1|Page User Manual EnviroPro Soil Moisture Sensor Version 4.7 August 2014 Copyright 2014 EnviroPro Dielectrics 2|Page Table of Contents 1 The EnviroPro Sensor .......................................................................... 4 2 Specifications ........................................................................................ 6 3 Ordering Details: ................................................................................... 8 4 Warranty ............................................................................................... 9 5 Wiring Details ...................................................................................... 10 5.1 Sensor Wiring ............................................................................. 10 5.2 Power Supply .............................................................................. 10 6 Sensor Setup and Configuration ........................................................ 11 7 Standard SDI-12 Commands ............................................................. 12 7.1 Initialisation and power up .......................................................... 13 7.1.1.1 Initialisation ..................................................................... 13 7.1.1.2 SDI-12 Operation ............................................................ 13 7.2 SDI-12 Info Command ................................................................ 14 7.3 SDI-12 Measurement.................................................................. 15 7.3.1 7.4 8 Limitations of M Command ................................................. 16 To Change a Probe Address ...................................................... 17 Installation ........................................................................................... 18 8.1 Soil Slurry.................................................................................... 18 8.2 Bentonite Slurry .......................................................................... 18 8.3 Installation ................................................................................... 19 8.4 Cable Considerations ................................................................. 19 8.4.1 At the Probe end ................................................................. 20 8.4.2 At the logger/telemetry end................................................. 20 8.5 Probe Removal ........................................................................... 21 8.5.1 Probe Removal Tool ........................................................... 21 3|Page 8.5.2 8.6 9 Removal with Vice Grips ..................................................... 23 Slurry and Probe Installation....................................................... 24 8.6.1 Large format (50mm) probes .............................................. 24 8.6.2 EnviroPro and Slurry........................................................... 26 Connection to Data Loggers / Telemetry Systems ............................. 27 10 Sensor Calibration .......................................................................... 28 10.1 Simple Calibration ....................................................................... 28 10.2 Bucket Calibration ....................................................................... 29 11 Regulatory Compliance Declaration ............................................... 30 4|Page 1 The EnviroPro Sensor This manual is written for the Series “D” EnviroPro family of soil moisture probes. Manuals for earlier versions of the product are available on request from EnviroPro Dielectrics Pty Ltd. The “D” version probes are recognizable by their green colour (earlier versions utilized a white outer tube). The EnviroPro sensor monitors changes in the dielectric properties of the soil to provide users with continuous monitoring of soil moisture, soil salinity and soil temperature. 5|Page Simple capacitance or dielectric sensors, which determine soil moisture levels using a frequency change, suffer from errors caused by the salinity of the water in the soil. By measuring the salinity of the soil, the EnviroPro soil probe can compensate for the salt induced changes and provide a more accurate soil moisture measurement. Measurement of soil salinity provides users with additional information to help them improve the application of nutrients and to deal with irrigation induced salt build up. The electronic components on the probe are fully encapsulated in epoxy – nothing can move, and all circuits are protected against moisture and chemicals, which makes it reliable, consistent and very stable. The EnviroPro is completely maintenance free with a long lifespan. The probe does not need to be removed and opened up for annual maintenance like competing sensors. To provide certainty to users, EnviroPro Dielectrics provides