System Components and Accessories for Sample Gas Conditioning Data Sheet

System Components and Accessories for Sample Gas Conditioning
Data Sheet
10/235.20 EN January 2004
A gas analysis system usually consists of two groups of devices:
• The gas analyzer measures the concentrations of the sample
gas components and converts them into an electrical signal.
• Supplementary devices sample the sample gas from the
process, condition it and feed it in the gas analysis system.
Contents
In most cases, the sample gas sampled from the process cannot
be processed by the gas analyzer without further conditioning. Too
high a dust content, temperature and dew point as well as too high
or too low pressure, and not least interference components in the
sample gas can impair the operational capability of the gas analyzer and falsify the measuring result.
Supplementary devices such as the sampling probe, the sample
gas line, the sample gas cooler and pumps and filters ensure that
the sample gas inlet conditions of the connected gas analyzers are
satisfied and faultless measuring results are obtained independently of the process sequence and the local conditions.
SCC-C Sample Gas Cooler
SCC-F Sample Gas Feed Unit
SCC-E Sample Gas Feed Unit
Electrical Cool Trap
CGKA 2 Automatic Condensate Drain
CGMF 2 Sample Gas Feed Unit
Diaphragm Pump 4N
Diaphragm Pump 2
Pre-pressure Controller
SCC-K NO2 / NO Converter
CGO-9 Tubular Furnace
Reaction Element
Membrane Filter
CGKW Condensate Monitor
ER-144A/Ex Switching Unit
Acid Filter
Disposable Filter
Flow Meters, Flow Monitors
KFA Isolated Switch Amplifier
Stripper for Cooling Water and Waste Water
Multi-way Test Gas Cocks
Cylinder Pressure Reducers
Page
2
7
11
14
16
18
21
23
25
27
31
33
34
35
36
37
38
39
41
42
43
44
SCC-C Sample Gas Cooler
! Compressor sample gas cooler
! Constant outlet dew point
! 1 or 2 gas paths
! Heat exchanger made from glass, PVDF or stainless steel,
exchangeable without tools
! Corrosion-resistant gas lines
! High pressure design up to 1 MPa (10 bar)
! Temperature display
! Suitable for ambient temperatures up to 50 °C
! Compact surface-mounting housing (width 1/2 19 inches,
6 height units)
! System solutions in connection with the SCC-F sample
gas feed unit
! Suitable for use in measuring equipment conforming with
17th BImSchV (Federal Immission Protection Law)
Pressure drop in heat exchanger
approx. 1 hPa (1 mbar)
approx. 4...8 hPa (4...8 mbar) for HE125
Sample gas inlet conditions
Sample gas pressure
Heat exchanger Sample gas pressure pabs
material
without
with
peristaltic pump
peristaltic pump
Glass
PVDF
Stainless steel
50...200 kPa
(0.5...2.0 bar)
50...250 kPa
(0.5...2.5 bar)
0.05...1 MPa
(0.5...10 bar)
50...150 kPa
(0.5...1.5 bar)
50...150 kPa
(0.5...1.5 bar)
50...150 kPa
(0.5...1.5 bar)
Sample gas flow rate
1 x 250 l/h (HE250) or 1 x 125 l/h (HE125) or 2 x 125 l/h, with
respect to the sample gas pressure pabs = 100 kPa (1 bar) and
25 °C
Sample gas inlet temperature
max. 140 °C
Sample gas inlet dew point
max. 70 °C,
max. 60 °C for HE250 at sample gas flow rate > 200 l/h
Dead volume in heat exchanger
Heat exchanger material
Glass
PVDF
Stainless steel
Dead volume in heat exchanger
HE125
HE250
40 ml
25 ml
30 ml
140 ml
100 ml
100 ml
Gas tightness
5 x 10–6 hPa l/s
Temperature display
Digital display in °C
Status signal
< 0 °C and > 8 °C, 2 potential-free two-way switches,
switching capacity 250 V AC, 2 A
Refrigerant
R 134a
Materials of gas-conducting parts
Operating data
PVDF, PTFE, PP, glass, Viton, stainless steel (high pressure
version)
Sample gas outlet temperature
Factory-set to +3 °C
Dew point stability
±0.3 °C per 10 °C temperature change,
±0.3 °C per 10 l/h change in flow rate
Refrigeration capacity
40 W at +10...+50 °C
Warm-up time
approx. 15 min
2
Power supply
Input voltage
230 V AC or 115 V AC –15...+10 %, 50...60 Hz
Power consumption
max. 200 VA
Starting current
2.5 A at 230 V AC
System Components and Accessories for Sample Gas Conditioning
10/23-5.20 EN January 2004
SCC-C Sample Gas Cooler
Construction
Electromagnetic compatibility
Design
Surface-mounting housing 1/2 19 inch, 6 height units
Radio interference suppression
Class B to EN 55011
Mounting
On the wall with mounting brackets supplied or
in a 19-inch cabinet / rack
Interference immunity
Testing to EN 50082-2
Housing material and color
Sheet steel; light gray (RAL 7035), basalt gray (RAL 7012)
Testing to EN 60068-2-27 and EN 60068-2-6
Degree of protection of housing
IP 20 to EN 60529
Transport
Shock: 30 g / 18 ms / 18 shocks
Vibration: 1 g / ±0.15 mm / 5...150 Hz, 3 x 5 cycles
Dimensions
see dimensional drawing on page 4
Weight
approx. 15.6 / 16.5 kg with 1/ 2 heat exchangers
Heat exchanger
Heat exchanger Sample gas inlets &
material
outlets
Condensate outlet
Glass
Pipe nipple GL25
Stainless steel
Screwed pipe or
hose connection
GL18
6 mm tube
G 1/4 inch
internal thread
G 3/8 inch
internal thread
G 3/8 inch
internal thread
Peristaltic pump (option)
Condensate outlet PVDF, DN 4/6
Electrical connections
Power supply
Cable, 3 m long, permanently fixed
Status signal
Cable, 3 m long, permanently fixed
Electrical safety
Ambient temperature
Operation: +10...+50 °C,
Storage and transport: –25...+60 °C
Relative humidity
≤ 75 % annual average, occasional and slight condensation
permitted, ≤ 95 % on 30 days per year
Options
Metering device
in gas path 1 by means of a peristaltic pump installed in the
SCC-C sample gas cooler or in the SCC-F sample gas feed
unit, feed tank as accessory, for small SO2 measuring ranges
(< 1000 mg/m3) or to avoid acid aerosols
Peristaltic pumps
1 or 2 peristaltic pumps for discharging condensate
(flow rate approx. 300 ml/h, 5 rpm) and metering (flow rate
approx. 15 ml/h),
Pressure resistance pabs = 50...150 kPa (0.5...1.5 bar),
Peristaltic tube service life approx. 5 months,
Power consumption 3.5 VA,
Weight approx. 0.6 kg
CSA certification
Testing to EN 61010-1
Protective class
I
Overvoltage category / degree of contamination
III / 2
Protective separation
Electrical isolation of the 115 / 230 V AC power supply from
the other current circuits by means of reinforced or double
insulation. Functional extra-low voltages (PELV) on the low
voltage side
10/23-5.20 EN January 2004
Operation
Vibration: 1 g / ±0.07 mm / 5...55 Hz, 3 x 2 cycles
Ambient conditions
Gas connections
PVDF
Mechanical capabilities
The SCC-C sample gas cooler is certified to
– Class 2258 02 Process Control Equipment – For Hazardous
Locations and
– Class 2258 82 Process Control Equipment – For Hazardous
Locations – Certified to U.S. Standards
Class 1, Div. 2, Groups A, B, C and D, Temperature Code T4,
Maximum Ambient Temperature +50 °C.
Approval includes testing per applicable Canadian (CSA) and
US directives.
Certificate No. 1105720
System Components and Accessories for Sample Gas Conditioning
3
SCC-C Sample Gas Cooler
Dimensional drawing (dimensions in mm)
13
1
2
1
3
4
190.5
266
3.0
Temperature display
Condensate outlet (for option with
peristaltic pumps)
Gas connections of heat exchangers
Conduits for the (permanentlyconnected) electrical connection lines
37.8
2
246
263
220
301.5
4
46
3
Circuit and connection diagram
Bridge = 230V Transformer = 115V
Mains L
F1
X1/4
X2/2
L
2A T
R2
U
X2/3
275V
L
X4/3
X4/2
M
~
X5/O/5
M1
1
Pt 100
X4/1
Option
Pt 100
X5/O/3
X2/1
T
T
R1
3
X4/4
C
R
S
X5/O/2
5+
X6/4
X3/3
X5/M/2
Solid
X3/1
6-
C1
4,7µF
A1
X6/3
X5/O/1
X5/U/2
X5/M/3
K1
Alarm
Status 1
X3/2
8
X6/2
9
X6/1
X5/O/4
Fan
M2
X3/6
N
X3/4
MV1
K1
M
~
X5/O/6
X5/O/7
SR25
M4
SR25
M3
M
~
M
~
A2
K1
Status 2
X5/M/5
X5/M/1
X5/M/4
X5/M/7
X5/M/6
X3/5
Mains N
X1/3
Mains PE
X1/2
X5/U/1
X5/U/4
X5/U/3
X1/1
4
System Components and Accessories for Sample Gas Conditioning
10/23-5.20 EN January 2004
SCC-C Sample Gas Cooler
Refrigerant circuit
1
2
3
4
5
6
7
8
1
2
Refrigerant compressor
Refrigerant condenser
Refrigerant dryer
Capillary tube
Heat exchanger
Evaporator
Cooling block with temperature sensor
Valve
3
8
4
9
°C
7
5
6
10/23-5.20 EN January 2004
System Components and Accessories for Sample Gas Conditioning
5
SCC-C Sample Gas Cooler
Ordering information
Catalog No. 23070-0↓
Design
Wall-mounted housing 1/2-19 Inch
↓
↓
↓
↓
↓
↓
0
↓
0
↓
0
↓
0
↓
↓
0
↓
0
↓
0
↓
0
1
Preparation
without
with dosing unit
0
1
Heat exchanger
Glass
PVDF
Stainless steel
1
2
3
Gas paths
1 gas path 250 l/h
1 gas path 125 l/h
2 gas paths 2 x 125 l/h
1
2
3
Condensate drain
without
1 peristaltic pump
2 peristaltic pumps (2 x condensate)
2 peristaltic pumps (1 x condensate, 1 x dosing)
2 peristaltic pumps (1 x condensate, 1 x precondensation)
0
1
2
3
4
Display
Temperature display and monitoring
Temperature display and monitoring and Pt 100
1
2
Power supply
230 V, 50/60 Hz
115 V, 50/60 Hz
1
2
Certification
without
with CSA certification
0
1
↓
↓
↓
↓
Catalog No. 23070-0-
6
↓
System Components and Accessories for Sample Gas Conditioning
↓
↓
↓
↓
10/23-5.20 EN January 2004
SCC-F Sample Gas Feed Unit
! Complete sample gas conditioning in connection with the
SCC-C sample gas cooler
! 2 independent gas paths
! Corrosion resistant, metal-free gas lines
! Compact surface-mounting housing (width 1/2 19 inches,
6 height units)
! Service-friendly due to easy access to the modules
Sample gas inlet conditions
Feed performance standard
Sample gas pressure
pabs = 70...105 kPa (0.7...1.05 bar)
Flow rate / l/h
-600
-400
h
-2 Pa
00
hP
a
a
hP
100
-1
00
-4
00
hP
a
-200
0
200
Pressure / hPa
400
600
800
A Inlet: negative pressure, outlet: atmospheric pressure
B Inlet: acc. to chart, outlet: gauge pressure
Operating data
Warm-up time
approx. 10 minutes (plus warm-up time of sample gas
cooler)
Feed performance bypass
350
300
Dead volume
approx. 10 cm3 (plus dead volume of heat exchanger)
Flow rate display and adjustment
Rotameter 10...100 l/h with needle valve (metal-free)
Status signals
Flow rate and condensate monitors: 2 potential-free two-way
switches, switching capacity 250 V AC, 2 A
250
Flow / l/h
Gas tightness
5 x 10–6 hPa l/s
10/23-5.20 EN January 2004
150
50
-800
B
00
-3
Sample gas inlet dew point
The dew point of the sample gas must be at least 5 °C lower
than the lowest ambient temperature in the whole sample
gas path. Otherwise a sample gas cooler or a condensate
trap is required.
