This publication is not intended for distribution in the USA. CLINICAL SUMMARY A Prospective, Randomized Study of Cross-Linked and Non-Cross-Linked Polyethylene for Total Hip Arthroplasty at 10-Year Follow-Up Engh CA Jr., Hopper RH Jr., Huynh C, Ho H, Sritulanondha S, Engh CA Sr. A, The Journal of Arthroplasty. 2012; 27: (8 Suppl 1) S2-7.10 230 hips randomized to receive either 5.0 Mrad crosslinked (MARATHON™) or non-cross-linked polyethylene liners with a DURALOC® 100 Series cup. The wear rate for cross-linked liners was 0.04 ± 0.06 mm/ yr, this is an 82% reduction in comparison to the reported wear rate for non-cross-linked liners at 0.22 ± 0.13 mm/y. The osteolysis threshold is generally considered to be 0.10 mm/yr. None of the hips with cross-linked liners had a lesion that was considered clinically important (at least 1.5cm2). In comparison 22% off the un-revised hips with noncross-linked liners had clinically important lesions. In the cross-linked group (116 MARATHON liners) the 10 year survivorship rate for wear related complications with reoperation as an endpoint was 100%. Polyethylene Wear and Osteolysis After Cementless Total Hip Arthroplasty with Alumina-onHighly Cross-Linked Polyethylene Bearings in Patients Younger Than Thirty Years of Age Kim YH, Park JW, Patel CP, Kim DY. The Journal of Bone and Joint Surgery Inc. 2013; 95: 1088-1093.12 All patients were under the age of 30 and received a cementless total hip replacement with a DURALOC cup, 28 mm alumina forte ceramic femoral head and a MARATHON liner – the majority of patients in the series continued to partake in high demand activities following their total hip replacement. The mean annual liner penetration rate was 0.031 ± 0.004 mm per year. At the mean follow up of 10.8 years there were two hips with a mean annual liner penetration rate over the osteolysis threshold of 0.10 mm/year; however there was no evidence of femoral or acetabular osteolysis in these two hips. There were no polyethylene liner fractures. At a mean follow up of 10.8 years there was no evidence of femoral or acetabular osteolysis observed. Kaplan-Meier survivorship analysis using revision as an end point showed 100% survivorship of the acetabular and femoral component at a mean 10.8 year follow up (95% CI, 98 to 100). Over the last 15 years, the PINNACLE® Acetabular Cup System has grown to become one of the most widely used and clinically successful modular acetabular cup systems for hip replacement. Since its first implantation, the PINNACLE Acetabular Cup System has been provided for more than 1.5 million patients1. In 2014 the PINNACLE Acetabular Cup System was awarded a 10A ODEP rating by the Orthopaedic Data Evaluation Panel.2 Australian Orthopaedic Association National Joint Replacement Registry Data for the PINNACLE Acetabular Cup System According to the 2014 AOANJRR (Table HT13 and HT14 shown below) annual report, the PINNACLE Acetabular Cup System displayed better 10-year cumulative percent revision results than the 5.4% cumulative percent revision reported for the cementless total hip replacement category in patients with osteoarthritis excluding metal/metal prostheses with head size larger than 32 mm.3,4 Acetabular Component Femoral Component Implantations Year 5 Year 7 Year 10 CORAIL 22250 3.1 (2.8,3.4) 3.6 (3.2,4.1) 4.8 (3.9,5.9) SUMMIT® 3244 1.6 (1.2,2.2) 2.2 (1.6,3.1) 2.2 (1.6,3.1) S-ROM® 2582 3.5 (2.8,4.4) 4.2 (3.3,5.2) 4.4 (3.5,5.5) C-STEM AMT 1147 1.8 (0.9,3.4) NA NA ® PINNACLE ™ PINNACLE Acetabular Cup System – Cumulative Percent Revision Estimates of All Bearing Combinations with CORAIL, SUMMIT, S-ROM and C-STEM AMT (2014 AOANJRR, Table HT13 and HT14).3,5 National Joint Registry for England, Wales and Northern Ireland Registry Data for the PINNACLE Acetabular Cup System Based on joint replacement registry data from 2014, in 2013 the PINNACLE Acetabular Cup System was the most implanted cementless cup in the United Kingdom.6 Combination CORAIL / PINNACLE Implantations Year 1 Year 3 Year 5 Year 7 Year 10 MoP 30,061 0.85 (0.75-0.96) 1.44 (1.30-1.61) 1.80 (1.61-2.01) 2.33 (2.04-2.66) 2.58 (2.19-3.04)* MoM 11,871 0.85 (0.70-1.03) 2.39 (2.13-2.69) 5.14 (4.73-5.58) 8.82 (8.13-9.57) - CoP 8,483 0.60 (0.45-0.80) 1.14 (0.90-1.45) 1.74 (1.34-2.25) 1.84 (1.41-2.40)* - CoC 27,920 0.85 (0.75-0.97) 1.86 (1.69-2.06) 2.57 (2.33-2.84) 3.24 (2.86-3.66) - CoM 