D02418_04_insert_CRP HS_MU 981798 1 x 4 ml THIS PACKAGE INSERT IS APPLICABLE FOR USE OUTSIDE THE US. ANY REFERENCE TO THE KONELAB SYSTEMS ALSO REFERS TO THE T SERIES. INTENDED USE For in vitro quantitative determination of low concentrations of CRP in human serum and plasma on Konelab analyzers. All test results must be interpreted with regard to the clinical context. SUMMARY (1, 5) C-reactive protein (CRP) is a normal constituent of serum present in healthy individuals in very low concentrations. Invasive bacterial infection and extensive tissue damage cause increased CRP levels. CRP is one of the most sensitive acute phase proteins; levels in plasma usually rise dramatically after myocardial infarction, trauma, infection, inflammation, surgery, or neoplastic proliferation. The increase begins within 6 to 12 hours, and levels may reach 2000 times normal. Determination of CRP is clinically useful for screening for organic disease, assessment of the activity of inflammatory disease, detection of intercurrent infections in systemic lupus erythematosus (SLE), in leukemia, or after surgery and management of neonatal septicemia and meningitis, when specimen collections for bacteriological investigations may be difficult. Measurement of CRP by sensitive CRP assays may add to the predicative value of other markers used to assess the risk of cardiovascular and peripheral vascular diseases. PRINCIPLE OF THE PROCEDURE The method is based on measurement of immunoprecipitation at 540 nm. Microparticles coated with anti-human CRP are added to buffered samples. The increase in absorbance caused by immunoprecipitation is recorded when the reaction has reached its end-point. The change in absorbance is proportional to the amount of antigen (CRP) in solution. REAGENT INFORMATION CRP HS Reagent CRP HS Buffer Specimen Diluent CRP HS Calibrator 1 x 4 ml 2 x 50 ml 1 x 60 ml 2 x 1 ml Concentrations CRP HS Reagent Microparticles coated with anti-human CRP (sheep) (lyophilized) NaN3 1 - < 2.5% CRP HS Buffer Tris 0.2 mol/l NaN3 < 0.1 % Specimen Diluent PBS 0.01 mol/l NaN3 < 0.1 % CRP HS Calibrator Human NaN3 < 0.1 % The CRP concentration of the calibrator is printed on a separate Value sheet. Precautions For in vitro diagnostic use only. Exercise the normal precautions required for handling all laboratory reagents. Calibrator contains material of human origin, which has been tested and found negative for HBsAg and for antibodies to HIV 1/2- and HCV. However, the product must be handled just as carefully as the patient specimens. The lyophilized CRP HS Reagent contains sodium azide as preservative (1 - < 2.5%). Avoid inhaling dust. Do not swallow. Avoid contact with skin and mucous membranes. If contact occurs, wash with copious amounts of water. Sodium azide may react with lead and copper in drainage systems to form highly explosive metal azides. On disposal flush with large volumes of water. See separate sheet for risk and safety phrases (T:R;23/24/25 - 52/53 S;22-28-45). After reconstitution, the liquid CRP HS Reagent contains sodium azide < 0.1% which is not considered as a harmful concentration. Preparation CRP HS Buffer, Specimen diluent and CRP HS Calibrator are ready for use. Reconstitution of CRP HS Reagent: Add exactly 4 ml of deionised water. Close the vial, mix gently and allow to stand for at least 30 minutes at room temperature. Mix the contents once again before use. Do not shake. Transfer the reagent into an empty bar-coded vial for use in Konelab analyzers. Note 1: Always use CRP HS Buffer, CRP HS Reagent and CRP HS Calibrator of the same reagent lot number because the components within lots have been matched. Note 2: Mix CRP HS Reagent gently before you insert the reagent in the Konelab analyser. Do not shake. Note 3: Check that there are no bubbles in the bottleneck or on the surface of the reagent when you insert the reagent vials or vessels in the Konelab analyzer. Storage and Stability Reagents in unopened vials are stable at 2…8 °C until the expiry date printed on the label. Refer to the Application Notes of your Konelab analyzer for the on board stability of reagents. The reconstituted CRP HS Reagent and opened CRP HS Buffer may be used for 8 weeks if stored closed at 2…8 °C when not in use. It is recommended that the reagent vials be taken out of the analyzer and kept closed in the refrigerator when not in use. Once opened, the CRP HS Calibrator is stable for 8 weeks at 2…8 °C. Store the calibrator tightly capped when not in use. Do not use any vial that has visible evidence of microbial growth. Konelab analyzer. Any application which has not been validated by Thermo Fisher Scientific Oy cannot be performance guaranteed and therefore must be evaluated by the user. Materials provided Reagents, as described above, and an empty bar-coded vial for use in Konelab analysers. Materials required but not provided Control as indicated below. Calibration Use CRP HS Calibrator included in the kit according to the instructions provided for your Konelab analyzer. Allow the calibrator to remain on the analyzer for a maximum of one hour. Traceability: The value of CRP has been assigned by using IFCC preparation CRM 470 as a primary reference. Quality Control Use quality control sample at least once a day and after each calibration and every time a new bottle of reagent is used. Available controls: CRP High Sensitivity Control, code 981852 The control intervals and limits must be adapted to the individual laboratory requirements. The result of the quality control sample should fall within the limits pre-set by the laboratory. CALCULATION OF RESULTS The results are calculated automatically by the Konelab analyzer using a calibration curve. The analyzer automatically prepares a dilution series from the stock calibrator according to the pre-set parameters. The calibration curve is generated from the measured calibrators using the spline fit. Calibration curve (example) CRP HS 1,6 1,4 1,2 1 0,8 0,6 0,4 0 5 10 15 20 Concentration (m g/l) Konelab 20/30/60. The calibration curve is lot dependent. LIMITATIONS OF THE PROCEDURE (6) Interference Criterion: Recovery within ± 10% of initial values. Bilirubin: No interference found up to 1000 μmol/l (58.5 mg/dl). Hemolysis: No interference found up to 10 g/l of hemoglobin in hemolysate. Lipemia: No interference found up to 4 g/l of Intralipid® (trademark of Fresenius Kabi AB). There is a poor correlation between turbidity and concentration of triglyserides. For other interfering substances, please refer to the reference 3. Heterophilic antibodies, monoclonal immunoglobulins (or parts of them) and rheumatoid factors may interfere with the assay. Results from patients suspected of having such antibodies should be carefully evaluated (4). Antigen excess No antigen excess was found up to a CRP concentration of 300 mg/l. EXPECTED VALUES (5) Below 3 mg/l The quoted values should serve as a guide only. It is recommended that each laboratory verify this range or derive a reference interval for the population that it serves. When using CRP to assess the risk of coronary heart disease, measurements should be compared to previous values. MEASURING RANGE 0.25 – 10 mg/l Extended measuring range after secondary dilution: 0.25 – 40 mg/l PERFORMANCE CHARACTERISTICS (6) Determination limit The determination limit is the lowest concentration that can be measured quantitatively: 0.25 mg/l SPECIMEN COLLECTION Sample type Serum, heparin plasma and EDTA plasma can be used. Any turbid sample should be centrifuged (e.g. 2000xg for 15 min) Precautions Human samples should be handled and disposed of as if they were potentially infectious. Storage (2) The sample can be stored for 8 days at 2…8 °C or for 3 months at – 20 °C (if frozen only once). TEST PROCEDURE Refer to the Reference Manual and Application Notes for an automated procedure on your Imprecision (result unit mg/l) Mean 0.80 mg/l Mean 2.41 mg/l SD CV% SD CV% Within run 0.02 3.2 0.02 0.8 Between day 0.03 3.7 0.04 1.8 Total 0.04 5.9 0.05 2.3 Mean 7.02 mg/l SD CV% Within run 0.04 0.6 Between day 0.08 1.2 Total 0.15 2.2 A precision study was performed according to guidelines in the NCCLS Document EP5-A, using Konelab 30 for at least 20 days, with the number of measurements being at least n = 80. © 2007 Thermo Fisher Scientific Inc. All rights reserved. Konelab™ / T Series CRP High Sensitivity Template: D01162_4 EN Response (A) Page 1 D02418_04_insert_CRP HS_MU BIBLIOGRAPHY 1) Burtis CA and Ashwood ER (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, WB Saunders Company, Philadelphia, 2001, pp. 332-333 and 688. 2) Guder WG, Narayanan S, Wisser H, Zawta B. List of Analytes; Preanalytical variables. Brochure in: Samples: From Patient to the Laboratory. GIT Verlag GmbH, Darmstadt, 1996. 3) Young DS, Effects of Drugs on Clinical Laboratory Tests, Volume One, Listing by Test, Fifth Edition, AACC Press, Washington, D.C., 3-233 – 3-234. 4) Selby C. Interference in immunoassay. Ann Clin Biochem 1999;36:704-721. 5) Pearson TA et al. Markers of Inflammation and Cardiovascular Disease Application to Clinical and Public Health Practice, A Statement for Healthcare Professionals From the Centers for Disease Control and the American Heart Association., Circulation 2003;107:499-511. 6) Data on file at Thermo Fisher Scientific Oy. MANUFACTURER Thermo Fisher Scientific Oy Clinical Diagnostics Finland Ratastie 2, P.O. Box 100, FI-01621 Vantaa Tel. +358 9 329 100, Fax +358 9 3291 0300 www.thermo.com/konelab Date of revision (yyyy-mm-dd) 2007-09-13 Changes from previous version Company name updated. DE Konelab™ / T Serie CRP High Sensitivity CRP Hohe Sensitivität 981798 1 x 4 ml DIESE PACKUNGSBEILAGE IST FÜR DEN GEBRAUCH AUSSERHALB DER USA VORGESEHEN. JEDER VERWEIS AUF KONELABSYSTEME BEINHALTET AUCH DIE T SERIES. ANWENDUNGSBEREICH Zur quantitativen In-vitro-Bestimmung geringer Konzentrationen von CRP in Humanserum und -plasma mit einem Konelab-Analysengerät. Alle Testergebnisse müssen mit Bezug zum klinischen Zusammenhang interpretiert werden. ZUSAMMENFASSUNG (1, 5) C-reaktives Protein (CRP) ist ein normaler Bestandteil des Serums, der bei gesunden Individuen in sehr niedrigen Konzentrationen vorliegt. Invasive bakterielle Infektionen und starke Gewebeschäden führen zu erhöhten CRP-Konzentrationen. CRP ist eines der empfindlichsten Akute-Phase-Proteine; die Konzentration im Plasma steigt in der Regel nach Myokardinfarkten, Traumata, Infektionen, Entzündungen, Operationen oder neoplastischer Proliferation stark an. Der Anstieg beginnt innerhalb von 6 bis 12 Stunden. Die Konzentration kann das 2000 fache der normalen Werte erreichen. Die Bestimmung von CRP ist klinisch sinnvoll für das Screening auf Organerkrankungen, für die Beurteilung der Aktivität von entzündlichen Erkrankungen, für die Erkennung von interkurrenten Infektionen bei systemischem Lupus erythematodes (SLE), bei Leukämie, nach Operationen und für das Management von Septikämie und Meningitis bei Neugeborenen, wenn sich die Probennahme für bakteriologische Untersuchungen schwierig gestaltet. Die Bestimmung von CRP durch sensitive CRP-Assays kann den prädikativen Werte anderer Marker erhöhen, die zur Bewertung des Risikos kardiovaskulärer und peripherer Gefäßkrankheiten herangezogen werden. TESTPRINZIP Das Verfahren beruht auf der Messung der Immunpräzipitation bei 540 nm. Mit antihumanem CRP beschichtete Mikroteilchen werden gepufferten Proben hinzugegeben. Die durch die Immunpräzipitation verursachte Zunahme der Extinktion wird aufgezeichnet, wenn die Reaktion den Endpunkt erreicht hat. Die Extinktionsänderung ist proportional zur Antigenmenge (CRP) in Lösung. REAGENZDATEN CRP HS-Reagenz 1 x 4 ml CRP HS-Puffer 2 x 50 ml Probenverdünnung 1 x 60 ml CRP HS-Kalibrator 2 x 1 ml Konzentrationen CRP HS-Reagenz Mit antihumanem CRP (Schaf, lyophilisiert) beschichtete Mikroteilchen NaN3 1 - < 2.5% CRP HS-Puffer Tris 0.2 mol/l NaN3 < 0.1 % Probenverdünnung PBS 0.01 mol/l NaN3 < 0.1 % CRP HS-Kalibrator Human NaN3 < 0.1 % Die CRP-Konzentration des Kalibrators ist auf einem separaten Blatt mit den Werten angegeben. Sicherheitsmaßnahmen Nur zur In-vitro-Diagnostik. Die üblichen Sicherheitsmaßnahmen beim Umgang mit Laborreagenzien befolgen. Der Kalibrator enthält Material humanen Ursprungs, das erfolgreich auf Negativität gegenüber HBsAg und den Antikörpern gegen HIV 1/2 und HCV getestet wurde. Dennoch muss das Produkt mit der gleichen Sorgfalt wie die Patientenproben behandelt werden. Das lyophilisierte CRP HS-Reagenz enthält Natriumazid (1 - < 2.5 %) als Konservierungsmittel. Staub nicht inhalieren. Nicht verschlucken. Berührung mit Haut und Schleimhäuten vermeiden. Ist es zum Kontakt gekommen, mit reichlich Wasser waschen. Das Natriumazid kann mit Blei und Kupfer in Abwassersystemen reagieren und dabei hochexplosive Metallazide bilden. Bei der Entsorgung mit großen Mengen Wasser spülen. Auf einem separaten Blatt sind die entsprechenden Hinweise auf die besonderen Gefahren (R-Sätze) und Sicherheitsratschläge (S-Sätze) enthalten (T:R;23/24/25 - 52/53 S;22-28-45). Nach der Wiederherstellung enthält das flüssige CRP HS-Reagenz < 0.1 % Natriumazid, eine als unbedenklich eingestufte Konzentration. Vorbereitung CRP HS-Puffer, Probenverdünnung und CRP HS-Kalibrator sind gebrauchsfertig. Wiederherstellung des CRP HS-Reagenzes: Genau 4 ml deionisiertes Wasser hinzugeben. Die Phiole schließen, vorsichtig mischen und mindestens 30 Minuten lang bei Zimmertemperatur stehen lassen. Den Inhalt vor dem Gebrauch erneut mischen. Nicht schütteln. Das Reagenz zur Verwendung im Konelab-Analysengerät in eine leere Phiole mit Barcode überführen. Hinweis 1: CRP HS-Puffer, CRP HS-Reagenz und CRP HS-Kalibrator stets von der gleichen Reagenz-Chargennummer verwenden, da die Komponenten einer Charge aufeinander abgestimmt wurden. Hinweis 2: Das CRP HS-Reagenz vor dem Einlegen in das Konelab-Analysengerät vorsichtig mischen. Nicht schütteln. Hinweis 3: Darauf achten, dass sich im Flaschenhals oder an der Reagenzoberfläche keine Luftblasen befinden, wenn die Phiolen bzw. Glasfläschchen mit dem Reagenz in das Konelab-Analysengerät eingelegt werden. Lagerung und Haltbarkeit In ungeöffneten Phiolen sind die Reagenzien bei 2...8 °C bis zum Ablauf des Verfallsdatums auf dem Etikett haltbar. Die Haltbarkeit der Reagenzien nach dem Einlegen in das Konelab-Analysengerät ist in den entsprechenden Anwendungshinweisen enthalten. Das wiederhergestellte CRP HS-Reagenz und der geöffnete CRP HS-Puffer können für bis zu 8 Wochen verwendet werden, wenn sie bei Nichtgebrauch im geschlossenen Zustand bei 2...8 °C gelagert werden. Die Reagenzphiolen sollten bei Nichtgebrauch aus dem Analysengerät entnommen und im geschlossenen Zustand im Kühlschrank aufbewahrt werden. Nach dem Öffnen ist der CRP HS-Kalibrator bei 2...8 °C für 8 Wochen haltbar. Den Kalibrator bei Nichtgebrauch dicht verschlossen lagern. Es dürfen keine Phiolen mit sichtbaren Anzeichen von Mikrobenwachstum verwendet werden. UNTERSUCHUNGSMATERIAL Probenart Es können Serum, Heparinplasma oder EDTA-Plasma verwendet werden. Alle trüben Proben sollten zentrifugiert werden (z. B. 2000xg für 15 min). Sicherheitsmaßnahmen Proben humanen Ursprungs sind als potenziell dementsprechend zu behandeln und zu entsorgen. infektiös zu betrachten und Lagerung (2) Die Probe kann 8 Tage lang bei 2...8 °C oder 3 Monate lang bei –20 °C (bei einmaligem Einfrieren) gelagert werden. TESTDURCHFÜHRUNG Angaben zur Automatisierung mit dem Konelab-Analysengerät dem Referenzhandbuch und den Anwendungshinweisen entnehmen. Bei Verwendung von Applikationen, die nicht durch Thermo Fisher Scientific Oy validiert wurden, kann keine Garantie für die angegebenen Leistungsdaten übernommen werden. Für die Validierung derartiger Applikationen ist der Anwender daher selbst verantwortlich. Lieferumfang Die oben beschriebenen Reagenzien und eine leere Phiole mit Barcode zur Verwendung im Konelab-Analysengerät. Erforderliche, jedoch nicht im Lieferumfang enthaltene Materialien Kontrolle, wie nachstehend beschrieben. Kalibrierung Den im Kit enthaltenen CRP HS-Kalibrator gemäß den Anweisungen zum KonelabAnalysengerät verwenden. Den Kalibrator für höchstens eine Stunde im Analysengerät lassen. Rückverfolgbarkeit: Der CRP-Wert wurde mit der IFCC-Präparation CRM 470 als Primärreferenz zugewiesen. Qualitätskontrolle Mindestens einmal täglich sowie nach jeder Kalibrierung und bei jeder Verwendung eines neuen Reagenzfläschchens eine Qualitätskontrolle durchführen. Lieferbare Kontrollen: CRP High Sensitivity Control (CRP HS-Kontrolle), Bestellnr. 981852 Die Intervalle und Grenzen der Kontrolle müssen an die Anforderungen der einzelnen Laboratorien angepasst werden. Die Ergebnisse der Qualitätskontrolle sollten innerhalb der vom Labor vorgegebenen Grenzwerte liegen. BERECHNUNG DER ERGEBNISSE Die Ergebnisse werden vom Konelab-Analysengerät mithilfe einer Bezugskurve automatisch berechnet. Das Analysengerät setzt von den Kalibrator-Stammlösungen gemäß den voreingestellten Parametern automatisch eine Verdünnungsserie an. Die Bezugskurve wird von den gemessenen Kalibratoren durch eine Polynomanpassung erzeugt. © 2007 Thermo Fisher Scientific Inc. All rights reserved. Method comparison A comparison study was performed according to guidelines in the NCCLS Document EP9A using a commercially available method as a reference. Linear regression (result unit mg/l): y = 1.01 x – 0.29 r = 0.983 n = 87 The sample concentrations were between 0.29 and 10 mg/l. The results obtained in individual laboratories may differ from the performance data given. Template: D01162_4 Page 2 Page 3 D02418_04_insert_CRP HS_MU HERSTELLER Thermo Fisher Scientific Oy Clinical Diagnostics Finland Ratastie 2, P.O. Box 100, FI-01621 Vantaa Tel. +358 9 329 100, Fax +358 9 3291 0300 www.thermo.com/konelab Bezugskurve (Beispiel) CRP HS Datum der Überarbeitung (JJJJ-MM-TT) 2007-09-13 Extinktion (A) 1,6 1,4 Änderungen gegenüber der vorherigen Fassung Name des Unternehmens aktualisiert. 1,2 1 0,8 0,6 0,4 FR 0 5 10 15 20 Konzentration (m g/l) Konelab™ / Gamme T CRP High Sensitivity CRP à haute sensibilité Antigenüberschuss Bis zu einer CRP-Konzentration von 300 mg/l wurde kein Antigenüberschuss gefunden. REFERENZBEREICHE (5) Unter 3 mg/l Die angegebenen Werte gelten nur als Richtlinie. Es wird empfohlen, dass jedes Labor diesen Bereich überprüft oder ein Referenzintervall für die betroffene Population ableitet. Bei der Bewertung des Risikos koronarer Herzkrankheiten auf der Grundlage des CRP-Wertes sollten die Messungen mit vorherigen Werten verglichen werden. MESSBEREICH 0.25 – 10 mg/l Erweiterter Messbereich nach weiterer Verdünnung: 0.25 – 40 mg/l LEISTUNGSDATEN (6) Nachweisgrenze Die Nachweisgrenze entspricht der niedrigsten, quantitativ messbaren Konzentration: 0.25 mg/l Impräzision (Ergebnisse in mg/l) Mittelwert 0.80 mg/l Mittelwert 2.41 mg/l SD % VK SD % VK In der Serie 0.02 3.2 0.02 0.8 Von Tag zu Tag 0.03 3.7 0.04 1.8 Gesamtwert 0.04 5.9 0.05 2.3 Mittelwert 7.02 mg/l SD % VK In der Serie 0.04 0.6 Von Tag zu Tag 0.08 1.2 Gesamtwert 0.15 2.2 Unter Verwendung von Konelab 30 wurde mindestens 20 Tage lang eine Präzisionsstudie gemäß NCCLS-Richtlinie EP5-A durchgeführt, bei der die Anzahl der Messungen mindestens n = 80 betrug. Vergleich der Methoden Es wurde eine Vergleichsstudie gemäß NCCLS Richtlinie EP9-A mit einer im Handel erhältlichen Methode als Referenz durchgeführt. Lineare Regression (Ergebnisse in mg/l): y = 1.01 x – 0.29 r = 0.983 n = 87 Die in den Proben enthaltenen Konzentrationen lagen zwischen 0.29 und 10 mg/l. Die Ergebnisse einzelner Laboratorien können von den angegebenen Leistungsdaten abweichen. LITERATURVERWEISE 1) Burtis CA and Ashwood ER (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, WB Saunders Company, Philadelphia, 2001, pp. 332-333 and 688. 2) Guder WG, Narayanan S, Wisser H, Zawta B. List of Analytes; Preanalytical variables. Brochure in: Samples: From Patient to the Laboratory. GIT Verlag GmbH, Darmstadt, 1996. 3) Young DS, Effects of Drugs on Clinical Laboratory Tests, Volume One, Listing by Test, Fifth Edition, AACC Press, Washington, D.C., 3-233 3-234. 4) Selby C. Interference in immunoassay. Ann Clin Biochem 1999;36:704-721. 5) Pearson TA et al. Markers of Inflammation and Cardiovascular Disease Application to Clinical and Public Health Practice, A Statement for Healthcare Professionals From the Centers for Disease Control and the American Heart Association., Circulation 2003;107:499-511. 