Dräger X-dock 5300 Dräger X-dock 6300/6600 Technical Manual i

Dräger X-dock 5300
Dräger X-dock 6300/6600
Technical Manual
i
Content
1
1.1
1.2
For your safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
General safety notes . . . . . . . . . . . . . . . . . . . . . . . . .4
Definitions of alert icons . . . . . . . . . . . . . . . . . . . . . .4
2
2.1
2.2
2.3
2.4
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Product overview . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Feature description . . . . . . . . . . . . . . . . . . . . . . . . . .6
Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
GPL (General Public Licence) . . . . . . . . . . . . . . . . .6
3
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
4
4.1
4.2
4.3
4.4
4.5
4.6
4.7
4.8
4.9
Basics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Switching the station on or off . . . . . . . . . . . . . . . . . .7
Initial setup of station . . . . . . . . . . . . . . . . . . . . . . . .7
Touchscreen display . . . . . . . . . . . . . . . . . . . . . . . . .7
Start and test screens . . . . . . . . . . . . . . . . . . . . . . . .7
Logging users in or out . . . . . . . . . . . . . . . . . . . . . . .8
Managing user profiles . . . . . . . . . . . . . . . . . . . . . . .8
Configuring the test gas inlet . . . . . . . . . . . . . . . . . .9
Managing tests . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Managing instrument profiles . . . . . . . . . . . . . . . . .12
5
5.1
5.2
5.3
5.4
5.5
5.6
Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
Conducting a visual inspection . . . . . . . . . . . . . . . .13
Inserting or removing the gas measuring device into
or from a module . . . . . . . . . . . . . . . . . . . . . . . . . . .13
X-am 125+ module with charging function (optional) 13
Station self-test . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
Conducting a test . . . . . . . . . . . . . . . . . . . . . . . . . .14
After use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
6
6.1
6.2
6.3
6.4
6.5
Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
Menu overview . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
Managing tests . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
Station gas configuration . . . . . . . . . . . . . . . . . . . . .17
System configuration . . . . . . . . . . . . . . . . . . . . . . .17
Data management . . . . . . . . . . . . . . . . . . . . . . . . .17
7
Permission level. . . . . . . . . . . . . . . . . . . . . . . . . . .19
8
Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . .20
9
9.1
9.2
9.3
9.4
9.5
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
Maintenance intervals . . . . . . . . . . . . . . . . . . . . . . .21
Performing a firmware update . . . . . . . . . . . . . . . . .21
Changing the sealing insert . . . . . . . . . . . . . . . . . .22
Changing the fresh air filter . . . . . . . . . . . . . . . . . . .22
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
10
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
11
Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
12
Order list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
13
Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
Dräger X-dock 5300/6300/6600
3
For your safety
1
For your safety
1.1
General safety notes
Before using this product, carefully read the Instructions for
Use.
Strictly follow the Instructions for Use. The user must fully
understand and strictly observe the instructions. Use the
product only for the purposes specified in the Intended use
section of this document.
Do not dispose of the Instructions for Use. Ensure that they
are retained and appropriately used by the product user.
Only trained and competent users are permitted to use this
product.
Comply with all local and national rules and regulations
associated with this product.
Only trained and competent personnel are permitted to
inspect, repair and service the product. Dräger recommend
a Dräger service contract for all maintenance activities and
that all repairs are carried out by Dräger.
Only trained and competent personnel are permitted to
inspect, repair and service the product as detailed in these
Instructions for Use.
Use only genuine Dräger spare parts and accessories, or
the proper functioning of the product may be impaired.
Do not use a faulty or incomplete product. Do not modify
the product.
Notify Dräger in the event of any component fault or failure.










