Genesis™ DM Manual

Genesis™ DM Manual
Honeywell HomMed Customer Service
Toll-free: 888-353-5404
Phone: 262-252-5840
Fax: 262-252-6105
Honeywell HomMed Corporate Offices
Honeywell HomMed
3400 Intertech Drive • Suite 200
Brookfield, WI 53045
U.S. and Canada: (888) 353-5404
U.K.: 854 602 9954
Copyright © 2011 by Honeywell HomMed. All rights reserved.
P4510EN.07
2/16/12
GENESIS™ DM MANUAL
Table Of Contents
INTRODUCTION
This manual is a reference document for the clinician to help in the understanding and proficient use of
Honeywell HomMed Genesis™ DM monitors. The information herein supports software versions 4.11.xxA
and later. When you encounter any difficulty using the Genesis DM Monitor, please consult this manual first
to answer your questions and resolve the issue.
Each section listed in the Table of Contents allows the clinician to quickly locate the desired information,
and each section contains extensive details, procedures, and advised methods for achieving the best results
with the Genesis DM monitor. For example, the Programming section contains an extensive programming
chart, a patient Questions Menu, programming tips, and more.
If you encounter any difficulties with the Genesis DM monitor that you cannot resolve, please call
Honeywell HomMed Customer Service at 1-888-353-5404, preferably from the monitor location.
Honeywell HomMed Customer Service will make every attempt to resolve the issue over the phone.
Table of Contents
Getting Started ........................................................................................................1 Safety Information ...................................................................................................1 Equipment Overview ...............................................................................................6 The Genesis DM monitor .............................................................................................................6 Monitor ports ...........................................................................................................................8 Side ports ...................................................................................................................................9 Lithium battery .........................................................................................................................9 Other system equipment ..............................................................................................................9 Scales ..........................................................................................................................................10 Non-invasive Blood Pressure (NIBP) ..................................................................................10 Power Accessories ..................................................................................................................11 Programming the Genesis DM Monitor.................................................................12 Operations and Communications ...............................................................................................12 Genesis DM Menus .........................................................................................................................12 Genesis™ DM Programming Guide....................................................................................13 Setup Wizard....................................................................................................................................15 Setup Wizard Flowchart ................................................................................................................16 Time Menu ........................................................................................................................................19 System Settings ................................................................................................................................23 Questions Menu ..............................................................................................................................26 Password ...........................................................................................................................................28 Transmit Menu .................................................................................................................................28 Option Menu ....................................................................................................................................31 P4510EN.07
2/16/12
i
GENESIS™ DM MANUAL
Table Of Contents
Service Menu ....................................................................................................................................31 Installing the Genesis DM Monitor .........................................................................33 Attaching the scale ..........................................................................................................................34 Connecting a HomMed Bluetooth (wireless) Scale ................................................................35 Attaching the blood pressure cuff ...............................................................................................48 Attaching the modem .....................................................................................................................48 Powering up ......................................................................................................................................49 Monitor Demonstration Mode ....................................................................................................50 Re-starting the Monitor .................................................................................................................50 Patient Training........................................................................................................51 Genesis DM Monitor Voice and Text Prompts .......................................................................51 Voice Prompts ..........................................................................................................................51 Patient Reminders....................................................................................................................55 Disease Specific Symptom Management.............................................................................56 Taking patient vital signs ................................................................................................................56 Concluding the monitoring session ............................................................................................60 Patient-initiated monitoring sessions ..........................................................................................60 Routine Maintenance ...............................................................................................62 Cleaners for the Genesis DM Monitor ......................................................................................63 Cleaning the Genesis DM Monitor .............................................................................................64 Inspection ..........................................................................................................................................64 Calibration verification...................................................................................................................65 Remote upgrades ............................................................................................................................66 Peripheral Medical Devices .....................................................................................67 Health Frontier ecg@home .........................................................................................................68 Glucose Meters ...............................................................................................................................72 Oximeter ...........................................................................................................................................76 Peak Flow Meters ............................................................................................................................78 Prothrombin Meters .......................................................................................................................80 Temporal Temperature .................................................................................................................83 APPENDICES ...........................................................................................................85 Appendix A: Subjective Questions..............................................................................................85 Appendix B: Sleep Apnea Questions ..........................................................................................95 Appendix C: Prompt Text ............................................................................................................96 Appendix D: Genesis DM System Supplies ...............................................................................98 Appendix E: Medical Device Specifications ...............................................................................99 Appendix F: Electromagnetic Guidance .....................................................................................102 P4510EN.07
2/16/12
ii
GENESIS™ DM MANUAL
Getting Started
Getting Started
This section provides important information that you need to know before you start using the
Honeywell HomMed Genesis™ DM Monitor. The information herein supports software versions
4.11.xxA and later.
Indications for use
The Honeywell HomMed Genesis DM Monitor is intended for personal use where use of the system
allows retrospective review of certain physiological functions. The Genesis DM Monitor collects vital
signs data (including noninvasive blood pressure, pulse rate, and weight plus data from optional add-on
devices) then can transmit the data to a central viewing station via a communication network. The
Genesis DM Monitor is intended for use with adult and pediatric patients over twelve years of age.
Contraindications
The Genesis DM monitor is not an emergency medical response device. The monitor does not have
auditory or visual alarms for out-of-limit parameters.
The Genesis DM monitor does not provide real-time, critical-care monitoring of patient vital signs. All
data is transmitted to your monitoring platform, such as to the LifeStream™ platform or to Central
Station 3.5 (Central Station) platform and examined in a retrospective manner by qualified health
professionals.
Who should read this manual
This Honeywell clinician manual has been specifically prepared for those using the Genesis DM Monitor
in healthcare applications, including but not limited to administrative management personnel, clinicians
such as medical care managers and nurses, and technical support staff. The information in this manual
provides the operational concepts of the unit along with the instructions needed to install, program, use,
and maintain the Honeywell HomMed Genesis DM Monitor.
Special conventions
The following table lists the conventions used in this manual.
This convention
Identifies
Function font
Keys that you press on the keyboard
+ sign between key names
Keys that you press simultaneously START + YES
Courier font
Text that displays in the monitor display
Italics
Menu names, publication titles, or references sections of this manual
Caution or warning
A note that provides an important piece of information
A tip that will help you work faster or more efficiently
P4510EN.07
2/16/12
1
GENESIS™ DM MANUAL
Getting Started
Content overview
The following chart provides a brief overview of the sections in the manual.
Section
Overview
Getting Started
Describes what you can expect to find in this manual and provides
important safety and customer support information
Equipment Overview
Explains and illustrates each component of the Genesis DM.
Programming the Genesis DM monitor
A chart of Genesis DM programming menus, explains programming
concepts, and provides step-by-step programming instructions
Installing the Genesis DM monitor
Explains and illustrates how to setup the Genesis DM monitor
Patient Training
Important information about how the monitor works; explains how to
teach patients to use equipment properly
Routine Maintenance
Instructions for routine maintenance and equipment calibration
Peripherals
Information and instructions for using compatible peripheral medical
devices
Appendices
Additional reference information
Honeywell HomMed Customer Service
The goal of Customer Service is to support your agency and clinical staff in meeting the needs of your
patients. We are available 24/7/365, and we will work quickly to resolve issues with you as they arise.
When you call, please have this manual and a copy of the Genesis DM Programming/Troubleshooting Guide
handy. We will walk you through various troubleshooting techniques to assist you in identifying causes
and resolutions.
Toll-free:
Phone:
Fax:
888-353-5404
262-252-5840
262-252-5795
A Customer Service Representative will make every attempt to resolve the issue over the phone. However,
if equipment must be sent back, we will issue a Return Material Authorization (RMA). For shipping
instructions, please refer to “Shipping equipment back to Honeywell HomMed” on the next page.
Repairing equipment
Honeywell HomMed recommends that authorized repair centers make all repairs for the monitor and
accessories. Repairs made by unauthorized personnel will invalidate your warranty. For product
warranty information, please consult with your Honeywell HomMed contact person.
If any equipment needs repair, contact a Customer Service Representative and we will be happy to
troubleshoot the problem with you to determine the appropriate correction. Please be prepared to
identify the equipment (monitor, scale, or accessory) and describe the problem in detail. If the damaged
item is a monitor, scale, pulse oximeter, or GPRS device, please provide the model and serial
number to the Customer Service Representative when you call.
P4510EN.07
2/16/12
2
GENESIS™ DM MANUAL
Getting Started
Model and serial number locations:



Monitor – model and serial numbers are located on the bottom of the monitor.
Scale – model and serial numbers are located on the side or bottom of the scale.
Peripheral medical devices – contact the manufacturer of the device.
NOTE: The information in this manual supports software versions 4.11.xxA and later.
De-infestation service
If your equipment also requires de-infestation service, Honeywell HomMed Service will provide you with
proper containment and transportation instructions. There is an additional charge for de-infestation service.
An insect resistance retrofit kit is also available; refer to publication P4043 or contact Customer Service.
DO NOT attempt to disassemble or open the DM Monitor at any time – you will void the warranty and
possibly damage the Monitor.
Shipping equipment back to Honeywell HomMed
If Customer Service requests that you ship equipment back to Honeywell HomMed, please follow the
directions below.
1.
Carefully pack the equipment in its original box with its original protective packaging materials. If
you do not have the original box, use any sturdy box with at least one-inch clearance around the
equipment for protective packing materials. Pack securely to help avoid damaging the unit.
2.
When Customer Service gives you a Return Material Authorization (RMA) number, include the
RMA number on the outside of the shipping container. Ship to the following address:
RMA # _______________
Customer Service Department, Repairs
Honeywell HomMed
3825 Ohio Ave.
St. Charles, IL 60174
CE Mark
Marking by the symbol
2007/47/EC.
0413
indicates compliance of this device to the Medical Device Directive
Authorized Representative:
MDSS GmbH
Schiffgraben 41
30175 Hannover
Germany
When you set up a monitor for demo use
When you use a Genesis DM monitor to demonstrate the monitor, the potential exists for demo data
to transmit to an actual patient record. LifeStream holds data packets for 24 hours, attempting to find
the patient that matches the monitor’s serial number. If you use a monitor to transmit demo data and
P4510EN.07
2/16/12
3
GENESIS™ DM MANUAL
Getting Started
then assign the monitor to a patient within the 24-hour window, LifeStream will treat the demo data as
that patient’s data. To prevent posting demo data to a patient record:
In Transmit Mode, set the monitor to Local Gateway so that the Genesis DM monitor does not
transmit a data packet.
Additional references
Telemonitor Programming/Troubleshooting Guide. Honeywell HomMed, 2009.
Genesis™ DM Monitor Quick Setup Guide. Honeywell HomMed, 2007.
Genesis™ Roach Resistance Retrokit (P4043). Honeywell HomMed, 2008.
LifeStream™ Manual. Honeywell HomMed, 2009.
Central Station 3.5 Manual. Honeywell HomMed, 2005.
For more information about Honeywell HomMed authorized peripheral devices, refer to the
appropriate manufacturer’s material.
Phone Service
The DM with at least 4.10.11A firmware works with the types of phone service and companies as
listed in the table below.
NOTE: Telephone companies, providers, and services undergo frequent change.
Please consult your telephone service provider with any questions about your type of service.
If you have questions, are unsure, or otherwise can not establish a phone connection while
attempting to allow the DM Monitor to transmit, contact Honeywell HomMed Customer Service.
Service Type
Examples
Status
The Bell System
Works
Normal Phone also using DSL
Internet
SOME but not all of the
following:
AT&T
Verizon
Embarq
. . . and many others
Works with DSL Filter between monitor
and wall jack
Hybrid VOIP*
Cable Companies such as:
Comcast
Charter
Time Warner (Road Runner)
Bright House
Qwest
Cox
Frontier
AT&T
. . . and many others
Works 4.10.11A and newer firmware
Normal Phone Line
P4510EN.07
2/16/12
4
GENESIS™ DM MANUAL
Getting Started
Service Type
Full VOIP
PBX , and full digital phone lines
Examples
Status
Services such as:
Vonage
Magic Jack
Skype
Verizon FiOS
. . . and many others
Will NOT work
Some office and some hotel
Will Not Work
phones/lines/systems
*WARNING* could damage the modem
*VOIP – Voice over Internet Protocol (VOIP) signifies a wide range of transmission technologies for delivering
voice communications over a network such as the Internet or other packet-switched networks. Other terms
frequently encountered and synonymous with VOIP include, but are not limited to, IP telephony, Internet
telephony, voice over broadband (VoBB), broadband telephony, and broadband phone.
Internet telephony refers to communications services—voice, facsimile, and/or voice-messaging applications—that are
transported via the Internet, rather than the public switched telephone network (PSTN). The basic steps involved
in originating an Internet telephone call are conversion of the analog voice signal to digital format and
compression/translation of the signal into Internet protocol (IP) packets for transmission over the Internet; the
process is reversed at the receiving end. VOIP systems employ session control protocols to control the
establishment (set-up) and termination (tear-down) of calls as well as audio “codecs” that encode speech, allowing
transmission over an IP network as digital audio via an audio stream. Codec use is varied between different
implementations of VOIP (and often a range of codecs are used); some implementations rely on narrowband and
compressed speech, while others support high fidelity stereo codecs
P4510EN.07
2/16/12
5
GENESIS™ DM MANUAL
Safety Information
Safety Information
To prevent injury to yourself or damage to any equipment, please read and observe all the safety
information in this section before you install the Genesis DM Monitor.
In addition, please read the safety and product materials supplied by individual manufacturers for all
authorized peripheral equipment.
The following chart explains the hazard conventions used in this manual.
Term
!
!
Level of
risk
Definition
WARNING
Major
Could cause death or serious personal injury
(serious personal injury implies permanent
impairment or any injury that requires medical or
surgical intervention to preclude permanent
impairment)
CAUTION
Moderate
May result in moderate or minor personal injury
CAUTION
Minor
May result in equipment damage
NOTE
None
Important information
!
WARNING
ELECTRICAL SHOCK
Computer components contain sufficient voltage to cause death or serious injury.
DO NOT immerse in water or other liquid.
DO NOT disassemble.
!
WARNING
NOT AN EMERGENCY RESPONSE DEVICE
The Genesis DM monitor is not intended to be an emergency response device. The
monitor does not alarm and it does not transmit alarms,
In case of a medical emergency, patients should call local Emergency
Medical Service and/or their health care provider.
P4510EN.07
2/16/12
1
GENESIS™ DM MANUAL
Safety Information
!
WARNING
EXPLOSION HAZARD
This unit is powered with electricity and could ignite highly flammable gases.
DO NOT use this device in the presence of explosive or flammable
anesthetic agents.
!
WARNING
NOT A NEONATAL UNIT
Using the Genesis DM monitor on infants or small children could cause serious injury.
DO NOT use the monitor, temperature probe, blood pressure cuff, or any
monitor accessory on infants or small children.
!
WARNING
NOT DEFIBRILLATOR PROOF
Using the Genesis DM monitor on patients undergoing defibrillation may damage the
monitor or peripheral equipment.
DO NOT use the monitor on patients undergoing defibrillation.
!
WARNING
PHYSICAL HAZARD
Please exercise care in the routing of external cables, wires and hoses to avoid creating
a tripping hazard.
DO NOT route cables across a walkway
!
CAUTION
USE ONLY HOMMED AUTHORIZED MEDICAL EQUIPMENT
Attaching unauthorized equipment to the Honeywell HomMed Genesis DM monitor
may cause damage or equipment failure, including increased emissions or decreased
immunity of the Genesis DM.
Vital sign measurements may be inaccurate if unauthorized equipment is used with the
Genesis DM monitor.
DO NOT attach unauthorized medical equipment to the Genesis DM
monitor. When connecting this device to any instrument, verify proper
operation before clinical use. Refer to the equipment's user manual for full
instructions.
Any peripheral medical devices connected to the data interface must be certified
according to the respective IEC standards: IEC 60950 for data-processing equipment or
IEC 60601-1 for electromedical equipment.
All combinations of equipment must comply with IEC 60601-1-1 systems requirements.
IEC 60950 approved equipment must be placed outside of the “patient environment.”
The patient environment is defined as an area 1.5 meters (4.92 feet) from the patient.
For a list of Honeywell HomMed equipment and compatible peripheral medical devices,
refer to Appendix E.
P4510EN.07
2/16/12
2
GENESIS™ DM MANUAL
Safety Information
CAUTION
EXPOSURE TO RADIO FREQUENCY RADIATION
USA – FCC:This device must not be co-located or operating in conjunction with any
other antenna or transmitter.
Canada – IC:The installer of this radio equipment must ensure that the antenna is
located or pointed such that it does not emit RF field in excess of Health Canada limits
for the general population; consult Safety Code 6, obtainable from Health Canada’s
website www.hc-sc.gc.ca/rpb.
CAUTION
SENSITIVE EQUIPMENT
The monitor contains sensitive components. Liquids or chemicals will seriously damage
the unit.
DO NOT autoclave or sterilize with ethylene oxide.
DO NOT immerse any equipment in liquid.
If liquid is spilled on the monitor, or if the unit has been dropped or damaged, contact
Honeywell HomMed Customer Service.
CAUTION
REQUIRES ANALOG PHONE LINE
The modem requires an analog phone line. A digital phone line will destroy the modem.
DO NOT use a digital phone line. Verify presence of analog phone line prior
to installation.
CAUTION
EQUIPMENT DAMAGE
Connecting the Honeywell HomMed scale or Fairbanks scale to any device not
mentioned in this manual could severely damage the scale or the connected equipment.
DO NOT connect the Honeywell HomMed scale to any device other than
the Honeywell HomMed Classic, Sentry, or Genesis DM monitors.
CAUTION
BLOOD PRESSURE HOSE DAMAGE
Bending, kinking, or otherwise restricting the blood pressure hose can damage the hose
or interfere with readings.
DO NOT bend or crimp the blood pressure hose. Keep the hose free from
furniture or other objects that could compress them.
P4510EN.07
2/16/12
3
GENESIS™ DM MANUAL
Safety Information
CAUTION
ELECTROMAGNETIC COMPATABILITY
Electromagnetic compatibility of electrical equipment at very close distances to the
Genesis DM has not been evaluated.
DO NOT use the Genesis DM adjacent to or stacked with other equipment.
If it is necessary to do so, observe the monitor and verify normal operation
prior to use with the patient.
CAUTION
KEYPAD DAMAGE
Sharp or pointed instruments may permanently damage the keypad.
DO NOT press keys with any instruments. Use your fingers to press
Honeywell HomMed Monitor keys.
CAUTION
SPILLAGE
Liquid spilled onto the monitor may cause damage to the monitor and may present a
safety hazard to the user/patient.
Should this monitor become wet, disconnect the power cord, wipe off all
moisture and allow sufficient time for drying before re-connecting power
and operating the monitor.
CAUTION
INCORRECT AC POWER RATINGS CAN DAMAGE MONITOR
Incorrect AC power ratings could damage the Genesis DM monitor.
Verify that the AC ratings noted on the power supply label match the AC
ratings at your installation site. If the site ratings do not match the power
supply, do not use the monitor. Contact Honeywell HomMed Customer
Service.
CAUTION
INCORRECT CONNECTION SEQUENCE CAN DAMAGE THE MONITOR
Plug the power supply into the Genesis DM Monitor first and into the power source
last. Failure to do so could cause damage to the monitor and/or power supply.
Verify that the power supply is NOT plugged into an AC source before
connecting the monitor. Contact Honeywell HomMed Customer Service if
you are unsure or have any questions about the correct power-up method.
P4510EN.07
2/16/12
4
GENESIS™ DM MANUAL
Safety Information
NOTES













