[Tools & Techniques] www.biosciencetechnology.com 8 A Rugged Capillary HPLC System for High Throughput, Sample-Limited Chromatographic Analysis System maximizes detection, supports high sample throughput. by Rainer Schuster and Cornelia Vad I In order to accelerate the discovery of new drug leads, techniques such as combinatorial synthesis and high throughput screening are increasingly being used to synthesize and evaluate limited amounts of large numbers of analytes (combinatorial libraries). The need to isolate and characterize these materials prior to screening is creating a demand for liquid chromatography systems appropriately scaled for sample-limited analyses. This scaling requirement is also evident in many other life science and biopharmaceutical applications, among them peptide and protein purification. A significant reduction in sample size can be problematic for analyses that require chromatographic separation using conventionally sized packed HPLC columns. Under these conditions, the volume of mobile phase needed to maintain satisfactory flow rates can dilute the already low level of sample analytes to a concentration below the limit of detection. A logical way to overcome this limitation is to perform the chromatography using much smaller volumes of mobile phase. To do this, the chromatographic column must be downsized to capillary dimensions (50 µm <1.0 mm ID) in the context of a high performance chromatographic system configured to operate under the constraints and conditions encountered at this operational scale. (Experiments have demonstrated that detection sensitivity actually increases as column ID decreases.) (Figure 2.) In addition to proper instrument scaling, an appropriate capillary chromatographic system must maximize detection sensitivity, provide precise, reproducible performance, and be sufficiently rugged to support requirements for high sample throughput with minimal downtime. The Agilent 1100 Series Capillary HPLC system meets these standards (Figure 1). The modular configuration and other user-friendly design features incorporated into this instrumentation make for easy set up, validation, routine operation, and maintenance. Figure 1. The Agilent 1100 Series Capillary HPLC system with solvent reservoir, capillary pump, micro vacuum degasser, thermostatted column compartment, microwell plate sampler, diode array detector, and 500 nl flow cell. A handheld controller simplifies setup and provides full control of all system modules and online display of operations. The graphic user interface with colorcoded system display shows the progress of an analysis. Agilent’s ChemStation chromatographic software performs all data management from acquisition to report generation. Screening libraries requires appropriately scaled Active flow rate control Until now, capillary HPLC systems have attempted to control flow rate by means of a passive splitter set to a desired split ratio. This approach has proven to be unreliable in maintaining a constant flow rate independent of changes in back pressure—changes which can be caused by variations in pump performance, change in a solvent gradient, or the buildup of obstructions anywhere in the system during an analysis. Since precise control of flow rate is essential for quality chromatographic separations and for properly characterizing and collecting analytes based on predictable chromatographic retention times, a method based on active flow control rather than a passive splitter is called for. The Agilent 1100 Series Capillary HPLC system employs a highpressure capillary pump with electronic flow control (EFC) to actively maintain constant flow. A special sensor monitors flow resistance in the column flow path and signals any changes to an electromagnetic proportioning valve (EMPV) which compensates by appropriate alterations in the split ratio (Figure 3). EFC has proven to be a highly robust method of flow control as demonstrated by retention time precision across replicate runs (Figure 4). The pump design also provides flexibility in solvent selection and mixing with independent sets of valves for each operation in each LC systems. Figure 2. The Relationship Between Decreasing Column I.D. and