1,774 0.45 (0.23-0.90) 2.67 (1.99-3.58) 4.57 (3.37-6.19) - - Kaplan-Meier estimates of the Cumulative Revision Rate Estimates by Bearing Articulation for CORAIL and PINNACLE at 1, 3, 5, 7 and 10 years after the primary operation (2014 NJR, Table 3.10), (95% CI)7 Combination C-STEM AMT Cemented Stem / PINNACLE Implantations Year 1 Year 3 Year 5 Year 7 Year 10 2,984 0.91 (0.62-1.34) 1.34 (0.95-1.89) 2.11 (1.47-3.03) 2.11 (1.47-3.03)* - Kaplan-Meier estimates of the Cumulative Revision Rate Estimates for C-STEM AMT and PINNACLE at 1, 3, 5, 7 and 10 years after the primary operation (2014 NJR, Table 3.9), (95% CI).8 * blue italics signify that fewer than 250 cases remained at risk at these time points Fixation and Wear With a Contemporary Acetabular Component and Cross-Linked Polyethylene at Minimum 10-Year Follow-Up Bedard N, Callaghan J, Stefl M, Williams T, Liu S, Goetz D. Fixation and Wear with Contemporary, The Journal of Arthroplasty. 2014; 29: 1961-1969.9 The PINNACLE Sector cup (3 holes) was used in 149 hips and the PINNACLE multi-hole cup was used in 1 hip | Supplementary fixation of two or three screws was used in all bar one case in which no screws were used | MARATHON liners were used in all cases. Average linear wear rate was 0.05 mm/year and average volumetric wear rate was 16.23 mm3 /year. The linear steady state wear rate was 0.042 mm/year. No cases of acetabular osteolysis reported and only one case of femoral osteolysis was reported. At 10 years follow up, survivorship with an end point of reoperation for any reason was 99.2 ± 2.9%. No PINNACLE porous coated acetabular shells were revised at 10 Year follow up. “Our results demonstrated that the cementless acetabular component utilized in this cohort had excellent durability at a minimum 10-year follow-up” A Comparison of Second and Third-Generation Modular Cup Design, Is New Improved? Powers CC, Ho H, Beykirch SE, Huynh C, Hopper RH, Engh CA Jr., Engh CA. The Journal of Arthroplasty. 2010; 25: 514-521.11 This study compared the clinical outcomes of a matched series of 42 DURALOC and 42 PINNACLE total hip replacements with a mean follow-up of 5.9 years. The median Harris Hip Score was excellent for both groups, 97 (57-100) for DURALOC and 98 (53-100) for PINNACLE. The mean wear rate for DURALOC was 0.04 ± 0.08 mm/y and 0.03 ± 0.09 mm/y for PINNACLE. (P=.81) The mean volumetric wear based on the 2-dimensional head penetration was 160 ± 98 mm3 for DURALOC and 185 ±132 mm3 for PINNACLE. (P=.33) The only difference in outcome between the two groups was the incidence of pelvic osteolysis seen on CT (P = 0.005, 2-tailed Fisher exact). No osteolysis was observed in the PINNACLE group. In the DURALOC group there were 9 osteolytic lesions identified in 8 hips. The authors conclude that the most likely reason for the reduced incidence of retroacetabular osteolysis is due to differences in the locking mechanisms. Corail F2L ABGII MoM SPH-Blind ASR ABGII (Shell/Insert) 976 49 44 2900 614 862 2.2 (1.7, 2.8) (10.0, (25.0, (37.1, 3.1 (3.5, 7.0) (4.5, 8.4) (5.1, 9.2) 1.5 (2.0, (0.9, 4.8) 2.6) 11.14.9 2.4 (1.5,12.4) 3.7) 26.66.1 3.1 (2.1,28.3) 4.5) 39.26.8 4.2 (3.0,41.3) 5.9) Corail H-Max ABGII DeltaMotion Delta TridentPF(Shell) 127 136 587 598 2313 0.8 (0.3, 2.1) 2.1 2.4 (1.1, (1.9, 3.7) 3.1) 1.5 4.1 (0.6, (3.3, 3.3) 5.0) 5.0 (4.2, 6.0) 6.0 (5.0, 7.2) 7.9 (6.5, 9.6) Corail M/L TaperI Accolade Duraloc Continuum Trident (Shell) 57 13 331 1433 570 8746 1.4 (0.9, 2.2) 2.0 1.6 (1.1, (1.3, 3.6) 1.9) 2.2 3.1) 2.9 (1.6, (2.5, 3.2) 2.8 3.8 (2.0, (3.4, 3.8) 4.2) 3.9 4.6 (2.9, (4.1, 5.3) 5.2) 5.6 5.9 (4.1, (5.1, 7.5) 6.9) Corail M/L TaperI Accolade Pinnacle Trilogy Trident/Tritanium (Shell) 513 13 11 22250 569 648 1.6 (1.4, 1.7) 1.4 1.3 (0.7, (0.6, 2.8) 2.5) 2.4 (2.2, 2.7) 1.7 1.9 (0.9, (1.0, 3.2) 3.4) 3.1 (1.3, (2.8, 4.4) 3.4) 2.4 3.6 (1.8, (3.2, 6.0) 4.1) 3.3 4.8 (3.9, 5.9) Corail M/L Taper Kinectiv Continuum Pinnacle Alloclassic Allofit 68 36 187 966 1402 5304 2.2 (1.4, 3.3) 2.1 1.4 (1.5, (1.1, 3.1) 1.8) 3.7 (2.6, 5.1) 3.0 2.3 (2.2, (1.9, 4.2) 2.7) 6.1 3.1 (4.7, (2.6, 7.9) 3.6) 9.0 3.6(6.8, (3.1,11.7) 4.2) 4.9 (4.2, 5.8) MoM 40 133 65 1020 2780 621 2.5 (1.7, 3.6) 1.9 1.3 (1.4, (0.7, 2.5) 2.6) 3.4 (2.4, 4.7) 2.4 5.0 (1.9, (3.5, 3.0) 7.0) 3.6 (2.6, 4.1(8.8, (3.0,14.5) 5.6) 3.2 3.9) 11.3 4.0 (3.3, 4.9) 6.9 (2.5, (5.1, 5.0) 9.3) MoM Epoch Mallory-Head Alloclassic