Operation of this device may be adversely affected in the presence of strong EMI or RF sources,
such as electrosurgery equipment.
Operation of this device may be adversely affected in the presence of computed tomography (CT)
equipment.
Ability to obtain blood pressure measurement may be impacted by patients with tremors.
Performance with common arrhythmias, such as atrial or ventricular premature heartbeats or atrial
fibrillation, has been verified by use of a patient simulator. Blood pressure measurements may not be
representative for patients experiencing mild or moderate to severe arrhythmias.
Dyes introduced into the bloodstream, such as methylene blue, indocyanine green, indigo carmine,
patent blue V (PBV), and fluorescein, may cause an inability to determine relevant SpO2 readings.
Any condition that restricts blood flow, such as use of a blood pressure cuff, extremes in systemic
vascular resistance, or low perfusion may cause an inability to determine relevant pulse.
Significant levels of dysfunctional hemoglobins, such as carboxyhemoglobin or methemoglobin, will
affect the accuracy of the SpO2 readings.
SpO2 is not for continuous monitoring (no alarms).
SpO2 measurements may be adversely affected in the presence of high ambient light. If necessary,
shield the sensor area.
SpO2 measurements may be adversely affected if the patient wears fingernail polish or false nails. If
necessary, remove polish or nails to obtain a reading.
Dashes displayed in any parameter indicate the measurement is invalid or unavailable.
To prevent cross-contamination, clean the monitor and all accessories prior to assigning equipment
to a new patient. Please see cleaning procedures located in this manual.
The Honeywell HomMed scale and the Fairbanks scale are not stand-alone devices. The scale must
be connected to a Honeywell HomMed monitor to operate.
.
P4510EN.07
2/16/12
5
GENESIS™ DM MANUAL
Equipment Overview
Equipment Overview
When you receive your shipment from Honeywell HomMed, carefully remove each piece of equipment
from the box. Save the carton and packing materials in case you need to return any of the contents. As
you inventory the contents of the box, check each piece of equipment for damage. If anything is
damaged or missing, contact Customer Service immediately. Your package should include the
following items:







Genesis DM monitor
Scale and scale cable
Blood pressure cuff and hose
Phone cord and duplex jack (not shipped to the UK)
AC adapter (power supply)
Power strip (not shipped to the UK)
External Pulse Oximeter and Finger Probe (optional)
The Genesis DM monitor
The front of the monitor has a vacuum fluorescent display (VFD) surrounded by six function keys. Each
key is labeled and color-coded for easy visual identification. When you press the keys correctly, the
monitor beeps. The volume of the beep matches the volume of the voice prompts.
The panel displays messages, bar graphs, and menus in a light font against a dark background for easy
reading.
If programmed for use by only one patient, the current day, date, and time display randomly over the
display face when the monitor is idle. If the monitor is programmed for multi-patient use, then a prompt
to swipe an ID card displays instead of the current time.
P4510EN.07
2/16/12
6
GENESIS™ DM MANUAL
Equipment Overview
The figure below and the following chart explain the monitor features and key functions.
#
Feature
Function
1
BP
2
Green START key
3
Red STOP key
4
Gray MANUAL key
5
WT
6
Blue RETEST key
7
Black NO key
Black NO key records a “no” answer; DOWN scroll key in Setup Mode
8
White YES key
White YES key records a “yes” answer; UP scroll key in Setup Mode
Enters/exits Setup Mode (pressed with START key)
9
BPM
BPM displays the pulse rate. If optional SpO2 is attached, a pulse strength bar
appears next to the pulse rate number during SpO2 collection.
10
SPO2
Percentage of oxygen saturation (%SpO2) – displays the oxygen saturation measurement value
(for optional device)
Displays the systolic and diastolic pressure readings in mmHg
Green START key starts a blood pressure reading;
Enters/exits Setup Mode (pressed with YES key)
Red STOP key stops blood pressure reading and deflates the cuff
Gray MANUAL key enters Manual Mode; Exits a menu in Setup Mode
Displays measured weight from the scale
Blue RETEST key starts an unscheduled monitoring session;
Selects a menu option in Setup Mode
After transmitting vitals, pressing MANUAL twice clears vitals and returns the display to idle mode. If
MANUAL is not pressed the vitals will be presented for 5 minutes and then cleared automatically.
P4510EN.07
2/16/12
7
GENESIS™ DM MANUAL
Equipment Overview
Monitor ports
There are numerous ports located on the rear and left side of the Monitor that connect to devices
including the scale, non-invasive blood pressure (NIBP), optional accessories, and peripheral medical
devices. The figure below shows the ports located on the rear of the monitor.
Genesis DM Rear Panel
(1) Modem connection
The modem connection is for a standard (RJ11) phone cord that runs from the monitor to a duplex
phone jack and allows the modem to dial and transmit data using a phone line.
(2) GPRS Communications / Non-isolated communication port (COM-A)
GPRS Communications
In addition to the standard modem, the Genesis DM monitor will support a wireless GPRS modem that
will automatically connect to available secure wireless networks without additional client setup. In the
event that the unit fails to connect automatically, the user will receive a relevant error message.
The GPRS modem is external, features a right angle antenna, and connects to the monitor unit with a
supplied cable to port COM A.
The wireless GPRS modem can ONLY be connected to the COM A port.
The wireless GPRS MUST be allowed to power up for a minimum of 20 seconds before the
monitor will recognize it.
Non-isolated communication port (COMA)
This is a non-isolated port with a Mini-DIN 6-pin connector that can supply up to +5VDC @ 250mA to
an external device. This port is for external devices that only transmit data. Patients should never
attempt to use a device connected to the monitor via a non-isolated port.
(3) Power jack
The power jack is the jack for the AC adapter which powers the monitor.
P4510EN.07
2/16/12
8
GENESIS™ DM MANUAL
Equipment Overview
(4) Isolated serial I/O ports (SCALE, COM1, COM2)
The three isolated ports with Mini-DIN 4 connectors supply up to +5VDC @ 250mA.
The first isolated port is a dedicated scale COM port. COM1 and COM2 allow a patient to use an
external device connected to the monitor. Isolation protects patients from any potential of shock.
Connector cables for external devices only fit the port appropriate for the device.
Side ports
The following figure shows the ports located on the side of the Genesis DM monitor.
(1) Genesis DM SmartMedia Card slot (NOTE: The card slot is not featured on all Genesis DM units.)
(2) Non-invasive blood pressure (NIBP) pneumatic port
The NIBP pneumatic port, with its quick-release connector, is the attachment site for the blood
pressure hose.
Lithium battery
The lithium battery maintains the memory, clock and configuration for approximately five years.
You cannot replace the lithium battery.
If the battery fails, contact Customer Service to have it replaced.
Other system equipment
In addition to the monitor, other equipment can include a scale with a scale cord or a wireless scale,
blood pressure cuff and hose, and power accessories. The SpO2 finger sensor with external oximeter
P4510EN.07
2/16/12
9
GENESIS™ DM MANUAL
Equipment Overview
module cable is an optional accessory. This section provides a general overview of each piece of
equipment. For detailed specifications, refer to Appendix D, Equipment Specifications.
Scales
There are two basic types of scales: a Fairbanks scale (used with product bearing CE Marking) and the
Honeywell HomMed scale in regular and Bluetooth wireless versions. The scales have a capacity of 500
lb (227 kg) and collect weight using four load cells, one in each corner of the scale. The accuracy of the
weight is +/- 0.5 lb (0.2kg). The Genesis DM supplies power and communication to the scale through
the scale cord.
NOTE: Only the Fairbanks scale is available in Europe.
Gravity Adjustment: The scale is adjusted for gravity in the location to which HomMed is shipping it.
Any relocation to a different gravity zone requires re-verification by HomMed or a European local
authority. Any damage or recalibration may also require re-verification. The following table identifies
distinguishing features of each scale. The Bluteooth scale does NOT have adjustable feet.
Feature
Size
Shape
Surface texture
Feet(excluding Bluetooth)
Scale cord
Fairbanks
Approximately 13.5" by 13.5"
Square
Linear textured pattern
Screw in/out; removable
(excluding Bluetooth)
4-pin MiniDIN connector to monitor
and tabbed jack to scale
Honeywell HomMed
Approximately 15.75" by 14"
Curved
Round textured pattern
Screw in/out: not removable
(excluding Bluetooth)
4-pin MiniDIN Blue to monitor and DB9 serial connector to
scale
Non-invasive Blood Pressure (NIBP)
The NIBP module inside the monitor measures systolic pressure and diastolic pressure. Blood pressure
measurements determined with this device are equivalent to those obtained by a trained observer using
the cuff/stethoscope auscultatory method within the limits prescribed by the AAMI SP10, Electronic or
Automated Sphygmomanometers. A copy of the Report Study Findings performed for the AAMI SP10
evaluation is available upon request. Pressures are calculated by a proprietary algorithm as follows:
Systolic:
The cuff pressure when an increase in oscillations is perceived
Diastolic:
The cuff pressure when oscillations are no longer decreasing
as pressure is released from the cuff
Note: Pulse rate (BPM) is also collected and transmitted with the blood pressure reading. The pulse rate is
displayed on the monitor if the oximeter is NOT programmed. If the oximeter is programmed and the pulse
cannot be obtained through the oximeter, the unit will display the pulse, sensed through the NIBP cuff.
SYSTEM
Mm Kg
BPM
79
–––/–––
%Bp02
94
0.0LBS
P4510EN.07
2/16/12
10
GENESIS™ DM MANUAL
Equipment Overview
The display also includes a pulse strength indicator:
The pulse strength indicator rises and falls with each pulse sensed by the system.
SYSTEM
Mm Kg
BPM
79
–––/–––
%Bp02
94
0.0LBS
The blood pressure cuff is a D-ring style cuff that complies with SP10 specifications for cuff length and width.
Cuffs are available in the following sizes:
Size
Fits limb size
Adult I
16 - 22 cm (6 - 8.5 inches)
Adult II
21 - 29 cm (8 - 11 inches)
Adult III
28 - 37 cm (11 - 14.5 inches)
Adult IV
36 - 46 cm (14 - 18 inches)
Note: Proper cuff size is a variable affecting the accuracy of NIBP measurements. To order a different
sized cuff, contact Customer Service (888) 353-5404.
Power Accessories
The AC adapter converts alternating current to direct current to power the monitor. The adapter is
CSA listed to IEC 60601-1. The power strip provides four (4) outlets.
P4510EN.07
2/16/12
11
GENESIS™ DM MANUAL
Programming the Genesis DM Monitor
Programming the Genesis DM Monitor
Operations and Communications
To program the Genesis DM monitor for patient use, you must enter Setup Mode and navigate through
the menus and menu options to customize patient settings. This section provides information about
communication providers, menus, and programming instructions, PIN numbers, and phone numbers.
Your Clinical Consultant will explain the best transmit mode for your area and give you the required
PIN and phone numbers. When you program a monitor for a patient, you should go through each menu
to verify that the entered numbers match the numbers given to you during your in-service.
Genesis DM Menus
The Genesis DM monitor displays menu names and options in the display face. Because of the limited
space, you can only see one menu or menu option at a time as you scroll through available choices.
Entering Setup Mode opens the Main menu which offers eight menu options: Time Menu, System
Settings, Questions Menu, Password, Transmit Menu, Option Menu, Service Menu, and Exit
Setup. Each menu, with the exceptions of Exit Setup and Password, has sub-menus with additional
options. Some menu options are password protected. If you do not enter your password, you will not
see password-protected sub-menus.
The following table shows the menu layout. Password-protected options are noted with (C).
P4510EN.07
2/16/12
12
GENESIS™ DM MANUAL
Programming the Genesis DM Monitor
Genesis™ DM Programming Guide
P4510EN.07
2/16/12
13
GENESIS™ DM MANUAL
Programming the Genesis DM Monitor
The Genesis DM monitor is fully compatible with Central Station 3.5 ONLY when Classic mode
has been activated. This mode is accessed during programming and eliminates conflicts with
LifeStream specific content. See the specific programming instructions to activate Classic mode
compatibility.
A configuration packet differs from a vital signs packet or a firmware upgrade packet. A configuration packet
is the set of Genesis DM data that describes the user’s vitals signs session.
For instance, the monitor may prompt the user in Canadian French three (3) times a day, 8:00 AM, 12:00
PM, and 4:00 PM. During the 8:00 AM session, the user is prompted for blood pressure, weight, and glucose
using a LifeScan meter. At 12:00, the user is prompted for blood pressure and weight only. Finally, at 4:00
PM, the user is prompted for blood pressure, weight, and prothrombin time using the ITC meter. All of this
data is found in the configuration packet.
A configuration packet can be sent from the Genesis DM monitor or the LifeStream product.
After a change is made to the Genesis DM at the monitor, the user is prompted to send a
configuration packet back to LifeStream. If you haven’t made a change to the Genesis DM, but still
wish to send a configuration packet to LifeStream, press MANUAL then YES.
NOTE: When the wireless scale is attached to the monitor to send a packet, press MANUAL, answer
NO to taking a weight measurement, then press YES.
Alternately, a change made at LifeStream will be sent to the customer’s mailbox for download by
the Genesis DM the next time it communicates to the mailbox. If you are at the Genesis DM and
are expecting that there is a configuration packet waiting, press MANUAL then YES.
NOTE: When the wireless scale is attached to the monitor to send a packet, press MANUAL, answer
NO to taking a weight measurement, then press YES.
Factory Default Reset
Users of the Genesis DM monitor will be able to easily reset the monitor to the original factory default
settings, such as agency PIN, default language, transmit settings, and time zone. This is especially helpful
when you need to reassign the monitor to another patient, and want to remove any custom settings
established for another patient. At any time following start up, the user will be able to reset the unit by:
1.
On the idle screen, press START. The Display will align at the center.
2.
When the Display is aligned at the center, press MANUAL. This brings up the Confirmation
Screen for Loading Factory Defaults.
NOTE: If YES is pressed, the factory defaults are restored. If NO is pressed, or after a timeout of
three (3) seconds, the operation quits and the system returns to the Idle Screen.
P4510EN.07
2/16/12
14
GENESIS™ DM MANUAL
Programming the Genesis DM Monitor
Setup Wizard
The Genesis DM monitor will launch an automatic setup wizard each time it is plugged in. The wizard
initially asks if you want to begin setup: responses include Yes, No or allowing it to timeout. The setup
wizard presents several screens that allow users to adjust settings for the monitor. The display will
prompt the user, and if a response is not received within 30 seconds, the wizard will be bypassed. The
unit will ALWAYS bypass the setup wizard when restarting following a firmware upgrade session.
The Setup Wizard Flowchart on the next page provides a graphical overview of the Setup Wizard.
P4510EN.07
2/16/12
15
GENESIS™ DM MANUAL
Programming the Genesis DM Monitor
Setup Wizard Flowchart
P4510EN.07
2/16/12
16
GENESIS™ DM MANUAL
Programming the Genesis DM Monitor
Setup Workflow
When the monitor starts up you will have the option of entering the start up wizard. The basic
workflow is as follows.
1.
2.
Initially, the monitor detects communications availability and displays “Detecting Comms”
o
If no communications are found, the user is prompted to check the modem connection
or plug in the GPRS unit (wait at least 20 seconds after powering the unit on for the
modem to initialize).
The user is then prompted to try the detection cycle again.
o
If communications are found then the monitor displays a message indicating that
communications have been found for either the GPRS unit or the phone line and
continues on to retrieve data.
Retrieve “Config Data” begins
o
3.
If no configuration data is received then the monitor continues on to auto-detect
peripherals.
After the peripheral detection sequence, the monitor prompts the user to make a voice setting
selection. The monitor then sends the final configuration to the monitoring application.
You can enter Setup Mode anytime the Genesis DM monitor is idle. However, if you are in Setup Mode
during a scheduled report time, the monitor will skip the monitoring session and will not transmit any
information. Your monitoring application will post a “Transmit Error” (orange alert) for the missed
reading.
To enter Setup Mode manually, quickly press and release START and YES simultaneously. Do not
hold the keys.
Setup Mode normally opens to the Time menu. However, if you press and hold instead of press and
release, you may scroll to a different menu.
Function keys
Once in Setup Mode, you can use the keys on the front panel to enter or exit menus, scroll up or down,
and select menu options. For an objective view, refer to the Programming Guide on page 15.
The following table summarizes the functions for each key or key combination:
START + YES
MANUAL
Enter/exit Setup Mode
 Back
Exit current menu or menu option
YES