the probes with a 5 year, return to base warranty. The EnviroPro is available in a number of standard segment lengths: 40cm, 80cm, 120cm and 160cm. Sensors are located at 10cm along the probe body. A 40cm probe this carries 4 sensors and a 160cm probe 16. Two variants of the probe are available: • The EnviroPro – which returns soil moisture and soil temperature from each sensor element • The EnviroPro EC – which returns soil moisture, soil temperature and soil conductivity from each sensor element. Upon payment of the difference in price between the standard and EC version, the former may be upgraded to return EC at any time. This requires use of the EnviroPro Probe Updater unit. When an upgrade is ordered (quoting the serial number), your distributor will be sent a probe specific unlock code, which is entered into the Probe Updater unit. The EnviroPro probe connects to any SDI-12 compatible data logger. If more than one probe is to be connected to the same logger, the probes need to be programmed to occupy different addresses: e.g. the first as Address 0 and the second as Address 1. The address can be changed at any time using the standard SDI-12 commands (Refer Section 6). 6|Page 2 Specifications The following table shows the specifications of the EnviroPro sensor. Parameter Sensor Depths Sensor Position on Probe Body Range / Specification Accuracy/ 10cm Increments - Qualifier Top sensor centred 5cm below top of probe body Dimensions Diameter Length 32mm 40cm – 47cm 80cm – 87cm 120cm – 127cm 160cm – 167cm Interface SDI-12 - +6 to +15 V DC The probe is protected against reverse polarity and incorrect connection Sleep 0.4 mA (per 40cm segment) During Measurement 60 mA For 0.6 sec During Data Transfer 18 mA For 0.4 sec 0-50% +/- 2 % Power Supply Power Consumption Soil Moisture (Volumetric Soil Moisture Content) (in standard calibration media) o Soil Temperature -10 to +60 Celsius Soil Salinity 0 to 6 dS/m o + / - 1 Celsius Relative value only. Requires 7|Page calibration on site to local soil Operating Range Cable Length Cable termination -10 to 100 Celsius o - 5m - Bare wire Connectors available on request for volume quantities Note: Soil Moisture Accuracy In common with all other capacitance probes, the EnviroPro responds to changes in the dielectric properties of the soil. The relationship between dielectric properties and volumetric moisture content is well understood and widely documented. However it varies from probe design to probe design and from one soil type to another. The EnviroPro is calibrated at the time of manufacture, to a standardized soil media (fine sand) and it is the accuracy of the sensors in this media which is shown in the table in the preceding section. Secondary calibrations are then used to adjust the probe output for specific soils. Further details are given in the section on probe calibration (Section 10). Electrical Conductivity Accuracy The EnviroPro responds to changes in the soil’s bulk conductivity which is influenced by soil type, moisture content, soil temperature and the total salt load of the soil. Bulk conductivity values should be compared between two periods where the soil moisture content is the same. The relationship between bulk conductivity and true pore water conductivity is site specific. To convert the indicated EC to pore water conductivity, the values obtained from the probe should be compared with those obtained from soil samples or from a suction sampler. 8|Page 3 Ordering Details: When ordering or identifying probes, please specify the model and version as per the table below: EP100DL-04 EP100DL-08 EP100DL-12 EP100DL-16 40cm 4 sensor probe for SM and ST 80cm 8 sensor probe for SM and ST 120cm 12 sensor probe for SM and ST 160cm 16 sensor probe for SM and ST EP100D-04 EP100D-08 EP100D-12 EP100D-16 40cm 4 sensor probe for SM ST and EC 80cm 8 sensor probe for SM ST and EC 120cm 12 sensor probe for SM ST and EC 160cm 16 sensor probe for SM ST and EC The Upgrader Unit is a distributor only tool and is required in order to enable EC on a standard probe. Your distributor can perform the upgrade for you. Part numbers for the upgrades are shown below: EP-UPG-04 EP-UPG-08 EP-UPG-12 EP-UPG-16 Upgrade 40cm probe to enable EC Upgrade 80cm probe to enable EC Upgrade 120cm probe to enable EC Upgrade 160cm probe to enable EC EP-UPDATER EnviroPro Upgrader Unit (required to upgrade Standard probes to enable the EC measurement, only available to distributors) 9|Page 4 Warranty The EnviroPro probes are covered by a five (5) year warranty. Warranty is available on a return to base basis only. Probe owners must pay for return shipment of faulty products either to EnviroPro Dielectrics or their local distributor. If the probe is assessed by EnviroPro Dielectrics and found to be a warranty failure, it will be replaced free of charge. EnviroPro Dielectrics will pay the return shipment to the owner. The warranty does not cover mechanical damage, damage inflicted during probe installation or removal or damage caused by animals. Prior to installing the product, please ensure that you read, understand and accept the Warranty Statement. If you do not accept the conditions of the Warranty Statement, please return the probe to the probe supplier for a refund. 