References Trilogy Mallory-Head Durom F2L Metafix Alloclassic SPH-Blind Trinity Fitmore 49 16 91 614 679 1695 3.1 (2.0, 4.8) 2.0 2.8 (1.2, (2.1, 3.4) 3.7) 4.9 (3.5, 7.0) 3.8 4.0 (2.0, (3.2, 7.3) 5.1) 6.1 5.0 (4.5, (4.0, 8.4) 6.2) 6.8 5.5 (5.1, (4.5, 9.2) 6.8) H-Max Nanos Alloclassic Delta PF Metal (Shell) R3 Trabecular 12 334 598 513 996 2.1 (1.1, 2.0) 3.7) 0.7 (0.2, 2.2 (1.4, 3.3) 1.0 3.0 (0.4, (2.0, 2.8) 4.3) 4.0 (2.8, 5.6) 4.0 (2.8, 5.6) 570 889 796 2.0 (1.1, 3.6) 1.0 0.5 (0.5, (0.2, 1.9) 1.4) 1.6 0.7 (0.9, (0.3, 2.7) 1.6) 2.2 1.0 (1.4, (0.5, 3.5) 2.2) 13 55 57 569 508 3236 1.4 (0.7, 2.8) 3.2 1.6 (1.9, (1.2, 5.1) 2.1) 1.7 (0.9, 3.2) 5.0 1.9 (3.4, (1.4, 7.3) 2.4) 2.4 (1.3, 4.4) 6.6 2.0 (4.7, (1.5, 9.2) 2.6) M/L Taper Continuum Natural Hip file, DePuy Fitmore 1. Data on Orthopaedics,29 139Inc. Alloclassic Trilogy 2.7 2.0 (1.8, (0.9, 4.0) 4.5) 7.6 7.2(5.7, (5.2,10.0) 9.9) 4.4 (4.7, (3.2, 6.9) 6.1) 5.7 9.4 (7.5, 11.8) 7.6 6.4(5.7, (5.2,10.0) 7.9) 4.2 (2.9, 6.1) M/L Taper Omnifit Anthology Trilogy Secur-Fit R3 3.3(5.9, (1.8,10.7) 6.0) 10.8 (8.2, 14.0) 8.0 Mallory-Head Polarstem Apex Mallory-Head R3 Fin II 133 36 29 2780 2099 940 1.9 (1.4, 1.7 2.4) 1.8 (1.2, (1.1, 2.5) 2.9) 2.4 (1.5, (1.9, 3.0) 2.3 (1.6, 3.4) 3.7) 3.2 3.9) 2.3 (1.5, 3.7 (2.5, (2.4, 3.4) 5.5) 4.0 5.0 (3.3, 4.9) 7.5) 5.7 (4.7, 6.9) Natural S-Rom Citation Hip Fitmore Pinnacle Trident (Shell) 29 87 40 889 2582 1147 1.0 (0.5, 2.1 2.7) 1.7 (1.6, (1.1, 1.9) 1.6 (0.9, 2.7) 3.2 2.5 (2.6, (1.7, 4.0) 3.5) 2.2 (1.4, 3.5) 3.5 3.0 (2.8, (2.2, 4.4) 4.2) 2.7 (1.8, 4.0) 4.2 3.2 (3.3, (2.3, 5.2) 4.4) 4.2 (2.9, 6.1) 4.4 3.9 (3.5, (2.8, 5.5) 5.3) Omnifit SL-Plus Corail Trident R3 ASR (Shell) 58 37 976 1245 1182 2900 1.9 (1.2, 2.8) 11.13.4 3.1 (2.3,12.4) 4.3) 26.64.0 (3.0,28.3) 5.3) 39.24.7 (3.6,41.3) 6.1) 2.0 (2.4, 4.8) (2.9, 5.6) 2.2 (1.3, (1.7, 3.0) (10.0, (25.0, (37.1, 5.3 (4.1, 6.9) Polarstem Secur-Fit Corail Femoral R3 Acetabular DeltaMotion Component Quadra-H Secur-Fit Corail Component Versafit Trident (Shell) Duraloc S-Rom Secur-Fit ABGII Plus Corail Duraloc Option Trident ABGII (Shell) Pinnacle 31 150 189 513 666 5200 2928 22250 S-Rom Summit ABGII Corail MoM MoM ABGII (Shell/Insert) ASR Pinnacle 87 353 44 68 SL-Plus Summit ABGII Epoch EPF-Plus Pinnacle Trident (Shell) Trilogy 94 50 136 40 SL-Plus Summit Accolade I F2L MoM R3 Trident (Shell) Pinnacle SPH-Blind 37 40 331 49 2. Orthopaedic Evaluation Panel. ratings. Available [Accessed 08/07/2014]. M/L Taper Kinectiv Data Continuum 2.1 (1.5, 3.1) 3.0 (2.2, 4.2) Omnifit Trident (Shell) 1.9 3.1 4.3) 4.0 (3.0, 4.7 (3.6, 5.3 (4.1, 6.9) 36 1402 58 1245 Anthology Reflection (Shell) 1.4 (1.2, (0.8, 2.8) 2.4) 1.6 (2.3, (1.0, 2.7) 1.9 from (1.2, 5.3) 3.1)URL: 2.1http//www.odep.org.uk (1.3, 6.1) 3.4) 17 ODEP 887 product 9.4 (7.5, 11.8) 3. Australian Orthopaedic Association Joint Replacement Annual Report. Adelaide: Metafix Trinity 2.0 (1.2, 3.4) 3.8 (2.0, 7.3) Quadra-H Versafit 1.9 (1.6, 3.1 3.2 16 679 140 6314 CLS Allofit 1.4 (0.8, 2.3) 2.6) 3.4 (2.6, (2.3, 3.7) 5.1) Registry. 3.8 (2.6, 3.8) 5.5) 5.2 (3.7, 7.3) 5.9 (4.2, 8.2) AOA; 2014. Available 38 National 780 Nanos R3 0.7 (0.2, 2.0) 1.0 (0.4, 2.8) S-Rom Duraloc 1.5 2.4 3.4 5.0) 4.0 5.8) 4.7 513 314 666 CLS URL: https://aoanjrr.dmac.adelaide.edu.au/annual-reports-2014 Fitmore Option 1.9 (0.8, (1.1, 2.8) 3.3) 4.0 (1.5, (2.7, 3.9) 5.9) 4.4 (2.2, (3.0, 6.4) HT13. 4.8 (2.7, (3.4,Page 6.9) 5.1 (3.5, 6.7) 7.3) 646 from Table 77.(3.3, Table Percent Revision of Primary Hip with Cementless Omnifit Secur-Fit 3.2 (1.9, 5.1) 5.0 (3.4, 7.3) 6.6 (4.7, 9.2) 8.0 10.8 (8.2, SL-Plus EPF-Plus 1.7 2.3) 2.8 3.6 (2.9, 4.5) 4.6 (3.7, 5.7) 6.6 (4.7,14.0) 9.3) 55 508 94 2256 Citation HT13: Cumulative Vitalock 0.5 (1.2, (0.2,Total 1.7) Conventional 2.2 (2.2, (1.2, 3.6) 3.8) 2.8 (1.7,Replacement 4.0(5.9, (2.6,10.7) 6.0) 6.5 9.2) 6.5 (4.7, 9.2) 32 555 Fixation MoM N 13 N 367 2099 713 587 Revised 140 6314 219 7580 57 Total 1433 1.7 2.4) 0.7 (0.3, 1.7) 0.8 (1.2, 2.1) 2.3 (1.5, 3.4) 2.1 1.5 (1.2, (0.6, 3.7) 3.3) 2.3 (1.5, 3.4) 1.9 (1.6, 2.3) 1.5 1.4 (1.3, (0.9, 1.8) 2.2) 3.1 (2.6, 3.7) 2.5 2.2 (2.2, (1.6, 2.9) 3.1) 3.2 (2.6, 3.8) 3.1 2.8 (2.7, (2.0, 3.6) 3.7 3.9 (3.2, (2.9, 4.3) 5.3) 4.0 5.6 (3.5, (4.1, 4.7) 7.5) 1.5 (0.8, 2.8) 1.2 1.8 (0.9, 2.4) 1.6 (1.4, 1.5) 1.7) 2.4 (1.5, 3.9) 1.9 3.1 (1.6, (2.5, 2.3) 3.8) (2.2, 2.7) 3.4 (2.2, 5.0) 2.4 4.1 (2.0, (3.4, 2.9) 4.9) 3.1 (2.8, 3.4) 4.0 (2.7, 2.7 