Scroll forward/up through numbers, menu options, or menus
NO

Scroll backward/down through numbers, menu options, or menus
RETEST
P4510EN.07
Enter
Enter a menu, accept a displayed value
2/16/12
17
GENESIS™ DM MANUAL
Programming the Genesis DM Monitor
MANUAL button
The MANUAL button allows you to back out one level from wherever you are. It is handy when you
want to quickly back out of a menu without scrolling to an exit option. For example, if you enter a menu
option and then realize that you are in the wrong menu, you can press MANUAL to back out. Keep in
mind however, that pressing MANUAL just backs up one level.
So if you are in Report Time #1 setting a value (a value is bracketed) and press MANUAL, the
displayed values are accepted and you back up or return to Report Time #1. If you press MANUAL a
second time, and you return to the Main menu.
Exiting menus and Setup Mode
Referring to the menu layout, you can see that the Main menu as well as every sub-menu has an Exit
option. The Exit option takes you out of the current menu and returns you to the Main menu. The
Exit Setup Menu in the Main menu takes you out of Setup Mode and returns the monitor to idle
mode.
To immediately exit Setup Mode at any time, press START + YES.
Scheduling frequency
You may schedule how often Genesis DM collects weight, questions, and optional device data.
Scheduling options include every time, one time per day, or not at all (off).
Once-a-day collection
If you choose one time per day, Genesis DM collects the scheduled data during the first monitoring
session initiated by a scheduled Report Time or by pressing RETEST. Once the scheduled data is
collected, the monitor will not collect it again until the next day. Genesis DM resets to a new day at
midnight.
If the monitor times out without a patient response (non-compliance), it will attempt to collect the
scheduled data the next monitoring session. Timing out or even powering down does not affect the
once-a-day schedule.
P4510EN.07
2/16/12
18
GENESIS™ DM MANUAL
Programming the Genesis DM Monitor
Time Menu
Enter Setup Mode and, if necessary, scroll until Time Menu displays on the monitor. Press RETEST
to enter the Time menu. Use YES or NO to scroll to the Time Menu option you wish to enter.
Report Time (1 – 4)
Options:
Time (in selected format), and Off
Default:
Time 1 defaults to 8:00; all other times are set to Off
Access Level:
General
The Report Time is the scheduled time for the Genesis DM monitor to collect and transmit patient
vitals. Patients can have up to four report times in one 24-hour period. If you set more than one time,
make sure that you schedule report times at least one hour apart to accommodate the Scheduled
Reading window in your monitoring application. For more information about scheduled readings, refer
to the manual for your application, such as the LifeStream User Manual.)
The Report Time displays in the format programmed in the Time Format menu (a.m. /p.m. or 24-hour
format).
If you assign multi-use status to a monitor, make sure all of the monitor’s Report Times are set to Off.
In the Time Menu, scroll to the Report Time option that you wish to set (1, 2, 3, or 4).
1.
Press RETEST to enter. Depending on the current setting, either On or Off is bracketed and the
other choice of the two may be selected.
2.
To change the setting, press either YES or NO.
3.
Press RETEST to accept the setting. The hour is bracketed.
4.
Scroll to the correct hour. Press RETEST to accept the hour setting. The minutes are bracketed.
5.
Scroll to the correct minutes. Press RETEST to accept the minute setting.
You can scroll to the next Time Menu option, press MANUAL to return to the Main menu, or
press START + YES to exit Setup Mode.
Report Time # (1- 4) Days
Options:
Every day or any combination of days
Default:
Every day (Su Mo Tu We Th Fr Sa)
Access Level:
General
P4510EN.07
2/16/12
19
GENESIS™ DM MANUAL
Programming the Genesis DM Monitor
Each report time is followed by this option that allows you to set a report time frequency. The default is
daily. However, you can program any combination of days.
1.
In the Time Menu, scroll to the Report Time # (1-4) Days option that you wish to set (1, 2, 3,
or 4).
2.
Press RETEST to enter. The first setting is bracketed.
3.
To change the setting, press either YES or NO, and then press MANUAL.
–OR-
1.
Press RETEST to pass the first setting and move to the next setting.
2.
Repeat step three until the days are completely set.
You can scroll to the next Time Menu option -or- press MANUAL to return to the Main menu, or
press START + YES to exit Setup Mode.
Time
Options:
Time in the format set in Time Format menu
Default:
Present time for your time zone
Access Level:
General
The Time menu allows you to set the correct time. If you need to reset the time, follow the steps
below.
1.
In the Time menu, scroll to the Time: option and press RETEST. The hour is bracketed.
2.
Scroll to the correct hour. Press RETEST to accept the hour setting and move to minutes.
3.
Scroll to the correct minutes. Press RETEST to accept the minute setting.
You can scroll to the next Time Menu option -or- press MANUAL to return to the Main menu, or
press START + YES to exit Setup Mode.
P4510EN.07
2/16/12
20
GENESIS™ DM MANUAL
Programming the Genesis DM Monitor
Date
Options:
Day of the week, month, day, and year (displays according to selected date format)
Default:
Today’s day and date
Access Level:
General
The Date menu allows you to set the correct date. If you need to change the date (MM/DD/YY format
only), follow these steps:
1.
In the Time Menu, scroll to the Date option and press RETEST to enter the menu. The day
of the week is bracketed.
2.
Scroll to the correct day of the week. Press RETEST to accept the day. The numeric month is
bracketed.
3.
Scroll to the correct month. Press RETEST to accept the setting and move to the numeric day.
4.
Scroll to the correct day. Press RETEST to accept the setting and move to the year.
5.
Scroll to the correct year. Press RETEST to accept the setting.
You can scroll to the next Time Menu option –or- press MANUAL to return to the Main menu, or
press START + YES to exit Setup Mode.
Time Format
Options:
a.m./p.m. or 24-hour format
Default:
a.m./p.m.
Access Level:
General
You can set the monitor to display either 24-hour or a.m./p.m. time. To change the format, follow the
steps below.
1.
In the Time Menu, scroll to Time Format, and press RETEST to enter. The currently programmed
format is bracketed.
2.
Press either YES or NO to change the format, and press RETEST to accept the setting.
You can scroll to the next Time Menu option -or- press MANUAL to return to the Main menu, or
press START + YES until you exit Setup Mode.
P4510EN.07
2/16/12
21
GENESIS™ DM MANUAL
Programming the Genesis DM Monitor
Date Format
Options:
MM/DD/YY (US) or DD/MM/YY (Europe)
Default:
MM/DD/YY
Access Level:
General
The Date Format allows you to set the Genesis DM monitor to a United States format of
month/day/year, or to a European format of day/month/year.
1.
In the Time Menu, scroll to Date Format, and press RETEST to enter. The currently programmed
format is bracketed.
2.
Press either YES or NO to change the format, and press RETEST to accept the new
setting.
You can scroll to the next Time Menu option -or- press MANUAL to return to the Main menu, or
press START + YES until you exit Setup Mode.
Daylight Savings
Options:
USA, EEC, OFF
Default:
Set to locale
Access Level:
General
If the monitor is powered on, Daylight Savings automatically adjusts the clock to accommodate the time
change. The USA default setting moves the clock forward one hour at 2:00 a.m. on the second Sunday in
March and back one hour at 2:00 a.m. on the first Sunday of November. The European setting (EEC)
moves the clock forward one hour at 1:00 a.m. on the last Sunday in March and back one hour at 1:00
a.m. on the last Sunday of October. If your locale does not have Daylight Savings, turn this setting to
OFF.
1.
In the Time Menu, scroll to Daylight Savings, and press RETEST to enter. The currently
programmed format time is bracketed.
2.
Press either YES or NO to change the format, and press RETEST to accept the setting.
You can scroll to the next Time Menu option -or- press MANUAL to return to the Main menu, or
press START + YES until you exit Setup Mode.
P4510EN.07
2/16/12
22
GENESIS™ DM MANUAL
Programming the Genesis DM Monitor
System Settings
After entering Setup Mode, press YES to scroll to System Settings, and then press RETEST to enter
the System Settings menu.
Volume
Options:
Lowest, Low, Med, High, and Highest
Default:
Low
Access Level:
General
The volume setting controls both the key beep and the voice prompt volumes. To change the volume,
follow the steps below.
1.
Press RETEST to enter the Volume option. The current setting is bracketed.
2.
Press either YES or NO to scroll to a new setting, and press RETEST to accept the setting. A
sample sound clip plays. If the volume level is not satisfactory, press RETEST to reenter Volume
and change the setting. There are five volume level settings to select from.
You can scroll to the next System Settings option -or- press MANUAL to return to the Main menu, or
press START + YES to exit Setup Mode.
Spoken Vitals
NOTE: Present in Genesis DM v4.10.14 and higher. Spoken vitals are not available in all languages (see
Appendix E for a list of available languages).
Options:
On, Off
Default:
Off
Access Level:
General
The Spoken Vitals setting controls whether the feature is on or off. Press either YES or NO to change
the setting and press RETEST to enter the setting.
Spoken Vitals will be audible at the end of the collection sequence, and before transmission of vital signs.
You can scroll to next System Settings option -or-press MANUAL to return to the Main menu, or
press START + YES to exit Setup Mode.
P4510EN.07
2/16/12
23
GENESIS™ DM MANUAL
Programming the Genesis DM Monitor
Scale Prompt
Options:
OFF, every time for or any combination of days ( _ _ Mo _ _ We _ _ Fr _ _), or only one
time each day ( _ _ Mo* _ _ We* _ _ Fr* _ _ )
Default:
Every day ( Su Mo Tu We Th Fr Sa)
Access Level:
General
Scale Prompt allows you to turn the audio and text scale prompts off or turn the prompts on for all
or any combination of days. The factory default is on for every day. If the prompt is turned on, the
monitor prompts for (and collects) the patient’s weight at the beginning of each monitoring session.
In the System Settings menu, scroll to Scale Prompt, and press RETEST to enter. The first
selection is bracketed. If the prompt is off, only underscores display.
1.
Press either YES or NO to change the setting, and press RETEST to accept the new setting.
2.
Continue through the seven settings until all are set correctly.
You can scroll to next System Settings option -or- press MANUAL to return to the Main menu, or
press START + YES to exit Setup Mode.
Language
NOTE: Languages include US and British English with female (default) and male (clinician-selectable)
voice choices. Languages other than English are available (see Appendix E). If a different language is
selected, available questions will be in that language. Configure all settings first, then change the language
last of all. The Clinician language may also be set to a different language than the Patient language.
Options:
English
Default:
English
Access Level:
General
The Language menu allows you to change the language of the audio and text prompts for the patient. To
change the language setting, follow the steps below.
1.
In the System Settings menu, scroll to Language, and press RETEST to enter. The current setting is
bracketed.
2.
Press either YES or NO to scroll to the desired language, and press RETEST to enter the setting.
You can scroll to the next System Settings option -or- press MANUAL to return to the Main
menu, or press START + YES to exit Setup Mode.
Units
Options:
English (lbs and °F) or metric (kg and °C)
Default:
Set to common usage for your locale
Access Level:
General
P4510EN.07
2/16/12
24
GENESIS™ DM MANUAL
Programming the Genesis DM Monitor
You can set the Genesis DM to display either English or metric units. Genesis DM monitors should
default to what is customary in your locale. However, if you need to change the units, follow the steps
below.
1.
In the System Settings menu, scroll to Units, and press RETEST.
The current selection is bracketed.
2.
Press either YES or NO to change the setting, and press RETEST to accept the new setting.
You can scroll to the next System Settings option -or- press MANUAL to return to the Main
menu, or press START + YES to exit Setup Mode.
Inflation Settings
Options:
50 – 260 mmHg
Default:
175 mmHg
Access Level:
General
You can use this menu to set the initial NIBP cuff inflation setting on a single-user monitor. (Otherwise,
set the initial NIBP on the patient ID card.)
Genesis DM uses this initial setting only for the first blood pressure collection because the monitor
stores the latest blood pressure (BP) reading in memory to use as a setting for the next blood pressure.
The monitor inflates the cuff to 35 mmHg higher than the last systolic reading.
Once the monitor is powered down the default setting of 175 mmHg is re-established.
In the System Settings menu, scroll to Inflation Settings, and press RETEST to enter.
The current setting is bracketed.
Press either YES or NO to scroll to the desired inflation setting, and press RETEST to accept the
new setting.
You can scroll to the next System Settings option -or- press MANUAL to return to the Main
menu, or press START + YES to exit Setup Mode.
Display Brightness
Options:
Low, Med, High
Default:
High
Access Level:
General
This menu allows you to adjust the brightness level of the VFD display. To adjust the brightness level,
follow the steps below.
P4510EN.07
2/16/12
25
GENESIS™ DM MANUAL
Programming the Genesis DM Monitor
1.
In System Settings, scroll to Display Brightness, and press RETEST to enter.
The current selection is bracketed.
2.
Press either YES or NO to scroll to the desired brightness, and press RETEST to accept the new
setting.
You can scroll to the next System Settings option -or- press MANUAL to return to the Main
menu, or press START + YES to exit Setup Mode.
Questions Menu
The Questions Menu allows you to program the monitor to ask subjective disease-state questions
similar to the questions you might ask during a patient visit. There is a block of fifty-one standard
questions from which you may select up to ten to include in the monitoring session (Classic mode only).
With LifeStream a number of duplicate questions are available that have Disease Specific Symptom
Management (DSSM) educational content attached.
For a list of all of the questions, refer to Appendix A: Subjective Questions or to the Genesis DM
Programming/Troubleshooting Guide.
The menu also offers a block of questions for sleep apnea. It is an on/off option only. If you turn the option
on, the monitor will ask all twelve questions in the block.
Question Education (Educational Content)
NOTE: The Question Education is not available in Classic mode.
The option applies only to Questions 1 – 51.
Options:
Off – no educational content will be displayed
On – educational content will be displayed for select questions
Default:
On
Access Level: General
Education content is an audio and text message that will display after specific question are answered “YES”.
Selecting Off will disable these messages. With Question Education turned Off, questions 1 – 51 will not
display their educational content. When the Question Education feature is On, questions 1 – 51 will display
their educational content. Questions 52 and following will always have educational content displayed.
NOTE: Only one piece of educational content will be played if more than one question is answered
Yes. Under this condition, there is a random nature involved with the educational content played.
Questions/Reminders
Options:
Default:
Access Level:
P4510EN.07
Off or Questions 1 – 51(Classic mode only)
Questions 1 – 61(Genesis DM with Disease Specific Symptom Management [DSSM] activated)
Off
General
2/16/12
26
GENESIS™ DM MANUAL
Programming the Genesis DM Monitor
As you scroll through the block of questions and/or reminders, you see bracketed cue text and the
question ID number. The ID number is located in the upper right-hand corner of the display. Each
question has its own unique ID number that separates it from the other questions in the list.
Each Question has an ID number and an associated question. For example, the prompt cue text (seen by
patient) reads. “Out of Medications?” This prompts’ ID is 07. It is the seventh question of the 51
total questions. The #1 designates the place order in which the question will be presented to the
patient. In this example, Question 7 of 51 is being presented first:
Question #1
07
Out of medications?
For the list of all questions, refer to Appendix A Subjective Questions.
Reminders are chosen the same way that questions are chosen, and the programming steps for
reminders are the same. Reminders are listed after the questions.
Question Time
Options:
OFF, every time for or any combination of days ( _ _ Mo _ _ We _ _ Fr _ _), or only one
time each day ( _ _ Mo* _ _ We* _ _ Fr* _ _ )
Default:
All days
Access Level:
General
Each question is followed by an option that allows you to set a question frequency. The default is daily.
However, you can program any combination of days. You must have at least one day selected or the
monitor will not ask the question.
To program a question and question time
1.
2.
3.
In Questions Menu, scroll to the desired question, and press RETEST. The Off text (or cue text) is
bracketed.
Press either YES or NO to change the setting, and press RETEST to accept the new setting.
Press YES to scroll to the question time, and press RETEST to enter. The first selection is
bracketed.
Press either YES or NO to change the setting, and press RETEST to accept the new setting.
Continue through the seven days until all are set correctly.
Scroll to the next question you want to program and repeat steps 1 – 5.
When finished, press MANUAL to return to the Main menu, or press START + YES to exit Setup
Mode.
Sleep Apnea
Sleep Apnea is one block of questions that are either on or off. The default is Off. To turn the prompt
on, enter a collection frequency.
1.
In the Questions menu, scroll to Sleep Apnea, and press RETEST. The first selection is bracketed. If
the prompt is off, only underscores display.
P4510EN.07
2/16/12
27
GENESIS™ DM MANUAL
Programming the Genesis DM Monitor
2.
Press either YES or NO to change the setting, and press RETEST to accept the new setting.
3.
Continue through the seven days until all are set correctly.
When finished, press MANUAL to return to the Main menu, or press START + YES to exit Setup Mode.
Password
Without entering a password, you will not be able to see or enter password-protected menus or menu
options. The password protected menu options are found in the Service and Transmit menus.
All Genesis DM monitors use the same password. If you do not know the password, contact your
supervisor. Follow the steps below to enter the password.
1.
In Setup Mode, scroll to the Password menu.
2.
Press RETEST to enter the menu. The first underscore is bracketed.
3.
Use either the YES or NO key to scroll to the first number of the password.
4.
Press RETEST to enter the number and move to the next underscore.
Repeat steps 4 and 5 until all four numbers are entered. Once you have pressed RETEST to enter the
fourth number, all menu options become available.
Transmit Menu
In Setup Mode, use YES or NO to scroll to the Transmit Menu, and press RETEST to enter. The
first option, PIN, displays.
PIN Number
Options:
7-digit number
Default:
Set to your organization’s PIN before shipment
Access Level:
Password protected
The PIN number identifies your organization’s mailbox where monitor vitals and configurations are
stored until retrieved by Honeywell HomMed. This PIN is provided by Honeywell HomMed. The
P4510EN.07
2/16/12
28
GENESIS™ DM MANUAL
Programming the Genesis DM Monitor
Genesis DM monitor sends data to this mailbox address and the monitoring application checks this same
address to retrieve the data.
You must make sure that this number is correct. Check the number against the one you received
during your in-service.
1.
In the Transmit Menu, scroll to the PIN Number option. If the number is correct, you can move to
your next task. However, if it is incorrect, press RETEST to enter the option. The first digit is
bracketed.
2.
Use YES or NO to scroll to the correct number, and press RETEST to accept the value and
move to the next digit.
3.
Repeat steps 2 and 3 until you enter or correct the entire PIN number.
You can scroll to the next Transmit Menu option, press MANUAL to return to the Main menu, or
press START + YES to exit Setup Mode.
If you just need to correct one digit, use the MANUAL key instead of the RETEST key. The
MANUAL key will allow entry of the new digit and also back you out of the menu.
Dial Type
Options:
Tone or pulse (rotary)
Default:
Tone
Access Level:
General
If tone dialing is unavailable and there is a rotary phone, change the dial type to pulse.
1.
In the Transmit Menu, scroll to the Dial Type option and press RETEST to enter.
2.
Press either YES or NO to change tone to pulse, and press RETEST.
You can scroll to the next Transmit Menu option, press MANUAL to return to the Main menu, or
press START + YES to exit Setup Mode.
Transmit Mode
Options:
ISP, Local Gateway, Remote Gateway
Default:
ISP
Access Level:
General
The transmit mode determines how the Genesis DM transmits data. When set to ISP, the monitor
sends data over a phone line to your mailbox. An ISP is required to transmit data to the monitoring
application via a phone line or the GPRS modem.
P4510EN.07
2/16/12
29
GENESIS™ DM MANUAL
Programming the Genesis DM Monitor
If the monitor is going to transfer patient data to a local server, set the transmit mode to Local Gateway.
For example, use Local Gateway when you demonstrate the Genesis DM monitor.
Set the transmit mode to Remote Gateway if you want the monitor to transmit data to a remote site
other than an ISP mailbox.
1.
In the Transmit menu, scroll to the Transmit Mode option, and press RETEST. The current setting is
bracketed.
2.
Press either YES or NO to scroll to the correct transmit mode, and press RETEST to accept it.
You can scroll to the next Transmit Menu option, press MANUAL to return to the Main menu, or
press START + YES to exit Setup Mode.
Test Message
Access Level:
General
Test Message allows you to verify transmission by sending an empty test data packet to the monitoring
application. Although you can send a test message at any time to verify transmission, always send a
test transmission after you complete both programming and setup to make sure the monitoring
application can receive data from the monitor.
It is recommended that you always send a test transmission following all new patient setups before
the vitals are sent.
There is nothing to program in Test Message.
Check For Updates
Access Level:
General
Check for Update allows you to communicate to Honeywell HomMed servers and check for new
firmware. Updates are automatically checked every evening. This command is to manually force the
update process.
Phone Prefix
The Phone Prefix feature allows you to add a prefix to the phone number to access an outside line, or
to disable call waiting. Prefixes supported are specific to the patient’s telephone service. For example, to
access an outside line that requires dialing 9 for connectivity, enter “9,” (9 comma) in the Phone Prefix
field.
Options:
Default:
Access Level:
P4510EN.07
x-digit alphanumeric
blank
General
2/16/12
30
GENESIS™ DM MANUAL
Programming the Genesis DM Monitor
Option Menu
The Option Menu allows you to assign communication ports to compatible peripheral medical devices.
You must assign a COM port to each device or the monitor will not prompt for the device and it will
not collect data from that device, even if the device is connected to the COM port.
After assigning a COM port, you can then program how often the Genesis DM collects test results from
the device. The default is daily. However, you can program any combination of days. You must have at
least one day selected or the monitor will not prompt for the device.
Further details are contained in the Peripheral Medical Devices section and in other areas of this manual.
Service Menu
This menu allows Honeywell HomMed technicians to access the Genesis DM monitor to obtain specific
diagnostic data and perform calibration tasks. If you are asked to perform any tasks that require you to
enter this menu, Customer Service will walk you through the steps.
Service Call
Choose this item to initiate a connection between the monitor and the Honeywell HomMed service
department computer.
Service Call Phone # (C)
This item contains the number the monitor uses to connect to the Honeywell HomMed service
department computer.
Calibration Verify (C)
This item is used to perform a calibration test on the NIBP measurement.
!
WARNING
RISK OF INJURY
Verifying calibration while the cuff is attached to a patient could cause bruising or other injury.
DO NOT verify NIBP calibration while the cuff is attached to a patient.
Report Menu (C)
Choose the Report Menu (C) to access the following information:
Serial Number (C)
This item reports the serial number assigned to the monitor.
P4510EN.07
2/16/12
31
GENESIS™ DM MANUAL
Programming the Genesis DM Monitor
Hardware Version (C)
This item reports the hardware version of the monitor.
Software Version (C)
This item reports the software version of the software currently installed on the
monitor.
Oximeter Version (C)
Reports the hardware version of the oximeter board installed in the monitor.
NIBP Version (C)
Reports the NIBP version of the monitor.
BP Safety Version (C)
Reports the BP Safety version of the monitor.
BP Hardware Version (C)
Reports the BP Hardware version of the monitor.
NIBP Cycles (C)
Reports the number of NIBP cycles since the last calibration was performed.
Last Calibration (C)
Reports the date of the last calibration of the monitor.
Exit Report Menu
Exit the Service menu and return to the previous menu.
Load Factory Default (C)
Returns the monitor to the original factory settings at the date of delivery.
Classic Mode (C)
Activates Classic Mode. This turns off any additional features specific to the Genesis DM monitor. Also
complies with Central Station 3.5.
P4510EN.07
2/16/12
32
GENESIS™ DM MANUAL
Installing the Genesis DM Monitor
Installing the Genesis DM Monitor
It is important to set up the Genesis DM monitor in an area where it can successfully measure vital signs
and transmit results. If a patient uses peripheral medical devices, ensure there will be enough room for
the patient to attach them without any difficulty.
To determine a suitable location, consider the following:
DO:

Choose an accessible place to encourage daily use.

Find a flat, stable location for the scale where it is moved minimally and where the patient can
easily stand on it.

Use a flat, stable surface for the monitor, located where the patient can sit comfortably at rest
for the readings.

Select an area that allows the free flow of air around the monitor.

Ensure that a bed, chair, or other seating is readily available.

Be certain that the patient can easily attach any peripheral devices.
DO NOT:

Place any equipment near sources of water such as sinks or bathtubs.

Place on or near a heat source (stove, fireplace, heaters, heat vents, etc.) or in direct sunlight.

Use an electrical wall outlet controlled by a light switch.

Power up until the monitor is setup and all of the equipment is properly attached.

