm2g - Repeat the operation for a total of 8 cycles - Load the spot container at 4 Kg/cm2g and close the discharge and the charge valves - Purge the holder cylinder through the vent valve - OPTA-PERIPH 19 LNG SAMPLING SYSTEM DATE 14/08/2008 REV. 0 The two charge containers are rack mounted inside of an heated casing at the back of holder cylinder cabinet temperature controlled at approx. 40°C.. Portable gas sample containers are designed as a cylinder made of stainless steel tubing and fitted with a stainless steel needle valve on each end. Pressure rating and capacity values: Volume 0.5 Liters/each Loading Press. 4.0 barg Maximum Press 30.0barg Design, manufacturing and verification of portable cylinder comply with the European Pressure Equipment Directive PED 97/23/CE. THI 7. SYSTEM INTEGRATION: The autosampling system is integrated in a SS cabinet ground mounted with all components accessibility. Electrical and pneumatic interconnections are supplied from the Control Cabinet EExp mounted on right side and certified ATEX for zone 1 group B T3. The control cabinet contents the Power distribution 220 VAC/ 24 VDC, the Solenoid OPTA-PERIPH 20 LNG SAMPLING SYSTEM DATE 14/08/2008 REV. 0 valve interface for processing the sampling valves with multiplexer unit and Modbus controller, the holder cylinder level converter, the FIC 1 and /2 mass flow controllers control board and the front screen system supervision. The solenoid valves manifold , Multiplexer I/O and Modbus communication interface are integrated in the Burkert/Wago modular system. This system simplifies wiring distribution and accepts additional elements to be integrated later on. A RS485 isolated interface is provided for long distance communication through the Modbus coupler to control room Application PC Controller with the functionality of a PLC. The active communication channel works independently of the solenoid valves manifold controller for full duplex operation over 19200 baud. Control cabinet A graphic screen is provided on the cabinet front door; the autosampler PID represented gives the network configuration of system with actual parameters collected from the Modbus hub. Two push button are provided for changing the window or zooming the graphic display. Graphic display zooming THI Solenoid manifold / Multiplexer / Modbus coupler OPTA-PERIPH 21 LNG SAMPLING SYSTEM DATE 14/08/2008 REV. 0 8. SYSTEM OPERATION DATA'S System operation : -At origin START of transfer is switched by operator and the sampling starts automatically when the stable nominal flow rate through import pipe-line is monitored by the system. The sampling will be temporary suspended for any sudden change in the flow rate. As well, the cargo size is programmed by operator on Application PC Controller for calculation of sampling periods. -Charge Holder cylinder empty detection is sensed for filling authorisation and motive air pressure is switched during 30 sec. on holder cylinder before original purging is made through the piston shaft bore . - According to cargo size filling cycle is switched during 6 to 12 H in charge holder. During this cycle the air loading pressure is controlled to 1.5 kg/cm2g in holder cylinder if process pressure is above 2.5 barg, if this pressure drops the air loading pressure is controlled at process pressure less 1bar . -At the end of the charge holder cycle (6 to 12 H), if the level of sample recovered in the Holder cylinder is corresponding to 40.5 SL of gas +/- 10%, the sample is transferred to the charge transportation containers within the procedure and if this operation is proceeded without pressure alarm detection on charge and discharge Pressure transmitter PT5 & PT3 , the information “Cylinder N°1/2 ready” is displayed; the Holder cylinder is switched –off with the piston on down position and the system is isolated from process . THI The operator must proceed with collection of the charge transportation containers. Further to containers changing the system is automatically checking at new transfer start-up by measuring gas pressure at PT5 & PT3 that containers have been mounted with valves opened. If case of unsuccessful validation the system is hold in standby even if a new Start of transfer is switched Or - If “Sampling standby” is automatically switched by the process flowrate dropping or manually switched by operator the system is hold in standby with alarm “Sampling standby” displayed until nominal flowrate is recovered of Start transfer is switched Or -If “Sampling cancellation” is manually switched by operator: Holder cylinder is vented up to empty detection and system is hold in standby with alarm “Sampling cancellation” displayed until Start transfer is switched Or : - If the holder cylinder high volume of 40.5 SL of gas + 10% OPTA-PERIPH 22 LNG SAMPLING SYSTEM DATE 14/08/2008 REV. 0 level alarm detects before end of sampling period, sample is transferred to relative portable container and the high level sampling alarm is displayed. Or : - If the sampling time is exceeded before the low volume of 40.5 SL of gas - 10% is found: sample is transferred to relative portable containers and the low level sampling alarm is displayed Or : - If the 4 Kg/cm2g pressure charge of portable container is not sensed at PT5 during Holder to portable container transfer: Sampling fault alarm is displayed and system is hold in standby . Or If the drop to BOG pressure (250 g/cm2g ) at portable container outlet is not sensed at the discharge pressure transmitter PT3 during Holder to Spot cylinder transfer : Sampling fault alarm is displayed and system is hold in standby 9. APPLICATION PC CONTROLLER: DESCRIPTIVE THI Supplied for installation in safe area (Control Room) with the communication architecture described on para 2 or for a local installation in explosive area zone , the Application PC Controller is processing the Sample take-off line data's acquisition, the spot autosampling system, controlling the Gas Chromatograph results statistical treatmen and the analyse of parameters influence. Software development is in the OPTA-PERIPH scope with calibration parameters as well as communication data-link to the control cabinet EExp multiplexer and to the Gas Chromatograph. Architecture is designed around PC hardware and Windows™ operating system : sampling processing and analysis method tables are easily changed modified or edited in the same operating format. OPTA-PERIPH 23 LNG SAMPLING SYSTEM DATE 14/08/2008 REV. 0 FUNCTIONS: PROBE VAPORIZER SUPERVISION The graphic screen copies are representing some steps of program for PROBE VAPORIZER SUPERVISION SPOT AUTOSAMPLER PROCESSING CHROMATOGRAPH RESULTS STATISTICAL TREATMENT / TRANSFER DATA-LOGGING CHARGE AUTOSAMPLER PROCESSING HOLDER FILLING PURGING HOLDER LINE PURGING THI STABILIZATION OPTA-PERIPH 24 LNG SAMPLING SYSTEM DATE 21/07/2007 REV. 0 FILLING BOMB FLUSH BOMB PURGING BOMB FILLING ALL BOMBS EMPTYING HOLDER THI END OF CYCLE OPTA-PERIPH 25 LNG SAMPLING SYSTEM DATE 21/07/2007 REV. 0 CHROMATOGRAPH RESULTS STATISTICAL TREATMENT: THI TRANSFER DATA LOGGING: OPTA-PERIPH 26 LNG SAMPLING SYSTEM DATE 21/07/2007 REV. 0
© Copyright 2024