SEAL “How To”
Change Lamp
3
Peristaltic Pumps
4
Change Syringe
5
Change Sample Probe and Flusher
6
SEAL Analytical - Mequon Technology Center 10520-C Baehr Road Mequon, WI 53092 262-241-7900 www.Seal-us.com
Copyright © 2010 SEAL Analytical
Reproduction or distribution of this document is expressly forbidden without written permission from SEAL Analytical
2
SEAL “How To”
Change Lamp
Disconnect
•
Turn Power Off
•
Remove Top Cover
•
Remove 3 mounting screws
•
Un-plug from Main Board
Reconnect
•
Plug new into Main Board
•
Place lamp into position
•
Note the screw holes only line
up one way
•
Insert and tighten screws
AQ1 Lamp
(11-0019)
Lamp Plug
Mounting Screws
Adjust Measurement System
•
Select Measurement Setup
•
Wash The Cuvette 3 times
•
Start Readings
•
Click on each wavelength
•
Select the filter with the highest
•
Adjust to 8.9V with Gain
•
Select Dark Current
•
Adjust to 0.0030 with Offset
Offset
Gain
Hover your Mouse Pointer over any
Icon To see the function.
SEAL Analytical - Mequon Technology Center 10520-C Baehr Road Mequon, WI 53092 262-241-7900 www.Seal-us.com
Copyright © 2010 SEAL Analytical
Reproduction or distribution of this document is expressly forbidden without written permission from SEAL Analytical
3
SEAL “How To”
Peristaltic Pump
Change Pump Tube
•
Loosen Knurled Nut
•
Slide Platen
•
Remove Pump Tubes
•
Disconnect from Nipple
•
Clean Pump Rollers
•
Clean Platen
•
Oil Rollers
•
Connect New Tube to Nipple
Knurled Nut
Pump Rollers
Pump Platen
Pump Tube
Pump Roller
Pin Point Oiler
Oil Here
Clean Platen Here
Clean Rollers
SEAL Analytical - Mequon Technology Center 10520-C Baehr Road Mequon, WI 53092 262-241-7900 www.Seal-us.com
Copyright © 2010 SEAL Analytical
Reproduction or distribution of this document is expressly forbidden without written permission from SEAL Analytical
4
SEAL “How To”
Change Syringe
1.
2.
3.
4.
Disconnect
•
Water in/out
•
4 Mounting Screws
•
Syringe Clamp
•
Clamp Screw
•
Rotate Screw Down
•
Valve Wire Connector
Mounting Screws
Syringe Clamp
Clamp Screw
Screw Drive
Replace
•
Syringe O-ring
•
Syringe Barrel
•
Syringe Piston
Valve Wire Connector
Top of Syringe Barrel
Re-Connect
•
Valve Wire Connector
•
4 Mounting Screws (loose)
•
Syringe Clamp
•
4 Mounting Screws (tighten)
•
Water in/out
Syringe Header Block
Inlet Valve
Syringe Glass Barrel
Header Block
Adjust Syringe Piston
•
Rotate Screw Up
•
Piston 1 -2 mm down
•
Tighten Clamp Screw
Clamp Screw
Needle Nose Pliers
Syringe Screw Base
SEAL Analytical - Mequon Technology Center 10520-C Baehr Road Mequon, WI 53092 262-241-7900 www.Seal-us.com
Copyright © 2010 SEAL Analytical
Reproduction or distribution of this document is expressly forbidden without written permission from SEAL Analytical
5
SEAL “How To”
Change Sample Probe & Flusher
1.
Disconnect Sample Probe
•
Unscrew thumbscrew
•
Disconnect tygon sample line
2.
Replace old Sample Probe
•
Reconnect tygon sample line
•
Place sample probe in arm
•
and tighten thumbscrew
3.
Disconnect Probe flusher
•
Using Allen wrench unscrew Probe
Flusher
•
Carefully disconnect silicon tubing
4.
Reconnect Probe Flusher
•
Attach silicon tubing to metal
prongs
•
Install Probe Flusher on the bottom of
Sample Arm
5.
In Mechanical Setup of Maintenance
check dilutor operation
•
See operator manual