Place the Genesis DM monitor on top of or next to electrical equipment (for example, a
television or microwave). If absolutely necessary to place next to an electrical device, verify that
the monitor works properly.
P4510EN.07
2/16/12
33
GENESIS™ DM MANUAL
Installing the Genesis DM Monitor
Attaching the scale
Make sure that you use the appropriate equipment for each monitor. Pay special attention to the scale
cords, keeping in mind that there are different cords depending on your configuration.
Before leaving for a patient site, read the labels on the scale cord(s) to verify that you have the correct
cords for the monitors and scales that you intend to install.
Feature
Fairbanks Scales
(used with product bearing CE Marking)
HomMed Scales and HomMed Bluetooth
Scales
Size
Approximately 13.5" by 13.5"
Approximately 15.75" by 14"
Shape
Square
Curved
Surface texture
Linear textured pattern
Round textured pattern
Feet (excluding Bluetooth)
Screw in/out; removable (excluding Bluetooth)
Some HomMed Scales: Screw in/out: not
removable (excluding Bluetooth)
Scale cord
4-pin MiniDIN connector to monitor and tabbed
jack to scale
4-pin MiniDIN to monitor and DB9 serial
connector to scale
The ideal location for the scale is on an uncarpeted, hard surface. If the floor is carpeted, make sure the
bottom of the scale clears the carpet so that when the patient steps on the scale, pressure is placed on
the four feet that contain the load cells for scales equipped with feet (the Bluetooth scale is not
equipped with adjustable feet). As required, adjust scale height by screwing the feet in or out.
All scales can handle from 0 to 500 pounds (0 to 227 kg). Assure that the person being weighed on the
scale utilizes proper positioning. Instruct the patient to be centered on the scale and to remain
motionless during scale use. Patients who are unable to stand motionless without assistance (a wall,
chair, helper, etc.) throughout the time it takes for the weight to be measured should not use the scale.
To attach the Honeywell HomMed scale
1.
Attach the scale cable to the scale first. Locate the scale’s DB-9 port at the top of the scale and
insert the DB-9 connector on the scale cable into the port. Tighten the screws until secure.
2.
Plug the round MiniDIN connector into the SCALE communication port on the monitor.
3.
When the scale is in position for the patient and level, connect it to the monitor.
To attach the Fairbanks scale
1.
Attach the scale cable to the scale first. Insert the cable’s tabbed plug into the scale’s output port
located on the back of the scale. You should hear it click into place.
2.
Plug the round MiniDIN connector into the SCALE communication port on the monitor.
3.
When the scale is in position for the patient and level, connect it to the monitor.
Honeywell HomMed and Fairbanks scales are interchangeable provided you have the appropriate
scale connector cable. If you exchange scales:
• Read the label on the scale connector cord to verify that you have the correct scale cable
for the scale you intend to attach to the monitor.
• Power-down the monitor before you exchange scales. Attach and position the new scale
before you turn the monitor on.
P4510EN.07
2/16/12
34
GENESIS™ DM MANUAL
Installing the Genesis DM Monitor
Connecting a HomMed Bluetooth (wireless) Scale
These instructions cover preparing the DM Monitor to use a Bluetooth adapter and wireless Honeywell
HomMed Bluetooth Scale. Three basic actions are involved:
1. Installing the Honeywell HomMed Bluetooth adapter.
2. Connecting one wireless scale.
3. Connecting when multiple wireless scales are in the area.
NOTE: Bluetooth is a proprietary, open wireless method for exchanging data within relatively short
distances using radio signals. For more information, see http://www.bluetooth.com.
Installing the Honeywell HomMed Bluetooth adapter
The wireless scale kit includes a Bluetooth Adapter and an Adapter Cable. As needed, securely connect
the Adapter Cable to the Bluetooth Adapter. For the Bluetooth Adapter, assure that the slide switch on
the top of the Adapter is positioned toward the Connector end of the adapter to enable proper
transmission and data flow between the wireless scale and the DM Monitor.
What is the slide switch?
The slide switch is part of the adapter. The slide switch allows different positons to be selected for
different adapter functionality. The slide switch is not an ON / OFF switch; therefore, the slide switch is
not a “power” switch per se, but must be in the proper position to enable data transmission.
Why does the slide swith need to be in the proper position?
The slide switch needs to be in the proper position in order to enable proper transmission and data
flow between the wireless scale and the DM Monitor.
Slide
Switch
toward
connector
end
Slide
Switch
Bluetooth Adapter
P4510EN.07
2/16/12
Connector
End
35
GENESIS™ DM MANUAL
Installing the Genesis DM Monitor
Adapter Cable
Before proceeding, check to make sure that the DM Monitor is powered OFF.
With the DM Monitor power OFF, plug the Adapter Cable into the Scale port at the back of the
monitor, or the wireless scale will not be sensed properly.
Scale
Port
Adapter Cable
Bluetooth
Adapter
The adapter is equipped with light emitting diodes (LEDs) to indicate various conditions:
• Blue Blinking = scanning for / communicating with BT devices
• Solid Red = adapter powered during data transfer
• Green = indicates data transfer
P4510EN.07
2/16/12
36
GENESIS™ DM MANUAL
Installing the Genesis DM Monitor
After connecting the Bluetooth Adapter and Adapter Cable to the DM Monitor and making sure that
the slide switch is toward the connector end, power ON the DM Monitor. When prompted, enter the
Setup Wizard by pressing YES.
Enter Setup Wizard?
(Yes/No)
Connecting one wireless scale
The monitor will prompt the user to place the scale in its final position.
Place Scale in
Final Position
Assure that there is power to the scale, supplied either by the scale batteries or the scale AC power
adapter. Place the scale in a position convenient for the patient.
NOTE: The scale should be less than 30 feet (10 meters) from the monitor for signal acquisition. The
reason that the scale should be less than 30 feet (10 meters) from the monitor is that the signal
between the two devices only works within such relatively short distances. If the scale and monitor are
too far apart, or if items or personnel are blocking the signal, then the scale and monitor will not be able
to connect and communicate with one another as needed for proper functionality.
P4510EN.07
2/16/12
37
GENESIS™ DM MANUAL
Installing the Genesis DM Monitor
Press YES to allow the monitor to detect the scale. The monitor will display the “Please Wait Scanning
Devices” message as it scans for suitable Bluetooth devices in the nearby vicinity. The DM Monitor
locates, senses, and pairs with a single scale automatically.
Please Wait
Scanning Devices . . .
NOTE: Establishing communications may take several minutes.
During the Paring process the Bluetooth adapter blue LED (light emitting diode) will flash on and off,
indicating activity. The red LED should be ON, indicating that the adapter is powered by the monitor.
Check the DM Monitor-to-Bluetooth adapter connection if the red LED is not ON.
P4510EN.07
2/16/12
38
GENESIS™ DM MANUAL
Installing the Genesis DM Monitor
The blue LED is labeled with the number, 1, in the line art drawing below, and is the LED furthest from
the cable end of the adapter.
When the blue diode is flashing, the Bluetooth adaptor is searching for devices in the vicinity.
When a single scale is detected and paired with the monitor, the monitor will display the Pairing
Successful message. If there is only one scale in the vicinity, that scale will be the one that the monitor
will pair with.
Pairing Successful
If the monitor is unable to detect the scale, it will display the “Pairing Failed” message.
Pairing Failed
P4510EN.07
2/16/12
39
GENESIS™ DM MANUAL
Installing the Genesis DM Monitor
If the pairing failed, check to make sure that the scale has power and that the adapter is properly
connected to the monitor and that the Bluetooth adapter red LED is ON. Also, verify that the scale is
located close enough to the monitor, and less than 10 meters or about 30 feet away.
After a failed attempt to detect a wireless scale, the DM Monitor screen will ask if you desire a scan for
available Bluetooth devices. After verifying that the scale has power and is the appropriate distance from
the monitor, press YES to retry.
Scan for Available
BT Devices?
[YES] or [No]
When the scale is detected, the monitor will display the Pairing Successful message.
Pairing Successful
After the Pairing Successful message appears, complete the Setup Wizard in the normal manner. The
monitor will then be ready for use, programmed with the wireless scale.
P4510EN.07
2/16/12
40
GENESIS™ DM MANUAL
Installing the Genesis DM Monitor
Connecting with multiple wireless scales in the area
To configure the DM Monitor for use when there are multiple wireless scales in the area, before beginning
the pairing process, verify the serial number of the scale that you wish to pair with the monitor. The serial
number is located on the bottom of the scale.
You may write the Serial Number here for reference purposes: ___________________________
After verifying the serial number, power ON the DM Monitor.
STARTING
GENESIS DM
When prompted, enter the Setup Wizard by pressing YES.
Enter Setup Wizard?
[YES] or [NO]
The DM Monitor and Bluetooth Adapter will sense the presence of multiple devices.
Press YES to continue.
Please Wait
Scanning Devices . . .
NOTE: Establishing communications may take several minutes.
P4510EN.07
2/16/12
41
GENESIS™ DM MANUAL
Installing the Genesis DM Monitor
The monitor will prompt the user to place the Scale in its final position.
Place Scale in
Final Position
Assure that there is power to the scale, supplied either by the scale batteries or the scale AC power
adapter. Place the scale in a position convenient for the patient.
NOTE: The scale should be less than 30 feet (10 meters) from the monitor for signal acquisition.
Press YES to allow the monitor to detect the scale.
The monitor will display the “Please Wait Scanning Devices” message as it scans for suitable Bluetooth
devices in the nearby vicinity.
Please Wait
Scanning Devices . . .
NOTE: Establishing communications may take several minutes.
During the Paring process the Bluetooth adapter blue LED (light emitting diode) will flash on and off,
indicating activity. The red LED should be ON, indicating that the adapter is powered by the monitor. Check
the DM Monitor-to-Bluetooth adapter connection if the red LED is not ON.
P4510EN.07
2/16/12
42
GENESIS™ DM MANUAL
Installing the Genesis DM Monitor
The blue LED is labeled with the number, 1, in the line art drawing below, and is the LED furthest from the
cable end of the adapter.
When the blue diode is flashing, the Bluetooth adaptor is searching for devices in the vicinity.
When the DM Monitor and Bluetooth Adapter have located the suitable devices in the area, a list will
appear and prompt the user to select a scale to pair to the monitor.
Only one device or serial number is displayed on the monitor at a time. Pressing the NO button, for
example, goes down the list, from 1 to 2 to 3 to 4 . . . back to 1. Pressing the YES button goes up the list,
from 1 to the end of the list . . . to 4 to 3 to 2 to 1.
The RETEST button selects the currently displayed serial number.
• IMPORTANT •
If a scale serial number appears on the monitor display screen for more than 30 seconds,
the monitor will automatically select and pair with that scale.
NOTE: If the monitor pairs to a different scale than the one desired, see the Bluetooth Wireless Scale
Troubleshooting section that appears after this section.
Select Scale To Pair. . .
1. 5094950014
P4510EN.07
2/16/12
43
GENESIS™ DM MANUAL
Installing the Genesis DM Monitor
Press the NO key to continue to the next number.
Select Scale To Pair. . .
2. 5094950020
Again, press the NO key to continue to the next number.
Select Scale To Pair. . .
3. 5094950021
When the selected scale is detected and paired with the monitor, the monitor will display the Pairing
Successful message.
Pairing Successful
After the Pairing Successful message appears, complete the Setup Wizard in the normal manner.
The monitor will then be ready for use, programmed with the wireless scale.
Using the Wireless Scale
During a scheduled reading, the monitor will display the following message when preparing the scale to take
a measurement:
P4510EN.07
2/16/12
44
GENESIS™ DM MANUAL
Installing the Genesis DM Monitor
Preparing Scale
Please Wait
When communication is established, the user will be prompted to step on the scale as in normal (wired)
operation. The audible message will be, “Please step on the scale.”
STEP
ON
SCALE
ooooo
ooooo
When the scale senses, through a series of signal transactions, that the person being weighed has stabilized
on the scale, then the weight will display. The display of the weight may take a moment.
220
LBS
When the weight sensing is complete, the audible message will be, “Please, step off the scale.”
P4510EN.07
2/16/12
45
GENESIS™ DM MANUAL
Installing the Genesis DM Monitor
STEP OFF
SCALE
ooooo
ooooo
This completes the installation and use of the Wireless Scale.
Bluetooth Wireless Scale Troubleshooting
If communication cannot be established
If communication cannot be established, the DM Monitor will display one of the following error messages:
Error Code E203, E204, or E205; Scale Disconnected.
E203 error code: The Bluetooth Adapter is not properly plugged in to the monitor.
ERROR CODE: E203
Scale Disconnected
Check the connection on the rear panel and retry.
E204 error code: Bluetooth Adapter is not paired to the scale.
ERROR CODE: E204
Scale Disconnected
P4510EN.07
2/16/12
46
GENESIS™ DM MANUAL
Installing the Genesis DM Monitor
Check the Bluetooth Adapter. Make sure that the Bluetooth Adapter is powered ON and that it is properly
connected to the Adapter Cable, and that the Adapter Cable is properly connected to the DM Monitor at
the Scale port on the rear panel of the DM Monitor. Retry.
E205 error code: No Bluetooth connection because the scale is not powered, the scale batteries are
absent, improperly installed, or need replacement, or the scale is too far away from the monitor.
ERROR CODE: E205
Scale Disconnected
To correct the error, check that there is power to the scale, or that it is located within range of the
monitor (approximately 10 meters, or 30 feet). Reschedule another vitals collection to retry.
Pairing to a new scale
If a different scale than the one desired is paired to, follow the procedure below.
1. Plug in the monitor to scale cable, run setup wizard and configure the monitor for wired mode, then
remove the scale cable and run setup wizard again and pair to new scale.
2. Run setup wizard, monitor will first attempt to pair to previous scale. After that fails, say YES to the Scan
for Additional Bluetooth Devices question. Monitor will now pair with the new scale.
If the monitor fails to pair and displays the Pairing Failed message:
The monitor may not move to the “Scan for additional devices” prompt for various reasons. If that should
happen, simply press YES when the Pairing Failed prompt is displayed.
Changing batteries
To change the batteries, turn the scale over. Use a Philips screwdriver to remove the screw retaining the
battery access panel to the scale. Observe the orientation of the four AA batteries inside the scale. Remove
the batteries and replace with four new AA batteries for best results.
NOTE: The DM Monitor MUST BE RESTARTED after changing the batteries. This is because the Bluetooth
wireless scale is not configured to give a zero weight reading (“zeroed”) on power up. The scale is only
zeroed during monitor startup when the monitor does the scale self test. This is unique to the Bluetooth
wireless scale – including when the AC adapter is being used. For other scales, the DM Monitor will
automatically zero the scale on scale power up.
Zeroing the scale
After changing batteries on the Bluetooth wireless scale, return the scale to its normal position and assure
that there are no items on the scale. Re-start the DM Monitor. The monitor will conduct a scale self-test
P4510EN.07
2/16/12
47
GENESIS™ DM MANUAL
Installing the Genesis DM Monitor
and zero the scale for optimal accuracy of weight readings. Check the zeroing to assure that zero ( 0 )
appears on the monitor display.
Attaching the blood pressure cuff
Push the metal connection on the blood pressure tubing firmly into the NIBP port. It should click when
it locks into place.
Attaching the modem
CAUTION
REQUIRES ANALOG PHONE LINE
The modem requires an analog phone line. A digital phone line will destroy the modem.
DO NOT use a digital phone line. Verify presence of analog phone line prior to
installation.
The modem connection is on the back of the Genesis DM monitor.
To connect the modem to the phone outlet, follow the steps below.
1.
Unplug the patient’s phone line from the wall telephone jack.
2.
Plug the provided duplex jack into the wall jack.
3.
Plug the patient’s telephone into one side of the duplex jack.
4.
Plug the Genesis DM phone cord into the other side of the duplex jack.
5.
Plug the remaining end of the Genesis DM phone cord into the MODEM plug located on the back
of the monitor.
Alternately, you can connect the GPRS modem to connect wirelessly.
If DSL is installed as the broadband connection, be certain that a DSL filter is installed between the
monitor and wall jack.
NOTE: If a patient wants to use GPRS rather than a plain old telephone service (POTS) connection, verify
that there is sufficient GPRS wireless network coverage available at the patient’s location prior to attempting
the installation. To check coverage, contact the wireless service provider(s) in the patient’s geographical
area.
P4510EN.07
2/16/12
48
GENESIS™ DM MANUAL
Installing the Genesis DM Monitor
Powering up
Before you power up the Genesis DM monitor:

Check to see that you have attached all of the equipment.

Make sure that the scale is in the proper location. Moving the scale is not recommended after it
is connected and power is applied.

Look to see that there is nothing on the scale. The scale will not zero correctly if there is any
weight on it when the monitor is turned on and activates the scale.

Position the monitor so that the cords are out of the way and do not create a tripping hazard.
Also make sure that the rear of the monitor has at least a two-inch (5cm) clearance from
surrounding objects or walls.
CAUTION
INCORRECT CONNECTION SEQUENCE CAN DAMAGE THE MONITOR
Plug the power supply into the Genesis DM Monitor first and into the power source last.
Failure to do so could cause damage to the monitor and/or power supply.
Verify that the power supply is NOT plugged into an AC source before
connecting the monitor. Contact Honeywell HomMed Customer Service if you
are unsure or have any questions about the correct power-up method.

Plug the AC adapter into the power input on the rear of the monitor.

Plug the AC adapter into a suitable power source, such as a wall outlet or power strip.

