INSTRUCTIONS Number Description

INSTRUCTIONS
Pierce™ Mass Spec Sample Prep Kit
for Cultured Cells
84840
2522.1
Number
Description
84840
Pierce Mass Spec Sample Prep Kit for Cultured Cells, sufficient reagents for processing 20 samples
of 100µg of cell lysate protein for MS analysis
Kit Contents:
Cell Lysis Buffer, 5mL
Digestion Buffer, 5mL
No-Weigh™ DTT, 24 microtubes, each containing 7.7mg of dithiothreitol (DTT)
Iodoacetamide, Single-Use, 24 microtubes, each containing 9.3mg of iodoacetamide (IAA)
Trypsin Storage Solution, 250µL
Pierce Digestion Indicator, 10µg
Lys-C Protease, MS Grade, 20µg
Pierce™ Trypsin Protease, MS Grade, 2 × 20µg
Storage: Upon receipt, remove Insert A (containing Pierce Digestion Indicator, Lys-C Protease and
Pierce Trypsin Protease, MS Grade) and store at -20°C. Store the remaining components at 4°C.
Product is shipped on dry ice.
Table of Contents
Introduction ................................................................................................................................................................................. 1
Procedure Summary..................................................................................................................................................................... 2
Important Product Information .................................................................................................................................................... 2
Additional Materials Required..................................................................................................................................................... 2
Procedure for Preparation of Peptides from Cultured Cells for MS Analysis ............................................................................. 2
A. Material Preparation ....................................................................................................................................................... 2
B. Cell Lysis ........................................................................................................................................................................ 2
C. Reduction, Alkylation and Acetone Precipitation........................................................................................................... 3
D. Enzymatic Protein Digestion .......................................................................................................................................... 3
Troubleshooting ........................................................................................................................................................................... 4
Additional Information ................................................................................................................................................................ 4
Related Thermo Scientific Products ............................................................................................................................................ 5
Introduction
The Thermo Scientific™ Pierce™ Mass Spec Sample Prep Kit for Cultured Cells enables reproducible processing of cultured
mammalian cells for proteomic mass spectrometry (MS) analysis. The kit contains all of the necessary buffers, reagents, MSgrade enzymes and an optimized protocol to generate MS-compatible peptide samples from whole-cell lysates. Consistent
sample preparation is vital, but difficult, for robust MS analysis. Although there are numerous MS sample preparation
methods, most are highly variable and time consuming, which results in excessive peptide modification (e.g., oxidation,
carbamylation, over-alkylation) and subsequent inefficient data analysis and interpretation. Enzymatic digestion
reproducibility is another source of sample prep variability and is especially critical to monitor for quantitative MS-proteomic
analysis. The kit uses a double-digestion protocol with Lys-C and trypsin to more completely proteolize complex protein
samples and generates peptides with < 10% miscleavages. Additionally, the kit contains Thermo Scientific™ Pierce™
Digestion Indicator, which is used as a standard protein digestion indicator to assess the digestion reproducibility for multiple
samples.
Pierce Biotechnology
PO Box 117
(815) 968-0747
3747 N. Meridian Road
Rockford, lL 61105 USA
(815) 968-7316 fax
www.thermoscientific.com/pierce
Procedure Summary
Important Product Information
•
Warm the Cell Lysis Buffer and Digestion Buffers to room temperature before use. Store buffers at 4°C.
•
Pierce Digestion Indicator contains 10µg of protein per vial in a suitable buffer. The actual concentration is printed on
the bottle label. Digestion efficiency may be verified by determining the sequence coverage of the Pierce Digestion
Indicator (see the Additional Information Section).
Additional Materials Required
•
Microcentrifuge tubes
•
Microtip probe sonicator or nuclease (e.g., Thermo Scientific™ Pierce™ Universal Nuclease for Cell Lysis, Product No.
88700)
•
Heating block
•
Chilled (-20°C) 100% acetone and 90% acetone
•
Trifluoroacetic acid (TFA)
•
Phosphate-buffered saline (PBS)
•
Protein assay kit (e.g., Thermo Scientific™ BCA Protein Assay Kit, Product No. 23227)
•
75-300µm capillary C18 reversed-phase column
•
Mass spectrometer with online or off-line nano-flow liquid chromatography (LC) system
•
Vacuum concentrator (e.g., Thermo Scientific™ SpeedVac™ Vacuum Concentrator)
•
Data analysis software (e.g., Thermo Scientific™ Proteome Discoverer™ Software)
Procedure for Preparation of Peptides from Cultured Cells for MS Analysis
A. Material Preparation
Pre-chilled 90% acetone
Prepare 90% acetone in ultrapure water (e.g., mix 45mL of 100% acetone with 5mL of
ultrapure water) and store at -20°C.