1.2
Definitions of alert icons
The following alert icons are used in this document to provide
and highlight areas of the associated text that require a greater
awareness by the user. A definition of the meaning of each
icon is as follows:
4
!
WARNING
!
CAUTION
ii
NOTICE
Indicates a potentially hazardous situation
which, if not avoided, could result in death or serious
injury.
Indicates a potentially hazardous situation which, if not
avoided, could result in physical injury, or damage to
the product or environment. It may also be used to
alert against unsafe practices.
Indicates additional information on how to use the
product.
Dräger X-dock 5300/6300/6600
Description
2
Description
2.1
Product overview
X-dock 5300 (X-am)
1
X-dock 5300 (Pac)
X-dock 6300/6600
2
4
0
5
3
X-dock 5300/6300
X-dock 6600
15
6
7
8
9
10
11
12
14
13
13
X-dock 6300/6600
max. 10
16
17
18
19
00133286.eps
Dräger X-dock 5300/6300/6600
5
Installation
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
Master
Module
Status LED
Touchscreen display
Function key
Fresh air inlet with filter
Antitheft slot
Power supply
USB ports
Ethernet port
Mini USB port
Exhaust
Gas inlets
Compressed air inlet
Type plate
X-am 125 module
Charge status LED
X-am 125+ module (with charging function)
Pac module
2.2
Feature description
2.2.1
Master
The master assumes control of the sequencing of the
maintenance station for functional testing, calibration and
adjustment, as well as functions for user management,
instrument management, printing of standard reports and
standard certificates (using PostScript printers only), and also
the user interface.
2.2.2
Modules
The instrument-specific interfaces such as, e. g., IR
communication, gas exposure unit and charging contact are
integrated into the modules. In addition, the modules contain
sensors for detecting the visual, audible and vibration alarms
on the instruments.
2.3
Intended use
The Dräger X-dock 5300/6300/6600 is a maintenance station
of modular construction. The X-dock can be used to perform
automated calibrations, adjustments and bump tests on
portable gas monitors in parallel and independently from one
another. A system consists of one master for 3 (X-dock 5300/
6300) or 6 (X-dock 6600) test gases. The X-dock 5300
consists of a master station with one module and cannot be
expanded. Up to 10 modules can be connected to the X-dock
6300 and 6600 master. The modules automatically detect
when an instrument has been inserted and control the gas
supply so that an appropriate supply to the instrument is
ensured at all times.
The following gas measuring devices can be used with the
X-dock and its modules:
X-dock 5300/6300/6600
with Pac module:
with X-am 125 (+) module:
Dräger Pac 3500
Dräger X-am 1700
Dräger Pac 5500
Dräger X-am 2000
Dräger Pac 7000
Dräger X-am 5000
Dräger X-am 5600
6
2.4
GPL (General Public Licence)
Some of the device software includes open-source software,
which has been published under GPL, LGPL or another opensource licence. They include GPL GPLv2, LGPL, MIT,
PostgreSQL, Apache, Apache 2, zlib. The source texts of the
software can be obtained from Dräger on a CD under material
number 83 21 874 for at least three years after purchase of the
software. The licence terms and conditions of the software are
included on CD.
3
Installation
ii
NOTICE
Ensure adequate space for the entire assembly.
The master and all modules must have the same
firmware version. If this is not the case, a firmware
update needs to be carried out (see chapter 9.2 on
page 21).
1. If applicable, install modules on the master station in
accordance with the assembly instructions (X-dock 6300/
6600 only).
 A maximum of 10 modules can be installed on one
master station.
 The available modules can be combined in any way
desired.
2. If applicable, install wall mount or cylinder holder in
accordance with the assembly instructions.
3. Remove the sleeves from the intended gas inlets and from
the gas outlet.
ii
NOTICE
If the sleeve is not removed from the gas outlet, the
station will be unable to conduct the self-test without
errors.
4. Fit the gas feed hoses to the gas inlets on the master and
connect them to the regulator valve on the test gas
cylinder.
5. If required, connect an exhaust hose (max. 10 m long) to
the exhaust outlet.
6. Ensure a supply of compressed air or fresh air:
 Connect the compressed air hose to the compressed
air port (outlet pressure of pressure regulator valve
0.5 bar, flow rate >3 l/min).
OR
 If required, connect the fresh air hose to the fresh air
filter.
7. Connect the power pack.
 Station with up to 3 modules: Power pack 24 V / 1.33 A
 Station with 4 to 10 modules: Power pack 24 V / 6.25 A
The entire system is supplied with power via the master.
Dräger X-dock 5300/6300/6600
Basics
ii
NOTICE
Initial setup of station
NOTICE
1. Switch on the station, see chapter 4.1 on page 7.
2. Log in with the preconfigured "admin" user
(User name: admin, password: 123456), see chapter 4.5
on page 8.
3. Configure the test gas inlet, see chapter 4.7 on page 9.
4. Change language if necessary:
a. Select
> System configuration > Language.
b. Selected the required language.
c. Confirm selection with OK.
5. Set date and time where necessary:
a. Select
> System configuration > Date & time.
b. Make the desired settings.
c. Confirm settings with OK.
6. If required, change standard tests (see chapter 4.8 on
page 11).
7. If required, set up network (see chapter 6.4.3 on page 17).
If no action has occurred for 10 minutes, the station
switches automatically to standby mode.
4.3
Dräger recommends the use of Dräger gas cylinders
and Dräger pressure regulator valves (see order list).
Alternatively there is the option of using a suitable
pressure regulator valve with 0.5 bar outlet pressure
and >3 l/min flow rate.
Dräger recommends connecting an exhaust hose
(max. 10 m long) to the exhaust outlet to discharge
the test gas into the open air.
4
Basics
4.1
Switching the station on or off
ii
4.2
To switch the station on:
 Press and hold the
key on the master for approx.
1 second.
The following information is displayed during the switch-on
process:
 Software version number
To switch the station off:
 Press and hold the
key on the master for approx.
3 seconds.
The station switches off.
Standby mode:
Standby mode is activated after approx. 10 minutes with no
activity on the station (input via touchscreen or opening/
closing of a module cover).
 When the station switches over to standby mode, any user
logged in will be logged off automatically. The user will
need to log back in again when switching back to operating
mode.
 The touchscreen is switched off in standby mode.
 The charging function of X-am 125+ modules with charging
function is not affected by standby mode. The charging
process is not interrupted.
 To switch to operating mode:
 Tap the function key or
 touch the touchscreen or
 open or close a module flap.

Dräger X-dock 5300/6300/6600
Touchscreen display
The buttons on the touchscreen display change dynamically
depending on the task being executed. To execute an action,
select the corresponding icon on the display.
Press the
screen.
4.4
key on the master at any time to access the start
Start and test screens
The keys on the start and test screen change dynamically
depending on the log-in status, single-mode status and the
number of modules in use.
Start screen: User logged in
(single mode deactivated)
Logged-on
user
►
Instrument field
►
Menu
►
◄ User
logon/logoff
◄ Favourites bar
7
Basics
4.5
Start screen: User not logged in
(single mode deactivated)
◄
Instrument field
ii
User
logon/logoff
►
Logging users in or out
NOTICE
A User-ID is required to log in. This must be created
beforehand by the Administrator (see chapter 4.6 on
page 8).
A user with administrator rights is created by default:
User name: admin
Password: 123456
ii
NOTICE
Dräger recommends changing the admin password
after initial start-up.
◄ Favourites bar
Start screen: single mode activated
Selected test
►
Instrument field
►
User
◄ logon/logoff
To log in a user:
1. Select .
a. Select .
b. Select the desired user name from the list.
or
a. Select Select user.
b. Enter desired user name.
2. Enter the password and confirm with .
To log out the current user:
1. Select .
Information about the current user will be displayed.
2. Select .
The current user will be logged out.
ii
4.4.1
Symbols
Menu
8
Select this button to access the
menu (see chapter 6 on
page 16).
Confirm
Select this button to confirm an
input or function.
Cancel
Select this button to cancel an
input or function.
Back
Select this button to access the
previous screen.
Log user in or out
Select this button to log users in
or out. The number in the
symbol indicates the permission
level (see chapter 7 on
page 19).
4.6