IF USING A POWER STRIP: Make sure that the AC adapter is flush with the strip. Plug the
power strip into a wall outlet. As needed, turn on the switch on the power strip.
As soon as the monitor powers up, it displays the current software version. If the monitor detects any
problems, it will display an error code. For a list of error codes, refer to the Genesis DM Monitor
Programming/Troubleshooting Guide. A serious error will halt the system. If a serious error occurs, contact
Customer Service. All error codes should be investigated as they may prevent proper operation.
After successful power up, the monitor enters the idle mode and the panel displays the current date,
day of week, and time randomly across the screen. If the Genesis DM monitor is for multi-patient use
and does not have any programmed report times, the idle mode displays a prompt that asks the user to
swipe the ID card.
P4510EN.07
2/16/12
49
GENESIS™ DM MANUAL
Installing the Genesis DM Monitor
Monitor Demonstration Mode
The Genesis DM Monitor includes a demonstration utility that will display a “simulated” vitals acquisition
sequence.
To access the demonstration mode:
1) Power up the monitor
2) Wait for the monitor to enter the “Idle Mode”
3) Press the “Stop” and “No” keys at the same time
The monitor will display this question:
Run the Demo
[yes] [no]
To run the demonstration:
1) Press the “YES” key.
2) The Demonstration Mode will display and run.
3) Upon completion, the monitor will return to the “Idle Mode”.
4) To repeat the demonstration, start again by pressing the “Stop” and “No” keys at the same time.
NOTE: The voiced segment of the monitor demonstration will be done in the currently selected
patient language.
Re-starting the Monitor
The Genesis DM Monitor can be started again.
To restart the monitor select “No” at the display:
Run the Demo
[yes] [no]
The monitor will display the question:
Restart the Monitor
[yes] [no]
To restart the monitor, select and press the “yes” key.
P4510EN.07
2/16/12
50
GENESIS™ DM MANUAL
Patient Training
Patient Training
It is important to review the benefits of using the Genesis DM with patients and caregivers. Explain how
the system empowers patients to take an active part in managing their health care. They can log and
track their vitals and have the security of knowing that their Genesis DM data is being viewed daily by
their health care providers.
!
CAUTION
NOT AN EMERGENCY RESPONSE DEVICE
The Genesis DM Monitor is not intended to be an emergency response device and does not
provide real time critical care data. Clinicians might only review data once a day.
In case of a medical emergency, patient should call local Emergency Medical
Service and/or health care provider.
It is vital that patients realize that the monitor is not a medical emergency response device. Instruct the
patient on what to do in a medical emergency.
Reinforce that the patient is the only one who should use the monitor. Explain that the monitoring
software assumes that all information received from the monitor comes from the patient and posts the
data in the patient’s records.
Genesis DM Monitor Voice and Text Prompts
Voice Prompts
The Genesis DM voice prompts feature female and male voices. Voice gender preference is configured
by the user at the monitor and this option is also available via the setup wizard.
The monitor uses friendly voice prompts and easy-to-read text prompts to guide the patient through
each step of the monitoring session. Text and graphical cue prompts display in the monitor’s display
panel at the same time that the audio prompt plays.
Prompt timing
A monitoring session generally includes collecting weight, blood pressure, heart rate, and answers to
customized patient questions. Patients may also use the optional temporal temperature, oximeter or
peripheral medical devices.
P4510EN.07
2/16/12
51
GENESIS™ DM MANUAL
Patient Training
Audio and text prompts are separated to allow the patient ample time to perform requested tasks. If
the patient does not complete the request within the time frame allowed after the first prompt, the
monitor repeats the prompt up to three times.
At the beginning of a session, the patient has thirty minutes to comply with the initial prompt to step on
the scale. During the thirty minutes, the scale prompt is repeated at the rate of once every five minutes
until the patient steps on the scale or the thirty-minute time frame ends. If the monitor times out, it
sends an empty packet of data to the monitoring application.
Once the session is started, the monitor allows time for each task as follows:

Thirty seconds to answer each question

Three minutes to attach a peripheral device
The Genesis DM monitor repeats prompts for each step until the patient complies with the instruction
and the unit gets a valid reading, or until time allowed for the task runs out. If the monitor does not get
a value for an enabled vital or peripheral device, it inserts a null value in the data packet for that vital or
device.
Monitor Prompt
Timing
Value Received
Value NOT Received
Question
30 seconds
Data received -advances to next
question
Null value inserted into data
packet
Attach a peripheral
device
3 minutes
Data received
Null value inserted into data
packet
General prompts and text displays
The monitor goes through the following procedure: says a greeting, prompts for weight and then blood
pressure, asks programmed questions, prompts for programmed peripheral medical devices, reminds
patient to take medications, and thanks patient for completing their vital signs. If the monitor is
programmed to collect temperature and/or oximetry and glucose, the prompts are incorporated into
the session.
NOTE: See Question Education on page 26.
The following table describes the basic prompts and text displays. For a list of all possible prompts,
refer to Appendix C: Prompt Text.
Basic Monitor
Actions
Text Display
Sequence
Audio Report Sequence
Greeting
Good morning
(Afternoon, Evening)
Good morning
(Afternoon, Evening)
It is now time to record your vital signs
Weight
Step on the scale
Please step on the scale
After weight is taken
Step off the scale
Please step off the scale and sit down on a chair
in front of the monitor.
P4510EN.07
2/16/12
52
GENESIS™ DM MANUAL
Patient Training
Basic Monitor
Actions
Blood Pressure
Text Display
Sequence
Put cuff on arm
Rest arm as
instructed
Press Start
Questions
Remove cuff
[cue text varies for
each question]
Audio Report Sequence
Place the blood pressure cuff on your right arm
above the elbow. Tighten it securely.
Rest your arm as instructed by your clinician.
Press the green start key.
Please remove the blood pressure cuff.
[varies for each question]
Receiving data, please wait.
Data upload is finished.
Thank you
Data Transmission
and Thank you
Transmitting
Transmission success
Please remember to take your medications as
prescribed by your physician and maintain the
proper diet.
Thank you for completing your vital signs
measurement.
No vital signs, your clinic will be notified.
The following list describes the basic prompts and text displays. For a list of all possible prompts, refer
to Appendix C: Prompt Text.
The monitor says a greeting and announces that it is time to take vital signs.
Cue text: Good Morning (Afternoon, Evening).
The monitor instructs the patient to stand on the scale.
The monitor skips this if the scale is not attached or if a collection frequency is not set.
Cue text: Step on the scale
After the patient’s weight is recorded, the monitor prompts the patient to step off the scale and sit
down on a chair in front of the monitor. Instruct the patient to wait for the prompt to ensure
that weight is collected.
Cue text: Step off the scale
Weight results display by the scale symbol. (Results exceeding the scale’s capacity, will be displayed as
dashes - -)
A voice prompt instructs the patient to place the blood pressure cuff on the right arm above the elbow
and tighten it securely.
P4510EN.07
2/16/12
53
GENESIS™ DM MANUAL
Patient Training
Cue text: Put cuff on arm
An additional blood pressure prompt instructs the patient to rest their arm as instructed. It is important
that you show them where to place their arm to obtain the most accurate blood pressure reading.
Cue text: Rest arm as instructed
The patient is prompted to press the START key.
Cue text: Press Start
Real-time inflation and deflation pressures display while the monitor measures blood pressure. When
done, the monitor displays the blood pressure readings.
The monitor instructs the patient to remove the blood pressure cuff.
Cue text: Remove cuff
The monitor then asks the patient’s selected questions requesting that the patient press either the YES
or NO key in response.
Cue text: <cue text for each question>
The monitor reminds the patient to continue with prescribed medications and diet and thanks them for
completing their vital signs.
Cue text: Data Transfer <and then> Transmitting
Patient data is transmitted automatically. If the transmission is successful, the display reads:
“Transmission success”
The Genesis DM monitor continues to display patient weight, blood pressure, and if collected, SpO2 for
five minutes to allow the patient to record vitals. To clear the display and return to idle mode right
away, press the green Start key on the left. Remember, pressing the Manual key twice brings you out of
the Menu, and pressing any key once simply clears the status.
Prompts for peripheral devices
Prompts for peripheral devices vary depending on the device. The monitor uploads data from each
programmed device and displays the following messages:

If this is the first time this monitor has collected readings from this device,
“new <device name> detected”.

If there is new data: “Receiving Data” followed by “Data upload finished”.
The monitor does not display test results collected from peripheral devices, only the text prompts
related to the device. Once it has collected data, it repeats the process with the next scheduled device.
For more information on peripheral medical devices, refer to Appendix E Medical Device Specifications and
Peripheral Medical Devices.
P4510EN.07
2/16/12
54
GENESIS™ DM MANUAL
Patient Training
If you exchange a patient’s monitor, the new monitor will consider the patient’s peripheral devices as
“new.” It is recommended that you send a test packet following a change of monitors and for every
patient change.
Patient Reminders
The Genesis DM monitor will play audio patient reminders that are presented after the questions are
asked and that can be scheduled with the monitor. These reminders will present audio and text prompts
at the conclusion of vitals acquisition. An example of a patient reminder would be:
“Please remember your physician appointment this week”
Reminders will be presented prior to educational modules where applicable. Reminders cannot be
scheduled with the monitor in Classic mode.
For the list of all reminders, refer to Appendix A Subjective Questions.
Questions and Reminders (Patient Messaging)
A clinician may periodically send a patient important healthcare questions and/or reminders during the
day to gather more information on the health status of the patient. The monitor will prompt the user to
respond if there is a pending message by displaying the following message:
Press [Yes] To View New Message
Simply follow the onscreen prompts and answer any message(s) by either pressing the YES key or the
NO key on the monitor. This information will be transmitted to the clinician, and the patient will see the
following Data Transfer message on the monitor:
P4510EN.07
2/16/12
55
GENESIS™ DM MANUAL
Patient Training
Disease Specific Symptom Management
The Genesis DM monitor currently stores 51 patient questions which prompt the user for a Yes or No
answer.
The Genesis DM monitor also features Disease Specific Symptom Management (DSSM) educational
components. These questions are duplicates of previously presented questions that have an educational
(DSSM) component attached. DSSM components are presented as audio and textual information
segments during data transmission. If the answer to any of these DSSM questions is “Yes” then a
SINGLE educational component will be presented.
These education modules will not work with a monitor in Classic mode.
For the list of all educational components, refer to Appendix G Education Components.
Taking patient vital signs
Before you take the first set of vital signs for a patient, make sure that the patient knows how to use
each piece of equipment. Demonstrate how to connect peripheral medical devices to the Genesis DM
monitor and answer any questions that the patient may have.
When the patient is comfortable with the equipment, go through the complete process with the patient
and actually transmit the first set of vital signs.
Teach the patient that the correct way to press keys is to hold the top of the monitor with the fingers
and to use the thumb to press the keys.
P4510EN.07
2/16/12
56
GENESIS™ DM MANUAL
Patient Training
Collecting weight
!
CAUTION
PERSONAL INJURY HAZARD
Weight placed on the edge of the scale may cause the scale to tip and can
cause a patient to fall.
DO NOT step onto or attempt to stand on the edge of the scale.
A patient must be able to stand on a scale without support for the time that it takes the monitor to
collect a weight, approximately three seconds without movement. If the patient is not able to do so,
disconnect the scale and turn the scale prompt off.
For stability and measurement accuracy, instruct the patient to stand in the middle of the scale and
not to move while the scale collects the patient’s weight. Stepping or standing on the edge may cause
the scale to tip and possibly result in personal injury.
Collecting blood pressure
Proper cuff size and placement are essential to the accuracy of the blood pressure reading. Every cuff
has an arm size range (in centimeters) printed on the outside label. The smallest and largest points in the
range are determined as follows:

The smallest size in the range is the point where the tab can no longer be pulled through the
ring. If the cuff is still loose, use the next smaller size cuff.

For the largest size in the range, the ring should fall in the natural break of the cuff. The natural
break is the seam where the two Velcro patches meet. If the arm is larger than this, use the
next larger cuff size.
The blood pressure cuff may be used on either the right or left arm, but the patient should use the same
arm consistently. Show the patient where to rest their arm so that it is at or below heart level in order
to obtain the most accurate reading. The cuff should be placed on the opposite arm from the oximeter,
if used.
The position of the patient’s arm can significantly affect the readings. In general, the arm should rest
supported at or below heart level. A consistent position should be used for each reading to increase the
accuracy of the data.
While taking blood pressure, the monitor displays the real-time cuff inflation pressure in the Systolic
display. When the reading is complete, the monitor displays systolic and diastolic results.
Safety notes
Any blood pressure measurement can be affected by the position of the patient, movement, talking,
physiological condition, and other factors.
Make sure that the BP hoses are not kinked, compressed, or restricted.
Check to ensure that the operation of the equipment does not result in the prolonged impairment of
the circulation of the patient. Point out that the patient can stop the inflation at any time by pressing the
red STOP key. To restart inflation, press the green START key.
P4510EN.07
2/16/12
57
GENESIS™ DM MANUAL
Patient Training
Inflation pressure
The Genesis DM monitor has a preset default inflation pressure of 175 mmHg. However, it only uses
this setting once because the Genesis DM monitor saves the last blood pressure reading in memory.
The next inflation pressure is 35 mmHg above the last saved systolic until the monitor’s power is cycled,
at which point the 175 mmHg preset is restored.
If the patient’s blood pressure is higher than the inflation pressure, the Genesis DM monitor increases
the pressure by 35 mmHg, re-inflates, and takes a new blood pressure reading. If the monitor cannot
obtain a blood pressure after a total of three attempts, the monitor displays an error.
If the patient has an ID card, the monitor uses the inflation setting on the card each time it takes a
blood pressure reading.
Cuff placement instructions
Teach the patient how to properly position and tighten the cuff:
1.
Teach the patient to hold the cuff so that the artery marker (white arrow) points toward the hand.
2.
Instruct the patient how to face the palm up and how to slip the hand through the cuff.
P4510EN.07
2/16/12
58
GENESIS™ DM MANUAL
Patient Training
3.
Teach the patient to position the cuff so that the artery marker arrows point to the ring finger and
slide the cuff up the arm, over the elbow. There should be about a two-finger space between the
bend in the elbow and the sphygmomanometer cuff.
4.
Encourage the patient to grasp the tab and pull down to tighten the cuff. Remind the patient to not
make the cuff too tight; the patient should be able to slide a finger between the cuff and the arm.
P4510EN.07
2/16/12
59
GENESIS™ DM MANUAL
Patient Training
5.
Now instruct the patient to wrap the cuff end under and around the arm and fasten it to the Velcro.
6.
Tell the patient to keep still and to rest the arm as the blood pressure reading is taken. Let the
patient know that moving or talking is to be avoided until the monitor completes the reading.
Using peripheral devices
After the base vitals and answers to subjective questions are collected, the Genesis DM monitor uploads
test results from the patient’s peripheral medical device(s). The only exceptions are oximetry and
temperature. Genesis DM collects oximetry with the blood pressure and temperature after blood
pressure.
The monitor prompts for each programmed device, allowing enough time for the patient to connect the
device if not already attached to the monitor. For more information about a specific peripheral device,
including temporal temperature, refer to the section Peripheral Medical Devices in this manual.
Concluding the monitoring session
After the monitor collects all of the patient’s vitals and any programmed peripheral medical device test
results, it transmits the data. The message, “Transmit Success” displays on the screen for 30 seconds.
If the monitor is programmed with settings for a single user, the vitals continue to display on the screen
for five minutes after transmission. If the monitor is for multi-users, the vitals only display for thirty
seconds. You can remove the information from the screen by pressing the MANUAL key twice. When
you train multi-users, instruct patients to remove their information from the screen before they leave
the monitor.
Patient-initiated monitoring sessions
Patients can send vitals to their health care provider at any time; they do not have to wait for a
scheduled report time. Patients can initiate unscheduled monitoring sessions by pressing the blue
RETEST key. If Yes is chosen for a retest of the full set of vital statistics, the Genesis DM monitor
prompts patients through the monitoring session in the same manner as a scheduled report time.
If the patient is collecting vitals during an unscheduled session that overlaps a scheduled report time, the
monitor will complete the unscheduled session and then immediately begin the scheduled one.
Point out to the patient that they normally use this feature if the monitoring application clinician
requests that they repeat a test. Inform the patient that vitals may only be reviewed during the day, so if
they send a reading in the middle of the night, their health care provider may not look at the reading
until much later. Stress again that this is not an emergency response device.
Partial Retest
The Genesis DM monitor allows users to select a portion of vital signs acquisition via the monitor’s
RETEST key. When retesting patients will be prompted with two options available:

For a complete retest of vital signs, press Yes

For a partial retest of vital signs, press No
When choosing a subset, missing vitals will display as no values in the LifeStream application.
P4510EN.07
2/16/12
60
GENESIS™ DM MANUAL
Patient Training
NOTE: Partial retests will not prompt for peripheral devices or play patient education. Full retests will
prompt for peripheral devices and play patient education.
Manual mode
Clinicians can use Manual mode to take vitals without transmitting the data to your monitoring
application. Technicians also use Manual mode when they calibrate the monitor.
When in Manual mode, the monitor only collects vitals. It does not prompt, display text, or collect test
results from peripheral devices. The results are not transmitted.
To enter or exit Manual mode, press the MANUAL button.
P4510EN.07
2/16/12
61
GENESIS™ DM MANUAL
Routine Maintenance
Routine Maintenance
This section provides instructions for routine maintenance. Follow the instructions and safety
information carefully. Failure to do so may void your warranty.
Honeywell HomMed offers preventative maintenance services at the fees and terms stipulated in your
contract. If you wish Honeywell HomMed to perform any services, contact Customer Service.
!
WARNING
RISK OF ELECTRIC SHOCK
Electrical components contain sufficient voltage to cause death or serious injury.
DO NOT disassemble.
DO NOT autoclave.
DO NOT sterilize with ethylene oxide.
DO NOT immerse in water or any other liquid.
DO NOT spray any cleaners on or into the unit.
UNPLUG before cleaning, disinfecting, or performing any maintenance.
For service, call Honeywell HomMed Customer Service.
P4510EN.07
2/16/12
62
GENESIS™ DM MANUAL
Peripheral Medical Devices
Cleaners for the Genesis DM Monitor
The outer case of HomMed monitors is made of polycarbonate. This is a durable plastic commonly used in
medical equipment. It is compatible with a variety of cleaners and surface sanitizers; however, there are
some chemicals that should not be used. The following is a summary of expected chemical compatibilities. It
is based on Honeywell HomMed’s testing and information verification.
Please note that the manufacturers of cleaning and sanitizing products may change their chemical
formulations at any time.
The monitor should never be immersed or soaked in any liquid.
Acceptable
Chemical Name
Acetic acid
Alcohols; isopropyl
Ammonium chlorides
Citric acid
Detergents
Formaldehyde
Hydrogen peroxide
Citric oils (but not orange)
Pine oils
Soap solutions
Sodium bicarbonate
Naptha
Ethanol
Ethylene glycol
Propylene glycol
Water
Enzymatic cleaners
Common Names or Example Brands
Vinegar
Septihol, Lysol spray
Formula 409 (also includes ethylene glycol)
Goo Gone (also includes Naptha)
Pine Sol
baking soda
Simple Green
Super Green
Enzol
Not Acceptable
Chemical Name
Acetone
Ammonia
Ketones (including methyl ethyl ketone)
Sodium hydroxide
Turpentine
Acetaldehydes(including glutaraldehyde)
Mineral spirits
Orange oil
Phenols
Sodium hypochlorite
Iodine
P4510EN.07
Common Brand Names
caustic soda
Cidex, Sporicidin
Vesphenell SE, LpH
bleach, Clorox
Wescodyne, Betadine
2/16/12
63
GENESIS™ DM MANUAL
Peripheral Medical Devices
Clean only the surfaces of the monitor and the accessories. Refer to the manufacturer’s cleaning
instructions for the accessories and peripheral devices. Qualified service technicians must perform any
cleaning, servicing, or repairs that require disassembly. Honeywell HomMed does supply peripheralspecific instruction sheets; contact Customer Service.
Cleaning the Genesis DM Monitor
Unplug the Genesis DM monitor. The monitor memory device retains all programmed settings even
when powered down.
Detach all accessories and peripheral medical devices.
To clean and disinfect, wipe the surfaces with isopropyl alcohol, and then wipe with a lint-free cloth
lightly moistened in water.
Cleaning the NIBP cuffs
1.
Remove bladders from each cuff.
2.
Using a mild disinfectant/detergent and warm water, wash cuffs by hand or in a washing machine.
3.
Hang cuffs to dry. DO NOT dry cuffs in a dryer.
4.
When cuffs are dry, replace the bladders.
To obtain accurate blood pressure readings, you must insert the correct bladder size for the cuff.
Inspection
Any time you visit a patient, inspect the Genesis DM monitor and accessories. Check the following:

Power cord is not frayed or connected to unauthorized equipment.

Cables are properly attached and in good condition.

All accessories are securely attached.

Monitor is not standing in or near water.

Monitor has not been moved to an unsuitable location.
If there is a frayed power cord or if the unit is attached to unauthorized equipment, unplug the unit and
notify Customer Service.
If a monitor has been dropped or damaged, call Customer Service. Qualified service personnel must
inspect any dropped or damaged units before they are assigned for use.
P4510EN.07
2/16/12
64
GENESIS™ DM MANUAL
Peripheral Medical Devices
Calibration verification
Your monitor components and scale are calibrated prior to shipment from the factory, so additional
calibration is not required at time of delivery, However, upon receiving your shipment from Honeywell
HomMed, carefully remove each piece of equipment from the box and verify that nothing has been
damaged and all pieces are in working order.
Honeywell HomMed recommends that you establish and routinely follow a calibration schedule for all
measurement devices, such as once annually as a minimum. If equipment malfunctions, has been dropped
or damaged, or appears to be out of calibration, you may choose to verify calibration using the
instructions in this section, or you may contact Customer Service. The Honeywell HomMed repair
policy is as follows:
If the item is under warranty and the repair is covered under the warranty, there is no charge for the
repair service.
If the item is no longer under warranty, or if the repair is not covered under the warranty, we will
charge a fee for the repair.
For additional warranty information, please refer to your contract.
This section provides instructions for calibration verification for Honeywell HomMed equipment. For
peripheral medical devices, follow the manufacturer’s service and maintenance instructions.
Scale calibration verification
This procedure requires the use of a calibrated weight greater than 30 pounds (14kg) and less than 500
pounds (227kg). To verify the scale’s calibration:
1.
Make sure that there is no weight on the scale.
2.
With the scale attached to the Genesis DM monitor, power up the monitor.
3.
From idle mode, press MANUAL to enter Manual Mode.
4.
Place a calibrated weight on the scale.
5.
Verify that the weight displayed is equal to the calibrated weight +/- 0.5 pounds (0.2 kg).
6.
If this procedure fails, please contact Customer Service to schedule an RMA.
P4510EN.07
2/16/12
65
GENESIS™ DM MANUAL
Peripheral Medical Devices
NIBP cuff calibration verification
!
CAUTION
RISK OF INJURY
Verifying calibration while the cuff is attached to a patient could cause bruising or other
injury.
DO NOT verify NIBP calibration while the cuff is attached to a patient.
To verify the NIBP cuff’s calibration:
1.
Secure the blood pressure cuff to a rigid fixture about the size of an arm.
2.
Tee-in a standard, calibrated mercury or digital manometer with the NIBP cuff.
3.
Press START + YES to enter the monitor’s Setup Mode.
4.
Scroll to Password menu and enter your password.
5.
Scroll to the Service Menu and press RETEST to enter the menu.
6.
Scroll to Calibration Verify, and press RETEST to enter the option.
Press START to begin the calibration verification. The NIBP pumps the pressure to 260 mmHg (+ 10
mmHg). Verify the monitor’s displayed blood pressure numbers to the manometer reading. The
manometer reading should agree to within +/- 3 mmHg or +/- 2%, whichever is greater.
Press STOP to cancel the reading and release the pressure.
Press START + YES to exit Setup Mode.
Optional finger sensor calibration verification
To verify oximeter calibration requires a pulse oximeter tester. Refer to the manufacturer’s manual for
testing procedures. For information about ordering testers, contact Customer Service.
Remote upgrades
Following each successful vital signs transmission, the home unit automatically checks the upgrade server
to identify any available upgrades. When available, the unit will attempt to download the new software
between 1:00 A.M. and 3:00 A.M. (local time).
P4510EN.07
2/16/12
66
GENESIS™ DM MANUAL
Peripheral Medical Devices
Peripheral Medical Devices
This section discusses each of the peripheral medical devices that are compatible with the Genesis DM
monitor. Specific instructions are included with the cables for each peripheral.
Please refer to the manufacturer’s manual for each peripheral for detailed care and maintenance
instructions.
The programming instructions in this section assume that you understand basic programming concepts
such as how to enter Setup Mode, use the monitor keys for programming, and navigate through the
menu options. If you do not know how to perform these tasks, please review the information in
Section 3: Programming the Genesis DM Monitor.

