Sp e c i a l Fe a t u r e s
‡ Reduced operating costs
‡ 100 % reliability and
availability on emission
and process analyzers
‡ High and constant dryness
rate (long term stability)
‡ Compact, small footprint
‡ Modular constructed
‡ Long term stability
‡ Replaceable heat
exchanger
‡ Designed for all emission
monitoring and process
analysis applications
‡ Very low maintenance
expenditure
SMAK–
Sample Gas
Conditioner
To provide clean dry gases to
extractive analyzers for
continuous emission monitoring and process control,
the SICK SMAK Sample Gas
Conditioner is an essential
part of any extractive gas
analyzer system. This
condition unit will continuously dehumidify a sample
stream and rapidly separate the condensable
liquids from the gases.
The SICK SMAK is ideal
for applications where a
constant sample dewpoint and dryness rate
must be maintained.
Spiral tube heat
exchanger
Spiral tube heat exchanger
and typhoon separator
excels by
‡ Maximized heat transfer
rate
‡ Quick formation of larger
separable droplets
‡ Reduced contact surface of
droplets
‡Lowest rate of soluble
component SO2
‡ Fast response on dew point
changes
‡ Self cleaning, no fouling, noclogging
‡ Antifreezing cooling storage
adsorbs peakloads
SMAK features
Compact, robust design in
accordance with environmental and approval standards
‡ High cooling efficiency
achieved by an internal
airflow system even in high
ambient temperatures
‡ Highly corrosion resistant
sample tubing materials
‡ Protected against invasion
by ambient corrosives
‡ Leakage-free operation
confirmed by shockproof
and field experience
‡ Monitoring system including
reading of actual cooling
temperature in °C or °F and
potential free contacts
providing unattended
operation
MKAS-64 analyzer cabinet
Control unit
H2O
In
Probe
Out
Out
In
Filter
Meas. gas line
Condensate
GME 64
Condensate
pump
Condensate
pump
Gas conditioner
Humidity
sensor
Condensate
Condensate
monitor
Condensate
container 10 l*
Reference
Air
* Option
SMAK Function Overview
Ordering system
General Data
Max. sample inlet temperature 140 ... 180 °C *
Max. sample inlet dewpoint
80 °C
Outlet gas temperature
3 °C
Outlet gas dewpoint
3 °C
Ambient temperature
5 ... 45 °C
Material of Gas Path: PVDF and Glass
Operating pressure (absolute)
Max. gas flowrate per gas path
Max. sample inlet temperature
Dead space per gas path
Pressure drop per gas path
0.5 ... 2.5 bar
300 l/h **
140 ... 180 °C
137 cm3
0.025 bar
Material of Gas Path: Stainless Steel
Operating pressure (absolute)
Max. gas flowrate per gas path
Max. sample inlet temperature
Dead space per gas path
Pressure drop per gas path
*
0.5 ... 100 bar
500 l/h **
180 °C
137 cm3
0.047 bar
PVDF = 140 °C, Stainless steel = 180 °C;
Glass = 180 °C are max. temperatures,
Standard condition: 140 °C
** in accordance with standard conditions inlet
gas temperature = 140 °C, inlet dewpoint =
65 °C, ambient temperature = 5-40 °C
Device versions
SMAK..
6-mini
6-1
6-2/1
6-2
6-3
6-2-1 Ex
6-2-2 Ex
dimensions[mm]
250x300x240
290x360x335
290x360x335
290x440x360
350x470x430
350x470x430
350x470x430
power consumption[W]
180 W
300 W
300 W
350 W
370 W
370 W
370 W
Gas Cooler GME 73-1
1
Cooler
1
Type
SMAK 6-mini
SMAK 6-1
SMAK 6-2/1
SMAK 6-2
SMAK 6-3
SMAK 6-2-1 Ex
SMAK 6-2-2 Ex
Gas path
1
1
2
2
3
1
2
1
2
3
4
5
6
7
Ex-version
Yes
No
1
2
Material
PVDF
Glass
Steel 1.4543
1
2
3
Power supply
230 V AC/50 Hz
230 V AC/60 Hz
115 V AC/50 Hz
115 V AC/60 Hz
1
2
3
4
Specials
No
Display via LED (not with Ex)
External alert relais (only Ex)
SICK AG • Environmental Monitoring• Nimburger Str. 11 • D-79276 Reute • Germany
Tel.: (+49) 0 76 41/4 69 0 • Fax: (+49) 0 76 41/4 69 11 49 • www.sick.de
1
2
3
8 008 356/03-98/GB3/BW • Subject to change without prior prior notice
Technical Data