NOTICE
When entering the user name, 3 already saved user
names will automatically be suggested and displayed.
For quick selection, please select the desired user
name.
Managing user profiles
Select
> Data management > User.
The user overview will be displayed.
Created user
►
Dräger X-dock 5300/6300/6600
Basics
To create a new user profile:
1. Select
> Data management > Users > New.
The input form will be displayed.
Enter last name
Enter first name
►
►
Enter User ID
Enter permission level and password
►
►
Enter company name
Enter location
►
►
Enter comments
►
Select the gas inlet
►
2. Select desired test gas inlet.
The configuration menu will appear.
2. The following information must be created:
 User name
 User ID
 Permission level (see chapter 7 on page 19)
 Password
3. The following information can optionally be given:
 Company name
 Location
 Comments
4. Save the new user profile with OK.
ii
NOTICE
Select Test gas inlet
Enter part no.
Enter Expiry date
►
►
◄ Enter batch no.
►
Create Test gas
◄ component
Select Test gas
►
◄ Delete configuration
Enter concentration
►
◄ Select unit
Further options
►
User profiles can only be created by the administrator
or by users with authorisation level 5.
When using a Dräger test gas cylinder:
To edit an existing user profile:
1. Select
> Data management > Users.
The user profiles will be displayed.
2. Select the user profile.
3. Select Edit.
4. Edit the desired settings.
5. Confirm the new settings with OK.
4.7
!
Configuring the test gas inlet
ii
NOTICE
Upon entering the part no. of a Dräger gas cylinder, a
gas cylinder level indicator is displayed automatically,
unless this function was deactivated before (see
chapter 4.7.2 on page 10).
1. Enter the part no. of the Dräger test gas cylinder.
All the necessary details for the configuration will be filled
in automatically. The batch number and the expiry date can
be entered manually in addition.
WARNING
The test gas concentrations entered must be identical
to the specifications on the test gas cylinder used.
Incorrect details will result in faulty measurement
results.
To configure a test gas inlet:
1. Select
> Station gas configuration.
An overview of the test gas connections is displayed.
Dräger X-dock 5300/6300/6600
ii
NOTICE
The values entered automatically must be matched
with those indicated on the test gas cylinder. If these
values are not identical, the value indicated on the gas
cylinder is to be considered and a manual correction of
the values must be made.
2. If necessary select Further options and to reset the gas
cylinder level indicator.
3. If required, configure other test gas inlets in the same way.
9
Basics
When using a test gas cylinder from another manufacturer:
1. Create or delete test gas component.
 Create a new test gas component with
.
 Delete the current test gas component with
.
ii
Status of test gas pressure:
Display Gas pressure Meaning
Deleting all test gas components
information of the test gas inlet.
deletes
all
Select test gas.
Enter the test gas concentration.
Select test gas units.
If required, create additional test gas components.
The following information can optionally be given:
 Part no. of the test gas cylinder
 Batch number of test gas cylinder
 Expiry date of the test gas cylinder
7. If required, enter Further options.
ii
0,4 - 0,6 bar
Test gas pressure correct.
>0.6 bar
Test gas pressure too high.
Gas cylinder level indicator:
Gas cylinder level
Display
61 - 100 %
30 - 60 %
Further options
< 30 %
NOTICE
Gas cylinder level indicator deactivated
It is necessary to specify the volume, volume type,
pressure and unit to use the gas manager.
4.7.2
Settings
1. Select
Configured test gas
components
►
Enter volume
►
Enter pressure
Status
Test gas pressure
►
►
◄ Volume type
◄ Enter unit for
pressure
State of test
gas pressure/
◄
gas cylinder level
indicator
1. Enter volume of test gas cylinder.
2. Set type of volume:
Comp. = cylinder volume
Relax. = gas volume (cylinder volume x pressure)
3. If required, refresh test gas pressure.
4. If necessary, activate Hose > 2 m if the length of the test
gas hose exceeds 2 m.
5. If required, configure other test gas inlets in the same way.
10
Test gas pressure too low.
NOTICE
2.
3.
4.
5.
6.
4.7.1
<0.4 bar
> Station gas configuration > Settings.
Select test gas for X-am 2000
►
Fresh air inlet setting
►
Set test behaviour for missing test gases
Select test gas for PAC 7000 OV
Gas cylinder level indicator
►
►
►
The OV sensor of the PAC 7000 OV allows to choose between
two different test gases that are also used for calibration and
testing. These test gases are carbon monoxide (CO) and
ethylene oxide (EO).
For the X-am 2000, it is possible to choose between three
different test gases that are also used for calibration and
testing. The three options are methane (CH4), propane (C3H8)
and pentane (C5H12). The sensor is calibrated with different
sensitivity levels depending on the gas selected. More
information on this can be found in the relevant sensor data
sheets.
Dräger X-dock 5300/6300/6600
Basics
ii
NOTICE
The corresponding gas must be connected to one of
the gas inlets and set in the gas configuration.
There is also the option of setting an "increased sensitivity"
option for propane and pentane. This option artificially
increases the sensitivity to calibrate the sensors so that these
have approximately a nonane sensitivity level (in other words,
a sensitivity level as if they were calibrated for nonane). More
information on the subject of cross-sensitivity calibration can
be found in the relevant sensor data sheets.
To select the test gas for X-am 2000:
1. Select X-am 2000.
2. Select the required test gas from the list.
The following selection is available:
 Methane - CH4 (default setting)
 Propane - C3H8
 Pentane - PENT
For propane and pentane, the "Increased sensitivity"
(vapour sensitivity) option can also be activated.
3. Confirm selection with OK.
To set the fresh air input:
1. Select Fresh air.
2. Select pump (fresh air inlet; default setting) or compressed
air inlet.
3. Confirm selection with OK.
1. Select Gas level monitoring.
2. Activate or deactivate the Gas level monitoring on check
box.
3. Confirm selection with OK.
To reset the gas cylinder level indicator for a new test gas
cylinder:
1. Connect a new test gas cylinder to a test gas connection.
2. Select
> Station gas configuration.
3. Select desired gas inlet.
Select Further options and select
to reset the gas cylinder
level indicator.
4.8