ECG

Glucose Meters
NOTE: In addition to assuring that the date and time are correctly set, also check for prior
readings in the device and clear/delete those readings as necessary.

Oximetry

Peak Flow Meters

Prothrombin Meters

Temporal Temperature
To ensure accurate trends, make sure that the device’s date and time are correct before transmitting
any test results. To set the date and time, refer to the manufacturer’s instructions.
P4510EN.07
2/16/12
67
GENESIS™ DM MANUAL
Peripheral Medical Devices
Health Frontier ecg@home
This section explains how to program the Genesis DM monitor to prompt for an ECG recording, setup
the ecg@home device, and instruct the patient in obtaining an ECG recording. For information about
viewing ECG recordings in LifeStream, refer to your LifeStream User Manual.
If you have any questions or concerns, call Customer Service. For repair or shipping information, refer
to the Getting Started section of this manual.
Assigning a COM port
To enable the monitor to prompt for and collect ECG data, you must assign a COM port. If you do not
assign a COM port, the monitor will not prompt for the device and it will not collect data from
that device even if the device is connected to the COM port.
It is also important to use an isolated port. Isolated ports protect patients from any potential of shock
when they use a device that is connected to the monitor.
The following instructions assume you know how to navigate in Setup Mode. If you have not
programmed a monitor, refer to Section 3 Programming the Genesis DM Monitor for complete
programming instructions.
P4510EN.07
2/16/12
68
GENESIS™ DM MANUAL
Peripheral Medical Devices
1.
Enter Setup Mode, scroll to the Option Menu and press RETEST to enter the menu.
2.
In the ECG Monitor menu, assign either COM1 or COM2.
3.
In the ECG Time menu, set the collection frequency.
4.
Exit Setup Mode.
The cable that connects the ecg@home to the Genesis DM monitor has a 4-pin connector on one end
and a jack on the other. The pin connector plugs into a com port on the Genesis DM, and the jack plugs
into the RS232 connection on the ecg@home. To attach the ecg@home to the Genesis DM monitor,
follow the steps below.
1.
Locate the RS232 connection on the side of the ecg@home device.
RS232
connection
Bottom
Top
2.
Plug cable jack into the ecg@home RS232 connection.
3.
Plug the 4-pin connector into the assigned COM port of the Genesis DM monitor.
Additional notes

At times the ECG reading may not be clear and the reading will show “noise,” a fuzzy, extra thick or
hard-to-determine ECG line:
The patient should attempt at least two times to obtain a reading without noise. Even if the second
reading has noise, the patient should still transmit the reading.

The ecg@home does not have an OFF button. The device automatically turns off after five minutes.
When connected to the Genesis DM monitor, the monitor turns the device off after a successful
transmission.
P4510EN.07
2/16/12
69
GENESIS™ DM MANUAL
Peripheral Medical Devices


The monitor erases the reading prior to turning the device off.
If the patient uses the device as a stand-alone (not connecting it to the monitor), the device erases
the old reading when the patient turns it on.
Please see the instructions from the manufacturer for approved use.
Hand-held method
Take the ECG reading with hand-held use of the ecg@home device. If this is too difficult, instructions are
also given for conducting the ECG reading with the device on a table top or other suitable surface.
The Genesis DM monitor will prompt the patient through each of the following steps. After the initial
prompt informing the patient that it is time to take an ECG reading, the prompts will continue to walk the
patient through the following steps:
1.
2.
3.
4.
5.
6.
7.
Apply 2-3 drops of electrode solution on each electrode of the device. Be careful to drop liquid
onto the electrode rather than touch the bottle tip to the electrode.
Press START on the ecg@home device.
An asterisk displays in the upper right corner of the screen and a heart symbol flashes on the
display.
Place your thumbs on the electrodes and press gently.
After a few seconds, the heart symbol disappears and the device beeps. The wave symbol moves
from left to right across the screen. Heart pulse rate also displays as the device collects the ECG
reading.
When the collection is complete, the wave symbol stops flashing, the device beeps twice, and a
second asterisk displays under the first one.
When the second asterisk displays, press YES on the Genesis DM monitor to upload the ECG
results. The number “32” displays to indicate that the communication port is open and the device is
transmitting.
When the upload is complete, the monitor displays the text: “Data Upload finished”
Table-top method
If the patient has difficulty obtaining a reading without artifact, instruct them to obtain a reading by placing
the device on a sturdy table.






Before taking a reading, make sure that both hands are clean.
Place the ecg@home device on a sturdy table low enough that the patient can comfortably place
his/her fingers on the device.
Sitting comfortably in a chair, place a pillow in the lap and position it so that it will support both
arms.
Using the pillow for support, extend both arms (as straight as possible without movement) and rest
the ring fingers gently on the electrodes. (The middle finger also works well.)
Relax and breathe normally.
Do not move or talk during the recording.
After the initial prompt informing the patient that it is time to take an ECG, the Genesis DM monitor
prompts the patient through the following steps:
P4510EN.07
2/16/12
70
GENESIS™ DM MANUAL
Peripheral Medical Devices
1.
Apply two to three drops of the electrode solution on both electrodes of the ECG device. Be
careful to drop liquid onto the electrode rather than touch the bottle tip to the electrode.
2.
Press the START key on the ECG device.
3.
An asterisk displays in the upper right corner of the screen and a heart symbol flashes on the
display.
4.
Place your ring (or middle) fingers gently on the two electrodes to take an ECG reading.
After a few seconds, the heart symbol disappears and the device beeps. The wave symbol moves from left to
right across the screen. Heart pulse rate also displays as the device collects the ECG reading. When the
collection is complete, the wave symbol stops flashing, the device beeps twice, and a second asterisk displays
under the first one. When the second asterisk displays, press YES on the Genesis DM monitor to upload the
ECG results. The number “32” displays to indicate that the communication port is open and the device is
transmitting. When the upload is complete, the monitor displays the text:
“Data Upload finished”
Using the external auxiliary electrode
If the patient is unable to use both hands, use the auxiliary electrode (leg lead). To obtain a recording
with the auxiliary electrode, instruct the patient to do the following:
1.
Put two to three drops of the electrode solution on the external electrode and on the right thumb
electrode. Be careful to drop liquid onto the electrode rather than touch the bottle tip to the
electrode.
2.
Hold the ecg@home device in the right hand with the thumb gently pressing the electrode.
3.
Place the external electrode on the upper left leg.
4.
With the left hand, press START to turn the unit on and then gently hold the external electrode in
place while the device takes a reading.
5.
When the upload is complete, the monitor displays the text, “Data Upload finished.”
Cleaning the ecg@home
Always disconnect the device from the Genesis DM monitor prior to cleaning. Clean the device with a
soft cloth moistened with the electrode solution. DO NOT use water, detergent or any other cleaning
product.
To clean the electrodes, moisten a cotton swab with the electrode solution and gently swab clean.
To disinfect the unit, use a solution of isopropyl alcohol (60-70%) and water (30-40%).
P4510EN.07
2/16/12
71
GENESIS™ DM MANUAL
Peripheral Medical Devices
Glucose Meters
Overview
The Genesis DM monitor transmits up to twelve of the most recent, valid glucose test results from a
glucose meter to the centralized LifeStream reporting database. A valid reading has the test time and
date. If any reading is missing a time and date stamp, the monitor will not collect it.
The monitor does not erase any data from the glucose meter and does not collect or transmit controlsolution test results obtained in test mode or designated as control results.
To help ensure accurate trends, verify that the glucose meter date and time settings are correct
before transmitting any test results or after replacing batteries. To set the date and time, refer to
the manufacturer’s instructions.
If you want to erase data from LifeScan or Roche Accu-Chek models, you can purchase software and a
customized cable from LifeScan or Roche.
The Genesis DM monitor is compatible with the following glucose meters:














Bayer Ascensia Elite® XL
Bayer Ascensia Breeze®
Bayer Ascensia Breeze 2™
Bayer Ascensia Contour™
Home Diagnostics Prestige IQ®
Home Diagnostics TrueTrack Smart System®
LifeScan One Touch Basic®
LifeScan One Touch Profile®
LifeScan One Touch Ultra®
LifeScan One Touch Induo®
LifeScan One Touch Ultra2®
Roche Accu-Chek Advantage®
Roche Accu-Chek Aviva®
Roche Accu-Chek Compact™ Plus
For further information, contact:
Bayer: http://www.bayerdiabetes.com
Home Diagnostics: http://www.homediagnosticsinc.com
LifeScan: http://www.lifescan.com
Roche Diagnostics: http://www.roche.com
NOTE: The above information is subject to change at any time and is provided as a courtesy. Refer to
the manufacturer’s materials to help assure the accuracy of contact information.
Safety information
Remind patients to thoroughly read the user documentation provided by the manufacturer and to pay
close attention to any cautions or warnings.
P4510EN.07
2/16/12
72
GENESIS™ DM MANUAL
Peripheral Medical Devices
Assigning a COM port
You must assign a COM port to the glucose meter or the monitor will not prompt the patient for data
or communicate with the device to collect data (even if the device is connected to the COM port).
Assigning a COM port triggers the monitor to include glucose meter prompts during the monitoring
session, to communicate with the glucose meter, and to upload test results.
The following instructions assume you know how to navigate in Setup Mode. If you have not
programmed a monitor, refer to Section 3 Programming the Genesis DM Monitor for complete
programming instructions.
1.
Enter Setup Mode, Scroll to the Option Menu and press RETEST to enter the menu.
2.
In the GlucometerType menu, select the appropriate glucose meter.
3.
In the Glucometer menu, assign either COM1, COM 2 or COM-A, whichever is appropriate for
cable.
4.
In the Glucometer Time menu, set the collection frequency.
5.
Exit Setup Mode.
Attaching the glucose connector cable
One end of the connector cable has a 4 or 6-PIN-connector that plugs into the com port on the rear of
the monitor. The other end of the cable either has a jack connector that plugs into the communication
port on the glucose meter, or a DB-9 connector that plugs into another cable. The second cable then
plugs into the glucose meter.
The pins in the connector cables can be damaged if you attempt to plug the cable into a receptor that
doesn’t match the pin configuration of the cable. To prevent damage to the pins in the connector cables,
always program the monitor first and then make sure you plug the cable into the port you programmed
for glucose.
Please see any instructions included with your specific cable for additional details.
P4510EN.07
2/16/12
73
GENESIS™ DM MANUAL
Peripheral Medical Devices
Patient training
The Genesis DM begins the process of uploading data as soon as it detects the device. The audio and
text prompts inform the patient that the upload is in progress and then announces when the data upload
is finished. If the patient fails to attach the device and the monitor times out, Genesis DM moves on to
the next scheduled activity in the monitoring session. The following table contains glucose meter
prompts.
Condition
Text and audio
Genesis DM is ready to initialize
communication
Please attach the glucose meter to the monitor.
Please make sure the cable is attached to the port marked COM (1, 2
or A) on the back of the monitor.
If this is the first time this Genesis DM has
received information from this
glucose meter
New glucose meter detected
Transmission
Receiving Data, please wait
The upload is complete
Data upload is finished
The Genesis DM monitor does not display glucose test results.
Glucose meter specifics
Some glucose meters require that the patient manually turn them on before they are able to transmit
data. The text message that displays on the face of the glucose meter differs from meter to meter. The
following table explains important differences between glucose monitors.
Meter
LifeScan One Touch
Basic
LifeScan One Touch
Profile
LifeScan One Touch
Ultra
Turn on to
transmit?
Glucose meter text display during transmission
Yes
Code ( ) Insert strip
No
PC
Bayer Ascensia Elite XL
Yes
- - - MEM followed by a count down of test results
Roche Accu-Chek
Home Diagnostics
Prestige
No
No display (unit does not turn on)
Normal display screen shows during transmission, device turns off
when transmission complete
P4510EN.07
No
2/16/12
74
GENESIS™ DM MANUAL
Peripheral Medical Devices
The following table contains important notes specific to glucose meter types.
Meter
Bayer Ascensia Elite XL
LifeScan One Touch
Roche Accu-Chek
Note
The meter must be OFF when you attach it to its connector cable. If not, the device
will not transmit readings.
Until the patient turns the meter on, the Genesis DM monitor CANNOT detect
the device. The monitor will prompt the patient to verify the cable connection to
the com port. However, as soon as the
device is on and monitor detects it, the prompt will stop.
The meter does not transmit readings when the batteries are low.
Instruct the patient to follow the manufacture’s recommendations and instructions
to maintain battery power.
The meter must remain off during transmission. It is also important that the patient
does not press any buttons during transmission.
Patient Instructions
Before the monitoring session the patient should finish collecting test results, turn their glucose meter
off, and place their glucose meter next to the monitor before the monitoring session.
When it is time for the monitor to collect glucose test results, the monitor will prompt for the device. If
the monitor does not detect the device after the initial prompt, it will also prompt the patient to make
sure that the cable is connected to the COM port programmed for the device. The monitor repeats
both prompts.
1.
2.
Complete all tests and place the device next to the connector cable before the monitoring session.
Make sure the glucose meter is off.
Genesis DM prompts:
“Please attach the glucose meter to the monitor. Be sure that the cable is attached to COM (1, 2 or A).”
3.
4.
Plug the connector cable into the communication port on the meter.
If using a model that requires that you manually turn it on, do so now.
The monitor collects the data and moves on to the next device or concludes the monitoring session.
P4510EN.07
2/16/12
75
GENESIS™ DM MANUAL
Peripheral Medical Devices
Oximeter
A finger sensor attaches to an external oximeter module cable which measures oxygen saturation
(SpO2) and heart rate. If a patient uses the optional oximeter, the oximeter heart rate over-rides the
NIBP heart rate in the monitor display. Genesis DM transmits both the NIBP and oximeter heart rate
values. LifeStream stores both values in the database but only displays the oximeter heart rate.
Assigning a COM port
You must assign an isolated COM port in order for the Genesis DM to prompt for collection during the
monitoring session.
To assign a COM port, follow the instructions below. The oximeter can also be programmed in the
monitor Setup Wizard. The COM port is automatically assigned at that time.
The following instructions assume you know how to navigate in Setup Mode. If you have not
programmed a monitor, refer to section 3 Programming the Genesis DM Monitor for complete
programming instructions. In addition to the monitor, the oximeter can also be assigned via the Quick
Setup Wizard available when the monitor is first turned on.
1.
In Setup Mode, scroll to the Option Menu and press RETEST to enter the menu.
2.
Scroll to Oximeter and press RETEST.
3.
Using either the YES or NO key, scroll to the correct COM port (COM1 or COM2), and press
RETEST.
4.
In the Oximeter Time menu, set the collection frequency.
5.
Exit Setup Mode.
Attaching the optional oximeter cable and finger sensor
Follow the steps below to attach the oximeter cable and finger sensor to the monitor.
1.
Plug the 4-pin connector (located on the oximeter module) into the isolated port you assigned in
programming.
P4510EN.07
2/16/12
76
GENESIS™ DM MANUAL
Peripheral Medical Devices
2.
To attach the finger sensor, hold the plug so that the letter B is on the bottom.
3.
Push the plug into the module port until it locks into place.
Patient training
If you have assigned a COM port and have the oximeter module cable connected to the Genesis DM, the
monitor includes oximeter prompts with the blood pressure prompts as follows:

Place the blood pressure cuff on the right arm above the elbow and tighten it securely.
Cue text: “Put cuff on arm”

Place the finger probe (sensor) on the middle finger of your left hand with the design facing up.
Cue text: “Put sensor on finger” (pulsing bar graph also displays)

Rest your arm as instructed by your clinician.
Cue text: “Rest arm as instructed”

When blood pressure and oximetry values are collected, the monitor instructs the patient to
remove the sensor and blood pressure cuff.
Cue text: “Remove sensor and cuff”
Patient Instructions
Stress the importance of always using the oximeter on the hand opposite of the arm wearing the blood
pressure cuff.
Point out the raised finger design on the top of the oximeter, and explain that the finger sensor is worn with
the design on the top of the finger. Honeywell HomMed recommends using the middle finger to make it
easier for the patient to press monitor keys. However, when necessary, the patient may use a different
finger.
Instruct the patient to insert the middle finger into the oximeter until it reaches the finger stop and
completely covers the detector.
The oximeter is properly placed if the pulse bar on the monitor display shows a strong pulse while collecting
the SpO2 reading. Once the monitor begins to collect the blood pressure and SpO2 readings, the patient
should remain still to ensure an accurate reading.
P4510EN.07
2/16/12
77
GENESIS™ DM MANUAL
Peripheral Medical Devices
Peak Flow Meters
This section provides information and instructions for programming the Genesis DM monitor, attaching
the peak flow cable and meter, and teaching patients how to upload test results. For information about
peak flow meter specifications and capabilities, or for instructions on how to use, test, or maintain
meters, refer to the manufacturer’s material.
Overview
The Genesis DM monitor is compatible with the PiKo-1 manufactured by Pulmonary Data Services
(PDS). The PiKo-1 can store up to 96 tests with the warning indicators and time and date stamp.
However, the Genesis DM monitor only downloads the 12 most recent readings.
Safety information

The PiKo-1 is intended for use as a single-user device. Instruct patients that no one else is to use
their peak flow meter.

Remind patients to thoroughly read the user documentation provided by the manufacturer and to
pay close attention to any cautions or warnings.