3.2) 5.3 (2.2, (4.5, 5.8) 6.2) 3.6 (3.2, 4.1) 4.7 (3.3, 6.7) 3.6 6.8 (3.1, (5.8, 4.3) 7.9) 4.8 (3.9, 5.9) 2582 1118 862 966 2.1 2.7) 1.2 (0.7, 1.5 (1.6, (0.9, 2.0) 2.6) 2.2 (1.4, 3.3) 3.2 (2.6, 4.0) 3.5 (2.8, 4.4) 4.2 (3.3, 5.2) 6.5 (17.6, (30.0, 2.4 (5.2, (1.5, 8.1) 3.7) 19.86.1 3.1 (2.1,22.3) 4.5) 32.9 (3.0,36.1) 5.9) 3.7 5.1) (4.7, 7.9) 9.0 (6.8, 11.7) 4.4 7.2 (3.5, (5.2, 5.5) 9.9) 2256 3244 2313 1020 1.7 (1.2, 2.3) 1.0 2.4 (0.7, (1.9, 1.4) 3.1) 2.5 (1.7, 3.6) 2.8 (2.2, 3.6) 1.4 4.1 (1.1, (3.3, 2.0) 5.0) 3.4 (2.4, 4.7) 3.6 (2.9, 4.5) 1.6 5.0 (1.2, (4.2, 2.2) 6.0) (2.6, 5.0) 4.6 (3.7, 5.7) 2.2 6.0 (1.6, (5.0, 3.1) 7.2) 4.1 (3.0, 5.6) 6.6 (4.7, 9.3) 2.2 7.9 (1.6, (6.5, 3.1) 9.6) 4.4 (3.2, 6.1) 1182 784 8746 614 2.0 (1.3, 3.0) 1.5 1.6 (0.9, 1.9) 3.1 (2.0, 2.7) 4.8) 3.4 (2.4, 4.8) 2.2 2.9 (1.4, (2.5, 3.5) 3.2) 4.9 (3.5, 7.0) 4.0 (2.9, 5.6) 3.2 3.8 (2.2, (3.4, 4.8) 4.2) 6.1 (4.5, 8.4) 4.9 4.6 (3.5, (4.1, 6.8) 5.2) 6.8 (5.1, 9.2) 7.6 5.9(5.4, (5.1,10.8) 6.9) (5.7, 10.0) Revised Total N 265 N 150 5200 7605 65 621 13 569 1.2 (0.9, 1.5) 1.5 1.31(1.3, 2.6) 1.4 (0.7, 2.8) Yr 1.8) 1.9 (1.6, 2.3) 2.3 (2.0, 5.0 (3.5, 2.7) 7.0) 1.7 (0.9, 3.2) 3 Yrs 2.4 (2.0, 2.9) 2.7 (2.2, 3.2) 2.6 (2.3, 3.1 (2.7, 3.5) 6.9 (5.1, 3.0) 9.3) 11.3 (1.3, 4.4) 3.3 (1.8,14.5) 6.0) 5 Yrs 7(8.8, Yrs 3.6 (3.1, 4.3) 4.3 10(3.7, Yrs 4.9) 1 Yr 3 Yrs 5 Yrs 7 Yrs 10 Yrs 13 Yrs 4.7(7.4, (3.6,13.7) 6.2) 10.1 4. Australian Orthopaedic Association Joint Replacement Annual Report. Adelaide: AOA; 2014. Available Secur-Fit DeltaMotion 0.7 (0.3, 1.7) 2.1 (1.2, 3.7) Synergy 1.6 (0.9, 3.1 4.6 (3.4, 6.4) 7.5 (5.7, 9.9) Accolade I Trident/Tritanium (Shell) 1.3 (0.6, 2.7) 2.5) 1.9 (2.1, (1.0, 4.5) 3.4) Registry. 13 713 BHR 55 817 H-Max Delta PF 2.1 (1.1, 3.7) 11 National 648 12 598 N 219 N Femoral Secur-Fit Trident (Shell) 1.5 (1.3, 1.8) 2.5 (2.2, 2.9) 3.1 (2.7, 3.6) HT18. 3.7 (3.2, 4.3) 4.0 Synergy R3 1.6 2.4 (1.8, 2.9 (2.2, 3.9) Alloclassic AllofitAcetabular 1.41(1.2, 2.3 (1.9, 3.0) 2.7) (2.6, 3.6 (3.1,Page 4.2) 4.9 (4.2, 5.8) 7580 70 3161 187 5304 M/L Taper Continuum 2.0 (1.1, 3.6) 13 570 from URL: https://aoanjrr.dmac.adelaide.edu.au/annual-reports-2014 Table 84. Yr 2.2) 3 Yrs 5 Yrs 7 Yrs 10(3.5, Yrs 4.7) 13 Yrs Component Component MoM Femoral Secur-Fit Synergy Alloclassic M/L TaperPlus MoM Acetabular Trident (Shell) Reflection (Shell) Durom Trilogy 4.7 (3.6, 6.2) 5.4 13 (4.5, Yrs 6.4) Table HT18: Percent Revision of Primary Hip Replacement by Fixation (Primary Component Component Trabecular Metal Cumulative Continuum 4.9 (3.3,Total 7.1) Conventional 6.1 (4.2, 9.0) Revised 28 Total 529 MoM Summit 1.2 2.0) 6.5 (5.2, 8.1) (17.6, (30.0, Taperloc 1.1 2.3 (1.7, 3.8) Alloclassic Fitmore 2.8 (0.7, (2.1, 1.8) 3.7) 4.0 (1.6, (3.2, 3.3) 5.1) 19.82.5 5.0 (4.0,22.3) 6.2) 32.95.5 (4.5,36.1) 6.8) 6.4 (5.2, 7.9) ASR 353 1118 31 1653 91 1695 M/L Taper Kinectiv Exceed Continuum 2.1 (1.5, 3.1) 3.0 (2.2, 4.2) 36 1402 OA) Tri-Lock DeltaMotion 0.3 0.3 663 Trabecular Metal Diagnosis Continuum 4.9 (0.1, (3.3, 1.2) 7.1) 6.1 (0.1, (4.2, 1.2) 9.0) 282 529 MoM 1.0 (0.7, 3.4) 1.4) 1.4 (1.1, 2.0) 1.6 (1.2, 2.2) 2.2 (1.6, 3.1) 2.2 (1.6, 3.1) 1.8 (0.9, 4.3 2.2 (1.4, 3.3) 3.0 (2.9, (2.0, 6.5) 4.3) 7.3 4.0(5.3, (2.8,10.0) 5.6) (2.8,11.7) 5.6) 13.1 3244 50 514 33 996 1.9 2.5) 2.4 (1.9, 3.0) 3.2 (2.5, 3.9) 8.7 4.0(6.5, (3.3, 4.9) 5.7(9.8, (4.7,17.6) 6.9) 9.4 (7.5, 11.8) 133 2780 2.4 (2.0, 2.9) 3.2 (2.7, 3.8) 3.7 (3.2, 4.3) 4.1 (3.5, 4.7) 4.7 (4.1, 5.5) 4.7 (4.1, 5.5) 4234 0.3 0.3 663 N175 1.5 (0.9, 2.7) 2.2 (1.4, 3.2 (2.2, 4.8) 4.9 (3.5, 6.8) 1.7 (1.1, 2.6) 2.4 (1.7, 3.5) 2.8 4.1 4.3(5.4, (3.1,10.8) 5.9) 0.5 (0.1, (0.2, 1.2) 1.4) 0.7 (0.1, (0.3, 1.2) 1.6) 1.0 (2.0, (0.5, 4.1) 2.2) 2.0 (2.9, (0.9, 5.6) 4.5) 7.6 40 784 41 1230 796 2.0 (1.2, 3.4) 3.8 (2.0, 7.3) 1629 679 1 Yr 3 Yrs 5 Yrs 7 Yrs 109.3 Yrs(8.9, 9.8) 12.4 13 Yrs 2.2 (2.1, 2.4) 3.9 (3.7, 4.2) 5.5 (5.3, 5.8) 7.3 (6.9, 7.6) (11.4, 2068 37376 2.4 3.2 3.7 4.1 (3.5, 4.7) 4.7 (4.1, 5.5) 4.7 (4.1,13.5) 5.5) 175 4234 Total 1.6 (0.9, 2.7) 3.1 (2.1, 4.5) 4.6 (3.4, 6.4) 7.5(5.8, (5.7,11.2) 9.9) 2.4 (1.4, 4.2) 4.1 (2.6, 6.2) 6.1 (4.3, 8.7) (1.2, 2.9) 2.1) 1.9 (2.7, (1.4, 3.8) 2.4) 2.0 (3.2, (1.5, 4.3) 2.6) 8.1 55 817 0.7 (2.0, (0.2, 2.0) 1.0 (0.4, 2.8) 35 502 574 3236 513 2.2 3.9 5.5 7.3 12.4(8.1, (11.4, 13.5) 2068 37376 7489 824 19695 1.3) 2.4) 2.3) 4.2) 3.3) 5.8) 4.4) 7.6) (5.5,(8.9, 6.4)9.8) 9.9) R3 1.6 (1.2, 2.2) 2.4 (1.8, 3.0) 2.9 (2.2, 