Managing tests
Select
> Define test.
The test overview will be displayed.
Select test mode
►
Select test 1
►
◄ Edit test 1
Select test 2
►
◄ Edit test 2
Select test 3
►
◄ Edit test 3
◄ Configure test mode
To set the test behaviour with missing test gases:
1. Select Missing gas.
2. Activate checkbox (default setting: activated).
3. Confirm selection with OK.
This function can be used to set whether or not a test or
calibration is carried out when a required test gas is not
connected.
!
WARNING
If this function is deactivated, the corresponding
channel is not tested or calibrated.
To select the test gas for PAC 7000 OV:
1. Select Pac 7000 OV.
2. Select the required test gas from the list.
The following selection is available:
 Ethylene oxide - EO (default setting)
 Carbon monoxide - CO
3. Confirm selection with OK.
To set the gas cylinder level indicator:
ii
NOTICE
The gas cylinder level indicator is only available for
cylinders that are configured via a Dräger part no.
Dräger X-dock 5300/6300/6600
4.8.1
Selecting the test mode
The following test modes are available. The test modes define
the behaviour of the system when the user is logged off. The
Test scheduler and Log-in mode test modes are configurable.
Description
The Favorites mode displays the predefined
tests. These can be selected via the favourites
bar.
Single mode In Single mode, a predefined test is started by
closing the module flap. Several tests can be
started and performed in parallel.
Test
The Test scheduler mode enables the user to
scheduler
configure the time and weekday when a
predefined test should be performed. To
configure the Test scheduler mode, select .
Manual tests can only be started when the user
is logged in.
Log-in mode Only logged-in users (level 2-5) can perform
tests in Log-in mode. The configuration enables
you to define whether a user should be
automatically logged off after the completion of
a test. To configure the Log-in mode, select .
Mode
Favorites
mode
11
Basics
To select the test mode:
1. Select .
2. Select the desired test mode.
3. If necessary, select the
of the test mode to be edited
(only possible with the Test scheduler and Log-in mode
test modes).
4. If necessary, edit tests (see chapter 4.8.2 on page 12).
5. Save the settings with OK.
4.8.2
ii
Creating or editing a test
NOTICE
A maximum of 3 favourites may be created.
The first 3 characters of the test name are displayed in
the list of favourites.
To create or edit a test:
1. Select icon
for the test to be edited.
The editing window will be displayed.
Enter test name
►
Select gas test
►
Further options
►
ii
►

4. If required, select Options.
The following options are available:
Response time test The response times of the sensors
(only for calibration) are tested.
The battery voltage is measured
Bat. check
and displayed.
All the alarm functions of the gas
Alarm test
monitor are checked (noise, light,
vibration).
The zero point is checked.
Zero check
12
Average shift values (time-weighted average) are generally
limited to eight hours exposure per day per workplace for
5 days a week during a work lifetime.
6. Save settings with OK.
4.9
2. Enter desired test name.
3. Select desired gas test.
The following gas tests are available:
No gas test, only the activated
options are run.
Test for exceeding A1
Quick bump test
concentration.
Test for reaching cylinder
Bump test
concentration within a tolerance
range.
Zero point and sensitivity
Calibration
adjustment
When the alarm functions are being tested, the
ambient noise must not be too loud as otherwise the
station will not be able to check the horn function.
5. If required, select Further options:
The time on the gas monitor is
Sync. clock
synchronised with the station.
If this option is selected, Reset DL
and Reset TWA are automatically
also activated to ensure that the DL
(data logger) is unique.
Download data memory and events
Download DL
memory data from the gas monitor
to the station.
The data in the data and events
Reset DL
memories are cleared.
Reset TWA
The TWA1 time is reset.
The gas monitor is switched off
Turn off
automatically after the test.
Perform automatic repair (e. g.
Autom. repair
calibrate after a faulty bump test).
The system is purged with fresh air
Fresh air
after every test.
Create certificate (stored in the
Generate
station as a PDF file).
Print out the certificate on a
Print
connected USB printer.
1
Further options
NOTICE
Managing instrument profiles
Select
> Data management > Devices.
The instrument overview is displayed.
Instrument no.
►
◄ Instrument type
To create an instrument profile:
 A new instrument profile will be created automatically as
soon as a gas instrument that is not yet stored in the station
is placed in a module.
Dräger X-dock 5300/6300/6600
Use
5
A defective pressure reducer on the gas cylinder can
lead to increased pressure in the station. The gas
hoses may loosen as a result and gas may escape.
X-am-125+ module with charging function only:
 After the instrument is inserted, the charging status is
displayed for approx. 5 seconds via the charging status
LED.
 The charging function starts automatically approx. 15
mintues after the last test.
Health hazard! Test gas must not be inhaled. Observe
the hazard warnings in the relevant Safety Data
Sheets.
Provide venting into a fume cupboard
or outside the building.
To remove the instrument from the module:
1. Push the lock up slightly and open the module cover
upwards.
2. Remove the instrument.
Use
!
ii