200
a
hP
Sample gas inlet temperature
+10...+50 °C
Inlet pressure
0
Sample gas flow rate
Gas paths 1 and 2 max. 100 l/h
Bypass
max. 200 l/h
250
A
200
150
100
50
-1000
-500
0
500
Pressure / hPa
System Components and Accessories for Sample Gas Conditioning
1000
1500
2000
2500
7
SCC-F Sample Gas Feed Unit
Material of gas-conducting parts
Electromagnetic compatibility
PVDF, PTFE, PVC, EPDM, Viton, Novoprene
Radio interference suppression
Class B to EN 55011
Power supply
Interference immunity
Testing to EN 50082-2
Input voltage
230 V (207...250 V) AC, 50 Hz or 60 Hz ± 2 Hz
115 V (100...135 V) AC, 50 Hz or 60 Hz ± 2 Hz
Mechanical capabilities
Power consumption during operation
Diaphragm pump approx. 50 VA,
metering pump approx. 3.5 VA
Testing to EN 60068-2-27 and EN 60068-2-6
Construction
Operation
Vibration: 1 g / ±0.07 mm / 5...55 Hz, 3 x 2 cycles
Design
Surface-mounting housing 1/2 19 inch, 6 height units
Mounting
On the wall with the mounting brackets supplied or
in a 19 inch cabinet / rack
Housing material and color
Sheet steel; light gray (RAL 7035), basalt gray (RAL 7012)
Degree of protection of housing
IP 20 to EN 60529
Dimensions
see dimensional drawing on page 9
Gas connections
PVDF coupling for tube of 6 x 4 x 1 mm
Electrical connections
Power supply
Screw clamp terminals for stranded or solid wire with
max. 1.5 mm2
Status signals
Screw clamp terminals for stranded or solid wire with
max. 1.5 mm2
Electrical safety
Transport
Shock: 30 g / 18 ms / 18 shocks
Vibration: 1 g / ±0.15 mm / 5...150 Hz, 3 x 5 cycles
Ambient conditions
Ambient temperature
Operation: +10...+50 °C,
Storage and transport: –25...+60 °C
Relative humidity
≤ 75 % annual average, occasional and slight condensation
permitted, ≤ 90 % on 30 days/year
Options
Metering unit
By means of peristaltic pump, flow rate 15 ml/h,
feed tank as accessory
for small SO2 measuring ranges (< 1000 mg/m3) or
to avoid acid aerosols
I/O card
To connect the SCC-F sample gas feed unit and the SCC-C
sample gas cooler to the AO2000 Series continuous gas
analyzers via the system bus
– Display and monitoring of the cooler temperature
– Monitoring and processing of status signals of the SCC-F
sample gas feed unit (condensate, flow rate)
– Control of an external solenoid valve for zero gas
connection, 230 V, 50 Hz
– Digital inputs for monitoring condensate collecting bottles
and reagent reservoirs
Testing to EN 61010-1
CSA certification
Protective class
I
Overvoltage category / degree of contamination
III / 2
Protective separation
Electrical isolation of the 115 / 230 V AC power supply from
the other current circuits by means of reinforced or double
insulation. Functional extra-low voltages (PELV) on the low
voltage side
The SCC-F sample gas feed unit is certified to
– Class 2258 02 Process Control Equipment – For Hazardous
Locations and
– Class 2258 82 Process Control Equipment – For Hazardous
Locations – Certified to U.S. Standards
Class 1, Div. 2, Groups A, B, C and D, Temperature Code T4,
Maximum Ambient Temperature +50 °C.
Approval includes testing per applicable Canadian (CSA) and
US directives.
Certificate No. 1105720
8
System Components and Accessories for Sample Gas Conditioning
10/23-5.20 EN January 2004
SCC-F Sample Gas Feed Unit
Dimensional drawing (dimensions in mm)
Z
Ansicht / View Z
20
220
2
6
7
4
37.8
Sample gas inlet
Sample gas outlet
Electrical connections
Flow rate monitor with needle valve
Metering pump
Condensate monitor
LED display and reset button of
switching amplifiers
301.5
266
190.5
6
1
1
2
3
4
5
6
7
4
5
246
46
263
Ordering information
Catalog No. 23212-0↓
Design
Wall-mounted housing 1/2-19 Inch
↓
↓
↓
↓
0
↓
0
↓
0
↓
0
↓
0
↓
0
↓
↓
0
↓
0
↓
0
↓
0
↓
0
↓
0
1
Gas paths
1 gas path, 1 condensate monitor, 1 flow monitor, 1 diaphragm pump
2 separate gas paths, 2 condensate monitors, 2 flow monitors, 2 diaphragm pumps
1 gas path, 1 condensate monitor, 2 flow monitors, 1 diaphragm pump
1 gas path, 1 condens. monitor, 2 flow monitors, 1 diaphr. pump, 1 pressure controller
2 separate gas paths, 2 condensate monitors, 2 flow monitors, 1 diaphragm pump
2 separate gas paths, 2 condensate monitors, 1 flow monitor, 1 diaphragm pump
1
2
3
4
5
6
Dosing
without
with dosing pump
0
1
Electronic processing unit
with 1 electronic processing unit
with 2 electronic processing units
with I/O interface board
1
2
3
Power supply
230 V, 50/60 Hz
115 V, 50/60 Hz
1
2
Certification
without
with CSA certification
0
1
↓
↓
↓
↓
Catalog No. 23212-0-
10/23-5.20 EN January 2004
↓
System Components and Accessories for Sample Gas Conditioning
↓
↓
9
SCC-F Sample Gas Feed Unit
Pneumatic diagrams
Catalog number 23212-0-11XXXX000000
Catalog number 23212-0-12XXXX000000
Catalog number 23212-0-13XXXX000000
Catalog number 23212-0-14XXXX000000
Catalog number 23212-0-15XXXX000000
Catalog number 23212-0-16XXXX000000
10
System Components and Accessories for Sample Gas Conditioning
10/23-5.20 EN January 2004
SCC-E Sample Gas Feed Unit
! Process-proven modules
! Ideal for processes with non-corrosive gases
! Integrated Peltier cooler and peristaltic pump (option)
! Simple installation und connection
! Portable housing for rack and wall mounting
(1/2 19 inch, 4 height units)
! Attractively priced system solution with
EL1020 gas analyzer
Sample gas inlet conditions
Power supply
Sample gas pressure
pabs = 70...105 kPa (0.7...1.05 bar)
Input voltage
230 V AC, 50 Hz or 115 V AC, 60 Hz
Sample gas flow rate
7...70 l/h
Power consumption during operation
230 V AC, 50 Hz: 1.05 A, 240 VA
Sample gas inlet temperature
+10...+50 °C
Construction
Sample gas inlet dew point (Peltier cooler)
max. 50 °C
Operating data
Outlet dew point (Peltier cooler)
+3 °C
Warm-up time
approx. 10 minutes with Peltier cooler
Gas tightness
5 x 10–6 hPa l/s
Flow rate display and adjustment
Rotameter 10...100 l/h with needle valve (metal-free)
Status signals
Cooler, flow and condensate monitoring: 1 potential-free
two-way switch, switching capacity 250 V AC, 2 A
Design
Surface-mounting housing 1/2 19 inch, 4 height units
Mounting
On the wall or in a 19 inch cabinet / rack with the mounting
brackets supplied
Housing material and color
Sheet steel; light gray (RAL 7035), basalt gray (RAL 7012)
Degree of protection of housing
IP 20 to EN 60529
Dimensions
see dimensional drawing on page 13
Weight
approx. 11.2 kg
Gas connections
PPH couplings for tube 6 x 4 x 1 mm
Material of gas-conducting parts
PPH, EPDM, Viton, Novoprene, glass
Electrical connections
Power supply
Cable, length 1.5 m, permanently fixed
Status signal
Cable, length 1.5 m, permanently fixed
10/23-5.20 EN January 2004
System Components and Accessories for Sample Gas Conditioning
11
SCC-E Sample Gas Feed Unit
Electrical safety
Pneumatic diagrams
Testing to EN 61010-1
Version without cooler
Protective class
I
2
3
Overvoltage category / degree of contamination
III / 2
Protective separation
Electrical isolation of the 115 / 230 V AC power supply from
the other current circuits by means of reinforced or double
insulation. Functional extra-low voltages (PELV) on the low
voltage side
1
Electromagnetic compatibility
Radio interference suppression
Class B to EN 55011
Version with Peltier cooler and condensate pump
Interference immunity
Testing to EN 50082-2
2
3
Mechanical capabilities
Testing to EN 60068-2-27 and EN 60068-2-6
1
Transport
Shock: 30 g / 18 ms / 18 shocks
Vibration: 1 g / ±0.15 mm / 5...150 Hz, 3 x 5 cycles