6) Datenbestände der Thermo Fisher Scientific Oy. 1 x 4 ml CETTE NOTICE EST VALABLE POUR UTILISATION EN DEHORS DES ÉTATS-UNIS. TOUTE RÉFÉRENCE AUX SYSTÈMES KONELAB FAIT ÉGALEMENT RÉFÉRENCE À LA GAMME T. INDICATION Pour la détermination quantitative in vitro de la concentration en protéine C réactive dans le sérum ou le plasma humains au moyen des analyseurs Konelab. Tous les résultats de tests doivent être interprétés en tenant compte du contexte clinique. RÉSUMÉ (1, 5) La protéine C réactive (CRP) est un constituant normal du sérum qui se rencontre à très faible concentration chez les sujets en bonne santé. Une infection bactérienne invasive ou des dommages étendus aux tissus se traduisent par une augmentation de la concentration en protéine C réactive. La protéine C réactive est l'une des protéines de la phase aiguë les plus sensibles ; sa concentration plasmatique présente généralement une augmentation importante suite à un infarctus du myocarde, à un traumatisme, à une infection, à une inflammation, à une intervention chirurgicale ou à une prolifération néoplasique. Cette augmentation débute après 6 à 12 heures et sa concentration peut atteindre 2000 fois la normale. Le dosage de la protéine C réactive est utile en clinique pour le dépistage des maladies organiques, l'évaluation de l'activité d'une pathologie inflammatoire, la détection des infections intercurrentes lors de lupus érythémateux systémique (LES), lors de leucémie ou après une intervention chirurgicale, ainsi que pour la prise en charge de la septicémie et de la méningite néonatales, lorsque le prélèvement d'échantillons pour des examens bactériologiques peut s'avérer difficile. La mesure de la protéine C réactive par des tests de sensibilité peut compléter d'autres marqueurs utilisés pour évaluer le risque de maladies cardiovasculaires et vasculaires périphériques. PRINCIPE DE LA PROCEDURE Cette méthode est basée sur la mesure de l'immunoprécipitation à 540 nm. Des microparticules revêtues d'une CRP anti-humaine sont ajoutées à des échantillons tamponnés. L'augmentation de l'absorbance associée à l'immunoprécipitation est enregistrée lorsque la réaction a atteint son point final. La modification de l'absorbance est proportionnelle à la quantité d'antigène (CRP) dans la solution. INFORMATIONS SUR LE REACTIF Réactif CRP HS Tampon CRP HS Diluant pour échantillon Calibreur CRP HS 1 x 4 ml 2 x 50 ml 1 x 60 ml 2 x 1 ml Concentrations Réactif CRP HS Microparticules revêtues d'une CRP antihumaine (mouton) (lyophilisée) 1 - < 2.5% NaN3 Tampon CRP HS Tris 0.2 mol/l NaN3 < 0.1 % Diluant pour échantillon PBS 0.01 mol/l NaN3 < 0.1 % Calibreur CRP HS Humain NaN3 < 0.1 % La concentration en CRP HS du calibreur est indiquée sur une Fiche de valeurs séparée. Précautions Usage diagnostique in vitro uniquement. Respecter les précautions habituelles requises lors de la manipulation de tout réactif de laboratoire. Ce produit a été testé et s'est avéré négatif pour le HBsAg et les anticorps dirigés contre le HIV-1, le HIV-2 et le HCV. Néanmoins, il convient de le manipuler comme un produit potentiellement infectieux. Le réactif CRP HS lyophilisé contient de l'azide de sodium, utilisé comme conservateur (1 - <2.5%). Eviter d'inhaler les poussières. Ne pas avaler. Éviter tout contact avec la peau et les muqueuses. En cas de contact accidentel, laver avec de grandes quantités d'eau. L'azide de sodium est susceptible de réagir avec le plomb et le cuivre des canalisations pour donner naissance à des nitrures métalliques fortement explosifs. Lors de son élimination, rincer à grande eau. Voir fiche séparée pour les phrases de risque et de sécurité (T:R;23/24/25 - 52/53 S;22-28-45). Après reconstitution, le liquide du réactif CRP HS contient de l'azide de sodium à < 0.1%, ce qui n'est pas considéré comme une concentration nocive. © 2007 Thermo Fisher Scientific Inc. All rights reserved. Störfaktoren Kriterium: Wiederfindung von ± 10 % der Ausgangswerte. Bilirubin: Keine Interferenzen bis zu 1000 μmol/l (58.5 mg/dl). Hämolyse: Keine Interferenzen bei bis zu 10 g/l Hämoglobin im Hämolysat. Lipämie: Keine Interferenzen bis zu 4 g/l Intralipid® (Marke von Fresenius Kabi AB). Es besteht eine schwache Korrelation zwischen der Trübung und der Triglyceridkonzentration. Siehe Literaturhinweis 3 für weitere Substanzen, die Störfaktoren darstellen können. Heterophile Antikörper, monoklonale Immunoglobuline (oder deren Bestandteile) und Rheumatoidfaktoren können die Testergebnisse verfälschen. Die Ergebnisse bei Patienten mit Verdacht auf derartige Antikörper müssen mit größter Sorgfalt ausgewertet werden (4). 981798 Template: D01162_4 Konelab 20/30/60. Die Bezugskurve ist chargenabhängig. GRENZEN DES VERFAHRENS (6) Page 4 D02418_04_insert_CRP HS_MU LIMITES DE LA PROCEDURE (6) Interférence Critère : Récupération se situant dans les limites de ± 10% des valeurs initiales. Bilirubine : Aucune interférence observée jusqu'à 1000 µmol/l (58.5 mg/dl). Hémolyse : Aucune interférence observée jusqu'à 10 g/l d'hémoglobine dans l'hémolysat. Lipémie : Aucune interférence significative jusqu'à 4 g/l d'Intralipid® (marque de Fresenius Kabi AB). Il existe une faible corrélation entre turbidité et concentration en triglycérides. Pour les autres substances interférentes, se reporter à la référence 3. Les anticorps hétérophiles, les immunoglobulines monoclonales (ou certaines parties de celles-ci) et les facteurs rhumatoïdes sont susceptibles d'interférer avec le dosage. Les résultats des patients que l'on suspecte de posséder ce type d'anticorps doivent être évalués avec circonspection (4). Excès d'antigène Aucun excès d'antigène n'a été trouvé jusqu'à une concentration de CRP de 300 mg/l. Conservation et stabilité Les réactifs contenus dans les flacons scellés sont stables à 2… 8 °C jusqu'à la date de péremption figurant sur l'étiquette. Sur la stabilité des réactifs dans l'appareil, se référer à la fiche d'application de l'analyseur Konelab. Stockés hermétiquement fermés à 2…8 °C lorsqu'ils ne sont pas utilisés, le réactif CRP HS reconstitué et le tampon CRP HS ouvert peuvent être utilisés pendant 8 semaines. Il est recommandé de sortir les flacons de l'analyseur et de les conserver fermés au réfrigérateur lorsqu'ils ne sont pas utilisés. Après ouverture, le calibreur CRP HS est stable pendant 8 semaines à 2…8 °C. Stocker le calibreur hermétiquement fermé lorsqu'il n'est pas utilisé. Ne pas utiliser le flacon si celuici présente des signes visibles de contamination microbienne. VALEURS ATTENDUES (5) Inférieures à 3 mg/l Les valeurs citées sont données à titre indicatif uniquement. Il est recommandé que chaque laboratoire établisse les valeurs physiologiques pour la population qui constitue sa clientèle. Lorsque la CRP est utilisée dans l'évaluation du risque de maladie coronarienne, les résultats doivent être comparés avec les valeurs précédentes. PRÉLÈVEMENT DE L'ÉCHANTILLON CARACTÉRISTIQUES RELATIVES AUX PERFORMANCES (6) Type d'échantillon Du sérum et du plasma hépariné ou EDTA peuvent être utilisés. Tout échantillon trouble doit être centrifugé (par exemple à 2000xg pendant 15 minutes). Limite de dosage La limite de dosage est la plus faible concentration qu'il est possible de mesurer de manière quantitative : 0.25 mg/l Précautions Les échantillons d'origine humaine doivent être manipulés et éliminés comme des matériaux potentiellement infectieux. Imprécision (unité du résultat : mg/l) Moyenne 0.80 mg/l Moyenne 2.41 mg/l ET CV % ET CV % Répétabilité 0.02 3.2 0.02 0.8 Reproductibilité 0.03 3.7 0.04 1.8 Total 0.04 5.9 0.05 2.3 Moyenne 7.02 mg/l ET CV % Répétabilité 0.04 0.6 Reproductibilité 0.08 1.2 Total 0.15 2.2 L'étude de précision a eu lieu conformément aux directives du document NCCLS EP5-A sur un analyseur Konelab 30 pendant 20 jours, le nombre de mesures étant de n = 80. PROCÉDURE D'ANALYSE Se référer au manuel de référence et à la fiche d'application pour une description de la procédure automatisée sur l'analyseur Konelab. Toute application n'ayant pas été validée par Thermo Fisher Scientific Oy ne peut pas être garantie en ce qui concerne ses performances et doit par conséquent être évaluée par l'utilisateur. Matériel fourni Réactifs décrits plus haut, et flacon vide à code-barres destiné aux analyseurs Konelab. Matériel requis mais non fourni Contrôle comme indiqué ci-dessous. Calibrage Utiliser le calibreur CRP HS inclus dans le kit selon les instructions fournies pour l'analyseur Konelab. Ne pas laisser le calibreur plus d'une heure sur l'analyseur. Traçabilité : La valeur de la CRP a été assignée en utilisant comme référence principale la préparation IFCC CRM 470. Contrôle de qualité Utiliser un échantillon de contrôle de qualité au moins une fois par jour, après chaque étalonnage et chaque fois que l'on entame un nouveau flacon de réactif. Contrôles disponibles : Contrôle haut CRP, code 981852 Les intervalles et les limites de contrôle doivent être adaptés aux besoins de chaque laboratoire. Les résultats de l'échantillon de contrôle de qualité doivent se situer dans la plage de tolérance prédéfinie par le laboratoire. CALCUL DES RÉSULTATS Les résultats sont calculés automatiquement par l'analyseur Konelab à l'aide d'une courbe de calibrage. L'analyseur réalise automatiquement une série de dilutions à partir du stock de calibreur conformément aux paramètres prédéfinis. Le tracé de la courbe de calibrage est assuré par la fonction SPLINE (fonction polynomiale d'ajustement par déformation sur une courbe). Courbe de calibrage (exemple) CRP HS BIBLIOGRAPHY 1) Burtis CA and Ashwood ER (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, WB Saunders Company, Philadelphia, 2001, pp. 332-333 and 688. 2) Guder WG, Narayanan S, Wisser H, Zawta B. List of Analytes; Preanalytical variables. Brochure in : Samples : From Patient to the Laboratory. GIT Verlag GmbH, Darmstadt, 1996. 3) Young DS, Effects of Drugs on Clinical Laboratory Tests, Volume One, Listing by Test, Fifth Edition, AACC Press, Washington, D.C., 3-233 3-234. 4) Selby C. Interference in immunoassay. Ann Clin Biochem 1999;36:704-721. 5) Pearson TA et al. Markers of Inflammation and Cardiovascular Disease Application to Clinical and Public Health Practice, A Statement for Healthcare Professionals From the Centers for Disease Control and the American Heart Association., Circulation 2003;107:499-511. 6) Données disponibles sur fichiers chez Thermo Fisher Scientific Oy. FABRICANT Thermo Fisher Scientific Oy Clinical Diagnostics Finland Ratastie 2, P.O. Box 100, FI-01621 Vantaa Tél. +358 9 329 100, télécopie +358 9 3291 0300 www.thermo.com/konelab Date de révision (aaaa-mm-jj) 2007-09-13 Modifications par rapport à la version précédente Mise à jour du nom de la société. 1,6 Résponse (A) Comparaison de méthodes Une étude comparative a été réalisée conformément aux directives du document NCCLS EP9-A en utilisant comme référence une méthode disponible dans le commerce. Régression linéaire (unités du résultat : mg/l) y = 1.01 x – 0.29 r = 0.983 n = 87 Les concentrations des échantillons se situaient entre 0.29 et 10 mg/l. Les résultats obtenus dans chaque laboratoire peuvent différer des données de performances indiquées. 1,4 1,2 1 0,8 0,6 0,4 0 5 10 15 Concentration (m g/l) Konelab 20/30/60. La courbe de calibrage dépend du lot. 20 © 2007 Thermo Fisher Scientific Inc. All rights reserved. Conservation (2) Les échantillons peuvent être conservés pendant 8 jours à 2…8 °C ou pendant 3 mois à –20 °C (à condition de n'être congelés qu'une seule fois). GAMME DE MESURE 0.25 – 10 mg/l. Domaine de mesure étendu après dilution secondaire : 0.25 – 40 mg/l. Template: D01162_4 Préparation Le tampon CRP HS, le diluant des échantillons et le calibreur CRP HS sont prêts à l'usage. Reconstitution du réactif CRP HS : Ajouter exactement 4 ml d'eau déminéralisée. Fermer le flacon, mélanger délicatement et laisser au moins 30 minutes à température ambiante. Mélanger de nouveau le contenu avant usage. Ne pas secouer. Transférer le réactif dans un flacon vide à code-barres pour analyseur Konelab. Remarque 1 : Toujours utiliser le tampon CRP HS, le réactif CRP HS et le calibreur CRP HS portant le même numéro de lot de réactif. En effet, ils ont été appariés à l'intérieur du lot. Remarque 2 : Mélanger délicatement le réactif CRP HS avant de l'introduire dans l'analyseur Konelab. Ne pas secouer. Remarque 3 : S'assurer de l'absence de bulles au niveau du goulot du flacon ou à la surface du réactif lors de la mise en place des flacons ou récipients de réactifs dans l'analyseur Konelab. D02418_04_insert_CRP HS_MU CS Uchovávání (2) Vzorek lze uchovávat po dobu 8 dnů při teplotě 2…8 °C nebo po dobu 3 měsíců při teplotě – 20 °C (vzorek je možné zmrazit pouze jednou). Konelab™ / Série T CRP High Sensitivity POSTUP TESTU Údaje o automatizovaném postupu práce na analyzátoru Konelab naleznete v Referenčním manuálu a aplikačních poznámkách. Nelze zaručit provedení žádné aplikace, která nebyla validována společnosti Thermo Fisher Scientific Oy. Taková aplikace proto musí být hodnocena uživatelem. C-reaktivní protein, stanovení s vysokou citlivostí 1 x 4 ml TENTO PŘÍBALOVÁ INFORMACE JE URČENA PRO POUŽITÍ MIMO ÚZEMÍ USA. KAŽDÁ ZMÍNKA O SYSTÉMECH KONELAB SE ROVNĚŽ TÝKÁ SÉRIE T. URČENÉ POUŽITÍ Pro kvantitativní stanovení nízkých koncentrací CRP in vitro v lidském séru a plazmě na analyzátorech Konelab. Všechny výsledky testů musejí být interpretovány s ohledem na klinický kontext. Dodávané materiály Reagencia uvedená výše a prázdná lahvička označená čárovým kódem určená pro použití v analyzátorech Konelab. Požadované, avšak nedodávané materiály Kontrolní materiály uvedené dále. Kalibrace Používejte kalibrátor CRP HS obsažený v soupravě podle návodu, který vám byl dodán s analyzátorem Konelab. Kalibrátory používejte v analyzátoru nejvýše po dobu jedné hodiny. SHRNUTÍ (1, 5) C-reaktivní protein (CRP) je normální složkou séra přítomnou u zdravých osob ve velmi nízkých koncentracích. Zvýšené hladiny CRP jsou způsobeny invazivní bakteriální infekcí a rozsáhlým poškozením tkání. CRP je jednou z nejcitlivějších bílkovin akutní fáze; hladiny v plazmě se obvykle dramaticky zvyšují po infarktu myokardu, úrazu, infekci, zánětu, chirurgickém zákroku nebo proliferaci novotvaru. Zvýšení začíná během 6 až 12 hodin a hladina může dosáhnout až 2000 násobku normálních hodnot. Stanovení CRP je klinicky užitečné pro screening organických chorob, posouzení aktivity zánětlivého onemocnění, stanovení přidružených infekcí u systémového lupus erythematodes (SLE), u leukémie nebo po chirurgickém zákroku a při léčbě novorozenecké septikémie a meningitidy, kdy odběr vzorků pro bakteriologická vyšetření může být obtížný. Měření CRP citlivým stanovením CRP může zvýšit predikční hodnotu dalších markerů použitých pro posouzení rizika kardiovaskulárních onemocnění a onemocnění periferních cév. Řízení jakosti Používejte vzorek pro řízení jakosti alespoň jednou denně, po každé kalibraci a vždy, když použijete novou lahvičku reagens. Dodávané kontrolní materiály: Kontrolní materiál CRP High Sensitivity, kód 981852 Intervaly a limitní hodnoty kontrolních materiálů musejí být upraveny podle požadavků jednotlivých laboratoří. Výsledek vzorku pro řízení jakosti by měl vyhovovat limitním hodnotám přednastaveným laboratoří. PRINCIP POSTUPU Tato metoda je založena na měření imunoprecipitace při 540 nm. K pufrovaným vzorkům se přidávají mikročástice potažené protilátkami proti lidskému CRP. Zvýšení absorbance způsobené imunoprecipitací je zaznamenáno, když reakce dosáhne koncového bodu. Změna absorbance je přímo úměrná množství antigenu (CRP) v roztoku. VÝPOČET VÝSLEDKŮ Výsledky jsou automaticky vypočteny analyzátorem Konelab pomocí kalibrační křivky. Analyzátor automaticky připraví řadu ředění ze základního kalibrátoru podle předem nastavených parametrů. Kalibrační křivka je generována na základě měřených kalibrátorů pomocí kubické interpolace pro vyhlazování křivek. INFORMACE O REAGENCIÍCH Reagens CRP HS Pufr CRP HS Činidlo pro ředění vzorků Kalibrátor CRP HS Kalibrační křivka (příklad) 1 x 4 ml 2 x 50 ml 1 x 60 ml 2 x 1 ml Mikročástice potažené protilátkami lidskému CRP (ovčí) (lyofilizované) NaN3 1 - < 2.5% Pufr CRP HS TRIS 0.2 mol/l NaN3 < 0.1 % Činidlo pro ředění vzorků PBS 0.01 mol/l NaN3 < 0.1 % Kalibrátor CRP HS Lidský NaN3 < 0.1 % Koncentrace CRP v kalibrátoru je vytištěna v samostatné tabulce hodnot. CRP HS Zvláštní opatření Určeno pouze pro diagnostické použití in vitro. Používejte běžná bezpečnostní opatření vyžadovaná pro manipulaci se všemi laboratorními reagencii. Kalibrátor obsahuje materiál lidského původu, který byl testován, a bylo zjištěno, že je negativní pro HBsAg a protilátky proti HIV1/2 a HCV. S výrobkem je však nutné manipulovat stejně opatrně jako se vzorky pacientů. Lyofilizované reagens CRP HS obsahuje jako konzervační látku azid sodný (1 - <2.5 %). Zabraňte vdechování prachu. Nepolykejte. Zabraňte kontaktu s kůží a sliznicemi. Dojde-li ke kontaktu, opláchněte velkým množstvím vody. Azid sodný může reagovat s olovem a mědí v odpadech a vytvářet vysoce výbušné azidy kovů. Při likvidaci spláchněte velkým množstvím vody. Standardní pokyny pro bezpečné zacházení s nebezpečnými látkami (Světy) a věty popisující charakter nebezpečnosti chemických látek (R-věty) naleznete na samostatném listu (T:R;23/24/25 - 52/53 S;22-28-45). Po rekonstituci obsahuje tekuté reagens CRP HS azid sodný v množství < 0.1%; tato koncentrace není považována za škodlivou. Příprava Pufr CRP HS, činidlo pro ředění vzorků a kalibrátor CRP HS jsou připraveny k použití. Rekonstituce reagens CRP HS: Přidejte přesně 4 ml deionizované vody. Lahvičku uzavřete, jemně ji zamíchejte a nechte ji stát 30 minut při pokojové teplotě. Před použitím obsah znovu promíchejte. S lahvičkou netřepejte. Reagens převeďte do prázdné lahvičky označené čárovým kódem, která bude použita v analyzátorech Konelab. Poznámka 1: Vždy používejte pufr CRP HS, reagens CRP HS a kalibrátor CRP HS, které mají stejné číslo šarže, protože složky v rámci jednotlivých šarží byly vzájemně sladěny. Poznámka 2: Před vložením reagens CRP HS do analyzátoru Konelab reagens jemně promíchejte. S lahvičkou netřepejte. Poznámka 3: Překontrolujte, zda při vkládání lahviček nebo nádob s reagencii do analyzátoru Konelab nejsou v hrdle lahvičky nebo na povrchu reagens bubliny. Uchovávání a stabilita Reagencia v neotevřených lahvičkách jsou stabilní při teplotě 2…8 °C, a to do data ukončení použitelnosti vytištěného na štítku. Údaje o stabilitě reagencií “on-board” naleznete v aplikačních poznámkách svého analyzátoru Konelab. Rekonstituované reagens CRP HS a otevřený pufr CRP HS lze použít po dobu 8 týdnů, pokud v době, kdy nejsou používány, jsou uchovávány při teplotě 2…8 °C. V době, kdy není reagens používáno, se doporučuje vyjmout lahvičky reagens z analyzátoru a uchovávat je uzavřené v chladničce. Otevřený kalibrátor CRP HS je stabilní po dobu 8 týdnů při teplotě 2…8 °C. V době, kdy kalibrátor nepoužíváte, ho uchovávejte dobře uzavřený. Nepoužívejte žádnou lahvičku, na níž jsou patrné stopy růstu mikroorganismů. ODBĚR VZORKŮ Typ vzorků Lze použít sérum, heparinizovanou plazmu a EDTA plazmu. Každý zakalený vzorek je nutné centrifugovat (např. 2 000 xg po dobu 15 minut). Zvláštní opatření S lidskými vzorky je nutné nakládat a likvidovat je, jako by byly potenciálně infekční. 1.6 proti 1.4 Odezva (A) Koncentrace Reagens CRP HS Identifikovatelnost: Hodnota CRP byla stanovena pomocí preparátu CRM 470 IFCC, jenž byl použit jako primární reference. 1.2 1 0.8 0.6 0.4 0 5 10 15 20 Koncentrace (mg/l) Konelab 20/30/60. Kalibrační křivka je závislá na konkrétní šarži. OMEZENÍ POSTUPU (6) Interference Kritérium: Recovery v rozpětí ± 10% počátečních hodnot. Bilirubin: Až do 1 000 μmol/l (58.5 mg/dl) nebyla zjištěna interference. Hemolýza: Až do 10 g/l hemoglobinu v hemolyzátu nebyla zjištěna interference. Lipémie: Až do 4 g/l přípravku Intralipid® (ochranná známka společnosti Fresenius Kabi AB) nebyla zjištěna interference. Mezi turbiditou a koncentrací triglyceridů existuje nízká korelace. Údaje o dalších interferujících látkách naleznete v odkazu č. 3. Kvantitativní rozbor může být narušen přítomností heterofilních protilátek, monoklonálních imunoglobulinů (nebo jejich částí) a revmatoidních faktorů. Výsledky pacientů, u nichž existuje podezření na přítomnost těchto protilátek, je nutné pečlivě hodnotit (4). Nadměrné množství antigenů Při koncentraci CRP do 300 mg/l nebyla zjištěna přítomnost nadměrného množství antigenů. PŘEDPOKLÁDANÉ HODNOTY (5) Méně než 3 mg/l Uvedené hodnoty by měly sloužit pouze jako vodítko. Doporučuje se, aby každá laboratoř toto rozpětí ověřila nebo odvodila referenční interval pro populaci, jíž poskytuje služby. Při použití CRP pro posouzení rizika ischemické choroby srdeční je nutné měření srovnat s předchozími hodnotami. ROZPĚTÍ MĚŘENÍ 0.25 – 10 mg/l Rozšířené rozpětí měření po sekundárním ředění: 0.25 – 40 mg/l CHARAKTERISTIKY ÚČINNOSTI (6) Limitní hodnota detekce Limitní hodnota detekce je nejnižší koncentrace, kterou lze kvantitativně stanovit. 0.25 mg/l © 2007 Thermo Fisher Scientific Inc. All rights reserved. 981798 Template: D01162_4 Page 5 D02418_04_insert_CRP HS_MU Srovnání metod Byla provedena srovnávací studie podle pokynů uvedených v dokumentu NCCLS EP9-A, při níž byla použita komerčně dostupná metoda jako metoda referenční. Lineární regrese (jednotky výsledku mg/l): y = 1.01 x – 0.29 r = 0.983 n = 87 Koncentrace vzorků se pohybovaly v rozmezí 0.29 až 10 mg/l. Výsledky získané v jednotlivých laboratořích se mohou od uvedených dat účinnosti lišit. SEZNAM LITERATURY 1) Burtis CA and Ashwood ER (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, WB Saunders Company, Philadelphia, 2001, pp. 332-333 and 688. 2) Guder WG, Narayanan S, Wisser H, Zawta B. List of Analytes; Preanalytical variables. Brochure in: Samples: From Patient to the Laboratory. GIT Verlag GmbH, Darmstadt, 1996. 3) Young DS, Effects of Drugs on Clinical Laboratory Tests, Volume One, Listing by Test, Fifth Edition, AACC Press, Washington, D.C., 3-233 – 3-234. 4) Selby C. Interference in immunoassay. Clin Biochem 1999;36:704-721. 5) Pearson TA et al. Markers of Inflammation and Cardiovascular Disease Application to Clinical and Public Health Practice, A Statement for Healthcare Professionals From the Centers for Disease Control and the American Heart Association., Circulation 2003;107:499-511. 