WARNING
NOTICE
To prevent loss of test gas, Dräger recommends
closing the test gas cylinders when the station is left
unattended for long periods.
5.3
!
Adjustment may not be possible due to instrument and
channel errors.
5.1
Conducting a visual inspection
A visual inspection of the gas measurement systems should be
conducted every time before being inserted into the station.
1. Check that the housing, external filters and the nameplates
are intact.
2. Check the battery contacts and sensor inputs for dirt.
ii
5.2

Devices that have not passed the visual inspection
must not be inserted into the station. Otherwise the
test cannot be fully evaluated correctly.
Inserting or removing the gas measuring
device into or from a module
WARNING
Risk of explosion! Do not charge underground or in
explosion hazard areas. The X-am 125+ modules with
charging function are constructed in compliance with
regulations for fire-damp weather and explosion
protection.
It is only possible to charge the instrument batteries using
the X-am 125+ module with charging function.
The charging time is approx. 4 hours for a completely
empty battery.
A new NiMH supply unit reaches its full capacity after 3 full
charge/discharge cycles. Never store the device for
extended periods (max. 2 months) without a power supply
as the internal buffer battery will run down.omatically once
the temperature is back in charging range.


NOTICE
X-am 125+ module with charging function
(optional)
If an error occurs:
Remove from the module and re-insert.
 If this does not correct the error, have the module repaired.

!
CAUTION
A short-circuit in the charging contacts in the modules,
e. g. due to metal objects falling into the device, will not
damage the station but should be avoided due to a
potential risk of overheating and error displays on the
module.
Overview of charge status LEDs
Colour
Status
Meaning
green
on continuously
Charge status 100 %
green
flashing
Battery charging.
red
flashing
Charging error
2
1
00233286.eps
To insert the instrument in the module:
1. If necessary, push the lock up slightly and open the module
cover upwards.
2. Place the instrument in the corresponding module.
3. Close the module cover.
The instrument will be detected automatically.
Dräger X-dock 5300/6300/6600
13
Use
5.4
Station self-test
A self-test is carried out:
 When the station is started up.
 When the last successful self-test was more than 24 hours
ago and a test is being carried out.


If necessary, log user out on the station (see chapter 4.5 on
page 8).
The preset test is performed according to the configured
time schedule.
Conducting a test
If Log-in mode is activated:
Log user in on the station (see chapter 4.5 on page 8).
 Select the desired test from the favourites bar.
The test will start automatically.
The status LED flashes blue.
The individual test phases are displayed.
!
WARNING
Test passed:
ii
NOTICE
The station is tested for leaks, pump function, and the software
version of the individual modules and the master.
5.5

When using methane, propane or butane in the range
>100 %LEL, an exhaust hose (max. 10 m long) must
be connected to the exhaust outlet to ensure the
extraction of excess explosive gas.
X-dock 5300
Single mode is activated by default.
X-am 5000
Several tests can be started and executed in parallel in
the Single mode.
0
A failure of a LED, horn or vibration test results in a
negative evaluation of the overall test, and thus in the
locking of the respective gas measurement system.
A test of the sensor reserve is only performed with
sensors that support this function. The results are
displayed under test details and give information about
the state of the sensor.
The following tests are preconfigured:
Test 1: QT
Test 2: EXT
Test 3: CAL
Fast bump test including alarm testing
Extended bump test incl. zero-point check and
alarm testing.
Calibration, alarm test, fresh air flushing and
certificate.
01033286.eps




Confirmation is shown on the display.
The Status LED flashes green.
If required, select the desired instrument field for additional
information.
Remove the instrument from the module.
Test not passed:
1. If necessary, open the test gas cylinders.
2. If necessary, switch on the X-dock.
3. Perform a visual inspection of the gas measurement
systems (see chapter 5.1 on page 13).
4. Insert the instruments into the modules (see chapter 5.2 on
page 13).
X-dock 5300
X-am 5000
0
If Single mode is activated:
The preset test is started automatically by closing the
module cover.
The Status LED flashes blue.
The individual test phases are displayed.

01133286.eps
If Favorites mode is activated:
 Select required test from favourites bar.
The test will be started automatically.
The Status LED flashes blue.
The individual test phases are displayed.




