SEAL “Technical Tips” Sample Addition

:
SEAL “Technical Tips”
Sample Addition
“Sample Addition Mode” has been incorporated into AQ1 Software, allowing
the user to stop the run in progress, add samples, and restart the run.
The included example demonstrates the entire process, from the set up of the
original schedule to the completion of the restarted run to which samples have
been added.
Below is the original “Scheduling” screen where the user first schedules the
tray:
SEAL Analytical - Mequon Technology Center 10520-C Baehr Road Mequon, WI 53092 262-241-7900 www.Seal-us.com
Copyright © 2010 SEAL Analytical
Reproduction or distribution of this document is expressly forbidden without written permission from SEAL Analytical
29
:
SEAL “Technical Tips”
Sample Addition
After the tray has been saved in ”Scheduling”, the run is initiated by doubleclicking the „Run‟ icon.
The “Tray Selection” screen then appears where the desired tray is selected for
analysis.
The “Options” screen then displays tick boxes where the user specifies
whether they want to auto standardize, perform water baselines, and/or
reset the reaction segment counter.
SEAL Analytical - Mequon Technology Center 10520-C Baehr Road Mequon, WI 53092 262-241-7900 www.Seal-us.com
Copyright © 2010 SEAL Analytical
Reproduction or distribution of this document is expressly forbidden without written permission from SEAL Analytical
30
:
SEAL “Technical Tips”
Sample Addition
If the „Auto Standardization‟ tick box is selected from the previous screen, the
“Auto Standard Preparation” screen appears. This screen allows the user to
confirm where the blank and top standard are located. Once the user has verified
that all the information is correct, clicking „OK‟ will start the run.
SEAL Analytical - Mequon Technology Center 10520-C Baehr Road Mequon, WI 53092 262-241-7900 www.Seal-us.com
Copyright © 2010 SEAL Analytical
Reproduction or distribution of this document is expressly forbidden without written permission from SEAL Analytical
31
:
SEAL “Technical Tips”
Sample Addition
The “Run” screen will now appear. The „Sample Addition‟ icon, located at the top
of the screen, will initiate “Sample Addition Mode.”
Sample Addition icon
SEAL Analytical - Mequon Technology Center 10520-C Baehr Road Mequon, WI 53092 262-241-7900 www.Seal-us.com
Copyright © 2010 SEAL Analytical
Reproduction or distribution of this document is expressly forbidden without written permission from SEAL Analytical
32
:
SEAL “Technical Tips”
Sample Addition
Once the user clicks the „Sample Addition‟ icon, the screen shown below appears
allowing the user to choose between the following two options: “Stop at the end
of the current cycle” or “Finish all work in progress.” If “Stop at the end of the
current cycle” is selected, the AQ1 will finish the current cycle, immediately go
through its wash cycle, and stop. The samples and/or standards that were
prepared, but not aspirated, will remain in the reaction wells and will be reprepared after restarting the run. If “Finish all work in progress” is chosen, the
AQ1 will continue to aspirate and read all samples and/or standards that have
been prepared and are incubating. The AQ1 will then go through its wash cycle
and stop.
SEAL Analytical - Mequon Technology Center 10520-C Baehr Road Mequon, WI 53092 262-241-7900 www.Seal-us.com
Copyright © 2010 SEAL Analytical
Reproduction or distribution of this document is expressly forbidden without written permission from SEAL Analytical
33
:
SEAL “Technical Tips”
Sample Addition
After clicking „Continue‟, the “Scheduling” screen appears which allows the user to
add samples. Both options (i.e. “Finish work in progress” and “Stop at the end of
current cycle”) allow samples to be added to the schedule while the AQ1 is
stopping.