4
Operation
Vibration: 1 g / ±0.07 mm / 5...55 Hz, 3 x 2 cycles
Ambient conditions
Ambient temperature
Operation: +10...+45 °C,
Storage and transport: –25...+60 °C
1
2
3
4
Sample gas inlet
Test gas inlet
Sample gas outlet
Condensate outlet
Relative humidity
≤ 75 % annual average, occasional and slight condensation
permitted, ≤ 90 % on 30 days/year
12
System Components and Accessories for Sample Gas Conditioning
10/23-5.20 EN January 2004
SCC-E Sample Gas Feed Unit
Dimensional drawing (dimensions in mm)
10
7
263
220
3.0
10
Alarm Pump
5
6
7
101,5
9
177
8
Sample Gas
Test Gas
SCC-E
37,75
11
246
1) only in version with Peltier cooler
and condensate pump
12
185
220
Sample gas inlet
Test gas inlet
Sample gas outlet
Power supply and status signal cables
Flow rate monitor with needle valve
Peristaltic pump 1)
Condensate monitor 1) / diaphragm filter
LED indicator condensate alarm red 1)
Pump switch
Temperature controller for cooler 1)
3-way cock sample gas / test gas
Condensate outlet 1)
4 x M5 threaded holes in ground plate
47,5
300
372,5
1
2
4
3
17,5
1
2
3
4
5
6
7
8
9
10
11
12
13
25
13
248
36
Ordering information
Catalog No. 23235-0↓
↓
↓
with 3-way cock, diaphragm filter, pump and flow monitor with needle valve
with 3-way cock, condensate monitor, pump and flow monitor with needle valve
1
2
0
1
Cooler and condensate pump
without
with Peltier cooler and condensate pump
1
2
0
1
Design
Wall-mounted housing 1/2-19 inch
↓
0
↓
0
↓
0
↓
0
↓
0
↓
0
↓
↓
0
↓
0
↓
0
↓
0
↓
0
↓
0
↓
0
↓
0
1
2
↓
↓
↓
↓
Catalog No. 23235-0-
10/23-5.20 EN January 2004
0
↓
1
Power supply
230 V, 50 Hz
115 V, 60 Hz
Accessories
Carry handle
0
↓
Catalog No. 23009-4-0730702
System Components and Accessories for Sample Gas Conditioning
13
Electrical Cool Trap
■ Sample gas cooling with Peltier element
■ Heat exchanger made of PVDF
■ Status signals for too high or too low temperature
■ Automatic condensate drain as an option
Sample gas inlet conditions
Construction
Sample gas pressure
pabs = 40...220 kPa (0.4...2.2 bar) at sample gas inlet
temperature 150 °C
Design
Surface-mounting housing (1/2 19-inch) for wall mounting
Sample gas flow rate
150 l/h related to 100 kPa (1 bar) and 25 °C at sample gas
inlet temperature 150 °C
Sample gas inlet dew point
≤ 65 °C
Housing material
Aluminum sheet, anodized without color
Degree of protection of housing
IP 20 to EN 60529
Dimensions
see dimensional drawing on page 15
Operating data
Weight
approx. 7 kg
Sample gas outlet temperature
+10 °C
Gas connections
Refrigeration capacity
> 54 kJ at ambient temperature 25 °C
Timer for the automatic condensate drain
Run time 2 min, off period 20 min
Dead volume of the heat exchanger
approx. 50 ml
Sample gas
Pipe, 6 x 1 mm
Condensate drain
Pipe, 12 x 1 mm
in version with peristaltic pump: hose 6 x 1 mm
Electrical connections
Buffer volume for condensate
approx. 10 ml
Power supply
Cable with grounding plug, length 5 m
Materials of gas-conducting parts
PVDF, FEP
Status signal (optional)
Sub-miniature socket, 9-pin, with plug and cable, length 5 m
Power supply
Input voltage
230 V AC –13...+5 %, 50 Hz
115 V AC –13...+5 %, 50 Hz or 60 Hz
with integral peristaltic pump only 50 Hz; 60 Hz on request
Power consumption
approx. 70 VA
14
Mechanical capabilities
Testing to EN 60068-2-27 and EN 60068-2-6
Transport
Shock: 30 g / 18 ms / 18 shocks
Vibration: 2 g / ±0.15 mm / 5...150 Hz, 3 x 5 cycles
Operation
Vibration: 1 g / ±0.07 mm / 5...55 Hz, 3 x 2 cycles
System Components and Accessories for Sample Gas Conditioning
10/23-5.20 EN January 2004
Electrical Cool Trap
Ambient conditions
Dimensional drawing (dimensions in mm)
Ambient temperature
Operation: +5...+35 °C
Storage and transport: –25...+60 °C
320
42 70
Relative humidity
≤ 75 % annual average, slight and occasional condensation
permissible
Climatic class
3K4 to EN 60721-3-3
Ordering information
230
190
21
ø8.5
10
300
208
Catalog No.
Electrical cool trap, 230 V, 50 Hz
23058-5-8018055
Electrical cool trap with automatic
condensate drain, 230 V, 50 Hz
23058-5-8018054
Electrical cool trap, 115 V, 60 Hz
23058-5-8018463
Electrical cool trap with automatic
condensate drain, 115 V, 60 Hz
23058-5-8018462
Circuit and connection diagram
10/23-5.20 EN January 2004
System Components and Accessories for Sample Gas Conditioning
15
CGKA 2 Automatic Condensate Drain
■ For connection to any sample gas cooler
■ Version with level contact as an option
■ Time relay available as an accessory
Condensate inlet conditions
Power supply
Inlet pressure
pabs = 50...150 kPa (0.5...1.5 bar) with unpressurized outlet
Input voltage
220 V AC –15...+10 %, 50 Hz, or –5...+10 %, 60 Hz
110 V AC –10...+10 %, 50 Hz or 60 Hz
Condensate temperature
+1...+70 °C
Power consumption
7 VA
Viscosity of the condensate
approx. 50...175 Pa s
Construction
Operating data
Outlet pressure
pabs = 50...150 kPa (0.5...1.5 bar) with unpressurized inlet
Delivery
1 l/h ± 10 %, max. 0.4 l/h at 40 % cyclic duration factor
Buffer volume
0.2 l
Design
Surface-mounting housing for wall mounting
Housing material and color
Polycarbonate, RAL 7032
Degree of protection of housing
Motor housing: IP 54, wheel head with plastic top: IP 20 to
EN 60529
Dimensions
see dimensional drawing on page 17
Rotational speed of pump
20 RPM at 50 Hz
Gas tightness
pabs = 50...150 kPa (0.5...1.5 bar) from the pressure to the
suction side in every wheel head position
Weight
approx. 1.7 kg
Service life of hose
Replace hose as a preventive measure every six months
Gas connections
Materials of gas-conducting parts
Condensate drain
Bulkhead screwed connection DN 4/6 for hose 6 ∅ 1 mm
Verderprene, PP, PVC, Norprene, LLDPE
Condensate inlet
Connecting branches for hose 15.9 ∅ 3.2 mm
Electrical connections
Terminals for stranded or solid wire with max. 1.5 mm2,
feed-through via membrane nipple PG 9
16
System Components and Accessories for Sample Gas Conditioning
10/23-5.20 EN January 2004
CGKA 2 Automatic Condensate Drain
Mechanical capabilities
Dimensional drawing (dimensions in mm)
Testing to EN 60068-2-27 and EN 60068-2-6
Transport
Shock: 30 g / 11 ms / 18 shocks
Vibration: 2 g / ±0.17 mm / 5...150 Hz, 3 x 5 cycles
Operation
Vibration: 0.5 g / ±0.04 mm / 5...55 Hz, 3 x 2 cycles
Ambient conditions
Ambient temperature
Operation: +5...+50 °C intermittent mode with cyclic
duration factor ≤ 40 %, i.e. run time max. 5 minutes and off
period approx. 7.5 minutes
Storage and transport: –25...+60 °C
Relative humidity
≤ 75 % annual average, occasional and slight condensation
permissible
Climatic class
3K4 to EN 60721-3-3
Level contact (optional)
1
2
3
4
5
6
Condensate inlet
Condensate outlet
Pump inlet
Power supply
Mounting holes
Outer link
“Overflow” status signal
Floating contact, rating 24 V / 1 A
Ordering information
Accessories for condensate disposal
CGKA 2 220 V, 50 Hz
23227-5-8018441
Time relay
CGKA 2 220 V, 60 Hz
23227-5-8018442
Run time and off period settings
0.6...60 sec., recommended setting: cyclic duration factor
40 %, i.e. run time 5 minutes and off period 7.5 minutes
CGKA 2 110 V, 50 Hz
23227-5-8018446
CGKA 2 115 V, 60 Hz
23227-5-8018443
CGKA 2 with level monitor 220 V, 50 Hz
23227-5-8018445
Degree of protection
Terminals IP 20, housing IP 50 to EN 60529
CGKA 2 with level monitor 220 V, 60 Hz
23227-5-8018444
Weight approx. 0.125 kg
CGKA 2 with level monitor 110 V, 50 Hz
23227-5-8018448
CGKA 2 with level monitor 115 V, 60 Hz
23227-5-8018447
Catalog No.
