ECG Training Edan SE-601 Cardioscan Pty. Ltd.

Cardioscan Pty. Ltd.
ECG Training
Edan SE-601
Cardioscan Pty. Ltd.
Company Information
 Cardioscan is the largest provider of cardiac reporting offering services
to Pathology companies, General Practitioners and Private Hospitals in
Australia and has been running since 1984.
 Backed by a team of dedicated Cardiologists and Technicians.
The mission:
 To deliver superior cardiac reporting services by providing excellent
reporting turnaround times, 2 hours for urgent turnaround and 4 hours
for normal ECG’s, 24 hour turnaround service for Holter and NBP’s.
 Our aim is for our customers to be able to transfer data electronically
and receive these reports back in a timely manner with the ability to
produce statistical reports reducing duplication and excessive
paperwork.
Overview of Equipment
Picture
•
•
•
•
•
Edan 601 Monitor
12 Lead Set
Trace Paper
Telstra WiFi Device
Netgear Device
TRANSMITTING DEVICES
• Netgear leads
• Netgear (modem)
• Telstra Wi Fi device
Cardioscan Pty. Ltd.
ELECTRODES AND ALIGATOR CLIPS
12 Lead set – coloured coded
Aligator clips with the correct fitting of
electrodes
Initial Set Up
 Ensure the Edan ECG machine is placed on a stable surface or
trolley
 Attach the Blue Alligator clips to the 12 lead set and plug cord into
the unit
 Insert the 2 NetGear cables into the NetGear device then connect
same cables to the right side of the Edan ECG unit
Netgear
cables
12 lead cable connection
Cardioscan Pty. Ltd.
ECG TRACING
STEP ONE
 Check Telstra WiFi device is on, turn on
if required by pressing and holding
power button.
 Unit should remain on at all times, do
not turn off.
 Check a number 1 is beside signal tower
at top centre of screen.
 Check signal strength indicated by
number of bars in top left corner.
Cardioscan Pty. Ltd.
STEP TWO
Press and hold power button until
“booting” appears on screen
.
Once the ECG machine is turned on check
that the following 3 steps occur.
• Power Light will be on
• LAN light will be on
• WLAN light will be amber and change to
green once Telstra device is connected.
STEP THREE
PLACEMENT OF LEADS
Lead Placement
Limb Leads
Apply limb ECG electrodes to the soft tissue of limbs or
outer torso
LA: black – Left Arm
LL: red – Left Leg
RA: white – Right Arm
RL: green – Right Leg
Chest Leads
V1: red
V2: yellow
V3: blue
V4: green
V5: orange
V6: purple
– 4th Intercostal space, right sternal border.
– 4th Intercostal space, left sternal border.
– Midway between V2 and V4.
– 5th Intercostal space, midclavicular line.
– Anterior axillary line, horizontal to V4
– Mid axillary line, horizontal to V4 and V5.
Cardioscan Pty. Ltd.
STEP FOUR
PATIENT DATA - PRESS F1
N.B: Lab Number must commence with
Pathology site prefix
PATIENT ID TO
COMMENCE WITH
“PREFIX”
Patient Details

To enter patient details, Tab across to each field.

If you tab past a field, press shift and tab together to
return to field required


When entering DOB use only one digit
For example: 1st March 1978 = 1.3.1978
DO NOT USE ZERO other than in the year field

If the ECG Trace is classed as URGENT, add # after the lab
number.
For example LV13765421#
Press shift and 3 down together to produce hash symbol

Once all fields have been completed, tab to OK then press
enter button.

This will return screen to ECG Tracing
TAB TO OK THEN
PRESS “ENTER”
Cardioscan Pty. Ltd.
STEP FIVE
RECORDING ECG
•
•
•
•
Check patient details are correct
Ensure all leads are connected to
patient
Check to ensure Netgear and
Telstra devices are connected
To commence recording of ECG
press stop / start button
Cardioscan Pty. Ltd.
STEP SIX - CHECK
•
At top centre of screen under auto the
following will appear;
• Analysing (ECG trace sent to Patient File)
• Printing (if programmed)
• ECG Printout will occur
automatically
• Transmitting
• The word Trans is visible until
transmission is complete.
• If transmitting NOT successful, ECG
is stored, accessed and manually
transmitted via F5, Patient Files (see
Access Patient Files)
Cardioscan Pty. Ltd.
* On completion of this final step the ECG will be automatically
sent to Cardioscan in Melbourne for analysing and reporting
* An ECG recording will be automatically saved in patient file.
To Access Patient Files
Use the blue menu strip at the bottom of the screen as a guide.
•
•
•
•
Press F5 button (>>) and then press F2 button (File)
Use F1/F2 buttons to scroll up and down to required patient.
Once patient has been selected press F4 Button (Select)
Blue Menu Strip at the bottom of the screen will change to allow the
following functions:
• F1 = Edit – This allows you to change the patient information
demographics
• F2 = Trans – This allows you to send the ECG for reporting if T
does not appear in the Column called STATE
• F3 = Export – This allows data to sent to external device
• F4 = Delete – This allows selected data to be deleted from
recorder
• F5 = Preview – This allows you to view ECG recording and if
required print new copy of tracing by pressing F5 again
TROUBLE SHOOTING
 Chest leads
 poor contact or alligator
clip is attached incorrectly
to dot
Correct
Incorrect
 Lead indicators are
highlighted in yellow if
incorrect
Troubleshooting
http://incenter.medical.philips.com
HINTS FOR DOING THE ECG
Prior





Ensure all equipment on ECG trolley: ECG electrodes, Chlor-Hex swabs, clippers and blades; check paper level in
machine
Explain ECG procedure to patient
Wash hands
Patient Prep - Electrodes need good skin contact for good ECG tracing
1 Position patient
2 Expose chest area
3 For males shave chest as required
Enter patient data into machine – all fields MUST be filled (ref p2)
Important




Be aware of any cardiac symptoms at the time of recording, such as chest pain, palpitations or shortness of breath
Symptoms must be hand written on the printout and reported to nurse in charge
Check Telstra Device connection – can be manually transmitted later
Ensure good quality ECG trace before printing/sending.

Remove electrodes: re-stick electrodes on plastic film from packet, place patient label on back of film and store in
patient folder
Cover the patient and return them to a position of comfort
Wash Hands
Return ECG trolley to storage. Plug In yellow power cord.
Wipe leads with detergent wipes, replace equipment as required and leave trolley clean




Additional Information
The small space between the clavicle and the 1st rib is NOT the 1st ICS
The center of the active surface of the electrode should be aligned with the
relevant anatomical landmark
When recording and ECG from female patients, it is convention that V4 – V6 are
placed under the breast. If correct anatomical positions are not possible,
then position electrode in best practical position.
Check for an inverted or negative QRS in lead aVR, as this indicates arm lead
reversal
Remember: “white is right”
There are many rhymes to remember the limb leads
– ask your senior nurses
Using out of date / old electrodes may degrade the ECG signal quality.