Installation & Operation Guide Firmware Version 1.01F & Later 6/2/14

Installation & Operation
Guide
Firmware Version 1.01F & Later
6/2/14
Command Alkon Inc.
5168 Blazer Parkway
Dublin, Ohio 43017
Service Dept: 614.792.0677
Fax: 614.793.0608
Part Number: 25018
© 2003-2014 Command Alkon Incorporated. All rights reserved.
Command Alkon Incorporated believes the statements contained herein are accurate as
of the date of publication of this document. HOWEVER, COMMAND ALKON
INCORPORATED HEREBY DISCLAIMS ALL WARRANTIES EITHER EXPRESSED OR IMPLIED,
INCLUDING WITHOUT LIMITATION ANY IMPLIED WARRANT OF MERCHANTABILITY OR
FITNESS FOR A PARTICULAR PURPOSE. In no event will Command Alkon Incorporated be
liable for any damages, including any lost profits, lost savings, or other incidental or
consequential damage arising out of the use of or inability to use any information
provided through this publication, even if Command Alkon Incorporated has been advised
of the possibility of such damages, or for any claim by any other party. Some states do
not allow the limitation or exclusion of liability or consequential damages, so the above
limitation may not apply.
This information is not intended to be an assertion of future action. The contents of this
document are for informational purposes only and are subject to change without notice.
Command Alkon Incorporated expressly reserves the right to change or withdraw current
products that may or may not have the same characteristics listed in this publication.
Should Command Alkon Incorporated modify its products in a way that may affect the
information contained in this publication, Command Alkon Incorporated assumes no
obligation whatever to inform any user of the modification.
This publication may contain technical inaccuracies or typographical errors. Command
Alkon Incorporated assumes no responsibility for any errors that may appear in this
manual. This publication is intended only for the direct benefit of users of Command
Alkon Incorporated products. This manual may not be used for any purposes other than
those for which it is provided. Information disclosed herein was originated by and is the
property of Command Alkon Incorporated, and, except for rights granted by written
consent, such information shall not be disclosed, disseminated, or duplicated in whole or
in part. Command Alkon Incorporated reserves all patent, proprietary, design,
manufacturing, use, and reproduction rights.
www.commandalkon.com
Command Alkon Incorporated
1800 International Park Drive, Suite 400
Birmingham, AL 35243-4232
(205) 879-3282
5168 Blazer Parkway
Dublin, OH 43017-1339
(614) 799-6650
For a list of all Command Alkon locations, go to:
http://www.commandalkon.com/locations.asp
COMMANDseries (and the names of its components, such as COMMANDconcrete and
COMMANDnetwork), Spectrum, Eagle, and COMMANDbatch are registered trademarks of
Command Alkon Incorporated. All rights reserved.
Windows is a registered trademark of Microsoft Corporation in the United States and
other countries.
Hydro-Control is a registered trademark of Hydronix Limited, United Kingdom.
Verifi is a registered trademark of RS Solutions, Cincinnati, Ohio, USA.
2
6/2/14
Contents
Contents
........................................................................................ 3
Introduction
.................................................................................. 5
Purpose ................................................................................................ 5
Audience ............................................................................................... 5
Revision Summary ................................................................................. 5
E-Z Cal Manual Station
................................................................... 8
Overview of the E-Z Cal Manual Station ..................................................... 8
Regulatory Compliance ......................................................................... 10
Offer of Source for GPL Software ............................................................ 11
E-Z Cal Major Hardware Components ...................................................... 12
Manual Station .................................................................................. 12
IO Box ............................................................................................. 15
Installation and Wiring Guidelines (E-Z Cal and J-Box) .............................. 19
Water and Moisture ............................................................................ 20
Mounting .......................................................................................... 20
Ventilation ........................................................................................ 20
Manual Station Power Source and Supply .............................................. 21
IO Box Supply Wiring ......................................................................... 23
IO Box Power Source 120VAC .............................................................. 23
IO Box Power Source 240VAC .............................................................. 23
IO Box Voltage Jumper Settings .......................................................... 23
IO Cable Connections ......................................................................... 25
120 Volt Power Connections ................................................................ 26
240 Volt Power Connections ................................................................ 27
E-Z Cal General Operation ..................................................................... 28
Powering On and Off .......................................................................... 28
Operating Plant Equipment ................................................................. 29
E-Z Cal Displays ................................................................................ 30
Clearing Bad Cell and Bad Module Messages .......................................... 31
If Communication to the Batching PC is Lost ......................................... 32
Setup and Calibration Using the E-Z Cal .................................................. 33
Operating Modes ............................................................................... 33
Setup Buttons ................................................................................... 34
Calibration Enable Switch ................................................................... 35
Setup Procedures ................................................................................. 39
Do This First ..................................................................................... 40
Setup Scale ...................................................................................... 41
Setup Counter ................................................................................... 44
Setup Moisture Probe ......................................................................... 46
6/2/14
3
Setup Generic Analog Input ................................................................ 47
Calibration Procedures .......................................................................... 49
Do This First ..................................................................................... 50
Scale Calibration on the E-Z Cal ........................................................... 51
Moisture Probe Calibration on the E-Z Cal ............................................. 54
Generic Analog Input Calibration on the E-Z Cal ..................................... 55
View Configuration Settings - Info Mode .................................................. 56
View System Settings and Select Certification Type ................................... 58
Technical Specifications ......................................................................... 61
Manual Station .................................................................................. 61
Scale Interface .................................................................................. 64
Meter Counter ................................................................................... 65
Moisture Probe Interface .................................................................... 65
IO Box ............................................................................................. 66
Input Modules ................................................................................... 68
Output Modules ................................................................................. 70
Load Cell Color Codes ........................................................................ 72
Mean Time Before Failure - Manual Station ............................................ 73
Mean Time Before Failure - IO Box ....................................................... 73
E-Z Cal Parts Replacement Guidelines ..................................................... 74
Manual Station .................................................................................. 74
IO Box Fuses .................................................................................... 75
Spare Parts List ................................................................................. 76
Index
4
........................................................................................... 77
6/2/14
Introduction
Topics in This Section
Purpose
Audience
Revision Summary
Purpose
This guide provides instructions for installing Command Alkon’s E-Z Cal
Manual Station and I/O box. It also provides procedures for configuring and
calibrating equipment (scales, counters, moisture probes, and generic
counters) that use displays on the manual station.
Audience
This guide is intended to be used by Command Alkon installation and service
personnel, as well as plant personnel authorized to configure batching
software.
Revision Summary
Date
Firmware
Version
Nov. 1, 2004
V1.01B+
Revision
Introduction section added, along with this Revision
Summary.
Removed note about a Command Alkon
representative inspecting the J-Box once a month.
Nov. 16, 2004
V1.01B+
Corrected version number on title page. Was “Version
1.02”. Now reads “Firmware Version 1.01B+”.
Apr. 5, 2005
V1.01B+
Added statement about the E-Z Cal supporting scale
capacities no larger than 60,000 lbs.
Dec. 2, 2005
V1.01B+
Added scale, meter, and moisture probe specifications
that were inadvertently left out when the manual was
reformatted.
Jul. 17, 2006
V1.01B+
Added block power wiring diagrams for 120V and
240V systems.
Aug. 8, 2006
V1.01B+
•
•
6/2/14
Corrected first sentence in Purpose section.
Reviewed proofs for print vendor.
5
Oct. 13, 2006
V1.01B+
•
•
•
•
•
Jul. 23, 2007
V1.01C+
•
Added new power supply photos.
Added corrected block power wiring diagrams for
120V and 240V systems.
Updated photos to show newer Manual Station
and J-Box.
Added regulatory conformance statements for
NTEP and CE (European Union).
Reformatted manual for use in online help.
•
Added “lb” and “ ” (blank) as UOM’s for U.S.
counters.
Added “kg” and “ ” (blank) as UOM’s for metric
counters.
Added sections on setting up and calibrating a
Generic Analog Input.
Improved Notes, Tips, Cautions, and Warnings
with better icons and formatting.
Minor formatting changes.
Fixed title page so the part number shows.
Prepared the manual for the print vendor.
•
•
•
Jul. 31, 2007
V1.01C+
•
•
Apr. 22, 2008
V1.01C+
Minor correction to section on calibrating a Generic
Analog Input.
Jul. 14, 2008
V1.01C+
Prepared the manual for the print vendor.
Nov. 3, 2008
V1.01C+
Corrected the text describing the drawing in the “IO
Cable Connections” section.
Jul. 1, 2009
V1.01C+
Corrections made to the “E-Z Cal General Operation”
section regarding powering on and off.
Sep. 3, 2009
V1.01D &
Later
Corrected some instructions in the “E-Z Cal General
Operation” section:
•
Press Setup/Calib twice quickly to bypass the
waiting period when powering on the manual
station.
• Press AGG ZERO button twice quickly to reset the
Agg scale to zero. Same for Cem scale.