6) Příslušné údaje jsou k dispozici u společnosti Thermo Fisher Scientific Oy. VÝROBCE Thermo Fisher Scientific Oy Clinical Diagnostics Finland Ratastie 2, P.O. Box 100, FI-01621 Vantaa Tel.: +358 9 329 100, Fax +358 9 3291 0300 www.thermo.com/konelab Datum revize (rrrr-mm-dd) 2007-09-13 Změny oproti předchozí verzi Název společnosti byl aktualizován. EL Konelab™ / Σειρά T CRP High Sensitivity CRP Υψηλής Ευαισθησίας 981798 1 x 4 ml ΑΥΤΟ ΤΟ ΕΝΘΕΤΟ ΣΥΣΚΕΥΑΣΙΑΣ ΠΡΟΤΙΘΕΤΑΙ ΓΙΑ ΧΡΗΣΗ ΕΚΤΟΣ ΤΩΝ ΗΠΑ. ΟΠΟΙΑΔΗΠΟΤΕ ΑΝΑΦΟΡΑ ΣΤΑ ΣΥΣΤΗΜΑΤΑ KONELAB ΑΝΑΦΕΡΕΤΑΙ ΕΠΙΣΗΣ ΣΤΗ ΕΙΡΑ T. ΠΡΟΟΡΙΖΟΜΕΝΗ ΧΡΗΣΗ Για τον in vitro ποσοτικό προσδιορισμό των χαμηλών συγκεντρώσεων τριγλυκεριδίων σε ανθρώπινο ορό ή πλάσμα χρησιμοποιώντας αναλυτές Konelab. Όλα τα αποτελέσματα της εξέτασης πρέπει να ερμηνεύονται σε σχέση με το κλινικό περιβάλλον. ΠΕΡΙΛΗΠΤΙΚΑ (1,5) Η C-δραστική πρωτεΐνη (CRP) αποτελεί ένα φυσιολογικό συστατικό του ορού και εντοπίζεται σε πολύ χαμηλή συγκέντρωση στα υγιή άτομα. Η επεκτατική βακτηριδιακή μόλυνση και η εκτεταμένη καταστροφή ιστού ενδέχεται να προκαλέσουν αυξημένα επίπεδα της CRP. H CRP είναι μία από τις πλέον ευαίσθητες πρωτεΐνες της οξείας φάσης. Τα επίπεδά της στο πλάσμα αυξάνονται συνήθως δραματικά μετά από έμφραγμα του μυοκαρδίου, τραυματισμό, μόλυνση, φλεγμονή, χειρουργική επέμβαση ή νεοπλασματική αναπαραγωγή. Η αύξηση ξεκινά σε διάστημα από 6 έως 12 ώρες και τα επίπεδα μπορούν να ξεπεράσουν κατά 2000 φορές τα φυσιολογικά επίπεδα. Ο καθορισμός της CRP είναι κλινικά χρήσιμος για τον εντοπισμό οργανικής πάθησης, την αξιολόγηση της δραστηριότητας μιας φλεγμονώδους ασθένειας, την ανίχνευση παρεμπιπτόντων μολύνσεων σε περίπτωση συστηματικού ερυθρηματώδους λύκου (SLE), λευχαιμίας ή μετά από χειρουργική επέμβαση και θεραπεία της νεογνικής σηψαιμίας και μηνιγγίτιδας, στην περίπτωση που η συλλογή δειγμάτων για βακτηριδιακή ανάλυση είναι δύσκολη. Η μέτρηση της CRP με ευαίσθητες μεθόδους CRP, αυξάνει την προγνωστική αξία άλλων δεικτών που χρησιμοποιούνται στην αξιολόγηση του κινδύνου σε καρδιαγγειακή και περιφερική αγγειακή νόσο. ΑΡΧΗ ΤΗΣ ΔΙΑΔΙΚΑΣΙΑΣ Η μέθοδος βασίζεται στη μέτρηση της ανοσοκαθίζησης στα 540 nm. Τα επιστρωμένα μικροσωματίδια με αντι-ανθρώπινη CRP προστίθενται στα δείγματα των ρυθμιστικών διαλυμάτων. Η αύξηση της απορρόφησης που προκαλείται λόγω της ανοσοκαθίζησης καταγράφεται, όταν η αντίδραση φτάνει πλέον στο τελικό σημείο. Η αλλαγή στην απορρόφηση είναι αναλογική προς το ποσοστό αντιγόνου (CRP) που υπάρχει στο διάλυμα. ΠΛΗΡΟΦΟΡΙΕΣ ΑΝΤΙΔΡΑΣΤΗΡΙΩΝ Αντιδραστήριο CRP HS Ρυθμιστικό διάλυμα CRP HS Αραιωτικό Δείγματος Βαθμονομητής CRP HS 1 x 4 ml 2 x 50 ml 1 x 60 ml 2 x 1 ml Συγκεντρώσεις Αντιδραστήριο CRP HS Επικαλυμμένα μικροσωματίδια με αντι-ανθρώπινη CRP (λυοφιλοποιημένη πρωτεΐνη προβάτου) NaN3 1 - < 2.5% Ρυθμιστικό διάλυμα CRP HS Tris 0.2 mol/l NaN3 < 0.1 % Αραιωτικό Δείγματος PBS 0.01 mol/l NaN3 < 0.1 % Βαθμονομητής CRP HS Ανθρώπινος NaN3 < 0.1 % Η συγκέντρωση του ρυθμιστικού διαλύματος της CRP βρίσκεται σε ξεχωριστό Φύλλο τιμών. Προφυλάξεις Για in vitro διαγνωστική χρήση μόνο. Εφαρμόζετε τις κανονικές προφυλάξεις που απαιτούνται για το χειρισμό όλων των εργαστηριακών αντιδραστηρίων. Ο Βαθμονομητής περιέχει υλικό ανθρώπινης προέλευσης, το οποίο έχει βρεθεί αρνητικό σε HBsAg και αντισώματα HIV 1/2 και HCV. Ωστόσο, ο χειρισμός του προϊόντος πρέπει να γίνεται με την ίδια προσοχή όπως και σε δείγματα ασθενών. Το λυοφιλοποιημένο Αντιδραστήριο CRP HS περιέχει αζίδιο του νατρίου ως συντηρητικό (1 - < 2.5%). Αποφύγετε την εισπνοή σκόνης. Μην καταπίνετε. Αποφύγετε την επαφή με το δέρμα και με τους βλεννογόνους. Αν συμβεί επαφή, ξεπλύνετε με άφθονη ποσότητα νερού. Το νατραζίδιο μπορεί να αντιδράσει με μόλυβδο και χαλκό στα αποχετευτικά συστήματα για να σχηματίσει ιδιαίτερα εκρηκτικά μεταλλικά αζίδια. Κατά την απόρριψη ξεπλύνετε με μεγάλη ποσότητα νερού. Δείτε το ξεχωριστό φυλλάδιο των παρατηρήσεων κινδύνων και ασφαλείας (Τ:R,23/24/25-52-52/53 S,22-28-45). Μετά την ανασύσταση το υγρό Αντιδραστήριο CRP HS περιέχει αζίδιο του νατρίου < 0.1 το οποίο δε θεωρείται ως βλαβερή συγκέντρωση. Προετοιμασία Το ρυθμιστικό διάλυμα CRP HS, ο διαλύτης του δείγματος και ο βαθμονομητής CRP HS είναι έτοιμα για χρήση. Ανασύσταση του αντιδραστηρίου CRP HS: Προσθέστε ακριβώς 4 ml απιονισμένου νερού. Κλείστε το φιαλίδιο και αφήστε το να ηρεμήσει για 30 τουλάχιστον λεπτά σε θερμοκρασία δωματίου. Αναμίξτε τα περιεχόμενα ακόμα μια φορά πριν τη χρήση. Μην αναταράξετε έντονα. Μεταφέρετε το αντιδραστήριο σε ένα άδειο φιαλίδιο με bar-codes για χρήση στους αναλυτές Konelab. Σημείωση 1: Χρησιμοποιείται πάντα ρυθμιστικό διάλυμα CRP HS, αντιδραστήρια CRP HS και βαθμονομητή CRP HS με τον ίδιο αριθμό παραγωγής εξαιτίας της συμβατότητας των συστατικών στις παρτίδες. Σημείωση 2: Αναμίξτε το αντιδραστήριο CRP HS ήπια πριν εισάγετε το αντιδραστήριο στον αναλυτή Konelab. Μην αναταράξετε έντονα. Σημείωση 3: Ελέγξτε ότι δεν υπάρχουν φυσαλίδες στο λαιμό ή στην επιφάνεια του φιαλιδίου αντιδραστηρίου όταν εισάγετε τα φιαλίδια ή τα δοχεία αντιδραστηρίων στην αναλυτή Konelab. Αποθήκευση και Σταθερότητα Τα αντιδραστήρια σε φιαλίδια που δεν έχουν ανοιχτεί είναι σταθερά στους 2...8 °C μέχρι την ημερομηνία λήξης που αναγράφεται στην ετικέτα. Ανατρέξτε στις Σημειώσεις Εφαρμογών του αναλυτή Konelab για την σταθερότητα των αντιδραστηρίων όταν αυτά είναι τοποθετημένα στο μηχάνημα. Το αναπαρασκευασμένο αντιδραστήριο CRP HS και το ρυθμιστικό διάλυμα CRP HS που έχει ήδη ανοιχθεί μπορεί να χρησιμοποιηθεί για διάστημα 8 εβδομάδων εφόσον έχει αποθηκευτεί στους 2…8 °C όταν δε χρησιμοποιείται. Προτείνεται τα φιαλίδια του αντιδραστηρίου να αφαιρούνται από τον αναλυτή και να διατηρούνται κλειστά στο ψυγείο, όταν δε χρησιμοποιούνται. Όταν ανοιχθεί, ο βαθμονομητής CRP HS είναι σταθερός για 8 εβδομάδες στους 2…8 °C. Διατηρήστε το βαθμονομητή καλά κλειστό, όταν δεν τον χρησιμοποιείτε. Μη χρησιμοποιείτε οποιοδήποτε φιαλίδιο έχει ορατές ενδείξεις για μικροβιακή ανάπτυξη. ΣΥΛΛΟΓΗ ΔΕΙΓΜΑΤΩΝ Τύπος δείγματος Μπορεί να χρησιμοποιηθεί ορός, ηπαρίνη, πλάσμα και πλάσμα EDTA. Κάθε θολό δείγμα πρέπει να φυγοκεντρηθεί (π.χ. 2000xg επί 15 λεπτά). Προφυλάξεις Τα ανθρώπινα δείγματα πρέπει να χρησιμοποιούνται και να απορρίπτονται σαν να ήταν δυνητικά μολυσματικά. Αποθήκευση (2) Το δείγμα μπορεί να αποθηκευτεί για 8 ημέρες στους 2…8 °C ή για 3 μήνες στους –20 °C (αν έχει καταψυχθεί μόνο μία φορά). ΔΙΑΔΙΚΑΣΙΑ ΕΞΕΤΑΣΗΣ Ανατρέξτε στα φυλλάδια του Εγχειριδίου Αναφοράς και των Σημειώσεων Εφαρμογών για μια αυτόματη διαδικασία του αναλυτή σας Konelab. Η καλή λειτουργία οποιασδήποτε εφαρμογής που δεν έχει επικυρωθεί από την Thermo Fisher Scientific Oy, δεν μπορεί να έχει εγγύηση απόδοσης και επομένως πρέπει να εκτιμηθεί από το χρήστη. Παρεχόμενα Υλικά Αντιδραστήρια, όπως περιγράφηκε προηγουμένως και ένα άδειο φιαλίδιο με bar-codes για χρήση στους αναλυτές Konelab. Υλικά που απαιτούνται αλλά δεν παρέχονται Δείγμα ποιοτικού ελέγχου όπως περιγράφεται παρακάτω. Βαθμονόμηση Χρησιμοποιήστε το βαθμονομητή CRP HS που περιλαμβάνεται στο κιτ σύμφωνα με τις οδηγίες που δίνονται στον αναλυτή Konelab. Αφήστε το βαθμονομητή σε ηρεμία στον αναλυτή, το πολύ μια ώρα. Ανιχνευσιμότητα: Η τιμή της CRP έχει προσδιοριστεί χρησιμοποιώντας παρασκεύασμα IFCC CRM 470 ως αρχική αναφορά. © 2007 Thermo Fisher Scientific Inc. All rights reserved. Nepřesnost (výsledky v jednotkách mg/l) Střední hodnota 0.80 mg/l Střední hodnota 2.41 mg/l SD % CV SD % CV V rámci série 3.2 0.02 0.8 0.02 Mezi dny 0.03 3.7 0.04 1.8 Celkem 0.04 5.9 0.05 2.3 Střední hodnota 7.02 mg/l SD % CV V rámci série 0.6 0.04 Mezi dny 0.08 1.2 Celkem 0.15 2.2 Byla provedena studie přesnosti podle pokynů v dokumentu NCCLS EP5-A pomocí analyzátoru Konelab 30 po dobu nejméně 20 dnů, v níž byl počet měření nejméně n = 80. Template: D01162_4 Page 6 Page 7 D02418_04_insert_CRP HS_MU Ποιοτικός Έλεγχος Χρησιμοποιήστε δείγμα ποιοτικού ελέγχου τουλάχιστον μια φορά την ημέρα και μετά από βαθμονόμηση καθώς επίσης και κάθε φορά που χρησιμοποιείτε καινούργια φιάλη αντιδραστηρίου. Διαθέσιμα υλικά ελέγχου: CRP High Sensitivity Control (Υψηλής Ευαισθησίας Υλικό ελέγχου CPR), κωδικός 981852 Τα διαστήματα και τα όρια των υλικών ελέγχου πρέπει να προσαρμόζονται στις απαιτήσεις του κάθε εργαστηρίου. Το αποτέλεσμα του δείγματος ποιοτικού ελέγχου πρέπει να εμπίπτουν εντός των ορίων που έχει προκαθορίσει το εργαστήριο. Σύγκριση μεθόδου Διεξήχθη συγκριτική μελέτη σύμφωνα με τις οδηγίες στο Έγγραφο NCCLS EP9-A με χρήση μιας εμπορικά διαθέσιμης μεθόδου ως αναφορά. Γραμμική παλινδρόμηση (μονάδα αποτελεσμάτων mg/I): y = 1.01 x – 0.29 r = 0.983 n = 87 Οι συγκεντρώσεις δείγματος ήταν μεταξύ 0.29 και 10 mg/l. Τα αποτελέσματα που λαμβάνονται σε ξεχωριστά εργαστήρια πιθανόν να διαφέρουν από τα δεδομένα απόδοσης. ΥΠΟΛΟΓΙΣΜΟΣ ΤΩΝ ΑΠΟΤΕΛΕΣΜΑΤΩΝ Τα αποτελέσματα υπολογίζονται αυτόματα από τον αναλυτή Konelab με χρήση καμπύλης βαθμονόμησης. Ο αναλυτής προετοιμάζει αυτόματα μια σειρά από αραιώσεις από το βαθμονομητή στοκ, ανάλογα με τις προρυθμισμένες παραμέτρους. Η καμπύλη βαθμονόμησης προκύπτει από τους υπολογισμένους βαθμονομητές χρησιμοποιώντας την διαδικασία εξομάλυνσης καμπύλης. ΒΙΒΛΙΟΓΡΑΦΊΑ 1) Burtis CA and Ashwood ER (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, WB Saunders Company, Philadelphia, 2001, pp. 332-333 and 688. 2) Guder WG, Narayanan S, Wisser H, Zawta B. List of Analytes; Preanalytical variables. Brochure in: Samples: From Patient to the Laboratory. GIT Verlag GmbH, Darmstadt, 1996. 3) Young DS, Effects of Drugs on Clinical Laboratory Tests, Volume One, Listing by Test, Fifth Edition, AACC Press, Washington, D.C., 3-233 – 3-234. 4) Selby C. Interference in immunoassay. Ann Clin Biochem 1999;36:704-721. 5) Pearson TA et al. Markers of Inflammation and Cardiovascular Disease Application to Clinical and Public Health Practice, A Statement for Healthcare Professionals From the Centers for Disease Control and the American Heart Association., Circulation 2003;107:499-511. 6) Δεδομένα στο αρχείο της Thermo Fisher Scientific Oy. Καμπύλη βαθμονόμησης (παράδειγμα) CRP HS Απόκριση Απόκριση (A) (A) 1,6 ΚΑΤΑΣΚΕΥΑΣΤΗΣ Thermo Fisher Scientific Oy Clinical Diagnostics Finland Ratastie 2, P.O. Box 100, FI-01620 Vantaa Τηλ. +358 9 329 100, Fax +358 9 3291 0300 www.thermo.com/konelab 1,4 1,2 1 0,8 Ημερομηνία αναθεώρησης (εεεε-μμ-ηη) 2007-09-13 0,6 0,4 0 5 10 15 20 Αλλαγές από την προηγούμενη έκδοση Ενημερωμένο όνομα εταιρίας. Συγκέντρωση (m g/l) Συγκέντρωση (mg/l) Konelab 20/30/60. Η καμπύλη βαθμονόμησης εξαρτάται από την παρτίδα. Για άλλες παρεμβαλλόμενες ουσίες, παρακαλούμε να ανατρέξετε στην παραπομπή 3. Τα ετερόφυλα αντισώματα, οι μονοκλωνικές ανοσοσφαιρίνες (ή μέρη αυτών) καθώς και οι ρευματοειδείς παράγοντες ενδέχεται να παρεμβληθούν στην ανάλυση. Τα αποτελέσματα των ασθενών, για τους οποίους υπάρχουν υποψίες ότι διαθέτουν τέτοια αντισώματα, πρέπει να αξιολογούνται προσεκτικά (4). Περίσσεια Αντιγόνου Δε βρέθηκε περίσσεια αντιγόνου σε συγκεντρώσεις CRP έως 300 mg/l. ΑΝΑΜΕΝΟΜΕΝΕΣ ΤΙΜΕΣ (5) Κάτω από 3 mg/l Οι προαναφερθείσες τιμές πρέπει να εκλαμβάνονται ως οδηγός μόνον. Συνιστάται το κάθε εργαστήριο να διεξάγει την επαλήθευση αυτού του εύρους τιμών ή να παράγει ένα διάστημα αναφοράς για τον πληθυσμό με τον οποίο ασχολείται. Όταν χρησιμοποιείτε την CRP για να αξιολογήσετε τον κίνδυνο στεφανιαίας νόσου οι τιμές (μετρήσεις) θα πρέπει να συγκριθούν με προηγούμενες μετρήσεις. ΕΥΡΟΣ ΜΕΤΡΗΣΗΣ 0.25 - 10 mg/l Εκτεταμένο εύρος μέτρησης μετά από δευτερογενή αραίωση: 0.25 - 40 mg/l ΧΑΡΑΚΤΗΡΙΣΤΙΚΑ ΑΠΟΔΟΣΗΣ (6) Όριο προσδιορισμού Το όριο προσδιορισμού είναι η χαμηλότερη συγκέντρωση που μπορεί να μετρηθεί ποσοτικά: 0.25 mg/l Ανακρίβεια (μονάδα αποτελεσμάτων mg/l) Μέση τιμή 0.80 mg/l Μέση τιμή 2.41 mg/l SD CV% SD CV% Εντός της εκτέλεσης 0.02 3.2 0.02 0.8 Μεταξύ ημερών 0.03 3.7 0.04 1.8 Συνολικό 0.04 5.9 0.05 2.3 Μέση τιμή 7.02 mg/l SD CV% Εντός της εκτέλεσης 0.04 0.6 Μεταξύ ημερών 0.08 1.2 Συνολικό 0.15 2.2 Διεξάχθηκε μελέτη ακριβείας σύμφωνα με τις οδηγίες στο Έγγραφο NCCLS EP5-A χρησιμοποιώντας το Konelab 30 για διάστημα τουλάχιστον 20 ημερών, με τον αριθμό των μετρήσεων n = 80 τουλάχιστον. ES Konelab™ / Serie T CRP High Sensitivity PCR de alta sensibilidad 981798 1 x 4 ml ESTE PROSPECTO ES PARA USO FUERA DE EE. UU. TODAS LAS REFERENCIAS A LOS SISTEMAS KONELAB SE REFIEREN TAMBIÉN A LA SERIE T. INDICACIONES Determinación cuantitativa in vitro en analizadores Konelab de bajas concentraciones de PCR en suero y plasma humanos. Todos los resultados del test deben interpretarse en función del contexto clínico. RESUMEN (1, 5) La proteína C reactiva (PCR) es un constituyente habitual del suero presente en concentraciones muy bajas en individuos sanos. Los niveles de PCR aumentan en caso de infecciones bacterianas agresivas y daños tisulares extensos. La PCR es una de las proteínas de fase aguda más sensibles; los niveles en plasma suelen incrementarse drásticamente con infartos de miocardio, traumatismos, infecciones, inflamaciones, cirugía o proliferación neoplástica. El aumento se produce durante las 6 12 horas siguientes y puede alcanzar un nivel 2000 veces superior al normal. La determinación de PCR es útil clínicamente para detectar enfermedades orgánicas, evaluar la actividad de enfermedades inflamatorias, detectar infecciones intercurrentes en lupus eritematoso sistémico (LES), en leucemia o en postoperatorios y para el tratamiento de meningitis y septicemia neonatal cuando resulta difícil recoger muestras para su investigación bacteriológica. La medición de PCR con ensayos sensibles a la PCR pueden aumentar el valor predictivo de otros marcadores utilizados para evaluar el riesgo de enfermedades cardiovasculares y enfermedades vasculares periféricas. PRINCIPIO DEL PROCEDIMIENTO El método se basa en la medición de la inmunoprecipitación a 540 nm. Se agregan micropartículas revestidas con PCR antihumana a muestras tamponadas. El incremento de absorbencia provocado por la inmunoprecipitación se registra cuando la reacción alcanza su punto final. El cambio de absorbencia es proporcional a la cantidad de antígeno (PCR) en la solución. INFORMACIÓN SOBRE EL REACTIVO Reactivo PCR alta sensibilidad 1 x 4 ml Tampón PCR alta sensibilidad 2 x 50 ml Diluyente de muestras 1 x 60 ml Calibrador PCR alta sensibilidad 2 x 1 ml Concentraciones Reactivo PCR alta sensibilidad Micropartículas revestidas con PCR antihumana (oveja) (liofilizada) NaN3 1 - < 2.5% Tampón PCR alta sensibilidad Tris 0.2 mol/l NaN3 < 0.1 % Diluyente de muestras PBS 0.01 mol/l NaN3 < 0.1 % Calibrador PCR alta sensibilidad Humano NaN3 < 0.1 % La concentración de PCR del calibrador aparece impresa en una hoja de valores aparte. © 2007 Thermo Fisher Scientific Inc. All rights reserved. Παρεμβολές Κριτήριο: Ανάκτηση εντός ±10% των αρχικών τιμών. Χολερυθρίνη: Δε βρέθηκε παρεμβολή έως 1000 μmol/l (58.5 mg/dl). Αιμόλυση: Δεν βρέθηκε παρεμβολή μέχρι 10 g/l αιμοσφαιρίνης. Λιπιδιαιμία: Καμία παρεμβολή δεν εντοπίστηκε μέχρι τα 4 g/l Intralipid® (εμπορικό σήμα της Fresenius Kabi AB). Υπάρχει μικρή συσχέτιση ανάμεσα στη θολερότητα και τη συγκέντρωση τριγλυκεριδίων. Template: D01162_4 ΠΕΡΙΟΡΙΣΜΟΙ ΤΗΣ ΔΙΑΔΙΚΑΣΙΑΣ (6) D02418_04_insert_CRP HS_MU Almacenamiento y estabilidad Los reactivos en viales sin abrir son estables a 2…8 °C hasta la fecha de caducidad impresa en la etiqueta. Consulte las notas de la aplicación del analizador Konelab para determinar la estabilidad de los reactivos una vez abiertos y puestos en el analizador. El reactivo PCR de alta sensibilidad reconstituido y el tampón PCR de alta sensibilidad abierto pueden utilizarse durante 8 semanas siempre que se almacenen cerrados a 2…8 °C cuando no se utilicen. Es recomendable extraer los viales de reactivo del analizador y guardarlos cerrados en el refrigerador cuando no se utilicen. Una vez abierto, el calibrador PCR de alta sensibilidad es estable durante 8 semanas a 2…8 °C. Guarde el calibrador herméticamente cerrado cuando no vaya a utilizarlo. No utilice viales que presenten indicios visibles de desarrollo microbiano. RECOGIDA DE MUESTRAS Tipo de muestra Puede utilizarse suero, plasma con heparina y plasma con EDTA. Las muestras túrbidas deben centrifugarse (por ejemplo, 2000xg durante 15 min.). Precauciones Las muestras de origen humano deben manejarse y desecharse como si se tratase de material potencialmente infeccioso. Almacenamiento (2) La muestra puede almacenarse durante 8 días a 2...8 °C o durante 3 meses a -20 °C ( si se congela una sola vez). PRINCIPIO DEL PROCEDIMIENTO Consulte el procedimiento automático para el analizador Konelab en el manual del analizador y las notas de la aplicación. No puede garantizarse la fiabilidad de ninguna aplicación no aprobada por Thermo Fisher Scientific Oy, por lo que deberá evaluarla el usuario. Materiales suministrados Los reactivos descritos anteriormente y un vial vacío con código de barras para analizadores Konelab. Materiales requeridos pero no suministrados El control que se indica a continuación. Calibración Utilice el cabibrador PCR de alta sensibilidad incluido en el kit según las instrucciones provistas para el analizador Konelab. No deje que el calibrador permanezca en el analizador más de una hora. Trazabilidad: El valor de PCR se ha asignado utilizando un preparado CRM 470 de IFCC como referencia primaria. Control de calidad Utilice una muestra de control de calidad al menos una vez al día, después de cada calibración y siempre que emplee un frasco de reactivo nuevo. Controles disponibles: CRP High Sensitivity Control, código 981852 Los intervalos y límites de control deben adaptarse a los requisitos de cada laboratorio. El resultado de la muestra de control de calidad debe estar dentro de los límites establecidos por el laboratorio. CÁLCULO DE RESULTADOS El analizador Konelab calcula los resultados automáticamente por medio de una curva de calibración. El analizador prepara automáticamente una serie de dilución de un stock de calibrador según los parámetros preestablecidos. La curva de calibración se genera a partir de los calibradores medidos utilizando un ajuste spline. CRP HS 1.6 1.4 1.2 1 0.8 0.6 0.4 0 5 10 15 20 Concentración (mg/l) Konelab 20/30/60. La curva de calibración depende del lote. LIMITACIONES DEL PROCEDIMIENTO (6) Interferencias Criterio: recuperación entre + 10% de los valores iniciales. Bilirrubina: sin interferencias encontradas hasta 1000 μmol/l (58.5 mg/dl). Hemólisis: sin interferencias encontradas hasta 10 g/l de hemoglobina en el hemolizado. Lipemia: sin interferencias encontradas hasta 4 g/l de Intralipid® (marca comercial de Fresenius Kabi AB). Hay escasa correlación entre la turbidez y la concentración de triglicéridos. Si desea obtener información sobre la interferencia de otras sustancias, consulte la referencia 3. En el ensayo pueden interferir anticuerpos heterofílicos, inmunoglobulinas monoclonales (o parte de ellos) y factores reumatoides. Deben evaluarse cuidadosamente los resultados de pacientes sospechosos de tener tales anticuerpos (4). Exceso de antígeno No se ha hallado exceso de antígeno en concentraciones de PCR de hasta 300 mg/l. VALORES PREVISTOS (5) Por debajo de 3 mg/l Los valores mencionados sirven sólo de referencia. Se recomienda que cada laboratorio verifique este rango o derive un intervalo de referencia para la población a la que atiende. Cuando se utilice PCR para evaluar el riesgo de cardiopatía coronaria, las mediciones deben compararse con valores anteriores. RANGO DE MEDIDA 0.25 – 10 mg/l Rango de medida aumentado después de la segunda dilución: 0.25 – 40 mg/l CARACTERÍSTICAS DEL RESULTADO (6) Límite de determinación El límite de determinación es la concentración más baja que puede medirse cuantitativamente: 0.25 mg/l Imprecisión (unidad de resultado mg/l) Media 0.80 mg/l Media 2.41 mg/l SD CV (%) SD CV (%) Intraserie 3.2 0.02 0.8 0.02 Interdiario 0.03 3.7 0.04 1.8 Total 0.04 5.9 0.05 2.3 Media 7.02 mg/l SD CV (%) Intraserie 0.6 0.04 Interdiario 0.08 1.2 Total 0.15 2.2 Se realizó un estudio de la precisión según las directrices del documento EP5-A del NCCLS, utilizando el analizador Konelab 30 durante al menos 20 días, siendo el número mínimo de medidas n = 80. Comparación de métodos Se realizó un estudio comparativo según las directrices facilitadas en el documento EP9-A del NCCLS usando como referencia un método comercialmente disponible. Regresión lineal (unidad de resultado mg/l): y = 1.01 x – 0.29 r = 0.983 n = 87 Las concentraciones de las muestras oscilaron entre 0.29 y 10 mg/l. Los resultados obtenidos en cada laboratorio pueden diferir de los datos de resultados presentados. BIBLIOGRAFÍA 1) Burtis CA and Ashwood ER (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, WB Saunders Company, Philadelphia, 2001, pp. 332-333 and 688. 2) Guder WG, Narayanan S, Wisser H, Zawta B. List of Analytes; Preanalytical variables. Brochure in: Samples: From Patient to the Laboratory. GIT Verlag GmbH, Darmstadt, 1996. 3) Young DS, Effects of Drugs on Clinical Laboratory Tests, Volume One, Listing by Test, Fifth Edition, AACC Press, Washington, D.C., 3-233 – 3-234. 4) Selby C. Interference in immunoassay. Ann Clin Biochem 1999;36:704-721. 5) Pearson TA et al. Markers of Inflammation and Cardiovascular Disease Application to Clinical and Public Health Practice, A Statement for Healthcare Professionals From the Centers for Disease Control and the American Heart Association., Circulation 2003;107:499-511. 