The User’s Manual contains the following warning: “If you use the PiKo-1 above 1000 feet elevation,
(300m), adjust your values by adding 1.5%. To do this, multiply the PEF & FEV1 by 1.015.”
Assigning a COM port
You must assign a COM port to the peak flow meter or the monitor will not prompt the patient for
data or communicate with the device to collect data (even if the device is connected to the COM port).
Assigning a COM port triggers the monitor to do the following:



Include peak flow meter prompts during the monitoring session
Communicate with the peak flow meter
Upload test results
The following instructions assume you know how to navigate in Setup Mode. If you have not
programmed a monitor, refer to Section 3 Programming the Genesis DM Monitor for complete
programming instructions.
1.
2.
3.
4.
5.
Enter Setup Mode, and enter the password.
Scroll to and enter the Option Menu.
In the Peak Flow menu, assign COM A.
In the Peak Flow Time menu, set the collection frequency.
Exit Setup Mode.
P4510EN.07
2/16/12
78
GENESIS™ DM MANUAL
Peripheral Medical Devices
Attaching the peak flow meter cradle to the Genesis DM monitor
The peak flow meter transmits data to the Genesis DM monitor via its cradle and a Honeywell
HomMed connector cable. The Honeywell HomMed cable has a black, 9-pin connector that matches the
cradle cable connector, and a round, 6-pin connector that matches COM-A on the rear of the Genesis
DM monitor. To attach the peak flow cradle to the monitor, follow the steps below:
1.
Plug the Honeywell HomMed cable 6-PIN connector into COM-A.
2.
Plug the Honeywell HomMed 9-pin male connector into the cradle cable connector. If you have
difficulty, make sure that the pins in the Honeywell HomMed connector properly align with the
holes in the cradle connector.
3.
Tighten the screws on the cradle connector until snug.
4.
Arrange the equipment so the patient can easily place their peak flow into the cradle and the cables
are out of the way.
5.
Place the peak flow meter into the cradle.
Patient training
The Genesis DM monitor begins the process of uploading data as soon as the monitor detects the
device. The audio and text prompts inform the patient that the upload is in progress and then
announces when the data upload is finished.
If the patient fails to turn the peak flow meter on and place it into the cradle, Genesis DM skips peak
flow collection and moves on to the next device. If the Genesis DM has collected all data from all
devices, the monitor concludes the monitoring session.
Once the Genesis DM establishes communication with the peak flow meter, Genesis DM displays and
says:
“Receiving data, please wait; Data upload is finished.”
The Genesis DM monitor does not display peak flow meter test results.
For a complete list of prompts, refer to Appendix B
Patient Instructions
Patients should complete all testing and place the peak flow meter in its cradle before the monitoring
session. The patient should then follow the prompts to transmit data.
1.
When the monitor prompts to activate the peak flow, pick up the device and press the button
twice.
2.
The meter turns on and displays a zero with dashes going through it.
P4510EN.07
2/16/12
79
GENESIS™ DM MANUAL
Peripheral Medical Devices
3.
Return the peak flow to its cradle. The Genesis DM communicates with the device and downloads
data. During the transmission, data flashes on the peak flow display but test results do not display on
the monitor.
4.
When data collection completes, the monitor prompts for the next programmed device or, if data
collection is complete, concludes the monitoring session.
Prothrombin Meters
This section provides information and instructions for programming the Genesis DM monitor, attaching
the prothrombin cable and meter, and teaching patients how to upload test results. For information
about prothrombin meter specifications and capabilities or for instructions on how to use, test, or
maintain meters, refer to the manufacturer’s material.
Overview
The Genesis DM monitor is compatible with both of the following prothrombin meters:
• ProTime prothrombin meter manufactured by International Technidyne Corporation (ITC).
• HemoSense® INRatio PT/INR testing system from HemoSense, Inc.
The prothrombin meter transmits only the latest reading to the Genesis DM monitor. The monitor
does not erase any data from the prothrombin meter.
Prothrombin meters automatically synchronize their date and time to the monitor. If you need to
change meter dates or times, refer to the manufacturer’s instructions.
Safety information
!
CAUTION
ELECTRICAL SHOCK
Using a prothrombin meter while it is attached to the Genesis DM monitor could
possibly result in receiving an electrical shock.
DO NOT attempt to take PT/INR readings with a prothrombin time meter when it
is connected to the Genesis DM monitor.
Disconnect the prothrombin time meter before use.
Always follow the manufacturer’s instructions for proper use.
Instruct the patient to never use their prothrombin meter when it is attached to the Genesis DM
monitor.
Remind patients to thoroughly read the user documentation provided by the manufacturer and to pay
close attention to any cautions or warnings.
P4510EN.07
2/16/12
80
GENESIS™ DM MANUAL
Peripheral Medical Devices
Assigning a COM port
You must assign a COM port to the prothrombin meter or the monitor will not prompt the patient for
data or communicate with the device to collect data (even if the device is connected to the COM port).
Assigning a COM port triggers the monitor to include prothrombin prompts during the monitoring
session, to communicate with the prothrombin meter, and to upload test results.
The following instructions assume you know how to navigate in Setup Mode. If you have not
programmed a monitor, refer to Section 3 Programming the Genesis DM Monitor for complete
programming instructions.
1.
Enter Setup Mode.
2.
Scroll to and enter the Option Menu.
3.
In the PT Monitor menu, assign COM-A.
4.
In the PT Monitor Time menu, set the collection frequency.
5.
Exit Setup Mode.
Attaching the prothrombin meter connector cable
Prothrombin meter cables are white with a green color band. One end of the cable has a 6-PIN
connector that plugs into the non-isolated ports on the rear of the Genesis DM monitor. The other end
of the cable has a 9-PIN connector that plugs into the prothrombin data port located on the back of the
prothrombin device.
1.
Plug the 6-PIN connector into COM-A.
2.
Place the cable so that the patient can easily attach the prothrombin meter.
P4510EN.07
2/16/12
81
GENESIS™ DM MANUAL
Peripheral Medical Devices
Patient training
Refer to the manufacturer’s written material for instructions on how to access and navigate to menu
options.
Ideally, before the monitoring session, the patient should place the prothrombin meter next to
the monitor and finish collecting test results. The monitor does not prompt or allow time for the
patient to take a test during the monitoring session because the time required for prothrombin testing is
too long.
When the monitor prompts the patient to attach the device, the patient has a few minutes to turn the
device on, navigate to the Print/Send menu, and transmit the latest test result.
The Genesis DM monitor does not display the test results.
For a complete list of possible prothrombin prompts, refer to Appendix C: Prompt Text.
Patient Instructions
Instruct the patient to complete all tests, and place the meter next to the Genesis DM monitor before the
session. Have the patient turn the meter on before connecting it to its cable. Otherwise, the meter will
open PC menus that the patient would not normally use.
Condition
Text and audio
Transmission
Receiving data, please wait
Data upload is finished
When prompted to attach the device, instruct the patient to follow these steps:
1.
2.
3.
4.
5.
6.
Turn the prothrombin time meter on.
Plug the prothrombin time meter cable into the data port located on the back of the device.
When the self-check is complete, press
to go to the Main Menu.
Select SHOW RESULTS and press
.
Go to PRINT/SEND DATA and press
.
Go to THIS RESULT and press
to transmit.
If the latest reading collected by the prothrombin time meter was an error reading, the Genesis DM
monitor collects the error reading inserting zeros for values.
A common prothrombin time meter error is Error #81 which occurs when there was an insufficient
amount of blood for accurate testing.
P4510EN.07
2/16/12
82
GENESIS™ DM MANUAL
Peripheral Medical Devices
Temporal Temperature
The Exergen TemporalScanner, available from Honeywell HomMed, can transmit temperature readings
to the Genesis DM monitor.
The scanner is activated when the patient presses the red ON button and continues to scan and
transmit until the patient releases the button. The device scans for and transmits the highest
temperature so the last reading is always the highest temperature reading.
The monitor waits five seconds after the receipt of the last temperature reading. If no further readings
are received within the five second period, the last reading is bundled into the data packet for
transmission to the monitoring application.
If the patient resumes using the scanner within the five-second period, the monitor waits until readings
once again stop, allows another five-second grace period, and then bundles the last reading received into
the data-packet.
Assigning a COM port
You must assign a COM port to the scanner or the monitor will not prompt the patient for data or
communicate with the device to collect data (even if the device is connected to the COM port).
Assigning a COM port triggers the monitor to include temperature prompts during the monitoring
session and to communicate with the scanner.
The following instructions assume you know how to navigate in Setup Mode. If you have not
programmed a monitor, refer to Section 3 Programming the Genesis DM Monitor for complete
programming instructions.
The TemporalScanner requires an isolated port (COM-1 or COM-2). To assign the COM port, follow
the steps below.
1.
Enter Setup Mode, scroll to and enter the Option Menu.
2.
Scroll to the External Temp menu and press MANUAL.
3.
Assign COM-1 or COM-2.
4.
In the External Temp Time menu, set the collection frequency.
5.
Exit Setup Mode.
Attaching the TemporalScanner to the monitor
The scanner has a cord with a 4-pin mini-DIN connector that plugs into a Genesis DM com port.
To attach the scanner:
1.
Plug the temperature cable’s connector into the COM port that you programmed for the
TemporalScanner (COM-1 or COM-2).
2.
Place the scanner to the side of the monitor so that it is easily accessible to the patient.
P4510EN.07
2/16/12
83
GENESIS™ DM MANUAL
Peripheral Medical Devices
Patient training
The Exergen TemporalScanner Reference Manual contains complete collection instructions as well as
preferred and alternate sites for measurement. Determine the best method of measurement for the patient,
and instruct the patient accordingly.
When prompted, the patient should use the scanner as instructed. The monitor displays the word
“Temperature” but DOES NOT display the temperature reading. The monitor then moves onto the next
scheduled activity in the monitoring session.
Prompts for temporal temperature
The monitor prompts for temperature after collecting blood pressure as follows:

