Sample Measurement Troubleshooting DS-11 Spectrophotometer Technical Note 127 Quick Help

DS-11 Spectrophotometer
Sample Measurement Troubleshooting
Technical Note 127
Quick Help
Most issues concerning accuracy and reproducibility are sample or measurement technique related and are
resolved by following the suggestions below:
• Ensure that the sample isolation procedure is optimized and samples are purified prior to making absorbance
measurements.
• Ensure all solutions are homogenous and well-mixed prior to sampling.
• Ensure sample concentrations fall within the linear range of the instrument.
• Establish a new blank using the appropriate buffer.
• Verify that the buffer does not have significance absorbance at the analysis wavelength.
Microvolume Mode
• Ensure both top and bottom microvolume measurement surfaces are clean prior to making the Blank
measurement.
• Remove sample solutions from both the top and bottom measurement surfaces immediately after each
measurement.
• Use calibrated pipettors and properly fitting tips to ensure the full 1 μL aliquot is delivered to the measurement
surface. Protein samples may wick to the outside of the tip and may not be properly dispensed.
• Use a clean pipette tip and a new aliquot of sample for each microvolume measurement.
Launch the Diagnostics app and perform a Microvolume mode Self Test to confirm instrument is working within
specifications. If the Self Test fails- clean both the top and bottom surfaces and repeat the test.
Cuvette Mode
• Use cuvettes with Z heights of 8.5 mm.
• Use quartz or UV transparent cuvettes for UV range measurements.
• Ensure cuvettes are clean prior to use.
• Fill cuvettes to manufacturer's specifications.
• Confirm the Mode selection has been changed to Cuvette mode on the Run screen prior to a measurement.
• Ensure cuvettes are properly inserted following the etched arrow light path.
Launch the Diagnostics app and perform a Cuvette mode Self Test to confirm instrument is working within
specifications. Remove any cuvettes prior to tapping the Self Test button.
Additional technical help bulletins are available at www.denovix.com. Contact DeNovix Customer Support if further
help is required. Please include the serial number of your instrument when contacting Customer Support by email.
Outside of the US, please contact your local distributor for assistance.
DeNovix, Inc.
3411 Silverside Road
Wilmington, DE 19810 USA
Phone: +1.302.442.6911
Email: [email protected]
www.denovix.com
Copyright 2014 DeNovix, Inc.
Version: Sept 2014