BD Microtainer MAP Microtube for Automated Process Instrument Compatibility Guide for Sysmex

BD Microtainer® MAP Microtube
for Automated Process
Instrument Compatibility Guide for Sysmex® Instrumentation
Every Drop of Blood Counts
The BD Microtainer MAP Microtube for Automated Process provides another opportunity to decrease iatrogenic blood loss
due to diagnostic testing. It can provide the capability for low volume blood draws, especially in patients with poor venous
access, and it can decrease the number of sample processing steps.
Sysmex supports blood conservation programs with a goal of helping laboratories make every drop of blood count while
delivering streamlined, standardized sample processing.
The following chart is provided to inform users of the compatibility of this new sample tube with various Sysmex
hematology lab systems.
XS-SERIES
XT-SERIES
XT-1800i
Compatible
Instrument
Compatibility
with MAP
Suggested Min.
Vol. (µL)
of Blood
to Run
Automated
XS-1000i
Compatible Manual
Open Mode Only
20μL
XT-2000i
Compatible
XE-SERIES
XE-2100
Standalone Compatible
XE-5000
Standalone Compatible
XT-4000i
Compatible
XE-Series on AlphaN and
HST-N Systems
Compatible
375μL
500μL
White paper VS8113-1 available
from the BD website.
Comments
Not compatible with
auto loader.
White paper VS8113-1
available from the
BD website.
HST-N– run in “single mode” setting.
Refer to your systems’ Operators
Manuals and Training Guides for
further details about modes
of operation.
Not compatible with the
SP-1000i slidemaker/stainer.
XN-SERIES
All XN-Series are compatible.
Instrument
Compatibility
with MAP
XN-1000
XN-2000
XN-3000
XN-9000
Suggested Min.
Vol. (µL)
of Blood
to Run
Automated
Comments
250μL
A dedicated rack with bar code must be used with
the MAP Tube on all configurations of the XN-Series.
System software enables automatic processing for the
tubes and bypassing of the SP-10 slide preparation unit
which is not compatible at this time. The minimum
sample size is 250μL with 88μL being sampled.
Supporting Documentation | Available Downloads
White Paper
A Comparison of BD Microtainer® MAP Microtube for Automated Process with the BD Microtainer® Tube Microgard™
Closure for Routine Hematology Testing on the Sysmex® XE-2100 Literature Number: VS8113-1
http://paswhitepapers.bd.com/
©2014 BD - BD, BD Logo and all other trademarks are property of Becton, Dickinson and Company.
©2014 Sysmex - Sysmex, Sysmex Logo and all other trademarks are property of Sysmex.
BD® Customer Support
BD Technical Support (US & Canada) 1.800.631.0174 or e-mail [email protected]
BD Customer Service 1.888.237.2762 or visit us anytime online at: www.bd.com/vacutainer
Sysmex Customer Support
Sysmex Corporation
1-5-1-Wakinohama-Kaigandori
Chuo-ku, Kobe 651-0073, Japan
Tel. +81 (78) 265-0521
www.sysmex.co.jp
Sysmex America, Inc.
577 Aptakisic Road
Lincolnshire, IL 60069, U.S.A.
Tel. +1 (800) 379-7639
www.sysmex.com/us
Sysmex Canada, Inc.
5045 Orbitor Drive
Building 9, Suite 401
Mississauga, ON, L4W 4Y4, Canada
Tel. +1 (905) 366-7900
www.sysmex.ca
Sysmex Latin America and the Caribbean
Rua Joaquim Nabuco 615 - Bairro Cidade Jardim
São José dos Pinhais
Paraná – Brasil – CEP 83040-210
Tel. +55 (41) 2104-1314
www.sysmex.com.br
©2014 Sysmex America, Inc.
MKT-10-1206 V03/05/14 P03/14