10 | P a g e 5 Wiring Details 5.1 Sensor Wiring The EnviroPro probes are normally shipped fitted with a 5m cable, the colour code for which is shown in the table below: Function Power Supply (+6 to +15V) Ground SDI 12 Data 3 Core Cable 4 Core Cable Brown Red Green/Yellow Black Blue Blue NOTE: The probes are protected against any combinations of incorrect connections up to 30V AC or DC. 5.2 Power Supply The EnviroPro sensors can operate from a power supply in the range of +6 to +15Volts. 11 | P a g e 6 Sensor Setup and Configuration The sensors should be configured and tested prior to installation in the soil. This may be done with one of 3 methods: From the telemetry or logging device to which the probe will be connected (using transparent mode) Using an SDI-12 Verifier or USB to SDI-12 converter such as the Tekbox TBS03 (which provides a complete terminal program, SDI Commander, for sending and receiving SDI-12 commands) Using an SDI-12 Sniffer. The Sniffer” can be connected across the SD-12 data line between the SDI-12 logging device and the probe. 12 | P a g e 7 Standard SDI-12 Commands The EnviroPro sensors are compatible with the basic requirements of the SDI-12 interface specification (SDI-12 Specification Version 1.2, October 1996). The commands can be sent to the sensor using a PC and terminal software – an SDI-12 verifier is not required. The sensor will respond to the following SDI-12 commands: Name Command Response Break Continuous spacing for at least 12 milliseconds Sensor / Interface unit wakes from Sleep mode Acknowledge Active a! a<CR><LF> Send Identification aI! allccccccccmmmmmmv vvxxx...xx<CR><LF> Change Address aAb! b<CR><LF> (Note 1) Address Query ?! a<CR><LF> (Note 2) Start Measurement aM! atttn<CR><LF> Send Data aD0! a<values><CR><LF> . a<values><CR><LF> . a<values><CR><LF> . a<values><CR><LF> aD9! a<values><CR><LF> aM1! atttn<CR><LF> . .atttn<CR><LF> . . atttn<CR><LF> . . atttn<CR><LF> Additional Measurements . aM9! atttn<CR><LF> Start Verification aV! atttn<CR><LF> Start Concurrent aC! atttnn<CR><LF> 13 | P a g e Measurement Additional Concurrent Measurements aC1! atttnn<CR><LF> . atttnn<CR><LF> . atttnn<CR><LF> . atttnn<CR><LF> aC9! Continuous Measurements aR0! ... aR9! a<values><CR><LF> (same as the D commands) In the above table: a Sensor address (0 to 9, a to z, A to Z) As a minimum, you should confirm that the sensor responds to an “I” command and preferably also perform a Concurrent (C) Measurement with the probe inserted into a suitable test medium. A test medium may be made up by placing a wetted sponge into a zip lock bag. The bag may be wrapped around the probe body and secured with a rubber band. It may then be slid up and down the probe body and measurements taken at each sensor position. 7.1 Initialisation and power up 7.1.1.1 Initialisation Under the SDI-12 protocol, the sensor interface normally remains permanently powered. The sensors connected to the interface go in to idle mode between reads and draw minimal current between measurements. After receipt of a Break Command, all sensors on the bus wake and listen for a command addressed to them. Note however that to save power, some logging devices / telemetry units support a Power Save mode, which shuts off sensor power between reads. The EnviroPro is compatible with Power Save mode. 7.1.1.2 Sleep SDI-12 Operation 14 | P a g e When the Interface is not executing any commands, it reverts to Sleep mode to minimize power consumption although the SDi_12 Bus Power line remains active at all times, the Interface shuts down power to the probes during sleep. Command Mode the SDI-12 logger wakes the Interface box by issuing a Break command • the data line rests in the mark state (typically 0 volts) • a break is signalled by setting the data line in the space state (3.5 to 5.5V) for at least 12 mSec upon receipt of the Break, the Interface will execute any SDI-12 commands issued from the logging device or computer in response to a Measure Command, the probe responds with the number of values it will return and the time after which they will be ready. The logging device / telemetry unit then queries the sensor again to collect the measurement values. 