3.9) Reflection (Shell) 1.4 (2.1, (0.8, 2.4)2.1 (1.9, (1.0, 1.9 (5.3, (1.2, 3.1)4.0 (3.7, 2.1 (6.9, (1.3, 3.4) 6.0 9.3 70 3161 17 164645 8871.1 (1.0, Fitmore 1.0 (0.5, 1.9) 1.6 (3.7, (0.9, 2.7)3.0 (2.8, 2.2 (1.4, 3.5) 2.7 (1.8, 4.0) 4.2 (2.9, 6.1) 9.0 29 889 TOTAL 7489 164645 Cementless 4551500 procedures 136311 (1.6, 1.8) 1.8) 2.8) 2.7) 3.5) 3.0) (4.0, 4.3)10.7) (5.2, 5.6) 7.6)6.4) Synergy Reflection 1.5 2.3 2.6 3.1 (2.7, 3.5) 4.3 (3.7, 4.9) 7.1 (6.7, 5.4 (4.5, Apex Only combinations Fin II with(Shell) 1.8 (1.3, (1.1, 2.9)2.7 (2.6, 2.4 (2.0, (1.6, 3.7)3.4 (3.3, 3.7 (2.3, (2.4, 5.5)4.1 8.0 5.0(5.9, (3.3, 7.5) 5.4 265 7605 29have 9401.7 Omnifit Secur-Fit 3.2 (1.9, 5.1) 5.0 (3.4, 7.3) 6.6 (4.7, 9.2) 10.8 (8.2, 14.0) Note: over been listed. 55 508 Page | 77 Procedures using metal/metal prostheses 77768 with head size1.2 than have been included Hybrid (1.1, 1.3)32mm 2.0) 3.3) 2.7) 3.8) 3.5) 6.1) (4.5,(4.1, 5.0)6.9) Taperloc Exceed 1.1 1.8) 2.3 2.5 CLS Allofit 1999 1.4 (0.7, (0.8, 2.6)1.9 (1.8, 3.4 (1.6, 5.1)2.6 (2.5, 3.8 (1.7, (2.6, 5.5)3.4 (3.2, 5.2 (3.6, (3.7, 7.3) 4.8 5.3 5.9 (4.2, 8.2) 6.7 (6.1, 7.3) 31 38have 1653 780 larger Omnifit Trident (Shell) 1.9 (1.2, 2.8) 3.1 (2.3, 4.3) 4.0 (3.0, 5.3) 4.7 Data Period September –2211 31 500 December 2013 58 1245 Note: Only 1 combinations with over procedures been listed. Summit Taperloc Alloclassic Mallory-Head VerSys Tri-Lock Summit Taperloc Alloclassic Metafix Pinnacle Trabecular Metal (Shell) M2a Mallory-Head Trilogy DeltaMotion Mallory-Head Trilogy MoM N Pinnacle Fixation Trinity Trilogy MoM Revised BHR R3 MoM Recap Other (1135) VerSys Synergy Taperloc Anthology Nanos Other (1135) TOTAL Cemented Synergy Anthology Natural Hip Models with both fixed and exchangeable neck stems are reported separately. Procedures using metal/metal with size larger 32mm have included MoM TOTAL 7586prostheses 233774 Taperloc 1.8 (0.9, 3.4) 4.3 (2.9, 7.3 CLS MoM Fitmore 1.9than (1.1, 3.3) 4.0 been (2.7, 6.5) 5.9) 4.4(5.3, (3.0,10.0) 6.4) 50 514 31 head 646 Polarstem R3 1.7 (1.2, 2.4) metal/metal 2.3 (1.5, 3.4) 2.3 (1.5, 3.4) 36 2099 denotes M2a prosthesis combinations that have used large heads (>32mm) bearings Models with both fixed and exchangeable neck stems are reported separately. Taperloc 1.7 2.6) 2.4 (1.7, 3.5) 2.8 CitationMoM Trident 2.7)32mm 2.5 3.0 (2.0, (2.2, 4.1) 4.2) 41 1230 40 1147 Quadra-H Versafit 1.9 (1.1, (1.6, 2.3) 3.1 (2.6, 3.7) 3.2 (2.6, 3.8) 140 6314 denotes Mallory-Head prosthesis combinations that have used large (>32mm) metal/metal bearings Note: All procedures using(Shell) metal/metal prostheses with head sizeheads larger than have been excluded MoM Taperloc Citation S-Rom 35 32 31 Vitalock Recap Duraloc Option 502 555 666 2.4 0.5 (1.4, (0.2, 4.2) 1.7) 1.5 (0.8, 2.8) 4.1 2.2 (2.6, (1.2, 6.2) 3.8) 2.4 (1.5, 3.9) 6.1 2.8 (4.3, (1.7, 8.7) 4.5) 3.4 (2.2, 5.0) 8.7 4.8(6.5, (3.4,11.7) 6.9) 13.1 5.1(9.8, (3.5,17.6) 7.3) 4.1 3.2 (2.9, (2.3, 5.6) 4.4) 4.3 3.9 (3.1, (2.8, 5.9) 5.3) 8.1 (2.6,11.2) 6.0) 4.0(5.8, (2.7, 5.8) 6.5 (3.3, (4.7, 6.7) 9.2) 4.7 6.5 (4.7, 9.2) Corail 2.2 (1.7, 2.8) 11.13.2 (10.0, 26.63.5 (25.0, 39.24.2 (37.1, 41.3) ASR 976 2900 S-Rom Pinnacle 2.1 (1.6, 2.7) (2.6,12.4) 4.0) Registry. (2.8,28.3) 4.4) (3.3, 5.2) 4.4 (3.5, 5.5) AOA; 2014. Available 87 National 2582 5. Australian Orthopaedic Association Joint Replacement Annual Report. Adelaide: Corail DeltaMotion 0.8 (1.2, (0.3, 2.3) 2.1) 1.5 (2.2, (0.6, 3.6) 3.3) 587 SL-Plus EPF-Plus 1.7 2.8 3.6 (2.9, 4.5) 4.6 (3.7, 5.7) 6.6 (4.7, 9.3) 947 2256 Page | 77 from URL: https://aoanjrr.dmac.adelaide.edu.au/annual-reports-2014 Table HT14. Page 78. Table HT14: Cumulative Percent Revision of Primary Total Conventional Hip Replacement with Hybrid Fixation MoM CorailPeriod 1 September Duraloc1999 – 31 December 2013 57 Data SL-Plus R3 37 1433 1182 1.4 (1.3, (0.9, 3.0) 2.2) 2.0 2.2 (2.4, (1.6, 4.8) 3.1) 3.4 2.8 (2.9, (2.0, 5.6) 3.8) 4.0 3.9 (2.9, 5.3) 5.6 (4.1, 7.5) Table Percent of Hip with Hybrid Fixation Corail HT14: Pinnacle 1.6 (0.3, (1.4,Total 2.4 (1.2, (2.2, 3.7) 2.7) 3.1 (2.8,Replacement 3.4) 3.6 (3.2, 4.1) 4.8 (3.9, 5.9) 513 Secur-Fit DeltaMotion 0.7 1.7) 2.1 Figure HT11: Cumulative Cumulative PercentRevision Revision ofPrimary TotalConventional Conventional Hip Replacement by Fixation (Primary 13N 22250 713 Femoral NPrimary Acetabular 1 Yr 1.8) CPR 2.5 3 Yrs CPR3.6) 10 Yrs MoM Component OA) Corail 