Added notes stating that the BAD LCIu and BAD LCIv
messages are only displayed if the E-Z Cal firmware
version is 1.01D and higher.
6
Jan. 4, 2010
V1.01F &
Later
•
•
Added description of BAD MPIu message.
Incremented version number to V1.01F & Later.
Jun. 10, 2011
V1.01F &
Later
Updated the technical specifications for input and
output modules.
Apr. 30, 2013
V1.01F &
Later
Corrections to the moisture probe calibration
procedure.
Nov. 13, 2013
V1.01F &
Later
Added notes about the calibration enable switch when
used with the V2.0 Processor Board.
Dec. 12, 2013
V1.01F &
Later
For systems with the V2.0 Processor Board, added
information about the SYSTEM Setup option and a
step for disabling AZT.
6/2/14
Feb. 27, 2014
V1.01F &
Later
Changed NTEP certificate approval number to the
latest one.
Jun. 2, 2014
V2.01E &
Later
Created a new E-Z Cal manual for the V2.x CPU
board.
6/2/14
7
E-Z Cal Manual Station
Overview of the E-Z Cal Manual Station
The E-Z Cal is used primarily as a backup control panel for plant devices in the
event of plant malfunction or batch computer failure.
Other uses for the E-Z Cal include:
8
•
Quick and easy configuration and calibration of scales, meters, and
moisture probes.
•
Convenient place to check scale, meter, and moisture probe readings
during automatic batching without the operator having to switch screens
on the batch computer.
6/2/14
•
Simplified system troubleshooting. For example, LED lights beside IO
modules in the IO Box indicate whether power is being provided or inputs
are being activated.
Caution!
If this equipment is used in a manner not specified in
this manual or marked on the equipment, the protection
provided by the equipment may be impaired.
6/2/14
9
Regulatory Compliance
NTEP Certification
Certificate Number: 97-124A3
Issue Date: December 20, 2013
This device was evaluated under the National Type Evaluation Program
(NTEP) and was found to comply with the applicable technical requirements of
Handbook 44, “Specifications, Tolerances, and Other Technical Requirements
for Weighing and Measuring Devices.”
CE (European Union) Certification
The Spectrum VI Batching System, consisting of the E-Z Cal Manual Station
and IO Junction Box, is in effective conformance to the Directives, “Low
Voltage Directive (73/23/EEC)” and “EMC Directive (89/336/EEC)”, including
all amendments, as well as the Standards listed in the Declaration of
Conformance.
10
6/2/14
Offer of Source for GPL Software
You may have received from Command Alkon, certain products that contain—
in part—some free software (i.e. software licensed in a way that allows you to
run, copy, distribute, change, and improve the software).
As part of these products, Command Alkon may have distributed to you
hardware and/or software, or made electronic downloads, that contain GPL
(General Public License) versions of Linux Kernel, LTIB, and Freescale BSD
which are free software developed by the Free Software Foundation.
With your purchase of the E-Z Cal Manual Station, Command Alkon has
granted you a license to the above-mentioned software under the terms of the
following licenses:
General Public License (GPL)
GNU GENERAL PUBLIC LICENSE
Version 3, 29 June 2007
Copyright (C) 2007 Free Software Foundation, Inc. <http://fsf.org/>
The full text of the General Public License can be found at:
http://www.gnu.org/licenses/gpl.html
Lesser General Public License (LGPL)
GNU LESSER GENERAL PUBLIC LICENSE
Version 3, 29 June 2007
Copyright (C) 2007 Free Software Foundation, Inc. <http://fsf.org/>
The full text of the Lesser General Public License can be found at:
http://www.gnu.org/licenses/lgpl.html
Command Alkon will provide the following information to anyone who
requests it (see our contact information below):
A copy of the complete corresponding machine-readable source code
that is distributed under the GNU GPL for the product(s) listed below.
Customers who choose to modify the original software delivered by Command
Alkon for the E-Z Cal product void all WARRANTY, SERVICE and SOFTWARE
SUPPORT agreements for said product.
Product(s) Incorporating GPL Software:
EZCal Manual Station
Contact Information for Requesting Source Code:
Command Alkon Inc.
5168 Blazer Parkway
Dublin, Ohio 43017-1339
Tel: +1 (614) 799-6650
Fax: +1 (614) 793-0608
Service Dept: +1 (614) 792-0677
6/2/14
11
E-Z Cal Major Hardware Components
The E-Z Cal hardware is comprised of two major components: the Manual
Station and the IO Box.
The Manual Station is a backup control panel that allows you to manually
operate any plant device without depending on the batch control computer.
The IO Box (also called a J-Box) is an integral part of the E-Z Cal. Whereas
the Manual Station is the “operator interface”, the IO Box is the “electrical
interface”.
Manual Station
CAUTIONS!
This equipment was manufactured to operate under
normal “dry” industrial conditions. Extreme
temperature or weather conditions can and will cause
damage to the equipment and bodily harm.
Per IEC 61010, no operator access is permitted in this
unit! Service should only be performed by qualified
personnel.
A Circuit Breaker (rated at 20 Amperes 240 Volts AC)
should be installed near the IO Box, be easily
distinguishable from other controls, and labeled to
indicate that it is the Disconnect device.
Warning!
HAZARDOUS LIVE VOLTAGE PRESENT! This equipment
must be isolated or disconnected before it is accessed.
Description
The E-Z Cal Manual Station is a backup control panel that allows you to
manually operate any plant device without depending on the batch control
computer.
Each button, light, and display on the Manual Station is labeled to indicate its
function. If this is your first time using the E-Z Cal, spend a few moments to
locate the most commonly used controls such as scale displays and discharge
gate pushbuttons (see the E-Z Cal Front Panel drawing).
12
6/2/14
E-Z Cal Front Panel
6/2/14
13
Features
•
Reliabilility: With easy to replace, low-voltage buttons, which can last up
to 20 times longer than the high-voltage types used by other companies,
and easily reassignable IO points, EZ-Cal is built for the long haul.
•
Digital signals are used throughout E-Z Cal, making drifting scale values
next to impossible.
•
Up to 12 Simultaneous Displays & 80 Manual Controls can be configured.
•
E-Z Cal automatically tests for load cell failure, potentially preventing bad
loads, wasted material, and unhappy customers.
•
E-Z cal uses a standard PC connected to the IO system with one proven,
reliable RS-232 cable.
•
E-Z Cal is enterprise-upgradeable, meaning it can easily be connected to
a Dispatch system via a TCP/IP connection.
Nameplate Label
The manual station’s nameplate label is located at the top of the left side
panel as shown in the following picture. The nameplate label contains
important information such as the voltage range, frequency, current,
operating temperature, CE mark and NTEP certification number.
Nameplate Label - Manual Station
14
6/2/14
IO Box
CAUTIONS!
This equipment was manufactured to operate under
normal “dry” industrial conditions. Extreme
temperature or weather conditions can and will cause
damage to the equipment and bodily harm.
Per IEC 61010, no operator access is permitted in this
unit! Service should only be performed by qualified
personnel.
A Circuit Breaker (rated at 20 Amperes 240 Volts AC)
should be installed near the IO Box, be easily
distinguishable from other controls, and labeled to
indicate that it is the Disconnect device.
Warning!
HAZARDOUS LIVE VOLTAGE PRESENT! This equipment
must be isolated or disconnected before it is accessed.
Description
The IO Box (also called a J-Box) is an integral part of the E-Z Cal. Whereas
the Manual Station is the “operator interface”, the IO Box is the “electrical
interface”. The IO Box fuses incoming plant power with a main fuse and
sectional hot power and neutral voltages. With the provided power cable, the
voltages of the plant power can be kept in phase and on the same source as
the computer power.
All plant control signals flow to and from the batch computer through the IO
Box. These signals include scale, meter, moisture probe and limit switch
inputs to the computer, as well as the on-off solenoid valve and motor control
signals from the computer to the plant. LED lights on the Power Supply Board
indicate whether power is on for the Supply Power and the Manual Station.
(See the IO Box Layout and Power Overview.)
The IO Box contains no electrical components such as motors, high-voltage
transformers, or solenoids (refer to your plant drawings for the location of
these components).
Plant devices not directly controlled by the batch computer can be wired to be
controlled from the IO Box. Examples are air compressors and cement
aerators.
6/2/14
15
IO Box Layout and Power Overview
16
6/2/14
Features
•
Modular design.
•
Pre-punched conduit holes (1/2” and 3/4”).
•
Pre-punched holes near the Emergency Stop switch for additional manual
plant operation switches.
•
LED lights to the left of each module to indicate that power is being
provided or that an input is being activated.
•
Key-lock security of both the front door and the Plant Power switch.
•
Sectional fuses with neon (Blown) lights per switch bank.
•
Factory-installed Relay placement strip.
•
Wire raceways to keep wiring neat and presentable.
