Read More - Acceler8 Media

THM-100 Manual
Section 2 Installation and Facilitization
2.1
Wall or Rack mounting
The THM-100 instrument weighs ~100 pounds. Four holes in the rear unistrut will accommodate up to
3/8” (10 mm) bolts or mounting studs. Spacing of the holes is 22.25” (565 mm) horizontally and 58” (1473
mm) vertically. See drawings following.
2.2
Clearance Requirements
Recommend at least 4” (10 cm) clearance on each side of the THM-100 and at least 2 feet (.6 m)
clearance in front of the THM-100. The doors are capable of swinging ~150 degrees.
2.3
Facilities
The instrument requires the following facilities:
Electrical Power
 Single phase 110 VAC (10 A) or 230 VAC (5 A) (automatically switches internally and will accept
voltages over a range of 85-264 VAC)
 The THM typically draws less than 1 Amp, with a maximum possible continuous amperage of 4.6
A (at 120 VAC) or 2.5 A (at 230 VAC).
 Installation should include an external, accessible breaker or switch. If a circuit breaker is
provided, it should be rated ≥ 10 A at 120 VAC or ≥ 5 A at 230 VAC and have current vs. time
delay curves such that it will allow for a cold start inrush current of 70 A for less than 1/60th of a
second. A 10 Amp 3-pole breaker with D tripping characteristics meets those requirements. (e.g.
ABB’s S 223-K 10 meets those requirements.)
 Electrical hookup is done internal to the top cabinet via hardwired made to the instrument’s
disconnect switch. Wiring used to the instrument should meet local code and should be at least
14 AWG (conductor diameter ≥ 1.63 mm). Power should include Line (L), Neutral (N), and
Ground/Earth (G/E). The power line should enter the cabinet through the right-hand conduit
fitting. L & N go to the disconnect switch and G/E goes to the ground lug.
Clean Dry Air or N2
 20 - 100 psig at inlet, less than 0.1 scfm (1.4 – 7 Bar, max of 10 Lpm, average of 1 Lpm)
 An SMC KQ2LE01-00 fitting, a “one-touch” bulkhead elbow fitting on the left side of the
instrument receives air through 1/8” tubing.
 Installation should include an external, accessible shutoff valve.
(Optional Compressor for Dry Air)
If there is no CDA supply facilitized close to the THM-100 instrument, a portable air-compressor can be
used, with the following strongly preferred specifications:
 Oil-free
 ~10 Liter (or larger) capacity tank
 Tank pressure >90 psig
The following models, among many others that meet these specifications, are:
 Sealey SAC05020
 Bostitch SFC50OL-U
 Ridgid OF45175
Installation and Facilitization
AMS CONFIDENTIAL – PROTECTED BY NDA
The information in this manual is the copyrighted, trade secret material of AMS. Possession and use of this information is restricted by contract; this information is not to be shared with third parties, and must be returned to AMS on demand. ©2010‐2012, all rights reserved.
5
THM-100 Manual
 Bostitch CAP2060
 Porter Cable C2002
 Rol-air JC10
An auto-draining moisture separator (AKA condensate trap) should also be installed, in-line between the
compressor and the THM-100 instrument. The exit of the moisture separator should be supplied with a
“one-touch” 1/8” tubing fitting.
Drain



Drain for sample water and any pre-sample purge water. The water contains minute amounts of
Quencher and Descaler (if needed). pH is typically 3-10 in water collected from the drain.
Drain should receive less than .5 liters per hour (normally much less) but at a .5 liters per minute
flowrate.
An SMC KQ2LE01-00 fitting, a “one-touch” bulkhead elbow fitting on the left side of the
instrument accepts 1/8” tubing for a drain line.
Water in (Sample)
 20-100 psig at inlet, capable of flow at 0.4 gph (1.4 – 7 Bar, capable of 0.25 lpm)
 Input should ideally have a regulator maintaining inlet pressure 30-40 psig (2.0-2.75 Bar), the
preferred pressure at the instrument. The instrument does have an internal water regulator.
 The sample inlet should have an external, accessible shutoff valve.
 An SMC KQ2LE01-00 fitting, a “one-touch” bulkhead elbow fitting on the left side of the
instrument accepts 1/8” tubing for the sample inlet.
Vent