2.2 (1.3, (1.4, 3.3) 3.7 (2.6,CPR 5.1) 5 Yrs 6.1 (2.7, (4.7, 7.9)7 Yrs 9.0 (6.8, Pinnacle(Shell) 68 966 Secur-Fit Trident (2.2, 2.9) 3.1 3.7CPR (3.2,11.7) 4.3) 4.0CPR (3.5, 4.7)13 Yrs CPR 219 7580 Component Diagnosis Revised Total1.5 Femoral N N HR adjusted for age and gender Acetabular Component 150 1 Yr 1.5) CPR 1.9 3 Yrs 5 Yrs CPR2.9) Yrs2.7 10 Yrs CPR 2.52.4 (1.7, 3.6) 3.4 (2.4,CPR 4.7) 4.0 3.6 (2.6, 5.0)7 4.1CPR (3.0, 4.4CPR (3.2, 4.3) 6.1)13 Yrs4.7 (0.9, (1.6, 2.3) 2.4 (2.0, (2.2, 3.2) 3.6 (3.1, (3.6, 6.2) 40 70 1020 5200 Revised Total 9811.2 (1.6, 3.5) 3.1 (2.2, 4.4) (2.9, 5.5) 5.2 (3.9, 6.9) 5.6) 8.1 (6.3, 10.4) Cemented vs Hybrid 0 - 1Mth: HR=0.61 (0.46, 0.81),p<0.001 Epoch Trilogy (Shell) Secur-Fit Plus Trident 20% Component C-Stem Duraloc Cemented 49 353 SPH-Blind ASR Cementless Pinnacle Duraloc Hybrid 12 50 Delta PF Pinnacle Pinnacle MoM 16% M/L Taper Continuum Summit Pinnacle CPCS R3 C-Stem AMT Joint Pinnacle National Registry for MoM M/L Taper Trilogy Synergy BHR Reflection (Shell) CPCS R3 14% Cumulative Percent Revision 6. MoM 3.12.1 (2.0, 4.8)3.5) 3.1 4.9 (1.8, (3.5, 4.5) 7.0) 19.8 (4.5, 8.4)4.9 6.8 (5.1, 9.2) 7.6 (5.7, 10.0) 614 700 (0.7, 2.0) 6.5 (5.2, 8.1) (17.6, 22.3) 32.9 (30.0, 36.1) 1118 22 (1.2, 2.8 2.8 (1.8, 4.5) (3.0, 7.9) 1Mth - 6Mth: (0.73, 1.15),p=0.469 70 9811.2 2.4 (1.6, (2.2, 4.4) 4.0 6.1 (2.9, 5.5) 5.2 (3.9, 6.9) 8.1HR=0.92 (6.3, 10.4) 2.10.5 (1.1, 3.7)1.1) 5981147 (0.7, 1.4) 1.4 (1.8, (1.1, 2.5) 2.0) 1.8 (1.2, 2.26Mth (1.6, 3.1) HR=1.49 2.2 (1.6, 3244 - 1.5Yr: (1.22, 3.1) 1.81),p<0.001 13 (0.2, (0.7, (0.9, 3.4) 22 7001.0 2.1 (1.2, 3.5) 1.4 2.8 4.5) 2.8 1.6 (1.8, 4.5)2.2)4.9 (3.0, 7.9) 1.5Yr+: HR=1.40 (1.26, 1.54),p<0.001 2.01.7 (1.1, 3.6)2.4) 13 42 5702077 1.5 (0.9, 2.7) 2.2 (1.7, (1.4, 3.3) 3.5) 2.5 (2.2, 4.9 (3.5, 6.8) 7.6 (5.4, 10.8) 40 (1.2, 2.4 (1.8, 3.5) 13784 1147 0.5 (0.2, 1.1) Ireland 1.4 (0.7, 2.5) 1.8 3.2 (0.9, 3.4)4.8) Report England, Wales & Northern 11th Annual Prostheses Cementless vs Hybrid 0 - 2Wk:2014. HR=1.78 (1.47, 2.14),p<0.001 used in hip, knee, ankle, elbow and 1.41.7 (0.7, 2.8)2.4) 1.7 (1.7, (0.9, 3.3) 3.2) 2.5 2.4 (1.3, 4.4)2.7 (2.0, 3.3 (5.7, (1.8, 6.0)5.3 (3.7, 7.5) 13 42 5692077 (0.9, 2.7) 3.1 (2.1, 4.5) (3.4, 7.5 55 67817 26681.6 1.0 (0.6, 1.4) 2.4 1.3 (0.9, 1.8) 1.8 4.6 (1.3, 2.4)6.4) 3.6) 9.9) (1.2, (1.8, 3.5) 2Wk - 3Mth: HR=1.39 (1.25, 1.55),p<0.001 shoulder replacement procedures70 2013. Available from URL: http://www.njrcentre.org.uk/njrcentre/default.aspx. Table 4 (a). Page 8. M/L Taper Kinectiv Continuum(Shell) 2.11.1 (1.5, 3.1) 3.0 (2.2, 2.8) 4.2) 2.6 36 1402 Synergy R3 (1.2, 2.2) 2.4 (1.8, 3.0) (2.2, 3161 CPT Allofit 15 8411.6 (0.6, 2.1) 1.6 (0.9, (1.5, 4.5) 4.5) CPCS Reflection 67 2668 1.0 1.4) 1.3 1.8) 1.8 2.9 (1.3, 2.4)3.9)2.6 2.7 (1.5, (2.0, 3.6) 5.3 (3.7, (0.94, 7.5) 1.37),p=0.187 3Mth - 6Mth: HR=1.13 F2L Summit C-Stem 18% H-Max AMT Summit C-Stem 12% Mallory-Head Synergy CPT 133 265 Mallory-Head Reflection (Shell) Continuum 1.92.4 (1.4, 2.5)3.4) 3.3 2.4 (2.4, (1.9, 4.6) 3.0) (1.3, 1.8) 2.3 (2.0, 2.7) 278014161.5 7605 40 (1.7, 3.2 (2.5, 3.9) 2.6 (2.3, 3.0) 4.06Mth (3.3, 4.9)HR=1.34 5.7(1.20, (4.7,1.50),p<0.001 6.9) 9.4 3.1 (2.7, 3.5) 4.3 (3.7, 4.9) 5.4(7.5, (4.5,11.8) 6.4) - 2Yr: CPT Allofit 15 841 1.1 (0.6, 2.1) 1.6 (0.9, 2.8) 2.6 (1.5, 4.5) 2.6 (1.5, 4.5) 7. National Registry for England, &2.0Northern Ireland 11th Annual Report 2014. Prostheses used in hip, knee, ankle, elbow and shoulder 2Yr - 3Yr: Metafix10% Joint Trinity (1.2, 3.4)3.3) 3.8 (2.7, (2.0, 5.1) 7.3) 16 Wales 6791112 Taperloc Exceed (0.7, 1.8) 2.3 (1.6, 3.3) (1.7, 31 1653 CPT Trabecular Metal (Shell) 47 2.2 (1.5, 4.9 2.5 (3.6, 6.5)3.8)6.2 (4.3, 8.8) HR=1.42 (1.21, 1.66),p<0.001 Continuum 40 14161.1 2.4 (1.7, 3.4) 3.7 3.3 (2.4, 4.6) 3Yr - 4Yr: HR=0.94 (0.79, 1.13),p=0.518 Nanos R3 0.71.5 (0.2, 2.0)1.9) 1.0 (2.1, (0.4, 5.1) 2.8) 3.1 513from Taperloc 1.8 (0.9, 3.4) 4.3 (2.9, 6.5) 7.3 (5.3,3.7) 10.0)3.8 8.7 (6.5, 11.7)5.113.1 replacement procedures 2013. Available URL: http://www.njrcentre.org.uk/njrcentre/default.aspx. Table 3.10. Page 51. M2a 504 196 514 CPT Trilogy 6131 (1.3, 2.4 2.9) (2.7, (3.3, 4.4) (4.3,(9.8, 6.1)17.6) Trabecular Metal (Shell) 47 1112 2.2 (1.5, 3.3) 3.7 (2.7, 4.9 (3.6, 6.5) 6.2 (4.3, 8.8) 8% MoM 4Yr - 4.5Yr: HR=1.38 (1.04, 1.81),p=0.023 Natural Hip Taperloc Elite Plus CPT Fitmore Mallory-Head Duraloc Trilogy 1.02.0 (0.5, 1.9)3.0) 1.6 (2.7, (0.9, 5.0) 2.7) 5.4 2.2 (1.4, 3.5)7.3 2.7 (2.9, (1.8, 4.0) 4.2 (3.1, (2.9, 6.1) 29 196 8891078 (1.1, 2.6) 2.4 (1.7, 3.5) (2.0, 4.1 5.6) 4.3 5.9) 41 1230 98 (1.3, (4.2, 7.0) 9.1) 9.9 14.2 (10.2, 19.4) 61311.7 1.5 1.9) 3.6 2.4 (2.1, 2.9) 3.1 2.8 (2.7, 3.7)4.1) 3.8 (5.8, (3.3, 4.4)HR=0.93 5.1(8.0, (4.3, 6.1) 4.5Yr+: (0.84,12.0) 1.02),p=0.121 Polarstem Exeter V40 R3 Hemispherical ABGII 36 Exeter V40 R3 Pinnacle 6% Omnifit Secur-Fit 3.2 (1.9, 5.1)2.5) 5.0 (1.6, (3.4, 3.4) 7.3) 6.6 (4.7, 9.2) 8.0 (5.9, 10.7) 10.8 (8.2, 14.0) 8. National Registry for England, &2.4 Northern 11th Annual Report 2014. Prostheses used 55 Wales 5081218 Taperloc (1.4, 4.2) 4.1 (2.6, 6.2) 6.1 (4.3, 8.7) 8.1 Recap 35 vs Cemented 0(5.8, - 1Mth: HR=2.52 (1.91, 3.33),p<0.001 Exeter Vitalock 59 1.6 (1.0, 2.3 2.5 (1.8, 3.6) 3.3 (2.4, 4.5)11.2) 4.7(8.0, (3.6, 6.2) (4.4, 7.4) in hip, knee, ankle, elbow and Elite Plus Joint Duraloc 98502 1078 2.0 (1.3, 3.0) Ireland 3.6 (2.7, 5.0) 5.4Cementless (4.2, 7.0) 7.3 (5.8, 9.1) 9.9 12.0) 14.25.7 (10.2, 19.4) - 3Mth: HR=1.36 (1.10, 6.9) 1.67),p=0.003 OmnifitV40 Trident (Shell) procedures58 (1.2, 2.8) 3.1 (2.3, 2.4) 4.3) 2.1 (3.0, 4.71Mth (3.6, 6.1)3.7 5.3 (4.1, 1245 Exeter ABGII 35 1064 (0.6, 2.0) (0.9, (1.4, 3.2) 4.5) 5.2) shoulder 2013. Available from URL: http://www.njrcentre.org.uk/njrcentre/default.aspx. Table 3.9. Page 50. Vitalock 59 12181.91.1 1.6 (1.0, 2.5) 1.4 2.3 (1.6, 3.4) 2.5 4.0 (1.8, 3.6)5.3)3.2 3.3 (2.2, (2.4, 4.7 (2.6, (3.6, 6.2) 5.7 (4.4, 7.4) 4% replacement 3Mth - 1.5Yr: HR=1.01 (0.87, 1.18),p=0.888 2% MoM (1.2, 2.4)3.5) 2.3 (0.9, (1.5, 4.9) 3.4) 2.1 (1.5, 20991064 20 6491.72.0 (1.2, (2.0, 3.4 2.3 (2.2, 5.3) 5.3) 35 1.1 (0.6, 2.0) 3.2 1.4 2.4) (1.4, 3.2)3.4)3.4 3.2 (2.2, 4.5) 3.7 (2.6, 5.2) 1.5Yr - 3Yr: HR=0.99 (0.85, 1.14),p=0.836Page | 77 Quadra-H Versafit1999 –J,31Stefl 1.90.5 (1.6, 2.3)1.1) 3.1 (0.5, (2.6, 3.7) 1.0 3.2 (2.6, 3.8)1.7 140 23 6314T,1249 Data Period 1 September December Exeter V40N, Mallory-Head (0.2, 1.6) (0.6, 1.8) (1.0, 3.3 (2.1,(0.54, 5.0)0.88),p=0.003 9. Bedard Callaghan M, 2013 Williams Liu S, Goetz D. Fixation Wear Contemporary Acetabular Components Hemispherical 20 649 2.0 (1.2, 3.5) 0.9 3.2 (2.0, 4.9) 3.4and (2.2, 5.3) 3.4with (2.2, 5.3) 3Yr2.9) - 3.5Yr: HR=0.69 S-Rom Duraloc Option 1.51.1 (0.8, 2.8)2.0) 2.4 (1.2, (1.5, 3.2) 3.9) 2.2 (2.2, 4.03.5Yr (2.7, 5.8)3.3 4.7(0.64, (3.3, 6.7) 31 19 0% - 5Yr: HR=0.76 ExeterCross-Linked V40 Pinnacle (0.6, 2.0 (1.4, 3.6) 5.8) Mallory-Head 236661037 1249 0.5 (0.2, 1.1)Journal 0.9 (0.5, 1.6) 1.0 3.4 (0.6, 1.8)5.0)3.1 1.7 (1.6, (1.0, 2.9) (2.1, 5.0)0.90),p=0.001 and Polyethylene at 10-Year Follow-Up. of Arthroplasty. 2014; 29: 1961-1969. 0 1 2 3 4 5 6 7 10 11 12 13 S-Rom Pinnacle 3.2 (2.6, 4.0) 3.5 (2.8, 4.4) 4.25Yr+: (3.3,HR=0.66 5.2) (0.59, 4.40.74),p<0.001 (3.5, 5.5) 87 2582 8 2.19(1.6, 2.7) 22 1021 1.1 (0.6, 2.0) 2.3 2.0 (1.4, (1.2, 3.7) 3.2) 3.8 2.2 (2.3, (1.4, 6.3) 3.6) 3.1 (1.6, 5.8) 19Procedure 1037 Years Since Primary SL-Plus EPF-Plus 1.71.1 (1.2, 2.3)1.3) 2.8 (1.6, (2.2, 1.9) 3.6) 2.4 3.6 (2.9, 4.6 (3.7, 6.6 (4.7, 225636720 Exeter V40 (Shell)RH Jr., Huynh94C,833 (1.0, (2.2, 2.6) (2.8, 3.3) 5.7)4.4Randomized (4.0, 4.9) 9.3) R3 22 1021 (0.6, 2.0) 1.8 2.3 (1.4, 3.7) 3.8CA (2.3, 6.3)4.5) 10.Engh CA Jr., Trident Hopper Ho H, Sritulanondha S, Engh Sr. A 3.1 Prospective Study of Cross-Linked and Non-CrossSL-Plus R3 (1.3, 3.0)1.6) 3.4 (1.0, (2.4, 2.9) 4.8) 2.4 4.0 (2.9, 37 833 118236720 Exeter V40 Trident/Tritanium 20 15092.01.0 (0.6, Trident (Shell) (Shell) 1.1 (1.0, 1.3) 1.7 1.8 (1.6, 1.9) (2.2, 2.6)5.6)3.1 (2.8, 3.3) 4.4 (4.0, 4.9) Linked Polyethylene for Total Hip Arthroplasty at 10 –Year Follow Up, Journal of Arthroplasty. 2012; 27: (8 Suppl 1) S2-7 Secur-Fit DeltaMotion 0.7 (0.3, 1.7) 2.1 (1.2, 3.7) 13 713 Exeter V40 Trilogy 18 598 1.7 (0.9, 3.1) 2.5 (1.5, 4.1) 2.7 (1.6, 4.5) 3.0 (1.8, 4.8) 4.9 (2.6, 9.1) Exeter V40 Trident/Tritanium (Shell) Secur-Fit Trident (Shell) Exeter V40 at Risk Vitalock Number 0 Yr 1219 Yr 20 1509 1.0 (0.6, 1.6) 1.7 (1.0, 2.9) (1.3, 1.8)1.5) 1.7 2.5 (1.2, (2.2, 2.9) 2.3 3.1 (2.7, 3.7 (3.2, 4.0 (3.5, 758019591.5 62 2.3) (1.7, 3.1) 3.7) 4.3) (2.5, 4.1) 4.7) 13 Yrs 30.9 Yrs(0.6, 5 Yrs 73.6) Yrs2.8 (2.2, 103.2Yrs Exeter V40 Trilogy 18 598 1.7 (0.9, 3.1) 2.5 (1.5, 4.1) 2.7 (1.6, 4.5) 3.0 (1.8, 4.8) 4.9 (2.6, 9.1) 11.Powers CC, Ho H, (Shell) Beykirch SE, Huynh C, Hopper RH, Engh CA2.8) Jr.,2.4Engh CA.3.1 A Comparison Second4.7and ThirdSecur-Fit Plus Trident (0.9, 1.5)2.1) 1.9 (1.3, (1.6, 2.3) (2.0, 2.7 (2.2, 3.6 of (3.1, 150 41 5200 MS 30 V40 Allofit 1365 1.4 (0.9, (1.7, 3.4) 4.4) (2.9, 5.7) Exeter Vitalock 62 19591.215104 0.9 (0.6, 1.5) 1.9 1.7 (1.2, 2.3) 2.3 2.4 3.1)2.9) 2.8 (2.2, 3.7) 3.2)4.0 3.2 (2.5, 4.1) 4.3) Cemented 19695 18129 12030 8969 4321 410 (3.6, 6.2) Summit (0.7, 2.0)2.5) 6.5Journal (5.2, 8.1) 19.8 (17.6,4.1) 22.3) 32.9 (30.0, ASR 353 1118 Generation Modular Cup Design, Is New Improved? The2.9 of (2.6, Arthroplasty. 