•
Clearly numbered module and plant connection points.
•
Emergency Stop switch located on the front door that quickly
disconnects all power, along with a factory pre-wired emergency stop
block for customer use.
Nameplate Label
The IO Box’s nameplate label is located at the top of the left side panel. This
label contains important information such as the voltage range, frequency,
current, operating temperature, and CE mark.
Nameplate Label - IO Box
6/2/14
17
Ground Label
The ground label (shown next) is located inside the IO Box in the lower left
corner of the back panel. This label indicates that the IO Box is earthgrounded.
Ground Label - IO Box
Caution/Warning Stickers
The following Caution and Warning stickers are attached to the front door of
the IO Box.
IO Box Caution and Warning Stickers
18
6/2/14
Installation and Wiring Guidelines (E-Z Cal and J-Box)
Topics
Water and Moisture
Mounting
Ventilation
Manual Station Power Source and Supply
IO Box Supply Wiring
IO Box Power Source 120VAC
IO Box Power Source 240VAC
IO Box Voltage Jumper Settings
IO Cable Connections
120 Volt Power Connections
240 Volt Power Connections
6/2/14
19
Water and Moisture
The E-Z Cal Manual Station and IO Box must not be installed where they will
be subjected to moisture (for example, outdoors, near a sink or in a nonweatherproof enclosure).
Mounting
Note: Use proper lifting techniques when lifting the Manual Station or
the IO Box. Always lift with your legs, keeping your back as
upright and as low as possible.
The Manual Station must be situated on a table or countertop large enough to
accommodate the four rubber feet. The surface must be of sufficient stability
and rigidity to prevent tilting or sagging.
The IO Box should be securely fastened to a rigid vertical structure such as a
wall. Alternatively, it may be mounted to some other permanent structure no
more than 45 degrees from vertical and sufficient in strength to support the
weight of the IO Box and associated wiring.
Ventilation
The Manual Station must be situated so that the following free-air clearances
can be maintained at all times:
—
Rear: 5cm
—
Top-Rear Edge: 15 cm (vertical)
The IO Box must be mounted so that the following free-air clearances can be
maintained at all times:
20
—
Left Side: 15 cm
—
Right Side, Top, and Bottom: 2 cm
—
Front (Door): 65 cm
6/2/14
Manual Station Power Source and Supply
The Manual Station, which is equipped with its own overcurrent protection, is
powered by an A.C. source from the IO Box.
See Also:
Power Cord Type and Rating
120 Volt Power Connections
240 Volt Power Connections
As shown in the following three drawings, the Manual Station power unit
consists of two power supplies:
•
The Base Board power supply
•
and a 12-Volt power supply for the AAEON 6810 and an optional Lanolinx
Ethernet switch (COMMANDbatch systems only).
Manual Station Power Supplies
6/2/14
21
Base Board Power Supply for Manual Station
12-VDC Power Supply for AAEON and Lanolinx Switch (COMMANDbatch only)
22
6/2/14
IO Box Supply Wiring
The IO Box must be connected to its power source with wiring which conforms
to the following specifications: 12 AWG, 300 volts, 105 oC. All wiring must
conform to local electrical codes. In the absence of applicable codes, the use
of grounded metal conduit is strongly encouraged.
IO Box Power Source 120VAC
The IO Box must be connected to a power source of 110 to 120 Volts AC, 20
Amps, 50/60 Hz. The power source must be equipped with a disconnect
switch and overcurrent protection rated at 240 Volts AC, 20 Amps, 50/60 Hz.
An integrated switch/circuit breaker is recommended for an overcurrent
protection device.
Caution!
A Circuit Breaker (rated at 20 Amperes 240 Volts AC)
should be installed near the IO Box, be easily
distinguishable from other controls, and labeled to
indicate that it is the Disconnect device.
IO Box Power Source 240VAC
The IO Box must be connected to a power source of 200 to 240 Volts AC, 20
Amps, 50/60 Hz. The power source must be equipped with a disconnect
switch and overcurrent protection rated at 240 Volts AC, 20 Amps, 50/60 Hz.
An integrated switch/circuit breaker is recommended for an overcurrent
protection device.
Caution!
A Circuit Breaker (rated at 20 Amperes 240 Volts AC)
should be installed near the IO Box, be easily
distinguishable from other controls, and labeled to
indicate that it is the Disconnect device.
IO Box Voltage Jumper Settings
Caution!
A Circuit Breaker (rated at 20 Amperes 240 Volts AC)
should be installed near the IO Box, be easily
distinguishable from other controls, and labeled to
indicate that it is the Disconnect device.
6/2/14
23
By making simple jumper changes on the Opto Board (see the 22821 Opto
Board Drawing), you can control which reference voltages (Hot Power,
Auxiliary, or Neutral) the modules in the IO Box will use.
22821 Opto Board Drawing
The non-detachable power cable (part number 21336) must be installed and
all peripherals must be connected to it. Also ensure that the Manual Station is
connected to the IO Box via power cable (part number 25409). See the 120
Volt Power Connections and 240 Volt Power Connections drawings.
During normal operation, the front door of the IO Box does not need to be
opened.
Each button, light, and display on the Manual Station has a corresponding LED
indicator light in the IO Box. Refer to the wiring diagrams shipped with your
system for the names and locations of these lights.
24
6/2/14
IO Cable Connections
The following drawing shows the IO Box connections for the A/B and C/D
cables (Part Number 23645 or 23644-50P).
The A/B and C/D cable ends marked “To J-Box” connect to the IO Box.
The opposite ends of the A/B and C/D cables go to their corresponding
connectors on the back of the Manual Station.
IO Box Cable Connections
6/2/14
25
120 Volt Power Connections
26
6/2/14
240 Volt Power Connections
6/2/14
27
E-Z Cal General Operation
Topics
Powering On and Off
Operating Plant Equipment
E-Z Cal Displays
Clearing Bad Cell and Bad Module Messages
If Communication to the Batching PC is Lost
Powering On and Off
Powering On
1.
If the E-Z Cal displays are not lit, press the I (On) power switch
located on the rear panel. Wait approximately five minutes for the
flashing “POWER ON” message to disappear.
Tip!
To bypass the waiting period, press the “Setup/Calib” button twice
quickly. This puts the system into Calibration mode. Press “Setup/
Calib” once more to exit Calibration mode and show the normal
displays.
Manual Station Power Switch
Manual Station
Power Switch
(behind plug connector)
28
6/2/14
2.
On the front of the IO Box door, turn the Plant Power key switch
to the “On” position (12 o’clock) and make sure the red
Emergency Stop switch is not engaged (pull it until it clicks).
Power Key Switch on IO Box Door
Powering Off
On the front of the IO Box door, turn the Plant Power key switch to the 'Off'
position (9 o'clock).
Important!
•
•
For highest reliability, scale stability, and convenience, leave EZ Cal power turned on 24 hours a day, 7 days a week, except
when servicing the equipment.
For plants that use a generator: Before shutting down the
generator, first turn off power to the IO box, then power off the
E-Z Cal. After starting up the generator, start the E-Z Cal, then
the IO Box.
Operating Plant Equipment
6/2/14
•
To operate a piece of plant equipment such as a gate or an admix valve,
press its corresponding pushbutton. Most buttons are “momentary”,
meaning you must push and hold down on the button to energize the
solenoid valve, then release the button to de-energize the valve.
•
To reset the Agg or the Cem scale value to zero, press the AGG ZERO or
CEM ZERO button twice quickly. This can be done to compensate for
material buildup on the scale gates. DO NOT do this while an
automatic batch is in progress, or batching errors may occur!
•
To reset all counters to a zero value, press the RESET COUNTERS button.
DO NOT do this while an automatic batch is in progress, or
batching errors may occur!
29
E-Z Cal Displays
Displays
•
Normally the scale and meter readouts show the amount of material that
has been batched, along with the unit of measure (kg, lb, ml, etc.). If you
are manually weighing or metering a material, watch its scale or meter
readout closely so you can release the button in time to deliver the right
amount.
•
Under certain conditions the readouts may display special messages,
which are described below.
POWER ON
Power was off, then turned on. Please wait a minute for warm-up.
BAD CELL
The load cell for that scale has failed, or the cable has been
unplugged. If you can’t quickly diagnose the problem, contact
Command Alkon Service.
--
--
(Dashes) No scale, probe, or counter has been configured for this
display.
NO CALIB
This display has been set up for a scale or moisture probe, but no
calibration has been performed.
BAD LCIu
Indicates that an update has not been received from the LCI (Load
Cell) module. Check the module.
Note: Displayed only if E-Z Cal firmware is V1.01D or higher.
BAD LCIv
Indicates that updates are being received from the LCI module,
but the raw values are not changing. Check the module.
Note: Displayed only if E-Z Cal firmware is V1.01D or higher.
BAD MPIu
Indicates that an update has not been received from the MPI
(Moisture Probe) module. Check the module.
