The Mini External Fixator Surgical Technique

The Mini External Fixator
Surgical Technique
2
The Mini External Fixator – Surgical Technique
Indications, Features and Components
Indications
•
•
•
•
•
•
Complex fractures of phalanges and metacarpals
Displaced intra-articular fractures
Segmental bone loss
Open fractures not allowing stable internal fixation
Fractures with associated complex soft tissue injuries
Tumour resections
Features
•
•
•
•
Module fits in the Modular Hand System
Low clamp profile
Compatible with the small External Fixator 107.982
Choice of threaded Kirschner wires of 1.25mm or 1.6mm dia.
with corresponding clamps
• Radiolucent Carbon-Fibre Rods
• Allows sequential clamping of two Kirschner wires in the clamp
Components
1.25mm
1.6mm
Clamp Bodies for 1.25mm and 1.6mm wires
Allow sequential tensioning of Kirschner wires to reduce the operation
time. The clamp-body nuts are captured.
Slender design contributes to overall low-profile frame for improved
patient comfort.
Connecting Clamp, 3.0/3.0mm dia.
Allows connection of 3.0mm Carbon-Fibre Rods for the treatment of
displaced intra-articular fractures. An additionally available 3.0/4.0mm
Connecting Clamp allows bridging to the small Synthes® External Fixator
for the treatment of complex hand and distal radius fractures.
3
The Mini External Fixator – Surgical Technique
Carbon-fibre rods
Lightweight for patient comfort.
Radiolucent for improved intra- and postoperative radiographic
visualization.
1.25mm
1.6mm
Threaded Kirschner wires with trocar tip, in two sizes
Provide more stability and better hold in cortical bone.
Eliminate need to predrill, for reduced operative time.
1.25mm and 1.6mm diameters optimise adaptation to fracture patterns
and patient anatomy.
Angle
Allows lateral application of pins while maintaining a dorsal frame construct, as determined by the fracture pattern and surgeon’s preference.
4
The Mini External Fixator – Surgical Technique
Surgical Technique
1
Insert first Kirschner wires
Insert one threaded Kirschner wire through the double drill sleeve
1.25/1.6 in each fragment. Place the first two Kirschner wires as close as
possible to the fracture in each fragment.
2
Add clamps and rods
Connect the appropriate clamp to each Kirschner wire through the open
hole of the clamp:
• 1.25mm Kirschner wires – Clamp Body for 1.25mm wires
• 1.6mm Kirschner wires – Clamp Body for 1.6mm wires
Insert a 3.0mm rod to connect the clamps.
Reduce the fracture. Lightly tighten the nuts of both pin and rod clamps
to provide sufficient stability for radiography and final fracture reduction.
Kirschner wire
through open hole
5
The Mini External Fixator – Surgical Technique
3
Insert remaining Kirschner wires
Insert the two remaining Kirschner wires as far as possible from the
fracture, using the remaining clamp holes as guides.
K-wire through
closed hole
4
Secure all clamp nuts
Tighten the nuts on both clamps, using the clamp holder to avoid
fracture displacement. Cut the Kirschner wires, if necessary.
6
The Mini External Fixator – Surgical Technique
Set contents
115.985
Mini External Fixator Instrument and Implant Set
304.564
Module for Mini External Fixator Instruments and
Implants, without contents
Item no.
Contents
292.600
Kirschner Wire, 1.25mm dia., with threaded tip,
length 100mm, stainless steel
Kirschner Wire, 1.6mm dia., with threaded tip,
length 100mm, Stainless steel
Angle, 3.0mm dia.
Carbon-Fibre Rod, 3.0mm dia., length 25mm
Carbon-Fibre Rod, 3.0mm dia., length 45mm
Carbon-Fibre Rod, 3.0mm dia., length 60mm
Carbon-Fibre Rod, 3.0mm dia., length 75mm
Clamp Body for 1.25mm wires
Clamp Body for 1.6mm wires
Connecting Clamp, 3.0/3.0mm dia.
Clamp Holder
Combination Wrench 5.5mm
Double Drill Sleeve 1.25/1.6
292.708
395.103
395.105
395.107
395.109
395.111
395.125
395.126
395.133
395.139
395.302
395.902
Units
10
10
2
2
2
2
2
6
6
4
1
1
1
Additionally available
292.710
395.134
395.610
395.640
392.040
Kirschner Wire, 1.6mm dia., with treaded tip,
length 150mm, trocar point (10 units per pack)
Connecting clamp, 3.0/4.0mm dia.
Connecting Bar, 4.0mm dia., carbon fibre, length 80mm
Connecting Bar, 4.0mm dia., carbon fibre, length 140mm
Hand Chuck for Kirschner wires
7
Manufacturer: Mathys Medical Ltd
Güterstrasse 5, P.O. Box, CH-2544 Bettlach
www.synthes.com
036.000.596
036.000.xxx
Presented by:
©
Medical
2004
© Stratec
edical 2004
Stratec M
Printed
Printed iin
witzerland
n SSwitzerland
LLAG
AG
Subject