6) Datos de archivo de Thermo Fisher Scientific Oy. © 2007 Thermo Fisher Scientific Inc. All rights reserved. Preparación El tampón PCR de alta sensibilidad, el diluyente para muestras y el calibrador PCR de alta sensibilidad están listos para su uso. Reconstitución del reactivo PCR de alta sensibilidad: Agregue exactamente 4 ml de agua desionizada. Cierre el vial, mezcle suavemente y deje reposar durante al menos 30 minutos a temperatura ambiente. Vuelva a mezclar el contenido antes de utilizarlo. No lo agite. Transfiera el reactivo a un vial vacío con código de barras para analizadores Konelab. Nota 1: Utilice siempre tampón PCR de alta sensibilidad, reactivo PCR de alta sensibilidad y calibrador PCR de alta sensibilidad con el mismo número de lote de reactivo, ya que los componentes se han agrupado por lotes. Nota 2: Mezcle suavemente el reactivo PCR de alta sensibilidad antes de insertarlo en el analizador Konelab. No lo agite. Nota 3: Compruebe que no haya burbujas en el cuello de la botella ni en la superficie del reactivo cuando inserte los viales o recipientes en el analizador Konelab. Curva de calibración (ejemplo) Template: D01162_4 Precauciones Sólo para uso en diagnósticos in vitro. Adopte las medidas de precaución habituales para manipular reactivos de laboratorio. El calibrador contiene material de origen humano sometido a pruebas que han determinado la ausencia de AgsHB y anticuerpos de VIH 1/2- y VHC. Sin embargo, es preciso manipular el producto con el mismo cuidado que las muestras del paciente. El reactivo PCR de alta sensibilidad liofilizado contiene azida sódica como conservante (1 - <2.5%). Evite inhalar polvo y no lo ingiera. Evite el contacto con la piel y las membranas mucosas. Si se produjera contacto, lávese con agua abundante. La azida sódica puede reaccionar con las cañerías de plomo y cobre de los sistemas de desagüe y formar azidas metálicas altamente explosivas. Antes de desechar el reactivo, enjuáguelo con agua abundante. Consulte la hoja aparte que incluye Frases sobre Riesgos y Seguridad (T:R;23/24/25 - 52/53 S;22-28-45). Una vez reconstituido, el reactivo PCR de alta sensibilidad líquido contiene azida sódica al < 0.1%, concentración que no se considera perjudicial. Reacción (A) Page 8 Page 9 D02418_04_insert_CRP HS_MU FABRICANTE Thermo Fisher Scientific Oy Clinical Diagnostics Finland Ratastie 2, P.O. Box 100, FI-01621 Vantaa Tel. +358 9 329 100, Fax +358 9 3291 0300 www.thermo.com/konelab Säilitamine ja stabiilsus Avamata viaalides püsivad reaktiivid temperatuuril 2…8 °C stabiilsena sildile trükitud aegumistähtajani. Teave laetud reaktiivide stabiilsuse kohta on esitatud Konelab analüsaatori tehnilistes märkustes. Valmis segatud CRP HS reaktiiv ja avatud CRP HS puhver tuleb ära kasutada 8 nädala jooksul, kui neid hoida temperatuuril 2…8 °C. Reaktiiviviaalid on siivitav analüsaatorist välja võtta ja suletuna külmikus hoida. Avatud CRP HS kalibraator püsib temperatuuril 2…8 °C stabiilsena 8 nädalat. Mittekasutatavat kaliibrimisreaktiivi tuleb hoida tihedalt suletuna. Mitte kasutada mikroobide kasvule viitavate märkidega viaale. Fecha de revisión (aaaa-mm-dd) 2007-09-13 Cambios desde la versión anterior Nombre de empresa actualizado. PROOVIDE VÕTMINE Proovi tüüp Kasutada võib seerumit, hepariniseeritud plasmat või EDTA plasmat. Hägused proovid tuleb tsentrifuugida (nt 2000xg 15 min jooksul) Hoiatused Inimpäritolu proovid nakkusallikateks. ET Konelab™ / T Seeria CRP High Sensitivity 1 x 4 ml PAKENDI INFOLEHT ON KOOSTATUD KASUTAMISEKS VÄLJASPOOL USA-D. KONELAB SYSTEMSI VIITED KEHTIVAD ÜHTLASI T SEERIA KOHTA. SIHTOTSTARVE Madala kontsentratsiooniga CRP kvantitatiivseks in vitro määramiseks inimese vereseerumis või -plasmas Konelabi analüsaatorite abil. Katsetulemuste tõlgendamisel tuleb alati arvesse võtta kliinilist tausta. KOKKUVÕTE (1, 5) C-reaktiivne proteiin (CRP) on seerumi tavapärane koostisosa, mis terve organismi puhul on väga madala kontsentratsiooniga. Invasiivsed bakteriinfektsioonid ja ekstensiivsed koekahjustused põhjustavad CRP taseme tõusu. CRP on üks tundlikemaid aktuutse etapi proteiine; ainehulk plasmas tõuseb enamasti olulisel määral müokardi infarkti, trauma, infektsiooni, inflammatsiooni, operatsioonide ja neoplastilise proliferatsiooni korral. Tõus algab 6 kuni 12 tunni vältel ja ulatub 2000 korrani normaaltasemest. CRP kliinilisest määramisest on kasu läbivaatuseks orgaaniliste haiguste puhul, põletikuliste haiguste intensiivsuse hindamiseks, süsteemsete lupus erythematosuste (SLE) korduvate infektsioonide avastamiseks, leukeemia puhul ning pärast kirurgilist lõikust ja neonataalse septitseemia ja meningiidi jälgimisel, kui proovide kogumine bakterioloogiliste uuringute jaoks võib raskendatud olla. CRP mõõtmine tundlike CRP analüüside abil võib suurendada muude markerite predikatiivset väärtust kardiovaskulaarsete ja perifeersete vaskulaarhaiguste riski hindamisel. MEETODI PÕHIMÕTE Katse põhineb immuunerituse mõõtmisel lainepikkusel 540 nm. Puhvriga segatud proovidele lisatakse mitte-inimpäritolu CRP-ga mikroosakesi. Imuuneritusesest tingitud neelduvuse tõus registreeritakse reaktsiooni lõppemise hetkel. Absorbeerumise muutus on proportsionaalne antigeeni (CRP) kogusega lahuses. kõrvaldamisel lugeda võimalikeks Kaasasolevad materjalid Eelnevalt kirjeldatud reaktiivid ja vöötkoodiga varustatud tühi viaal Konelabi analüsaatori jaoks. Vajalikud materjalid, mida kaasas pole Allpool kirjeldatud kontroll. Kaliibrimine Kasutatakse CRP HS kalibraatorit, järgides Konelab analüsaatori Kaliibrimisreaktiivi pole soovitav hoida analüsaatoris üle ühe tunni. juhendit Jälgitavus: CRP väärtus on leitud, kasutades esmase etalonina IFCC valmistist CRM 470. Kvaliteedikontroll Kvaliteedikontrolliproove kasutatakse vähemalt kord päevas, pärast iga kaliibrimist ja iga kord, kui kasutatakse uut reaktiivipudelit. Saadaval on järgmised kontrollproovid: Kõrgtundlik CRP kontrollreatkiiv, kood 981852 Kontrollide vahemikud ja piirid tuleb kohaldada individuaalsete laborite vajadustele. Kvaliteedi kontrollproovide abil saadud tulemused peavad jääma labori poolt eelnevalt kehtestatud piiridesse. TULEMUSTE ARVUTAMINE Tulemused arvutab Konelab analüsaator kaliibrimiskõvera järgi. Analüsaator valmistab põhikalibraatorist automaatselt lahjendusseeria vastavalt etteantud parameetritele. Kaliibrimiskõver koostatakse mõõdetud kalibaatorite järgi, kasutades lekaalpaigutust. Kaliibrimiskõver (näide) 1 x 4 ml 2 x 50 ml 1 x 60 ml 2 x 1 ml CRP HS 1.6 CRP-ga mikroosakesed NaN3 1 - < 2.5% Tris 0.2 mol/l NaN3 < 0.1 % Proovilahjendaja PBS 0.01 mol/l NaN3 < 0.1 % CRP HS kalibraator inipäritolu NaN3 < 0.1 % Kalibraatori CRP kontsentratsioon on prinditud eraldi väärtuste lehele. CRP HS puhver Hoiatused Kasutamiseks ainult in vitro diagnostikas. Kõigi laborireaktiivide käsitsemisel tuleb rakendada tavapäraseid ettevaatusabinõusid. Kaliibraator sisaldab inimpäritoluga materjali, mis on kontrollimisel osutunud HBsAg ning HIV 1/2- ja HCV antikehade suhtes negatiivseks. Siiski tuleb seda käsitseda sama ettevaatlikult, nagu patsiendilt pärinevat proovi. Lüofoolne CRP HS reaktiiv sisaldab säilitusainena naatriumasiidi (1 - < 2.5%). Vältida tolmu sissehingamist. Mitte neelata. Vältida kokkupuudet naha ja limaskestadega. Kokkupuute korral pesta rohke veega. Naatriumasiid võib reageerida torustikus leiduva plii ja vasega ning moodustada äärmiselt plahvatusohtlikke metalliasiide. Pärast kasutamist lahjendada suure koguse veega. Vt R- ja S-lausete kaarti (T:R;23/24/25 - 52/53 S;22-2845). Valmis segatuna sisaldab vedel CRP HS reaktiiv naatriumasiidi < 0.1%, mida ei loeta ohtlikuks kontsentratsiooniks. Ettevalmistamine CRP HS puhver, proovi lahjendus ja CRP HS kalibraator on valimis kasutamiseks. CRP HS rekaktiivi rekonstrueerimine: Lisada täpselt 4 ml deioniseeritud vett. Sulgeda viaal, segada kergelt ja lasta toatemperatuuril seista 30 min. Enne kasutamist sisu veelkord segada. Mitte loksutada. Valada reaktiiv Konelab analüsaatori tühja vöötkoodiga viaali. Märkus 1: Kasutada alati sama reaktiivi partiinumbriga CRP HS puhvrit, CRP HS reaktiivi ja CRP HS kalibraatorit, sest komponendid on partii piires ühtlustatud. Märkus 2: Segada CRP HS reaktiivi kergelt enne Konelabi analüsaatorisse sisestamist. Mitte loksutada. Märkus 3: Enne reaktiiviviaalide või -nõude viimist Konelab analüsaatorisse tuleb kontrollida, et pudelikaelas ega reaktiivi pinnal ei oleks mulle. 1.4 1.2 1 0.8 0.6 0.4 0 5 10 15 20 Concentration (mg/l) Konelab 20/30/60. Kalibreerimiskõver oleneb partiist. PROTSEDUURI PIIRANGUD (6) Segavad mõjud Tingimus: Saagis algväärtuse ± 10% suhtes. Bilirubiin: Kontsentratsioonini kuni 1000 μmol/l (58.5 mg/dl) segavat mõju ei täheldatud. Hemolüüs: hemoglobiini kontsentratsioonini kuni 10 g/l segavat mõju ei täheldatud. Lipeemia: Intralipid® (Fresenius Kabi AB kaubamärk) sisalduseni kuni 4 g/l segavat mõju ei täheldatud. Korrelatsioon hägususe ja triglütseriidide sisalduse vahel oli väike. Muid segavaid aineid on käsitletud viites 3. Analüüsi võivad mõjutada heterofiilsed antikehad, monokloonsed immunoglobuliinid (või nende osad) ja reumatoidfaktorid. Nimetatud antikehadega patsientide puhul saadud tulemusi tuleb hoolikalt hinnata (4). Antigeeni liia määramine CRP kontsentratsioonini kuni 300 mg/l antigeenide liiga täheldatud ei ole. © 2007 Thermo Fisher Scientific Inc. All rights reserved. Mitte-inimpäritolu Template: D01162_4 Kontsentratsioonid CRP HS reaktiiv (lambalt) (lüofiliseeritud). ja KATSEPROTSEDUUR Teave automaatse katseprotseduuri kasutamise kohta Konelab analüsaatoril on esitatud juhendis ja tehnilistes märkustes. Thermo Fisher Scientific Oy poolt valideerimata rakendusviiside sooritusnäitajaid tagada ei saa, seetõttu peab neid hindama lõppkasutaja. Response (A) TEAVE REAKTIIVIDE KOHTA CRP HS reaktiiv CRP HS puhver Proovilahjendaja CRP HS kalibraator käsitsemisel Säilitamine (2) Proovi võib temperatuuril 2…8 °C säilitada kuni 8 päeva, temperatuuril – 20 °C kuni 3 kuud (ühekordsel külmutamisel). Kõrgtundlik CRP 981798 tuleb Page 10 D02418_04_insert_CRP HS_MU OODATAVAD TULEMUSED (5) Alla 3 mg/l Toodud väärtused on mõeldud suunitluslikena. Kõigil laboritel on soovitatav seda vahemikku kontrollida, või tuletada ise etalonvahemik teenindatava populatsiooni seas. CRP kasutamisel koronaari haiguste diagnoosimisel tuleb tulemusi võrrelda eelnevalt saadud väärtustega. AZ ELJÁRÁS ALAPELVE A módszer immunprecipitáció 540 nm-en történő mérésén alapul. Anti-humán CRP-vel bevont mikrorészecskék kerülnek hozzáadásra pufferelt mintákhoz. Az immunprecipitáció okozta abszorbancia-növekedés kerül rögzítésre a reakció végpontján. Az abszorbarnciában bekövetkezett változás arányos az oldatban lévő antigén (CRP) mennyiségével. MÕÕTEPIIRKOND 0.25 – 10 mg/l Pikendatud mõõtepiirkond pärast sekundaarset lahjendust: 0.25 -40 mg/l REAGENSADATOK CRP HS reagens CRP HS puffer Mintadiluens CRP HS kalibrátor SOORITUSNÄITAJAD (6) Avastamispiir Avastamispiiri määrab madalaim kvantitatiivselt mõõdetav kontsentratsioon: 0.25 mg/l Ebatäpsus (tulemuse ühik (mg/l) Keskmine 0.80 mg/l Keskmine 2.41 mg/l SD CV% SD CV% Katseseeriasisene 0.02 3.2 0.02 0.8 Eri päevad 0.03 3.7 0,04 1.8 Summarne 0.04 5.9 0.05 2.3 Keskmine 7.02 mg/l SD CV% Katseseeriasisene 0.04 0.6 Eri päevad 0.08 1.2 Summarne 0.15 2.2 Vastavalt NCCLS dokumendi EP5-A suunistele viidi Konelab 30 peal vähemalt 20 päeva jooksul läbi täpsusuuring, mõõtmiste arv uuringus oli minimaalselt n = 80. Meetodite võrdlus NCCLS dokumendi EP9-A suuniseid järgides viidi läbi võrdlev uuring, kasutades etalonina müügilolevat meetodit. Lineaarne regressioon (tulemuse ühik (mg/l): y = 1.01 x –0.29 r = 0.983 n = 87 Proovikontsentratsioonid jäid vahemikku 0.29 ja 10 mg/l. Individuaalsetes laborites saadavad tulemused võivad erineda esitatud sooritusnäitajatest. KIRJANDUS 1) Burtis CA and Ashwood ER (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, WB Saunders Company, Philadelphia, 2001, pp. 332-333 and 688. 2) Guder WG, Narayanan S, Wisser H, Zawta B. List of Analytes; Preanalytical variables. Brochure in: Samples: From Patient to the Laboratory. GIT Verlag GmbH, Darmstadt, 1996. 3) Young DS, Effects of Drugs on Clinical Laboratory Tests, Volume One, Listing by Test, Fifth Edition, AACC Press, Washington, D.C., 3-233 – 3-234. 4) Selby C. Interference in immunoassay. Ann Clin Biochem 1999;36:704-721. 5) Pearson TA et al. Markers of Inflammation and Cardiovascular Disease Application to Clinical and Public Health Practice, A Statement for Healthcare Professionals From the Centers for Disease Control and the American Heart Association., Circulation 2003;107:499-511. 6) Thermo Fisher Scientific Oy andmefail. TOOTJA Thermo Fisher Scientific Oy Clinical Diagnostics Finland Ratastie 2, P.O. Box 100, FI-01621 Vantaa Tel. +358 9 329 100, Faks +358 9 3291 0300 www.thermo.com/konelab Teksti läbivaatamise kuupäev (aaaa-kk-pp) 2007-09-13 Muudatused võrreldes eelmise versiooniga Ettevõtte nimi uuendatud. 1 x 4 ml 2 x 50 ml 1 x 60 ml 2 x 1 ml Koncentrációk CRP HS reagens Anti-humán CRP-vel (birka) mikrorészecskék (liofilizálva) NaN3 1 - < 2.5% CRP HS puffer Tris 0.2 mol/l NaN3 < 0.1 % Mintadiluens PBS 0.01 mol/l < 0.1 % NaN3 CRP HS kalibrátor Humán NaN3 < 0.1 % A kalibrátor CRP koncentrációja megtalálható egy külön értéklapon. bevont Óvintézkedések Kizárólag in vitro diagnosztikus használatra. Tartsa be a laboratóriumi reagensek kezelésére vonatkozó szokásos előírásokat. A kalibrátor emberi eredetű anyagokat tartalmaz, amelyek a vizsgálatok során negatívnak bizonyultak HBsAg-re, valamint HIV 1/2 és HCV antitestekre nézve. Ennek ellenére a terméket ugyanolyan gondossággal kell kezelni, mint a betegből származó mintákat. A liofilizált CRP HS reagens tartósítószerként nátrium-azidot tartalmaz (1 - <2.5%). Kerülje a por belégzését! Ne nyelje le. Kerülje érintkezését a bőrrel és a nyálkahártyákkal. Érintkezés esetén bő vízzel öblítse le. A nátrium-azid a csatorna-hálózatban lévő ólommal, illetve rézzel reakcióba lépve rendkívül robbanékony fém-azidokat képezhet. Kidobáskor öblítse le nagy mennyiségű vízzel. Lásd a kockázati és biztonsági megfontolások című külön adatlapot (T:R;23/24/25 - 52/53 S;22-28-). A feloldást követően a folyékony CRP HS reagens tartósítószerként nátrium-azidot tartalmaz < 0.1%-ban, mely nem tekinthető ártalmas koncentrációnak. Előkészítés A CRP HS puffer, a mintadiluens, valamint a CRP HS kalibrátor használatra kész. A CRP HS reagens feloldása: Adjon hozzá pontosan 4 ml desztillált vizet. Zárja le az üveget, óvatosan keverje fel, majd hagyja állni szobahőmérsékleten legalább 30 percig. Használat előtt keverje fel ismét az üveg tartalmát. Ne rázza fel. Tegye át a reagenst egy üres, vonalkóddal ellátott üvegbe a Konelab analizátoron történő használathoz. 1. megjegyzés: Mindig ugyanazon reagens-tételszámmal rendelkező CRP HS puffert, CRP HS reagenst és CRP HS kalibrátort használjon, mert az egy tételszámhoz tartozó termékek összetevői megfelelnek egymásnak. 2. megjegyzés: Óvatosan keverje fel a CRP HS reagenst, mielőtt betenné a Konelab analizátorba. Ne rázza fel. 3. megjegyzés: Ellenőrizze, hogy nincs-e buborék az üveg nyakánál vagy a reagens felszínén, amikor reagenst tartalmazó üvegeket, illetve edényeket tesz a Konelab analizátorba. Tárolás és stabilitás Felbontatlan üvegben tárolt reagensek 2...8 °C között a címkén feltüntetett lejárati időpontig stabilak. A reagensek elemzőkészüléken belüli stabilitásához olvassa el a Konelab analizátorhoz tartozó Alkalmazási tudnivalókat. A feloldott CRP HS reagens és a felnyitott CRP HS puffer használaton kívül, zárt állapotban, 2…8 °C között tárolva 8 hétig marad stabil. Használaton kívül ajánlott a reagensüvegeket az analizátoron kívül, zárt állapotban, hűtőszekrényben tartani. A felnyitott CRP HS kalibrátor 2…8 °C hőmérsékleten 8 hétig stabil. Szorosan lezárva tárolja a kalibráló anyagot, amennyiben nem használja. Ne használjon olyan üveget, amelyen mikróbanövekedésre utaló szemmel látható jeleket lát. MINTAVÉTEL Minta típusa A vizsgálathoz szérum, heparinos plazma és EDTA-s plazma használható. A zavaros mintát centrifugálni kell (pl. 2000xg beállítással 15 percig) Konelab™ / T Sorozat CRP High Sensitivity Magas szenzitivitású CRP 981798 1 x 4 ml EZ A TÁJÉKOZTATÓ AZ EGYESÜLT ÁLLAMOKON KÍVÜLI HASZNÁLATRA VONATKOZIK. A KONELAB RENDSZEREKRE TETT MINDEN UTALÁS A T SOROZATRA IS VONATKOZIK. RENDELTETÉS Emberi szérum és plazma alacsony CRP koncentrációjának in vitro kvantitatív meghatározásához Konelab analizátorokban. Minden vizsgálati eredményt a klinikai képpel összefüggésben kell értékelni. ÖSSZEGZÉS (1, 5) A C-reaktív protein (CRP) igen alacsony koncentrációban az egészséges egyén szérumának normális összetevője. Az invazív bakteriális fertőzés és a kiterjedt szövetkárosodás emelkedett CRP szintet eredményez. A CRP az egyik legérzékenyebb akutfázis fehérje: szintje általában drasztikusan megemelkedik miokardiális infarktust, traumát, fertőzést, gyulladást, műtétet, illetve neoplasztikus proliferációt követően. Az emelkedés 6-12 órán belül kezdődik, és a szint elérheti a normális érték 2000-szeresét. A CRP meghatározása klinikailag hasznos organikus betegségek szűrésében, a gyulladásos betegség aktivitásának felmérésében, szisztémás lupus erythematosushoz (SLE), leukémiához társuló, illetve műtét után kialakuló fertőzések kimutatásában, valamint az újszülöttkori szepszis és meningitisz kezelésében, amikor nehézkes lehet a bakteriológiai mintavétel. A CRP szenzitív vizsgálatokkal való mérése a kardiovaszkuláris és perifériás érbetegségek rizikójának felmérésekor kiegészítheti más markerek értékét. Tárolás (2) A minta 2…8 °C -on 8 napig, –20 °C -on pedig (egyszeri lefagyasztással) 3 hónapig tárolható. A VIZSGÁLATI ELJÁRÁS Automatizált eljárás kivitelezéséhez olvassa el a Konelab analizátorhoz tartozó Használati kézikönyvet és Alkalmazási tudnivalókat. Nem garantálható semmilyen olyan alkalmazás eredménye, amelyet a Thermo Fisher Scientific Oy nem hagyott jóvá, ezért ezeket a felhasználónak kell értékelnie. Szolgáltatott anyagok A fent leírt a reagensek, valamint egy üres, vonalkóddal ellátott üveg a Konelab analizátoron történő használathoz. Szükséges, de nem szolgáltatott anyagok Az alább leírt kontrollok. Kalibrálás A készletben lévő CRP HS kalibrátort a Konelab analizátorhoz tartozó tájékoztató szerint használja fel. Legfeljebb egy órára hagyja az analizátorban a kalibrátort. Nyomonkövethetőség: A CRP értékének meghatározásához elsődleges referenciaként az IFCC CRM 470 típusú preparátuma került felhasználásra. Minőségellenőrzés Használjon minőségellenőrző mintát naponta legalább egyszer, minden kalibrálás után, illetve minden új üveg reagens felnyitásakor. Rendelkezésre álló kontrollminták: CRP magas szenzitivitású kontroll, kód 981852 Az ellenőrzési intervallumokat és határértékeket az aktuális laboratóriumi követelményekhez kell igazítani. A minőségellenőrző minta eredményének a laboratórium által előre beállított határértékek közé kell esnie. Template: D01162_4 HU © 2007 Thermo Fisher Scientific Inc. All rights reserved. Óvintézkedések Az emberi mintákat úgy kell kezelni és kidobni, mintha azok fertőzőek lennének. Page 11 D02418_04_insert_CRP HS_MU AZ EREDMÉNYEK KISZÁMÍTÁSA Az eredményeket a Konelab analizátor automatikusan kiszámítja egy kalibrációs görbe segítségével. Az analizátor az előre beállított paraméterek alapján automatikusan hígítási sort készít a szabvány kalibrátor felhasználásával. A kalibrációs görbe kialakítása a mért kalibrátorokból történik harmadrendű görbeillesztési módszerrel (spline fit). Kalibrációs görbe (példa) A Thermo Fisher Scientific Oy-nál nyilvántartott adatok. GYÁRTÓ Thermo Fisher Scientific Oy Clinical Diagnostics Finland Ratastie 2, P.O. Box 100, FI-01621 Vantaa Tel. +358 9 329 100, Fax +358 9 3291 0300 www.thermo.com/konelab Ellenőrzés időpontja (éééé-hh-nn) 2007-09-13 CRP HS Változtatások az előző változathoz képest A cég neve frissítésre került. 1.6 1.4 1.2 1 IT Konelab™ / Serie T CRP High Sensitivity 0.8 0.6 CRP ad alta sensibilità 0 5 10 15 20 Concentration (mg/l) Konelab 20/30/60. A kalibrációs görbe cikkszámtól függ. AZ ELJÁRÁS KORLÁTAI (6) Interferencia Kritérium: Az értékek a kezdeti értékek ± 10%-án belül maradnak. Bilirubin: Nem észleltek interferenciát 1000 μmol/l (58.5 mg/dl) értékig. Hemolízis: Nem észleltek interferenciát a hemolizátumban 10 g/l hemoglobinszintig. Lipaemia: Nem észleltek interferenciát Intralipid® (a Fresenius Kabi AB védjegye) esetén 4 g/l értékig. Gyenge összefüggés van a zavarosság és a trigliceridkoncentráció között. Egyéb interferenciát okozó anyagok tekintetében lásd a 3. referenciát. Heterofil antitestek, monoklonális immunglobulinok (vagy azok részei), valamint rheumatoid faktor jelenléte zavarhatják a vizsgálatot. Amennyiben előfordulhat a betegnél ilyen antitestek jelenléte, akkor az eredményeket óvatosan kell értékelni (4). Antigéntöbblet Nem észlelhető antigéntöbblet 300 mg/l CRP koncentrációig. VÁRHATÓ ÉRTÉKEK (5) 3 mg/l alatt A feltüntetett értékek kizárólag tájékoztató jellegűek. Minden laboratórium esetében ajánlott a normál tartomány megerősítése, illetve a kiszolgált populációra jellemző referenciaintervallum kialakítása. A szívkoszorúér-betegség kockázatának felmérésére való használatakor a CRP mért értékét a korábbi értékekhez kell hasonlítani. MÉRÉSI TARTOMÁNY 0.25 – 10 mg/l Kiterjesztett mérési tartomány másodlagos hígítás után: 0.25 – 40 mg/l TELJESÍTMÉNYI JELLEMZŐK (6) Meghatározási határérték A meghatározási határérték az a legalacsonyabb koncentráció, amely még kvantitatíven mérhető: 0.25 mg/l Pontatlanság (az eredmény mértékegysége mg/l) Középérték 0.80 mg/l Középérték 2.41 mg/l SD CV% SD CV% Mérés közben 0.02 3.2 0.02 0.8 Napok közt 0.03 3.7 0.04 1.8 Teljes 0.04 5.9 0.05 2.3 Középérték 7.02 mg/l SD CV% Mérés közben 0.04 0.6 Napok közt 0.08 1.2 Teljes 0.15 2.2 Az NCCLS EP5-A dokumentumának előírásai alapján pontossági vizsgálat történt legalább 20 napon keresztül a Konelab 30 analizátorral, minimum 80-as mérésszámmal (n=80). Módszerösszehasonlítás Összehasonlítási tanulmány készült az NCCLS EP9-A dokumentumának előírásai alapján, referenciaként egy másik forgalomban lévő módszert felhasználásával. Lineáris regresszió (az eredmény mértékegysége mg/l): y = 1.01 x – 0.29 r = 0.983 n = 87 A mintakoncentrációk 0.29 és 10 mg/l között voltak. Az egyes laboratóriumokban nyert értékek különbözhetnek a megadott teljesítmény adatoktól. BIBLIOGRÁFIA 1) Burtis CA and Ashwood ER (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, WB Saunders Company, Philadelphia, 2001, pp. 332-333 and 688. 