7.2 SDI-12 Info Command Use the “I” command to query the sensor and return the sensor information – model, serial number etc. Example: From a telemetry unit which supports Transparent SDI-12 commands: Command 0I! Response 012EP100DL041120529185012 29185012 Serial No 205 Version 04011 Model EP100DL Vendor 12 SDI-12 version 0 SDI-12 Address 15 | P a g e 7.3 SDI-12 Measurement To make an SDI-12 Measurement you should always use the C (Concurrent Measure) commands rather than the Measure (M) commands. Only use M if your logging device does not support C. When a measure command is issued, the probe will make a measurement of all available parameters and retain them in its memory for 12 seconds. If a subsequent read is issued within this time, the values will be returned from memory rather than making a new measurement. If a command is received after 12 seconds, a fresh measurement will be made. C! C1! C2! C3! C4! C5! Returns salinity compensated soil moisture Note: the “0” is omitted rather than sending 0C0! Returns soil salinity Returns soil temperature (degrees Celsius) Returns un-compensated soil moisture Returns raw salinity Note: the manufacturer reserves the right to alter this command in future versions Returns soil temperature (degrees Fahrenheit) The results are returned using the D<n>! command: Under the SDI-12 protocol, after receiving a C or M commands, the sensor must return the number of values which will be sent. When a D command is issued, the SDI-12 logger checks to see if all of the required values were received. If not, it continues to execute D commands (incrementing the suffix each time) until all values are received. In response to an M command, the EnviroPro returns 4 values with each D command • A 40cm probe will thus return all of its values with D0 • an 80cm probe will require D0 and D1 • a 120cm probe will require D0, D1, D2 (See notes below on limitation on number of readings) • a 160cm probe will require D0, D1, D2 (See notes below on limitation on number of readings) the C (Concurrent Measure) command allows more values to be returned using a single D command hence reducing the time taken to return results which in turn reduces total power consumption. 16 | P a g e M Command with D0 to Dn: Example from a telemetry unit which supports Transparent SDI-12 commands, for a sensor on Address 0, using the Measure (M) command: Command 0M! Response 00018 8 measurements in 1 sec Command Response Command Response 0D0! 0+23.45+23.30+25.49+25.3 0D1! 0+0.001+0.023+0.032+0.042 C Command with D0: Example from a telemetry unit which supports Transparent SDI-12 commands, for a sensor on Address 0, using the Concurrent Measure (C) command: Command 0C! Response 000108 8 measurements in 1 sec Command 0D0! Response 0+008.67+000.85+000.77+000.75+001.43+000.55+000.17+000.00 7.3.1 Limitations of M Command The M Command has a number of limitations: The M command only has one character value with which to return the number of readings being given. • the response to an M is <address><nnn sec><x readings> • the response to a C is <address><nnn sec.<xx readings> Hence only 9 measurements can be returned from a sensor. This means that on a 120 or 160cm probe, you will not be able to retrieve all of the sensor values • The 120cm will return the 10, 20, 30, 40, 50, 60, 80, 100 and 120cm values • The 160cm probe will return the 10, 20, 40, 60, 80, 100, 120, 140 and 160cm values In response to an M command, the D command will only return 4 values at a time, increasing the time taken to get values from the longer probes. 17 | P a g e 7.4 To Change a Probe Address The address of a probe may be changed with the a / A command. Example: From a telemetry unit which supports Transparent SDI-12 commands, change probe address from “0” to “s”: Command 0As Response s 18 | P a g e 8 Installation Installation should only be completed once the probes have been tested and the probe address set. The probe is installed into an over-sized hole and must be set into a slurry mix. See section 8.6 for a discussion on the impact of slurry on narrow versus large format probes. Installation augers are available through EnviroPro Dielectrics or your local Distributor. EP-AUGER Installation auger 36mm EP-INST-EXT10 Extension bar 1m EP-INST-EXT05 Extension bar 0.5m Note: the EP-AUGER kit comprises a 36m gouge auger, T-Handle, 1.0m extension bar and auger cleaning tool. Additional extension bars may be required for probes longer than 80cm. The slurry may be made up of soil removed from the site or using a bentonite-sand mix. 8.1 Soil Slurry In most cases, the slurry may be made up using soil removed from the hole. Preparation: Pass the soil through a sieve to remove any rocks and organic material Mix the sieved soil with water to form a creamy paste. 