Sample Gas Conditioner JCC
FEATURES
uStable dew point
+3°C..+/- 0,2°C
uAmbient temperature
+ 5 ... +45°C (50°C)
uProportional temperature & pressure control
uBuilt-in pump & condensate discharge pump
uDigital temperature display and flowmeter
uHigh performance heat exchanger
uProven and reliable technology
uCompact and robust
u100% environmentally friendly (CFC free)
GENERAL
The JCC Sample Gas Conditioner is designed
to lower the sample dew point and supply
dehumidified sample gas with a pump to the
moisture intolerant analysis equipment. The
operation capability at high ambient temperature
enables also applications i.e. in boiler houses.
friendly R134a. The high performance heat
exchanger guaranties maximum heat transfer
rate and leak free operation. A built-in
condensate discharge pump removes
continuously the condensate and respectively a
sample filter the impurities.
TECHNICAL
A constant dew point independent of varying
sample specifications and ambient temperature
is ensured by an improved proportional
temperature & pressure control . The cooling
system features a continuously running non-CFC
compressor motor filled with environmentally
FEATURES
The JCC Sample Gas Conditioner features a
single or dual gas path made of PVDF. A
flowmeter allows the adjustment of the required
flow rate. A digital temperature display, an
isolated contact and interlocked pump provide
unattended trouble free operation.
Technical Data:
Gas flow rate:
100Nl/h one gas path
200Nl/h two gas paths
Gas inlet dew point max.: 65°C
Gas inlet temp. max.:
140°C
Gas outlet dew point:
+3°C (factory set point)
Dew point stability:
+/-0.2°C
Ambient temperature:
+5....+45°C (50°C)
Cooling power:
160W at ambient 25°C;
evaporationtemp. 0°C
Dead volume:
approx. 160 ccm
(60ccm heat exchanger)
Condensate discharge: 1 cond. pump / gas path
capacity 300ml/h
Gas filter:
one way filter 1mµ
Nylon /Borosilicatefibre
Flowmeter:
Flowmeter range:
Operating control:
with needle valve PP/Glass/PVDF
10..100Nl/h / 15..150Nl/h
Digital indication of
heat exchanger temperature
Built-in temp. alarm with
potential free switch over contact,
rating 250VAC 2A with cable,
interlocked pump at >8°C
Gas inlet: Fitting DN4/6 PVDF
Gas outlet: Fitting DN4/6 PP
2 x 3p cable 2m; Mains; Status
steel sheet; coated RAL 7035
430 x 225 x 300mm (WxDxH)
approx. 20kg
Contact rating:
Gas connection:
Mains:
Housing:
Dimension:
Weight:
Gas Flow
Gaslaufplan
Gas flow schematic
1
3
4
Gaslaufplan
Gas flow schematic
OUT
5
6
1
3
1
4
5
OUT
1
2
3
4
5
6
6
IN
IN
2
2
CONDENSATE
2
Heat exchanger
Peristaltic pump
Diaphragm pump
Condensate sensor (Option)
Filter
Flowmeterw. needle valve
CONDENSATE
Dimension Basic
SAMPLEIN
16
27
7,6
483 ( 19" )
465
R
354 ( 8HU )
182
5
SAMPLEOUT
POWERCORD
STATUSCORD
4
40
431
Typ
J C C -
Housing
Gas path
Fine dust filter
Wall mounting on rear panel
L
Wall mounting on side panel
Q
19" Rack mounting
R
WWT-1 gas path; PVDF
1 Condensate pump
1
WWT-2 gas path; PVDF
2 Condensate pumps
4
One way filter
1
Special on request
2
Gas Sample pump
<200Nl/h
Gas flow control
Flowmeter PE w. needle valve 10..100Nl/h
1
Flowmeter PE w. needle valve 15..150Nl/h
3
Condensate detector
Power supply
2
Without
0
With KW-1 (interlocks pump by condensate breakthrough)
1
230V 50Hz
1
230V 60Hz
2
115V 50Hz
3
115V 60Hz
4
JCT
Analysentechnik GmbH
Werner Heisenberg-Straße 4 A-2700 Wiener Neustadt
Tel. +43 (0) 2622 / 87201 Fax +43 (0) 2622 / 872011
E-Mail: [email protected] Web: www.jct.at
01/02