If Test scheduler mode is activated:
14

An error message is shown on the display.
The Status LED flashes red.
If required, select the desired instrument field for additional
information.
Identify and rectify the error.
Repeat the test if necessary.
Dräger X-dock 5300/6300/6600
Use
Overview of status LEDs
Colour
Status
blue
flashing
green
flashing
red
flashing
5.6
Meaning
Process in progress
Test passed
Test failed/cancelled
After use
1. If required, remove instruments from modules.
2. Close the test gas cylinders.
ii
NOTICE
To keep energy consumption low, Dräger
recommends switching off the equipment after use
according to the Instructions for Use.
Dräger X-dock 5300/6300/6600
15
Menu
6
Menu
6.1
Menu overview
ii
NOTICE
The menu is only available for users with permission
levels 5.
Define test
Station gas configuration
System configuration
Date & time
Data management
Language
Network
System address
Database
DBMS address
Info screen
Update
Service interval
Devices
Users
Station event logger
Export data logger
Export certificates
Mobile synchronization
MSD1
1
16
Mass storage device
Dräger X-dock 5300/6300/6600
Menu
6.2
Managing tests
In this menu, the existing tests can be edited, new tests can be
created and the standard test can be defined.
 Select
> Define test.
The test overview will be displayed.
For more information see chapter 4.8 on page 11.
6.3
Station gas configuration
The individual gas connections can be configured in this menu.
 Select
> Station gas configuration.
An overview of the test gas connections is displayed.
For more information see chapter 4.7 on page 9.
6.4
System configuration
6.4.1
Setting Date & time
ii
NOTICE
If the station is connected to a server, only place and
format can be changed. The station synchronises time
and date automatically.
The following settings can be made in this menu:
 Time zone
 Date format
 Date
 Time
 Location
To set the date and time:
1. Select
> System configuration > Date & time.
2. Make the desired settings.
3. Confirm settings with OK.
6.4.2
Setting the language
1. Select
> System configuration > Language.
2. Selected the required language.
3. Confirm selection with OK.
6.4.3
Network
ii
NOTICE
If the IP or DBMS address is changed, the station must
be re-started to make the change active.
To set System address:
1. Select
> System configuration > Network > System
address.
2. Select dynamic or static IP address.
3. Where necessary, enter the IP of the system address.
4. Where necessary, enter the Subnet Mask (default:
255.255.255.0).
Dräger X-dock 5300/6300/6600
To set the central database address:
1. Select
> System configuration > Network > DBMS
address.
2. Enter the server IP #.
6.4.4
Cleaning the database
1. Select
> System configuration > Database > Clean
up.
2. Select Start clean up.
3. Confirm deletion of the database entries with OK.
The progress of the deletion is displayed.
ii
6.4.5
NOTICE
Database entries, data loggers and certificates are
deleted.
Displaying info screen
1. Select
> System configuration > Info screen.
The following information is displayed:
 Application version
 Hardware version
 Kernel version
 MAC address
 IP address
6.4.6
Performing an update
To perform a firmware update, see chapter 9.2 on page 21.
6.4.7
Service interval
1. Select
> System configuration > Service interval.
 This function displays how many days have passed
since the last service was performed. If the service
interval has been exceeded, an additional check box to
deactivate the self check note for 365 days is displayed.
6.5
ii
6.5.1
Data management
NOTICE
All entries are stored in UTC time (Universal Time
Coordinated).
Managing instruments
The individual instrument profiles can be displayed in this
menu.
 Select
> Data management > Devices.
The instrument profiles will be displayed.
For more information see chapter 4.9 on page 12.
6.5.2
Managing user profiles
User profiles can be created, edited or deleted in this menu.
 Select
> Data management > Users.
The user profiles will be displayed.
For more information see chapter 4.6 on page 8.
17
Menu
6.5.3
Exporting Station event logger
The saved data of the event memory can be saved to a
connected USB storage device with this function.
To copy data from the event memory to a USB data storage
device:
1. Connect the USB data storage device to the USB port on
the station.
The USB icon will appear in the status bar.
2. Select
> Data management > Station event logger.
The event memory is displayed.
3. Select the required files from the list.
4. Confirm selection with OK.
The selected files are copied onto the USB data storage
device.
6.5.4
Exporting data memory
The saved data of the data memory can be saved to a
connected USB storage device with this function.
To copy data from the data memory to a USB data storage
device:
1. Connect the USB data storage device to the USB port on
the station.
The USB icon will appear in the status bar.
2. Select
> Data management > Export data logger.
The data memory is displayed.
3. Select the required files from the list.
4. Confirm selection with OK.
The selected files are copied onto the USB data storage
device.
6.5.5
ii
NOTICE
Dräger recommends cleaning the local data after
successfully importing the data into the database (see
chapter 6.4.4 on page 17).
Dräger recommends exporting the local data on a
regular basis to keep the amount of data to be
exported, and thus copied, small.
The export via the MSD area is faster than via a USB
storage device.
6.5.7
Activating or deactivating MSD mode
This function allows the station to be used as a USB mass
storage device and the data saved in the station can be copied
to a connected PC.
To activate MSD mode:
1. connect PC with USB mini B cable to the X-dock.
2. Select
> Data management > MSD mode.
3. Activate MSD mode.
4. Confirm setting with OK.
The station is now visible on the PC as a drive.
ii
NOTICE
No other actions can be performed with the station
while MSD mode is active.
Exporting certificates
This function can be used to save the stored certificates to a