8.2 Bentonite Slurry The bentonite slurry can be used is the soil on site is unsuitable. In this case, the following items are required for installation of the sensor 1kg of fine sand (e.g. Unimin AFS85 grade) 100g Bentonite (civil grade e.g. Unimin Active Gel 150) 1L water Small bucket with lid 19 | P a g e Funnel 2 L plastic bottle with lid Note: The volume of the mixture is about 1.4L, the weight is 2.1kg Preparation Mix the dry bentonite and sand together in a bucket Pour the dry mix through the funnel into the 2 L plastic bottle Pour in the water, close the lid and shake until thoroughly mixed Continue to shake the mix at regular intervals until it is used. 8.3 Installation After identifying a site for the sensor, drill a hole down to 10cm greater than the probe length (e.g. 95cm for an 80cm probe) Pre-wet the soil in the hole by pouring in enough water to fill it to close to the surface. Wait until the water has drained in to the surrounding soil before proceeding. If you do not pre-wet the hole, the soil may draw the water from the slurry too quickly and you will not be able to insert the probe. If this happens, you will have to remove the probe, clear the slurry out with the installation auger and repeat the process (Pre-wetting is usually not required with the bentonite slurry Pour approximately 7000 mL of the mixture into the hole (for an 80cm probe) Push the sensor down into the hole: as the sensor moves down, it will displace the slurry from the bottom • The slurry mix will fill any air gaps around the sensor and provide a consistent contact with the soil • For the probe to measure the top 10cm of soil, the head of the probe should sit 2.5cm below the surface Remove any surplus slurry mix from above ground level If probe does not push down to required depth on first attempt (15kg maximum pressure allowed on top of probe), pull out the probe, clean the hole with the auger and repeat the process. 8.4 Cable Considerations If you are going to bury the EnviroPro cable – for example in turf or broadacre crops – you need to be conscious of the risk of damage to the probe if either the cable is snagged by machinery or stretched due to compaction of the ground. 20 | P a g e The simplest way to avoid damage is to leave a goose neck at each end of the cable. If the cable is pulled, the loop can tighten, preventing the cable form being stressed at the junction with the probe body. 8.4.1 At the Probe end As you install the probe, orient the probe body so that the cable entry point is pointing in the opposite direction to your trench. 8.4.2 At the logger/telemetry end Where the cable leaves the ground, you should consider placing it in a protective sheath to prevent damage form vermin or machinery. The sheath can be rigid conduit, flexible conduit or blind poly tube. Leave a goose neck in the cable prior to routing the cable end to the telemetry unit or data logger. 21 | P a g e Once you have made the goose neck, strap the cable to the post to which the telemetry unit or data logger is attached using zip ties. 8.5 Probe Removal The EnviroPro probes may be removed using a Probe Removal Tool. If one is not available, Vice Grips and insertion foam may be used. 8.5.1 Probe Removal Tool An EnviroPro Probe Removal Tool is available to order through EnviroPro Dielectrics and your local Distributor. Part Number EP-REMOV Removal Process: Using a hand trowel or small spade, carefully remove the soil around the head of the probe down to a depth of 15cm, being extra careful not to damage the probe cable Place the probe removal tool around the probe 10mm below the cable Tighten the wing nut firmly by hand 22 | P a g e Grip the two handles of the probe removal tool and rock it clockwise and anticlockwise After several attempts you should be able to gain 1 to 2mm of movement You should then be able to lift the probe vertically from the soil Note: If the soil is very dry, you may wish to pre-wet the soil 24 hours prior to attempting the removal 23 | P a g e Ensure that you lift the probe vertically as any off vertical force may damage the probe and such damage is not covered by the manufacture warranty. 8.5.2 Removal with Vice Grips If you do not have a Probe Removal Tool, you can use a pair of locking pliers (Vice Grips). To avoid damaging the probe body you must first wrap the unit a strip of insertion rubber. Removal Process: Using a hand trowel or small spade, carefully remove the soil around the head of the probe down to a depth of 15cm, being extra careful not to damage the probe cable Place the insertion rubber around the probe just below the cable. It may be easier if you fix it in place with insulation tape Lock the Vice Grips firmly on to the insertion rubber Grasp the Vice Grips firmly with two hands and carefully rock clockwise and anticlockwise After several attempts you should be able to gain 1 to 2mm of movement You should then be able to lift the probe vertically from the soil 24 | P a g e 8.6 Slurry and Probe Installation 8.6.1 Large format (50mm) probes When the large format (50mm) capacitance