It is now time to take your temperature.
Cue text: Time to take your temperature

Please take your temperature as instructed.
Cue text: Take temperature as instructed
The patient may skip collecting temperature by pressing the NO key after the temperature prompt.
Patient Instructions
If appropriate, remind the patient to brush hair away from the forehead so the probe can maintain contact
with skin as it slides across the forehead.
1.
When the monitor prompts for temperature, pick up the TemporalScanner and place the probe on
the center of your forehead.
2.
Press and hold the red ON button.
3.
Slowly slide the probe across your forehead to your hairline. Do not go down the side of your face.
4.
Lift the probe and place it on the skin just behind your ear lobe.
5.
Release the button and put the scanner down.
6.
Continue following the prompts to complete your monitoring session.
P4510EN.07
2/16/12
84
GENESIS™ DM MANUAL
Appendix A: Subjective Questions
APPENDICES
Appendix A: Subjective Questions
The following chart lists the fifty-one subjective questions offered in the Genesis DM Questions menu.
Some of the questions include Educational Content, which is shown below the relevant question.
Subjective Questions
Number
00
01
Questions
Off
Are you experiencing more difficulty breathing today compared to a normal day?
Stress MAY cause difficulty breathing
IF ORDERED - Use oxygen
Pace yourself - don't overdo
If breathing difficulty continues
OR New cough - Can't do activities
Discuss with CARE PROVIDER
02
Are you more tired today compared to a normal day?
More tired ~ limited activities:
May be a change in\n medical condition.
Still tired? Can't do usual activities?
Chest pain, difficult breathing, cough
Discuss with CARE PROVIDER
03
Have you been using your inhalers more than usual?
Increased need for inhalers signals\n a change in condition.
Use inhalers AS PRESCRIBED
If difficult to breathe, chest tight or in
pain, or sudden cough that continues
Contact your CARE PROVIDER
04
Did you need extra pillows to sleep comfortably last night?
Using extra pillows to sleep may\n signal change. Use oxygen &
inhalers AS PRESCRIBED
Contact your CARE PROVIDER:
EXTRA pillows CONTINUES
Tired - trouble breathing - cough
05
Have you had to limit your activities more than usual?
Limited activities and more tired:
ay be a change in
medical condition.
Still tired? Can't do usual activities?
More difficult breathing or new cough
Discuss with CARE PROVIDER
P4510EN.07
2/16/12
85
GENESIS™ DM MANUAL
Appendix A: Subjective Questions
06
07
08
Do you need your clinician to contact you today?
Are you out of any of your medications?
Are you having difficulty taking any of your medications?
Problems taking medicines\n-side effects
Take medicines as directed.
Read pharmacy directions.
Discuss with CARE PROVIDER
09
Are you having difficulty following your diet?
Diet is important. Diet affects how
you feel. Many reasons for not
following diet
Work with CARE PROVIDER
for recommendations.
That will help
10
Are you having any nausea today?
Blood sugar affected by: Diet – meds
- illnesses - infections - physical
activity. Readings: too low/too high
Followed diet - taken medications?
Still not under control.
Call Care Provider.
11
12
13
14
15
Are you having trouble controlling your blood sugar?
Are you having any pain?
Have you been having any chest pain?
Are you having any headaches?
Have your ankles been swollen more than usual?
Ankle swelling caused by:
What you eat & medicines you take
Elevate legs, Limit salt IF ORDERED
Call CARE PROVIDER:
Feet or legs more swollen
16
Are you having any difficulty standing or walking?
Difficulty standing or walking.
At risk for falls.
Learn cause - what to do.
Discuss with Care Provider.
17
18
19
20
21
22
23
P4510EN.07
Have you been having any new numbness or tingling?
Are you having any dizziness?
Have you noticed any bleeding today?
Have you noticed any new bruising today?
Are you having any trouble with urination?
Are you having any trouble with your bowel movements?
Has your wound changed in size or color?
2/16/12
86
GENESIS™ DM MANUAL
Appendix A: Subjective Questions
24
25
26
27
Have you noticed a change in odor from your wound?
Have you noticed a change in the drainage from your wound?
Have you had to use your nitroglycerin in the last day?
Have you had to use your oxygen in the last day?
Need for oxygen signals a change.
Low oxygen can make you feel:
Short of breath, tired, foggy or dizzy
Improving oxygen can help
DON'T turn oxygen flow up
CHECK with CARE PROVIDER
28
Have you developed a cough?
New cough?
Usual treatment not working?
Possible change in condition.
Short of breath too?
Report to Health Care Provider.
29
30
31
Have you noticed a decrease in your appetite?
Are you having difficulty swallowing?
Have you had a fall in the last day?
Falls are serious
Weakness, pain
Change in heart rate or BP
Learn the cause
Report falls to Care Provider
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
P4510EN.07
Are you having difficulty taking care of your wound?
Have your hands or face been more swollen in the last day?
Has there been any protein in your urine in the last day?
Do you have severe heartburn today?
Have you noticed a decrease in your baby’s movement today?
Have you had any blurred vision today?
Have there been any changes in the medication you are taking?
Are you having difficulty following your weight-loss plan?
Are you having difficulty following your exercise plan?
Do you feel more anxious or upset today?
Has your mood been more depressed this week compared to a normal week?
Are you having difficulty managing stress this week compared to a normal week?
Are you having difficulty following your smoking cessation plan?
Are you having difficulty following your alcohol reduction plan?
Are you having difficulty understanding your diagnosis?
2/16/12
87
GENESIS™ DM MANUAL
Appendix A: Subjective Questions
Have you had 2 plus or greater protein in your urine today?
47
Have you been to the emergency room this week?
48
Were you admitted to the hospital any time this week?
49
Has your doctor added, deleted, or changed any of your medications this week?
50
51
Did you have an unexpected visit to your physician this week?
Ten (10) Questions and Disease Specific Symptom Management (DSSM)
These question are repeats of existing questions and will trigger the DSSM Educational Components
(Educational Content) to be displayed during vitals transmission.
52
Are you experiencing more difficulty breathing today compared to a normal day? (same as Question Number 1)
Common causes for increased difficulty breathing may be caused by strenuous activity, anxiety or not following
recommendations made by your health care provider. Ensure you remain compliant with your medications, your
oxygen, or other respiratory treatment as prescribed. If your breathing difficulty progresses, you suddenly develop a
new cough, or have to severely limit your usual activity, it may signal a serious change in your medical condition
which should be discussed with your health care provider
53
Have you had to limit your activities more than usual? (same as Question Number 5)
Limiting your activities may signal a change in your medical condition. If you are unable to resume normal activities,
have difficulty breathing, or suddenly develop a new cough, contact your health care provider
54
Are you having difficulty taking any of your medications? (same as Question Number 8)
Difficulty taking medications, taking medications at irregular times or not taking medications consistently may affect
your medical condition and in some cases may
make your symptoms worse. Some medications have serious side effects if taken incorrectly or if stopped abruptly.
Contact your health care provider if you are having difficulty taking your medications or have questions
55
Are you having trouble controlling your blood sugar? (same as Question Number 11)
Blood sugar control may be affected by diet, weight, medications, physical activity, illnesses and infections.
Uncontrolled blood sugar, over time, may lead to medical complications. If you have taken your medications as
prescribed and followed your diet, adjustments to your medication may be needed. Always ensure you report any
symptoms of high or low blood sugar to your health care provider
56
Are you having any headaches? (same as Question Number 14)
There are many reasons people develop a headache. Sometimes, headaches may be related to elevations in blood
pressure, also known as hypertension. Diet, weight control, exercise, and medications are used to treat high blood
pressure. If you experience a headache that is persistent, contact your health care professional.
57
Have your ankles been swollen more than usual? (same as Question Number 15)
Swollen ankles may occur with heart and lung conditions. Increased swelling may be related to changes in diet,
medications, or activities. Limiting your salt intake and elevating your feet may be helpful. If swelling continues to
increase, your breathing becomes more difficult, or you develop a cough; contact your health care provider
P4510EN.07
2/16/12
88
GENESIS™ DM MANUAL
Appendix A: Subjective Questions
58
Are you having any difficulty standing or walking? (same as Question Number 16)
Difficulty walking or standing may be caused by medications, weakness, and/or physical pain. Difficulty standing or
walking could also be due to a change in your medical condition; and may increase your risk for falls
59
Are you having any dizziness? (same as Question Number 18)
Dizziness can be caused by many medical conditions. Sudden changes in blood pressure, heart rate or blood sugar can
cause dizziness as can side effects from certain medications. Dizziness, when accompanied by headache or blurred
vision, may indicate a more serious condition which should be discussed with your health care provider
60
Have you developed a cough? (same as Question Number 28)
A newly developed cough may signal a change in your medical condition. If your cough is unrelieved with prescribed
medication or if you develop difficulty breathing contact your health care provider.
61
Have you had a fall in the last day? (same as Question Number 31)
Falls can be a major cause of injury, and can lead to serious complications. Falls may be caused by medications, pain,
weakness, or changes in heart rate or blood pressure or
be accidental. Report any falls to your health care provider, even if it seems like you did not injure yourself.
62
Have you experienced any problems with your balance in the last day?
63
Have you almost fallen in the last day?
64
Did you skip your prescribed diabetes medication today?
Missing your medication may cause high blood sugar later. Contact your doctor if you have questions about your
insulin dose.
65
Did you adjust your medications based on your blood glucose results in the past day?
66
Have you not been able to take your blood sugar readings as prescribed?
Taking your blood sugar readings as instructed by your health care provider is an important part of managing your
health. Please follow your health care provider’s instructions.
67
Are you feeling stressed about having or managing your diabetes?
Diabetes can affect you physically and emotionally. Living with it every day can make you feel discouraged, stressed,
or even depressed. It’s natural to have mixed feelings about your diabetes management and experience highs and
lows. The important thing is to recognize these emotions as normal. Take steps to reduce the negative impact they
could have on your self care. If symptoms persist, please call your health care provider.
68
P4510EN.07
Did you have difficulty eating healthy in the past 24 hours?
2/16/12
89
GENESIS™ DM MANUAL
Appendix A: Subjective Questions
To keep your blood sugar steady, follow your recommended diet.
69
Did you have difficulty meeting your activity or exercise goals in the past 24 hours?
Regular activity is important to every day health. Increase your activities as tolerated. include regular activity as part
of your schedule. Have fun by doing activity with a friend. Start small and build over time. Reward yourself for
meeting goals and sticking with the plan. Don’t get discouraged. Keep at it.
70
Have you had to limit your daily activities or movement due to pain in the past 24 hours?
71
Have you noticed a new rash in the past 24 hours?
72
Has your rash gotten worse in the past 24 hours?
73
Have you used any over-the-counter medications in addition to those prescribed for you in the past 24 hours?
74
Have you noticed any increases in thirst in the past 24 hours?
75
Have you had any recent fever or chills?
76
Do you feel worse since starting your new medications?
77
Are you running low on any medical supplies?
78
Is your temperature higher than 38 C?
79
Is your temperature higher than 100 F?
80
Is your Zoe Fluid Monitor reading under 18?
81
Is your Zoe Fluid Monitor reading over 30?
82
Is your PiKo-6 FEV1/FEV6 ratio under 70%?
83
Do you have any new or increased sputum, mucus or phlegm?
84
Is your sputum, mucus or phlegm thicker today?
85
Do you feel that you may have a chest infection?
86
Did you monitor your vitals without your oxygen?
87
Did you have blood work drawn today?
P4510EN.07
2/16/12
90
GENESIS™ DM MANUAL
Appendix A: Subjective Questions
88
Did you go to the doctor this past week?
89
Are you taking any prescribed or over-the-counter blood thinning medications?
90
Do you feel dizzy or light headed today?
91
Do you have increased sputum, mucus or phlegm?
92
Is your sputum, mucus or phlegm a different color today?
93
Are you experiencing any burning with urination or increased frequency?
94
Was your most recent blood sugar lower than usual?
95
Did you seek medical treatment?
96
Has your doctor added, deleted, or changed any of your medications this week?
Note:
Questions 62 through 96 are available in English and Spanish. If you program your monitor
for another language, these questions will appear in English with no audio.
Reminders
Number
Reminders
101
You have an appointment with your doctor today. Please write down any questions you may have and bring them to
your appointment.
102
You have an appointment with your doctor this week. Please write down any questions you may have and bring them
to your appointment.
103
You have an appointment with your health care provider today. Please write down any questions you may have and
bring them to your appointment.
104
You have an appointment with your health care provider this week. Please write down any questions you may have
and bring them to your appointment.
105
You are scheduled to have a visit from your home care nurse today.
106
You are scheduled to have a visit from your nursing assistant today.
107
You are scheduled to have a visit from your respiratory therapist today.
108
You are scheduled to have a visit from your physical therapist today.
P4510EN.07
2/16/12
91
GENESIS™ DM MANUAL
Appendix A: Subjective Questions
109
Please check your medications to see if you need any refills.
110
Remember you have an event with your family this week.
111
Remember your social club is having a meeting this week.
112
Remember your grandchildren are coming over this week, have fun.
113
Please remember to change your dressing today.
114
Please remember to keep feet and legs elevated while sitting.
115
Please remember to eat foods low in Sodium.
116
Please remember to wear your oxygen as prescribed.
117
Please remember to use your nebulizer as prescribed.
118
Please remember to do your exercises as prescribed.
119
Please remember to take your pain medications as needed and as prescribed.
120
Flu season - avoid large crowds
121
Immunizations up to date
122
Get your flu shot
123
Happy Birthday!
124
Happy Anniversary!
125
Read medicine labels
126
Know YOUR symptom of low gluco
127
Know YOUR symptom of high gluco
128
Schedule mammogram
129
Schedule annual physical
130
Congratulations on following your diet
131
Well done, gluco is under control
132
Well done, BP is under control
133
Please use your oxygen
P4510EN.07
2/16/12
92
GENESIS™ DM MANUAL
Appendix A: Subjective Questions
134
Remember to maintain your oxygen equipment as instructed.
135
You are scheduled to have a video visit with your health care provider today.
136
You are scheduled to have a video visit with your health care provider tomorrow.
Note:
Questions 134 through 136 are available in English and Spanish. If you program your
monitor for another language, these questions will appear in English with no audio.
Non-Alerting Messages
Number
Message
300
Has your care provider talked with you about all the prescription and over-the-counter medications you were
taking?
301
Have you take any new prescription medications or changed your prescription medications?
302
Has your care provider explained the purpose of taking your new or changed medications?
303
Has your care provider explained what the side effects of your medications could be?
304
Has your care provider explained when you need to take your medications?
305
When you started getting home health care from this agency, did someone from the agency ask to see all the
prescriptions and over-the-counter medications you were taking?
306
Have you and a home health provider from this agency talk about pain?
307
When you first started getting home health care from this agency, did someone from the agency tell you what
care and services you would get?
308
When you first started getting home health care from this agency, did someone form the agency talk with you
about how to set up your home so you can move around safely?
Note:
Questions 300 through 308 are available in English and Spanish. If you program your
monitor for another language, these questions will appear in English with no audio.
P4510EN.07
2/16/12
93
GENESIS™ DM MANUAL
Appendix B: Sleep Apnea Questions
Multi-Level Questions
Level 1
Level 2
Level 2 questions appear if Level 1 questions receive a YES response. If the patient responds with
a YES the next Level 2 question appears, and so on, until a No answer is received or there are no
more Level 2 questions.
Are you having any problems with urination?
309
310
311
Are you having trouble controlling your blood sugar?
312
313
319
322
Note:
P4510EN.07
Did you skip your prescribed diabetes medication today?
Are you having any dizziness?
314
315
Are you experiencing any burning with urination or increased frequency?
Was your most recent blood sugar lower than usual?
Have you almost fallen in the last day?
316
Are you having any difficulty walking or standing?
317
Are you having any dizziness or light headedness today?
318
Did you seek medical treatment for your dizziness or light-headedness?
Has your doctor added, deleted, or changed any of your medication this week?
320
Do you feel worse since starting your new medication?
321
Have you used any over the counter medications in addition to those prescribed for you in the
past 24 hours?
Have you developed a cough?
323
Is your cough productive?
324
Is your sputum, mucus, or phlegm thicker today?
325
Is your sputum, mucus, or phlegm a different color today?
326
Do you feel that you may have a chest infection?
Multi-level questions (309 through 322) are available in English and Spanish. If you program
your monitor for another language, these questions will appear in English with no audio.
2/16/12
94
GENESIS™ DM MANUAL
Appendix A: Subjective Questions
Appendix B: Sleep Apnea Questions
Sleep apnea questions are a single block of twelve questions that are either on or off. When turned on,
the Genesis DM monitor will go through all of the questions. The questions are listed below.
1.
Do you snore nearly every day?
2.
Do you snore 3-4 times a week?
3.
Have you been told that your snoring is louder than talking?
4.
Can others hear your snoring in another room?
5.
Has your snoring ever bothered other people?
6.
Has anyone noticed that you quit breathing nearly every day during your sleep?
7.
Has anyone noticed that you quit breathing during sleep 3-4 times a week?
8.
Do you feel tired or fatigued today after your sleep?
9.
Do you feel tired, fatigued, or not up to par during your wake time nearly every day?
10.
Do you feel tired, fatigued, or not up to par during your wake time 3-4 times a week?
11.
Do you nod off or fall asleep while driving a car nearly every day?
12.
Do you nod off or fall asleep while driving a car 3-4 times a week?
P4510EN.07
2/16/12
95
GENESIS™ DM MANUAL
Appendix C: Prompt Text
Appendix C: Prompt Text
The following is a compilation of all Genesis DM monitor audio prompts. What the patient hears
depends on monitor programming.
Good Morning (Afternoon, Evening), it is now time to record your vital signs.
Please swipe your card.
Please verify that this is your card using the yes and no keys.
Please swipe your card again.
Please step on the scale.
Please step off the scale and sit down on a chair in front of the monitor.
Please sit down on a chair in front of the monitor.
Place the blood pressure cuff on your right arm above the elbow. Tighten it securely.
Place the finger sensor on the middle finger of your left hand with the finger design facing up.
Rest your arm as instructed by your clinician.
Press the green start key.
Please remove the blood pressure cuff.
Please remove the finger sensor and the blood pressure cuff.
It is now time to take your temperature.
Please take your temperature as instructed by your clinician.
Please answer the following questions using the yes and no keys.
Please attach the glucose meter to the monitor.
No glucose meter detected.
New glucose meter detected.
Please attach the prothrombin time meter to the monitor.
No prothrombin time meter detected.
Please activate your peak flow meter and place it in the cradle.
P4510EN.07
2/16/12
96
GENESIS™ DM MANUAL
Appendix C: Prompt Text
No peak flow detected.
It is now time to take your ECG reading.
Apply two to three drops of the electrode solution on both electrodes of the ECG device.
Press the start key on the ECG device.
Place your thumbs gently on the two electrodes to take an ECG reading.
Press the yes key when you are finished.
Please attach the ECG device to the monitor.
No ECG device detected.
Please make sure the cable is attached to the port marked COM (A, 1, 2) on the back of the monitor.
Receiving data, please wait.
Data upload is finished.
Please remember to take your medications as prescribed by your physician and maintain the proper diet.
Thank you for completing your vital signs measurement.
No vital signs, your clinic will be notified.
P4510EN.07
2/16/12
97
GENESIS™ DM MANUAL
Appendix E: Medical Device Specifications
Appendix D: Genesis DM System Supplies
Use only supplies authorized by Honeywell HomMed, as other equipment may result in increased emissions
or decreased immunity to emissions of the Genesis DM. To order supplies, contact Customer Service and
include the part number and name.
Honeywell HomMed
Suite 200
3400 W Intertech Drive
Brookfield, WI 53045
Toll-free: (888) 353-5404 (U.S. and Canada)
Phone: 262-252-5840 • Fax: 262-783-5441
P4510EN.07
2/16/12
98
GENESIS™ DM MANUAL
Appendix E: Medical Device Specifications
Appendix E: Medical Device Specifications
The following peripheral medical devices are compatible with the Genesis DM monitor. For instructions on
how to use any peripheral medical device, please refer to manufacturer’s product materials. Specifications
may be changed by individual manufacturers without notice from Honeywell HomMed.
ECG device
HealthFrontier
ecg@home (available in the U.S. and Canada only)
Glucose meters
Bayer
Ascensia Breeze®
Ascensia Breeze2®
Ascensia Contour®
Ascensia Elite XL®
Home Diagnostics, Inc.
Prestige IQ®
TrueTrack Smart System®
LifeScan
One Touch Basic®
One Touch Profile®
One Touch Ultra®
One Touch Ultra InDuo®
One Touch Ultra2®
Roche
ACCU-CHEK Advantage®
ACCU-CHEK Aviva®
ACCU-CHEK Compact™ Plus
Peak flow meters
Ferraris
Piko-1 Peak Flow Meter®
Prothrombin meters
HemoSense
InRatio
ITC
ProTime Microcoagulation System®
Exergen
Temporal Scanner
Temporal temperature
Specifications
Refer to manufacturer manual for specifications for any approved Honeywell HomMed medical devices.
Genesis DM monitor
Compliance approvals
UL60601, CSA601, IEC 60601-1
Physical
Dimensions
Width:
Height:
Depth:
Weight
7.0 inches (17.8 cm)
2.85 inches (7.25 cm)
7.25 inches (1842 cm)
Approximately 1.95 pounds (0.89 kg)
Power
AC Power
Wall mount style power supply
Input of 105-125VAC, 60 Hz
Input of 215-240VAC, 50/60 Hz
Output of 24VDC @ 500mA with 4kV isolation
Battery
LI+ (lithium-Ion), 3.0DC non-rechargeable for memory backup and real-time clock power; not a primary power
source when AC power is not present
P4510EN.07
2/16/12
99
GENESIS™ DM MANUAL
Appendix E: Medical Device Specifications
Environment
The monitor may not meet its performance specifications if stored or used outside the temperature and humidity ranges listed below.
Temperature
Operation:
Storage:
0°C - 50C (32°F – 122F)
-40°C - 75C (-40°F - 167F)
Relative humidity
Operation:
Storage:
15 - 95% (non-condensing)
10 – 95% (non-condensing)
EMC
EN 60601-1-2 with scale cables lot code 1103-01 or higher
Languages
NOTE: English is generally the default language, although other languages are available. Contact Customer Service with any questions.
Clinician:
US English, Dutch, Canadian French
Patient:
Menu selectable; standard is US English Audio (Male and Female voice) and Text
UK English
Portuguese
Italian
German
French
Canadian French
Hindi
Welsh
Spanish
Russian
Armenian
Polish
Urdu
Bengali
Punjabi
Dutch
Modem Type:
33.6K baud
Compliance approvals:
FCC Part 68
FCC Part 15B
Connection:
RJ-11
Non-Invasive Blood Pressure
Method of measurement:
Oscillometric
Blood pressure range:
20 to 260 mmHg (40 – 260 mmHg systolic, 20 – 200 mmHg diastolic)
Pressure Transducer Accuracy; +/- 3 mmHg between 0 mmHg and 300 mmHg, for operating conditions between -20 C and 60 C and less than
95% relative humidity.
Patient safety:
Maximum cuff inflation time is limited to 50 seconds. Duration of blood pressure reading is limited to 130 seconds.
Redundant safety circuitry aborts reading if cuff pressure exceeds 300 mmHg or if cuff has been inflated for 180
seconds.
Calibration:
Recommended yearly
Safety standards:
Meets all relevant parts of AAMI SP-10, IEC-60601-1, IEC-60601-2-30, EN1060-1, EN1060-3
Initial inflation pressure:
120-250 mmHg adult mode
Default inflation pressure:
175 mmHg (120-280 mmHg)
Oximeter module
Range:
0 - 99% Functional SpO2 (1% increments)
Accuracy:
±2 at 70 - 99% less than 70% is unspecified
Alarms:
None
Averaging:
8 pulse beat average
P4510EN.07
(continued)
2/16/12
100
GENESIS™ DM MANUAL
Appendix E: Medical Device Specifications
Calibration:
Factory calibrated over the range of 70% to 99% SpO2 using human blood samples to functional saturation. Test
methods available upon request. No in-service calibration is required.
Sensor:
Red - 660nm, 2mW (typical)
Infrared 905nm, 2-2.4mW (typical)
Display Update Rate:
1Hz (Maximum age of SpO2 data is 20 sec.)
Heart rate (oximeter)
Range:
30-254 BPM (1 BPM increments)
Accuracy:
±2% or 2 BPM, which ever is greater
Alarms:
None. Not for continuous monitoring
Averaging:
8 second average
Display Update Rate:
1 Hz (Maximum age of SpO2 data is 20 sec.)
Heart rate (NIBP)
Range:
40 – 200 BPM (adult mode)
Accuracy:
+ 5 BPM or 5%, whichever is greater
Pulse Strength (requires oximeter)
Range:
0-6 segment bar graph, indicates logarithmic strength of patient’s pulse
Display Update Rate:
60Hz
Scale
HONEYWELL
Manufacturer:
Regular and Bluetooth Scales
Honeywell HomMed
Range:
0 to 500 lbs. (0 to 227 kg)
Accuracy:
+/- 0.5 lbs. (+/- 0.2 kg) or +/- 0.5%, whichever is greater
Update Rate:
< 0.5 seconds
Stability of Calibration:
Calibration to zero (“zeroing”) is performed at power-up. Multipoint span factory calibration is preserved through
power cycles. Factory calibration is good for 5 years of normal use.
FAIRBANKS (used with product bearing CE Marking)
Manufacturer:
Fairbanks
Range:
0 to 500 lbs. (0 to 227 kg)
Accuracy:
+ 0.5 lbs. (+ 0.2 kg) or + 0.5%, whichever is greater
Update Rate:
< 0.5 seconds
Stability of Calibration:
Zero calibration is performed at power-up. Multipoint span factory calibration is preserved through power cycles.
Factory calibration is good for 5 years of normal use.
Gravity Adjustment:
The scale is adjusted for gravity in the location to which HomMed is shipping it. Any relocation to a different gravity
zone requires re-verification by HomMed or a European local authority. Any damage or recalibration may also
require re-verification.
P4510EN.07
2/16/12
101
GENESIS™ DM MANUAL
Appendix F: Electromagnetic Guidance
Appendix F: Electromagnetic Guidance
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The Genesis DM is intended for use in the electromagnetic environment specified below. The customer
or the user of the Genesis DM should assure that it is used in such an environment.
Emissions Test
Compliance
RF Emissions
CISPR 11
Group 1
RF Emissions
CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage Fluctuations/ Flicker Emissions
IEC 61000-3-3
Electromagnetic Environment – Guidance
The Genesis DM uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
Class B
The Genesis DM is suitable for use in all establishments, including
domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings
used for domestic purposes.
Class A
Complies
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The Genesis DM is intended for use in the electromagnetic environment specified below. The customer
or the user of the Genesis DM should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electrostatic Discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Electrical Fast Transient/Burst
IEC 61000-4-4
±2 kV
for power supply lines
±1 kV
for input/output lines
±2 kV for power supply
lines
±1 kV for input/output lines
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Voltage Dips, Short Interruptions and
Voltage Variations on Power Supply Input
Lines
IEC 61000-4-11
<5% UT (>95% dip in UT)
for 0.5 cycle
40% UT (60% dip in UT)
for 5 cycles
70% UT (30% dip in UT)
for 25 cycles
<5% UT (>95% dip in UT)
for 5 sec
0% UT (95% dip in UT)
for 0.5 cycle
40% UT (60% dip in UT)
for 5 cycles
70% UT (30% dip in UT)
for 25 cycles
<5% UT (>95% dip in UT)
for 5 sec
Power Frequency (50/60 Hz) Magnetic
Field
IEC 61000-4-8
3 A/m
3 A/m
Electromagnetic Environment
– Guidance
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
Genesis DM requires continued
operation during power mains
interruptions, it is recommended
that the Genesis DM be powered
from an uninterruptible power
supply or a battery.
Power frequency magnetic fields
should be at levels characteristic of
a typical commercial or hospital
environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.
P4510EN.07
2/16/12
102
GENESIS™ DM MANUAL
Appendix F: Electromagnetic Guidance
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The Genesis DM is intended for use in the electromagnetic environment specified below. The customer
or the user of the Genesis DM should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance Level
Conducted RF
3 Vrms
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
150 kHz to 80 MHz
Radiated RF
3 V/m
3 V/m
Portable and mobile RF communications
equipment should be used no closer to any
part of the Genesis DM, including cables, than
the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
IEC 61000-4-3
80 MHz to 2.5 GHz
80 MHz to 2.5 GHz
Recommended Separation Distance
Electromagnetic Environment –
Guidance
d = 1.17√P
d = 1.17√P
80
MHz to 800 MHz
d = 2.33√P
800
MHz to 2.5 GHz
where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in meters
(m).
Field strengths from fixed RF transmitters, as
determine by an electromagnetic site survey a,
should be less than the compliance level in
each frequency rangeb.
Interference may occur in the vicinity of
equipment marked with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Field
strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, and electromagnetic site survey should be considered. If the
measured field strength in the location in which the Genesis DM is used exceeds the applicable RF compliance level above,
the Genesis DM should be observed to verify normal operation. If abnormal performance is observed, additional measures
may be necessary, such as reorienting or relocating the Genesis DM.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
P4510EN.07
2/16/12
103
GENESIS™ DM MANUAL
Appendix F: Electromagnetic Guidance
Recommended Separation Distances Between Portable and Mobile RF Communications
Equipment and the Genesis DM
The Genesis DM is intended for use in an electromagnetic environment where radiated radio frequency
(RF) signals are controlled. The customer or the user of the Genesis DM can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Genesis DM as recommended below, according to
the maximum output power of the communications equipment.
Separation Distance According to Frequency of Transmitter
M
800 MHz to 2.5 GHz
150 kHz to 80 MHz
80 MHz to 800 MHz
W
d = 1.17√P
d = 1.17√P
d = 2.33√P
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.70
3.70
7.37
100
11.70
11.70
13.30
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated
using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
Rated Maximum Output Power
of Transmitter
USA-Federal Communications Commission (FCC)
This equipment has been tested and found to comply with the limits for a Class B digital device,
pursuant to Part 15 of FCC Rules. These limits are designed to provide reasonable protection against
harmful interference in a residential installation.
This equipment generates, uses, and can radiate radio frequency energy. If not installed and used in
accordance with the instructions, it may cause harmful interference to radio communications. However,
there is no guarantee that interference will not occur in a particular installation.
If this equipment does cause harmful interference to radio or television reception, which can be
determined by tuning the equipment off and on, the user is encouraged to try and correct the
interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the distance between the equipment and the receiver.
• Connect the equipment to outlet on a circuit different from that to which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
Any changes or modifications not expressly approved by the party responsible for compliance could
void the user’s authority to operate the equipment.
Caution: Exposure to Radio Frequency Radiation.
This device must not be co-located or operating in conjunction with any other antenna or transmitter.
P4510EN.07
2/16/12
104
GENESIS™ DM MANUAL
Appendix F: Electromagnetic Guidance
Canada – Industry Canada (IC)
This device complies with RSS 210 of Industry Canada.
Operation is subject to the following two conditions:
(1) this device may not cause interference,
and
(2) this device must accept any interference, including interference that may cause undesired operation
of this device.
Caution: Exposure to Radio Frequency Radiation.
The installer of this radio equipment must ensure that the antenna is located or pointed such that it
does not emit RF field in excess of Health Canada limits for the general population; consult Safety Code
6, obtainable from Health Canada’s website www.hc-sc.gc.ca/rpb.
P4510EN.07
2/16/12
105