2) Guder WG, Narayanan S, Wisser H, Zawta B. List of Analytes; Preanalytical variables. Brochure in: Samples: From Patient to the Laboratory. GIT Verlag GmbH, Darmstadt, 1996. 3) Young DS, Effects of Drugs on Clinical Laboratory Tests, Volume One, Listing by Test, Fifth Edition, AACC Press, Washington, D.C., 3-233 – 3-234. 4) Selby C. Interference in immunoassay. Ann Clin Biochem 1999;36:704-721. 5) Pearson TA et al. Markers of Inflammation and Cardiovascular Disease Application to Clinical and Public Health Practice, A Statement for Healthcare Professionals From the Centers for Disease Control and the American Heart Association., Circulation 2003;107:499-511. 981798 1 x 4 ml IL PRESENTE INSERTO NELL’IMBALLO PUO’ ESSERE APPLICATO AL DI FUORI DEGLI STATI UNITI. EVENTUALI RIFERIMENTI A KONELAB SYSTEMS SI RIFERISCONO ANCHE ALLA SERIE T. USO CONSIGLIATO Prodotto impiegato per la determinazione quantitativa in vitro di basse concentrazioni di proteina C-reattiva (CRP) nel siero e nel plasma umani con analizzatori Konelab. I risultati del test devono essere sempre interpretati in riferimento al contesto clinico specifico. SOMMARIO (1, 5) La proteina C-reattiva (CRP) è un normale componente del siero presente in concentrazioni molto basse negli individui sani. Infezioni batteriche invasive e danno tessutale esteso possono determinare un aumento dei livelli di CRP. La CRP è una delle proteine più sensibili, rilevabile solamente durante la fase attiva di alcune patologie acute; i valori plasmatici aumentano raggiungendo livelli molto elevati, di norma, in seguito a infarto miocardico, trauma, infezione, infiammazione, intervento chirurgico o proliferazione neoplastica. L’aumento si verifica entro 6 - 12 ore e i livelli possono raggiungere anche un valore 2000 volte superiore la norma. La determinazione della CRP è clinicamente utile nello screening di patologie organiche, nella valutazione dell’attività della patologia infiammatoria, nel rilevamento di infezioni intercorrenti nel lupus eritematoso sistemico (LES), nella leucemia, o dopo un intervento chirurgico e il trattamento di una setticemia e meningite neonatale, quando può rivelarsi difficoltosa la raccolta di campioni per indagini batteriologiche. La misurazione della CRP mediante dosaggi sensibili alla CRP può aggiungersi al valore predicativo di altri marker impiegati nella valutazione del rischio di patologia cardiovascolare e vascolare periferica. PRINCIPIO DELLA PROCEDURA Il metodo si basa sulla determinazione a 540 nm dell’immunoprecipitazione. Microparticelle rivestite con CRP anti-umana vengono aggiunte a campioni tamponati. L’aumento di assorbanza prodotto dalla formazione di immunoprecipitati viene registrato nel momento in cui la reazione ha raggiunto il suo punto finale. La differenza di assorbanza della reazione è proporzionale alla quantità di antigene (CRP) in soluzione. INFORMAZIONI SUI REAGENTI Reagente CRP HS Tampone CRP HS Diluente del campione Calibratore CRP HS 1 x 4 ml 2 x 50 ml 1 x 60 ml 2 x 1 ml Concentrazioni Reagente CRP HS Microparticelle rivestite con CRP anti-umana (pecora) (liofilizzata) NaN3 1 - < 2.5% Tampone CRP HS Tris 0.2 mol/l NaN3 < 0.1 % Diluente del campione PBS 0.01 mol/l NaN3 < 0.1 % Calibratore CRP HS Umano NaN3 < 0.1 % La concentrazione di CRP del calibratore è stampata su un foglio dei valori a parte. Precauzioni Solo per uso diagnostico in vitro. Rispettare le normali precauzioni previste per l’utilizzo di tutti i reagenti di laboratorio. Il calibratore contiene materiale di origine umana che è stato testato ed è risultato negativo per la presenza di anticorpi HBsAg, HIV 1/2 e HCV. l prodotto dovrà tuttavia essere usato con la stessa cura dedicata ai campioni dei pazienti. Il reagente CRP HS liofilizzato contiene sodio azide come conservante (1 - < 2.5%). Evitare di inalare la polvere. Non ingerire. Evitare il contatto con la cute e le membrane mucose. In caso di contatto, lavare con acqua abbondante. La sodio azide contenuta può reagire a contatto con tubature e scarichi in piombo e rame formando azidi metalliche altamente esplosive. Al momento dello smaltimento lavare con abbondante acqua. Vedere il foglio a parte per le frasi di rischio e di sicurezza (T:R;23/24/25 - 52/53 S;22-28-45). Dopo la ricostituzione, il reagente CRP HS liquido contiene sodio azide a una concentrazione < 0.1%; tale concentrazione non è classificata come nociva. © 2007 Thermo Fisher Scientific Inc. All rights reserved. 0.4 Template: D01162_4 Response (A) 6) Page 12 D02418_04_insert_CRP HS_MU RACCOLTA DEL CAMPIONE Tipo di campione Possono essere utilizzati siero, plasma eparinizzato e plasma con EDTA. Ogni campione torbido deve essere centrifugato (es. 2000xg per 15 min) Interferenze Criterio: Recupero entro ± 10% del valore iniziale. Bilirubina: Non sono state rilevate interferenze fino a 1000 μmol/l (58.5 mg/dl). Emolisi: Non sono state rilevate interferenze fino a 10 g/l di emoglobina in emolisato. Lipemia: Non sono state riscontrate interferenze significative fino a 4 g/l di Intralipid® (marchio registrato di Fresenius Kabi AB). Vi è una scarsa correlazione tra torbidità e concentrazione di trigliceridi. Per le altre sostanze interferenti, vedere la voce biblio-grafica 3. Anticorpi eterofilici, immunoglobuline monoclonali (o parti di essi) e fattori reumatoidi possono interferire con il saggio. I risultati ottenuti da pazienti per i quali si sospetti la presenza di questi anticorpi devono essere valutati con attenzione (4). Eccesso di antigene Non sono stati osservati eccessi di antigene a una concentrazione di CRP pari a 300 mg/l. VALORI PREVISTI (5) Inferiori a 3 mg/l I valori citati dovranno servire esclusivamente come riferimento. Si raccomanda ad ogni laboratorio di verificare l'applicabilità di questo intervallo alla propria popolazione di pazienti o determinare appositi intervalli di riferimento. Quando si usa la CRP per valutare il rischio di coronaropatia, le misurazioni effettuate devono essere poste a confronto con valori precedenti. INTERVALLO DI MISURAZIONE 0.25 – 10 mg/l Intervallo di misurazione ampliato dopo la seconda diluizione: 0.25 – 40 mg/l CARATTERISTICHE PRESTAZIONALI (6) Precauzioni I campioni umani devono essere maneggiati e smaltiti come campioni potenzialmente infetti. Limite di rilevamento Il limite di rilevamento rappresenta quantitativamente: 0.25 mg/l Conservazione (2) Il campione può essere conservato per 8 giorni a 2...8 °C o per 3 mesi a –20 °C (se congelato una sola volta). Imprecisione (risultati espressi in mg/l) Media 0.80 mg/l Media 2.41 mg/l DS CV% DS CV% Entro la serie 0.02 3.2 0.02 0.8 Fra la serie 0.03 3.7 0.04 1.8 Totale 0.04 5.9 0.05 2.3 Media 7.02 mg/l DS CV% Entro la serie 0.04 0.6 Fra la serie 0.08 1.2 Totale 0.15 2.2 È stato eseguito uno studio di precisione conformemente al protocollo NCCLS, documento EP5-A come linea guida e utilizzando un analizzatore Konelab 30 per 20 giorni, con un numero di misurazioni pari a n = 80. PROCEDURA ANALITICA Per le procedure automatiche, consultare il manuale d’uso e le note applicative dell'analizzatore Konelab impiegato. Tutte le applicazioni non esplicitamente approvate da Thermo Fisher Scientific Oy, non possono essere garantite in termini di prestazioni e dovranno pertanto essere valutate dall'utilizzatore. Materiali inclusi nel kit Reagenti sopra descritti e un vial vuoto recante il codice a barre, idoneo all’uso su analizzatori Konelab. Materiali necessari ma non inclusi nel kit I controlli indicati di seguito. Calibrazione Utilizzare il Calibratore CRP HS incluso nel kit conformemente alle istruzioni dell'analizzatore Konelab in uso. Si raccomanda di lasciare il calibratore sull’analizzatore per un periodo massimo di un’ora. Tracciabilità: Il valore della CRP è stato assegnato mediante procedure riconducibili al riferimento primario IFCC CRM 470. Controllo di qualità Utilizzare il campione del controllo di qualità almeno una volta al giorno e dopo ogni calibrazione e ogni volta che si utilizza un nuovo flacone di reagente. Controlli disponibili: CRP High Sensitivity Control, codice 981852 Gli intervalli e i limiti del controllo devono essere adattati ai requisiti dei singoli laboratori. I risultati del campione del controllo di qualità devono rientrare nei limiti stabiliti a priori dal laboratorio. CALCOLO DEI RISULTATI I risultati vengono calcolati automaticamente dall’analizzatore Konelab come segue utilizzando una curva di calibrazione: L’analizzatore prepara automaticamente una serie di diluizioni da uno stock di calibratori in base ai parametri prestabiliti. La curva di calibrazione viene ottenuta interpolando con una spline i punti misurati. Curva di calibrazione (esempio) CRP HS PRODUTTORE Thermo Fisher Scientific Oy Clinical Diagnostics Finland Ratastie 2, P.O. Box 100, FI-01621 Vantaa Tel. +358 9 329 100, Fax +358 9 3291 0300 www.thermo.com/konelab Modifiche rispetto alla versione precedente Ragione sociale aggiornata. 1.4 1.2 1 0.8 0.6 0.4 0 5 10 15 Concentrazione (m g/l) Konelab 20/30/60. La curva di calibrazione dipende dal lotto. 20 più bassa concentrazione misurabile Metodo di confronto È stato eseguito uno studio comparato secondo il protocollo NCCLS, documento EP9-A come linea guida, utilizzando una metodica disponibile in commercio come riferimento. Regressione lineare (risultato espresso in mg/l): y = 1.01 x – 0.29 r = 0.983 n = 87 Le concentrazioni dei campioni erano comprese tra 0.29 e 10 mg/l. I risultati ottenuti nei singoli laboratori possono differire dai dati sulle prestazioni riportati. BIBLIOGRAFIA 1) Burtis CA and Ashwood ER (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, WB Saunders Company, Philadelphia, 2001, pp. 332-333 and 688. 2) Guder WG, Narayanan S, Wisser H, Zawta B. List of Analytes; Preanalytical variables. Brochure in: Samples: From Patient to the Laboratory. GIT Verlag GmbH, Darmstadt, 1996. 3) Young DS, Effects of Drugs on Clinical Laboratory Tests, Volume One, Listing by Test, Fifth Edition, AACC Press, Washington, D.C., 3-233 – 3-234. 4) Selby C. Interference in immunoassay. Ann Clin Biochem 1999;36:704-721. 5) Pearson TA et al. Markers of Inflammation and Cardiovascular Disease Application to Clinical and Public Health Practice, A Statement for Healthcare Professionals From the Centers for Disease Control and the American Heart Association., Circulation 2003;107:499-511. 6) Dati in archivio presso Thermo Fisher Scientific Oy. Data della revisione (aaaa-mm-gg) 2007-09-13 1.6 Risposta (A) la © 2007 Thermo Fisher Scientific Inc. All rights reserved. Conservazione e stabilità Se conservati a una temperatura di 2...8 °C, i reagenti in vial intatti sono stabili fino alla data di scadenza indicata sull'etichetta. Per la stabilità dei reagenti sullo strumento, fare riferimento alle Note applicative dell'analizzatore Konelab in uso. Il reagente CRP HS ricostituito e il tampone CRP HS aperto, ove rimangano inutilizzati, possono essere utilizzati per 8 settimane se conservati chiusi a 2...8 °C. Si raccomanda di togliere i vial di reagente dall’analizzatore e di conservarli chiusi in frigorifero se inutilizzati. Una volta aperto, il calibratore CRP HS è stabile per 8 settimane se conservato a una temperatura di 2...8 °C. Se inutilizzato, conservare il calibratore ermeticamente chiuso. Non utilizzare vial che presentino evidenti segni di crescita microbica. LIMITI DELLA PROCEDURA (6) Template: D01162_4 Preparazione Tampone CRP HS, diluente del campione e calibratore CRP HS sono pronti all’uso. Ricostituzione del reagente CRP HS: Aggiungere esattamente 4 ml di acqua deionizzata. Chiudere il vial, mescolare delicatamente e lasciare riposare per almeno 30 minuti a temperatura ambiente. Mescolare il contenuto ancora una volta prima dell’uso. Non agitare energicamente. Trasferire il reagente in un vial vuoto recante il codice a barre, idoneo all’uso su analizzatori Konelab. Nota 1: Usare sempre tampone CRP HS, reagente CRP HS e calibratore CRP HS con lo stesso numero di lotto reagente poiché i componenti dei lotti sono stati accoppiati. Nota 2: Mescolare delicatamente il reagente CRP HS prima di collocarlo sull’analizzatore Konelab. Non agitare energicamente. Nota 3: Controllare che non siano presenti bolle sul collo del flacone o sulla superficie del reagente durante l’inserimento di vial o recipienti di reagente nell'analizzatore Konelab. Page 13 D02418_04_insert_CRP HS_MU 981798 1 x 4 ml ŠIOS PAKUOTĖS INFORMACINIS LAPELIS TINKAMAS NAUDOTI UŽ JAV RIBŲ. BET KOKIA NUORODA Į KONELAB SYSTEMS APIMA IR T PRODUKTŲ SERIJĄ. PASKIRTIS Skirtas in vitro kiekybiniam mažų CRP koncentracijų nustatymui žmogaus serume ir plazmoje, naudojant Konelab analizatorius. Visi tyrimų rezultatai turi būti interpretuojami atsižvelgiant į klinikinį kontekstą. SANTRAUKA (1,5) C-reaktyvus proteinas (CRP) yra normalaus serumo sudedamoji dalis, sveikuose individuose randama labai mažomis koncentracijomis. CRP kiekio padidėjimą sąlygoja invazyvinės bakterinės infekcijos arba didelis audinių pažeidimas. CRP yra vienas jautriausių ūmios fazės proteinų; jo kiekis plazmoje paprastai ženkliai pakyla dėl miokardo infarkto, traumų, infekcijų, uždegimų, chirurginių operacijų ar auglių. Koncentracijos didėjimas prasideda 6 - 12 valandų bėgyje, pasiekiama iki 2000 kartų didesnė nei įprastinė koncentracija. CRP koncentracijos nustatymas yra kliniškai naudingas organų ligų, uždegiminių ligų raidos stebėjimui, sisteminės eritematozinės vilkligės (SLE) nustatymui, leukemijos nustatymui ar po chirurginių operacijų bei kūdikių septicemijos ir meningito operacijų, kuomet gali būti sudėtinga paimti mėginį bakteriologiniam tyrimui. PROCEDŪROS PRINCIPAS Šis metodas paremtas imuninio nusodinimo tyrimu prie 540 nm. Į buferinius mėginius pridedama nežmogiškais CRP padengtų mikrodalelių. Absorbcijos padidėjimas, kurį sukelia imuninis nusodinimas yra registruojamas reakcijai pasibaigus. Absorbcijos pokytis yra proporcingas antigeno (CRP) kiekiui tirpale. INFORMACIJA APIE REAGENTUS CRP HS reagentas 1 x 4 ml CRP HS buferis 2 x 50 ml Mėginio tirpalas 1 x 60 ml CRP Plus Kalibratorius, 2 x 1 ml Koncentracijos CRP HS reagentas mikrodalelės, padengtos ne žmogaus CRP (avies) (liofilizuotos) NaN3 1 - < 2.5% CRP HS buferis Tris 0.2 mol/l NaN3 < 0.1 % Mėginio tirpalas PBS 0.01 mol/l NaN3 < 0.1 % CRP HS kalibratorius Žmogaus NaN3 < 0.1 % CRP koncentracija kalibratoriuje yra išspausdinta atskirame Verčių sąraše. Atsargumo priemonės Tik in vitro diagnostiniam naudojimui. Laikykitės įprastų atsargumo priemonių, taikomų dirbant su laboratorijos reagentais. Kalibratoriuje yra žmogiškos kilmės medžiagų, kurios buvo patikrintos ir gauti neigiami atsakymai dėl HBsAg ir dėl ŽIV 1/2- ir HCV antikūnių. Tačiau su šia medžiaga reikia elgtis taip pat atsargiai, kaip ir su paciento mėginiais. Reagento sudėtyje yra natrio azido (1 - < 2.5%), kaip konservanto. Venkite įkvėpti dulkių. Neprarykite. Saugokite, kad nepatektų ant odos ir gleivinės. Patekus gausiai praplaukite vandeniu. Natrio azidas gali reaguoti su švinu ir variu vandentiekio sistemoje ir sudaryti labai sprogius metalų azidus. Išpildami lauk, nuplaukite dideliu kiekiu vandens. Žiūrėkite atskirą Rizikos ir saugos frazių lapą (T:R;23/24/25 - 52/53 S;22-28-45). Atkurto skysto CRP HS reagento sudėtyje yra < 0,1% natrio azido, kas nėra laikoma žalinga koncentracija. Paruošimas CRP HS buferis, mėginio tirpalas ir CRP HS kalibratorius yra parengti naudojimui. CRP HS reagento atkūrimas: Įpilkite lygiai 4 ml distiliuoto arba dejoniuoto vandens. Užkimškite ir leiskite pastovėti bent 30 min. kambario temperatūroje. Prieš naudojimą dar kartą sumaišykite reagentus. Nekratykite. Prieš naudodami Konelab analizatoriuje, perkelkite reagentą į tuščią, brūkšniniu kodu pažymėta buteliuką. 1 Pastaba: visuomet naudokite CRP HS buferį, CRP HS reagentą ir CRP HS kalibratorių su tuo pačiu serijos numeriu, nes serijos viduje komponentai sutampa. 2 Pastaba: švelniai sumaišykite CRP HS reagentą prieš įdėdami jį į Konelab analizatorių. Nekratykite. 3 Pastaba: Dėdami reagento buteliukus ar indus į Konelab analizatorių, patikrinkite ar prie buteliuko kaklelio arba reagento paviršiuje nėra burbuliukų. Saugojimas ir pastovumas Reagentai uždarytuose buteliukuose yra pastovūs prie 2…8 °C iki ant etiketės nurodytos galiojimo termino pabaigos. Su informacija apie prietaise patalpintų reagentų pastovumą susipažinkite savo Konelab analizatoriaus Naudojimo pastabose Atkurtą CRP HS reagentą ir atidarytą CRP HS buferį galima naudoti 8 savaites, nenaudojant jį laikant uždarą 2…8 °C temperatūroje. Rekomenduojama, jog nenaudojamo reagento buteliukai, išimti iš analizatoriaus, būtų laikomi uždaryti šaldytuve. Atidarytas kalibratorius yra pastovus 8 savaites 2…8 °C temperatūroje. Nenaudojamas kalibratorius turi būti gerai uždarytas. Nenaudokite jokių buteliukų, kurie akivaizdžiai pažeisti ar kuriuose matomas mikrobinis augimas. MĖGINIŲ SURINKIMAS Mėginio rūšis Galima naudoti serumą, heparino plazmą arba EDTA plazmą. Bet kokie drumsti mėginiai turi būti apdorojami centrifugoje (pvz. 2000xg išlaikant 15 min.) Atsargumo priemonės Su žmonių mėginiais reikia dirbti ir juos šalinti taip, lyg jie būtų potencialiai užkrečiami. Laikymas (2) Mėginius galima laikyti 8 dienas esant 2…8 °C, arba 3 mėnesius esant -20 °C temperatūrai (jei užšaldyta tik kartą). TYRIMO TVARKA Apie automatizuotą procedūrą naudojant Konelab analizatorių skaitykite Nuorodų vadove ir Pastabose apie Pritaikymą. Jei prietaisas naudojamas procedūroms, kurių nepatvirtino Thermo Fisher Scientific Oy, negalima garantuoti tų procedūrų sėkmės ir tokį pritaikymą turi įvertinti vartotojas. Būtinos, bet netiekiamos medžiagos Toliau aprašytos kontrolinės medžiagos. Kalibravimas Naudokite CRP HS kalibratorių, esantį pakuotėje; laikykitės jūsų Konelab analizatoriui pateiktų instrukcijų. Rekomenduojama kalibratorių laikyti analizatoriuje daugiausiai vieną valandą. Sietis: CRP vertė priskirta kaip pirminę nuorodą naudojant IFCC preparatą CRM 470. Kokybės kontrolė Naudokite kokybės kontrolės mėginius bent kartą per dieną ir po kiekvieno kalibravimo, taip pat kiekvieną kartą, kai naudojate naują reagento buteliuką. Kontrolinės medžiagos: CRP didelio jautrumo kontrolinė medžiaga, kodas 981852 Kontrolės intervalai ir apribojimai turi būti pritaikyti individualios laboratorijos poreikiams. Kokybės kontrolės mėginių rezultatai turi patekti į iš anksto laboratorijos nustatytas ribas. REZULTATŲ APSKAIČIAVIMAS Konelab analizatorius rezultatus apskaičiuoja automatiškai naudodamas kalibravimo kreivę. Analizatorius automatiškai pagal iš anksto nustatytus parametrus iš turimų atsargų paruošia skiedinį. Kalibravimo kreivė generuojama automatiškai pagal matuotus kalibratorius. Kalibravimo kreivė (pavyzdys) CRP HS 1,6 1,4 1,2 1 0,8 0,6 0,4 0 5 10 15 20 Koncentracija (mg/l) Konelab 20/30/60 Kalibravimo kreivė priklauso nuo partijos. PROCEDŪROS APRIBOJIMAI (6) Interferencija Kriterijus: pradinių verčių atstatymas ± 10% tikslumu. Bilirubinas: Iki 1000 μmol/l (58.5 mg/dl) interferencija nepastebėta. Hemolizė: Iki 10 g/l hemoglobino hemolizate interferencija nepastebėta. Lipemija: Iki 4 g/l Intralipid® (Fresenius Kabi AB prekinis ženklas) interferencija nepastebėta. Koreliacija tarp drumstimosi ir trigliceridų koncentracijos labai nedidelė . Daugiau apie interferuojančias medžiagas skaitykite 3 nuorodoje. Su bandiniu gali interferuoti heterofiliniai antikūniai, monokloniniai imunoglobulinai (ar jų dalys) ir reumatoidiniai veiksniai. Pacientų, galinčių turėti šių antikūnių, tyrimų rezultatai turi būti vertinami atsargiai (4). Antigeno pertekliaus aptikimas CRP koncentraciją didinant iki 300 mg/l antigenų pertekliaus nepastebėta NUMATOMOS VERTĖS (5) Žemiau 3 mg/l Šios vertės turi būti naudojamos tik kaip orientyras. Rekomenduojama, kad kiekviena laboratorija patikrintų šį diapazoną arba išvestų atskaitos intervalą populiacijai, kuriai ji tarnauja. Naudojant CRP įvertinti karūninės širdies ligos pavojų, matavimai turi būti lyginami su ankstesnėmis vertėmis. MATAVIMO SRITIS 0.25 – 10 mg/l Praplėsta matavimo sritis antrą kartą atskiedus: 0.25 – 40 mg/l EKSPLOATACIJOS CHARAKTERISTIKOS (6) Aptikimo riba Aptikimo riba žymi žemiausią išmatuojamą koncentraciją/ aktyvumą, kurį galima atskirti nuo nulio: 0.25 mg/l Paklaida (rezultatų matavimo vienetas (mg/l) Reikšmė 0.80 mg/l Reikšmė 2.41 mg/l 1 sn CV% 1 sn CV% Serijoje 0.02 3.2 0.02 0.8 Kas dieną 0.03 3.7 0.04 1.8 Viso 0.04 5.9 0.05 2.3 Reikšmė 7.02 mg/l 1 sn CV% Serijoje 0.04 0.6 Kas dieną 0.08 1.2 Viso 0.15 2.2 Tikslumo tyrimas atliktas pagal gaires, išdėstytas NCCLS dokumente EP5-A, 20 dienų naudojant Konelab 30, matavimų skaičius buvo bent n= 80. © 2007 Thermo Fisher Scientific Inc. All rights reserved. CRP didelio jautrumo Tiekiamos medžiagos Aukščiau aprašyti reagentai ir tuščias brūkšniniu kodu pažymėtas buteliukas, skirtas naudoti Konelab analizatoriuose. Template: D01162_4 Konelab™ / T Series CRP High Sensitivity Rezultatas (A) LT Page 14 D02418_04_insert_CRP HS_MU Metodų palyginimas 1. Lyginamasis tyrimas atliktas kaip gaires naudojant NCCLS dokumentą EP9-A ir kaip etaloną naudojant kitą komerciškai prieinamą metodą. Linijinė regresija (rezultatas mg/l): y = 1.01 x – 0.29 r = 0.983 n = 87 Mėginio koncentracijos buvo nuo 0,29 iki 10 mgl/l. Atskirose laboratorijose gauti rezultatai gali skirtis nuo pateiktų duomenų. BIBLIOGRAFIJA 1) Burtis CA and Ashwood ER (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, WB Saunders Company, Philadelphia, 2001, pp. 332-333 and 688. 2) Guder WG, Narayanan S, Wisser H, Zawta B. List of Analytes; Preanalytical variables. Brochure in: Samples: From Patient to the Laboratory. GIT Verlag GmbH, Darmstadt, 1996. 