Pre-chilled 100% acetone
Store 100% acetone at -20°C.
B. Cell Lysis
1.
Culture cells to harvest at least 100µg of protein. For best results, culture a minimum of 1 × 106 cells.
Note: Rinse cell pellets 2-3 times with 1X PBS to remove cell culture media. Pellet cells using low-speed centrifugation
(i.e., < 1000 × g) to prevent premature cell lysis.
2.
Lyse the cells by adding five cell-pellet volumes of Cell Lysis Buffer (i.e., 100µL of Cell Lysis Buffer for a 20µL cell
pellet). Pipette sample up and down to break up the cell clumps and gently vortex sample to mix.
3.
Incubate the lysate at 95°C for 5 minutes.
Pierce Biotechnology
PO Box 117
(815) 968-0747
3747 N. Meridian Road
Rockford, lL 61105 USA
(815) 968-7316 fax
2
www.thermoscientific.com/pierce
4.
Cool the lysate on ice for 5 minutes.
5.
Sonicate lysate on ice using a microtip probe sonicator to reduce the sample viscosity by shearing DNA. Alternatively,
use Pierce Universal Nuclease for Cell Lysis (Product No. 88700) to enzymatically digest DNA and RNA. If using
nuclease, add 25 units of nuclease to 1mL of cell lysate and incubate at room temperature for 15 minutes.
6.
Centrifuge lysate at 16,000 × g for 10 minutes at 4°C.
7.
Carefully separate the supernatant and transfer into a new tube.
8.
Determine the protein concentration of the supernatant using established methods such as the BCA Protein Assay Kit
(Product No. 23227).
C. Reduction, Alkylation and Acetone Precipitation
Note: This procedure is optimized for 100µg of cell lysate protein at 1mg/mL concentration; however, the procedure may be
used for 10-200µg of cell lysate protein with an appropriate amount of reagents (DTT, IAA, Lys-C and trypsin). When using
10µg of cell lysate, a protein concentration of 0.2-1mg/mL may be used.
1.
Warm and equilibrate the Pierce Digestion Indicator to room temperature.
2.
Add 100µg of lysate protein to a polypropylene microcentrifuge tube and adjust the sample volume to 100µL using Cell
Lysis Buffer to a final concentration of 1mg/mL.
3.
Add 0.5µg (0.5% w/w) of Pierce Digestion Indicator to the sample.
Note: The actual concentration is printed on the bottle label. Refer to the label to determine the required volume.
4.
Immediately before use, puncture the foil covering of the Thermo Scientific™ No-Weigh™ DTT tube with an empty
pipette tip. Add 100µL of ultrapure water to the tube and gently pipette up and down to dissolve the contents of the tube.
The final concentration of DTT is ~500mM.
Note: To preserve DTT stability between uses, return unused microtubes to the pouch containing the desiccant pack.
5.
Add 2.1µL of DTT solution to the sample (final DTT concentration is ~10mM). Mix and incubate at 50°C for
45 minutes. Discard any unused DTT solution.
6.
Cool the sample to room temperature for 10 minutes.
7.
Immediately before use, puncture the foil covering of the Single-Use Iodoacetamide tube with an empty pipette tip. Add
100µL of Cell Lysis Buffer to the tube and gently pipette up and down to dissolve the contents of the tube. The final
concentration of IAA is ~500mM. Protect solution from light.
8.
Add 11.5µL of IAA solution to the sample (final IAA concentration is ~50mM). Mix and incubate at room temperature
for 20 minutes protected from light. Discard any unused IAA solution.
9.
After alkylation with IAA, immediately add 460µL (4 volumes) of pre-chilled (-20°C) 100% acetone to sample. Vortex
tube and incubate at -20°C for one hour to overnight to precipitate proteins.