Process vapors are vented through an activated charcoal filter in the middle cabinet.
The middle cabinet has positive ventilation. If any odors are detected (except during periodic
replenishment of chemical consumables), it indicates there may be a minor internal leak. Call for
service.
Environment
 The instrument requires a ventilated space with temperature 5-35 C, humidity <80% (less above
31 C), and access to the front and both sides of the tool (see installation drawing following).
 The space should have some turnover of air with outside (fresh) air, especially in the case of
fumes/odors.
Communication
 For user convenience, the instrument is equipped with external Ethernet and USB ports.
Note: Unless IP67 ethernet or USB connections are made, the instrument loses its NEMA-4X rating
on the top cabinet when these connections are uncapped and in use. Replacing the caps restores the
ratings.
 The instrument has 4 channels of 4-20 mA output available as follows
o Channel 1, Level of dominant species (typically chloroform), connections between +24
VDC and Iout0.
o Channel 2, Level of Total THMs, connections between +24 VDC and Iout1.
o Channel 3, Tool Status, connections between +24 VDC and Iout2.
o Channel 4, Concentration Alarm status, connections between +24 VDC and Iout3.
o AGND on the USB-3102 needs to be connected to -24 VDC (Ground)
 The instrument is capable of email notification of events and results (only if hooked up to the LAN
and allowed external privileges).
Installation and Facilitization
AMS CONFIDENTIAL – PROTECTED BY NDA
The information in this manual is the copyrighted, trade secret material of AMS. Possession and use of this information is restricted by contract; this information is not to be shared with third parties, and must be returned to AMS on demand. ©2010‐2012, all rights reserved.
6
7
8
6
4
5
3
2
1
REVISIONS
DESCRIPTION
ZONE REV.
A DAKOTA CABINET (STARTS DW-5)
STATUS
DATE
REL TO CUSTOMERS01/04/11
APPROVED
APPR
F
F
2X 20 MM CONDUIT
FITTINGS (T&B 9362)
0
(1
40
0.
EXTERNAL RJ45 & USB
CONNECTIONS
.1
6)
13.53 (345)
11.77 (299)
4X
22.25 (565.15)
E
E
D
D
60.00 (1524)
58.00 (1473.2)
3X 1/8" ONE-TOUCH
TUBING FITTINGS
C
D THM-100-001 A
REV
C
B
B
26.71
UNLESS OTHERWISE SPECIFIED:
DO NOT SCALE DRAWING
DIMENSIONS ARE IN INCHES
TOLERANCES:
FRACTIONAL 1/64
ANGULAR: .5
BEND .015
.XX
.01
63
.XXX .005
A
PROPRIETARY AND CONFIDENTIAL NOTICE:
THE DESIGN AND RELATED INFORMATION CONTAINED HEREIN AND/OR
ATTACHED IS THE SOLE PROPERTY OF METARA INCORPORATED. THE
DISCLOSURE OF THIS INFORMATION DOES NOT CONSTITUTE THE RELEASE
OF ANY PROPRIETARY RIGHTS THEREIN. PERMISSION TO REPRODUCE
THIS INFORMATION OR THE PRODUCTS DISCLOSED HEREIN AND/OR
ATTACHED MUST BE OBTAINED IN WRITING FROM AQUA METROLOGY SYSTEMS.
8
7
6
5
4
INTERPRET GEOMETRIC
TOLERANCING PER: ASME Y14.5M-1994
MATERIAL
VARIOUS
DRAWN
NAME
J. GARVEY
CHECKED
ENGR
M. METCALFE VARIOUS
SEE COMPONENT DWGS
3
TITLE:
A
THM MONITORING SYSTEM
MFG
Q.A.
THIRD ANGLE
PROJECTION
TTHM
SIZE
D
FINISH
NEXT ASSY
Aqua Metrology
Systems Limited
DATE
01/04/11
WEIGHT: 75.84
2
REV
DWG. NO.
A
THM-100-001
LB. SCALE: 1:5
SHEET 1 OF 2
1
filename: THM-100 DAKOTA CABINET
8
7
6
4
5
3
2
1
F
F
4.00
4.00
CLEARANCE FOR
VENTILATION
CLEARANCE FOR CONNECTIONS
AND VENTILATION
E
E
D
D
CLEARANCE FOR OPERATOR
AND FOR DOORS TO OPEN
C
DESIRED CLEARANCE FOR
DOOR SWING, NOT REQUIRED
C
.00
D THM-100-001 A
REV
24
B
B
6 0°
Aqua Metrology
Systems Limited
A
TITLE:
A
24.00
THM MONITORING SYSTEM
THIRD ANGLE
PROJECTION
8
7
6
5
4
3
PROPRIETARY AND CONFIDENTIAL NOTICE:
THE DESIGN AND RELATED INFORMATION CONTAINED HEREIN AND/OR
ATTACHED IS THE SOLE PROPERTY OF METARA INCORPORATED. THE
DISCLOSURE OF THIS INFORMATION DOES NOT CONSTITUTE THE RELEASE
OF ANY PROPRIETARY RIGHTS THEREIN. PERMISSION TO REPRODUCE
THIS INFORMATION OR THE PRODUCTS DISCLOSED HEREIN AND/OR
ATTACHED MUST BE OBTAINED IN WRITING FROM METARA INCORPORATED.
DO NOT SCALE DRAWING
TTHM
SIZE
REV
DWG. NO.
D THM-100-001
SCALE: 1:10
1
2
filename: THM-100 DAKOTA CABINET
A
SHEET 2 OF 2