See Also:
Clearing Bad Cell and Bad Module Messages
30
6/2/14
Clearing Bad Cell and Bad Module Messages
The E-Z Cal performs a quick check of the load cells, cables, and electronics
approximately every 10 minutes. If a problem is detected, one of the following
messages flashes on the affected display, alternating with the scale weight or
moisture reading (which might be incorrect!):
BAD CELL
BAD LCIu
BAD LCIv
BAD MPIu
Notes: •
•
See E-Z Cal Displays for a description of each message.
The BAD LCIu and BAD LCIv messages will only be displayed if
the E-Z Cal firmware version is 1.01D or higher.
The message continues to be displayed after the problem is corrected. This
allows you to be aware of a problem that comes and goes by itself.
After you have diagnosed and fixed the problem, press the “Setup” button.
The displays show something like:
Bad cell
found on
Disp 1,
Module 1
Is the
problem
fixed?
No
Press the Up/Down buttons to select Yes and press Enter.
A message will be displayed for each scale or moisture probe that has a
problem. After all of the messages have been accepted with Yes or No, press
the “Setup” button to return the system to normal operation.
6/2/14
31
If Communication to the Batching PC is Lost
Turn the ‘Manual Power’ key switch on the front of the E-Z Cal to the 3 o'clock
position. (Keyswitch is located left of the first switch bank).
Manual Station Key Switch
32
•
This allows feed buttons to function without communicating with the
batching PC.
•
Depending on local regulations and your company's policies, you may
have to turn the key to the 12 o'clock position before starting an
automatic load with the batching PC.
6/2/14
Setup and Calibration Using the E-Z Cal
Topics
Operating Modes
Setup Buttons
Calibration Enable Switch
Setup Procedures
Calibration Procedures
Operating Modes
•
In Setup mode, each E-Z Cal display can be configured to show readings
for one of the following: a scale, a meter, or a moisture probe. A display
can be configured to show scale, meter, or moisture probe information
with a variety of options such as unit of measure, decimal point position,
etc. Active displays can be disabled.
•
In Calibrate mode, you adjust the E-Z Cal displays to match the actual
weight on the scales, or the actual value of aggregate moisture.
•
In Info mode, you can view configuration settings for each display, view
system information such as firmware and software versions, and turn
scale certification (i.e. NTEP) on or off.
Note: Do not proceed beyond this point unless you intend to
reconfigure one or more scales or meters.
Caution!
If you plan to change any scale parameters, make sure you
have the proper test weights, training, and time to re-check
the weighing accuracy of those scales. Careless setup will
cause serious batching errors that the batch computer will
not be able to correct!
6/2/14
33
Setup Buttons
You will use five special buttons in the upper right-hand row of the E-Z Cal to
enter the Setup, Calibration, and Info modes.
E-Z Cal Setup Buttons
Buttons used for
Setup and Calibration
Setup/Calib
Press this button to enter one of the operating modes. Press this button again
to exit the mode you are in and return to normal display, however, any
information you were entering/changing will be discarded.
Up/Down
During Setup, you will be presented with a list of items or with a numerical
parameter that you can change. Press the Up/Down buttons to move (“scroll”)
through the list of items, or to increase or decrease the number. Hold the
button down to scroll very fast through the choices.
Enter
After selecting the desired item or setting a parameter to the right value,
press this button to accept the setting and move on to the next option.
Check Grad
This button is used to view each scale’s full scale capacity and grad size
without entering Setup. Any time during normal display operation, you can
press and hold the “Grad” button to show each scale’s full scale capacity in its
display window. When you release the button, the grad size will be shown for
one second before the display reverts back to normal.
34
6/2/14
Calibration Enable Switch
To prevent accidental changes to scale setup and calibration data, the E-Z Cal
has an internal Calibration Enable Switch that can be locked. Normally this
switch is set to the “locked” position by your system installer.
Note: You do not need to unlock the Calibration Enable Switch to change
the setup or calibration of counters or moisture probes.
To Unlock the Calibration Enable Switch
1.
(Optional, depending on government regulations for your area):
Remove the calibration wire seal from the small blue metal tab on
the rear panel (see the following picture).
Caution!
Removing this seal invalidates your local Weights and
Measures certification and may prevent you from
shipping product to some or all customers.
Location of Weights & Measures Seal
Tab where
calibration wire
is attached.
2.
6/2/14
Remove the two thumbscrews shown in the following picture and
lift off the top cover.
35
Removing the Top Cover
Top Cover Thumbscrew
Locations
3.
Swing the front panel and open until it stops as shown in the
following picture.
Swinging Open the Front Panel
4.
36
Remove the 8 screws (4 along the top and 4 along the bottom)
from the RF Shield Cover shown in the following picture. Remove
the shield cover and set it aside, along with the 8 screws.
6/2/14
Removing the RF Shield Cover
5.
Carefully remove the LCI Module Cover Plate (shown in the
following picture) without touching the delicate circuit boards
underneath. Set the cover plate aside.
Removing the LCI Module Cover Plate
6.
6/2/14
Locate the small blue and white slide switch located on the far left
side of the main circuit board (see the following picture). With a
small screwdriver or paper clip, slide the white switch downward
(toward you if you are at the front of the E-Z Cal).
37
Calibration Enable Switch
7.
Swing the front panel closed so you can get to the Setup
pushbuttons.
Notes: When the Calibration Enable Switch is in the “unlocked”
position:
• All displays that are set up for “Scales” will show a “C” as
the first character and an “E” as the last character. This
signifies that the switch is not in the locked position.
• Also when the Calibration Enable Switch is in the
“unlocked” position, and a scale is at “Center Zero” (less
than ¼ grad up or down from exact zero), an “@” symbol
is displayed before the weight.
8.
Proceed with the Setup and Calibration Using the E-Z Cal
procedures of this guide.
9.
When you are finished with setup and calibration, slide the white
calibration enable switch back up to the locked position.
10. Reinstall the LCI Module Cover Plate (no screws), then the RF
Shield Cover using the 8 screws.
11. Swing the front panel closed.
12. Reinstall the top cover and tighten the two thumbscrews.
13. Schedule a visit by a certified scale company or DOT
representative to re-seal the cover plate.
38
6/2/14
Setup Procedures
Use these procedures to configure (set up) scales, counters, moisture probes,
and generic analog inputs.
•
These devices should be configured in the E-Z Cal before you set them up
in the batch control software.
•
For Scales, you can assign the measuring system (U.S. or SI), unit, grad
size, decimal point location, and allowed grads for re-zeroing the scale.
•
For Counters, you can assign the measuring system (US or SI), unit,
grad size, and decimal point location.
•
For Moisture Probes, all you need to do is select the probe module
position (choices are 1 - 8).
•
For Generic Analog Inputs, you can assign the measuring system (U.S.
or SI), unit, grad size, decimal point location, and allowed counts for rezeroing the reading.
Topics in This Section
Do This First
Setup Scale
Setup Counter
Setup Moisture Probe
Setup Generic Analog Input
6/2/14
39
Do This First
1.
Press the “Setup/Calib” button. The displays show:
Info
Select
Menu
Option
•
If the Calibration Enable switch is unlocked, you can set up any scale,
meter, or moisture probe.
•
If this switch is locked, you can only configure counters and moisture
probes. You will not be permitted to cofigure scales.
2.
Use the Up/Down buttons to select “Setup”, press Enter, and
continue with the following steps. The displays show:
Disp. 1
Select
3.
display
to set
up
Use the Up/Down buttons to move through the displays. (For
displays 7 -12 the message moves to the top line.) Select the
appropriate display and press Enter. The displays show the name
of the setup parameter, and its value. For example, if “Disp. 2” is
selected, the displays show:
None
Select
4.
40
Scale,
Counter,
Probe or
Generic
Use the Up/Down buttons to select None, Scale, Counter, or
Moisture Probe then press Enter.
—
If you selected Scale, continue with Setup Scale.
—
If you selected Counter, continue with Setup Counter.
—
If you selected Probe, continue with Setup Moisture Probe.
—
If you selected Generic, continue with Setup Generic Analog Input.
—
If you selected None, you will be given the choice of selecting
another display or quitting the Setup procedure.
6/2/14
Setup Scale
This procedure assumes that the Calibration Enable switch is unlocked. If
this switch is locked, the display will flash “Setup Locked” for about one
second, and then will display:
Yes
Setup
another
display?
Yes / No
(Refer to the Calibration Enable Switch section for instructions on unlocking
the Calibration Enable switch.)
1.
If you selected Scale in step 4 of Do This First, the displays show:
1
Select
2.