3) Young DS, Effects of Drugs on Clinical Laboratory Tests, Volume One, Listing by Test, Fifth Edition, AACC Press, Washington, D.C., 3-233 – 3-234. 4) Selby C. Interference in immunoassay. Ann Clin Biochem 1999;36:704-721. 5) Pearson TA et al. Markers of Inflammation and Cardiovascular Disease Application to Clinical and Public Health Practice, A Statement for Healthcare Professionals From the Centers for Disease Control and the American Heart Association., Circulation 2003;107:499-511. 6) Duomenys iš Thermo Fisher Scientific Oy. GAMINTOJAS Thermo Fisher Scientific Oy Clinical Diagnostics Finland Ratastie 2, P.O. Box 100, FI-01621 Vantaa Tel. +358 9 329 100, Faksas +358 9 3291 0300 www.thermo.com/konelab Peržiūros data (mmmm-mm-dd) 2007-09-13 Pokyčiai nuo ankstesnės versijos Atnaujintas kompanijos pavadinimas. Piesardzības pasākumi Tikai in vitro diagnostikai. Ievērojiet parastos piesardzības pasākumus par apiešanos ar visiem laboratorijas reaģentiem. Kalibrators satur humānas izcelsmes materiālu, kas ir pārbaudīts un nesatur HBsAg, HIV ½ antivielas un HVC. Tomēr ar produktu jārīkojas tikpat uzmanīgi, kā ar pacienta paraugu. Liofilizēts CRO HS reaģents kā konservantu satur nātrija azīdu (1 - < 2.5%). Izvairieties no putekļu ieelpošanas. Nenorīt. Izvairieties no kontakta ar ādu un gļotādām. Kontakta gadījumā, skalojiet ar lielu ūdens daudzumu. Nātrija azīds var reaģēt ar svinu un varu drenēšanas sistēmā un veidot viegli uzliesmojošus metālu azīdus. Pirms iznīcināšanas skalot ar lielu ūdens daudzumu. Riska un drošības frāzes sk. atsevišķā lapā (T:R;23/24/25 - 52/53 S;22-28-45). Pēc atšķaidīšanas CRO HS reaģenta šķīdums satur < 0.1% nātrija azīdu; šāda koncentrācija netiek uzskatīta par bīstamu. Sagatavošana CRO HS buferis, parauga atšķaidītājs un CRO HS kalibrators ir gatavi lietošanai. CRO HS reaģenta atšķaidīšana: Pievienojiet precīzi 4 ml dejonizēta ūdens. Noslēdziet pudelīti, uzmanīgi sajauciet un atstājiet istabas temperatūrā vismaz 30 min. Pirms lietošanas vēlreiz sajauciet pudelītes saturu. Nekratiet. Pārvietojiet reaģentu tukšā pudelītē, kas kodēta ar svītrkodu un paredzēta izmantošanai Konelab analizatoros.. 1. piezīme: Vienmēr izmantojiet CRO HS buferi, CRO HS reaģentu un CRO HS kalibratoru ar vienādu partijas numuru, jo komponentiem jāsaskan pēc partijas numura. 2. piezīme: Pirms ievietošanas Konelab analizatorā, rūpīgi sajauciet CRO HS reaģentu. Nekratiet. 3. piezīme: Ievietojot reaģenta pudelītes vai traukus Konelab analizatorā, pārliecinieties, ka pudelītes kakliņā vai uz reaģenta virsmas nav gaisa burbuļu. Uzglabāšana un stabilitāte Reaģenti neatvērtās pudelītēs ir stabili 2…8 °C temperatūrā līdz derīguma termiņa beigām; derīguma termiņš ir uzdrukāts uz etiķetes. Lai iegūtu informāciju par ievietoto reaģentu stabilitāti, sk. Konelab analizatora “Pielietojuma piezīmes”. Atšķaidīts CRO reaģents un atvērts CRO HS buferis var tikt izmantots 8 nedēļas, ja nelietošanas gadījumā tiek uzglabāts noslēgts 2…8 °C temperatūrā. Ja reaģenti netiek lietoti, ir ieteicams reaģentu pudelītes izņemt no analizatora un uzglabāt ledusskapī noslēgtā veidā. Pēc atvēršanas CRO HS kalibrators ir stabils 8 nedēļas 2…8 °C temperatūrā. Ja kalibrators netiek lietots, uzglabājiet to cieši noslēgtu ar vāciņu. Neizmantojiet pudelīti, ja tajā redzamas mikrobu vairošanās pazīmes. PARAUGU IEGŪŠANA Parauga veids Var izmantot serumu, heparīna plazmu vai EDTA plazmu. Jebkurš duļķains paraugs ir jācentrifugē (piem., 2000xg 15 min) Augstas jutības CRO 981798 1 x 4 ml ŠIS IEPAKOJUMA PIELIKUMS IR PIELIETOJAMS ĀRPUS ASV. JEBKURA ATSAUKSME UZ KONELAB SISTĒMU ATSAUCAS ARĪ UZ T SĒRIJU. PAREDZĒTĀ IZMANTOŠANA In vitro kvantitatīvai zemas koncentrācijas CRO noteikšanai cilvēka serumā vai plazmā ar Konelab analizatoriem. Visi testa rezultāti ir jāņem vērā, vadoties pēc klīniskā konteksta. KOPSAVILKUMS (1, 5) C reaktīvais olbaltums (CRO) ir normāla seruma sastāvdaļa, kas ļoti zemās koncentrācijās sastopams veselu cilvēku serumā. Invazīva bakteriāla infekcija un plašs audu bojājums izraisa CRO koncentrācijas pieaugumu. CRO ir viens no visjutīgākajiem akūtas fāzes proteīniem, tā koncentrācija plazmā parasti izteikti pieaug pēc miokarda infarkta, traumas, infekcijas, iekaisuma, operācijas vai audzēja proliferācijas. Koncentrācijas pieaugums sākas 6 līdz 12 stundas pēc bojājuma un vērtības var 2000 reizes pārsniegt normu. CRO noteikšana klīnikā palīdz organisku slimību skrīningā, iekasuma slimību aktivitātes noteikšanā, interkurentu infekciju atklāšanā sistēmiskās sarkanās vilkēdes, leikēmijas gadījumā vai pēc operācijas un jaundzimušo septicēmijas un meningīta kontrolē, kad var būt sarežģīta paraugu paņemšana bakterioloģiskai izmeklēšanai. CRO noteikšana, izmantojot jutīgas CRO noteikšanas metodes, var palielināt citu marķieru prognostisko nozīmi, nosakot sirds-asinsvadu un perifēro asinsvadu slimību risku. PROCEDŪRAS PRINCIPS Šī metode ir balstīta uz imūnprecipitācijas mērījumu pie 540 nm. Bufera paraugiem tiek pievienotas mikrodaļiņas, kas pārklātas ar antihumāno CRO. Reakcijai sasniedzot beigu punktu, tiek atzīmēts imūnprecipitācijas izraisītais absorbcijas palielinājums. Izmaiņas absorbcijā ir proporcionālas antigēna (CRO) daudzumam šķīdumā. REAĢENTU INFORMĀCIJA CRO HS reaģents CRO HS buferis Parauga atšķaidītājs CRO Plus kalibrators, Koncentrācijas CRO HS reaģents 1 x 4 ml 2 x 50 ml 1 x 60 ml 2 x 1 ml Mikrodaļiņas, kas pārklātas ar antihumāno CRO (auna) (liofilizēts) NaN3 1 - < 2.5% CRO HS buferis Tris 0.2 mol/l NaN3 < 0.1 % Parauga atšķaidītājs PBS 0.01 mol/l NaN3 < 0.1 % CRO HS kalibrators cilvēku NaN3 < 0.1 % Kalibratora CRO koncentrācijas ir uzdrukātas atsevišķas Referento vērtību lapas. Piesardzības pasākumi Ar no cilvēkiem iegūtajiem paraugiem jārīkojas un no tiem jāatbrīvojas tā, it kā tie būtu iespējami infekciozi. Uzglabāšana (2) Paraugu var uzglabāt 8 dienas 2…8 °C temperatūrā vai 3 mēnešus -20 °C temperatūrā (ja tiek sasaldēts tikai vienu reizi). PĀRBAUDES PROCEDŪRA Lai iegūtu informāciju par Konelab analizatora automātisko procedūru, sk. Uzziņas rokasgrāmatu un Pielietojuma piezīmes. Jebkāds pielietojums, kuru nav apstiprinājis Thermo Fisher Scientific Oy, nevar sniegt garantētu rezultātu, tādēļ tas jāizvērtē pašam lietotājam. PIEGĀDĀTIE MATERIĀLI Reaģenti, kā aprakstīts iepriekš, un tukša pudelīte, kas kodēta ar svītrkodu un paredzēta izmantošanai Konelab analizatoros. Nepieciešamie materiāli, kas netiek piegādāti Kontroles, kā aprakstīts tālāk. Kalibrēšana Izmantojiet komplektam pievienoto CRO HS kalibratoru, atbilstoši jūsu Konelab analizatora instrukcijām. Ļaujiet kalibratoram atrasties analizatorā maksimāli vienu stundu. Izsekojamība: CRO vērtība ir noteikta, kā primāro atsauci izmantojot IFCC sagatavoto CRM 470. Kvalitātes kontrole Izmantojiet kvalitātes kontroles paraugu vismaz reizi dienā un pēc katras kalibrācijas, kā arī ikreiz, kad tiek izmantota jauna reaģenta pudele. Pieejamās kontroles: CRO augstas jutības kontrole, kods 981852 Kontroles intervāli un robežlielumi ir jāpiemēro individuālajām laboratorijas prasībām. Kvalitātes kontroles paraugu rezultātiem jābūt laboratorijas noteikto vērtību robežās. REZULTĀTU APRĒĶINĀŠANA Konelab analizators rezultātus aprēķina automātiski, izmantojot kalibrācijas līkni. Analizators automātiski sagatavo atšķaidījumu sērijas no kalibratora krājumiem, atbilstoši iepriekš iestatītajiem parametriem. Kalibrācijas līkne tiek veidota, izmantojot kalibratoru mērījumus, nosakot ierievja atbilstību. © 2007 Thermo Fisher Scientific Inc. All rights reserved. Konelab™ / T Sērijas CRP High Sensitivity Template: D01162_4 LV Page 15 D02418_04_insert_CRP HS_MU RAŽOTĀJS Thermo Fisher Scientific Oy Clinical Diagnostics Finland Ratastie 2, P.O. Box 100, FI-01621 Vantaa Tālr. +358 9 329 100, Fakss +358 9 3291 0300 www.thermo.com/konelab Kalibrācijas līkne (piemērs) CRP HS Reaģēšana (A) 1.6 Pārskatīšanas datums (gggg-mm-dd) 2007-09-13 1.4 Veiktās izmaiņas Atjaunināts uzņēmuma nosaukums. 1.2 1 0.8 0.6 PL 0.4 0 5 10 15 20 Koncentrācija (mg/l) Konelab™ / Seria T CRP High Sensitivity Wysokoczuły test na CRP Mijiedarbība Kritērijs: Atjaunošana ±10% robežās no sākotnējām vērtībām. Bilirubīns: Līdz koncentrācijai 1000 μmol/l (58.5 mg/dl) interference nav novērota. Hemolīze: Hemolizātā, kura hemoglobīna koncentrācija ir līdz 10 g/l, interference nav novērota. Lipēmija: Līdz Intralipid® (Fresenius Kabi AB preču zīme) koncentrācijai 4 g/l interference nav novērota. Pastāv neliela saistība starp duļķainumu un triglicerīdu koncentrāciju. Informācijai par citām interferējošām vielām, lūdzu, sk. 3. atsauci. Ar metodi var mijiedarboties heterofīlas antivielas, monoklonāli imūnglobulīni (vai to daļas) un reimatoīdais faktors. Pacientu rezultāti, kuru paraugos iespējama šādu antivielu klātbūtne, ir rūpīgi jāizvērtē (4). Antigēna pārpalikums Antigēna pārpalikums nav novērots līdz CRO koncentrācijai, kas nav augstāka par 300 mg/l. VĒLAMĀS VĒRTĪBAS (5) Zem 3 mg/l Minētās vērtības ir tikai ar rekomendējošu nozīmi. Ieteicams, lai katra laboratorija apstiprinātu šo diapazonu vai izstrādātu referentos lielumus populācijai, kuru tā apkalpo. Ja CRO tiek izmantots, lai novērtētu koronārās sirds slimības risku, mērījums ir jāsalīdzina ar iepriekšējiem mērījumiem. MĒRĪŠANAS DIAPAZONS 0.25 – 10 mg/l Paplašinātais mērīšanas diapazons pēc otrreizējas atšķaidīšanas: 0.25 – 40 mg/l VEIKTSPĒJAS RAKSTUROJUMS (6) Uztveršanas robeža Uztveršanas robeža ir zemākā kvantitatīvi izmērāmā koncentrācija: 0.25 mg/l Neprecizitāte (rezultāta vienība mg/l) Vidējais 0.80 mg/l Vidējais 2.41 mg/l SD CV% SD CV% Darbības laikā 0.02 3.2 0.02 0.8 Ar dienas starpību 0.03 3.7 0.04 1.8 Kopā 0.04 5.9 0.05 2.3 Vidējais 7.02 mg/l SD CV% Darbības laikā 0.04 0.6 Ar dienas starpību 0.08 1.2 Kopā 0.15 2.2 Precizitātes izpēte tika veikta vismaz divdesmit dienas, saskaņā ar NCCLS EP5-A dokumenta noteikumiem, izmantojot Konelab 30, mērījumu skaits vismaz n = 80. Metodes salīdzinājums Salīdzināšanas izpēte tika veikta saskaņā ar NCCLS EP9-A dokumenta noteikumiem, izmantojot citu komerciāli pieejamu metodi kā atsauci. Lineārā regresija (rezultāta vienība mg/l): y = 1.01 x – 0.29 r = 0.983 n = 87 Paraugu koncentrācijas bija diapazonā no 0.29 līdz 10 mg/l. Rezultāti, kas iegūti atšķirīgās laboratorijās, var atšķirties no dotajiem veiktspējas datiem. BIBLIOGRĀFIJA 1) Burtis CA and Ashwood ER (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, WB Saunders Company, Philadelphia, 2001, pp. 332-333 and 688. 2) Guder WG, Narayanan S, Wisser H, Zawta B. List of Analytes; Preanalytical variables. Brochure in: Samples: From Patient to the Laboratory. GIT Verlag GmbH, Darmstadt, 1996. 3) Young DS, Effects of Drugs on Clinical Laboratory Tests, Volume One, Listing by Test, Fifth Edition, AACC Press, Washington, D.C., 3-233 – 3-234. 4) Selby C. Interference in immunoassay. Ann Clin Biochem 1999;36:704-721. 5) Pearson TA et al. Markers of Inflammation and Cardiovascular Disease Application to Clinical and Public Health Practice, A Statement for Healthcare Professionals From the Centers for Disease Control and the American Heart Association., Circulation 2003;107:499-511. 6) Dati Thermo Fisher Scientific Oy failos. 1 x 4 ml NINIEJSZA ULOTKA OPAKOWANIA PRZEZNACZONA JEST DO STOSOWANIA POZA GRANICAMI USA. WSZELKIE ODNIESIENIA DO SYSTEMÓW KONELAB ODNOSZĄ SIĘ RÓWNIEŻ DO SERII T. PRZEZNACZENIE Do analiz ilościowych białka C-reaktywnego (C-Reactive Protein, CRP) w niskich stężeniach w ludzkiej surowicy lub osoczu, w analizatorach Konelab. Wyniki wszystkich testów należy interpretować w kontekście danych klinicznych. OMÓWIENIE (1, 5) CRP, normalny składnik surowicy, u zdrowych osób występuje w bardzo niskich stężeniach. Na skutek zakażeń inwazyjnymi gatunkami bakterii lub rozległego uszkodzenia tkanek następuje zwiększenie poziomu CRP. CRP należy do białek ostrej fazy pozwalających na uzyskanie oznaczenia o najwyższej czułości. Poziom CRP zwykle gwałtownie wzrasta w odpowiedzi na zawał serca, uraz, zakażenia, zapalenia, zabiegi operacyjne lub szerzenie się nowotworu. Wzrost rozpoczyna się po 6 - 12 godzinach i może osiągać wartości przekraczające 2000-krotnie normę. Test na CRP pozwala na wykonanie szeregu oznaczeń, przydatnych z klinicznego punktu widzenia – badań przesiewowych chorób somatycznych, oceny aktywności chorób zapalnych, wykrywania współistniejących zakażeń w toczniu rumieniowatym układowym oraz białaczce. Oznaczenie jest również przydatne po zabiegach chirurgicznych i w leczeniu posocznicy oraz zapalenia opon mózgowych u noworodków, gdy pobranie próbek do badań bakteriologicznych może być utrudnione. Ponadto oznaczenie CRP prowadzone za pomocą wysokoczułych testów może zwiększać skuteczność predykcji innych markerów ryzyka schorzeń układu sercowo-naczyniowego i naczyń obwodowych. ZASADA DZIAŁANIA TESTU Test przebiega w oparciu o pomiar immunoprecypitacji przy długości fali 540 nm. Do próbek zawieszonych w buforze dodaje się mikrocząstki opłaszczone przeciwciałami skierowanymi przeciwko ludzkiemu CRP. Po osiągnięciu punktu końcowego reakcji rejestruje się zwiększenie absorbancji wywołane immunoprecypitacją. Zmiana absorbancji jest proporcjonalna do zawartości antygenu (CRP) w roztworze. INFORMACJE O ODCZYNNIKACH Odczynnik CRP HS Bufor CRP HS Rozcieńczalnik do próbek Kalibrator CRP HS 1 x 4 ml 2 x 50 ml 1 x 60 ml 2 x 1 ml Stężenia Odczynnik CRP HS Mikrocząstki opłaszczone przeciwciałami owczymi skierowanymi przeciwko ludzkiemu CRP. NaN3 1 - < 2.5% Bufor CRP HS Tris 0.2 mol/l NaN3 < 0.1 % Rozcieńczalnik do próbek PBS 0.01 NaN3 < 0.1 % Kalibrator CRP HS (ludzki) NaN3 < 0.1 % Stężenie CRP kalibratora podano na oddzielnym Arkuszu wartości. liofilizowanymi mol/l Środki ostrożności Do stosowania wyłącznie w diagnostyce in vitro. Należy przestrzegać standardowych środków ostrożności obowiązujących podczas pracy ze wszystkimi odczynnikami analitycznymi. Kalibrator zawiera materiał pochodzenia ludzkiego, w przypadku którego uzyskano ujemny wynik badań na obecność antygenów HBs i przeciwciał skierowanych przeciwko HIV 1/2 i HCV. Niemniej jednak podczas pracy z produktem należy zachować środki ostrożności obowiązujące dla wszystkich próbek pochodzących od pacjentów. Liofilizowany odczynnik CRP HS zawiera środek konserwujący – azydek sodu w stężeniu (1 - < 2.5%). Należy unikać wdychania pyłu. Nie połykać. Unikać kontaktu ze skórą i błonami śluzowymi. Jeżeli dojdzie do takiego kontaktu, zmyć obfitą ilością wody. Azydek sodu może reagować z ołowianymi i miedzianymi rurami instalacji wodno-kanalizacyjnej, tworząc bardzo wybuchowe azydki metali. Wylewane odpady należy spłukać dużą ilością wody. Informacje dotyczące zagrożeń i bezpieczeństwa pracy przedstawiono w odrębnym dokumencie (T:R;23/24/25 - 52/53 S;22-28-45). Po przygotowaniu płynny odczynnik CRP HS zawiera środek konserwujący – azydek sodu w stężeniu < 0,1%, uznawanym za bezpieczne. Przygotowanie oznaczenia Odczynnik CRP HS, rozcieńczalnik do próbek i kalibrator CRP HS są gotowe do użycia. Przygotowanie odczynnika CRP HS: Dodać dokładnie 4 ml wody dejonizowanej. Zamknąć fiolkę, ostrożnie zmieszać i odstawić na 30 minut w temperaturze pokojowej. Przed użyciem ponownie dokładnie zmieszać odczynniki. Nie wstrząsać. Przenieść odczynnik do pustej fiolki do stosowania w analizatorach Konelab, znakowanej kodem kreskowym. Uwaga 1: Ponieważ składniki są dobierane w obrębie jednej partii, należy zawsze stosować bufor CRP HS, odczynnik CRP HS oraz kalibrator CRP HS oznaczone tym samym numerem partii. © 2007 Thermo Fisher Scientific Inc. All rights reserved. PROCEDŪRAS IEROBEŽOJUMI (6) 981798 Template: D01162_4 Konelab 20/30/60. Kalibrācijas līkne ir atkarīga no partijas numura. Page 16 D02418_04_insert_CRP HS_MU POBIERANIE PRÓBEK Typ próbki Można stosować próbki surowicy, heparynizowanego osocza lub osocza z dodatkiem EDTA. W przypadku zmętnienia należy odwirować próbki (np. przez 15 min z przeciążeniem 2000xg.) Środki ostrożności Próbki ludzkiego materiału biologicznego powinny być traktowane jako potencjalnie zakaźne. Przechowywanie (2) Próbki mogą być przechowywane w temp. 2...8 °C przez 8 dni, a w temp. -20 °C (jeśli zastosowano jednokrotne zamrożenie) przez 3 miesiące. SPOSÓB WYKONANIA TESTU Informacje dotyczące automatycznego przeprowadzenia testu przedstawiono w Źródłowej instrukcji obsługi i Uwagach dotyczących obsługi analizatora Konelab. Nie gwarantuje się odpowiedniej wydajności żadnych zastosowań, które nie otrzymały atestu firmy Thermo Fisher Scientific Oy. Tego typu zastosowania powinny zostać poddane walidacji przez Użytkowników. Dostarczone materiały Odczynniki, zgodnie z opisem powyżej, oraz fiolka do stosowania w analizatorach Konelab, znakowana kodem kreskowym. Materiały niezbędne do wykonania badania, lecz niedostarczane Odczynnik kontrolny, zgodnie z opisem poniżej. Kalibracja Należy stosować kalibrator CRP HS wchodzący w skład zestawu, zgodnie z instrukcjami dołączonymi do analizatora Konelab. Nie należy pozostawiać kalibratora w analizatorze na dłużej niż 1 godz. Zgodność z normami: Ilość CRP oznaczono przy użyciu preparatu IFCC CRM 470 jako głównego odczynnika wzorcowego. Kontrola jakości Należy badać próbkę do kontroli jakości co najmniej raz dziennie oraz po każdej kalibracji i po rozpoczęciu stosowania nowej butelki odczynnika. Dostępne odczynniki kontrolne: CRP High Sensitivity Control (wysokoczuły odczynnik kontrolny CRP), nr kat. 981852. Przedziały czasowe i limity kontroli jakości należy dostosować do wymogów poszczególnych pracowni analitycznych. Wynik próbki używanej do kontroli jakości powinien się zawierać w granicach ustalonych w danej pracowni analitycznej. OBLICZANIE WYNIKÓW Wyniki są obliczane automatycznie w analizatorze Konelab na podstawie krzywej kalibracyjnej. W analizatorze jest automatycznie przygotowywana seria rozcieńczeń stężonego roztworu kalibratora, zgodnie z zaprogramowanymi parametrami. Krzywa kalibracyjna jest sporządzana na podstawie pomiaru kalibratorów metodą wygładzenia połączenia punktów pomiarowych. Krzywa kalibracyjna (przykład) Zmiana absorbiancji (A) CRP HS Zakłócenia Kryterium: wartość odzysku wynosi ± 10% wartości początkowych. Bilirubina: nie stwierdzono zakłóceń w przypadku stężeń do 1000 μmol/l (58.5 mg/dl). Hemoliza: nie powoduje zakłóceń przy stężeniu poniżej 10 g/l hemoglobiny w hemolizacie. Lipemia: nie stwierdzono zakłóceń w przypadku stężeń Intralipidu® (znak towarowy firmy Fresenius Kabi AB) do 4 g/l. Istnieje słaba korelacja pomiędzy stopniem zmętnienia a stężeniem triglicerydów. Informacje dotyczące innych substancji zakłócających przedstawiono w punkcie 3. piśmiennictwa. Przeciwciała heterofilne, immunoglobuliny monoklonalne (lub ich składowe) oraz czynniki reumatoidalne mogą zakłócać opisywane oznaczenie. Należy zachować ostrożność, oceniając wyniki próbek pochodzących od pacjentów, u których podejrzewa się obecność wymienionych przeciwciał (4). Nadmiar antygenu Nie opisywano nadmiaru antygenu w stężeniach CRP do 300 mg/l. WARTOŚCI OCZEKIWANE (5) Poniżej 3 mg/l Przedstawione wielkości należy traktować jedynie jako wartości orientacyjne. Zaleca się, by w każdej pracowni analitycznej zweryfikowano podane zakresy lub ustalono własny przedział referencyjny dla badanej populacji. Stosując oznaczenie CRP w ocenie ryzyka choroby wieńcowej, należy analizować wyniki w odniesieniu do wartości uzyskanych uprzednio. ZAKRES POMIAROWY 0.25 – 10 mg/l Rozszerzony zakres pomiarowy po powtórnym rozcieńczeniu: 0.25 – 40 mg/l CHARAKTERYSTYKA WYDAJNOŚCIOWA (6) Granica oznaczalności Granica oznaczalności (najniższe stężenie oznaczalne ilościowo): 0.25 mg/l Niedokładność (wynik wyrażony w mg/l) Średnia 0.80 mg/l Średnia 2.41 mg/l SD CV% SD CV% W jednej serii 0.02 3.2 0.02 0.8 W różnych dniach 0.03 3.7 0.04 1.8 Sumarycznie 0.04 5.9 0.05 2.3 Średnia 7.02 mg/l SD CV% W jednej serii 0.04 0.6 W różnych dniach 0.08 1.2 Sumarycznie 0.15 2.2 Badanie precyzji przeprowadzono na podstawie wytycznych zawartych w dokumencie NCCLS EP5-A, w analizatorze Konelab 30, wykonując co najmniej 80 oznaczeń w czasie nie krótszym niż 20 dni. Porównanie metod Przeprowadzono badanie porównawcze w oparciu o wytyczne zawarte w dokumencie NCCLS EP9-A, wykorzystując dostępną w handlu metodę referencyjną. Regresja liniowa (wynik w mg/l): y = 1.01 x – 0.29 r = 0.983 n = 87 Stężenia związków w próbkach wynosiły od 0.29 do 10 mg/l. Wyniki uzyskane w poszczególnych pracowniach analitycznych mogą się różnić od podanych parametrów wydajnościowych. PIŚMIENNICTWO 7) Burtis CA and Ashwood ER (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, WB Saunders Company, Philadelphia, 2001, pp. 332-333 and 688. 8) Guder WG, Narayanan S, Wisser H, Zawta B. List of Analytes; Preanalytical variables. Brochure in: Samples: From Patient to the Laboratory. GIT Verlag GmbH, Darmstadt, 1996. 9) Young DS, Effects of Drugs on Clinical Laboratory Tests, Volume One, Listing by Test, Fifth Edition, AACC Press, Washington, D.C., 3-233 – 3-234. 10) Selby C. Interference in immunoassay. Ann Clin Biochem 1999;36:704-721. 11) Pearson TA et al. Markers of Inflammation and Cardiovascular Disease Application to Clinical and Public Health Practice, A Statement for Healthcare Professionals From the Centers for Disease Control and the American Heart Association., Circulation 2003;107:499-511. 12) Dane znajduja sie w dokumentacji firmy Thermo Fisher Scientific Oy. Producent Thermo Fisher Scientific Oy Clinical Diagnostics Finland Ratastie 2, P.O. Box 100, FI-01621 Vantaa tel. +358 9 329 100; faks +358 9 3291 0300 www.thermo.com/konelab 1,6 1,4 1,2 Data zatwierdzenia (rrrr-mm-dd) 2007-09-13 1 Zmiany w stosunku do poprzedniej wersji Zaktualizowano nazwę firmy. 