SOLAμ for Twenty-fold Sample Concentration Prior to Analysis

SOLAμ for Twenty-fold Sample
Concentration Prior to Analysis
Jon Bardsley and Ken Meadows
Thermo Fisher Scientific, Runcorn, UK
Introduction
Despite advances in analytical detection technology achieving required limits of
sensitivity, whilst maintaining reproducibility, can still be an issue for many bioanalytical
laboratories.
Results
The assay gave a lin
0.998. The chromato
above the acceptabl
Traditional scale SPE helps to clean up the sample to minimize matrix effects, however in
order to pre-concentrate the sample a lengthy dry down and reconstitution step needs to
be employed. This process is not only time consuming but can have a detrimental effect
on the recovery of the analyte due to volatility or non-specific binding.
FIGURE 2. Linear
2.6
2.5
By using Thermo Scientific™ SOLAµ™ a twenty-fold concentration of sample can be
achieved without the need for a lengthy dry-down step, increasing sensitivity without
compromising assay performance.
2.4
2.3
2.2
2.1
2.0
1.9
FIGURE 1. Summary of workflow required to achieve a twenty-fold increase in
sample concentration prior to analysis without altering the work-flow or
compromising assay performance.
1.8
1.7
1.6
Area Ratio
1.5
1.4
1.3
1.2
1.1
1.0
0.9
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0.0
0
5000
Low, mid and high
30000 pg/mL respe
Assay is reproducib
TABLE 1. Assay
2 SOLAμ for Twenty-fold Sample Concentration Prior to Analysis
By loading 500 µL
twenty-fold conce
even with low elut
cleanliness were a
By loading 500 µL o
twenty-fold concent
even with low elutio
cleanliness were ac
FIGURE 3. Twentysample concentrat
SOLAµ allows use
allowing greater li
analyte recovery,
Thermo Scientific Poster Note • PN20950_HPLC_2014_E_05/14S 3
TABLE 2. Matr
against pure r
Results
red limits of
many bioanalytical
The assay gave a linear dynamic range from 40 to 40000 pg/mL with an r2 coefficient of
0.998. The chromatography for the limit of quantitation sample at 40 pg/mL is significantly
above the acceptable signal to noise limit.
ix effects, however in
itution step needs to
a detrimental effect
g.
FIGURE 2. Linear calibration line for Niflumic acid from 40 to 40000 pg/mL
Niflumic Acid
Y = 0.00375684+5.93882e-005*X R^2 = 0.9975 W: 1/X^2
2.6
2.5
of sample can be
ensitivity without
2.4
Analyte recov
fortified blank
against pure r
less than 9%
2.3
2.2
2.1
2.0
1.9
old increase in
k-flow or
1.8
1.7
TABLE 3. Ass
against fortifi
1.6
Area Ratio
1.5
1.4
1.3
1.2
1.1
1.0
0.9
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0.0
0
5000
10000
15000
20000
25000
30000
35000
40000
Low, mid and high QC samples were prepared at concentrations of 400, 20000 and
30000 pg/mL respectively. Good levels of accuracy at all QC levels were observed.
Assay is reproducible for replicate extractions (n=18) at both high and low QC levels.
TABLE 1. Assay accuracy and precision data
Discussi
This poster de
analysis while
cleanliness.
By loading a s
decrease the
evaporation p
SOLAμ provid
utilizing the re
technology. T
ensures highl
SOLAμ deliv
By loading 500 µL of sample onto the SOLAµ plate and eluting in a total of 25 µL a
twenty-fold concentration of the analyte was achieved. The results demonstrate that
Concentration
Prior to Analysis
4 SOLAμ for Twenty-fold Sample
even
with low elution
volume high levels of accuracy, precision, recovery and sample
cleanliness were achieved.
•
Lower sam
•
Increased
•
The ability
•
Improved
By loading 500 µL of sample onto the SOLAµ plate and eluting in a total of 25 µL a
twenty-fold concentration of the analyte was achieved. The results demonstrate that
even with low elution volume high levels of accuracy, precision, recovery and sample
cleanliness were achieved.