Scale
module
position
(1 - 8)
The first time you set up a scale, the number of the first
unassigned module is displayed. Use the Up/Down buttons to
select the Scale Module Position and press Enter. The displays
show:
SI
Select
system
of units
Scale vs. Probe Modules:
Scale and Probe Modules are small printed circuit boards which are
mounted in a row across the inside rear of the E-Z Cal. Probe Modules are
marked “MPI Module” (for Moisture Probe Interface); Scale Modules are
marked “LCI Module” (for Load Cell Interface). Modules are numbered 1
through 8, beginning at the left. Make sure you know which modules are
for scales and which are for moisture probes.
3.
Use the Up/Down buttons to select the system of units. The
choices are “SI” (for metric), or “U.S.” Press Enter. The displays
show:
kg
Select
4.
6/2/14
unit
Use the Up/Down buttons to select the unit and press Enter.
Choices for U.S. units are “lb” and “oz”. Choices for SI (metric)
41
units are “g”, “kg”, and “Mg”. If you selected “kg”, the displays
show:
5 kg
Enter
5.
grad
size
Determine the proper “grad size”, which is the increment by
which the display will count. In the U.S., select 1, 2, 5, 10, or 20
to comply with Handbook-44. Use the Up/Down buttons to set the
grad size then press Enter. Grad size range is 0 to 125. If you
need a fractional grad size (e.g. .5 kg) enter it as a whole number
(5) and set the decimal point position in the next step.
5 kg
Select
6.
decimal
places
Decimal place range is 0 to 0.000. For example, if you enter “1” as
the grad size, it can represent 1, 0.1, 0.01, or 0.001, depending on
the decimal placement. Use the Up/Down buttons to select the
decimal place and press Enter. The displays show:
20000 kg
Enter
7.
scale
capacity
The maximum allowable scale capacity is 60000 lbs but can be set
to a smaller value as long as it is a multiple of the grad size.
(Example: 40000 lbs is a multiple of 1, 2, 5, 10, or 20 grads.) Use
the Up/Down buttons to set the value then press Enter. The
displays show:
20
Enter
42
allowed
grads to
re-zero
6/2/14
8.
Use the Up/Down buttons to select the maximum allowed grads
to re-zero then press Enter. This parameter represents the
maximum number of scale divisions that can be re-zeroed up or
down via the front panel “Zero” button, without requiring
calibration. The displays show:
No
Disable
9.
AZT?
Use the Up/Down buttons to select “Yes” or “No” then press
Enter. AZT (Automatic Zero-Tracking) is an Automatic means
provided to maintain the zero balance indication, within specified
limits, without the intervention of an operator. The displays
show:
No
Setup
another
display?
Yes / No
10. Use the Up/Down buttons to select “Yes” or “No” then press
Enter. “Yes” allows you to select another display to assign to a
scale. “No” exits setup.
11. Return the Calibration Enable switch to the “locked” position as
explained in the Calibration Enable Switch procedure to prevent
accidental changes.
12. Schedule a visit by a certified scale company or DOT
representative to re-seal the cover plate.
6/2/14
43
Setup Counter
1.
If you selected Counter in step 4 of Do This First, the displays
show:
A1
Select
2.
j-box
input
point
Each counter can be assigned up to five IO Box input points for
meter pulses. In other words, up to five water or admix meters
can share the same display if they have the same calibration (i.e.
the same number of ounces, milliliters, etc. per pulse). Use the
Up/Down buttons to select an input point for the meter wire and
press Enter. The displays show:
No
Select
3.
another
input
point ?
Use the Up/Down buttons to select “Yes” or “No” and press Enter.
“Yes” returns you to the previous step to select another input
point. This process can be repeated for up to five input points.
Select “No” when you are finished. The displays show:
SI
Select
4.
system
of units
Use the Up/Down buttons to select the system of units. The
choices are “SI” (for metric), or “U.S.” Press Enter. The displays
show:
mL
Select
5.
unit
Use the Up/Down buttons to select the unit and press Enter.
Choices for U.S. liquid units are “oz”, “pt”, “qt”, “gl”, “ig”, “lb”,
and “ ” (blank). Choices for SI (metric) liquid units are “L”, “mL”,
“cL”, “dL”, “kg”, and “ ” (blank). The displays show:
1 mL
Enter
44
grad
size
6/2/14
6.
Determine the “grad size” (number of units per pulse) of the
meter(s) which will be connected to this counter. Use the Up/
Down buttons to select the grad size and press Enter. Grad size
range is 0 to 125. If you need a fractional grad size (e.g. .5 oz per
pulse) enter it as a whole number (e.g. 5) and select a decimal
point location in the next step.
10 mL
Select
7.
decimal
place
Decimal place range is 0 to 0.000. For example, if you enter “1”, it
can represent an actual grad size of 1, 0.1, 0.01, or 0.001,
depending on the decimal placement. Use the Up/Down buttons
to select the decimal place and press Enter. The displays show:
Yes
Setup
8.
6/2/14
Another
display?
Yes / No
Use the Up/Down buttons to select “Yes” or “No” then press
Enter. “Yes” allows you to choose another display to assign to a
counter. “No” exits setup.
45
Setup Moisture Probe
1.
If you selected Probe in step 4 of Do This First, the displays show:
Select
Probe
module
(1 - 8)
8
Scale vs. Probe Modules:
Scale and Probe Modules are small printed circuit boards which are
mounted in a row across the inside rear of the E-Z Cal. Probe Modules are
marked “MPI Module” (for Moisture Probe Interface); Scale Modules are
marked “LCI Module” (for Load Cell Interface). Modules are numbered 1
through 8, beginning at the left. Make sure you know which modules are
for scales and which are for moisture probes.
2.
The first time you set up a probe, the number of the first
unassigned module is displayed. Use the Up/Down buttons to
select the Probe Module Position and press Enter.
There is nothing else to set up for a Moisture Probe. Probes are
automatically assigned a unit of “%”, a grad size of 0.01, and a maximum
capacity of 20.00 %. The displays show:
Setup
another
display?
Yes
3.
46
Use the Up/Down buttons to select “Yes” or “No” then press
Enter. “Yes” allows you to choose another display to assign to a
probe. “No” exits setup.
6/2/14
Setup Generic Analog Input
This procedure assumes that the Calibration Enable switch is unlocked. If
this switch is locked, the display will flash “Setup Locked” for about one
second, and then will display:
Yes
Setup
another
display?
Yes / No
(Refer to the Calibration Enable Switch section for instructions on unlocking
the Calibration Enable switch.)
1.
If you selected Generic in step 4 of Do This First, the displays
show:
1
Select
Analog
Input
module
(1 - 8)
Notes:
•
Analog Input Modules are small printed circuit boards mounted in a row
across the inside rear of the E-Z Cal. Analog Input Modules are marked
“LCI Module” and are numbered 1 through 8, beginning at the left. Your
E-Z Cal may also have one or more Moisture Probe Interface Modules,
marked “MPI Module.” Make sure you know which modules are used for
generic analog inputs and which are used for moisture probes.
•
Part numbers for Generic analog boards are: 24854 for 0-10v, 25503 for
4-20 ma.
2.
The first time you set up a generic analog input, the number of
the first unassigned module is displayed. Use the Up/Down
buttons to select the Analog Input Module position and press
Enter. The displays show:
%
Select
3.
unit
Use the Up/Down buttons to select the unit and press Enter.
Choices are “%” and “ ” (blank). If you selected “%”, the displays
show:
1%
Enter
6/2/14
grad
size
47
4.
Press Enter to accept “1”. If you need a fractional grad size (.1 %,
for example), enter it as a whole number (1) and set the decimal
point position in the next step.
1%
Select
5.
decimal
places
Decimal place range is 0 to 0.000. For example, if you enter “1” as
the grad size, it can represent a value of 1, 0.1, 0.01, or 0.001,
depending on the decimal placement. Use the Up/Down buttons
to select the decimal place and press Enter. The displays show:
100.0%
Enter
6.
max
reading
Use the Up/Down buttons to set the maximum display reading
then press Enter. The displays show:
20
Enter
7.
allowed
grads to
re-zero
Use the Up/Down buttons to select the maximum number of
counts that can be re-zeroed up or down via the front panel
“Zero” button, without requiring calibration. After selection, the
displays show:
No
Setup
8.
48
another
display?
Yes / No
Use the Up/Down buttons to select “Yes” or “No” then press
Enter. “Yes” allows you to select another display to assign to a
scale. “No” exits setup.
6/2/14
Calibration Procedures
Use this procedure to calibrate scales and moisture probes. You cannot
calibrate a counter.
•
For Scales, you can zero the scale and set the correct weight using test
weights.
•
For Moisture Probes, you can set the expected moisture and the span
value. (The span value determines the relative sensitivity of the Moisture
display.)
•
For Generic, you can zero the display and set the correct display reading
using test signals.
Topics in This Section
Do This First
Scale Calibration on the E-Z Cal
Re-Span Scale
Full-Scale Calibration Completed
Moisture Probe Calibration on the E-Z Cal
Generic Analog Input Calibration on the E-Z Cal
6/2/14
49
Do This First
1.