2010; 514-521.839 Omnifit Trident 1.8 (1.3, (2.3, 3.7) 3.2 3.5 (2.8, 4.4)36.1)14938 3.9 (3.1, MS 30 Allofit (Shell) 41 2334 13651.282931 1.4 (0.9, 2.1) 1.9 (1.3, 2.8) 2.4 (1.7, 3.4) 3.1 (2.2, 4.025: (2.9, 4.9) 5.7) Cementless 136311 116379 73 56072 37444 MoM Summit Spectron Omnifit EF MoM Pinnacle BHR Trident (Shell) Hybrid 77768 MoM Summit 12.Kim YH,EF ParkR3 JW, Pinnacle MoMPatel CP, Spectron BHR (0.7, 1.4)2.0) 1.4 (2.3, (1.1, 4.8) 2.0) 3.2 (1.2, 2.2) 2.2 12.8) (1.6, 2.2 (1.6, 50 39 32442334 5321.051070 0.8 (0.3, (1.8, 6.6 1.6 (4.7, 9.4) 9.2 73 1.8 (1.3, 2.5) 2.9 3.7) (2.6, 4.1) 3.5(6.6, (2.8, 4.4) 3.1)3.9 (3.1, 4.9) 3.1) 67968 36475 24233 9968 817 (0.9, 2.7)2.4) 2.2 (1.8, (1.4, 3.5) 2.6 (2.2, 4.9 12.8) (3.5, 6.8) Total 7.6 (5.4,Hip 10.8)Arthroplasty with Alumina-onKim DY.40Polyethylene Wear Osteolysis After Cementless 27 (1.0, 2.6 3.8) (1.8, 3.8) 397841197 5321.51.5 0.8 (0.3, 2.0)and 2.9 4.8) 6.6 3.2 (4.7, 9.4)4.8)9.2 (6.6, Synergy (0.9, 2.7)1.4) 3.1 (1.6, (2.1, 2.4) 4.5)Than (3.4, 7.5 (5.7, 9.9)6.2 (5.3, BHR Spectron EF Reflection (Shell) 214 1.1 (0.8, 2.0 2.8 (2.3, 3.3) 3.8 (3.2, 4.4)Age. 7.2)2013; 11.1 (8.7, Highly Cross-Linked Polyethylene 55 Bearings in1.6Patients Thirty of JBJS. 95:14.1) 1088-1093 R3 278174969 1197 1.5 (1.0, 2.4)Younger 2.6 (1.8, 3.8) 2.6 4.6 (1.8, 3.8)6.4)Years Note: All procedures using MoM metal/metal prostheses with head size larger than 32mm have been excluded Synergy Other (764) Spectron EF Synergy Other (764) TOTAL Taperloc R3 Reflection (Shell) Reflection (Shell) Exceed (1.2, 2.2)1.4) 70 824 316115668 1.8 (1.6, 2.0) 214 49691.61.1 (0.8, (1.3, 1.8)2.0) 265 2939 7605 824 156681.51.8 (1.6, 91240 31 1653 1.1 (0.7, 1.8) 2.4 (2.9, (1.8, 3.5) 3.0) 2.8 (2.2, 3.2 4.5 2.9 (4.1, 4.9) 9.1 (8.2, 10.0) 2.0 (1.6, 2.4) (2.3, 3.3)3.9)5.7 3.8 (5.3, (3.2, 6.1) 4.4) 7.5 6.2 (6.9, (5.3, 8.0) 7.2) 11.1 (8.7, 14.1) 2.3 (2.9, (2.0, 3.5) 2.7) 4.5 2.6 (2.3, 3.1 (2.7, 4.3 (3.7, 5.4 10.0) (4.5, 6.4) 3.2 (4.1, 4.9)3.0)5.7 (5.3, 6.1) 3.5)7.5 (6.9, 8.0) 4.9)9.1 (8.2, 2.3 (1.6, 3.3) 2.5 (1.7, 3.8) TOTALOnly combinations with over 500 procedures have 2939 91240 Note: been listed. Taperloc 1.8than (0.9,32mm 3.4) have 4.3 been (2.9, 6.5) 7.3 (5.3, 10.0) M2a MoM 50 head 514 Procedures using metal/metal prostheses with size larger included Note: MoM Only combinations with over 500 procedures have been listed. Taperloc denotes Mallory-Head 1.7 (1.1, 2.6) metal/metal 2.4 (1.7, 3.5) 2.8 (2.0, 4.1) prosthesis combinations that have used large heads (>32mm) bearings 41 1230 Procedures using metal/metal prostheses with head size larger than 32mm have been included MoM MoM Taperloc 2.4 (1.4, 4.2) metal/metal 4.1 (2.6, 6.2) 6.1 (4.3, 8.7) 35 used502 denotes Recap prosthesis combinations that have large heads (>32mm) bearings 8.7 (6.5, 11.7) 13.1 (9.8, 17.6) 4.1 (2.9, 5.6) 4.3 (3.1, 5.9) 8.1 (5.8, 11.2) All analysis was carried out by DePuy Synthes, the NJR do not vouch for the accuracy of the interpretation. Page | 77 Data Period 1 September 1999 – 31 December 2013 Johnson & Johnson Medical Limited PO BOX 1988, Simpson Parkway, Livingston, West Lothian, EH54 0AB, United Kingdom. Page | 78 Data Period 1 September 1999 – 31 December 2013 Page | 78 Incorporated and registered in Scotland under company number SC132162. Data Period 1 September 1999 – 31 December 2013 Data Period 1 September 1999 – 31 December 2013 DePuy Orthopaedics, Inc. 700 Orthopaedic Drive Warsaw, IN 46582 USA Tel: +1 (800) 366 8143 Fax: +1 (574) 267 7196 Page | 84 DePuy International Ltd St Anthony’s Road Leeds LS11 8DT England Tel: +44 (0)113 270 0461 Fax: +44 (0)113 272 4101 depuysynthes.com ©Johnson & Johnson Medical Limited. 2015. All rights reserved. CA#DSEM/JRC/1214/0218 Issued: 04/15 DePuy (Ireland) Loughbeg Ringaskiddy Co. Cork Ireland Tel: +353 21 4914 000 Fax: +353 21 4914 199 0086
© Copyright 2024