FIGURE 3. Twenty-fold increase in response achieved by using SOLAµ plate for
sample concentration
© 2014 Thermo Fi
Scientific and its s
This information is
intellectual propert
SOLAµ allows users to pre-concentrate the sample up to 20-times prior to injection,
allowing greater limits of sensitivity to be achieved whilst maintaining a high level of
analyte recovery, accuracy and precision.
Thermo Scientific Poster Note • PN20950_HPLC_2014_E_05/14S 5
TABLE 2. Matrix effects results calculated by comparing fortified blank samples
against pure reference standards
coefficient of
L is significantly
2
0 pg/mL
Analyte recovery was shown to be greater than 89.9% by comparison to post extraction
fortified blank samples. Post extraction fortified blank samples were also compared
against pure reference standards to demonstrate matrix effects which were calculated at
less than 9% at both high and Low QC levels.
TABLE 3. Assay recovery results calculated by comparing extracted samples
against fortified blank samples
40000
20000 and
observed.
w QC levels.
Discussion
This poster demonstrates the advantages of SOLAµ for sample concentration prior to
analysis while maintaining high levels of precision, accuracy, recovery and sample
cleanliness.
By loading a sample volume of 500 µL and eluting in a volume of 25 µL it is possible to
decrease the lower limit of quantitation by a factor of twenty without the need for lengthy
evaporation procedures that may compromise sample integrity.
SOLAμ provides reproducibility, robustness and ease of use at low elution volumes by
utilizing the revolutionary Thermo Scientific™ SOLA™ , Solid Phase Extraction (SPE)
technology. This removes the need for frits delivering a robust, reproducible format which
ensures highly consistent results at low elution volumes.
SOLAμ delivers:
•
Lower sample failures due to high reproducibility at low elution volumes
•
Increased sensitivity due to lower elution volumes
•
The ability to process samples which are limited in volume
•
Improved stability of bio-molecules by reduction of adsorption and solvation issues
l of 25 µL a
onstrate 6that
SOLAμ for Twenty-fold Sample Concentration Prior to Analysis
ry and sample
technology. This removes the need for frits delivering a robust, reproducible format which
ensures highly consistent results at low elution volumes.
SOLAμ delivers:
•
Lower sample failures due to high reproducibility at low elution volumes
•
Increased sensitivity due to lower elution volumes
•
The ability to process samples which are limited in volume
•
Improved stability of bio-molecules by reduction of adsorption and solvation issues
of 25 µL a
nstrate that
and sample
Aµ plate for
© 2014 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher
Scientific and its subsidiaries.
This information is not intended to encourage use of these products in any manners that might infringe the
intellectual property rights of others.
to injection,
high level of
www.thermoscientific.com
PO20950_E 04/14S
©2014 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization.
All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as
an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products
in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to
change. Not all products are available in all countries. Please consult your local sales representative for details.
Africa +43 1 333 50 34 0
Australia +61 3 9757 4300
Austria +43 810 282 206
Belgium +32 53 73 42 41
Brazil +55 11 3731 5140
Canada +1 800 530 8447
China 800 810 5118 (free call domestic)
400 650 5118
Denmark +45 70 23 62 60
Europe-Other +43 1 333 50 34 0
Finland +358 9 3291 0200
France +33 1 60 92 48 00
Germany +49 6103 408 1014
India +91 22 6742 9494
Italy +39 02 950 591
Japan +81 6 6885 1213
Korea +82 2 3420 8600
Latin America +1 561 688 8700
Middle East +43 1 333 50 34 0
Netherlands +31 76 579 55 55
New Zealand +64 9 980 6700
Norway +46 8 556 468 00
Thermo Fisher Scientific, Sunnyvale, CA
USA is ISO 9001:2008 Certified.
Russia/CIS +43 1 333 50 34 0
Singapore +65 6289 1190
Sweden +46 8 556 468 00
Switzerland +41 61 716 77 00
Taiwan +886 2 8751 6655
UK/Ireland +44 1442 233555
USA +1 800 532 4752
PN20950_E 05/14S