Press the “Setup/Calib” button. The displays show:
Info
Select
Menu
Option
•
If the Calibration Enable switch is unlocked, you can calibrate any scale
or moisture probe.
•
If this switch is locked, you can only calibrate moisture probes. You will
not be permitted to calibrate scales.
2.
Use the Up/Down buttons to select “Calib” and press Enter. The
displays show:
Disp. 1
Select
50
display
to calib
3.
Use the Up/Down buttons to move through the displays. (For
displays 7 -12 the message moves to the top line.) Select the
appropriate display and press Enter.
•
If the display is set up for a Scale, continue with Scale Calibration on the
E-Z Cal.
•
If the display is set up for a Probe, continue with Moisture Probe
Calibration on the E-Z Cal.
•
If the display is set up for Generic, continue with Generic Analog Input
Calibration on the E-Z Cal.
6/2/14
Scale Calibration on the E-Z Cal
This procedure assumes that the Calibration Enable switch is unlocked. If
this switch is locked, the display will flash “Calib Locked” for about one
second, and then display:
No
Calib
another
display?
(Refer to the Calibration Enable Switch section for instructions on unlocking
the Calibration Enable switch.)
1.
If you selected Scale in step 3 of Do This First, and if the scale
was calibrated previously, you will be asked if you want to zero
the scale (see the following displays):
Yes
Zero
2.
scale?
Use the Up/Down buttons to select “Yes” or “No”.
—
If you select “No”, go to the Re-Span Scale section for a simplified
procedure to readjust the full-scale calibration.
—
If you select “Yes”, or if this is the first time this scale has been
calibrated, the displays show:
XXXX
Empty
scale &
press
zero
Where: XXXX is the live raw scale reading.
3.
Empty the scale, attach chains or a weight platform to hold the
test weights, and press the “Zero” or “Enter” button for that
scale. The displays show “Wait...” for several seconds then show:
2000 kg
Apply
6/2/14
weight &
enter
value
4.
Add calibrated test weights to the scale, and use the Up/Down
buttons to set the displayed weight to the proper value. When the
scale is stable, press Enter. The displays show “Wait . . .” for
several seconds.
5.
Go to the Full-Scale Calibration Completed section.
51
Re-Span Scale
If the scale was previously calibrated, and you selected “No” in step 2 of Scale
Calibration on the E-Z Cal, the displays show:
2000 kg
Apply
1.
weight &
press
enter
Add calibrated test weights to the scale. When the scale is stable,
press Enter. The displays show:
2000 kg
Enter
2.
correct
value
Use the Up/Down buttons to set the displayed weight to the
proper value. Press Enter. The displays show “Wait . . .” for
several seconds then show:
No
Calib
3.
52
another
display?
Use the Up/Down buttons to select “No” and exit calibration.
6/2/14
Full-Scale Calibration Completed
At this point, one of three things can occur:
•
The displays may show “Cell output too low.” This means the load cell
is grossly oversized or that there is a wiring problem. In technical terms,
it means that the load cell’s “span” (the difference in its output voltage
between zero and full scale) is less than 1.6 millivolts. A properly sized
load cell will have a span in the range of 5 to 15 millivolts. (The E-Z Cal
uses an excitation voltage of 5 volts.) Contact your scale manufacturer.
•
The displays may show “Cell output too high.” This means the load cell
is grossly undersized or that there is a wiring problem. In technical terms,
it means that the load cell’s full-scale output voltage is greater than 23
millivolts. A properly sized load cell will never have an output more than
about 16 millivolts. Contact your scale manufacturer.
•
Or, if all goes well, the displays show:
No
Calib
another
display?
If the scale is being calibrated with test weights whose total weight is less
than the full scale capacity of the scale, it is typical to perform a “build-up”
test at this point:
6/2/14
1.
Use the Up/Down buttons to select “No” then press Enter. “No”
exits setup and returns to the normal weight display mode.
2.
Remove the test weights from the scale, and verify that the
display goes back to its starting point.
3.
Manually batch an amount of material which is approximately the
same as the total weight of the test weights.
4.
Add the test weights back onto the scale. The E-Z Cal display
should show a value equal to the sum of the material weight in
step 3 plus the test weights.
5.
Repeat steps 2 - 4 until the scale’s capacity is reached.
6.
If the displays do not track with the expected weight values, you
may want to repeat the entire calibration process.
7.
When calibration is completed, remove all material from the
hopper, and remove the chains or weight platform from the scale.
The displays show a negative value. Press the “Zero” button for
that scale.
8.
When all scales and probes have been calibrated, go to
Calibration Enable Switch and finish the procedure, starting from
step 9.
53
Moisture Probe Calibration on the E-Z Cal
1.
If you selected Probe in step 3 of Do This First, the displays show:
Enter
current
moisture
value
0.00 %
2.
Use the Up/Down buttons to set the moisture to the expected
value and press Enter. The displays show:
Enter
probe
factor
1.00
3.
Use the Up/Down buttons to set the probe factor. A value of 1.00
works for most concrete sands. Decrease this to .30 for coarse
aggregates. Increasing the probe factor makes the display more
sensitive. Press Enter when finished. The displays show:
No
Calib
4.
54
another
display?
Use the Up/Down buttons to select “Yes” or “No” and press Enter.
“Yes” allows you to choose another display to calibrate. “No”
exits the calibration procedure.
6/2/14
Generic Analog Input Calibration on the E-Z Cal
1.
If you selected Generic in step 3 of Do This First, and if the Analog
Input was previously calibrated, the displays show:
Yes
Zero
signal?
2.
Use the Up/Down buttons to select “Yes” or “No”.
•
If you select “Yes”, the displays show:
Zero
—
signal &
press
zero
Apply a signal to the Analog Input that corresponds to a
display value of zero then press Enter. The displays show
“Wait...” for several seconds then show:
100.0%
Apply
•
signal &
enter
value
If you select “No”, the displays show the information above.
—
3.
Apply a known analog signal to the input module and use the
Up/Down buttons to set the correct display reading. Press
Enter when finished. The displays show “Wait...” for several
seconds.
If the signal is too low or too high, the appropriate warning
message is displayed. Adjust the signal then press Enter. The
displays should now show:
No
Calib
4.
6/2/14
another
display?
Use the Up/Down buttons to select “Yes” or “No” and press Enter.
“Yes” allows you to choose another display to calibrate. “No”
exits the calibration procedure.
55
View Configuration Settings - Info Mode
In Info mode, you can view configuration settings without changing them.
To check configurations without changing any settings:
1.
Press the “Setup” button. The displays show:
Info
Info,
2.
Setup,
Calib,
or Exit?
Press Enter to go into Info mode. The displays show:
Disp. 1
Select
3.
display
to view
Use the Up/Down buttons to move through the displays. (For
displays 7 -12 the message moves to the top line.) Select the
appropriate display and press Enter. The displays show the name
of the setup parameter, and its value. For example, if “Disp. 1” is
selected, the display might show:
Scale
Display
4.
type
Press Enter successively to scroll through the parameters and
their values. You will see different parameters as listed in the
following table, depending on the type of device (scale, counter,
moisture probe, or generic).
Scale
Parameters
Counter
Parameters
Moisture Probe
Parameters
Scale module
J-box input point (up to 5)
System of units
System of units
System of units
Unit
Unit
Unit
Grad size
Grad size
Grad size
Probe module
Generic Parameters
Input module
Scale Capacity
Max Display
Allowed grads to re-zero
Allowed grads to re-zero
Ch. B value
0
Disable AZT (Yes/No)
56
6/2/14
5.
When you reach the end of the parameter list for the device, the
displays show:
No
View
6.
6/2/14
another
display?
Use the Up/Down buttons to select “Yes” or “No” then press
Enter. “Yes” allows you to choose another display to view. “No”
exits Info mode.
57
View System Settings and Select Certification Type
Under the System menu, you can view software and hardware version
numbers, view the E-Z Cal’s IP Address, and select the certification type
(NTEP or OFF).
To view Version numbers:
1.
Press the “Setup/Calib” button. The displays show:
Info
Select
2.
Menu
Option
Use the Up/Down buttons to select “System”. The displays show:
System
Select
3.
System
Option
Press Enter. The displays show:
Version
Select
System
Option
4.
Press Enter to see the Firmware Version.
5.
Press Enter again to see the Scale Version.
6.
Press Enter again to see the CPLD Version.
7.
Press Enter to exit and go back to the main menu.
To view the IP Address:
Note:
1.
This procedure is reserved for Future Use.
Press the “Setup/Calib” button. The displays show:
Info
Select
58
Menu
Option
6/2/14
2.
Use the Up/Down buttons to select “System”. The displays show:
System
Select
3.
System
Option
Press Enter. The displays show:
Version
Select
4.
System
Option
Press the Up/Down buttons until the displays show:
IPCONFIG
Select
5.