Power supply
12...240 V AC/DC, 40...440 Hz
Accessories
Timer relay
90805-4-0741607
Reagent and condensate bottle
23234-5-8328972
Sample gas pressure pabs = 70...150 kPa (0.7...1.5 bar)
Level monitor condensate
23005-4-0741239
Material LDPE / PVC
Level monitor reagent
Condensate collecting bottle
Volume 10 l
Connections
Sample gas: 1 connecting piece for hose with ID 4–5 mm
Condensate: 2 connecting pieces for hose with ID 12–13 mm
Condensate drain: SW 22
Condensate trap, filter surface 20 cm
90805-5-8308994
2
23065-5-8018512
Weight approx. 1 kg
Level monitor
Switching voltage 30 V
Switching current 1 A
Switching power 30 VA
Connecting cable 2 x 0.25 mm2, length 5 m
10/23-5.20 EN January 2004
System Components and Accessories for Sample Gas Conditioning
17
CGMF 2 Sample Gas Feed Unit
! Feeding-in the sample gas
! Micro-porous filtration of the sample gas
! Monitoring of the sample-gas path for condensate
breakthrough with fault status signaling
! Adjustment and monitoring the sample gas flow rate with
alarm value signaling
! Compact unit with degree of protection of housing IP 20
Operating data
Structure
Inlet pressure
pabs = 70...105 kPa (0.7...1.5 bar)
Sample gas feed-in
Diaphragm pump 4N (see page 21)
Flow rate
max. 250 l/h at atmospheric pressure
Micro-porous filtration
Diaphragm filter in PVC
Inlet temperature
+5...+50 °C
Condensate monitoring
Diaphragm filter in PVC, monitoring of the diaphragm filter
for moisture by means of electrodes (conductivity measurement); fault status signaling with separate amplifier relay
ER 144A/Ex (see page 36)
Inlet dew point
below room temperature
Delivery
see delivery diagram of the diaphragm pump 4N on page 21
Flow meter
Flow rate 7...70 l/h, Bypass 25...250 l/h
Diaphragm filter
Retention rate 100 % for particles > 1 µm
Tightness
Pressure drop 0.1 hPa/min
Flow rate monitoring
Hovering body flow meter with inductive sensor for alarm
value monitoring and needle valve; alarm value signaling
with separate switching unit KFA (see page 41)
Construction
Design
Surface-mounting case, motor axis horizontal
Materials of the parts in contact with medium
Housing material and color
Plastic, RAL 7035
ETFE (ethylene-tetrafluorethylene), PP (polypropylene), FPM
(Viton), PVC (polyvinyl chloride), EPDM (ethylene-propylene),
special steel
Degree of protection of housing
IP 20 to EN 60529
Power supply
Dimensions
see dimensional drawing on page 19
AC voltage supply
230 V ± 10 %, 50 Hz or 60 Hz
115 V ± 10 %, 50 Hz or 60 Hz
Power consumption
approx. 50 VA
Class of protection
I
Weight
approx. 2.3 kg
Gas connections
G 1/4 female thread (DIN ISO 228/1) for male fittings, nozzles
for hose internal diameter 4 mm are supplied with the unit
18
System Components and Accessories for Sample Gas Conditioning
10/23-5.20 EN January 2004
CGMF 2 Sample Gas Feed Unit
Electrical connections
Ambient conditions
Terminals via cable connection PG 13.5
Ambient temperature
Operation: +5...+50 °C
Storage and transport: –25...+65 °C
Mechanical capabilities
Testing to DIN EN 60068-2-27 and DIN EN 60068-2-6
Transport
Shock 30 g / 18 ms / 18 shocks
Vibration: 2 g / ±0.15 mm / 5...150 Hz, 3 x 5 cycles
Climatic class
3K4 to EN 60721-3-3
Operation
Vibration: 2 g / ±0.07 mm / 4...7 Hz, 3 x 2 cycles
Dimensional drawing (dimensions in mm)
Electrical connection diagram
~165
45
11
Relative humidity
≤ 75 % annual mean, infrequent and light condensation
permitted
Z
3
32
17
60
32
95
140
Z
Switching Unit
ER 144A
BN +
BU -
Switching Unit
KFA.-SR2-Ex1.W
~150
2
1
3
29
RD
WH
+
16
148
160
6
L1
N
~150
88
120
Ø 4,2
1
Gas connections
2
Electrical connections
3
Needle valve
U = 115 / 230 V AC
L1
N
PE
Z-15193
Z-15199
Ordering information
Catalog No. 23229-0↓
Degree of protection
IP 20 (ambient temperature max. +50 °C)
↓
↓
↓
0
↓
0
↓
0
↓
0
↓
0
↓
0
↓
0
↓
↓
0
↓
0
↓
0
↓
0
↓
0
↓
0
↓
0
2
Gas outlets
1 sample gas outlet
1 sample gas outlet + 1 bypass outlet
1
2
Flow monitoring
without
with (switching unit KFA required, see page 41)
0
1
Condensate monitoring
without
with (switching unit ER 144A/Ex required, see page 36)
0
1
Power supply
230 V, 50/60 Hz
115 V, 50/60 Hz
1
2
↓
↓
↓
↓
Catalog No. 23229-0-
10/23-5.20 EN January 2004
↓
System Components and Accessories for Sample Gas Conditioning
↓
19
CGMF 2 Sample Gas Feed Unit
Pneumatic diagrams
Catalog number 23229-0-2100X0000000
Catalog number 23229-0-2101X0000000
Catalog number 23229-0-2110X0000000
Catalog number 23229-0-2111X0000000
Catalog number 23229-0-2200X0000000
Catalog number 23229-0-2201X0000000
Catalog number 23229-0-2210X0000000
Catalog number 23229-0-2211X0000000
20
System Components and Accessories for Sample Gas Conditioning
10/23-5.20 EN January 2004
Diaphragm Pump 4N
! Feed in of the sample gas
! Compact unit with degree of protection of housing IP 20
or IP 54
Operating data
Materials of the parts in contact with medium
Inlet pressure
pabs = 70...105 kPa (0.7...1.5 bar)
Valve plate, diaphragm: EPDM (ethylene-propylene);
pump body: PP (polypropylene); hose: FPM (Viton);
nozzles: ETFE (ethylene-tetrafluorethylene)
Flow rate
max. 250 l/h at atmospheric pressure
Power supply
Inlet temperature
+5...+50 °C
AC voltage
230 V ± 10 %, 50 Hz or 60 Hz
115 V ± 10 %, 50 Hz or 60 Hz
Inlet dew point
below room temperature
Feed performance
Power consumption
approx. 50 VA
250
Flow rate / l/h
100
-400
Construction
a
a
a
hP hP hP
00 200 300
-1
-
150
-600
B
a
hP
200
50
-800
Inlet pressure
0
A
-4
00
Design
Surface-mounting case for wall attachment, motor axis
horizontal
Housing material and color
Plastic, RAL 7035
hP
a
-200
0
200
Pressure / hPa
400
600
800
A Inlet: negative pressure, outlet: atmospheric pressure
B Inlet: according diagram, outlet: negative pressure
Tightness
Pressure drop 0.1 hPa/min
Degree of protection of housing
IP 20 or IP 54 to EN 60529
Class of protection
I
Dimensions
see dimensional drawing on page 22
Weight
approx. 1.8 kg
Gas connections
G 1/4 female thread (DIN ISO 228/1) for male fittings, nozzles
for hose inside diameter 4 mm are supplied with the unit.
10/23-5.20 EN January 2004
System Components and Accessories for Sample Gas Conditioning
21
Diaphragm Pump 4N
Electrical connections
Dimensional drawing (dimensions in mm)
Terminals via cable connection PG 13.5
Mechanical capabilities
Z
Testing to DIN EN 60068-2-27 and DIN EN 60068-2-6
Transport
Shock: 30 g / 18 ms / 18 shocks
Vibration: 2 g / ±0.15 mm / 5...150 Hz, 3 x 5 cycles
Operation
Vibration: 2 g / ±0.07 mm / 4...7 Hz, 3 x 2 cycles
2
1
17
60
32
32
95
140
Z
88
120
16
Ambient temperature
Operation: IP 54 +5...+40 °C, IP 20 +5...+50 °C
Storage and transport: –25...+65 °C
Relative humidity
≤ 90 % annual average, condensation permitted
Climatic class
3K4 to EN 60721-3-3
~150
ø 4,2
Ambient conditions
148
160
6
1
Gas connections
2
Electrical connections
Z-14185
Ordering information
Catalog No.
Diaphragm Pump 4N, IP 20; 230V, 50/60 Hz 23134-5-8018545
Diaphragm Pump 4N, IP 20; 115V, 50/60 Hz 23134-5-8018546
Diaphragm Pump 4N, IP 54; 230V, 50/60 Hz 23134-5-8018547
Diaphragm Pump 4N, IP 54; 115V, 50/60 Hz 23134-5-8018548
22
System Components and Accessories for Sample Gas Conditioning
10/23-5.20 EN January 2004
Diaphragm Pump 2
! Extremely robust version thanks to cast metal cladding
! Very reliable operationally and easy to maintain thanks to
the simple method of operation
! Delivery adjustable using magnetic shunt
Operating data
Power supply
Delivery
max. 300 l/h at atmospheric pressure, under other pressure
conditions: see delivery diagram.
The diaphragm is moved by an electric magnet. The rate of
flow can be changed by using a magnetic shunt.
Power supply
220 V AC, 50 Hz
300
l/h
Construction
A
I
I
B
200
II
II
150
100
Z-2777
50
-0+
50
Atmospheric pressure
100 hPa 150
p
Diaphragm pump 2
Diaphragm pump 2-Ex (not available)
Inlet: negative pressure, outlet: atmospheric pressure
Inlet: atmospheric pressure, outlet: negative pressure
Time-related behavior
Flow rate
30 l/h
60 l/h
90 l/h
Dead time
0.9 s
0.7 s
0.6 s
Time constant
1.1 s
1.0 s
0.9 s
Design
Surface-mounting case for wall attachment
Housing material
Cap and lower part in light alloy casting
100
I
II
A
B
Power consumption
max. 13 VA, depending on adjustment of the delivery
Pressure drop
0.4 hPa
1.1 hPa
1.8 hPa
Degree of protection of housing
IP 54 to EN 60529
Weight
approx. 3 kg
Gas connections
G 1/4 female thread (DIN ISO 228/1) for male fittings, nozzles
for hose inside diameters 4 mm are supplied with the unit
Electrical connections
Terminals via cable connection PG 11
Materials of the parts in contact with medium
Ambient conditions
Diaphragm: EPDM; valves: FPM (Viton); pump body: PVDF;
hose: PVC; nozzles: PTFE (Hostaflon)
Ambient temperature
Operation: +5...+45 °C, storage and transport: –25...+65 °C
Relative humidity
≤ 90 % annual average, condensation permitted
Climatic class
3K4 to EN 60721-3-3
10/23-5.20 EN January 2004
System Components and Accessories for Sample Gas Conditioning
23
Diaphragm Pump 2
Ordering information
Dimensional drawing (dimensions in mm)
Catalog No.