10. Centrifuge at 16,000 × g for 10 minutes at 4°C. Carefully remove acetone without dislodging the protein pellet.
11. Add 50µL of pre-chilled (-20°C) 90% acetone, vortex to mix and centrifuge at 16,000 × g for 5 minutes at 4°C.
12. Carefully remove acetone without dislodging the protein pellet. Allow the pellet to dry for 2-3 minutes and immediately
proceed to Section D. Enzymatic Protein Digestion.
Note: Do not dry the acetone-precipitated protein pellet for more than 2-3 minutes; excess drying will make the pellet
difficult to re-suspend in the Digestion Buffer.
D. Enzymatic Protein Digestion
1.
Add 100µL of Digestion Buffer to the acetone-precipitated protein pellet and resuspend by gently pipetting up and down
to break the pellet.
Note: An acetone-precipitated protein pellet may not completely dissolve; however, after proteolysis at 37°C, all the
protein will be solubilized.
Pierce Biotechnology
PO Box 117
(815) 968-0747
3747 N. Meridian Road
Rockford, lL 61105 USA
(815) 968-7316 fax
3
www.thermoscientific.com/pierce
2.
Immediately before use, add 40µL of ultrapure water to the bottom of the vial containing Lys-C and incubate at room
temperature for 5 minutes. Gently pipette up and down to dissolve. Store any remaining Lys-C solution in single-use
volumes at -80°C.
3.
Add 2µL of Lys-C (1µg, enzyme-to-substrate ratio = 1:100) to the sample. Mix and incubate at 37°C for 2 hours.
4.
Immediately before use, add 40µL of Trypsin Storage Solution to the bottom of the vial containing trypsin and incubate
at room temperature for 5 minutes. Gently pipette up and down to dissolve. Store any remaining trypsin solution in
single-use volumes at -80°C for long-term storage.
5.
Add 4µL of trypsin (2µg, enzyme-to-substrate ratio = 1:50) to the sample. Mix and incubate overnight at 37°C.
6.
Freeze samples at -80°C to stop digestion. (Optional: stop digestion by acidifying with TFA)
7.
Speed vac sample to remove the Digestion Buffer.
8.
Resuspend the sample in an appropriate buffer (e.g., 0.1% TFA) and subject for LC-MS analysis.
Note: Proteolytic digests prepared using this protocol are directly compatible with LC-MS analysis. Clean-up of samples
with C18 spin tips or columns (see Related Thermo Scientific Products) is optional.
Troubleshooting
Problem
Possible Cause
Solution
Low protein yield
Insufficient cells
Increase number of cells used for lysis
Difficult to re-suspend the pellet
following acetone precipitation
Over-dried protein pellet
Air dry the acetone-precipitated pellet for 2-3 minutes
Incomplete digestion
Inactive enzyme
Store any remaining enzyme solution in single-use
volumes at -80°C
Sequence coverage of the Pierce
Digestion Indicator is < 50%
Working at the limit of
detection of the MS
instrument
Increase concentration of the Pierce Digestion
Indicator to 1% (w/w)
Over-alkylation
Alkylation was allowed to
proceed for too long
Alkylate the reduced protein protected from light for
20 minutes at room temperature and immediately
remove the excess reagents by acetone precipitation
Quench alkylation reactions by adding 10mM DTT
(final concentration)
Additional Information
A. MS Data Analysis
Identify proteins by searching the LC-MS/MS data with a database-search software (e.g., Thermo Scientific™ Matrix
Science Mascot™ or Proteome Discoverer™ Software) including fixed modification of cysteines (Average mass: 57.0513Da,
Monoisotopic mass: 57.021464Da) and variable oxidation of methionines (Average mass: 15.9994Da, Monoisotopic mass:
15.994915Da). To assess digestion efficiency and instrument sensitivity, the protein database should include the Pierce
Digestion Indicator protein sequence:
MESDESGLPAMEIECRITGTLNGVEFELVGGGEGTPEQGRMTNKMKSTKGALTFSPYLLSHVMGYGFYHFGTYPSGYENPFLHAI
NNGGYTNTRIEKYEDGGVLHVSFSYRYEAGRVIGDFKVMGTGFPEDSVIFTDKIIRSNATVEHLHPMGDNDLDGSFTRTFSLRDG
GYYSSVVDSHMHFKSAIHPSILQNGGPMFAFRRVEEDHSNTELGIVEYQHAFKTPDADAGEE
The five digestion indicator peptides most commonly detected are underlined above. The Pierce Digestion Indicator peptides
are quantified manually with extracted ion chromatograms of the raw LC-MS/MS data or automatically with Thermo
Scientific™ Pinpoint™ Software (Table 1).