0,8 0,6 0,4 0 5 10 15 20 Stezenie (m g/l) Konelab 20/30/60. Przebieg krzywej kalibracyjnej zależy od partii odczynników. © 2007 Thermo Fisher Scientific Inc. All rights reserved. Przechowywanie i stabilność Odczynniki zachowują stabilność do terminu ważności podanego na etykiecie, o ile są przechowywane w temperaturze 2...8 °C w zamkniętych fiolkach. Dane dotyczące stabilności odczynników znajdujących się w analizatorze Konelab przedstawiono w Uwagach dotyczących obsługi analizatora Konelab. Przygotowany odczynnik CRP HS i raz otwarty bufor CRP HS można stosować przed upływem 8 tygodni, o ile niewykorzystane odczynniki przechowywano w zamkniętych naczyniach w temperaturze 2…8 °C. Jeżeli odczynniki nie są wykorzystywane, zaleca się, by wyjąć fiolki z odczynnikami z analizatora, zamknąć i przechowywać w chłodziarce. Raz otwarty kalibrator CRP HS zachowuje trwałość przez 8 tygodni w temperaturze 2…8 °C. Jeżeli odczynnik nie jest używany, powinien być przechowywany w szczelnie zamkniętym naczyniu. Nie stosować fiolek, w których są widoczne objawy wzrostu bakterii. OGRANICZENIA PROCEDURY (6) Template: D01162_4 Uwaga 2: Przed wprowadzeniem do analizatora Konelab należy ostrożnie zmieszać odczynnik CRP HS. Nie wstrząsać. Uwaga 3: Wkładając fiolki lub naczynia z odczynnikami do analizatora Konelab, należy się upewnić, że w szyjce naczynia ani na powierzchni odczynnika nie ma pęcherzyków powietrza. Page 17 D02418_04_insert_CRP HS_MU PT 981798 1 x 4 ml ESTE FOLHETO INFORMATIVO É APLICÁVEL PARA USO FORA DOS E.U.A. QUALQUER REFERÊNCIA AOS SISTEMAS KONELAB TAMBÉM SE REFERE À SÉRIE T. USO PRETENDIDO Para a determinação quantitativa in vitro de concentrações reduzidas de PCR no soro e plasma humanos nos analisadores Konelab. Todos os resultados dos testes devem ser interpretados tendo em conta o contexto clínico. SUMÁRIO (1, 5) A proteína C-reactiva (PCR) é um constituinte normal do soro presente nos indivíduos saudáveis, em concentrações bastante reduzidas. As infecções bacterianas invasivas e danos extensos nos tecidos provocam o aumento dos níveis de PCR. A PCR é uma das proteínas de fase aguda mais sensíveis; os níveis presentes no plasma aumentam, normalmente, de forma substancial após um enfarte do miocárdio, traumas, infecções, inflamações, cirurgias ou proliferações neoplásticas. O aumento de nível começa a verificar-se decorridas entre 6 a 12 horas, podendo chegar a cerca de 2000 vezes acima do nível normal. A determinação da PCR é clinicamente útil para a visualização de doenças orgânicas, a avaliação da actividade das doenças inflamatórias, a detecção de infecções intercorrentes de lúpus eritematoso sistémico (SLE), na leucemia ou após uma cirurgia e na gestão de septicemias e meningites neonatais, quando a recolha de amostras para investigação bacteriológica se pode tornar mais complicada. A medição da PCR através de ensaios sensitivos da PCR pode servir de complemento ao valor predicativo de outros marcadores utilizados para avaliar o risco de doenças vasculares cardiovasculares e periféricas. PRINCÍPIO DO PROCEDIMENTO O método é baseado na medição da imunoprecipitação a 540 nm. Micropartículas revestidas a PCR anti-humana são adicionadas a amostras tamponadas. O aumento em absorvência provocado pela imunoprecipitação é medido no ponto final da reacção. A mudança da absorvência é proporcional à quantidade de antígeno (PCR) na solução. INFORMAÇÕES DOS REAGENTES Reagente PCR HS Tampão PCR HS Diluente para amostras Calibrador PCR HS 1 x 4 ml 2 x 50 ml 1 x 60 ml 2 x 1 ml Armazenamento (2) A amostra pode ser armazenada durante 8 dias a 2...8 °C ou durante 3 meses a - 20 °C (se congelada apenas uma vez). PROCEDIMENTO Consulte o Manual de referência e as Notas de aplicação para obter mais informações sobre o procedimento automático do analisador Konelab. Qualquer aplicação não validada pela Thermo Fisher Scientific Oy não pode ter o desempenho garantido e, por isso, tem de ser avaliada pelo utilizador. Materiais fornecidos Reagentes, tal como descritos acima, e um frasco vazio e com código de barras para ser usado nos analisadores Konelab. Materiais necessários mas não incluídos Controlo conforme indicado abaixo. Calibragem Utilize o calibrador PCR HS incluído no kit de acordo com as instruções fornecidas com o analisador Konelab. Deixe o calibrador no analisador durante um período máximo de uma hora. Rastreabilidade: O valor da PCR foi atribuído mediante a utilização de preparação CRM 470 (IFCC) como referência primária. Controlo de qualidade Utilize amostras de controlo de qualidade pelo menos uma vez por dia, depois de cada calibragem e sempre que utilizar uma garrafa de reagente nova. Controlos disponíveis: CRP High Sensitivity Control (Controlo PCR de sensibilidade elevada), código 981852 Os intervalos e limites do controlo devem ser adaptados aos requisitos individuais de cada laboratório. Os resultados da amostra de controlo de qualidade devem ficar dentro dos limites predefinidos pelo laboratório. CÁLCULO DE RESULTADOS Os resultados são calculados automaticamente pelo analisador Konelab com uma curva de calibragem. O analisador prepara automaticamente uma série de diluições a partir do calibrador de acordo com os parâmetros pré-definidos. A curva de calibragem é gerada a partir dos calibradores medidos, usando o ajuste de "spline". Curva de calibragem (exemplo) Concentrações Reagente PCR HS Precauções Só para uso diagnóstico in vitro. Adopte as precauções habitualmente requeridas para o manuseamento dos reagentes de laboratório. O calibrador contém material de origem humana que foi testado e dado como negativo ao nível do HBsAg e dos anticorpos HIV-1/2 e HCV. O produto deve, contudo, ser manuseado com a mesma cautela com que se manuseiam as amostras dos pacientes. O reagente PCR HS liofilizado contém azida de sódio como conservante ( 1 - < 2.5%). Evite a inalação de poeiras. Não ingira. Evite o contacto com a pele e com as membranas mucosas. Se ocorrer contacto, lave com quantidade abundante de água. A azida de sódio pode reagir com o chumbo e com o cobre presente nos canos formando azidas metálicas altamente explosivas. Durante o processo de eliminação, utilize uma quantidade abundante de água. Consulte a folha separada para conhecer os Riscos e as Advertências de Segurança (T:R; 23/24/25 – 52/53 S; 22-28-45). Após a reconstituição, o reagente PCR HS líquido contém < 0.1% de azida de sódio, a qual não é considerada uma concentração perigosa. Preparação O tampão PCR HS, o diluente para amostras e o calibrador PCR HS estão prontos a usar. Reconstituição do reagente PCR HS: Adicione exactamente 4 ml de água desionizada. Feche o frasco, agite ligeiramente e deixe assentar durante 30 minutos à temperatura ambiente. Misture o conteúdo novamente antes de usar. Não agite. Transfira o reagente para um frasco vazio e com código de barras, específico para analisadores Konelab. Nota 1: Utilize sempre tampão PCR HS, reagente PCR HS e calibrador PCR HS do mesmo lote de reagentes porque os seus componentes têm correspondência. Nota 2: Misture ligeiramente o reagente PCR HS antes de inserir o reagente no analisador Konelab. Não agite. Nota 3: Certifique-se de que não há nenhuma bolha no gargalo do frasco ou na superfície do reagente quando inserir os frascos ou as ampolas do reagente no analisador Konelab. Armazenamento e estabilidade Os reagentes que se encontram em frascos fechados permanecem estáveis até à data de expiração indicada no rótulo, desde que mantidos a 2...8 °C. Consulte as Notas de aplicação do analisador Konelab para obter mais informações sobre a estabilidade dos reagentes quando inseridos no analisador. O reagente PCR HS reconstituído e o tampão PCR HS aberto podem ser usados durante 8 semanas se forem armazenados fechados a 2...8 °C quando não estiverem a ser utilizados. Recomenda-se que retire os frascos dos reagentes do analisador e que os mantenha fechados no frigorífico sempre que não forem utilizados. Depois de aberto, o calibrador PCR HS permanece estável durante 8 semanas a 2...8 °C. Conserve o calibrador hermeticamente fechado quando este não estiver a ser utilizado. Não utilize frascos que apresentem sinais visíveis de desenvolvimento microbiano. COLHEITA DA AMOSTRA Tipo de amostra Pode utilizar-se soro, plasma de heparina ou plasma de EDTA. As amostras turvas devem ser centrifugadas (por ex. 2000xg durante 15 min) CRP HS 1.6 1.4 Resposta (A) Micropartículas revestidas a PCR anti-humana (ovelha) (liofilizadas). NaN3 1 - < 2.5% Tampão PCR HS Tris 0.2 mol/l NaN3 < 0.1 % Diluente para amostras PBS 0.01 mol/l NaN3 < 0.1 % Calibrador PCR HS Humana NaN3 < 0.1 % A concentração de PCR do calibrador está indicada numa folha de Valores separada. 1.2 1 0.8 0.6 0.4 0 5 10 15 20 Concentração (mg/l) Konelab 20/30/60. A curva de calibragem depende do lote. LIMITAÇÕES DO PROCEDIMENTO (6) Interferência Critério: Recuperação dentro de ± 10% dos valores iniciais. Bilirrubina: Nenhuma interferência detectada até 1000 μmol/l (58.5 mg/dl). Hemólise: Nenhuma interferência detectada até 10 g/l de hemoglobina no hemolisado. Lipemia: Nenhuma interferência detectada até 4 g/l de Intralipid® (marca comercial da Fresenius Kabi AB). Há uma correlação fraca entre a turvação e a concentração de triglicéridos. Para outras substâncias interferentes, consulte a referência 3. Anticorpos heterofílicos e imunoglobulinas monoclonais (ou partes deles) e factores reumatóides podem interferir com o ensaio. Os resultados de pacientes com suspeita de tais anticorpos devem ser cuidadosamente avaliados (4). Excesso de antigénio Não foi encontrado qualquer excesso de antigénio numa concentração máxima de PCR de 300 mg/l. VALORES DE REFERÊNCIA (5) Abaixo de 3 mg/l Os valores indicados devem servir apenas como uma referência. Recomendamos que cada laboratório verifique este intervalo ou obtenha um intervalo de referência para a população que serve. Sempre que utilizar a PCR para avaliar o risco de doenças cardíacas coronárias, as medições devem ser comparadas com os valores anteriores. INTERVALO DE MEDIÇÃO 0.25 – 10 mg/l Intervalo de medição alargado depois da diluição secundária: 0.25 – 40 mg/l © 2007 Thermo Fisher Scientific Inc. All rights reserved. PCR de sensibilidade elevada Precauções As amostras humanas devem ser manuseadas e eliminadas como se fossem potencialmente infecciosas. Template: D01162_4 Konelab™ / Série T CRP High Sensitivity Page 18 D02418_04_insert_CRP HS_MU CARACTERÍSTICAS DE DESEMPENHO (6) Limite da determinação O limite da determinação é a concentração mais baixa que pode ser medida quantitativamente: 0.25 mg/l Imprecisão (unidade dos resultados mg/l) Média 0.80 mg/l Média 2.41 mg/l DS CV% DS CV% Ensaio 0.02 3.2 0.02 0.8 Vários dias 0.03 3.7 0.04 1.8 Total 0.04 5.9 0.05 2.3 Média 7.02 mg/l DS CV% Ensaio 0.04 0.6 Vários dias 0.08 1.2 Total 0.15 2.2 Efectuou-se um estudo de precisão de acordo com as directrizes do Documento EP5-A da NCCLS mediante a utilização do analisador Konelab 30 durante pelo menos 20 dias, com um número de medições equivalente a, pelo menos, n = 80. Comparação de métodos Realizou-se um estudo comparativo de acordo com as indicações contidas no Documento EP9-A da NCCLS, tendo-se utilizado como método de referência um método comercialmente disponível. Regressão linear (unidade dos resultados mg/l): y = 1.01 x – 0.29 r = 0.983 n = 87 Concentrações das amostras entre 0.29 e 10 mg/l. Os resultados obtidos em laboratórios individuais podem diferir dos dados de desempenho fornecidos. BIBLIOGRAFIA 1) Burtis CA and Ashwood ER (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, WB Saunders Company, Philadelphia, 2001, pp. 332-333 and 688. 2) Guder WG, Narayanan S, Wisser H, Zawta B. List of Analytes; Preanalytical variables. Brochure in: Samples: From Patient to the Laboratory. GIT Verlag GmbH, Darmstadt, 1996. 3) Young DS, Effects of Drugs on Clinical Laboratory Tests, Volume One, Listing by Test, Fifth Edition, AACC Press, Washington, D.C., 3-233 – 3-234. 4) Selby C. Interference in immunoassay. Ann Clin Biochem 1999;36:704-721. 5) Pearson TA et al. Markers of Inflammation and Cardiovascular Disease Application to Clinical and Public Health Practice, A Statement for Healthcare Professionals From the Centers for Disease Control and the American Heart Association., Circulation 2003;107:499-511. 6) Dados num ficheiro da Thermo Fisher Scientific Oy. FABRICANTE Thermo Fisher Scientific Oy Clinical Diagnostics Finland Ratastie 2, P.O. Box 100, FI-01621 Vantaa Tel. +358 9 329 100, Fax +358 9 3291 0300 www.thermo.com/konelab Data de revisão (aaaa-mm-dd) 2007-09-13 Alterações à versão anterior Nome da empresa actualizado. INFORMAÞII PRIVIND REACTIVII Reactiv PCR HS (cu sensibilitate ridicatã) Tampon PCR HS (cu sensibilitate ridicatã) Diluant pentru probã Calibrator PCR HS (cu sensibilitate ridicatã) 1 x 4 ml 2 x 50 ml 1 x 60 ml 2 x 1 ml Concentraþii Reactiv PCR HS Microparticule învelite cu PCR anti-uman (oaie) (liofilizat) NaN3 1 - < 2.5% Tampon PCR HS Tris 0.2 mol/l NaN3 < 0.1 % Diluant pentru probã PBS 0.01 mol/l NaN3 < 0.1 % Calibrator PCR HS Uman NaN3 < 0.1 % Concentraþia de PCR din calibrator este tipãritã pe o fişã separatã cu Valorile. Precauþii Numai pentru diagnostic in vitro. Adoptaþi mãsurile normale de precauþie necesare pentru mânuirea reactivilor de laborator. Calibratorul conþine material de origine umanã, care a fost testat şi a fost gãsit negativ la HBsAg şi la anticorpi pentru HIV 1/2- şi HCV. Cu toate acestea, produsul trebuie mânuit cu tot atâta grijã ca şi probele de la pacienþi. Reactivul PCR HS liofilizat conþine azidã de sodiu drept conservant (1 - < 2.5%). Evitaþi sã inhalaþi praful. Nu înghiþiþi. Evitaþi contactul cu pielea şi membranele mucoasei. În caz de contact, spãlaþi cu foarte multã apã. Azida de sodiu poate reacþiona cu plumbul şi cuprul din þevile de evacuare, formând azide de metal explozive. La aruncare, lãsaþi sã curgã cantitãþi mari de apã. Vezi fişa separatã pentru Fraze privind riscul şi siguranþa (T:R;23/24/25 - 52/53 S;22-28-45). Dupã reconstituire, reactivul PCR HS lichid conþine azidã de sodiu < 0.1%, care nu este consideratã o concentraþie nocivã. Pregãtirea Tamponul PCR HS, diluantul pentru probã şi calibratorul PCR HS sunt gata de folosire. Reconstituirea reactivului PCR HS: Adãugaþi exact 4 ml de apã deionizatã. Închideþi flaconul, omogenizaþi delicat şi lãsaþi sã se odihneascã cel puþin 30 de minute la temperatura camerei. Amestecaþi conþinutul încã o datã înainte de folosire. Nu agitaþi. Transferaþi reactivul într-un flacon gol cu cod de bare, care poate fi folosit în analizoarele Konelab. Nota 1: Folosiþi întotdeauna tampon PCR HS, reactiv PCR HS şi PCR HS cu acelaşi numãr al lotului de reactiv, deoarece componentele din cadrul lotului au fost potrivite. Nota 2: Amestecaþi delicat reactivul PCR HS înainte de a introduce reactivul în analizorul Konelab. Nu agitaþi. Nota 3: Verificaþi sã nu existe bule în gâtul flaconului sau la suprafaþa reactivului când introduceþi fiolele sau recipientele cu reactiv în analizorul Konelab. Depozitarea şi stabilitatea Reactivii din fiolele nedeschise sunt stabili la 2…8 °C pânã la data de expirare tipãritã pe etichetã. Consultaþi Notele de Aplicare ale dispozitivului dv. de analizã Konelab pentru stabilitatea reactivilor în cazul respectiv. Reactivul PCR HS reconstituit şi tamponul CRP HS deschis pot fi utilizate timp de 8 sãptãmâni dacã sunt depozitate la 2…8 °C când nu sunt folosite. Se recomandã ca fiolele de reactiv sã fie scoase din analizor şi sã fie þinute închise în frigider când nu sunt folosite. Dupã deschidere, calibratorul PCR HS este stabil timp de 8 sãptãmâni la 2…8 °C. Depozitaþi calibratorul bine închis atunci când nu e folosit. Nu folosiþi flacoane care prezintã urme vizibile de creştere microbianã. RECOLTAREA PROBELOR Tipul de probe Se pot folosi ser, plasmã heparinizatã şi plasmã EDTA. Toate probele tulburi trebuie centrifugate (de ex. la 2000 x g timp de 15 min.) Depozitarea (2) Proba poate fi depozitatã timp de 8 zile la 2…8 °C sau timp de 3 luni la – 20 °C (dacã a fost congelatã o singurã datã). PCR, sensibilitate ridicatã PROCEDURA DE ANALIZARE Consultaþi Manualul de Referinþã şi Notele de Aplicare pentru procedura automatã cu dispozitivul dv. de analizã Konelab. Toate aplicaþiile care nu au fost validate de Thermo Fisher Scientific Oy nu garanteazã performanþa şi, prin urmare, trebuie evaluate de cãtre utilizator. Konelab™ / T Series CRP High Sensitivity 981798 1 x 4 ml Materiale furnizate ACEST PROSPECT INTRODUS ÎN PACHET ESTE APLICABIL PENTRU FOLOSIREA ÎN AFARA S.U.A. ORICE REFERINÞÃ LA KONELAB SYSTEMS SE REFERÃ ªI LA SERIA T. UTILIZAREA PREVÃZUTÃ Pentru determinarea cantitativã in vitro a concentraþiilor scãzute de PCR din serul şi plasma umane, cu dispozitivele de analizã Konelab. Toate rezultatele analizei trebuie interpretate în context clinic. SUMAR (1, 5) Proteina C reactivã (PCR) e un compus normal al serului, prezent la persoanele sãnãtoase în concentraþii foarte mici. Infecþiile bacteriene invazive şi alterarea tisularã determinã creşterea nivelului de proteinã C reactivã. PCR e una dintre proteinele cele mai sensibile de fazã acutã; nivelul plasmatic creşte mult dupã infarctul de miocard, traumatisme, infecþii, inflamaþii, operaþii sau proliferare neoplazicã. Creşterea nivelului începe la 6-12 ore şi nivelul poate ajunge pânã la de 2000 de ori valorile normale. Determinarea PCR e utilã clinic pentru screeningul bolilor organice, evaluarea bolilor inflamatorii, determinarea infecþiilor intercurente în lupusul eritematos sistemic (LES), în leucemie sau dupã operaþii, ori în septicemia şi meningita neonatale, când e greu de recoltat probe pentru analizele bacteriologice. Mãsurarea PCR cu ajutorul unor analize PCR sensibile poate mãri valoarea predictivã a altor markeri folosiþi pentru a evalua riscul de boli cardiovasculare şi vasculare periferice. PRINCIPIUL DETERMINÃRII Metoda se bazeazã pe mãsurarea gradului de imunoprecipitare la 540 nm. Microparticulele învelite cu PCR anti-uman CRP se adaugã în probele cu tampon. Creşterea absorbanþei produsã de imunoprecipitare se înregistreazã când reacþia a ajuns la sfârşit. Modificarea absorbanþei e proporþionalã cu cantitatea de antigen (PCR) din soluþie. Reactivii mai sus-menþionaþi şi un flacon gol cu cod de bare, care poate fi folosit în analizorul Konelab. Materiale necesare, dar care nu sunt furnizate Martorii indicaþi mai jos. Calibrarea Folosiþi Calibratorul PCR HS inclus în kit, conform instrucþiunilor furnizate pentru dispozitivul dv. de analizã Konelab. Lãsaþi calibratorul sã stea în analizor maximum o orã. Urmãrire: Valoarea PCR a fost stabilitã folosindu-se preparatul IFCC CRM 470 ca referinþã principalã. Controlul de calitate Folosiþi probe pentru controlul de calitate cel puþin o datã pe zi şi dupã fiecare calibrare şi de fiecare datã când se foloseşte un flacon nou de reactiv. Martori disponibili: Martor PCR cu sensibilitate ridicatã, cod 981852 Intervalele şi limitele Martorului trebuie adaptate la necesitãþile individuale ale laboratorului. Rezultatul probei pentru controlul de calitate trebuie sã se încadreze în limitele stabilite de laborator. CALCULAREA REZULTATELOR Rezultatele sunt calculate automat de dispozitivul de analizã Konelab folosind o curbã de calibrare. Analizorul preparã automat o serie de diluþii din soluþia din calibrator, conform parametrilor prestabiliþi. Curba de calibrare e generatã de calibratorii mãsuraþi prin folosirea unui grafic tip "spline". Template: D01162_4 RO © 2007 Thermo Fisher Scientific Inc. All rights reserved. Precauþii Probele umane trebuie mânuite şi aruncate ca şi cum ar fi potenþial infecþioase. Page 19 D02418_04_insert_CRP HS_MU Data reviziei (aaaa-ll-zz) 2007-09-13 Curbã de calibrare (exemplu) Modificãri faþã de versiunea precedentã Numele companiei a fost actualizat. CRP HS 1.6 Răspuns (A) 1.4 1.2 SK Konelab™ / Séria T CRP High Sensitivity 1 0.8 0.6 CRP stanovenie s vysokou citlivosťou 0.4 0 5 10 15 20 981798 1 x 4 ml Concentraţie (mg/l) Interferenþe Criteriu: Recuperare în interval ± 10% din valorile iniþiale. Bilirubinã: Nu s-au gãsit interferenþe pânã la 1000 μmol/l (58.5 mg/dl). Hemolizã: Nu s-au gãsit interferenþe pânã la 10 g/l de hemoglobinã în hemolizat. Lipemie: Nu s-au gãsit interferenþe pânã la 4 g/l de Intralipid® (marcã înregistratã a firmei Fresenius Kabi AB). Existã o corelaþie scãzutã între turbiditate şi concentraþia trigliceridelor. Pentru alte substanþe care pot interfera, vã rugãm sã consultaþi referinþa 3. Anticorpii heterofili, imunoglobulinele monoclonale (sau pãrþi din acestea) şi factorii reumatoizi pot interfera cu analiza. Rezultatele obþinute la pacienþii suspectaþi cã ar avea asemenea anticorpi trebuie evaluate cu discernãmânt (4). Excesul de antigen Nu s-a gãsit exces de antigen pânã la o concentraþie a PCR de 300 mg/l. VALORI ESTIMATE (5) Sub 3 mg/l Valorile menþionate au doar caracter indicativ. Se recomandã ca fiecare laborator sã verifice acest interval sau sã elaboreze un interval de referinþã pentru populaþia pe care o deserveşte. Când se utilizeazã PCR pentru a evalua riscul unei boli coronariene, mãsurãtorile trebuie comparate cu valorile anterioare. INTERVAL DE MÃSURARE 0.25 – 10 mg/l Interval extins de mãsurare dupã a doua diluþie: 0.25 – 40 mg/l CARACTERISTICILE PERFORMANÞEI (6) Limita de determinare Limita de determinare reprezintã cea mai micã concentraþie care poate fi mãsuratã cantitativ: 0.25 mg/l Imprecizie (unitatea rezultatului mg/l) Medie 0.80 mg/l Medie 2.41 mg/l SD CV% SD CV% În cadrul aceleiaşi analize 0.02 3.2 0.02 0.8 Între zile 0.03 3.7 0.04 1.8 Total 0.04 5.9 0.05 2.3 Medie 7.02 mg/l SD CV% În cadrul aceleiaşi analize 0.04 0.6 Între zile 0.08 1.2 Total 0.15 2.2 Un studiu asupra preciziei a fost efectuat conform liniilor directoare din Documentul NCCLS EP5-A folosindu-se Konelab 30 într-un interval de cel puþin 20 de zile, cu numãrul mãsurãtorilor de cel puþin n = 80. Comparaþia metodelor Un studiu comparat a fost efectuat conform liniilor directoare din Documentul NCCLS EP9A folosindu-se o metodã disponibilã în comerþ ca referinþã. Regresia liniarã (unitatea rezultatului mg/l): y = 1.01 x – 0.29 r = 0.983 n = 87 Concentraþiile probelor au fost între 0.29 şi 10 mg/l. Rezultatele obþinute în laboratoarele individuale pot diferi de datele referitoare la performanþã menþionate. BIBLIOGRAFIE 1) Burtis CA and Ashwood ER (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, WB Saunders Company, Philadelphia, 2001, pp. 332-333 and 688. 