connected USB storage device.
To copy certificates to a USB data storage device:
1. Connect the USB data storage device to the USB port on
the station.
The USB icon will appear in the status bar.
2. Select
> Data management > Export certificates.
The event memory is displayed.
3. Select the required certificates from the list.
4. Confirm selection with OK.
The selected certificates are copied onto the USB data
storage device.
6.5.6
4. Confirm the selection with OK.
The data is copied to the target location.
To deactivate MSD mode:
 Cancel MSD mode with X.
The station is ready for operation again.
Mobile synchronization
This function can be used to copy the local database, reports,
certificates and device data loggers to a connected USB
storage device or the internal MSD area. This is useful if the
data cannot be exchanged via the network connection. The
data can then be imported into the X-dock Manager (see
X-dock Manager Online Help).
1. If necessary, connect the USB storage device to the USB
port on the station.
2. Select > Data management > Mobile synchronization .
The available target locations are displayed.
3. Select desired target location.
18
Dräger X-dock 5300/6300/6600
Permission level
7
Permission level
The availability of functions, tests or menu items depends on the privileges for the particular permission level.
Function
Perform preset tests
1
X
2
X
Permission level
3
X
4
X
5
X
Selection of tests for logged-in state
Change your password
X
X
Define test
Export data memory (station)
Set Station gas configuration
Show devices
Mode selection (Single/Group mode)
X
X
X
X
X
Manage users
Set date and time
Perform SW update (station)
X
X
X
Set language
Set network configuration
Database operations
X
X
X
Dräger X-dock 5300/6300/6600
19
Troubleshooting
8
Troubleshooting
Faults
Station does not start.
Cause
Remedy
Power supply not connected.
Check connections to mains supply,
power pack and station.
Power pack defective.
Replace power pack.
Check that the station can be started by
pressing the function key.
Module dirty.
Gas monitor not detected.
Measuring chamber leaky.
No communication with PC.
Gas monitors do not pass the gas tests
(test gas not reaching the instrument).
Test gas pressure is too low.
Clean module.
Instrument dirty.
Clean instrument.
Module defective.
Contact DrägerService.
SW version of gas monitor out of date.
Update gas monitor SW.
Instrument and/or module dirty.
Clean instrument and/or module.
Seal worn.
Replace seal (see chapter 9.3 on Page
22)
Ethernet cable not connected correctly.
Check cabling.
Ethernet cable defective.
Replace Ethernet cable.
IP address incorrectly configured.
Reconfigure IP address.
Firewall is blocking data traffic.
Configure the firewall correctly.
Hose connections incorrect.
Check hose connections.
Station gas configuration not correct.
Check Station gas configuration.
Test gas cylinder empty or closed.
Check test gas cylinder.
Pressure regulator valve incorrectly set.
Check setting of pressure regulator valve.
Pressure regulator valve incorrectly set.
Check setting of pressure regulator valve.
Hose connections incorrect.
Test gas cylinder empty or closed.
Test gas pressure is too high.
Pressure regulator valve incorrectly set.
Alarm element test not carried out.
Check whether or not external light
External light sources are interferring with sources are creating interference, reposition light sources or station as
the alarm element test.
necessary.
The system does not recognize the
connected USB storage device.
The formatting of USB storage devices
Dräger recommends using FAT32 and a
varies, depending on the manufacturer.
partitioning that contains partitioning
This applies to the file system (e.g. FAT32
tables (USB-ZIP or USB-HDD).
or NTFS) as well as the partitioning.
Module not found or Wrong
(incompatible) SW version error
message (although everything is
connected correctly).
Module X does not support firmware
version error message.
20
This error can occur if a master is
connected to modules with different
software versions.
Check setting of pressure regulator valve.
In this case, a firmware update has to be
performed via the master, even if the most
current firmware is already installed on
the master.
If the modules are not connected
Check whether all connections are
correctly, the electrical connection may be
correctly connected and screwed tight.
faulty.
The firmware version of the affected gas Perform an update with the CC Vision PC
monitor is not compatible with X-dock.
software.
Dräger X-dock 5300/6300/6600
Maintenance
9
Maintenance
9.1
Maintenance intervals
ii
9.1.1
NOTICE
The maintenance intervals must be established in
each individual case and shortened if necessary,
depending on safety considerations, process
conditions, and the technical requirements of the
equipment. Dräger recommend a Dräger service
contract for all maintenance activities and that all
repairs are carried out by Dräger.
Before every start-up
The following work must be carried out before every start-up of
the equipment:
 Check the hoses for dirt, brittleness and damage and
replace if necessary.
 Check the hoses are secure, to prevent escapes of gas.
 Check that all cable connections are secure.
 Visual inspection of the modules and sensor seals. If very
dirty or if there are visible defects, the sensor seal must be
replaced.
9.1.2
3. Select
>
System
configuration > Update.
A list of all firmware updates
available on the USB storage
device will be displayed.
4. Select the desired firmware update from the list. Inst is
displayed next to the firmware update.
5. Start the firmware update with OK.
The progress of the installation is
displayed:
Annually
Inspection of the entire X-dock station by competent
personnel.
9.2
Performing a firmware update
!
CAUTION
ii
NOTICE
The station power supply must not be disconnected
during the installation process. The station may be
damaged if this is not observed.
The station does not support any USB data storage
device with an NTFS file system.