Pierce Biotechnology
PO Box 117
(815) 968-0747
3747 N. Meridian Road
Rockford, lL 61105 USA
(815) 968-7316 fax
4
www.thermoscientific.com/pierce
Table 1. Properties of the five Thermo Scientific Pierce Digestion Indicator peptide sequences.
Digestion Indicator Peptide Sequence
Observed Mass/Charge
Observed Charge
Hydrophobicity Factor
1209.1007
871.9189
610.2701
648.3367
587.0315
+2
+2
+3
+3
+4
40.59
40.24
27.24
42.42
35.13
ITGTLNGVEFELVGGGEGTPEQGR
VMGTGFPEDSVIFTDK
DGGYYSSVVDSHMHFK
SAIHPSILQNGGPMFAFR
VEEDHSNTELGIVEYQHAFK
To assess LC-MS/MS instrument performance and to assist with the quantitative analysis, spike Thermo Scientific™
Pierce™ Retention Time Calibration Mixture (Product No. 88321) into the sample at a 1:100 dilution. This calibration
mixture is used to normalize autosampler variability, to verify elution-time consistency and mass calibration, and to provide
retention time controls for data analysis using the Pinpoint™ Software.
Related Thermo Scientific Products
84841
Pierce Digestion Indicator
23227
BCA Protein Assay Kit
88320
Pierce Peptide Retention Time Calibration Mixture
28904
Trifluroacetic acid
87784
Pierce C18 Tips, 100µL bed
89870
Pierce C18 Spin Columns
90061
TMTsixplex™ Isobaric Label Reagent Set
90051
Lys-C Protease, MS Grade
90057
Pierce Trypsin Protease, MS Grade
Products are warranted to operate or perform substantially in conformance with published Product specifications in effect at the time of sale, as set forth in
the Product documentation, specifications and/or accompanying package inserts (“Documentation”). No claim of suitability for use in applications regulated
by FDA is made. The warranty provided herein is valid only when used by properly trained individuals. Unless otherwise stated in the Documentation, this
warranty is limited to one year from date of shipment when the Product is subjected to normal, proper and intended usage. This warranty does not extend to
anyone other than Buyer. Any model or sample furnished to Buyer is merely illustrative of the general type and quality of goods and does not represent that
any Product will conform to such model or sample.
NO OTHER WARRANTIES, EXPRESS OR IMPLIED, ARE GRANTED, INCLUDING WITHOUT LIMITATION, IMPLIED WARRANTIES OF
MERCHANTABILITY, FITNESS FOR ANY PARTICULAR PURPOSE, OR NON INFRINGEMENT. BUYER’S EXCLUSIVE REMEDY FOR NONCONFORMING PRODUCTS DURING THE WARRANTY PERIOD IS LIMITED TO REPAIR, REPLACEMENT OF OR REFUND FOR THE NONCONFORMING PRODUCT(S) AT SELLER’S SOLE OPTION. THERE IS NO OBLIGATION TO REPAIR, REPLACE OR REFUND FOR PRODUCTS
AS THE RESULT OF (I) ACCIDENT, DISASTER OR EVENT OF FORCE MAJEURE, (II) MISUSE, FAULT OR NEGLIGENCE OF OR BY BUYER,
(III) USE OF THE PRODUCTS IN A MANNER FOR WHICH THEY WERE NOT DESIGNED, OR (IV) IMPROPER STORAGE AND HANDLING OF
THE PRODUCTS.
Unless otherwise expressly stated on the Product or in the documentation accompanying the Product, the Product is intended for research only and is not to
be used for any other purpose, including without limitation, unauthorized commercial uses, in vitro diagnostic uses, ex vivo or in vivo therapeutic uses, or
any type of consumption by or application to humans or animals.
Current product instructions are available at www.thermoscientific.com/pierce. For a faxed copy, call 800-874-3723 or contact your local distributor.
© 2013 Thermo Fisher Scientific Inc. All rights reserved. Mascot is a trademark of Matrix Science. All (other) trademarks are the property of Thermo Fisher
Scientific Inc. and its subsidiaries. Printed in the USA.