2) Guder WG, Narayanan S, Wisser H, Zawta B. List of Analytes; Preanalytical variables. Brochure in: Samples: From Patient to the Laboratory. GIT Verlag GmbH, Darmstadt, 1996. 3) Young DS, Effects of Drugs on Clinical Laboratory Tests, Volume One, Listing by Test, Fifth Edition, AACC Press, Washington, D.C., 3-233 – 3-234. 4) Selby C. Interference in immunoassay. Ann Clin Biochem 1999;36:704-721. 5) Pearson TA et al. Markers of Inflammation and Cardiovascular Disease Application to Clinical and Public Health Practice, A Statement for Healthcare Professionals From the Centers for Disease Control and the American Heart Association., Circulation 2003;107:499-511. 6) Date deþinute de Thermo Fisher Scientific Oy. PRODUCÃTOR Thermo Fisher Scientific Oy Clinical Diagnostics Finland Ratastie 23335, P.O. Box 100, FI-01621 Vantaa Tel. +358 9 329 100, Fax +358 9 3291 0300 www.thermo.com/konelab POUŽITIE Na in vitro kvantitatívne stanovenie CRP v ľudskom sére a plazme v analyzátoroch Konelab. Všetky výsledky testov musia byť interpretované s ohľadom na klinický kontext. ZHRNUTIE (1,5) C-reaktívny proteín (CRP) je normálnou zložkou séra prítomnou u zdravého jedinca vo veľmi nízkej koncentrácii. Invazívna bakteriálna infekcia a rozsiahle poškodenie tkaniva spôsobuje zvýšené hladiny CRP. CRP je jedným z najcitlivejších proteínov akútnej fázy; hladiny v plazme sa zvyčajne dramaticky zvyšujú po infarkte myokardu, úraze, infekcii, zápale, chirurgickom zákroku alebo po neoplastickej proliferácii. Zvyšovanie hladiny začína do 6 až 12 hodín, pričom môže dosiahnuť až 2000-násobok normálu. Stanovenie CRP je klinicky užitočné pri skríningu organických chorôb, zhodnotení aktivity zápalového ochorenia, detekcii interkurentných infekcií pri systémovom lupus erythematosus (SLE), pri leukémii alebo po chirurgickom zákroku a pri manažovaní neonatálnej septikémie a meningitídy, v prípade, kedy môže byť ťažkým odobrať vzorky pre bakteriologické vyšetrenie. Meranie CRP citlivým CRP testom môže zvýšiť výpovednú hodnotu ďalších markerov používaných na zhodnotenie rizika kardiovaskulárnych ochorení a ochorení periférnych ciev. PRINCÍP METÓDY Test je založený na meraní imunoprecipitácie pri 540 nm. Do pufrovaných vzoriek sa pridávajú mikročastice potiahnuté anti-ľudským CRP. Zvýšená absorbancia spôsobená imunoprecipitáciou sa zaznamenáva po dosiahnutí koncového bodu reakcie. Zmena absorbancie je úmerná množstvu antigénu (CRP) v roztoku. NFORMÁCIE O ČINIDLÁCH CRP HS činidlo CRP HS pufer Diluent vzorky CRP HS kalibrátor, Koncentrácie CRP HS činidlo 1 x 4 ml 2 x 50 ml 1 x 60 ml 2 x 1 ml Mikročastice potiahnuté anti-ľudským (ovčím) (lyofilizované) NaN3 1 - < 2.5% CRP HS pufer Tris 0.2 mol/l NaN3 < 0.1 % Diluent vzorky PBS 0.01 mol/l NaN3 < 0.1 % CRP HS kalibrátor Ľudský NaN3 < 0.1 % Koncentrácie CRP kalibrátora sú uvedené na samostatnom Hárku hodnôt. CRP Varovanie Len na in vitro diagnostiku. Dodržujte normálne bezpečnostné opatrenia, ktoré sú nevyhnutné pri manipulácii so všetkými laboratórnymi činidlami. Kalibrátor obsahuje materiál ľudského pôvodu, ktorý bol testovaný a má negatívne HBSAg a HIV-1, HIV-2 a HCV protilátky. S produktom sa však musí zaobchádzať s rovnakou opatrnosťou ako so vzorkami pacienta. Lyofilizované CRP HS činidlo obsahuje azid sodný ako konzervačnú látku (1 - < 2.5%). Nevdychujte prach. Neprehĺtajte/nepožívajte. Zabráňte styku s pokožkou a sliznicami. V prípade kontaktu vymyte veľkým množstvom vody. Azid sodný môže reagovať s olovom a meďou v potrubí, pričom vytvára vysoko výbušné kovové azidy. Pri likvidácii spláchnite veľkým množstvom vody. Výrazy súvisiace s rizikom a bezpečnosťou nájdete v samostatných dokumentoch (T:R;23/24/25 - 52/53 S;22-28-45). Po rekonštitúcii obsahuje kvapalné činidlo CRP HS azid sodný < 0.1%, čo nie je považované za škodlivú koncentráciu. Príprava CRP HS pufer, Diluent vzorky a CRP HS kalibrátor sú pripravené na použitie. Rekonštitúcia CRP HS činidla: Pridajte presne 4 ml deionizovanej vody. Zatvorte nádobku, jemne premiešajte a a nechajte odstáť aspoň 30 min. pri izbovej teplote. Pre použitím obsah ešte raz jemne premiešajte. Nepretrepávajte. Činidlo presuňte do prázdnej nádoby označenej čiarovým kódom, určenej na použitie v analyzátoroch Konelab. Poznámka 1: Vždy používajte CRP HS pufer, CRP HS činidlo a CRP HS kalibrátor s tým istým číslom šarže činidla, pretože jednotlivé zložky sú navzájom zladené v rámci jednej šarže. Poznámka 2: Premiešajte CRP HS činidlo jemne pred vložením do analyzátora Konelab. Nepretrepávajte. Poznámka 3: Pred vložením nádobiek do analyzátora Konelab skontrolujte, či sa v hrdle nádobky, alebo na povrchu činidla nenachádzajú bubliny. Uskladnenie a trvanlivosť Činidlá v neotvorených nádobkách sú stabilné pri 2…8 °C až do dátumu exspirácie vytlačeného na štítku. Informácie o stabilite činidiel nájdete v Používateľskej príručke vášho analyzátora Konelab. Rekonštituované CRP HS činidlo a otvorený CRP HS pufer možno používať 8 týždňov ak sa nepoužíva a je uskladnené uzavreté pri 2…8 °C. Ak sa činidlo nepoužíva, odporúča sa vybrať nádobky s činidlom von z analyzátora a udržiavať ich zatvorené v chladničke. Otvorený CRP HS kalibrátor je stabilný 8 týždňov pri 2…8 °C. Kalibrátor uskladňujte pevne uzavretý ak ho nepoužívate. Nepoužívajte žiadne nádobky s viditeľnou mikrobiálnou kontamináciou. © 2007 Thermo Fisher Scientific Inc. All rights reserved. LIMITELE PROCEDURII (6) TENTO PRÍBALOVÝ LETÁK JE URČENÝ PRE KRAJINY MIMO USA. AKÁKOĽVEK ZMIENKA SYSTÉMOV KONELAB SA VZŤAHUJE AJ NA SÉRIU T. Template: D01162_4 Konelab 20/30/60. Curba de calibrare depinde de lot. Page 20 D02418_04_insert_CRP HS_MU ODBER VZORIEK Typ vzorky Môžete používať sérum, heparinizovanú plazmu, alebo EDTA plazmu. Akákoľvek zakalená vzorka by mala byť centrifugovaná (napr. 2000xg na 15 min) Varovanie Manipulujte a znehodnocujte ľudské vzorky ako potenciálne infekčný materiál. Uskladnenie (2) Vzorka môže byť uskladnená 8 dní pri 2...8 °C alebo 3 mesiace pri -20 °C, po jednorazovom zmrazení. TESTOVACÍ POSTUP Automatický postup použitia analyzátora Konelab nájdete v manuáli a Používateľskej príručke. Meranie postupom, ktorý nie je validizovaný firmou Thermo Fisher Scientific Oy nie je garantované a musí byť vyhodnotené užívateľom. Poskytnuté materiály Činidlá popísané vyššie a prázdna nádobka s čiarovým kódom, na použitie v analyzátoroch Konelab. Materiály potrebné, ale neposkytované Kontroly uvedené nižšie. Kalibrácia CRP HS kalibrátor priložený v súprave používajte v súlade s pokynmi pre analyzátor Konelab. Odporúča sa nechať kalibrátor v analyzátore max. 1 hodinu. Odvoditeľnosť: Hodnota CRP sa určila pomocou IFCC preparátu CRM 470 ako primárnej referencie. Kontrola kvality Používajte vzorky na kontrolu kvality aspoň raz denne, po každej kalibrácii a vždy, keď použijete novú fľašku s činidlom. Dostupné kontroly: CRP Horná kontrola citlivosti (CRP HS), kód 981852 Intervaly a rozmedzia kontrolných hodnôt sa musia prispôsobiť individuálnym požiadavkám laboratória. Výsledné hodnoty kontrolných vzoriek by mali byť v rozmedzí referenčných hodnôt vopred stanovených laboratóriom. VÝPOČET VÝSLEDKOV Výsledky sa automaticky vypočítajú Konelab analyzátorom pomocou kalibračnej krivky. Analyzátor automaticky pripraví sériu riedení zo zásobného kalibrátora podľa vopred stanovených parametrov. Kalibračná krivka sa vytvára zo zmeraných kalibrátorov metódou spline fit. Kalibračná krivka (príklad) Porovnanie metódy Vykonala sa porovnávacia štúdia v súlade so smernicami v NCCLS dokumente EP9-A s použitím komerčne dostupnej metódy ako referencie. Lineárna regresia (výsledok v jednotkách (mg/l): y = 1.01 x – 0.29 r = 0.983 n = 87 Koncentrácie vzoriek sa pohybovali medzi 0.29 a 10 mg/l. Výsledky získané v jednotlivých laboratóriách sa môžu líšiť od daných prevádzkových dát. BIBLIOGRAFIA 1) Burtis CA and Ashwood ER (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, WB Saunders Company, Philadelphia, 2001, pp. 332-333 a 688. 2) Guder WG, Narayanan S, Wisser H, Zawta B. List of Analytes; Preanalytical variables. Brochure in: Samples: From Patient to the Laboratory. GIT Verlag GmbH, Darmstadt, 1996. 3) Young DS, Effects of Drugs on Clinical Laboratory Tests, Volume One, Listing by Test, Fifth Edition, AACC Press, Washington, D.C., 3-233 – 3-234. 4) Selby C. Interference in immunoassay. Ann Clin Biochem 1999;36:704-721. 5) Pearson TA et al. Markers of Inflammation and Cardiovascular Disease Application to Clinical and Public Health Practice, A Statement for Healthcare Professionals From the Centers for Disease Control and the American Heart Association., Circulation 2003;107:499-511. 6) Dáta sú k dispozícii v spoločnosti Thermo Fisher Scientific Oy. VÝROBCA Thermo Fisher Scientific Oy Clinical Diagnostics Finland Ratastie 2, P.O. Box 100, FI-01621 Vantaa Tel. +358 9 329 100, Fax +358 9 3291 0300 www.thermo.com/konelab Dátum revízie (RRRR-MM-DD) 2007-09-13 CRP HS Zmeny od predchádzajúcej verzie Meno spoločnosti aktualizované. 1.6 1.4 1.2 1 SV 0.8 Konelab™ / T Serien CRP High Sensitivity 0.4 0 5 10 15 20 Konentrácia (mg/l) Konelab 20/30/60. Kalibračná krivka závisí od šarže. OBMEDZENIA METODIKY (6) Interferencie Kritérium: Návrat k východziemu stavu ± 10% počiatočných hodnôt. Bilirubín: Bez interferencie do 1000 μmol/l (58.5 mg/dl). Hemolýza: Bez interferencie do 10 g/l hemoglobínu v hemolyzáte. Lipémia: Bez interferencie do 4 g/l Intralipid® (registrovaná značka Fresenius Kabi AB). Korelácia medzi turbiditou a koncentráciou triglyceridov je nízka. Ďalšie interferujúce substancie nájdete v Bibliografii č.3. Heterofílne protilátky a monoklonálne imunoglobulíny (alebo ich časti) ako aj reumatoidné faktory môžu interferovať s testom. Výsledky u pacientov s predpokladanou prítomnosťou týchto protilátok je potrebné starostlivo vyhodnotiť (4). Detekcia prebytku antigénu (AE) Nebol pozorovaný žiadny prebytok antigénu až do koncentrácie CRP 300 mg/l OČAKÁVANÉ HODNOTY (5) Nižšie ako 3 mg/l Tieto hodnoty slúžia len ako smernica. Odporúča sa, aby si každé laboratórium prispôsobilo rozsah referenčných hodnôt na konkrétnu populáciu. Pri použití CRP na zhodnotenie rizika koronárnych ochorení srdca by mali byť namerané hodnoty porovnané s prechádzajúcimi meraniami. MERACÍ ROZSAH 0.25 – 10 mg/l Rozšírený merací rozsah po sekundárnom rozriedení: 0.25 – 40 mg/l PREVÁDZKOVÉ CHARAKTERISTIKY (6) Hranica merateľnosti Hranica merateľnosti reprezentuje najnižšiu merateľnú koncentráciu, ktorá sa dá odlíšiť od nuly: 0.25 mg/l CRP Hög känslighet 981798 1 x 4 ml DENNA BIPACKSEDEL GÄLLER UTOM USA. EVENTUELLA HÄNVISNINGAR TILL KONELAB SYSTEMS AVSER ÄVEN T SERIEN. AVSEDD ANVÄNDNING För kvantitativ bestämning in vitro av låg komcentraion av CRP i humant serum och plasma på Konelab-analysatorer. Alla testresultat måste tolkas med hänsyn till det kliniska sammanhanget. SAMMANFATTNING (1, 5) C-reaktivt protein (CRP) är en normal beståndsdel i serum, som hos friska individer finns i mycket låg koncentration. Invasiv bakteriell infektion och omfattande vävnadsskada orsakar förhöjd CRP-nivå. CRP är en av de mest känsliga akutfasproteinerna. Nivån i plasma stiger normalt dramatiskt efter hjärtinfarkt, trauma, infektion, inflammation, kirurgi eller neoplastisk proliferation. Ökningen börjar normalt inom 6 till 12 timmar och nivån kan komma upp i 2000 gånger normalt. Bestämning av CRP är kliniskt användbart för screening för organisk sjukdom, bedömning av aktivitet hos inflammatorisk sjukdom, detektering av interkurrenta infektioner i systemisk lupus erythematosus (SLE), vid leukemi eller efter kirurgi och behandling av septikemi och meningit hos nyfödda, när provtagning för bakteriologiska undersökningar kan vara svårt. Mätning av CRP med känsliga CRP-tester kan ge förbättrad predikativ information hos andra markörer som används för att bedöma risken för hjärt-kärl-sjukdomar och sjukdomar i perifera kärl. TESTPRINCIP Metoden baseras på mätning av immunoprecipitation vid 540 nm. Mikropartiklar belagda med anti-humant CRP tillsätts i buffrade prover. Ökningen i absorbans orsakad av immunprecipitation registreras när reaktionen har nått sin slutpunkt. Förändringen i absorbans står i proportion till mängden antigen (CRP) i lösningen. REAGENSINNEHÅLL CRP HS-reagens CRP HS-buffert Provspädningsmedel CRP HS-kalibrator 1 x 4 ml 2 x 50 ml 1 x 60 ml 2 x 1 ml © 2007 Thermo Fisher Scientific Inc. All rights reserved. 0.6 Template: D01162_4 Odpoveď (A) Nepresnosť (výsledok v jednotkách (mg/l) Priemer 0.80 mg/l Priemer 2.41 mg/l SD CV% SD CV% Počas procesu 0.02 3.2 0.02 0.8 Medzi jednotlivými 0.03 3.7 0.04 1.8 dňami Celkovo 0.04 5.9 0.05 2.3 Priemer 7.02 mg/l SD CV% Počas procesu 0.04 0.6 Medzi jednotlivými 0.08 1.2 dňami Celkovo 0.15 2.2 Vykonala sa štúdia presnosti v súlade so smernicami NCCLS dokumentu EP5-A, s prístrojom Konelab 30 počas minimálne 20 dní, s počtom meraní n = 80. Page 21 D02418_04_insert_CRP HS_MU Beredning CRP HS-buffert, provspädningsmedel och CRP HS-kalibrator är klara att använda. Rekonstitution av CRP HS-reagens: Tillsätt exakt 4 ml avjoniserat vatten. Stäng flaskan, blanda försiktigt och låt stå i 30 minuter i rumstemperatur. Blanda innehållet på nytt före användning. Får ej skakas. Överför reagenset till en tom flaska med streckkodsetikett för användning i Konelab-analysatorer. Anm. 1: Använd alltid CRP HS-buffert, CRP HS-reagens och CRP HS-kalibrator med samma reagens-batch-nummer, eftersom komponenterna inom ett batch är matchade. Anm. 2: Blanda CRP HS-reagens försiktigt innan reagenset placeras i Konelabanalysatorn. Får ej skakas. Anm. 3: Kontrollera att det ej finns bubblor i flaskhalsen eller på ytan av reagenset då behållarna placeras i Konelab-analysatorn. Förvaring och hållbarhet Reagens i oöppnad behållare är hållbar vid 2…8 °C till på etiketten angivet utgångsdatum. Se Application Notes för aktuell Konelab-analysator för information om reagensernas hållbarhet i instrumentet. CRP HS-reagens som rekonstituerats och CRP HS-buffert som öppnats kan användas i 8 veckor om de förvaras vid 2…8 °C när de inte används. Vi rekommenderar att reagensflaskorna tas ut ur analysatorn och hålls tillslutna i kylskåp när de inte används. När CRP HS-kalibratorn öppnats är den hållbar i 8 veckor vid 2…8 °C. Kalibratorn ska förvaras ordentligt tillsluten när den inte används. Flaskor med innehåll som uppvisar tillväxt av mikroorganismer får inte användas. PROVTAGNING Provmaterial Serum, heparinplasma och EDTA-plasma kan användas. Eventuellt grumliga prover ska centrifugeras (t.ex. 2000xg i 15 min.) Försiktighetsåtgärder Humana prover ska behandlas som potentiellt smittförande, både vid hantering och kassering. Förvaring (2) Provet kan förvaras i 8 dagar vid 2...8 °C eller i 3 månader vid – 20 °C (om de frysts endast en gång). TESTUTFÖRANDE Se Handhavandemanualen och Application Notes för automatiskt utförande på aktuell Konelab-analysator. Varje applikation som ej har validerats av Thermo Fisher Scientific Oy, kan ej garanteras vad gäller prestanda och måste därför utvärderas av användaren. Bifogat material Reagenser enligt ovan och en tom flaska med streckkodsetikett för användning i Konelabanalysatorer. Erforderligt material som ej medföljer Kontroll enligt nedan. Kalibrering Använd den CRP HS-kalibrator som ingår i satsen, enligt instruktionerna för Konelabanalysatorn. Kalibratorn för stanna kvar i analysatorn i högst en timme. Spårbarhet: Värdet för CRP har fastställts med hjälp av IFCC-preparat CRM 470 som primär referens. Kvalitetskontroll Använd kvalitetskontrollprov minst en gång om dagen, efter varje kalibrering och varje gång en ny reagensflaska används. Tillgängliga kontroller: CRP hög känslighet-kontroll, artikelnr. 981852 Kontrollens intervall och gränser måste anpassas till laboratoriets egna krav. Resultatet från kvalitetskontrollprovet bör ligga inom de av laboratoriet fastställda gränserna. RESULTATBERÄKNING Resultaten beräknas automatiskt av Konelab-analysatorn med hjälp av en kalibreringskurva. Analysatorn framställer automatiskt en spädningsserie från stamkalibratorn enligt de angivna parametrarna. Kalibreringskurvan genereras från de uppmätta kalibratorerna med hjälp av kurvanpassning med en spline-funktion. CRP HS 1,6 1,4 1,2 1 0,8 0,6 0,4 0 5 10 15 20 Koncentration (m g/l) Konelab 20/30/60. Kalibreringskurvan är batchberoende. BEGRÄNSNINGAR I UTFÖRANDET (6) Interferens Kriterium: Utbyte inom ± 10 % av initialvärdena. Bilirubin: Ingen interferens har konstaterats upp till 1000 μmol/l (58.5 mg/dl). Hemolys: Ingen interferens har konstaterats upp till 10 g/l hemoglobin i hemolysat. Lipemi: Ingen interferens har konstaterats upp till 4 g/l Intralipid® (varumärket tillhör Fresenius Kabi AB). Det är dålig korrelation mellan grumlighet och triglyceridkoncentration. För andra interfererande ämnen, se referens 3. Heterofila antikroppar, monoklonala immunglobuliner (eller delar av dem) och reumafaktorer kan interferera med testet. Resultat från patienter hos vilka sådana antikroppar misstänks måste noggrant utvärderas (4). Antigen excess Ingen antigen excess konstaterades upp till en CRP-koncentration på 300 mg/l. REFERENSOMRÅDE (5) Under 3 mg/l Angivna värden är endast avsedda som vägledning. Vi rekommenderar att varje laboratorium verifierar detta område eller fastställer ett referensintervall för populationen som betjänas. När CRP används för att bedöma risken för hjärtinfarkt ska mätningar jämföras med tidigare värden. MÄTOMRÅDE 0.25 – 10 mg/l Utökat mätområde efter sekundär spädning: 0.25 – 40 mg/l UTFÖRANDETS KARAKTERISTIKA (6) Detektionsgräns Detektionsgränsen är den lägsta koncentration som kan mätas kvantitativt, i detta fall: 0.25 mg/l Imprecision (resultatenhet mg/l) Medel 0.80 mg/l SD CV% Inom serie 0.02 3.2 Mellandags 0.03 3.7 Totalt 0.04 5.9 Medel 7.02 mg/l SD CV% Inom serie 0.04 0.6 Mellandags 0.08 1.2 Totalt 0.15 2.2 Medel 2.41 mg/l SD CV% 0.02 0.8 0.04 1.8 0.05 2.3 En precisionsstudie har utförts enligt anvisningarna i NCCLS Dokument EP5-A med Konelab 30 under minst 20 dagar. Antalet mätningar var minst n = 80. Metodjämförelse En jämförelsestudie har utförts enligt anvisningarna i NCCLS Dokument EP9-A med en kommersiellt tillgänglig metod som referens. Linjär regression (resultatenhet mg/l): y = 1.01 x – 0.29 r = 0.983 n = 87 Provens koncentration låg mellan 0.29 och 10 mg/l. Resultaten som erhålls vid varje enskilt laboratorium kan skilja sig från angivna data för prestanda. REFERENSER 1) Burtis CA and Ashwood ER (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, WB Saunders Company, Philadelphia, 2001, pp. 332-333 and 688. 2) Guder WG, Narayanan S, Wisser H, Zawta B. List of Analytes; Preanalytical variables. Brochure in: Samples: From Patient to the Laboratory. GIT Verlag GmbH, Darmstadt, 1996. 3) Young DS, Effects of Drugs on Clinical Laboratory Tests, Volume One, Listing by Test, Fifth Edition, AACC Press, Washington, D.C., 3-233 – 3-234. 4) Selby C. Interference in immunoassay. Ann Clin Biochem 1999;36:704-721. 5) Pearson TA et al. Markers of Inflammation and Cardiovascular Disease Application to Clinical and Public Health Practice, A Statement for Healthcare Professionals From the Centers for Disease Control and the American Heart Association., Circulation 2003;107:499-511. 6) Data finns på fil hos Thermo Fisher Scientific Oy. © 2007 Thermo Fisher Scientific Inc. All rights reserved. Försiktighetsåtgärder Endast för in vitro-diagnostik. Vidta normala försiktighetsåtgärder som vid all hantering av laboratorireagenser. I kalibratorn finns humanbaserat material, som har testats och konstaterats vara negativt för HbsAg och för antikroppar mot HIV 1/2- och HCV. Produkten ska dock hanteras lika försiktigt som patientprover. Frystorkat CRP HS-reagens innehåller natriumazid som konserveringsmedel (1 - < 2.5%). Damm får ej inandas. Får ej sväljas. Undvik kontakt med hud och slemhinnor. Om kontakt skulle inträffa, skölj med mycket rikligt med vatten. Natriumazid kan reagera med bly och koppar i avloppssystem så att högexplosiva metallazider bildas. Vid nedspolning, spola med mycket stor mängd vatten. Se separat blad med risk- och säkerhetstexter (T:R;23/24/25 - 52/53 S;22-28-45). Efter rekonstitution innehåller CRP HS-reagens i vätskeform < 0.1% natriumazid, vilket inte anses vara en farlig koncentration. Kalibreringskurva (exempel) Template: D01162_4 Mikropartiklar belagda med anti-humant CRP (får) (frystorkat) NaN3 1 - < 2.5% CRP HS-buffert Tris 0.2 mol/l NaN3 < 0.1 % Provspädningsmedel PBS 0.01 mol/l NaN3 < 0.1 % CRP HS-kalibrator Human NaN3 < 0.1 % Kalibratorns CRP-koncentration finns tryckt på ett separat referensblad. Svar (A) Koncentrationer CRP HS-reagens Page 22 D02418_04_insert_CRP HS_MU TILLVERKARE Thermo Fisher Scientific Oy Clinical Diagnostics Finland Ratastie 2, P.O. Box 100, FI-01621 Vantaa Tel. +358 9 329 100, Fax +358 9 3291 0300 www.thermo.com/konelab Revisionsdatum (åååå-mm-dd) 2007-09-13 Template: D01162_4 © 2007 Thermo Fisher Scientific Inc. All rights reserved. Ändringar från tidigare utgåva Företagsnamnet är uppdaterat.
© Copyright 2024