Sensitivity. The results demonstrate that detection sensitivity increases as the I.D. of the column decreases. Since column permeability also decreases with decreasing column I.D., less pressure is required to pump solvent at a specific flow rate. This means that in comparison with normal-bore columns, capillary columns allow a fuller pressure range of the pump to be utilized for solvent delivery. Stationary phase: ZORBAX® SB-C18; length: 150 mm; solvent: water/acetonitrile, 40/60; flow rate: see diagram; sample: isocratic checkout sample; injection volume: 0.1 µl; third peak: biphenyl, 200 ng; temperature: 25 C; detection wavelength: 230 nm chamber, enabling precise gradient delivery from 0-100 %. An inline micro vacuum degasser employs new gas exchange membrane technology that provides a level of degassing efficiency comparable to larger-volume solvent degassers. The degasser’s low internal volume (1 ml) facilitates a 10-fold increase in rate of system purging compared to a conventional degasser and fast solvent changeover while minimizing solvent usage. Automated sample introduction and handling Samples can be introduced into the system by means of a micro autosampler or a micro well-plate sampler, both of which enable sampling over a 3-order of magnitude range from 30 nl to 40 µl. Each sampler can be equipped with Peltier temperature control (4 to Bioscience TECHNOLOGY 7 • 2002 10 [Tools & Techniques] www.biosciencetechnology.com Figure 3. Schematic Representation of the Agilent 1100 Series Capillary Pump with Electronic Flow Control (1). (Source: Page 3, First figure: Agilent 1100 Series Capillary LC System: A one-vendor solution for highest sensitivity and robustness - Brochure, Agilent publication number: 5988-1729EN. Solvent from each reservoir passes through the micro vacuum degasser and into one of the two pump chambers and from there into a gradient mixing chamber. One set of valves controls the solvent selection and a second set of valves determines the gradient ratio. Solvent downstream from the mixing chamber is split between the solvent waste path and the column flow path, the latter of which contains a flow sensor. (In practice, two separate sensors are used for two different capillary flow rate ranges-one for 1-20 µl/min and the other for 10-100 µl/min.) Active flow control is maintained by the Electromagnetic Proportioning Valve (EMPV) which compensates for changes in back pressure signaled by the flow sensor by altering the split ratio, thereby restoring the original flow rate. The relatively high solvent flow rate upstream of the EMPV (0.2-1.0 ml/min) relative to the downstream capillary flow rate (1-100 µl/min) coupled with the very small dead volume between sensor and capillary column, minimize the delay in the response to gradient induced flow rate changes, providing highly reproducible flow rates. (The Agilent 1100 Series Capillary HPLC system is designed to bypass EFC sensor and operate with wider-bore columns at flow rates up to 2.5 ml/min with appropriate hardware changes.) 40 C ) for protection of thermally-labile samples. Up to 100 2 ml sample vials can be accommodated in this way. The micro well-plate sampler has the ability to handle different types of sample containers e.g., thermostatted 96 and 384 well plates as well as 2 ml vials, and a large sampling capacity (runs up to 768 samples unattend- ed using 384 well plates), gives users considerable flexibility in sample containment and delivery while accelerating throughput. A micro Rheodyne® valve increases the speed of sample introduction while improving take-up and delivery precision tenfold compared with a standard autosampler. Together with an optimized design of nee- dle seat, loop and seat capillaries, the new metering device also minimizes dispersion. Further increases in throughput are gained by the sampler’s ability to overlap sampling and chromatographic separation cycles. A significant reduction in dead volume further limits dispersion and the resulting band broadening that can decrease Table 1. Analysis of Depressants– Equipment and Conditions. Conditions Equipment Agilent 1100 capillary LC system components: Sample mobile phase standard mixture (A=125 pg/µl, B=10 