Diaphragm Pump 2, 220 V, 50 Hz
23121-5-5861212
~170
2
50
1
PG 11
65
160
192
~90
Z-20374
24
O 8,4
1
Sample gas connections
2
Gate valve for regulating
the feed quantity
System Components and Accessories for Sample Gas Conditioning
10/23-5.20 EN January 2004
Pre-pressure Controller
! Keeps the sample gas pressure constant irrespective of
flow resistances such as filter blockage and long samplegas lines
! Prevents jumps in pressure in the sample gas feed-in
system, so that the pump service life is increased
! Use as a bypass controller with a sample gas flow rate of
60 l/h (normal case)
! Use as flow-off controller, where short dead times must
be achieved
Operating data
Construction
Inlet pressure
pabs = 750...1000 hPa (mbar)
Mounting
Wall mounting with support bracket (supplied),
vertical or horizontal
Inlet temperature
+5...+70 °C
Housing material
Brass, nickel plated
Flow rate (outflow volume or bypass)
50...250 l/h with an adjusted outlet pressure of 100 hPa
(mbar)
Dimensions
see dimensional drawing on page 26
Outlet pressure
80...120 hPa (mbar) positive pressure, adjustable
Weight
0.68 kg
Leak rate
≤ 10–4 mbar l/s
Gas connections
G 1/8 female thread (DIN ISO 228/1) for male fittings, nozzles
for hose internal diameter 4 mm are supplied with the unit
120
l/h
100
Q = 120 l/h
Mechanical capabilities
Testing to DIN EN 60068-2-27 and DIN EN 60068-2-6
Transport
Shock: 30 g / 11 ms / 18 shocks
Vibration: 2 g / ±0.15 mm / 5...150 Hz, 3 x 5 cycles
80
Q = 60 l/h
60
40
Z-20077
Sample gas flow rate Q
Flow rate characteristic of a bypass controller
750
800
850
900
950 hPa 1000
Pump inlet pressure pabs
Materials of the parts in contact with medium
PVDF, PTFE glass-fiber reinforced, Viton
10/23-5.20 EN January 2004
Operation
Vibration: 1 g / ±0.07 mm / 5...55 Hz, 3 x 2 cycles
Ambient conditions
Ambient temperature
Operation: +5...+70 °C,
Storage and transport: –25...+70 °C
Relative humidity
≤ 90 % annual average, condensation permitted
System Components and Accessories for Sample Gas Conditioning
25
Pre-pressure Controller
Dimensional drawing (dimensions in mm)
Use options
50
40
0
ø4
17
27
Bypass controller
Q1
G 1/8
5
ø40
40
65
ø6
14
ø9
30
max. 115
5
Q2
MG
PG 1
PG 2
PG 3
Z-20075/1
M5
PG 4
Flow-off controller
Q1
50
Ordering information
Catalog No.
Pre-pressure Controller
26
MG
PG 1
PG 2
PG 3
PG 4
Q2
Z-20075/2
Z-20076
23034-4-0801710
System Components and Accessories for Sample Gas Conditioning
10/23-5.20 EN January 2004
SCC-K NO2 /NO Converter
! Catalytic converter based on carbon-molybdenum
! High catalyst service life
! Catalyst cartridge can be changed without tools
! Option: Version with heated sample gas inlet
Operating data
Construction
Sample gas flow rate
max. 150 l/h
Design
19 inch slide-in housing, 3 height units
Temperature range
depending on sample gas flow rate:
30 l/h: 320 °C; 60 l/h: 320 °C; 90 l/h: 340 °C, 150 l/h: 360 °C
Mounting
In 19 inch cabinet / rack or with mounting brackets
(accessories) on the wall, distance above min. 1 height unit
Effectivity
≥ 95 % with new catalyst
Housing material and color
Sheet steel; light gray (RAL 7035), basalt gray (RAL 7012)
Sample gas pressure pabs
≤ 200 kPa (2 bar)
Degree of protection of housing
IP 20 to EN 60529
Pressure drop
≤ 0.01 kPa (10 mbar) at 90 l/h
Dimensions
see dimensional drawing on page 29
Warm-up time
approx. 30 min
Weight
approx. 8...9 kg
90% time T90
≤ 10 s at 60 l/h
Gas connections
Status signal for temperature deviation
1 NO potential-free contact,
capacity 24 V DC, 1 A
Materials of gas-conducting parts
Sample gas inlet and outlet unheated
G 1/4 female thread to DIN ISO 228/1
Heated sample gas inlet
Swagelok® screwed connection for pipes with 6 mm outer
diameter
Stainless steel W-No. 1.4571, PVDF, PTFE, Viton
Electrical connections
Power supply
Input voltage
230 V AC or 115 V AC –15...+10 %, 48...62 Hz
Power consumption
230 V AC: 350 VA; 115 V AC: max. 170 VA
10/23-5.20 EN January 2004
Power supply
Connector for non-heating apparatus, cable with grounding
plug, length 2 m
Status signal
9 pin Sub-D female connector with counter-plug
System Components and Accessories for Sample Gas Conditioning
27
SCC-K NO2 /NO Converter
Electrical safety
Service life of catalyst
Testing to EN 61010-1
The service life of the catalyst depends on its operating temperature as well as the flow rate and the NO2 and O2 content of
the sample gas. It can be estimated with the help of the charts
below.
Protective class
I
Protective separation
Electrical isolation of the 115 / 230 V AC power supply from
the other current circuits by means of reinforced or double
insulation. Functional extra-low voltages (PELV) on the low
voltage side
Service life / days
Overvoltage category / degree of contamination
III / 2
700
600
500
400
300
200
100
0
Electromagnetic compatibility
30 l/h
60 l/h
90 l/h
5
Radio interference immunity
Class B to EN 55011
10 15 20 25 30 35 40 45 50
NO2 / ppm
5 Vol.-% O2
Service life / days
Interference immunity
Testing to EN 50082-2
Ambient conditions
Ambient temperature
Operation: +10...+50 °C,
Storage and transport: –25...+65 °C
Relative humidity
≤ 75 % annual mean, occasional and slight condensation
permitted
450
400
350
300
250
200
150
100
50
0
30 l/h
60 l/h
90 l/h
5
10 15 20 25 30 35 40 45 50
NO2 / ppm
10 Vol.-% O2
The SCC-K NO2 / NO converter is certified to
– Class 2258 01 Process Control Equipment and
– Class 2258 81 Process Control Equipment – Certified to
U.S. Standards
Approval includes testing per applicable Canadian (CSA) and
US directives.
Certificate No. 1105720
Service life / days
CSA certification
250
200
30 l/h
60 l/h
150
100
90 l/h
50
0
5
10 15 20 25 30 35 40 45 50
NO2 / ppm
Ordering information
21 Vol.-% O2
Catalog No.
Converter SCC-K, cock PVDF, 230 VAC
23093-4-0801974
Converter SCC-K, 2 solen. valves, 230 VAC 23093-4-0801975
Converter SCC-K, heated, 1 s.v., 230 VAC
23093-4-0801976
Converter SCC-K, cock PVDF, 115 VAC
23093-4-0801977
Converter SCC-K, 2 solen. valves, 115 VAC 23093-4-0801978
Converter SCC-K, heated, 1 s.v., 115 VAC
23093-4-0801979
Accessories
Wall mounting bracket
23009-4-0801980
Carbon catalyst set
23009-4-0801981
28
System Components and Accessories for Sample Gas Conditioning
10/23-5.20 EN January 2004
SCC-K NO2 /NO Converter
Dimensional drawing (Dimensions in mm)
483
2
325
132,5 (3HE)
260
44
3
4
5
400
400
NO
NOx
1 Power supply input X1
Status signal output X2
(9-pin Sub-D female connector)
Heated sample gas inlet 6 mm
Sample gas outlet G1/4 inch
Sample gas inlet G1/4 inch
Ansicht von hinten
465.6
57
120
435
1
2
3
5
4
Circuit and connection diagram
Bei Steuerung nur intern
Brücke X2/1 nach X2/6
Internal controlled only
link X2/1 to X2/6
Converter
Statussignal NO
Status signal NO
X1 F1/2 3,15A
internal
external
X2/5
X2/9
X2/1
X8/4
X8/6
X8/10
X3/2
X2/3
X2/6
X8/1
X8/8
X9/3
X9/1
X9/4
A1 ASP 204M
K1
PE
X2/2
X8/9
Netzfilter
mains suppression filter
X3/3
Netz
Power
supply
Bypass
K2
K3
S1
X3/1
X4/3
X5/4
X5/3
X5/2
X5/5
X5/6
F4
Y1
Y2
+
-
E1
X6/2
X6/1
L1
83
80
142
111
143
112
E1.1
243
X6/5
242
X6/4
X4/2
230V
81
115V
X6/6
N
D1 1N4007
T1
115V
X6/3
D2 1N4007
F3
Frontplatte
front panel
+
C1
D3
470µF
35V
K1
X9/6
D4
X9/5
K2
X5/1
X7/1
0 sw./bl.
X7/2
+ rot/red
M1
10/23-5.20 EN January 2004
System Components and Accessories for Sample Gas Conditioning
29
SCC-K NO2 /NO Converter
Pneumatic layouts
Standard version with 4-way ball valve
1 Power supply
2 Status signal
4 Sample gas outlet
5 Sample gas inlet
E1 Tube furnace
Y1 4-way ball valve
4
360
2
E1
1
Y1
5
Option with heated sample gas inlet
1 Power supply
2 Status signal
3 Heated sample gas inlet
4 Sample gas outlet
E1 Tube furnace
Y1 3/2-way solenoid valve
3
360
2
E1
1
Y1
4
4
360
2
Option with two solenoid valves
1 Power supply
2 Status signal
4 Sample gas outlet
5 Sample gas inlet
E1 Tube furnace
Y1 3/2-way solenoid valve
Y2 3/2-way solenoid valve
E1
Y2
1
Y1
5
30
System Components and Accessories for Sample Gas Conditioning
10/23-5.20 EN January 2004
CGO-9 Tubular Furnace
Application and description
The tubular furnace is used to convert a non-measurable gas
component into a measurable compound which can be measured by a gas analyzer. This is for example necessary with
certain carbon, sulfur and halogen (such as chlorine) compounds. The conversion is performed in a reaction tube which
is filled with a catalyst and heated by the tubular furnace.
A further application of the tubular furnace is to remove interfering components without the other gas components being
affected.
Technical data
Sample gas pressure
pabs ≤ 120 kPa (1.2 bar)
Operating data
Dead time
in s at
30 l/h 60 l/h
Without filling 1.5
0.5
With filling
2
0.7
90% time
Pressure drop
in s at
in hPa/mbar at
30 l/h 60 l/h 30 l/h 60 l/h
2
0.75
1.5
3.6
4.5
2.5
6
12
Dimensional drawing (dimensions in mm)
Reaction tube
Material quartz, volume 20 ml, filling volume 8 ml
Thermocouple
Chromel-Alumel (corresponds to NiCr-Ni)
Furnace temperature
max. 800 °C
The intermediate temperatures are set using an external
on/off controller.
