serum digoxinthat allows the complete automation of this analysis. Up to now, the application of the EMIT method to automated instruments, including the Roche Cobas analyzers, necessitates a preliminary manual step consisting of pre-treatment of the samples with the sodium hydroxide reagent. In the proposed protocol, every step of the analysis is taken in charge by the instrument, including the pretreatment step. The reagents are reconstituted exactly as in the original EMIT procedure, except for the buffer, which is prepared by diluting the concentrate to 160 mL instead of 225 mL; this higher concentration is necessary to counteract the dilution effect introduced by the no-default water addition used in the system for flushing the delivery needles after addition of the start reagent and the sample. There is no need to prepare, by dilution, a working B reagent, and this ensures better reagent stability. The reaction conditions were kept as close as possible to the ones of the original EMIT protocol. The instrumental conditions are as follows: Sample: 20 4, water 7 4 Incubation: 5 s Start reagent 1 (pre-treatment reagent): 8 4, water 5 Incubation: 300 s Reagent (working A): 125 4 Incubation: 900 s Start reagent 2 (B reagent): 5 4, water 38 4 Incubation:5 s, Wait time: 5 8 Readings: first 120 s, interval180 s, number 10 The method shows very good inter-run 6.6% for a 1.6 ng/mL concentration, 4 precision (CV of seven single determina- tions). When various digoxin-free sera were supplemented with known amounts of digoxin and analyzed with the new protocol and the results correlated with the true values (the amount added), the following regression equation was obtained: y = 1.07x 0.061, r = 0.99, n = 28, where the y values are the measured ones. The Cobas FARA values are significantly higher than the true values (standard error on the slope, 0.041) and this difference is of unknown origin. When patients’ samples submitted for digoxin determination were analyzed simultaneously by various methods and the results compared with those obtained with the Cobas FARA, the following lines of regression were obtained: - Reference methods manual (Syva) FPI (Abbott) Enzymun (BMC) EMIT Equation y y y = = = 1.08x 1.09x 1.09x - + + 0.104 0.179 0.102 r n 0.96 0.87 0.86 35 39 28 Major advantages of this new protocol are improved precision, reduced reagent consumption (#{188}o that in the original procedure), and simplicity of operation. Microcomputer Program for Internal Quality Control with Use of Westgard’s Multi-Rule Shewhart Chart, Mario Pandin (Laboratorio di analisi chimico-climche, Ospedale Generale Provinciale, U. S. S. L. n. 19 del “Mediobrenta,” 35013 Cittadella (PD), Italy) As reported (1), I developed a microcomputer program, written in BASIC, for internal quality control in a laboratory that is not equipped with a main computer with a built-in automatic quality-control facility. This new version of that program presents new features and more flexibility. The hardware consists of an HP-86 microcomputer with a built-in user memory of 128K bytes and with the option of a 12-in. monitor; two disc-drives HP-9130A, storing 270K bytes on one disc (5#{188} in.); a bidirectional thermal printer HP-2671A; a graphic plotter HP-7470A with two built-in pen stalls for two-color plotting (all from Hewlett-Packard Co., Personal Computer Div., Corvallis, OR 97330). In addition to the previous features, the software program, which uses two disc-drives, now has these characteristics: 1. The use of the Westgard algorithm (2) as a criterion used in deciding whether the values indicate the analytical run is in or out of control. 2. Display on the screen and printing of error messagesof the control rule that causes the rejection of the run. 3. Simultaneous error messages storage on floppy-disc, with the indication if the error is probably a systematic error or a casual error. 4. Cumulative reports of error messages, selected by periods (day, month, year) and by single or all tests. 5. Possibilityto modify any parameter used in the quality-control program. 6. A new graphic presentation: the cusum chart in addition to Shewhart plot and Youden plot already available. 7. Visualization on the Youden plot, using two keys, the normal and pathological value of the control serum and the date, for an easy inspection of the plot. 8. The program is not copyrighted, so the user may modify or add other features as needed. This program is available, preferably on floppy disc, upon request from the author or from the Editorial Office of this journal. References 1. Pandin M. Computer program for quality control, especially suited for the small laboratory. Clin Chem 1985;31:166-7. 2. Westgard JO, Barry FL, Hunt MR. A multi-rule Shewhart chart for quality control in clinical chemistry. Clin Chem 1981;27:493501. Reference Intervals for Alkaline Phosphatase Activity Determined by the IFCC and AACC Reference Methods, Norbert W. Tietz and Denise F. Shuey (University Kentucky, College of Medicine, Department of Pathology, Lexington, KY 40536-0084) of We determined the reference interval for alkaline phosphatase (EC 3.1.3.1) activity in serum for populations between 20 and 50 years old and above 60 years old. Enzyme activity was determined at 30 and 37#{176}C by the method as recommended by the International Federation of Clinical Chemistry (IFCC) (1) and the American Association for Clinical Chemistry (AACC) (2). We used two spectrophotometers (Model 25 and Acta CIII; Beckman Instruments, Inc., Fullerton, CA 92634). The individuals between 20 and 50 years of age were selected from apparently healthy hospital employees; those older than 60 years were part of a group of healthy volunteers who participated in an extensive study of reference ranges for individuals above 60 years of age (3). All participants studied had fasted for at least 10 hand had sat for 20-30 mm before the blood collection, to CLINICALCHEMISTRY, Vol. 32, No. 8, 1986 1593
© Copyright 2024