System
Option
Press Enter. The displays show:
IPCONFIG
192.168
6.
79.200
Press Enter to exit and go back to the main menu.
To select the scale Certification type:
1.
Press the “Setup/Calib” button. The displays show:
Info
Select
2.
Menu
Option
Use the Up/Down buttons to select “System”. The displays show:
System
Select
3.
System
Option
Press Enter. The displays show:
Version
Select
6/2/14
System
Option
59
4.
Press the Up/Down buttons until the displays show:
Certify
Select
5.
System
Option
Press Enter then press the Up/Down buttons until the displays
look like one of the following examples, depending on whether
you want to turn on NTEP certification:
NTEP
Certify
Type
OFF
Certify
6.
60
Type
Press Enter to exit and go back to the main menu.
6/2/14
Technical Specifications
Topics
Manual Station
Scale Interface
Meter Counter
Moisture Probe Interface
IO Box
Input Modules
Output Modules
Load Cell Color Codes
Mean Time Before Failure - Manual Station
Mean Time Before Failure - IO Box
Manual Station
CAUTIONS!
This equipment was manufactured to operate under
normal “dry” industrial conditions. Extreme
temperature or weather conditions can and will cause
damage to the equipment and bodily harm.
Per IEC 61010, no operator access is permitted in this
unit! Service should only be performed by qualified
personnel.
A Circuit Breaker (rated at 20 Amperes 240 Volts AC)
should be installed near the IO Box, be easily
distinguishable from other controls, and labeled to
indicate that it is the Disconnect device.
Warning!
HAZARDOUS LIVE VOLTAGE PRESENT! This equipment
must be isolated or disconnected before it is accessed.
Note: A certified representative of Command Alkon should make
monthly inspections of this unit’s interior.
Voltage - when used with 120VAC IO Box
110 to 120 VAC.
6/2/14
61
Voltage - when used with 240VAC IO Box
200 to 240 VAC.
Frequency
50/60 Hz.
Current
5 Amp.
Power Supply Range
90 to 260 VAC, 49.5 to 60.5 Hz.
Power Cord Type and Rating
SJTO, 3-conductor plastic detachable, black, 26’ with C19 female plug
Temperature Rating: 60 oC
AWG Stranding 14
Rating: 1250 Watts 10A-125V
Radio Frequency Interference
Displays will not change by more than one division in the presence of
“normal” RF fields. A normal scenario is defined as a 5-watt hand-held
radio operated at least one meter from the E-Z Cal, within the
commercial radio bands between 27 and 460 MHz.
Operating Temperature
+10oC to +40oC.
Storage Temperature
-20oC to +50oC.
Relative Humidity
10% to 90%, non-condensing.
Altitude
0 to 3000 m.
Weight
25 lbs.
Dimensions
Height: 17 5/8”, Width: 18 1/4”, Depth: 3 7/8” (5 1/4” with power
supply).
62
6/2/14
Maximum Display Capacity
6 characters, including a decimal point.
Maximum scale capacity that can be entered:
—
60000 for whole integer precision graduations
—
6000.0 for one decimal precision graduations
—
600.00 for two decimal precision graduations
—
60.000 for three decimal precision graduations
Minimum Scale Division Size
0.001.
Maximum Scale Division Size
125 (You must limit your choice of scale division size to decimal multiples
of 1, 2, 5, 10, or 20 to comply with NIST Handbook - 44.)
Agency Sealing Method
Lead-wire seal, inserted through the protruding edge of a metal shield on
the underside of the enclosure. The shield prevents access to any
components that can affect the weighing integrity of the device.
Labeling
The names of the scale materials, scale capacities, and division sizes are
printed on labels which are inserted from the edge of the front control
panel, and are protected from abuse by a clear overlay.
All operator controls and field-installable connectors are labeled
according to their use.
The equipment identification label is a permanent, adhesive, destructible
mylar label at the top of the left side panel. A unique serial number is
added at the Command Alkon factory with an indelible ink marker or
stamp.
Cleaning
Clean the exterior of the Manual Station with a mild soap solution and a
soft towel. No liquids should be introduced to the interior of the
Manual Station!
6/2/14
63
Scale Interface
Number of Scale Divisions
4000 maximum.
Load Cell Excitation
5 VDC, 60 mA (4 350-ohm load cells).
Load Cell Wiring
Each load cell must be wired with a shielded 6-conductor cable,
terminating at the Command Alkon Manual Station or Calibration Box.
The cable must meet the following specifications:
—
Wire Gauge: #20 - 24
—
Overall Cable Diameter: 0.32" max
—
Insulation Material: PVC, teflon
—
Shield: 85% min, 100% preferred
—
Length: 1300' max
—
Suggested Type: Belden 9536 (Command Alkon PN-19320)
All load cell connections or cable splices must be made in waterproof
enclosures in accordance with accepted scale industry methods.
Dead-Load Offset
0 mV to 13 mV (0% to 85% of a 3 mV/V load cell).
Full-Scale Input
1.6 mV to 23.5 mV (11% to 155% of a 3 mV/V load cell).
Under-Zero
The display shows negative weight values below the calibrated zero point.
The value "- 0" does not occur.
Over-Capacity
The display shows "OVER" if the weight value exceeds the programmed
scale capacity value.
64
6/2/14
Meter Counter
Maximum Number of Counts
Limited only by display capacity.
Maximum Count Rate
Approximately 30 pulses per second, depending on duty cycle of meter
contacts.
Moisture Probe Interface
Capacity
20.00% x .01%.
Moisture Probe Wiring
The Moisture Probe must be wired with a shielded 6-conductor cable,
terminating at the Command Alkon Manual Station. The cable must meet
the following specifications:
—
Wire Gauge: #l8 (1300' max) - #24 (300' max)
—
Overall Cable Diameter: 0.32" max
—
Insulation Material: PVC, teflon
—
Shield: 85% min, l00 preferred
—
Suggested Type:
Belden 9536 (Command Alkon PN-l9320) 300' max
Belden 9506 (Command Alkon PN-20393) l300' max
6/2/14
65
IO Box
Caution!
This equipment was manufactured to operate under
normal “dry” industrial conditions. Extreme
temperature or weather conditions can and will cause
damage to the equipment and bodily harm.
Per IEC 61010, no operator access is permitted in this
unit! Service should only be performed by qualified
personnel.
A Circuit Breaker (rated at 20 Amperes 240 Volts AC)
should be installed near the IO Box, be easily
distinguishable from other controls, and labeled to
indicate that it is the Disconnect device.
Warning!
HAZARDOUS LIVE VOLTAGE PRESENT! This equipment
must be isolated or disconnected before it is accessed.
Note: An earth ground connection must be installed according to local
building code and industrial requirements.
Voltage - 120VAC IO Box
110 to 120 VAC.
Voltage - 240VAC IO Box
200 to 240 VAC.
Frequency
50/60 Hz.
Current
20 Amp.
Plant Outputs - 120VAC IO Box
3A Each Max. 10A Total.
Plant Outputs - 240VAC IO Box
1A Each Max. 10A Total.
Power Supply Range
90 to 260 VAC, 49.5 to 60.5 Hz.
66
6/2/14
Weight
95 lbs.
Dimensions
Height: 30 1/4”, Width: 24 3/8”, Depth: 8 5/8”.
Note: Use proper lifting techniques when lifting the Manual Station or
the IO Box. Always lift with your legs, keeping your back as
straight up and down as possible. Two people should be used to
lift the IO Box.
Operating Temperature
+10oC to +40oC.
Storage Temperature
-20oC to +50oC.
Relative Humidity
10% to 90%, non-condensing.
Altitude
0 to 3000 m.
Maximum Number of Plant Connections per Board
24.
Cleaning
Clean the exterior of the IO Box with a mild soap solution and a soft
towel. No liquids should be introduced to the interior of the IO
Box!
Labeling
All operator controls and field-installable connectors are labeled
according to their use.
The equipment identification label is a permanent, adhesive, destructible
mylar label at the top of the left side panel. A unique serial number is
added at the Command Alkon factory with an indelible ink marker or
stamp.
6/2/14
67
Input Modules
Input Module
Voltage
CAI Part No.
Crydom M-IDC5N
12 VDC
24900-A
Crydom M-IDC5N
24 VAC
24900-A
Grayhill 70M-IAC5
120 VAC
22828
Grayhill 70M-IAC5A
240 VAC
22830
Crydom Input Modules - General Specs
Crydom Input Module M-IDC5N
68
6/2/14
Grayhill Input Module 70M - General
Grayhill Input Module 70M-IAC5
Grayhill Input Module 70M-IAC5A
6/2/14
69
Output Modules
Output Module
Voltage
CAI Part No.
