Quality Management in CT CT Prof. Stelmark CT Prof. Stelmark CT Prof. Stelmark AVG. CT NUMBER • WATER – CT# = 0 , +- 3 HU • AIR – CT# = -1,000, +- 5 HU CT Prof. Stelmark AVERAGE CT NUMBER CT Prof. Stelmark Uniformity and noise When imaged, a phantom of uniform material such as water should demonstrate consistent CT values regardless of a pixel's position on the image matrix. Pixels at the center of the image should measure as water density (∼ 0 HU), as should pixels at various other points in the image. Beam hardening occurs as the x-ray beam passes through the object (phantom, patient). The beam exposing the center of the object has a higher average photon energy than the beam exposing the periphery of the object. This beam-hardening effect may slightly alter the pixel values at different points across an image of an object of uniform inherent density. The spatial uniformity of a CT system describes its ability to maintain relatively consistent CT values across the entire image of an object of equal density. CT Prof. Stelmark 4. Uniformity may be evaluated by positioning several ROI measurements at different locations along the center and periphery of an image. The CT values should not differ by more than 2 HU from one location to another CT Prof. Stelmark CT Prof. Stelmark CAUSE OF FAILURE • NOISE CT Prof. Stelmark FREQUENCY • DAILY CT Prof. Stelmark Linearity The relative accuracy between calculated CT numbers and their respective linear attenuation coefficients is termed linearity. The CT number for water is 0 HU. According to the formula for the calculation of CT number relative to water discussed previously (see “Image Display”), the CT number for air is −1000 HU, and the CT number for water is 0 HU. The CT system is calibrated according to these two known values. Daily calibration procedures should be performed using a water-filled phantom or a gantry cleared of all objects and exposing only room air. CT Prof. Stelmark CT Prof. Stelmark CT Prof. Stelmark Linearity Analysis of the values of the five pins should show a linear relationship between the Hounsfield unit and electron density. The coefficient of correlation for this linear relationship should be at least 0.96%, or 2 standard deviations. CT Prof. Stelmark LOW CONTRAST PHANTOM CT Prof. Stelmark CT Prof. Stelmark EXPECTED RESULTS • SMALLEST HOLES THAT COULD BE IMAGED SHOULD BE 3 mm IN DIAMETER OR SMALLER FOR 0.5% densityu difference. CT Prof. Stelmark CAUSE OF FAILURE • NOISE CT Prof. Stelmark HIGH CONTRAST PHANTOM CT Prof. Stelmark CT Prof. Stelmark CT Prof. Stelmark SPATIAL RESOLUTION • 0.45 –0.15 lp/mm (4.5 –15 lp/cm) CT Prof. Stelmark CAUSE OF FAILURE • EXCESSIVE FSS • MECHANICAL WEAR • DETECTOR VIABRATIONS CT Prof. Stelmark FREQUENCY • ONCE A MONTH CT Prof. Stelmark Slice Thickness Slice thickness (sensitivity profile) is measured with the use of a specially designed test object that incorporates a ramp, a spiral, or a step wedge. This assessment should be done semiannually; the slice thickness should be within 1 mm of the intended slice thickness for a thickness of 5 mm or greater. For an intended slice thickness of less than 5 mm, the acceptable tolerance is 0.5 mm. CT Prof. Stelmark Accuracy of Distance-Measuring Device EXPECTED RESULTS: The distance indicated by the CT scanner should agree with the true distance as determined by counting the spaces between the two holes. ACCEPTANCE LIMITS: Disagreement of 1 mm or less is good. Disagreement of greater than 2 mm should be corrected. POSSIBLE CAUSES OF FAILURE: Reconstruction algorithm may be improperly calibrated. If the manufacturer has not provided the user with a means to recalibrate the algorithm, a service person should be notified. FREQUENCY: This should be performed at the time of installation and annually thereafter. CT Prof. Stelmark Couch Incrementation With automatic maneuvering of the patient through the CT gantry, the patient must be precisely positioned. This evaluation should be done monthly. During a clinical examination with a patient-loaded couch, note the position of the couch at the beginning and at the end of the examination with the use of a tape measure and a straightedge on the couch rails. Compare this with the intended couch movement. It should be within ±2 mm. CT Prof. Stelmark
© Copyright 2025