probes were introduced to the market, the manufacturers adopted a design wherein the sensors would be hung on a central column (the “Probe column” or “Probe rod”), which was dropped into a PVC tube (the “Access Tube”). The PVC tube would be sealed at the bottom and have a cap fitted at the top. As the sensors work by sending an electric field into the soil, any discontinuities (such as air gaps) have a large impact on the sensor performance. Installation techniques had to be developed which reduced this risk. To install the tubes in the ground, a slightly under-size hole would be augered into the ground and a metal cutting edge would be fitted to the end of the tube. The tube would then be hammered into the soil. As the tube was driven down, the cutting edge would shave the hole to the correct dimension, theoretically ensuring a tight fit of access tube to the soil. Two different sized augers were designed: a sand auger, which could be inserted inside the access tube allowing the hole to be augered “through the tube” and a larger clay auger, which would not fit inside the tube. After the hole was drilled and the access tube hammered in, the inside of the tube would, be cleaned out and an expanding bottom stopper installed. The direct install technique has a number of pitfalls: Any gravel or organic matter in the hole can bind against the edge of the access tube as it is driven in, creating air gaps The method can be very slow in heavy soils (average rates of drilling are 1m per hour in heavy soils) If a stabilization tripod is not used or if the tripod is not perfectly stable, the top of the hole elongates, creating air gaps around the neck of the probe Over time the expandable bottom stoppers degrade and water starts to leak into the access tube (a large contributor to sensor failure) In crops such as cotton, it is impractical to install large numbers of probes in the time window available using this method. To get around these problems, a “Slurry” installation technique was developed. Rather than an under-size hole, an over-size hole would be 25 | P a g e drilled in the soil. The end of the access tube would be closed permanently with a glued on stopper. A slurry mix would be poured into the hole and the access tube pushed into the slurry. Criticism of slurry installations tend to centre on the impact of the slurry on the sensor performance. A quick look at the dimensions of the large format probes explains why. The sensor design was initially based around the dimensions of 50mm “DWV” PVC pipe, but later, custom pipe was extruded in order to obtain tighter control over wall thickness and smoothness of the pipe interior. Tube Dimensions 50 DWV Nominal size (mm) 50 Outside Dimension (mm) 56 Auger Dimensions Supplier Sand Auger C-Probe Sentek C-Probe Sentek C-Probe Sentek Clay Auger Slurry Auger Wall thickness (mm) 2.2 Outside Dimension (mm) 53 47 56 53, 56 67 61 The diagram below shows a representation of a large format sensor installed using a slurry technique. 26 | P a g e There are two important factors to note: assuming a perfectly augered hole (which is rarely the case) there is a minimum of 2.5m of slurry on either side of the access tube (if a C-Probe slurry auger is used this can be up to 6mm) the undisturbed soil (which after all is what we are trying to measure is) is 5 to 6mm distant from the sensor ring. Why are these numbers so important? Because the electric field which radiates from the sensor decays in proportion to the distance from the sensor ring: 90 % of the sensor’s measurement capacity comes from the first 10mm out from the sensor ring. If 6mm of this is taken up by air, plastic and slurry, then the proportion of the reading taken in the undisturbed soil is indeed very small. This is why slurry installation of large format probes has always been a bone of contention. 8.6.2 EnviroPro and Slurry The EnviroPro is a “slim-line” probe, with an outer diameter of 32mm. The installation auger is 36mm and there is no air gap between the sensor plate and the probe tube. The probe tube is also only 1.5mm in thickness. The net impact of all this is that not only is the slurry much thinner, but far more of the electric field is in the undisturbed soil than is the case with a large format probes. 