6. Following successful transfer to the station, the start is restarted automatically and the firmware update is installed
immediately afterwards. During the installation process,
the status LEDs on the modules will be white.
7. After the installation is complete, the station changes to
operating mode. The station is ready for operation.
1. Download the firmware update from the internet:
a. Go to www.draeger.com.
b. Go to the X-dock product page and unzip the firmware
update to the root directory of an empty USB data
storage device.
!
CAUTION
There must not be any older firmware files on the USB
data storage device.
2. Connect the USB data storage device with firmware update
to the USB port on the station.
The USB icon will appear in the status bar.
Dräger X-dock 5300/6300/6600
21
Disposal
9.3
ii
Changing the sealing insert
9.4
NOTICE
ii
The sealing inserts must be changed at regular
intervals (e. g. at each inspection) or sooner as
required.
Changing the fresh air filter
NOTICE
With regular use and depending on the conditions of
use, the fresh air filter should be changed typically
every 2 months.
1. Unscrew the old fresh air filter.
2. Screw in the new fresh air filter.
9.5
!
Cleaning
CAUTION
Abrasive cleaning implements (brushes, etc.),
cleaning agents and cleaning solvents can destroy the
fresh air filter.
The device does not need any special care.
 If very dirty, the equipment can be carefully wiped down
with a damp cloth.
 Carefully dry the device using a cloth.
10
Disposal
Dispose of the product in accordance with the applicable rules
and regulations.
00633286.eps
1. Open the module cover.
2. Squeeze the two external
locking lugs in and
withdraw the sealing insert
downwards.
3. Release the hoses from
the sealing insert.
4. Replace the sealing insert.
5. Fit the hoses to the new
sealing insert (note the
arrows on the sealing
insert and the hose).
6. Squeeze the external
locking lugs in and insert
the sealing insert into the
module cover until the
locking lugs engage.
7. Check that the sealing
insert is fitted correctly in
the module cover.
Disposing of electric and electronic equipment:
In accordance with EU Directive 2002/96/EC this
product must not be disposed of as household waste.
This is indicated by the adjacent icon.
You can return this product to Dräger free of charge.
For information please contact the national
marketing organisations and Dräger.
00733286.eps
22
Dräger X-dock 5300/6300/6600
Technical data
11
Technical data
Dimensions (H x W x D):
Master
approx. 3.54 x 5.70 x 9.84 inches
(120 x 130 x 250 mm)
Module
approx. 3.54 x 5.70 x 9.84 inches
(90 x 145 x 250 mm)
12
Order list
Name and description
Order No.
Dräger X-dock 5300 X-am 125
83 21 880
Dräger X-dock 5300 Pac
83 21 881
Dräger X-dock 6300 Master
83 21 900
Dräger X-dock 6600 Master
83 21 901
Dräger X-dock Module X-am 125
83 21 890
Dräger X-dock Module X-am 125+
(with charging function)
83 21 891
Dräger X-dock Module Pac
83 21 892
Dräger X-dock Module X-am 125,
AA version
83 24 260
Dräger X-dock Module X-am 125+,
AA version (with charging function)
83 24 261
Dräger X-dock Module Pac, AA version
83 24 262
Single wall mount
83 21 922
Comfort wall mount
83 21 910
Cylinder holder (table-top version)
83 21 918
Cylinder holder for top-hat rail
83 21 928
Power pack 24 V / 1.33 A (up to 3
modules)
83 21 849
Power pack 24 V / 6.25 A (up to 10
modules)
83 21 850
X-dock car adapter
83 21 855
Pressure regulator valve 0.5 bar
83 24 250
Pump filter set (includes filter and hose
connector)
83 19 364
Fluoroelastomer hose
12 03 150
Sealing insert (X-am)
83 21 986
Weight:
Master
Module
approx. 2.57 pounds
(30.86 ounces; 1500 g)
approx. 2.57 pounds
(30.86 ounces; 960 g)
Ambient conditions:
During operation
During storage
32 °F to +104 °F (0 °C to +40 °C)
-20-4 °F to +122 °F (-20 °C
to +50 °C)
700 to 1300 hPa
max. 95% relative humidity
X-dock 5300/6300
X-dock 6600
1x fresh air connection
1x compressed air inlet
1x exhaust outlet
3x gas inlets
6x gas inlets
Inlet pressure:
for the measured gas
for compressed air
0,5 bar ±20 %
0,5 bar ±20 %
Power supply:
11 V - 28 V DC, 6.25 A
Gas connections:
Connections:
Serial no. (year of
manufacture):
CE mark:
3x USB 2.0 standard A
connection, (host, cable <3 m)
1x USB 2.0 mini B connection,
(device, cable <3 m)
1x Ethernet port RJ45
Data transmission rate
10/100 Mbit
The year of construction is given
by the 3rd letter in the factory
number located on the nameplate:
B=2010, C=2011, D=2012,
E=2013, F=2014, G=2015,
H=2016, etc.
Example: Serial number ARFH0054, the 3rd letter is F, so the
year of construction is 2014.
Electromagnetic compatibility
(Directive 2004/108/EC)
Dräger X-dock 5300/6300/6600
Sealing insert (Pac)
83 21 987
X-dock Master display protector film
83 21 804
Stickers for module numbering
83 21 839
Barcode label, exterior
(22 x 8 mm, 500 labels)
AG02551
Barcode scanner
83 18 792
Dräger X-dock Manager Basic
83 21 860
Dräger X-dock Manager Professional
83 21 870
Dräger X-dock Manager Licence
(1x, both versions)
83 21 857
Dräger X-dock Manager Licence
(5x, both versions)
83 21 858
ii
NOTICE
Dräger recommends the use of Dräger test gas
cylinders.
23
Glossary
13
Glossary
Abbreviation
Description
ALARM
Alarm element test
FBT
Fast bump test
ABT
Accelerated bump test
CAL
Calibration
CALFBT
Calibration followed by fast bump test
CALEBT
Calibration followed by extended bump test
EXT
Extended bump test incl. zero-point check and alarm testing
FAV
Favourite
FRESH
Fresh air calibration
HORN
Horn
LED
Light-emitting diode
MST
Master
QT
Fast bump test including alarm testing
SPAN
Sensitivity calibration
T90
T90 test
T90_Reset
Reset T90 test results
UNDEF
Unknown
UNK
Unknown info
VIB
Vibration
ZCHECK
Zero-point check
ZERO
Zero-point calibration
24
Dräger X-dock 5300/6300/6600
Glossary
Dräger X-dock 5300/6300/6600
25
90 33 286 - TH 4634.600 en
© Dräger Safety AG & Co. KGaA
Edition 06 - September 2013 (Edition 01 - August 2012)
Subject to alteration
Product information
Dräger Safety AG & Co. KGaA
Revalstraße 1
23560 Lübeck, Germany
Tel +49 451 882 0
Fax +49 451 882 20 80
www.draeger.com