Document 197027

How to Use FUJI DRI-CHEM 4000 Series
NOTE: For details, read the Instruction Manual carefully.
* Depending on the unit equipped, applicable tests and
functions may differ. For details, refer to the Instruction Manual.
1
When using a new slide box,
the QC card must be read.
Load slides in a slide cartridge.
* QC cards are packed with slides
in the box.
* For ISE slides, QC cards are not
packed together.
CM slides: test name facing up
(no specific orientation).
ISE slides: the arrow indications
facing up and pointing to the
right.
Example of setting slides
CRP
(diluent)
CM
ISE
2
Sample
Place tips on the sample rack.
ISE
CRP
diluent
Place FUJI DRI-CHEM AUTO TIPS on the
sample rack.
3
CM
Set a sample tube in the sample rack.
When the amount of the sample is insufficient,
divide it and use FUJI PLAIN TUBEs.
d
c
MIX CUP
RE
b
4
For ISE tests:
For ISE tests, pour reference fluid
into the FUJI PLAIN TUBE 0.5
(0.5 mL).
5
Required reference
fluid volume
1.5 mm
a
0.5
400µL (MAX)
300µL (MIN)
0.3
At least
100µL
For CRP tests and tests requiring dilution:
For CRP tests and tests requiring dilution, use diluent.
Required diluent
volume
0.1
d
c
MIX CUP
RE
b
a
6
Close the sampler cover and press the START key.
ISE unit (P5-4)
C
R
Light source lamp (P5-13)
When a lamp replacement
error occurs, or the lamp’s
cumulative illumination time
exceeds 1000 hours
Incubator (P5-4)
C
Once every three months
or when inaccurate test
results occur
I :Inspection
R :Replacement
I
C
Air filter (P5-3)
Sampler O-ring (P5-16)
Once a month
R
Once a year
Spotting part (P5-9)
C
Slide reader (P5-9)
When reading errors occur frequently
C
When slide transfer error or
contamination with sample
etc. occurs
* ( ): Section to be referred to in the Instruction Manual
C :Cleaning
Cleaning/Inspection/Replacement
25
DBIL
13
UA
14
TCHO
18
21
TBIL
TP
15
NH3
16
28
Mg
TG
24
IP
human
only
Immunological
Electrolytes
Enzymes
35
test
code
CRP
Cl
51
93
92
91
Na
K
34
37
LDH
LAP
32
31
GOT/
AST
GPT/
ALT
30
33
38
39
GGT
CPK
CKMB
CHE
AMYL 43
ALP
test
name
human
only
NOTE: Test codes are subject to change without notice.
Test codes are printed out on the slide’s outer package.
General
Chemistry
17
CRE
26
23
Ca
HDLC
11
BUN
10
20
ALB
GLU
test
code
test
name
Test Code Table