SEAL Analytical - Mequon Technology Center 10520-C Baehr Road Mequon, WI 53092 262-241-7900 www.Seal-us.com
Copyright © 2010 SEAL Analytical
Reproduction or distribution of this document is expressly forbidden without written permission from SEAL Analytical
34
:
SEAL “Technical Tips”
Sample Addition
Notes:
1) New samples can only be added using this feature. AQ1 software does not
allow the user to edit or add tests to previously scheduled samples!
2) If running multiple tests on samples and if scheduling multiple tests for the
samples that are added on, please note:
If ”Sample Addition Mode” is initiated while the AQ1 is still performing Test 1 and
has not started Test 2, the AQ1 will complete the work in the following order:
a) Remaining „PND‟ samples for Test 1
b) Samples added on for Test 1
c) All samples for Test 2
If ”Sample Addition Mode” is initiated while the AQ1 is performing Test 2, after
having completed Test 1, the AQ1 will complete the work in the following order:
a) Samples added on for Test 1
b) Remaining „PND‟ samples for Test 2
c) Samples added on for Test 2
3) Neither auto-calibrants nor manual calibrants can be added to the schedule
in “Sample Addition Mode”.
4) If the frequency counting feature of QC Pro is being used for a test, the QC
Pro settings in Test Parameters must have „1‟ located in the „Start Before‟
field. This will ensure the QC is inserted prior to the first added sample.
SEAL Analytical - Mequon Technology Center 10520-C Baehr Road Mequon, WI 53092 262-241-7900 www.Seal-us.com
Copyright © 2010 SEAL Analytical
Reproduction or distribution of this document is expressly forbidden without written permission from SEAL Analytical
35
:
SEAL “Technical Tips”
Sample Addition
The “Scheduling” screen below shows the samples that have been added in
cups 6 – 9.
SEAL Analytical - Mequon Technology Center 10520-C Baehr Road Mequon, WI 53092 262-241-7900 www.Seal-us.com
Copyright © 2010 SEAL Analytical
Reproduction or distribution of this document is expressly forbidden without written permission from SEAL Analytical
36
:
SEAL “Technical Tips”
Sample Addition
After the addition of necessary samples, click on the „Run Screen‟ icon, located
on the left side on the screen, to view the status of the initial run. Once the
status indicator bar displays “Finished”, click the „Restart Run‟ icon to initiate the
run with the added samples.
SEAL Analytical - Mequon Technology Center 10520-C Baehr Road Mequon, WI 53092 262-241-7900 www.Seal-us.com
Copyright © 2010 SEAL Analytical
Reproduction or distribution of this document is expressly forbidden without written permission from SEAL Analytical
37
:
SEAL “Technical Tips”
Sample Addition
After clicking the „Restart Run‟ icon, the “Options” screen appears where the
user has the option to reset the reaction segment counter and/or perform water
baselines.
Please note, although it is possible to perform water baselines at this point, it is
not recommended.
SEAL Analytical - Mequon Technology Center 10520-C Baehr Road Mequon, WI 53092 262-241-7900 www.Seal-us.com
Copyright © 2010 SEAL Analytical
Reproduction or distribution of this document is expressly forbidden without written permission from SEAL Analytical
38
:
SEAL “Technical Tips”
Sample Addition
Once the user has clicked „OK‟, the “Run” screen will appear with the additional
samples appended to the tray. Below is the updated “Run” screen which
includes the added samples as well as the proper insertion of QC‟s.
SEAL Analytical - Mequon Technology Center 10520-C Baehr Road Mequon, WI 53092 262-241-7900 www.Seal-us.com
Copyright © 2010 SEAL Analytical
Reproduction or distribution of this document is expressly forbidden without written permission from SEAL Analytical
39
:
SEAL “Technical Tips”
Sample Addition
Once the status indicator bar displays “Finished”, the run is complete.
SEAL Analytical - Mequon Technology Center 10520-C Baehr Road Mequon, WI 53092 262-241-7900 www.Seal-us.com
Copyright © 2010 SEAL Analytical
Reproduction or distribution of this document is expressly forbidden without written permission from SEAL Analytical
40