Simulation This menu allows for simulation of flow rate and

Simulation
Simulation
This menu allows for simulation of flow rate and temperature. It should only be used
for testing and demonstration purposes. Make sure to return all of these simulation
values to zero, before returning to the normal mode of operation.
Caution: If the 4 to 20mA and/or the pulse/alarm outputs are connected to controllers,
set the controllers to “manual”. This will ensure that the simulated signals do not
cause false controller action.
The menu is accessible from the main programming menu by pressing FLO, and DGN
(F1):
FLOW PARAMETER 1
DGN
UNT
PRM EXIT
F1F2F3F4
Pressing DGN (F1) will show:
DIAGNOSTIC
SIM
ZRO
EXIT
F1F2F3F4
Pressing SIM (F1) will show:
Simulate Flow?
YES
NO
F1F2F3F4
Press YES (F1) to continue.
FloSim = 0 SCFM
CHG
OK
F1F2F3F4
Enter the value and then press OK (F4).
i
Note: Enter zero to disable this feature.
Simulate Temp?
YES
NO
F1F2F3F4
Press YES (F1) to continue.
TmpSim = 0 C
CHG
OK
F1F2F3F4
Enter the value and then press OK (F4).
i
Note: Enter zero to disable this feature.
ENABLE SIM?
YES
NO
F1F2F3F4
Press YES (F1) to start the simulation mode, otherwise press NO (F4). Upon pressing
either key, the program will return to the FLOW PARAMETER 1 menu.
i
Note: Simulation Mode will be cleared if the power is cycled.