Grayhill 70M-ODC5
12 VDC
24465
Crydom M-OACUA
(M-OAC5A)
24 VAC
25524
Crydom M-OACUA
(M-OAC5A)
120 VAC
25524
Crydom M-OACUA
(M-OAC5A)
240 VAC
25524
Grayhill Output Module 70M - General
Grayhill Output Module 70M-OCD5
70
6/2/14
Crydom Output Module M-OAC5A
6/2/14
71
Load Cell Color Codes
Manufacturer
+Excitation
- Excitation
+Signal
- Signal
Shield
A&D Engineering
RED
WHITE
GREEN
BLUE
YELLOW
Allegany
GREEN
BLACK
WHITE
RED
Bare
Artech
RED
BLACK
GREEN
WHITE
Bare
Beowulf
GREEN
BLACK
WHITE
RED
Bare
BLH
GREEN
BLACK
WHITE
RED
YELLOW
Celtron
RED
BLACK
GREEN
WHITE
Bare
Cardinal
GREEN
BLACK
WHITE
RED
Bare
Dillon
GREEN
WHITE
BLACK
RED
Orange
Electroscale
RED
BLACK
GREEN
WHITE
Bare
Flintec
GREEN
BLACK
WHITE
RED
YELLOW
Force Measurement
RED
BLACK
GREEN
WHITE
Bare
GSE
RED
BLACK
WHITE
GREEN
Bare
HBM
GREEN
BLACK
WHITE
RED
YELLOW
Interface
RED
BLACK
GREEN
WHITE
Bare
Kubota
RED
WHITE
GREEN
BLUE
YELLOW
Mettler Toledo
WHITE
BLUE
GREEN
BLACK
ORANGE
National Scale
GREEN
BLACK
WHITE
RED
YELLOW
NCI
RED
BLACK
WHITE
GREEN
Bare
Pennsylvania
ORANGE
BLUE
GREEN
WHITE
Bare
Philips
RED
BLUE
GREEN
GRAY
Bare
Revere Transducers
RED
BLACK
GREEN
WHITE
ORANGE
Rice Lake Weighing
RED
BLACK
GREEN
WHITE
Bare
Sensortronics
RED
BLACK
GREEN
WHITE
Bare
Sensotec
RED
BLACK
WHITE
GREEN
Bare
Strainsert
RED
BLACK
GREEN
WHITE
Bare
T-Hydronics
RED
BLACK
GREEN
WHITE
Bare
Tedea Huntleigh
GREEN
BLACK
RED
WHITE
Bare
Toledo
GREEN
BLACK
WHITE
RED
YELLOW
Totalcomp
RED
BLACK
GREEN
WHITE
Bare
Transducers Inc.
RED
BLACK
GREEN
WHITE
ORANGE
Weigh-Tronix
GREEN
BLACK
WHITE
RED
WHITE/ORANGE
72
+Sense
- Sense
YELLOW
RED
YELLOW
BLUE
BLUE
BROWN
•
Color code complies with ISA S37.8-1975 “Specifications for Tests for
strain gauge force transducers” also per Western Regional Strain Gauge
Committee Standard.
•
This cross reference chart allows you to replace one brand load cell with
another when they have the same output (mv/v) and the same physical
size.
•
To correct an instrument that displays a weight in the wrong direction,
interchange the +Output and -Output connections.
6/2/14
Example: If the load on the scale is increasing and the instrument
display is decreasing, reverse the Output connections. Do not reverse
the Excitation connections. The readout should now display increasing
readings with increasing load.
This information was reprinted with the permission of Totalcomp, Inc., 13-01
Pollitt Drive, Fair Lawn NJ 07410
201-797-2718, 800-631-0347
©Copyright 1996 Totalcomp, Inc.
Not to be copied or duplicated for any purpose without consent of copyright
owner. All rights reserved.
Mean Time Before Failure - Manual Station
The total Mean Time Before Failure (MTBF) for the Manual Station, as
determined by Commercial/Bellcore Method 1, is 39,742.83 Hours.
Mean Time Before Failure - IO Box
The total Mean Time Before Failure (MTBF) for the IO Box, as determined by
Commercial/Bellcore Method 1, is 36,529.52 Hours.
6/2/14
73
E-Z Cal Parts Replacement Guidelines
Topics
Manual Station
Boards
IO Modules
IO Box Fuses
Spare Parts List
Manual Station
Caution!
Parts located in the interior of the Manual Station are low
voltage and static sensitive. Only authorized service
personnel should replace these parts.
Boards
Risk of Electrical Shock!
High voltage is present in the power supply area once the
back cover has been removed! Use extreme caution when
working on the power supply. Improper replacement could
result in equipment damage or bodily injury and should
only be done by qualified service personnel!
Power down the Manual Station and disconnect the power cord. All boards
except for the base board and the power supply are attached with standoffs
and connectors. The base board and power supply are attached with screws.
IO Modules
Risk of Electrical Shock!
Before replacing any components in the Manual Station,
disconnect the power. Failure to comply with these
guidelines can result in severe equipment damage or bodily
injury! Only qualified service personnel should replace
components in this unit.
1.
74
The output and input modules in the IO Box are located in the
center of the 22821 board. Carefully remove the clear milar strip
then gently pull out the old module. (See the 22821 Opto Board
Drawing for reference.)
6/2/14
2.
Insert a new module with the SAME electrical specifications as
the one being removed.
Input Module: Insert the Yellow IAC module into the socket and place
the three-position black jumper in the correct reference position.
Connect the proper gauge wire to the corresponding terminal at the
bottom of the board.
Output Module: Insert the Black OAC module into the socket and place
the three-position black jumper in the correct reference position.
Connect the proper gauge wire to the corresponding terminal at the
bottom of the board.
3.
Mark up the appropriate wiring prints and send them to the
Command Alkon Service Department.
IO Box Fuses
As a troubleshooting aid, a neon bulb is located near each fuse or holder and
is lit if the fuse blows.
Warning!
Disconnect the power before replacing blown fuses.
Caution!
Use caution when removing the fuse so that the glass
doesn’t break and cause you bodily harm.
The new fuse must not be rated higher than what is
required for that fuse location. Fuse types and ratings
are shown in the Spare Parts List.
6/2/14
75
Spare Parts List
Item
Command Alkon
P/N
Manufacturer’s
P/N
Fuse, Time Delay, 10 Amp 250 Volt, 1/ 23626
4” x 1 1/4”
BUSS MDA-10
Fuse, GMA, Quck, 5 Amp 250 Volt, 5 x
20 mm
23045
Little Fuse 217005
Fuse, GDB, Quick Acting, 5 Amp 250
Volt, 5 x 20 mm
23045
BUSS GDB-5A
Relay, 8 Blade, DPDT, 120 VAC 16 Amp 23777
76
Relay, 11 Blade, EPDT, 120 VAC 16
Amp
23778
Socket, 8 Pin
23781
Socket, 11 Pin
23782
Module, mini opto, IAC5, 120 VAC,
Input, Yellow
22828
Module, mini opto, IAC5, 220 VAC,
Input, Yellow
22830
Module, mini opto, OAC5, 120 VAC,
Input, Black
22829
Module, mini opto, OAC5, 220 VAC,
Input, Black
22831
6/2/14
Index
E
E-Z Cal 8
Calibration Procedure
Do This First 50
Generic Analog Input Calibration 55
Moisture Probe Calibration 54
Scale Calibration 51
Calibration Procedures 49
Certification Type - Selecting 58
Command Alkon Offer of Source for GPL Software 11
Components 12
I/O Box 15
Manual Station 12
Configuration Settings - Viewing 56
General Operation 28
Clearing a Bad Cell Message 31
E-Z Cal Displays 30
If Communication to the Batching PC is Lost 32
Operating Plant Equipment 29
Powering On/Off 28
Installation and Wiring 19
120 Volt Power Connections 26
240 Volt Power Connections 27
I/O Box Power Source 120VAC 23
I/O Box Power Source 240VAC 23
I/O Box Supply Wiring 23
I/O Box Voltage Jumper Settings 23
I/O Cable Connections 25
Manual Station Power Source and Supply 21
Mounting 20
Ventilation 20
Water and Moisture 20
Overview 8
Parts Replacement Guidelines 74
I/O Box 75
Manual Station 74
Spare Parts List 76
Regulatory Compliance 10
Setup and Calibration 33
Calibration Enable Switch 35
Operating Modes 33
6/2/14
77
Setup Buttons 34
Setup Procedure
Do This First 40
Set Up Counter 44
Set Up Generic Analog Input 47
Set Up Moisture Probe 46
Set Up Scale 41
Setup Procedures 39
System Settings - Viewing 58
Technical Specifications 61
I/O Box 66
Input Modules 68
Load Cell Color Codes 72
Manual Station 61
Mean Time Before Failure - I/O Box 73
Mean Time Before Failure - Manual Station 73
Meter Counter 65
Moisture Probe Interface 65
Output Modules 70
Scale Interface 64
I
Introduction 5
Audience 5
Purpose 5
Revision Summary 5
78
6/2/14