Heat-up time
approx. 120 min.
Power supply
AC voltage 220 V +10...–12 %, 48...62 Hz
Power consumption 150 W
Case material
Steel sheet
Degree of protection of housing
IP 50 to EN 60529
A protective cap (accessory) is required if the tubular furnace
is not protected against touching by design measures such
as installation in a cabinet or a case.
Weight
approx. 1.0 kg
4 Terminal box
5 Thermocouple
6 Protective cap
Ordering information
Mounting
Wall fastening
Catalog No.
Sample gas connection
Glass tube 5 mm outside diameter for hose with 4 mm
inside diameter
Electrical connections
Terminals via cable gland PG 9
Ambient temperature
Operation: +5...+50 °C,
Storage and transport: –25...+65 °C
10/23-5.20 EN January 2004
1 Gas inlet
2 Gas outlet (offset by 60°)
3 Mounting bracket or strap
CGO-9 with reaction tube,
with thermocouple
23092-4-0806560
CGO-9 with reaction tube,
with catalyst LF 316, with thermocouple
23092-4-0856545
Accessories
CGO-9 reaction tube, quartz
23005-4-0806559
CGO-9 reaction tube, stainless steel with
catalyst LF 316
23004-4-0856544
CGO-9 stainless steel catalyst LF 316
23004-4-0801715
System Components and Accessories for Sample Gas Conditioning
31
CGO-9 Tubular Furnace
Examples of applications
Application
Furnace
Temperature Reagent
temperature controller
Sample gas Mounting
throughput
Oxidation of sulfur compounds
Combustion of organic chlorine compounds
Conversion of H2 to HCl using Cl2, where
Cl2 concentration ≥ H2 concentration
(O2 concentration > 4 Vol.-%, otherwise use
reaction element, see page 33)
NO2/NO conversion in the exhaust air of
nitric acid or pickling plants
Combustion of NH3 in air
Combustion of organic nitrogen compounds,
such as emissions from animal keeping or
carcass incineration
Application range: 100...3000 ppm NO2
800 °C
800 °C
800 °C
no
yes
yes
none
none
none
60 l/h
60 l/h
60 l/h
750 °C
yes
Stainless steel catalyst LF 316 60 l/h
Service life: change catalyst
every 6 months as a preventative measure.
The stainless steel catalyst is
not suitable for SO2containing combustion gases
32
System Components and Accessories for Sample Gas Conditioning
arbitrary
arbitrary
arbitrary
vertical
10/23-5.20 EN January 2004
Reaction Element
Application and description
The reaction element with UV lamp is used for gases which
react chemically under the influence of UV light. A typical
application is the conversion of hydrogen (H2) with chlorine (Cl2)
into hydrochloric acid (HCl). For chlorine concentrations of less
than 85 Vol.-%, however, the tubular furnace CGO-9 should be
used in principle.
The sample gas is irradiated in a quartz reaction vessel by a
UV lamp. The lamp is mounted in the centre of the reaction
vessel so that the UV light radiates to all sides. This device is
housed in a plastic case. The formation of ozone outside the
reaction vessel is prevented by purging the case with nitrogen
or instrument air.
Technical data
Sample gas pressure
pabs = 80...150 kPa (0.8...1.5 bar)
Dead volume of the reaction tube
approx. 250 cm3
Service life of the UV lamp
approx. 5 months
Dimensional drawing (dimensions in mm)
Power supply
230 VAC, 50 Hz
Power consumption 0.5 A during start-up, 0.35 A in
continuous operation
6
Case material
Glass-fiber-reinforced polyester
Weight
2 kg
1
Degree of protection of housing
IP54 to EN 60529
2
3
Materials of the gas-conducting parts
Viton, quartz
4
Electrical connection
Terminals via PG 11 cable gland
5
Gas connection
Glass tube with 7 mm outer ∅ for tube with 6 mm inner ∅
and 9 mm outer ∅
Case purging
approx. 10 l/h nitrogen or instrument air
1
2
3
Purge gas outlet
Sample gas outlet
Power supply
4
5
6
Sample gas inlet
Purge gas inlet
Cable gland
Ordering information
Catalog No.
Reaction element with UV lamp
10/23-5.20 EN January 2004
23095-5-5854124
System Components and Accessories for Sample Gas Conditioning
33
Membrane Filter
Application
Dimensional drawings (dimensions in mm)
Membrane filters are used in gas conditioning systems of
analyzer equipment for fine filtering of dust particles ≥ 1 µm, in
order to protect the analysis instruments from contamination in
the event of disturbances in the sample gas conditioning chain.
Membrane filter, PVC version
Technical data
Flow rate (air)
max. 300 l/h
Gas pressure pabs
PVC: max. 150 kPa (1.5 bar),
stainless steel: max. 200 kPa (2 bar)
Gas temperature
PVC: max. 55 °C, stainless steel: max. 100 °C
Retention rate
100 % for particles ≥ 1 µm
Pressure drop
2.5 to 20 hPa (mbar) at a flow rate of 30...250 l/h (air)
Materials of the gas-conducting parts
Case: PVC or stainless steel (Mat. No. 1.4571),
gas connections: Hostaflon, round cord rings: Viton,
filter membrane: glass fiber
Filter surface
approx. 24 cm2
Dead volume
approx. 15 cm3
Membrane filter, stainless steel version
Weight
PVC: approx. 0.59 kg, stainless steel: approx. 1.4 kg
Mounting
Panel and wall mounting (with mounting bracket)
Gas connections
G 1/4 (DIN lSO 228/1) or nozzles for hose internal diameter
4 mm
Scope of delivery
Membrane filter, mounting bracket, 2 nozzles, 2 round cord
rings, 50 filter membranes
Ordering information
Catalog No.
Membrane filter
Membrane filter, PVC
23145-5-8018438
Membrane filter, stainless steel
23147-5-8018439
Condensate monitor (see page 35)
Condensate monitor, PVC
23158-5-8018440
Condensate monitor, stainless steel
23157-5-8018039
34
System Components and Accessories for Sample Gas Conditioning
10/23-5.20 EN January 2004
CGKW Condensate Monitor
Application
The condensate monitor CGKW is used to indicate moisture
(condensate) or any other electrically-conducting liquid which
penetrates into the sample gas line to the gas analyzer. At the
same time, the condensate monitor is a membrane filter and it
therefore satisfies the double function of fine filtering the sample gas as well as of signaling moisture. The filter membrane
holds back the penetrated concentrate for a short while. An
independent condensate barrier (see application examples)
can be constructed with the condensate monitor as sensor
element and the switching unit ER-144A/Ex (see page 36).
switching unit ER-144A/Ex, the apparatus is permitted in
hazardous areas of Zones 1 and 2 up to a gas and ambient
temperature of 55 °C without restriction of the explosion groups
and temperature classes.
Application examples
The process gas is present at zero pressure or with
negative pressure
Sample gas feeding by pump. Condensate barrier by switching
off the pump. Condensate monitor and analyzer must be
arranged above the sample gas conditioning chain.
Technical data
Flow rate (air)
max. 300 l/h
Gas pressure pabs
PVC: max. 150 kPa (1.5 bar),
stainless steel: max. 200 kPa (2 bar)
Gas temperature
max. 55 °C
Retention rate
100 % for particles ≥ 1 µm
Pressure drop
2.5 to 20 hPa (mbar) at a flow rate of 30...250 l/h (air)
Materials of gas-conducting parts
Case: PVC or stainless steel (Mat. No. 1.4571), gas connections: Hostaflon, electrodes: stainless steel (Mat. No. 1.4305),
round cord rings: Viton, filter membrane: glass fiber
The process gas is present at zero pressure or with weak
positive pressure
Sample gas feeding by pump. Condensate barrier by switching
off the pump and switching over a 3/2-way solenoid valve.
Condensate monitor and analyzer can be arranged as you wish.
Filter surface
approx. 24 cm2
Dead volume
approx. 15 cm3
Dimensions
as for membrane filter (see page 34)
Weight
PVC: approx. 0.71 kg, stainless steel: approx. 1.5 kg
Mounting
Panel and wall mounting (with mounting bracket)
Gas connections
G 1/4 (DIN lSO 228/1) or nozzles for hose inside diameter
4 mm
The process gas is present with positive pressure
Sample gas feeding by positive pressure. Condensate barrier
by switching over a 3/2-way solenoid valve. Condensate
monitor and analyzer can be arranged as you wish.
Electrical connections
Two-core cable, approx. 1 m, for connecting to the switching
unit ER-144A/Ex (see page 36)
Mounting orientation
Safety nose at top, electrode below
Scope of delivery
Condensate monitor, mounting bracket, 2 nozzles, 2 round
cord rings, 25 filter membranes, operating manual
Explosion protection
(only with stainless steel case)
The condensate monitor is a device without voltage source
(VDE 0165/2.91 Section 6.1.3.1.3). If the condensate monitor is
being operated in the intrinsically-safe control circuit of the
10/23-5.20 EN January 2004
1
2
3
1)
Membrane pump
4 Switching unit
7 Analyzer
Condensate monitor
5 Pilot lamp
6 Flow meter
Solenoid valve 1)
The sample gas path to the analyzer is blocked in the deenergized condition
System Components and Accessories for Sample Gas Conditioning
35
ER-144A/Ex Switching Unit
Application and description
Dimensional drawing (dimensions in mm)
The switching unit ER-144A/Ex acts as switching amplifier for
the condensate monitor CGKW (see page 35). The intrinsicallysafe circuit of the switching unit ER-144A/Ex enables the
condensate monitor also to be installed in hazardous areas.
The ER-144A/Ex switching unit operates in closed-circuit
operation. The alarm positions of the output contacts correspond to those of the device in a non-voltage state or exceeding the adjustable measurement current value (electric conductance) or disconnection of the sensor lead. In the “Ready”
status, the output contacts have changed over to the corresponding counter-position (internal relay picked-up).
144/A
99
A 100-kΩ resistor (1 % / 0.25 W) must be installed at the sensor
in order to monitor the signal line between the sensor and the
electrode relay. Without the resistor, the switching unit will
change to the “Line break” position.