ng/µl) A = 25 mM NaH 2 P0 4 , pH 2.5 (H2 S04) • Capillary pump • Micro vacuum degasser B = acetonitrile 15 µl/min Flow rate gradient • Thermostatted column compart 0 min 22 % B, 5 min 35 % B, 7 min 35 % • Micro autosampler B,8 min 22 % B, 11 min 2 % B • Diode array detector Injection volume 100 nl • Agilent ChemStation Column compartment temp 40 C • Column: ZORBAX SB C-18, 150 ⫻ 0.5 mm, 3.5 µm Diode array detection signal 225/10 nm, reference • 500 nl flow cell 450/80 nm, 500-nl volume Table 2. Analysis of Sulfonamides—Equipment and Conditions Equipment Agilent 1100 Capillary LC system components: Conditions Sample mobile phase A = 0.1 % formic acid in water • Capillary pump • Micro vacuum degasser • Thermostatted column compart. • Micro autosampler • Diode array detector • 500 nl flow cell • Agilent ChemStation sulfonamide mixture B = 0.1 % formic acid in acetonitrile Flow rate 13 ml/min Gradient 0 min 15 % B, 10 min 95 % B, 11 min 15 % B Injection volume 100 nl Column compartment temp 25 C Diode array detection 278/10 nm (reference 450/80 nm) • Agilent 1100 Series LC/MSD SL • Electrospray source MS conditions: • Capillary LC nebulizer Source ESI • Column: ZORBAX SB C-18, 150 3 0.5 mm, 3.5 mm Ionization mode positive Vcap 4000V Nebulizer 10 psig Drying gas flow 7 l/min Drying gas temp 150 C peak resolution, sensitivity, and reproducibility. To enhance retention time stability, the system incorporates a column oven equipped with fast Peltier cooling and heating capability (10 degrees below ambient to 80 C). A micro column switching option is also available. High performance capillaries Connections are an important component of an HPLC system and can contribute significantly to band spreading. For this reason, system designers counter the dispersive effects in connections by minimizing dead volume. Normally, in a modular system such as the Agilent 1100 Series Capillary HPLC, the relatively long lengths of connection tubing required would be viewed as having a detrimental effect on band broadening. Fortunately, the effect on overall band spreading is negligible at capillary diameters (1) and the i.d. of the connection capillaries need be reduced less than that of the column diameter. HPLC capillaries are designed to be both rugged and to facilitate uniform flow. This is accomplished with PEEK (polyetherether ketone) coated fused silica capillaries with the required internal diameter. The polymer sheath protects the capillary against handling and breakage, while the fused silica inner tube provides the precision bore and clog-resistant smooth inner wall needed for consistently reproducible flow. Capillaries are secured with standard swagelok fittings. Optimum sensitivity The need to minimize extra column band broadening is an important constraint in the design of a detection flow cell. The Agilent 1100 Series Capillary HPLC system employs a low-volume 500 nL flow with geometry optimized to minimize peak dispersion while maximizing path length for high sensitivity (Figure 5). The configuration of the flow cell and the material from which it is constructed are selected for low RI sensitivity and reduced light scattering to minimize background noise, ensuring low baselines with long-term stability and good peak separation (Figure 6). Application examples Analysis of nanoliter volumes of antidepressants by capillary LC. To demonstrate the ability of the Agilent 1100 Series Capillary LC system to handle small sample volumes accurately and reproducibly, mixtures of tricyclic antidepressants were used as exemplary analytes. Figure 7 shows the analysis of the analytes near the method detection limit. Excellent data on precision, linearity and repeatability are given for 100 nl injections on a 500 µm ID capillary column. This pseudo non-destructive analysis makes the system ideal in situations where sample volumes are limited and multiple analyses are necessary on a single sample. Table 2 lists equipment and experimental conditions. Analysis of sulfonamides in the low pg range by capillary LC using diode array and mass spectrometric detection. continued on page 16 Bioscience TECHNOLOGY 7 • 2002 [Tools & Techniques] www.biosciencetechnology.com 10 ng/µl Repeatability - 100 nl injections 0.18 1.20 0.27 1.08 mAU 20 0.20 1.17 15 0.17 1.34 0.18 1.23 10 5 0 3.5 % RSD RT % RSD RT 3.75 4 4.25 4.5 4.75 5 5.25 5.5 0.23 1.49 mAU 2 ng/µl 16 0.30 1.49 4 0.24 1.28 3 5.75 min 0.21 1.76 0.24 1.26 2 1 0 3.5 4 4.5 5 5.5 min Figure 4. Flow Rate Reproducibility as Indicated by Retention Time Precision. In the illustration shown, retention time precision is 0.21-0.30% and 0.17-0.27% at 2 ng/ml and 10 ng/ml sample loadings at a flow rate of 4 ml/min. Figure 6. Analysis of nanoliter volumes of antidepressants by capillary LC. An absolute injection of 12.5 pg gave signal-tonoise values of 3.3 - 8.7. In figure 1B an overlay of 10 runs at the 10 ng/µl level is depicted. Relative standard deviation (RSD) was <0.5 % for retention time and <2 % for area, respectively. Analyzing a 2 ng/µl sample resulted in RSD for retention time of <0.5 % and RSD for area of <3 %. The assay was linear in a range from 125 pg/µl to 4 ng/l (r 2 >0.999). Column: ZORBAX SBC18, 3.5 µm. Equipment: Agilent 1100 Capillary LC system with capillary pump, micro vacuum degasser, thermostatted column compartment, micro autosampler, diode array detector, 500 nl flow cell, Agilent ChemStation software. Figure 5. Agilent 1100 Series Capillary HPLC Flow Cell. The 0.7 ⫻ 0.7 ⫻ 10 mm dimensions combine low-volume for minimal dispersion with a long detection path length for maximum sensitivity. Spectral integrity for diode array spectrophotometric detection is maintained by the use of flat detection cell windows. The configuration and materials used to construct the cell were selected for low RI sensitivity in order to achieve flat baselines at low flow gradients. A combination of deuterium and tungsten lamps provides an extended detection range from 190 to 950 nm. When dealing with extremely small sample volumes, adding mass spectrometric capability to a capillary LC system can often enhance the sensitivity and specificity of the analysis. Figure 8 shows the analysis of seven sulfonamides with 100 pg and 10 pg on-column with diode-array detection. With a signal-to-noise ratio of 3, the 10 pg injection is at the detection limit. Figure 9A shows the MS signal (total ion chromatogram) for the same analysis. The injected amounts for MS are 10 pg and 1 pg (10 times less than with DAD). Several of the sulfonamides elute closely and extraction of the ion of interest enables better integration and calibration, especially at low levels. In addition, using extracted ion chromatograms (EICs) can enhance the visibility of smaller components in a mixture. Figure 9B illustrates the use of EICs for three of the components of the sulfonamide mixture at 1 pg on-column. Relative standard deviation for the sulfonamides was <0.1 % for retention time and <2.8 % for area respectively, based on ten runs with 100 pg and 100 nl injection volumes. The assay was linear in a range from 1 pg/ml to 1 ng/ml (r2 >0.9999) for the MSD and 10 Figure 7. Analysis of a mixture of 7 sulfonamides with diode array at 10 pg and100 pg absolute with 100 nl injection volume. pg/ml - 100 ng/ml for DAD. Table 2 lists equipment and experimental conditions. About the authors Rainer Shuster is an application chemist and Cornelia Vad is a product manager at Agilent Technologies, Waldbronn, Germany. More information is available from: Agilent Technologies, Palo Alto, CA. 800-227-9770; agilent.com. ■ Write In 116 Or Reply Online References 1. Gerard Rozing, Maria Serwe, HansGeorg Weissgerber, and Bernd Glatz, A system and columns for capillary HPLC, American Laboratory, Vol. 33, page 26 38, May 2001. 2. Maria Serwe, Analysis of nanoliter volumes of antidepressants by capillary LC, Agilent Technologies Application Note, Pub. Number 5980-2172E, July, 2000. 3. Rainer Schuster and Christine Miller, Analysis of sulfonamides in the low pg range by capillary LC using diode array and mass spectrometric detection, Agilent Technologies Application Note, Pub. Number 5980-2499EN, September, 2000. 5988-8255EN (KAISER, ASTRID, mer. 06 nov. 2002) Figure 8. Analysis of the standard mix with 10 pg and 1 pg with MSD and SIM - total ion chromatogram from SIM (A) and extracted ion chromatogram from SIM(B). Bioscience TECHNOLOGY 7 • 2002
© Copyright 2024