27 | P a g e 9 Connection to Data Loggers / Telemetry Systems The EnviroPro sensor may be used with any equipment which supports the SDI-12 interface. If the loggers / telemetry system in use supports the creation of standard drivers for the sensors to which they are connected, use the table below to configure the sensors: Parameter Soil Moisture Soil Salinity Sensor Sensor type Engineering Unit SM xxxcm SDI Soil volume percentage Address Measurement Type Measurement Number Index 0 Concurrent (C) EC xxxcm SDI Conductivity (DeciSiemens/m) 0 Soil Temperature ST xxcm SDI Temperature (Celsius) 0 0 1 2 0 to 15 0 to 15 0 to 15 NOTE: 1. The default probe address is 0 but this can be changed to any valid SDI-12 address (0 to 9, a to z, A to Z) 2. xxxcm = sensor depth in cm 3. the standard salinity units are S/m. The EnviroPro reports dS/m where 1 S/m = 10 dS/m = 10mS/cm 28 | P a g e 10 Sensor Calibration The EnviroPro soil moisture sensors are pre-loaded with a calibration based on the unit’s performance in a standardized calibration media. This media is a fine sand. The values returned in other soils will differ in accordance with the average particle size distribution in the soil which is in turn dependent on the percentages of sand, loam and clay in the soil (ignoring other factors such as organic matter and rock fraction which will alter the net holding capacity). The calibrations take the following form: y = a * ln(x) - b The Coefficients for a range of generic soils are as follows: Soil Type Sand Loam Clay Default a 8 10.6 13.4 12.9 b 7 9.5 13 13.6 10.1 Simple Calibration A simple calibration can be completed by referencing the probe readings against a known accurate point sensor such as a HydraProbe or Theta Probe. This should be completed at 2 or 3 different moisture levels – the wider apart the moisture is at each, the more representative will be the calibration. Process: Install the probes and allow sufficient time for the probe readings to equilibrate to those of the surrounding soil Using a 15 or 20cm post-hole digger, begin digging a separate hole about 1m away from the probe site Starting from a depth of 10cm and then repeating at every change in soil texture: • stab the point sensor into the soil and take a measurement • Repeat the measurement at 2 or 3 more positions around the hole and then take the average of the readings (reject any outliers) • Record the date and time of the reading 29 | P a g e Continue down through the soil profile Return to the site and repeat the process in new sampling holes at 2 more points in the moisture range (i.e. close to field capacity, mid range, close to refill point Plot the Point Sensor readings in a spreadsheet against the EnviroPro readings for the same time Using the Solver function of the spreadsheet, calculate the coefficients to relate the two. You can perform a line of best fit by using the coefficients in the table above as a starting point. 10.2 Bucket Calibration A more complete calibration can be obtained by taking samples of the soil in 20 litre containers. Take the soil samples using a post hole digger, ensuring you sample at 10cm and then at each change in the soil texture. Procedure: Oven dry the 20 litre soil samples Using a calibrated water beaker, fill your containers with 10 litres of water and mark the inside of the container at this level. Empty the containers Drill drain 4 holes in the bottom of the containers and fit them with blanking grommets Mount the EnviroPro probe in the centre of the bucket Fill the containers to the 10 litre mark with soil – the bottom two sensors should be in the soil. Gently compress the soil to compact it to the original bulk density Weigh the full containers to get dry bulk density and starting mass Remove the blanking grommets from the bottom of the containers Add 100ml of water to the container and allow it to equilibrate through the soil Record the Soil Moisture content indicated by the EnviroPro Add another 100ml of water and repeat the measurement Continue until water first emerges from the bottom of the container. Record the EnviroPro reading – this will be a good estimate of Field Capacity Continue adding water until water pools on the surface – this will give an indication of Saturation. Record the EnviroPro reading Now plot the water contents and EnviroPro readings on a spreadsheet. Use the Solver function to determine the relationship between the two. You can perform a line of best fit by using the coefficients in the table above as a starting point. 30 | P a g e 11 Regulatory Compliance Declaration The EnviroPro Soil Moisture Probe has been tested for Electromagnetic Compatibility (EMC) and Radio Frequency Interference (RFI). Test Authority Date of Testing Test Standard Manufacturer Model Number Type of Equipment EMC Technologies Pty Ltd 176 Harrick Rd KEILOR PARK VIC 3042 AUSTRALIA th 4 May 2010 EN61326-1: 2006 (IEC61326-1: 2005 Ed 1) Electrical requirements for measurement, control and laboratory use – EMC requirements. Part 1 : General Requirements APCOS Pty Ltd PO Box 380 MOUNT COMPASS SA 5210 AUSTRALIA EP100D EnviroPro dielectric soil probe This is to certify that the EnviroPro dielectric soil moisture probe, manufactured by EnviroPro Dielectrics Pty Ltd of South Australia, meets or exceeds the standards for CE compliance as per the Council Directives noted above.
© Copyright 2025