Technical data
Electrical connections
Indication
One LED each for line breakage and condensate penetration
NO2
12 11 14 22 21 24
COM2
21 22 24 12
NO1
Output
2 potential-free changeover contacts, max. 250 V AC, 5 A
14 11 N L1
NC2
“ib”
≤ 13.1 V
≤ 5 mA
≤ 65 mW
≤ 6 µF
≤1H
NC1
“ia”
≤ 13.1 V
≤ 5 mA
≤ 65 mW
≤ 0.97 µF
≤ 0.9 H
COM1
Input
No-load voltage U0
Short-circuit current I0
Power P0
Permissible external capacitance C0
Permissible external inductance L0
Power supply
230 V AC or 115 V AC ±10 %, 48...62 Hz, ≤ 1 VA
Construction
Surface-mounting case
Degree of protection
Housing IP 40, terminals IP 20 to EN 60529
Electrical safety
Protection class II,
Over-voltage category III,
Degree of pollution 2
E1 E2 E1 E2
Weight
approx. 0.2 kg
Mounting
Snap fastening on standard rail 35 mm (EN 50022)
Relays are shown in the unpowered state.
Ambient temperature
–25...+60 °C
Ordering information
EC type examination certificate
TÜV 00 ATEX 1604
Designation
II (1) G [EEx ia] IIC
Catalog No.
Switching unit ER-144A/Ex, 230 V AC
23371-4-0730637
Switching unit ER-144A/Ex, 115 V AC
23371-4-0730638
Installation location
Outside the hazardous area
36
System Components and Accessories for Sample Gas Conditioning
10/23-5.20 EN January 2004
Acid Filter
Application and description
Dimensional drawing (dimensions in mm)
The acid filter is used in the gas conditioning system. Its purpose is to collect the sulfuric acid aerosols entrapped in the
sample gas. The acid filter may always be required where the
SO3 content in the sample gas is greater than 1 mg/m3 or the
SO2 content is greater than 1000 mg/m3 – for example in flue
gas from heavy-oil or hard coal firings.
140
20
65
6,5
60
85
The filter element is made up of borosilicate fibers. An influencing and “hang-up” of the measuring component in the sample
gas is not detectable with this material.
78
36
50
Technical data
DN4/6
ca. 25
Flow rate (air)
max. 250 l/h
23.5
Gas pressure
pabs = 50...200 kPa (0.5...2 bar)
Ordering information
Sample gas temperature
max. 150 °C
Catalog No.
Acid filter
Sample gas dew point
max. 70 °C
23045-5-8018419
Retention rate
99.99 % for particles ≥ 0.1 µm
Pressure drop
10 hPa at 250 l/h, 50 hPa at 250 l/h
Dead time
approx. 20 s
Materials of gas-conducting parts
Connection cap: PVDF; vessel: glass;
filter element: borosilicate glass micro fiber
Dead volume
approx. 130 ml
Dimensions
see dimensional drawing
Weight
approx. 0.57 kg
Mounting
Wall mounting with mounting bracket, mounting orientation
vertical
Gas connections
Male fitting (PVDF) for 4/6/1 mm, drainage connection GL14
with terminal fitting 4/6/1 mm (PVDF)
Ambient temperature
+5...+50 °C
Service life
Change element if there is contamination and perceptibly
higher pressure drop
Scope of delivery
Filter vessel with filter element, two male fittings, drainage
connection GL14, mounting bracket, mounting and replacement instructions
10/23-5.20 EN January 2004
System Components and Accessories for Sample Gas Conditioning
37
Disposable Filter
Application and description
Dimensional drawing (dimensions in mm)
The disposable filter is used in the gas conditioning system.
It is mounted in the sample gas line upstream of the gas
analyzer.
The disposable filter consists of a micro fiber filter pipe which
is welded into a plastic case.
Technical data
Flow rate (air)
max. 250 l/h
Gas pressure
pabs ≤ 200 kPa (2 bar)
Sample gas temperature
+5...+50 °C
Retention rate
99.99 % for particles ≥ 0.1 µm
Ordering information
Catalog No.
Pressure drop
approx. 2 hPa at 60 l/h
Disposable filter
23044-5-8018418
Dead time
approx. 3 s at 60 l/h
Materials of gas-conducting parts
Case: nylon, filter element: borosilicate glass micro fiber
Case volume
11.5 cm3
Dimensions
Height 80 mm, diameter 28 mm
Weight
approx. 0.1 kg
Mounting
Mounting orientation preferably vertical
Gas connections
Two pipe nipples, outside diameter 6.3 mm
Ambient temperature
+5...+50 °C
Service life
Change filter if there is condensate accumulation (acid), and
in any case every 6 months
Scope of delivery
Disposable filter, mounting and replacement instructions
38
System Components and Accessories for Sample Gas Conditioning
10/23-5.20 EN January 2004
Flow Meters, Flow Monitors
!
!
!
!
Flow measurement using float-type flow meters
Replaceable measuring tube
Integral needle valve
Flow monitor with inductive contact
Flow meter 7...70 l/h
Dimensional drawing (dimensions in mm)
Measuring range
7...70 l/h nitrogen at 20 °C and 1013 hPa (mbar)
The following flow rates are obtained with other gases under
the same conditions at the 60 l/h mark (approximate values):
Sulfur dioxide
SO2 48 l/h
Carbon dioxide
CO2 51 l/h
Argon
Ar
54 l/h
Air
59 l/h
63 l/h
Oxygen
O2
Carbon monoxide CO
68 l/h
Methane
CH4 98 l/h
156 l/h
Hydrogen
H2
Helium
He 213 l/h
Measuring deviation
±5 % of the full-scale value
Gas inlet conditions
Pressure max. 16 MPa (16 bar), temperature 100 °C
Flow rate adjustment
with integral needle valve
Materials
Measuring tube
Float
Float catcher
Fitting
Needle valve
Seals
Nozzles
borosilicate glass
stainless steel Mat. No. 1.4401
PTFE
stainless steel Mat. No. 1.4571
stainless steel Mat. No. 1.4571
Viton
Hostaflon
1 Gas inlet
2 Gas outlet
Weight
approx. 0.5 kg
Mounting
Panel or wall mounting by means of mounting bracket
Gas connections
1/4 NPT female thread or sleeve for hose 4 mm ID
Scope of delivery
Flow meter, mounting bracket, 2 nozzles
10/23-5.20 EN January 2004
System Components and Accessories for Sample Gas Conditioning
39
Flow Meters, Flow Monitors
Flow monitor 1.6...16 l/h, 7...70 l/h
Dimensional drawing (dimensions in mm)
Measuring range
1.6...16 l/h or 7...70 l/h nitrogen at 20 °C and 1013 hPa (mbar)
Measuring deviation
±5 % of the full-scale value
Flow rate adjustment
with integral needle valve
Materials
as with flow meter 7...70 l/h
Weight
approx. 0.8 kg
Mounting
as with flow meter 7...70 l/h
Gas connections
as with flow meter 7...70 l/h
Signaling
with inductive contact RC 10-14-NO
Float catcher set to 5 l/h or 20 l/h
Electrical connections
Terminal boxes with EMC filter and cable gland PG 11
Scope of delivery
Flow monitor, mounting bracket, 2 nozzles, O-ring
Flow monitor for chlorine 7...70 l/h
Measuring range
7...70 l/h chlorine at 20 °C and 1013 hPa (mbar)
1 Gas inlet
2 Gas outlet
Ordering information
Catalog No.
Flow meter 7...70 l/h
23151-5-8018474
Measuring deviation
±5 % of the full-scale value
Flow monitor 1.6...16 l/h
23155-5-8018476
Flow monitor 7...70 l/h
23155-5-8018475
Gas inlet conditions
Pressure max. 10 MPa (10 bar), temperature 20 °C
Flow monitor for chlorine 7...70 l/h
23155-5-8018477
Flow rate adjustment
with integral needle valve PEEK
Materials
Measuring tube
Float
Float catcher
Case
Seals
borosilicate glass
PEEK with lining
PTFE
PVDF
FFKM
Weight
approx. 0.5 kg
Mounting
Panel or wall mounting by means of mounting bracket
Gas connections
1/4 NPT female thread
Signaling
with inductive contact RC 10-14-NO
Float catcher set to 20 l/h
Electrical connections
Terminal boxes with EMC filter and cable gland PG 11
Scope of delivery
Flow monitor, mounting bracket, 2 nozzles, O-ring
40
System Components and Accessories for Sample Gas Conditioning
10/23-5.20 EN January 2004
KFA Isolated Switch Amplifier
Application and description
The KFA isolated switch amplifier is used as switch amplifier
for the flow monitor (see page 39).
The control circuit is monitored for line breakage and short
circuit. External faults are indicated by a red flashing LED.
Technical data
Input (intrinsically safe)
U0 ≤ 10.6 V
I0 ≤ 19.1 mA
P0 ≤ 51 mW
Permissible connection values
Explosion group
IIA
External capacitance
72 µF
External inductance
780 mH
IIB
16.2 µF
390 mH
IIC
2.32 µF
97 mH
Output relays
Potential-free changeover contacts, max. 253 V AC, 2 A;
cos ϕ > 0.7; switching frequency < 10 Hz
Power supply
207...253 V AC or 103.5...126 V AC, 45...65 Hz, ≤ 1 VA
Electrical connections
Degree of protection of housing
IP 30 to EN 60529
Input I EEx ia IIC
Input II EEx ia IIC
1+
4+
Switch S3 in position I
3-
6-
400 Ω ≤ R ≤ 2 kΩ
Weight
approx. 150 g
1+
Mounting
Panel or wall mounting with screw fastening (EN 50022/3) or
snap fastening on standard rail 35 mm (EN 50022)
10 kΩ
400 Ω ≤ R ≤ 2 kΩ
3-
Switch S3 in position I
10 kΩ
4+
6-
Switch S3 in position I
without SC
10 kΩ
2+
without
LB, SC 1+
1+
10 kΩ
3-
2+
3-
4+
3-
4+
5+
6-
5+
6-
Hazardous area,
zone 0, zone 1
Construction
Surface-mounting case
Switch S3 in position II
6-
Ambient temperature
–20...+60 °C
I
II
S1
Designation
II (1) G D [EEx ia] llC
Installation location
Outside the hazardous area
LB
SC
S3
red
I
II
S2
&
S3
red
+ -
&
1
Safe area
LB
SC
EC type examination certificate
PTB 00 ATEX 2081
green
yellow
yellow
+
7 8 9
Output I
10 11 12
Output II
14 15
Power
supply
-
Power LB/SC
Rail collective
error message
only
KFD2-SR2-Ex2.W
Ordering information
Catalog No.
10/23-5.20 EN January 2004
KFA6-SR2-Ex1.W, 1 channel, 230 V AC
23372-5-8328644
KFA6-SR2-Ex2.W, 2 channels, 230 V AC
23372-5-8328645
KFA5-SR2-Ex1.W, 1 channel, 115 V AC
23372-5-8328851
KFA5-SR2-Ex2.W, 2 channels, 115 V AC
23372-5-0730651
System Components and Accessories for Sample Gas Conditioning
41
Stripper for Cooling Water and Waste Water
Application and description
The stripper enables traces of organic solvents or pollution in
waste water or cooling water to be monitored continuously.
The stripper is used in combination with a flame-ionization
detector (FID). The device operates on the flow principle.
100...300 l/h of the water to be analyzed are sent through a
glass body in the center of which is located a glass frit which is
continuously purged with the instrument air. Hydrocarbons
change into the gas phase and are driven out of the medium by
the air bubbles. The air stream is then analyzed with an FID.
The stripper unit is operated on the bypass principle without a
pump. The cross-sections of the entire unit are designed in
such a way that clogging is prevented to a greater or lesser
extent. A threefold safeguard against flooding and water breakthrough to the FID is achieved by an overflow and an
emergency overflow as well as a check valve. The feed can be
shut off by means of a gate valve. To obtain long-term
measurement stability, the strip air is fed via flow regulators.
Technical data
Instrument air
100...700 kPa (1...7 bar), consumption approx. 100 l/h
Minimum pressure difference
10 kPa (0.1 bar)
Medium temperature
+10...+50 °C
Connections
Connection, test water:
1 inch thread connection
Connection, sample water:
DN 25 using DIN 4-hole flange with flat seal
Discharge, sample water:
Discharge point DN 50 via DIN 4-hole flange with flat seal
Contacts
Potential-free contact for flow monitoring of the sample
water
Flow rate sample water
100...300 l/h
Heat-up time
< 7 s with 3 m feed line between stripper and FID
Materials of gas-conducting parts
Rigid PVC, glass
Power supply
230 VAC, 50/60 Hz or 120 VAC, 60 Hz
Ambient temperature
+10...+50 °C
Ordering information
Catalog No.
Stripper IP 65, unheated
23841-5-4560001
Stripper IP 65, heated, 230 VAC
23841-5-0768362
Stripper IP 65, heated, 120 VAC
23841-5-0768361
System Components and Accessories for Sample Gas Conditioning
10/23-5.20 EN January 2004
Case material
G.R.P.
Degree of protection of housing
IP 65 to EN 60529
Dimensions
1000 x 460 x 340 mm (H x W x D)
Weight
24 kg
42
Multi-way Test Gas Cocks
Multi-way test gas cock, PVDF version
Flow rate
approx. 1000 l/h at 100 hPa pressure difference and 20 °C
Gas pressure
pabs = 50...200 kPa (0.5...2 bar)
Leakage rate
≤ 5 x 10–3 hPa I/s
Materials of gas-conducting parts
PVDF, PTFE, Viton
Weight
approx. 0.7 kg
Mounting
Wall or panel mounting
Gas connections
G1/4 (DIN ISO 228/1)
Ambient temperature
+5...+80 °C
Multi-way test gas cock, stainless steel version
Dimensional drawing (dimensions in mm)
Leakage rate
≤ 5 x 10–3 hPa l/s
Materials of gas-conducting parts
PTFE, stainless steel SS 316, suitable for pure oxygen
94
62,5
~110
Gas pressure
pabs = 17.6 MPa (176 bar)
~124
Flow rate
approx. 1000 l/h at 50 hPa pressure difference and 20 °C
75
82
Weight
approx. 0.8 or 0.45 kg
Mounting
Wall mounting (with accessories) or panel mounting
Nullgas
Zero gas
Messgas
Meas.gas
Gas connections
1/8-NPT and screw fitting for 6 mm OD pipes
Ambient temperature
+10...+65 °C
Endgas 1
Span gas
83
Ordering information
Catalog No.
Test gas cock PVDF
23177-4-0839859
Test gas cock stainless steel
90805-4-0890875
Montagebohrung
Bore holes
45°
63
M6
Test gas cock stainless steel for wall mount. 90805-5-8308581
10/23-5.20 EN January 2004
System Components and Accessories for Sample Gas Conditioning
43
Cylinder Pressure Reducers
Application and description
Cylinder pressure reducers are used to reduce the filling
pressure in pure gas and test gas cylinders to the required
operating pressure. Material compatibility must be observed in
respect of the test gas (see technical data). Two-stage cylinder
pressure reducers are employed where very constant pressure
and flow are required at low operating pressure – for example
where the subsequent gas paths are switched over using
solenoid valves. Pressure reducers with minimum contact at
the high-pressure gauge give an indication when the cylinder
pressure has dropped to such a level that it is necessary to
replace the gas cylinder.
Ordering information
Catalog No.
2-stage for nitrogen
23422-5-8018376
2-stage for non-corrosive test gases
23422-5-8018377
2-stage for corrosive test gases
23422-5-8018378
1-stage for nitrogen, brass
23422-4-0730654
1-stage for non-corrosive test gases, brass 23422-4-0730655
2-stage for hydrogen, brass
23422-4-0730652
2-stage for hydrogen, stainless steel
23422-4-0730653
Technical data
Catalog No.
Gas type
Construction
Inlet pressure
Outlet pressure (adjustable)
Safety valve set to
Relief valve set to
Pressure gauge indication
Inlet pressure
Outlet pressure
Flow rate
Control action
Status signal
Materials
Membranes
Seal
Case
Dimensions (W x H x D)
Weight
Connections
Connection to cylinder
(thread to DIN 477)
Gas outlet
Safety valve
Relief valve
Ambient temperature
Scope of delivery
Dimensional drawing
44
23422-5-8018376, 23422-5-8018377
Nitrogen, non-corrosive test gases
2-stage, brass, with minimum contact
max. 21 MPa (210 bar)
10...50 kPa (0.1...0.5 bar)
approx. 1.6 MPa (16 bar)
approx. 70 kPa (approx. 0.7 bar)
23422-5-8018378
Corrosive test gases
2-stage, stainless steel, with min. contact
max. 21 MPa (210 bar)
10...50 kPa (0.1...0.5 bar)
approx. 1.6 MPa (16 bar)
approx. 70 kPa (approx. 0.7 bar)
0...25 MPa (0...250 bar)
–100...+150 kPa (–1...+1.5 bar)
25...500 l/h
Back pressure increase with falling cylinder pressure down to 3 bar equals zero
Reed contact 100 V, 0.5 A
0...25 MPa (0...250 bar)
–100...+150 kPa (–1...+1.5 bar)
25...500 l/h
Back pressure increase with falling cylinder pressure down to 3 bar equals zero
Reed contact 100 V, 0.5 A
Stainless steel
ETFE, PVDF, EPDM
Brass chromium plated
approx. 160 x 185 x 167 mm
approx. 2.1 kg
Stainless steel
ETFE, PVDF, EPDM
Stainless steel 1.4401
approx. 160 x 185 x 167 mm
approx. 2.1 kg
Hexagon
Hexagon
Clamping ring screw fitting Swagelok®
for pipe with 6 mm outside diameter
Clamping ring screw fitting Swagelok®
for pipe with 6 mm outside diameter
1/4-NPT male thread
–5...+75 °C
Pressure reducer, pressure gauge, safety
valve and clamping ring screw fitting
attached; replacement seals, manual
see above
Clamping ring screw fitting Swagelok®
for pipe with 6 mm outside diameter
Clamping ring screw fitting Swagelok®
for pipe with 6 mm outside diameter
1/4-NPT male thread
–5...+75 °C
Pressure reducer, pressure gauge, safety
valve and clamping ring screw fitting
attached; replacement seals, manual
System Components and Accessories for Sample Gas Conditioning
10/23-5.20 EN January 2004
Cylinder Pressure Reducers
Technical data
Catalog No.
Gas type
Construction
Inlet pressure
Control range
Control action
Pressure gauge
Reed contact
pressure gauge
Contact
Outlet pressure gauge
indication range
Materials
Body
Membrane
Case seal
Seat
Filter
Connection cable
Bottle connection
Safety relief valve
Ambient temperature
Weight
Special features
23422-4-0730652
Hydrogen
2-stage, brass
21 MPa (210 bar)
0...350 kPa (3.5 bar)
Back pressure increase
with falling cylinder
pressure down to 3 bar
equals zero
–
25 MPa (250 bar)
23422-4-0730653
Hydrogen
2-stage, stainless steel
21 MPa (210 bar)
0...350 kPa (3.5 bar)
Back pressure increase
with falling cylinder
pressure down to 3 bar
equals zero
–
25 MPa (250 bar)
23422-4-0730654
Nitrogen
1-stage, brass
21 MPa (210 bar)
0...250 kPa (2.5 bar)
–
23422-4-0730655
Non-corrosive test gas
1-stage, brass
21 MPa (210 bar)
0...250 kPa (2.5 bar)
–
31.5 MPa (315 bar)
–
31.5 MPa (315 bar)
–
falling opening
–100...+500 kPa
(–1...+5 bar)
falling opening
–100...+500 kPa
(–1...+5 bar)
–
0...+400 kPa
(0...+4 bar)
–
0...+400 kPa
(0...+4 bar)
Brass
Stainless steel
Perbunan
PVDF/PTFE
50 µm
3m
DIN 477, No. 1
2
–5...+75 °C
2.1 kg
Outlet with throttle
approx. 40 l/h and
6 mm screwed fitting
Stainless steel 1.4404
Stainless steel
EPDM
PVDF/PTFE
50 µm
3m
DIN 477, No. 1
2
–5...+75 °C
2.1 kg
Outlet with throttle
approx. 40 l/h and
6 mm screwed fitting
Brass
Perbunan
Perbunan
–
100 µm
–
DIN 477, No. 10
1
–5...+75 °C
2.1 kg
Brass
Perbunan
Perbunan
–
100 µm
–
DIN 477, No. 14
1
–5...+75 °C
2.1 kg
Dimensional drawings
10/23-5.20 EN January 2004
System Components and Accessories for Sample Gas Conditioning
45
The IndustrialIT word mark and all mentioned
product names in the form XXXXXIT are
registered or pending trademarks of ABB.
ABB has Sales & Customer Support expertise
in over 100 countries worldwide.
The Company’s policy is one of continuous product
improvement and the right is reserved to modify
the information contained herein without notice.
Printed in the Fed. Rep. of Germany (01.04)